Philips V60 Service Manual

0 (0)

Respironics V60 Ventilator

Service Manual

For Technical Support and Customer Service, contact:

USA and Canada: 1-800-345-6443 (toll free) or 724-387-4000

Respironics Europe, Africa, Middle East: +33-1-47-52-30-00

Respironics Asia Pacific: +852-3194-2280

Facsimile: 724-387-5012

United States of America

Respironics California, Inc.

2271 Cosmos Court

Carlsbad, CA 92011

USA

Email and Web Addresses service@respironics.com clinical@respironics.com www.respironics.com

Authorized European Representative

Respironics Deutschland GmbH Gewerbestrasse 17

D-82211 Herrsching Germany +49-8-15-29-30-60

Copyright© 2009 Respironics California, Inc. and its affiliates.

All rights reserved.

This work is protected under Title 17 of the United States copyright code and is the sole property of Respironics California, Inc. No part of this document may be copied or otherwise reproduced, or stored in any electronic information retrieval system, except as specifically permitted under United States copyright law, without the prior written consent of Respironics California, Inc.

Table of Contents

1 Introduction and Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1

1.1 Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 1.2 Recommended Test Equipment, Tools, and Supplies . . . . . . . . . . . . . . . . . . 1-3 1.3 Where to Go for Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5

2 Warnings and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1

2.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1 2.2 Preparing for Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4 2.3 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6 2.4 Alarms and Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6 2.5 Care and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6 2.6 First-Time Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7 2.7 Communications Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7 2.8 Diagnostic Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7

3 Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1

3.1 Pneumatics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1

3.1.1 Air Inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2

3.1.2 Air Inlet Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2

3.1.3 Air Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3

3.1.4 Machine and Proximal Pressure Transducers . . . . . . . . . . . . . . . . . . . . . 3-3

3.1.5 Barometric Pressure Transducer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3

3.1.6 Oxygen Pressure Transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3

3.1.7 Manifold, Oxygen Inlet Filter, Filter Element . . . . . . . . . . . . . . . . . . . . . 3-4

3.1.8 Oxygen Solenoid Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4

3.1.9 Oxygen Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4

3.1.10 Blower. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4

3.1.11 Solenoid Valves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5

3.2 Electronics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6

3.2.1 Power management (PM) PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7

3.2.2 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7

3.2.3 Internal Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7

3.2.4 CPU PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8

3.2.5 Motor Controller (MC) PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10

3.2.6 Data Acquisition (DA) PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10

3.2.7 Flow Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10

1049766 Rev A

V60 Ventilator Service Manual

iii

Table of Contents

3.2.8 User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11

3.2.9 LCD Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11

3.2.10 Backlight Inverter PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11

3.2.11 Touch Screen Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12

3.2.12 Nav-Ring Assembly (Rotary Adjustment) . . . . . . . . . . . . . . . . . . . . . . 3-12

3.2.13 Power Switch Overlay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12

3.2.14 Switch PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12

3.2.15 UI PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13

3.3 Electronic Signal Paths . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14

4 Periodic Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1

4.1 Annual Preventive Maintenance Instructions . . . . . . . . . . . . . . . . . . . . . . . . 4-2

5 Diagnostic Mode and Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . 5-1

5.1 System Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2

5.1.1 Selecting a Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3

5.1.2 Setting Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5

5.1.3 Restoring Default Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6

5.1.4 Selecting Pressure Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7

5.1.5 Enabling Software Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8

5.1.6 Baud Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9

5.2 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10

5.2.1 Viewing Ventilator Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11

5.2.2 Pneumatic Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12

5.2.3 Output Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13

5.2.4 Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14

5.3 Touch Screen Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18

5.4 Diagnostic Codes, Alarms, and Troubleshooting . . . . . . . . . . . . . . . . . . . . . 5-19

5.5 Miscellaneous Troubleshooting Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-31

6 Reports and Software Downloads . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-1

6.1 Setting Up the Service PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1 6.1.1 Setting Up the Serial Interface Using HyperTerminal . . . . . . . . . . . . . . . 6-2 6.1.2 Setting Up the Serial Interface Using Tera Term . . . . . . . . . . . . . . . . . . 6-6 6.2 Generating a Diagnostic Report (DRPTA) . . . . . . . . . . . . . . . . . . . . . . . . . 6-17 6.3 Clearing the Significant Event Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-20 6.4 Downloading Ventilator Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-21 6.5 Programming the Ventilator Serial Number and Power-On Hours . . . . . . . . . 6-21 6.5.1 Programming the Ventilator Serial Number . . . . . . . . . . . . . . . . . . . . . 6-21 6.5.2 Programming Ventilator Power-On Hours. . . . . . . . . . . . . . . . . . . . . . . 6-23

iv

V60 Ventilator Service Manual

1049766 Rev A

Table of Contents

7 Component Removal/Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1

7.1 Disconnecting Power. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2 7.2 Air Inlet Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3 7.3 Internal Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3 7.4 Top Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5 7.5 AC Inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6 7.6 Fan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-7 7.7 Oxygen Inlet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8 7.8 Motor Controller (MC) PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9 7.9 Separating the UI from the Base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10 7.10 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11 7.11 Gas Outlet Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12 7.12 UI Retainer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-13 7.13 Proximal Pressure Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14 7.14 Speakers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-15 7.15 Power Management (PM) PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-16 7.16 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17 7.17 CPU PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-18 7.18 Real-Time Clock Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-19 7.19 Left Side Wall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-20 7.20 Gas Delivery Subsystem (GDS). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-21 7.21 Right Side Wall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-23 7.22 Oxygen Inlet Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-24 7.23 Data Acquisition (DA) PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-25 7.24 Air and Oxygen Flow Sensor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . 7-27 7.25 Oxygen Solenoid Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-28 7.26 Solenoid Valves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-29 7.27 Blower. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-30 7.28 Opening the User Interface (UI)/ Rear Bezel . . . . . . . . . . . . . . . . . . . . . . 7-31 7.29 Power Switch Overlay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-32 7.30 Switch PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-33 7.31 Nav-Ring Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-34 7.32 Front Bezel, Touch Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-35 7.33 LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-36 7.34 User Interface (UI) PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-38 7.35 Backlight Inverter PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-39 7.36 LCD Tray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-40 7.37 Bottom Feet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-41 7.38 Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-42

1049766 Rev A

V60 Ventilator Service Manual

v

Table of Contents

8 Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1

8.1 Required Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5 8.2 Preliminary Cleaning, Inspection and Setup . . . . . . . . . . . . . . . . . . . . . . . . 8-7 8.3 View and Record Ventilator Information . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-8 8.4 Pneumatic Calibration Analyzer Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9 8.4.1 Measurement Selection Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9 8.4.2 Averaging Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10 8.4.3 Trigger Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-11 8.4.4 Configurations Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12 8.5 Performance Verification Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-15 8.5.1 Electrical Safety (Test 1). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-15 8.5.2 Leak Tests (Test 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-17 8.5.3 Controls (Test 3). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-20 8.5.4 Pressure Accuracy (Test 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-21 8.5.5 Air Delivery/Flow Accuracy (Test 5) . . . . . . . . . . . . . . . . . . . . . . . . . . 8-24 8.5.6 Oxygen Flow Accuracy (Test 6). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-27 8.5.7 Oxygen Accuracy (Test 7) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-29 8.5.8 S/T Performance (Test 8) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-31 8.5.9 Alarms (Test 9) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-33 8.5.10 Power Fail (Test 10) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-35 8.5.11 Internal Battery (Test 11) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-36 8.6 Returning Ventilator to Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-37 8.7 Performance Verification Troubleshooting/Repair . . . . . . . . . . . . . . . . . . . . 8-37 8.7.1 Test 1: Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-38 8.7.2 Test 2: Leak Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-38 8.7.3 Test 3: Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-39 8.7.4 Test 4: Pressure Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-40 8.7.5 Test 5: Air Delivery/Flow Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-40 8.7.6 Test 6: Oxygen Flow Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-41 8.7.7 Test 7: Oxygen Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-41 8.7.8 Test 8: S/T Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-41 8.7.9 Test 9: Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-42 8.7.10 Test 10: Power Fail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-42 8.7.11 Test 11: Internal Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-43 8.8 Electrical Safety Data Form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-44 8.9 Performance Verification Data Form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-45

vi

V60 Ventilator Service Manual

1049766 Rev A

Table of Contents

9 Replacement Parts List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1

9.1 Complete Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1

9.2 Recommended Inventory Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7

9.3 Ventilator Chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-10

9.4 Pneumatics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-12

9.5 Electronics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13

10 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1

10.1 Control settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1 10.2 Patient data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3 10.3 Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4 10.4 Menu window settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4 10.5 Operator-accessible diagnostic mode functions . . . . . . . . . . . . . . . . . . . . 10-5 10.6 Physical characteristics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5 10.7 Environmental specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-6 10.8 Pneumatic specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-6 10.9 Electrical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-7 10.10 Other specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8 10.11 Contact Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8

Appendix A Respi-Link . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1

A.1 Downloading Ventilator Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1

A.2 Installing Ventilator Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-6

A.3 Reinitiate a Software Package. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-8

A.4 Remote Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-13

Appendix B Field Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1

Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Index-1

1049766 Rev A

V60 Ventilator Service Manual

vii

Table of Contents

(This page is intentionally blank.)

viii

V60 Ventilator Service Manual

1049766 Rev A

Chapter 1. Introduction and Intended Use

The V60 ventilator is a microprocessor-controlled, positive pressure ventilator assist system. The ventilator provides noninvasive and invasive ventilatory support for spontaneously breathing adult and pediatric patients.

The ventilator has a variety of modes and monitoring capabilities to assist in assessing performance and patient-to-ventilator synchrony. The safety features include in-depth alarms and a variety of integrated safety and self-diagnostic features. Many system functions are automatically checked at startup and during operation.

The ventilator includes a touch screen user interface (UI) and navigation ring (nav-ring) that allows the operator to select ventilator and alarm settings and displays of ventilator and patient data.

The ventilator is designed to be upgradeable, and features communications capabilities and an internal battery backup option.

Read this manual thoroughly before performing service or maintenance on the V60 ventilator. This manual includes advanced troubleshooting, calibration, and maintenance instructions for the ventilator. All maintenance and repair work should be performed by qualified biomedical technicians who have appropriate training and authorization to provide maintenance, repair, and service for the V60.

Review the operating instructions for the V60 ventilator before running tests, checking operational readiness, or initiating patient use. These instructions include important information about ventilator safety and operation.

For additional information about accessories or related equipment, such as humidifiers and remote alarm systems, refer to the appropriate instruction manual prior to operating the V60 ventilator. Review the applicable warnings and cautions in the V60 User Manual before operating the ventilator.

1049766 Rev A

V60 Ventilator Service Manual

1-1

Chapter 1

Introduction and Intended Use

1.1 Intended Use The Respironics V60 ventilator is an assist ventilator that is intended to augment patient breathing. It is intended for spontaneously breathing individuals who require mechanical ventilation: patients with respiratory failure, chronic respiratory insufficiency, or obstructive sleep apnea in a hospital or other institutional settings under the direction of a physician.

The ventilator is intended to support pediatric patients weighing 20 kg (44 lb.) or greater to adult patients. It is also intended for intubated patients meeting the same selection criteria as the noninvasive applications. The ventilator is intended to be used by qualified medical professionals such as physicians, nurses, and respiratory therapists. The ventilator is intended to be used only with various combinations of Respironics recommended patient circuits, interfaces (masks), humidifiers, and other accessories.

1-2

V60 Ventilator Service Manual

1049766 Rev A

1.2 Recommended

Test Equipment,

Tools, and Supplies

Chapter 1

Introduction and Intended Use

Table 1-1 lists the recommended tools, test equipment, and materials required to service and maintain the V60 ventilator.

Table 1-1: Recommended Test Equipment, Tools, and Materials

Description

Manufacturer and Model

 

 

Test Equipment

 

 

 

Digital multimeter (DMM) accurate to three decimal

Local Supplier

places

 

 

 

Electrical safety analyzer

Dale LT 5440 or equivalent

 

 

Pneumatic calibration analyzer capable of measuring

Respironics P/N 1040311 or equivalent

low pressure (cmH2O), flow rate (LPM), and volume

 

(liters)

 

 

 

Temperature/humidity monitor

Fisher Scientific 11661-14 or

 

equivalent

 

 

Test lung

IngMar QuickLung or equivalent

 

 

Service Tools and Supplies

 

 

 

V60 Service Kit, which includes:

Respironics P/N 1054291

 

 

Adapter, 22-mm OD, both ends

Respironics P/N 1002505

 

 

Adapter, 25-pin to 9-pin

Respironics P/N 1058403

 

 

Adapter, torque, cap/collar

Respironics P/N 1056005

 

 

BiPAP test adapter, 0.25-in.

Respironics P/N 332353

 

 

Cable, null modem

Respironics P/N 1022815

 

 

Cable, TTL communications

Respironics P/N 1058778

 

 

Circuit tube, 18-in. smooth-bore (qty. 2)

Respironics P/N 1000060

 

 

Coupling, straight, silicone

Respironics P/N 500-1000-43

 

 

Forceps, locking, red plastic

Respironics P/N 1058430

 

 

Plug, low-pressure

Respironics P/N 1058270

 

 

Plug, tapered 23/32 - 61/64 in., silicone (qty. 2)

Respironics P/N 1055322

 

 

Plug, tapered 9/16 - 3/4 in., silicone

Respironics P/N 1055323

 

 

Pressure pick-off port (oxygen enrichment

Respironics P/N 312710

attachment SNGL)

 

 

 

Proximal pressure line tubing

Respironics P/N 312114

 

 

Remote alarm test cable

Respironics P/N 1027818

 

 

Remote alarm test cable adapter

Respironics P/N 1027817

 

 

1049766 Rev A

V60 Ventilator Service Manual

1-3

Chapter 1

Introduction and Intended Use

Table 1-1: Recommended Test Equipment, Tools, and Materials

Description

Manufacturer and Model

 

 

Syringe, system leak test

Respironics P/N 1058271

 

 

Valve, ball

Respironics P/N 1058431

 

 

Valve, oxygen/regulator shut-off

Respironics P/N 1058380

 

 

Whisper Swivel II

Respironics P/N 332113

 

 

Adapter, USB to serial

Respironics P/N 1022895 or equivalent

 

 

Cleaning cloth

Local supplier

 

 

Fitting, system leak test syringe (replacement)

Respironics P/N 1060263

 

 

Isopropyl alcohol

Local supplier

 

 

Lubricant, KRYTOX GPL 226

Respironics P/N 1021021 or equivalent

 

 

Metric hex key set (rounded ends), 1.5 to 4 mm

Local supplier

 

 

Mild detergent or antiseptic wipes

Local supplier

 

 

Needle nose pliers

Local supplier

 

 

PC or laptop (required for downloading software and

Required: Windows XP operating

capturing diagnostic codes)

systems, serial port, and USB port

 

 

Pliers

Local supplier

 

 

Screwdriver, #0 Phillips

Local supplier

 

 

Screwdriver, #1 Phillips

Local supplier

 

 

Screwdriver, #2 Phillips

Local supplier

 

 

Screwdriver, #3 Phillips

Local supplier

 

 

Screwdriver, pen size, Phillips

Local supplier

 

 

Screwdriver, pen size, slotted

Local supplier

 

 

Socket, deep, 5/16-in.

Local supplier

 

 

Socket, deep, 9/16-in.

Local supplier

 

 

Torque driver capable of 11.2 to 283 N cm /

Local supplier

1 to 25 in.-lbf

 

 

 

Torque driver capable of 226 to 1130 N cm/

Local supplier

20 to 100 in.-lbf

 

 

 

Vacuum, ESD-safe

3M Model 497-AJM or equivalent

 

 

Workstation, antistatic

3M Model 725 or equivalent

 

 

1-4

V60 Ventilator Service Manual

1049766 Rev A

Chapter 1

Introduction and Intended Use

1.3 Where to Go for For Technical Support and Customer Service, contact:

Help

USA and Canada: 800-345-6443 or 724-387-4000

Respironics Europe, Africa, Middle East: +33-1-47-52-30-00

Respironics Asia Pacific: +852-3194-2280

Facsimile: +1-724-387-5012

1049766 Rev A

V60 Ventilator Service Manual

1-5

Chapter 1

Introduction and Intended Use

(This page is intentionally blank.)

1-6

V60 Ventilator Service Manual

1049766 Rev A

Chapter 2. Warnings and Cautions

Before servicing the Respironics V60 ventilator, read and understand this service manual, especially safety considerations. These safety considerations are for reference only, and are not intended to supersede your institution’s protocol for service or safe use of noninvasive ventilation.

The instructions in this manual are primarily reserved for use by a qualified service technician.

 

WARNING:

Alerts the user to the possibility of injury, death, or other serious adverse

 

 

 

reactions associated with the use or misuse of the device.

 

 

 

 

 

 

 

 

 

CAUTION: Alerts the user to the possibility of a problem with the device associated with its use or misuse, such as device malfunction, device failure, damage to the device, or damage to other property.

NOTE:

Emphasizes information of particular importance.

 

 

2.1 General

WARNING:

An alternative means of ventilation shall be available whenever the ventilator

 

 

is in use. If a fault is detected in the ventilator, disconnect the patient from it

 

 

and immediately start ventilation with such a device. The ventilator must be

 

 

removed from clinical use and serviced by Respironics-authorized service

 

 

personnel.

 

WARNING:

Use the Respironics V60 ventilator on spontaneously breathing patients only.

 

 

It is an assist ventilator and is intended to augment the ventilation of a

 

 

spontaneously breathing patient. It is not intended to provide the total

 

 

ventilatory requirements of the patient.

 

WARNING:

We do not recommend you use the Respironics V60 ventilator on patients who

 

 

require ventilation at predetermined tidal volumes. The ventilator provides

 

 

continuous positive airway pressure (CPAP) and positive pressure ventilation

 

 

(S/T, PCV, and AVAPS) and is indicated for assisted ventilation only. These

 

 

modes do not provide ventilation with guaranteed tidal volume delivery.

 

WARNING:

We do not recommend you use AVAPS on patients who require rapid and

 

 

frequent IPAP adjustments to maintain a consistent tidal volume. AVAPS, a

 

 

volume targeted mode, changes the IPAP setting in order to achieve the target

 

 

tidal volume. During AVAPS setup, there may be a period of time before the

 

 

target tidal volume is achieved. AVAPS is ideal for more stabilized patients.

1049766 Rev A

V60 Ventilator Service Manual

2-1

Chapter 2

Warnings and Cautions

WARNING:

To reduce the risk of CO2 rebreathing, make sure EPAP pressures and

 

exhalation times are sufficient to clear all exhaled gas through the exhalation

 

port. In noninvasive ventilation continuous air flow through the port flushes

 

exhaled gases from the circuit. The ability to completely exhaust exhaled gas

 

from the circuit depends on the EPAP setting and I:E ratio. Higher tidal

 

volumes further increase the volume of CO2 rebreathed by the patient.

WARNING:

To reduce the risk of CO2 rebreathing, monitor the patient for changes in

 

respiratory status at the start of ventilation and with each change in ventilator

 

settings, circuit configuration, or patient condition. Pay attention to ventilator

 

alarms that warn of increased CO2 rebreathing risk.

WARNING:

Be aware of the possibility of contamination from patient exhalate being

 

exhausted into the room through the exhalation port.

WARNING:

To ensure accuracy of oxygen administration and to monitor for the presence

 

of contamination (incorrect gas connected), use an external oxygen monitor

 

to verify the oxygen concentration in the delivered gas.

WARNING:

To reduce the risk of fire, use the ventilator in well-ventilated areas away from

 

flammable anesthetics. Do not use in a hyperbaric chamber or other similarly

 

oxygen-enriched environments. Do not use near an open flame.

WARNING:

To reduce the risk of electric shock from liquid entering the device, do not put

 

a container filled with a liquid on the ventilator.

WARNING:

To reduce patient risk of hypoxemia, keep free-flowing oxygen away from air

 

inlet of ventilator.

WARNING:

Connect to the ventilator only items that are specified as part of or compatible

 

with the ventilator system. Additional equipment connected to medical

 

electrical equipment must comply with the respective IEC or ISO standards.

 

Furthermore, all configurations shall comply with the requirements for

 

medical electrical systems (see IEC 60601-1-1 or clause 16 of edition 3 of

 

IEC 60601-1, respectively). Anybody connecting additional equipment to

 

medical electrical equipment configures a medical system and is therefore

 

responsible for ensuring that the system complies with the requirements for

 

medical electrical systems. Also be aware that local laws may take priority

 

over the above mentioned requirements. If in doubt, consult Respironics.

WARNING:

To reduce the risk of fire, explosion, leakage, or other hazard, take these

 

precautions with respect to the battery:

 

Do not attempt to disassemble, open, drop, crush, bend or deform, insert

 

 

foreign objects into, puncture, or shred the battery pack; modify or

 

 

remanufacture it; immerse or expose it to water or other liquids; expose

 

 

it to fire, excessive heat (including soldering irons); or put it in a

 

 

microwave oven.

 

Replace the battery only with another battery specified by the

 

 

manufacturer.

 

Follow all instructions for proper use of the battery.

 

Do not short-circuit the battery or allow metallic or conductive objects to

 

 

contact the battery connector housing.

 

Use the battery with the Respironics V60 ventilator only.

2-2

V60 Ventilator Service Manual

1049766 Rev A

Chapter 2

Warnings and Cautions

WARNING:

The nurse call/remote alarm should be considered a backup to the ventilator’s

 

primary alarm system.

WARNING:

To ensure that the alarm will be heard, make sure the alarm loudness is

 

adequate and avoid blocking the alarm speakers beneath the ventilator.

CAUTION:

Federal law (USA) restricts this device to sale by or on the order of a

 

physician.

CAUTION:

The Respironics V60 ventilator is designed to operate in the temperature

 

range of 5 to 40 ºC (41 to 104 ºF). To minimize the risk of overheating the

 

device, do not operate adjacent to heaters or other heat sources.

NOTE:

The displays shown in this manual may not exactly match what you see on

 

your own ventilator.

NOTE:

Pressures are indicated on the ventilator in cmH2O. Millibars and

 

hectopascals (hPa) are used by some institutions instead. Since 1

 

millibar equals 1 hPa, which equals 1.016 cmH2O, the units may be

 

used interchangeably.

NOTE:

The ventilator is not intended for use as an ambulance transport ventilator

 

or as an Automatic Transport Ventilator as described by the American

 

Hospital Association and referenced by the FDA. It is intended to allow

 

the patient to be transported within the hospital setting using a cart to

 

move the ventilator.

NOTE:

When attachments or other components or subassemblies are added to

 

the ventilator breathing system, the pressure gradient across the

 

ventilator breathing system, measured with respect to the ventilator

 

outlet, may increase.

NOTE:

The Respironics V60 ventilator parts that have patient contact are free of

 

latex.

NOTE:

If an alarm persists for no apparent reason, discontinue ventilator use and

 

contact Respironics.

NOTE:

If you detect any unexplained changes in the performance or visual

 

displays of the ventilator, discontinue ventilator use and contact

 

Respironics.

NOTE:

The Respironics V60 ventilator does not support automatic record

 

keeping.

1049766 Rev A

V60 Ventilator Service Manual

2-3

Chapter 2

Warnings and Cautions

2.2 Preparing for Ventilation

WARNING:

To ensure the correct performance of the ventilator and the accuracy of

 

patient data, we recommend you use only Respironics-approved accessories

 

with the ventilator.

WARNING:

To prevent possible asphyxia and to reduce the risk of CO2 rebreathing, take

 

these precautions with respect to mask and exhalation port use:

 

Use only a mask with an exhalation port or a nasal mask for noninvasive

 

 

ventilation.

 

Do not occlude the exhalation port.

 

Turn on the ventilator and verify that the port is operational before

 

 

application. Pressurized gas from the ventilator should cause a

 

 

continuous flow of air to exhaust from the leak port, flushing exhaled gas

 

 

from the circuit.

WARNING:

Never leave the mask on the patient while the ventilator is not operating.

 

When the ventilator is not operating, the exhalation port does not allow

 

sufficient exhaust to eliminate CO2 from the circuit. Substantial CO2

 

rebreathing may occur.

WARNING:

To ensure normal air circulation and exchange, do not cover or block the

 

ports on the ventilator or ventilator circuit. Do not block the air inlet panel on

 

the right side of the ventilator.

WARNING:

To prevent possible patient injury and possible water damage to the ventilator,

 

make sure the humidifier is set to appropriate temperature and humidification

 

settings.

WARNING:

To prevent the possibility of inadequate humidification, pay close attention to

 

the humidifier’s functioning when operating the ventilator at an ambient

 

temperature > 30 ºC (86 ºF). The ventilator warms the air delivered to the

 

patient above ambient temperature, which may impair the humidifier’s

 

performance.

WARNING:

To reduce the risk that the patient will aspirate condensed water from the

 

breathing circuit, position any humidifier lower than both the ventilator and

 

the patient.

WARNING:

To prevent possible patient injury and equipment damage, do not turn the

 

humidifier on until the gas flow has started and is regulated. Starting the

 

heater or leaving it on without gas flow for prolonged periods may result in

 

heat build-up, causing a bolus of hot air to be delivered to the patient. Circuit

 

tubing may melt under these conditions. Turn the heater power switch off

 

before stopping gas flow.

WARNING:

To reduce the risk of fire, use only patient circuits intended for use in oxygen-

 

enriched environments. Do not use antistatic or electrically conductive

 

tubing.

WARNING:

To prevent patient or ventilator contamination, recommend you use a

 

Respironics-approved main flow bacteria filter on the patient gas outlet port.

 

Filters not approved by Respironics may degrade system performance.

2-4

V60 Ventilator Service Manual

1049766 Rev A

Chapter 2

Warnings and Cautions

WARNING:

To reduce the risk of bacterial contamination or damage, handle bacteria

 

filters with care.

WARNING:

To reduce the risk of strangulation from patient tubing, use a tubing support

 

arm and secure the proximal pressure line with clips.

WARNING:

To reduce the risk of electric shock, connect the ventilator to an AC supply

 

mains with protective earth only.

WARNING:

Do not use extension cords, adapters, or power cords with the ventilator that

 

are not approved by Respironics.

WARNING:

To prevent unintentional disconnection of the power cord, always use the

 

correct, Respironics-supplied power cord and lock it into place with the

 

power cord retainer before you switch the ventilator on. The retainer is

 

designed to hold the connector end of the Respironics-supplied cord securely

 

in place.

WARNING:

To reduce the risk of electric shock, regularly inspect the AC power cord and

 

verify that it is not frayed or cracked.

WARNING:

To reduce the risk of strangulation, route the power cord to avoid

 

entanglement.

WARNING:

To reduce the risk of power failure, pay close attention to the battery’s charge

 

level. The battery’s operation time is approximate and is affected by ventilator

 

settings, discharge and recharge cycles, battery age, and ambient

 

temperature. Battery charge is reduced at low ambient temperatures or in

 

situations where the alarm is continuously sounding.

WARNING:

To ensure the ventilator’s safe operation, always run the full preoperational

 

check described in the operator’s manual before using the ventilator on a

 

patient. If the ventilator fails any tests, remove it from clinical use

 

immediately. Do not use the ventilator until necessary repairs are completed

 

and all tests have passed.

WARNING:

To prevent possible patient injury, disconnect the patient from the ventilator

 

before running the preoperational check. Make sure another source of

 

ventilatory support is available.

WARNING:

To prevent possible patient injury due to nonannunciating alarms, verify the

 

operation of any remote alarm device before use.

CAUTION:

To prevent possible damage to the ventilator, ensure that the connection

 

to the oxygen supply is clean and unlubricated, and that there is no water

 

in the oxygen supply gas.

CAUTION:

For 120 V equipment, grounding reliability can only be achieved when it

 

is connected to an equivalent receptacle marked “hospital only” or

 

“hospital grade.”

1049766 Rev A

V60 Ventilator Service Manual

2-5

Chapter 2

Warnings and Cautions

2.3Operation

2.4Alarms and Messages

2.5Care and Maintenance

WARNING:

To prevent possible patient injury, avoid setting alarm limits to extreme

 

values, which can render the alarm system useless.

WARNING:

If AC power fails and the backup battery is not installed or is depleted, an

 

audible and visual alarm annunciates for at least 2 minutes. Immediately

 

discontinue ventilator use and secure an alternative means of ventilation. As

 

in most ventilators with passive exhalation ports, when power is lost,

 

sufficient air is not provided through the circuit and exhaled air may be

 

rebreathed.

WARNING:

WARNING: To reduce the risk of electric shock, power down the ventilator and

 

disconnect it from AC power before cleaning or servicing it.

WARNING:

WARNING: To prevent patient or ventilator contamination, inspect and replace

 

the main flow bacteria filter between patients and at regular intervals (or as

 

stated by the manufacturer).

WARNING:

WARNING: To prevent possible patient injury, inspect and verify the proper

 

operation of the exhalation port regularly during use.

CAUTION:

Do not attempt to sterilize or autoclave the ventilator.

CAUTION:

To prevent possible damage to the ventilator, use only those cleaning

 

agents listed in this manual.

CAUTION:

To prevent possible damage to the touchscreen, take care when cleaning

 

it. Do not drip water and/or soap solution. After cleaning and rinsing,

 

remove all moisture with a dry, soft cloth. Never clean the touchscreen

 

with an abrasive brush or device, since this will cause irreparable

 

damage.

CAUTION:

To avoid introducing foreign matter into the ventilator and to ensure

 

proper system performance, change the air inlet filter at regular intervals

 

(or as stipulated by your institution).

CAUTION:

To ensure proper system performance, use a Respironics-approved air

 

inlet filter.

CAUTION:

Because some environments cause a quicker collection of lint and dust

 

than others, inspect the filters more often when needed. The air inlet

 

filter should be replaced; the cooling fan filter should be cleaned.

CAUTION:

To prevent possible damage to the ventilator, always ship it with the

 

original packing material. If the original material is not available, contact

 

Respironics to order replacements.

2-6

V60 Ventilator Service Manual

1049766 Rev A

2.6 First-Time

Installation

2.7 Communications

Interface

Chapter 2

Warnings and Cautions

WARNING:

Never attempt to disconnect or reconnect the battery during operation.

CAUTION: To prevent possible damage to the ventilator, always secure it to its stand or securely place it on a flat, stable surface that is free of dirt and debris. Do not use the ventilator adjacent to, or stack it with, other equipment.

WARNING:

Connect to the ventilator only items that are specified as part of or compatible

 

with the ventilator system. Additional equipment connected to medical

 

electrical equipment must comply with the respective IEC or ISO standards.

 

Furthermore, all configurations shall comply with the requirements for

 

medical electrical systems (see IEC 60601-1-1 or clause 16 of edition 3 of

 

IEC 60601-1, respectively). Anybody connecting additional equipment to

 

medical electrical equipment configures a medical system and is therefore

 

responsible for ensuring that the system complies with the requirements for

 

medical electrical systems. Also be aware that local laws may take priority

 

over the above mentioned requirements. If in doubt, consult Respironics.

WARNING:

It is the responsibility of the end user to validate the compatibility and use of

 

information transmitted from the ventilator with the device to be connected to

 

the ventilator.

WARNING:

The data provided through the communications interface is for reference only.

 

Decisions for patient care should be based on the clinician’s observations of

 

the patient.

WARNING:

To prevent possible patient injury due to nonannunciating alarms, verify the

 

operation of any remote alarm device before use.

WARNING:

To ensure the functionality of the remote alarm, connect only Respironics-

 

approved cables to the remote alarm port.

CAUTION:

The remote alarm port is intended to connect only to an SELV (safety

 

extra-low voltage and ungrounded system with basic insulation to

 

ground), in accordance with IEC 60601-1. To prevent damage to the

 

remote alarm, make sure the signal input does not exceed the maximum

 

rating of 24 VAC or 36 VDC at 500 mA with a minimum current of

 

> 1 mA.

2.8 Diagnostic Mode

WARNING:

To prevent possible patient injury, do not enter the diagnostic mode while a

 

 

patient is connected to the ventilator. Verify that the patient is disconnected

 

 

before proceeding.

1049766 Rev A

V60 Ventilator Service Manual

2-7

Chapter 2

Warnings and Cautions

(This page is intentionally blank.)

2-8

V60 Ventilator Service Manual

1049766 Rev A

Chapter 3. Theory of Operation

3.1 Pneumatics

The V60 ventilator is a microprocessor-controlled gas flow control and monitoring system that can deliver air and oxygen to augment or replace the work normally performed by the patient’s respiratory system. The ventilator uses electromechanical control circuits, flow and pressure monitors, and software programs to deliver pressure controlled breaths.

The ventilator includes a user interface (UI), internal blower, and gas delivery subsystem (GDS) that mixes air and oxygen. The ventilator can operate from a 40 to 87 psig (276 to 600 kPa) medical grade oxygen source for enriched oxygen operation. The internal power supply that can operate from mains (100 to 240 V AC 50/60 Hz) or internal battery (14.4 V DC) power. The ventilator also includes several communications interfaces.

Schematic diagrams of the V60 ventilator are available upon request.

The pneumatic subsystem delivers and monitors pressurized gas to the patient in response to commands from the CPU subsystem. The pneumatic subsystem includes the:

Manifold

Blower

Oxygen solenoid valve

Air and oxygen flow sensors

Pressure transducers

Solenoid valves

Motor controller (MC) PCBA

Data acquisition (DA) PCBA

The ventilator uses ambient air and high-pressure oxygen. Air enters through an inlet filter. Oxygen enters though a high-pressure inlet, and a proportional valve provides the operator-set concentration. The system mixes the air and oxygen, pressurizes it in the blower, and then regulates it to the user-set pressure. To do this, the ventilator compares the proximal (patient) pressure measurement with the ventilator outlet (machine) pressure, and adjusts the machine pressure to compensate for the pressure drop across the inspiratory filter, patient circuit, and humidifier. This helps ensure accurate and responsive pressure delivery and leak compensation.

The ventilator delivers gas to the patient through a main flow (inspiratory) bacteria filter, a single-limb patient breathing circuit, a humidification device (optional), and a patient interface such as a mask or ET tube. A pressure tap

1049766 Rev A

V60 Ventilator Service Manual

3-1

Philips V60 Service Manual

Chapter 3: Theory of Operation

proximal to the patient is used to monitor patient pressure. The internal exhalation port continually clears gas from the ventilator airway to ensure delivery of an accurate oxygen mixture.

Figure 3-1 shows a pneumatic schematic of the V60 ventilator.

Proximal pressure sensor

Machine pressure sensor

Ambient pressure

 

 

 

 

 

 

 

 

 

Air inlet

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ambient

 

 

 

 

 

 

 

 

 

filter

 

 

 

Bypass

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

pressure

 

 

 

 

 

 

 

 

 

 

 

 

element

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ambient air

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Purge

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Machine

 

inlet

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

solenoid

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

pressure

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

line

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Air flow

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Proximal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

sensor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

pressure

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Mixing

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Patient port

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

air & O2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2 inlet

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

pressure

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2 flow

 

 

 

 

Blower

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Barometric

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

sensor

 

 

 

 

 

 

 

 

 

 

pressure sensor

 

 

 

 

 

 

 

 

O2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

inlet

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Proportional

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

High pressure

 

 

filter

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

solenoid valve

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

oxygen inlet

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Bypass

 

 

 

 

 

 

 

 

 

 

 

Ambient

 

 

 

 

 

 

 

 

 

Sintered flow

 

 

 

 

 

 

 

 

 

 

 

 

pressure

 

 

 

 

 

 

 

 

 

 

 

element

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

normalizers

Internal exhalation port

 

Figure 3-1: V60 Ventilator Pneumatic Subsystem Schematic

3.1.1Air Inlet

Ambient air is entrained through the air inlet.

Air inlet

3.1.2Air Inlet Filter

The air inlet filter is designed to filter 5-micron particles at 70% efficiency at 150 SLPM flow.

Air inlet filter

3-2

V60 Ventilator Service Manual

1049766 Rev A

Chapter 3: Theory of Operation

 

3.1.3 Air Flow Sensor

 

The air flow sensor measures a subset (bypass flow) of total flow in the

 

pneumatic air path and interpolates the measurements according to constants

Manifold

that are calculated during gas delivery subsystem (GDS) calibration. The air

Air flow sensor

flow sensor also helps provide closed-loop control of gas flow during oxygen

 

blending.

3.1.4 Machine and Proximal Pressure Transducers

The machine and proximal pressure transducers on the DA PCBA measure the machine and proximal pressure over a range of -20 to +65 cmH2O.

3.1.5 Barometric Pressure Transducer

The barometric pressure transducer on the DA PCBA measures barometric pressure over a range of 525 to 850 mmHg.

3.1.6 Oxygen Pressure Transducer

The oxygen pressure transducer on the DA PCBA measures inlet pressure over

Oxygen pressure

Proximal

pressure

transducer

transducer

 

Barometric

Machine pressure

pressure

transducer

transducer

 

DA PCBA

a range of 0 to 87 psig. An alarm results if oxygen supply pressure is below 40 psig (276 kPa) or above 92 psig (634 kPa).

1049766 Rev A

V60 Ventilator Service Manual

3-3

Chapter 3: Theory of Operation

Oxygen filter element

3.1.7 Manifold, Oxygen Inlet Filter, Filter Element

Oxygen

inlet

filter The manifold includes a connection for the oxygen inlet, allowing countryspecific oxygen connections to be attached to the manifold. The manifold provides the pneumatic interfaces to the air inlet, oxygen inlet, blower inlet, and proximal and machine pressure lines.

Manifold

The oxygen inlet filter removes 5-micron particles from the oxygen gas supply. A 40-micron sintered bronze filter element acts to reduce turbulence in the oxygen flow. The 40-micron oxygen breather vent reduces noise in the oxygen flow.

Oxygen breather vent

3.1.8Oxygen Solenoid Valve

The oxygen solenoid valve and valve driver circuitry control the flow of oxygen according to the set O2 and flow. The oxygen solenoid valve is closed when there is a loss of power or system reset.

Manifold

Oxygen solenoid valve

Manifold

Oxygen flow sensor

Blower

3.1.9Oxygen Flow Sensor

The oxygen flow sensor measures a subset (bypass flow) of the total flow. These measurements are interpolated according to constants that are calculated during GDS calibration. Together, the oxygen solenoid valve and flow sensor provide closed-loop control for delivered oxygen flow.

3.1.10 Blower

The blower is controlled by the MC PCBA, and generates flow and pressure for the system. The blower includes an impeller, housing, and a three-phase brushless DC motor. The blower delivers a maximum pressure of less than 125 cmH2O in a dead-head condition, can accelerate from 10,000 to 22,500 revolutions per minute (RPM) in 120 msec from a nominal 5- to 25-cmH2O pressure rise. Maximum motor speed is approximately 40,000 RPM. The blower motor has internal Hall Effect sensors that are monitored by the MC PCBA and measure impeller speed.

3-4

V60 Ventilator Service Manual

1049766 Rev A

Chapter 3: Theory of Operation

3.1.11 Solenoid Valves

Manifold

Four solenoid valves are mounted on the manifold and controlled by the DA Solenoid valves (x4) PCBA. These three-way autozero solenoids include SOL1 (purge solenoid), SOL2 and SOL4 (machine pressure autozero solenoids), and SOL3 and SOL4

(proximal pressure autozero solenoids).

Solenoid valves

SOL3, SOL4: Proximal pressure autozero solenoids. SOL4: Also connects machine pressure to the proximal pressure transducer during autozero.

SOL2, SOL4: Machine pressure autozero solenoids.

SOL1: Purge solenoid, uses the machine pressure line to purge the proximal pressure line.

1049766 Rev A

V60 Ventilator Service Manual

3-5

Chapter 3: Theory of Operation

3.2 Electronics

The electronics system provides software-based control and monitoring, power management, user input, display, subsystem I/O, external communication, and alarms. The electronics system includes a Cirrus EP9307 microcontroller for control and monitor processing.

Control tasks include breath delivery, patient data calculation, and alarm detection/response. Monitoring tasks include controlling the LCD, front panel keys and indicators, inputs, and primary alarm output. An independent watchdog control provides safety monitoring.

The electronics system includes:

Power management (PM) PCBA

Power supply

Internal battery

CPU PCBA

Motor controller (MC) PCBA

Data acquisition (DA) PCBA

Flow sensors

User interface

Liquid crystal display (LCD)

Backlight inverter PCBA

Touch screen assembly

Nav-ring assembly

Power switch overlay

Switch PCB

User interface (UI) PCBA

3-6

V60 Ventilator Service Manual

1049766 Rev A

PM PCBA

Power supply

Chapter 3: Theory of Operation

3.2.1Power management (PM) PCBA

Battery charging and management circuitry

Internal supply voltages (supply voltages: 3.3 V, 5 V, 12 V, 35 V).

Fan power and tach monitoring.

Backup alarm control circuitry, including power fail detection.

Power switch control circuitry.

System alarm and reset management.

Electrical interfaces between the CPU PCBA, LCD, and the user interface (UI) PCBA.

Includes an EEPROM for calibration data, board identification information, and PM PCBA power-on hours.

3.2.2Power Supply

Provides ventilator and battery recharging power from AC line voltage.

Converts AC line voltage (100 to 240 VAC, 50 to 60 Hz) into 24 VDC power.

Includes input over-current, output over-voltage, and output currentlimiting protection.

3.2.3 Internal Battery

• 14.4-V, 11.5-Ah lithium-ion battery has a run-time of at least six hours under normal conditions.

 

• Provides operating power when AC power is not available.

Internal battery

• Provides charge and temperature status to the PM PCBA.

 

 

• Internal circuitry monitors battery status, provides self-contained

 

fault control features, and communicates this information to the PM

 

PCBA.

1049766 Rev A

V60 Ventilator Service Manual

3-7

Chapter 3: Theory of Operation

3.2.4

CPU PCBA

• Microprocessor: once in a run state, the microprocessor can only be

 

reset by a watchdog timeout or out-of-specification power condition.

 

Monitors operation of the ventilator and controls delivery of air and

 

oxygen to the patient. Verifies safe ventilator operation.

• Flash memory: 8 MB program storage.

• RAM: 8 MB for program execution and volatile data storage.

• EEPROM: 4 KB of storage for board-specific information (including

 

operating hours, time since last service, serial numbers, part

 

numbers, and software and hardware revisions).

• Watchdog timer: disables the blower and oxygen flow if not strobed by

CPU PCBA

software within a predefined time window that is independent of the

 

CPU master clock. In addition, ensures that software is operating.

• Real-time clock (RTC): a time of day clock that provides the date and

 

time to the ventilator, and is powered by a dedicated 3-V lithium coin

 

cell battery.

• LCD interface supports a display of 1024 horizontal x 768 vertical

 

pixels in 256 colors with a refresh rate of at least 50 Hz. The CPU

 

PCBA controls LCD brightness by varying a control voltage over a

 

range of 0 to 3.5 V (minimum to maximum brightness).

• Touch screen interface supports a five-wire type touch screen.

• Nav-ring rotary adjustment interface with a minimum resolution of 24 ticks per revolution.

• User key switches: interfaces to front panel keys.

• Alarm subsystem: includes a speaker driver circuit for the two main speakers, a backup piezo alarm, and a three-wire relay-controlled remote alarm interface (normal open, NO, or normal closed, NC) on the ventilator back panel.

• Blower speed monitor measures blower speeds from 3,000 to 50,000 RPM with 2% accuracy.

• Electrical interfaces to the power management (PM) and motor controller (MC) PCBAs.

• Two USB ports and an ethernet connection are designed for future enhancements. Table 3-1 summarizes hospital information system (HIS) RS-232 port pinout.

3-8

V60 Ventilator Service Manual

1049766 Rev A

Loading...
+ 216 hidden pages