Philips Optimus 50, Optimus 65, Optimus 80 User manual

0 (0)
SRM 06/12
kV dependent max mA values for exposure times 100ms in combination with Optimus 50kW
kV SF [µA] LF [µA] kV SF [µA] LF [µA] kV SF [µA] LF [µA] kV SF [µA] LF [µA]
40 250000 650000 70 485110 650000
100 350000 500000
130 269230 384615
41 257837 650000 71 492957 650000 101 346534 495049 131 267175 381679
42 265674 650000 72 486111 650000 102 343137 490196 132 265151 378787
43 273511 650000 73 479452 650000 103 339805 485436 133 263157 375939
44 281348 650000 74 472972 650000 104 336538 480769 134 261194 373134
45 289185 650000 75 466666 650000 105 333333 476190 135 259259 370370
46 297022 650000 76 460526 650000 106 330188 471698 136 257352 367647
47 304859 650000 77 454545 649350 107 327102 467289 137 255474 364963
48 312696 650000 78 448717 641025 108 324074 462962 138 253623 362318
49 320533 650000 79 443037 632911 109 321100 458715 139 251798 359712
50 328370 650000 80 437500 625000 110 318181 454545 140 250000 357142
51 336207 650000 81 432098 617283 111 315315 450450 141 248226 354609
52 344044 650000 82 426829 609756 112 312500 446428 142 246478 352112
53 351881 650000 83 421686 602409 113 309734 442477 143 244755 349650
54 359718 650000 84 416666 595238 114 307017 438596 144 243055 347222
55 367555 650000 85 411764 588235 115 304347 434782 145 241379 344827
56 375392 650000 86 406976 581395 116 301724 431034 146 239726 342465
57 383229 650000 87 402298 574712 117 299145 427350 147 238095 340136
58 391066 650000 88 397727 568181 118 296610 423728 148 236486 337837
59 398903 650000 89 393258 561797 119 294117 420168 149 234899 335570
60 406740 650000 90 388888 555555 120 291666 416666 150 233333 333333
These are the default values of the tube data file TUBE_R2.TDL or TUBE_R3.TDL.
Typically there are no changes after adaptation except there are reductions due to the max generator power (Optimus 50kW = 650mA max; Optimus 65kW = 900mA
max) or if the filament does not have its full emission anymore (e.g. from shorted filament loops).
The red marked emission current values (at 71kV SF and 40kV LF) are the max emission currents at a kV value from which on the filament current can be reduced
to achieve its max load performance.
srm0612-max-mA-OM50.doc August – 12 - 2003