Olympus CLV-180 User manual

5 (1)

MAINTENANCE MANUAL

CLV-180

ISSUE 1

FR-1700

CLV-180

INTRODUCTION

Introduction

This manual is intended for Olympus-certified technicians. Use of this manual by other individuals is prohibited.

Precautions

Before Repair

(1)Notify the site manager of the intent to repair the unit and obtain his/her approval before commencing.

(2)During repair, there is a risk of injury from the various tools and parts in the vicinity, as well as from fluid leakage on the floor. Inform the relevant people to restrict access to the repair area.

(3)To prevent potentially dangerous health risks, avoid working in a closed room, i.e., select a well-ventilated location when using organic solvents.

(4)In general, it is advisable to record the function and operation settings before repair, to enable restoration of the settings after service.

(5)If the original settings cannot be known due to mechanical problems present at the time the unit was accepted for repair, apply the factory-set values, or the safest settings (such as the lowest output levels). In such a case, inform the user that the settings have been changed.

(6)Guard against static electricity.

Use a conductive mat or wristband to discharge static electricity to prevent damaging the boards or other electrical components if it is necessary to touch them.

During Repair

(1)To prevent potentially dangerous health risks, thoroughly rinse any bodily areas that have come into contact with organic solvents as soon as possible.

(2)When using organic solvents, handle flames such as those in alcohol lamps with caution, because these solvents may ignite if exposed to flame.

In addition, always replace the lids back onto organic solvent containers before leaving the work-bench.

(3)Beware of electric shock.

Turn off the power and unplug the power cord before removing the cover of the unit.

(4)Beware of residual voltages.

The unit may contain residual charges in capacitor components. Take care to avoid electric shock when opening the top cover.

(5)To avoid personal injury and damage to the unit, heavy units should always be assembled or disassembled by at least two people.

(6)Repair with extreme caution to avoid injury.

Use extra caution around metal parts because the edges may be sharp.

(7)Reassemble parts according to their original configuration. This regards the following items especially:

a)Insulators, such as insulating tubes and mylar sheets.

b)Cable rerouting, clamps, and cores.

 

1

ISSUE1

INTRODUCTION

CLV-180

c)Shield parts and cover screws with toothed washers.

Failure to attach parts in their original configurations, even if it does not impair product function, poses the risk of noise radiation and reduced electrical safety.

(8)Use specified parts.

The parts and components of this product are designed to operate under certain anticipated vibration, heat, chemical exposure, and voltage conditions. Always replace parts with those specified in the parts list.

(9)Always tighten nuts/screws to the specified torque and only use components of the specified dimensions.

(10)Do not reuse O-rings, E-rings, or packing. Always use new ones.

(11)When reusing components, remove sealing tape or sealing compound, and clean the components before use.

(12)Check that no screws or nuts are loose.

After Repair

(1)After repair, inform the site manager of the nature of the problems, the cause, the countermeasures taken, and the parts replaced.

(2)Always notify the site manager if, during repair, liquid disinfectants, cleaners, or alcohol were used. Inform the site manager that before using the unit, it is necessary to verify the concentration levels of disinfectants to ensure that they have not been diluted.

(3)Restore the original unit settings as they were recorded before repair. Request the presence of site manager to verify the settings with you.

Copyright

© 2005 Olympus Medical Systems Corp. All rights reserved. Unauthorized reproduction or distribution in part or in whole is prohibited.

Trademarks

OLYMPUS is a registered trademark of the Olympus Corporation.

The company names, product names, and proprietary technical terms in this document are the trade-marks or registered trademarks of their respective owners.

 

2

ISSUE1

INTRODUCTION

Contents

Chapter 1: Product Specifications………………………………………...1-1

1Product Outline…………………………………..…………………………………………..1-2

2Features………………………….…………………..………………………………………..1-2

3Operational Conditions……………………………..……………………………………...1-3

4Specifications………….………………….…………..………………………………….…1-4

5Nomenclature and Functions…………………………………………..……………………..1-9 5-1 Front panel………………………………………………….…………………………….... 1-9 5-2 Control panel(buttons)……………………………………………………………………1-10 5-3 Control panel(indicators)………………………………………………………………...1-12 5-4 Rear and side panels……………………………………………………………………..1-14

Chapter 2: Troubleshooting………………………………………..…….2-1

1Contents……………………………………………………………………………………….2-2

2Troubleshooting…………………………...…………………………………………………….2-2 2-1 Power input failure…………………………………………………………..………….2-3 2-2 Lamp failure………………………………………………………………..………….2-4 2-3 Lamp malfunction…………….……………………………….…………..………….2-4 2-4 Scope connection failure………………………………………………………………….2-5 2-5 High-brightness malfunction……………………………………………………………..2-6 2-6 Cooling Fan malfunction………………………………………………………………….2-7 2-7 Shield malfunction at Scope removal…………………………………………………..2-7 2-8 Pump malfunction………………………………………………………………………….2-8 2-9 Filter Change Malfunction………………………………………………………………..2-9 2-10 Outgoing light failure……………………………………………………………………2-11 2-11 Manual Brightness failure……………………………………………………………...2-11 2-12 Automatic Brightness failure………………………………………………………….2-12 2-13 Panel Malfunction……………………………………………………………………….2-13 2-14 Emergency Lamp malfunction………………………………………………………..2-15 2-15 Back-up malfunction…………………………………………………………………..2-16 2-16 Lamp Life Meter Reset malfunction…………………………………………………2-16 2-17 Transillumination malfunction……………………………………………………….2-16

Chapter 3: Precaution on disassemble and reassembly……..…..….3-1

1Warn ing ……………………..……………………………………..……………………….3-2

2Cau t io n……………………………………………….…………………………………….3-2

Chapter 4: Assemble procedure…………………………..…..….4-1

1Jigs and Tools……………………………………………………………………………….4-2

2Areas……………………………………………………….…………………………………….4-2

3Disassembling and Reassembling procedure……………………………………………….4-3

Ch ap t er 5: Par t s L i s t ……....…………………………..…..….5-1

1Exploded Parts Diagram…………………………..…………………………………………….5-2

2Parts Li st …………………………………….…………………….….…..…………..………….5-10

Specification CLV-180

Chapter 1: Products Specifications

1-1

ISSUE1

Specification CLV-180

PRODUCT SPECIFICATIONS

1 Product Outline

Intended use:

When used in conjunction with an OLYMPUS endoscope, video system center, video monitor, accessories, and other related equipment, this product supports examination and endoscopic treatment by enabling observation through a fiberscope or through a video monitor.

Compatibility:

This product may be used in conjunction with EVIS 100/130/140 series videoscopes, EVIS EXERA 145/160 series videoscopes, EVIS EXERA II 165/180 series videoscopes, V series videoscopes, VISERA series videoscopes, rigid light guides, fiberscopes, as well as EVIS EXERA II CV-180/165, and endoscopic ultrasound systems.

Durability:

With adherance to the conditions listed below, this product has a usable life of 6 years from the manufactured/shipped date.

Conditions: During this product’s usable life, the user must perform inspection before using, and regular subsequent inspections as outlined in the attached materials and the operation manual. Should the inspections reveal that repairs are required, the user must have those repairs performed.

2Features

(1)This is a light source unit for exclusive use with the EVIS EXERA II VIDEO SYSTEM CENTER CV-180/165. However, this product is not to be used for surgery if used with the CV-165 (high-intensity function is not available).

(2)A mechanical detection unit is installed in the endoscope socket, enabling automatic switching to the maximum amount of light appropriate for any endoscope connected. It is capable of distinguishing between GI-type scope connectors, LG connectors for use with the SP high-intensity function, and LG connectors for use with the non-SP, high-intensity function.

(3)When videoscopes or OES fiberscopes (including OES connectors used with BF and CHF type endoscopes) are attached, the unit is capable of generating 1.3 times the output light intensity of the CLV-U4 EVIS 100 mode and the OES mode (OAI, OLA). It produces 1.6 times more maximum permissible illumination (white light) for the LG (non-OAI, OLA regions).

(4)When connected to a surgical fiberscope not compatible with the high-intensity RF connector, it produces an output light intensity of the maximum permissible illumination (white light) for the LG equivalent to the CLV-U4 EVIS 100 mode and the OES mode.

(5)When used with an endoscope and light guide compatible with the high-intensity RF connector, the high-intensity mode is available, creating a brighter illumination. Compared to the illumination of the CLV-U40 EVIS 100 mode or OES mode, approximately 2 times the maximum illumination (white light) is produced. In high-intensity mode, it is also possible to select equal brightness or double brightness of the CLV-U40 through the brightness mode switch.

(6)The optical filter used in NBI (narrowband imaging) observation has been loaded. NBI observations are enabled when an NBI-compatible endoscope is connected to the unit.

(7)The optical filter used in PDD (photodynamic diagnosis) observation can be loaded (OE market). One other optical filter for specialized observations may be loaded if required.

(8)The following modes may be selected:

a.Auto-ignition mode: the examination lamp automatically illuminates when the power is turned on.

b.Manual ignition mode: the examination lamp will illuminate when the lamp switch is pressed once

power has been turned on.

(9)The examination lamp can be manually turned off by continously pressing the LAMP switch for approximately 1 second.

(10)Heat exhaust is through the rear (rear heat exhaust).

(11)The system automatically switches to the emergency lamp if the examination lamp does not come on after pressing the lamp switch. As an emergency response, this will provide sufficient brightness required for endoscope removal in situations when the examination lamp does not come on or if it turns off. This will be indicated on the panel.

(12)Brightness can be adjusted using the 17 steps of automatic and manual brightness available.

(13)The air supply can be stopped completely as well as changed between high, medium or low supply.

(14)Disconnection of the emergency lamp will be automatically indicated on the panel.

(15)The transillumination switch enables the transmitted light illumination function (for GI endoscopes

1-2

ISSUE1

Specification CLV-180

only).

(16)Panel settings are retained, and the light source unit will display at the previous settings the next time the unit is turned on. However filter switch settings are not backed up, and will revert to the regular filter.

(17)The exterior and the operating panel may be disinfected with ethanol (70 ethyl or isopropyl alcohol) or with a mild detergent.

(18)This product has passed the OLYMPUS Eco-Product standard (Environmentally-friendly design standard).

3Operational Conditions

(1)Applicable video system centers: CV-180, CV-165

(2)Operational environment:

i)Use in a medical facility under the supervision of a medical doctor.

ii)Do not apply this light source directly to the heart.However, it may be used with the heart when

used in combination with a TYPE CF applied part indicated by a

mark on the instrument.

iii)The outer casing of the unit must be grounded for safety.

iv)Do not use the light source in a combustible atmosphere.

v)Do not use the light source with the following electronic equipment or endoscopes:

a.Any devices designed to apply electronic treatment of a patient for which safe usage with the CLV-180 has not yet been confirmed.

b.Any devices not designed to apply electronic treatment of a patient for which safety (e.g., leakage currents) has not yet be confirmed.

vi)Power supply:

 

Voltage:

100

– 120 V AC

 

Frequency:

50/60Hz

 

Input current:

500

VA

 

Power fluctuation:

Within ±10%

 

Frequency fluctuation:

Within ±1 Hz

vii)Environment

 

 

 

a.

Operating environment

 

 

 

 

Ambient temperature:

10

– 40°C

 

Relative humidity:

30

– 85%

 

Air pressure:

700

– 1060 hPa

b.

Storage environment

 

 

 

 

Ambient temperature:

25

– 70°C

 

Air pressure:

700

– 1060 hPa

1-3

ISSUE1

 

 

 

 

 

Specification

CLV-180

 

4

Specifications

 

 

 

 

 

 

 

Item

Specification

 

 

 

 

 

 

 

1.

Videoscopes

1. EVIS 100, 130, 140 series videoscopes

 

 

 

 

2. EVIS EXERA 145, 160 series videoscopes

 

 

 

 

3. V series videoscopes

 

 

 

 

4. EVIS EXERA II 165, 180 series videoscopes

 

 

 

 

5. VISERA series videoscopes

 

 

2.

EUS Videoscopes

1. EUS 130, 140, 160 series videoscopes

endoscopesCompatible

endoscopesofType

3.

Fiberscopes

1. OES 10, 20, 30, 40, 60 series fiberscopes

 

 

 

 

2. E, E2, E3 series fiberscopes

 

 

 

 

 

 

 

4.

Light guide for use

High-intensity compatible light guides

 

 

 

with rigid

Note:

The high-intensity mode can be set by pressing the

 

 

 

endoscopes

 

high-intensity mode selection switch when connecting the

 

 

 

 

 

light guide.

 

 

5.

Rigid endoscopes

Not compatible with rigid products other than those with

 

 

 

(Model names

high-intensity compatible light guides.

 

 

 

omitted)

 

 

 

 

 

1.

6.

High-intensity

CHF-CB30L, CHF-CB30S, URF-P3

1.

 

compatible

(LF-TP, LF-DP, LF-GP) *1

 

 

 

 

fiberscopes

(ENF-GP) *2

 

 

 

 

 

 

(for stiff eyepiece

*1

Applicable when used in combination with the following

 

 

 

endoscopes)

 

light guides: A3290, A3291, A3292, A3293, A3294, A3295,

 

 

 

 

 

A3296, A3297, A3298.

 

 

 

 

*2

Applicable when using in combination with light guide

 

 

 

 

 

A3293.

 

 

7.

High-intensity

LTF-V, LTF-V2, LTF-V3

 

 

 

compatible

 

A4940A, A4941A, A4942A, A4943A, A4800A, A4801A,

 

 

 

videoscopes

 

A4802A, A4803A, A4804A, A4805A, A50000A, A50001A,

 

 

 

 

 

A50010A, A50011A, A50020A, A50021A

 

 

 

Illumination light path

Emergency lamp

 

 

 

 

Condenser lens

 

 

 

 

Optical filters

 

 

 

 

Turret board (emergency lamp, filter, etc.)

Illuminationfunction

Optics1.

 

 

Diaphram

 

 

 

above, the mesh filter with desired aperture ratio mounted

 

 

 

 

Light-adjustment mesh turret

 

 

 

 

(1)

Each of the optical filters on the turret plate are

 

 

 

 

 

automatically placed in the light path according to the

 

 

 

 

 

normal observation mode (GI/SP), special observation

 

 

 

 

 

mode (GI/SP), or high-intensity mode (SP) settings.

 

 

 

 

(2)

In accordance with the various observational modes

2.

 

 

 

 

on the mesh turret can be automatically placed in the light

 

 

 

 

path.

 

 

 

 

 

 

 

 

 

(3)

The emergency lamp is automatically placed in the light

 

 

 

 

 

path when the examination lamp malfunctions.

 

 

 

 

(4)

There are four special filter frames provided to enable the

 

 

 

 

 

use of special filters.

1-4

ISSUE1

Specification CLV-180

2. Illumination light

2. Illumination function

Brightness adjustment

 

3.

Examination lamp

(1)

Type: Xenon short arc lamp with an eliptic mirror

 

 

 

 

(2)

Model: MD-631

 

(3)

Life: 500 hours (when lit continuously)

 

(4)

The lamp will either illuminate when the power is switched

 

 

on (auto-ignition funtion), or it can be turned on by

 

 

pressing the LAMP switch on the front panel. The lamp

 

 

can be turned off by pressing the LAMP switch

 

 

continuously for one second.

Emergency lamp

(1)

Type: Halogen 35 W (with reflector)

 

(2)

Life: Average of at least 500 hours

Method

(1)

Manual adjusment only for fiberscopes and rigid-type

 

 

endoscopes (excluding when used with a CV-180 and

 

 

camera head).

 

(2)

Automatic and manual adjustment available for

 

 

videoscopes.

 

(3)

Automatic adjustment is available when the CV-180 and

 

 

camera head are used with rigid endoscopes and

 

 

light-source cables.

NBI special observation

(1)

The filter mode switch enables selection of the NBI special

mode

 

observation mode when a light adjustment cable and the

 

 

CV-180 are connected to an NBI-compatible endoscope.

 

(2)

When the NBI special observation mode is selected, the

 

 

NBI special filter is placed in the illumination light path.

 

 

Note: The CV-180 power switch must be ON.

PDD normal observation

(1)

PDD normal observation mode will be activated when the

mode (OE)

 

light guide is connected to a light adjustment cable

PDD is option 1 for

 

equipped with the PDD function, and a camera head for

non-OE regions

 

exclusive use during PDD special observation are

 

 

connected to the CV-180.

 

(2)

During PDD normal observation mode, the PDD normal

 

 

filter is placed in the illumination light path.

 

 

Note: The CV-180 power switch must be ON.

 

 

The PDD normal filter and PDD special filter are

 

 

required for the PDD function.

 

 

 

PDD special observation

(1)

From the PDD normal observation mode, the PDD special

mode (OE)

 

observation mode may be selected by pressing the filter

PDD is option 1 for

 

mode switch.

non-OE regions

(2)

During PDD special observation mode, the PDD special

 

 

filter is inserted in the illumination light path.

 

(3)

Press the switch again to insert the PDD normal filter. The

 

 

filters can be exchanged by pressing the switch.

 

 

Note: The CV-180 power switch must be ON.

PDD special observation

(1)

With the CV turned OFF, attach the light-supply cable with

mode (OE)

 

the PDD function. When the light guide is attached,

(Fiberscope)

 

pressing the filter mode switch will select the PDD special

PDD is option 1 for

 

observation mode.

non-OE regions

(2)

During PDD special observation mode, the PDD special

 

 

filter is inserted in the illumination light path.

 

 

Note: The normal filter is used for PDD observation during

 

 

observation through a fiberscope, rather than the

 

 

PDD normal filter.

Optional observation

(1)

This function is reserved for experimental use.

mode (OE)

 

 

PDD is option 2 for

 

 

non-OE regions

 

 

1-5

ISSUE1

 

 

 

 

 

 

Specification

CLV-180

 

 

 

 

 

 

 

 

 

 

 

High-intensity mode

(1)

When a high-intensity compatible endoscope is inserted, the

 

 

 

 

 

high-intensity mode may be selected by pressing the

 

 

 

 

 

high-intensity switch. High-intensity mode may not be

 

 

 

 

 

selected with endoscopes that are not high-intensity

 

 

 

 

 

compatible.

 

 

 

 

 

 

 

(2)

When a high-intensity compatible endoscope is inserted, the

 

adjustment

 

 

 

high-intensity mode switch illuminates. This enables the

 

 

 

 

selection of the high-intensity mode (LED above the switch

 

 

 

 

 

 

 

 

 

 

illuminates) or the normal mode (LED above the switch goes

 

 

 

 

 

out).

 

 

 

 

 

 

 

(3)

The current settings for the high-intensity mode switch are

 

Brightness

 

 

 

retained when the power is turned OFF. When a

 

 

 

 

high-intensity compatible endoscope is inserted, illumination

 

 

 

 

 

 

 

 

 

 

will either be the high-intensity mode or normal mode,

 

 

 

 

 

depending on the settings retained by the system. Initial

 

 

 

 

 

factory setting: normal mode.

function

3.

 

(4)

When the high-intensity mode is terminated (normal mode),

 

 

 

illumination is equivalent to the CLV-U40 EVIS 100 and OES

 

 

 

 

 

 

 

 

 

 

mode.

 

 

 

 

 

 

 

(5)

High-intensity mode provides approximately twice the amount

Illumination

 

 

 

 

of illumination as the SP normal observation mode.

 

 

 

 

(image illumination)

 

 

Manual adjustment

(1)

Mechanical diaphram (set on the panel)

 

 

 

(2)

17 steps

 

 

 

 

 

 

Automatic adjustment

(1)

Mechanical diaphram with constant illumination control

 

 

 

(2)

17 steps

 

 

 

 

2.

 

Front panel display

 

 

 

 

 

 

 

 

 

method in special

 

Observation mode

Front panel display method

 

 

 

observation mode

 

 

 

 

 

 

 

 

 

 

NBI special

The NBI LED lights up (white)

 

 

 

 

 

 

 

 

 

 

 

observation mode

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PDD normal

The PDD LED lights up (green)

 

 

 

 

 

observation mode

(OE only)

 

 

 

 

 

 

 

 

 

 

 

 

PDD special

The PDD LED lights up (white)

 

 

 

 

 

observation mode

(OE only)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4. Cooling

Equipment

(1)

Forced air cooling with a fan (rear exhaust)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Air supply pump

(1)

Diaphragm system

supply

 

 

 

 

 

 

 

Air

Air supply pressure

(1)

Three air supply pressure levels (high, medium, low and

 

(2)

Maximum air supply pressure: less than 53.9 kPa

 

 

 

 

 

stop)

 

 

 

 

Air/Water3. function

1.Water

 

 

 

 

 

 

 

 

Control

(1)

Through air supply switch settings

 

 

 

 

 

 

 

 

 

 

 

 

Water supply

(1)

Supply is available through the endoscope distal end when

 

 

 

 

 

a water supply container is attached.

 

2.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1-6

ISSUE1

Specification CLV-180

 

Endoscope and light guide

 

One-touch connection

Connection4.

Light-supply cable (MAJ-1411)

Connector on the rear-panel

 

 

 

 

 

 

(Light-supply cable for CV connection)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Expansion cable (MAJ-202/972)

System connector on the rear-panel

 

 

 

 

 

 

 

(compatible with Endo GATE/Endo ALPHA)

 

 

 

 

 

 

 

 

Foot switch (MAJ-1391)

 

Foot switch connector on the rear-panel

Emergency indication

Emergency lamp indicator

 

Indicates if the emergency lamp is disconnected or if the

 

 

 

 

 

 

emergency lamp is in use (when the main lamp will not illuminate).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Lit up: Emergency lamp in use (the main lamp will not illuminate).

 

 

 

 

 

 

 

Blinking: Emergency lamp is disconnected or has been removed.

5.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Setting retention

 

 

 

(1)

Settings prior to powering off the unit are retained.

 

 

 

 

 

 

 

 

Manual / automatic brightness settings

 

 

 

 

 

 

 

 

Airflow settings

Panel

 

 

 

 

 

 

 

Light intensity level

 

 

 

 

 

 

 

High-intensity mode setting

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Special observation function display

6.

 

 

 

 

 

 

 

Lamp life indicator (with the power ON, the indicator may

 

 

 

 

 

 

 

 

be reset when the examination lamp is extinguished by

 

 

 

 

 

 

 

 

continuously pressing the counter reset switch.)

 

 

 

 

 

 

 

 

 

Switch illumination system

 

(1)

All switches illuminated

 

(not including counter reset

 

 

(not including counter reset switch)

 

switch)

 

 

 

 

 

 

 

Dimming1.

 

On endoscope removal

(1)

When the endoscope is removed, the light emanating from

 

 

 

 

 

 

 

the scope connector is dimmed by the dimmer plate

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

located in the endoscope socket.

 

 

 

 

 

 

 

 

 

 

Disinfection2.

 

Exterior

 

 

 

(1)

Disinfect with ethanol for disinfection (70 ethyl or

 

 

 

 

 

 

 

isopropyl alcohol) or with a mild detergent.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Safety

Alarm

 

Temperature switch

 

(1)

To ensure safety, when the internal temperature rises

 

 

 

 

 

 

above the specified value, the temperature switch turns off

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

and the electrical current supply to the lamp is terminated.

7.

3.

 

 

 

 

 

 

The unit detects that the temperature switch has shut off,

 

 

 

 

 

 

and sounds an alarm.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4. Electrical

 

Protection

 

 

 

Class 1 unit (3P power supply)

 

shock

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Region

Sold

MDD

 

 

 

 

Applicable5.

regulations

1. EU/ EFTA

Y

 

Class:II a (rule: 11-1)

 

 

CE marking: CEO197

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2. Japan

N

 

Pharmaceutical Law (Application of the Pharmaceutical law to assist Asian

 

 

 

 

market response)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3. USA

Y

 

FDC laws

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ISSUE1

 

 

 

 

 

 

1-7

 

 

 

 

 

 

 

 

 

6. Applicable external regulations

8. Other

Usage restrictions

 

1.

Specification CLV-180

4. Canada

 

Y

 

Canadian Medical Device and Equipment Regulation

 

 

 

Class: II

 

 

 

 

 

 

 

 

 

 

5. Singapore

 

Y

 

Singaporean medical device regulations

 

6. Other

 

Y

 

K-FDA (Korea)

 

 

 

 

SDA (China)

 

 

 

 

 

 

 

 

Common to

all

 

 

IEC 60601-1 (Medical electrical equipment - Part 1: General requirements

countries

 

 

for safety): 1988 + A1, A2

 

 

 

 

 

IEC 60601-1-1 (Medical electrical equipment - Part 1-1: General

 

 

 

 

requirements for safety): 2000

 

 

 

 

 

IEC 60601-1-2 (Medical electrical equipment - Part 1-2: General

 

 

 

 

requirements for safety - Collateral standard: Electromagnetic compatibility

 

 

 

 

- requirements and tests): 2001

 

 

 

 

 

IEC 60601-2-18 (Medical electrical equipment - Part 2: Particular

 

 

 

 

requirements for the safety of endoscopic equipment): 1996 + A1

 

 

 

 

ISO 14971(Medical devices -- Risk management): 2000

 

 

 

 

 

ISO 9000-3(Software engineering -- Guidelines for the application of ISO

 

 

 

 

9001:2000 to computer software): 1997

 

 

 

 

 

ISO 8600-1(Optics and photonics -- Medical endoscopes and endotherapy

 

 

 

 

devices -- Part 1: General requirements): 1997

 

 

 

 

 

ISO 7000(Graphical symbols for use on equipment)

: 2004

 

 

 

 

IEC 60417-1: 2002, -2: 1998(Graphical symbols for use on equipment)+

 

 

 

 

A1, A2

 

 

 

EU EFTA

 

 

EN 980 (Graphical symbols for use on equipment): 2003

 

 

 

 

 

 

2. Japan

 

 

JIS T1005 (Style manual for instruction manuals): 1983

 

 

 

 

 

JIS T0601-1 (Medical electrical equipment -- Part 1: General requirements

 

 

 

 

for safety): 1999

 

 

 

 

 

 

JIS T0601-1-1(Medical electrical equipment -- Part 1-1: General

 

 

 

 

requirements for safety -- Collateral standard: Safety requirements for

 

 

 

 

medical electrical systems): 1999

 

 

 

 

 

JIS T0601-1-2(Medical electrical equipment -- Part 1: General

 

 

 

 

requirements for safety -- 2. Collateral standard: Electromagnetic

 

 

 

 

compatibility -- Requirements and tests): 2002

 

3. USA

 

 

UL 60601-1: 2003

 

 

 

 

 

 

 

 

4. Canada

 

 

CAN/CSA-C22.2 No.601.1-M90: 1990

 

 

 

 

 

CAN/CSA-C22.2 No.601.1-S1: 1994

 

5. Other

 

 

Olympus Eco-Product Standard

 

Ambient temperature

 

 

10 – 40°C

 

 

 

 

 

 

 

 

 

 

 

 

Relative humidity:

 

 

30 – 85%

 

 

 

 

 

 

 

 

 

Air pressure

 

 

 

 

700 to 1060hPa

 

Usage environment

 

 

Do not use in a combustible atmosphere

 

Type of protection against

 

Class 1

 

 

electrical shocks

 

 

Degree of protection: Type BF applied part

 

However, this instrument may be applied to the heart if it is used with equipment classified as

TYPE CF applied part. (indicated by the

symbol)

 

Voltage

 

 

 

 

100 – 240 V AC

 

 

 

 

 

 

 

 

 

Frequency

 

 

 

 

50/60 Hz

 

 

Power fluctuation

 

Within ±10%

 

 

 

 

 

Frequency fluctuation

 

Within ±1 Hz

 

Power consumption

 

500VA

 

 

1-8

ISSUE1

 

 

 

 

 

 

Specification

 

 

 

 

 

 

 

 

CLV-180

 

 

 

 

 

 

 

 

 

replacment

User servicing

Examination lamp, fuses

 

 

 

 

 

 

 

 

 

 

2. Par need

OLYMPUS servicing

Emergency lamp replacement, special filter attachment

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Weight

Approximately 15.4

 

 

 

 

 

 

 

 

 

 

Dimensions

383 (W) × 536 (D) × 162 (D) mm (maximum)

 

 

 

 

370 () × 475 (D) × 150 (D) mm (standard)

 

 

 

 

 

 

 

 

 

 

Panel

Panel selection varies by region

 

 

Other

 

AC 100 - 240 English (3E, 6E, 6LA)

 

 

 

AC 100 - 120 English (3OA)

 

 

 

 

AC 100 - 240 Symbol (6S)

 

 

3.

 

AC 100

English (1J)

 

 

Power cord

Cord set with a 3-prong hospital grade

 

 

 

 

 

 

 

With plug (1J, 3OA, 3E)

 

 

 

 

Plug-less (6E, 6S,6LA)

 

 

 

 

 

 

 

 

 

 

 

 

Fuse

Fuse capacity

 

 

 

 

 

 

Littel (manufacturer) 215008: 8A

 

 

 

 

 

 

 

 

 

 

Warranty period

As specified in the applicable laws for each country

 

 

 

 

 

 

 

 

 

 

5 Nomenclature and Functions

5-1 Front panel

1-9

ISSUE1

Olympus CLV-180 User manual

Specification CLV-180

5-2 Control panel (buttons)

1-10

ISSUE1

Specification CLV-180

(1) Transillumination button

When pressing this button, light emitted from the endoscope’s distal end becomes brighter for 7 seconds, then returns automatically to its original brightness level.

(2) Intensity mode selector

Press to switch between the high-intensity mode and normal intensity mode when using an endoscope compatible with the high-intensity mode.

(3) Filter mode button

Pressing this button activates the NBI observation function.

(4) Air feed button

Pressing this button starts or stops the air feed from the endoscope’s distal end.

(5) Airflow regulator button

This button is use to control the pressure of the air being fed from the endoscope.

(6) Brightness adjustment buttons

These buttons are pressed to adjust the brightness level.

(7) Auto/manual brightness selector

This selector is pressed to select automatic or manual brightness control.

(8) Lamp button

This button is pressed to turn ON or OFF the examination (xenon) lamp.

(9) Counter reset button

After replacing the examination (xenon) lamp, the lamp hour meter is reset by pressing this button for 1.

1-11

ISSUE1

Specification CLV-180

5-3 Control panel (Indicators)

1-12

ISSUE1

Specification CLV-180

(1) Transillumination indicator

This indicator lights when the transillumination function is activated.

(2) High intensity mode indicator

The indicator lights when high-intensity mode is selected.

(3) Air feed ON/OFF indicator

This indicator lights when the air feed from the endoscope’s distal end is activated.

(4) NBI observation mode indicator

This indicator lights in green to indicate that the light source can perform the NBI observation. This indicator lights in green continuously when an endoscope with the NBI observation compatibility is connected and the light source is ready for the NBI observation. Also, the indicator lights in white when NBI observation mode is active.

The "OP.1" and "OP.2" are prepared for future use.

(5) Airflow regulator indicators

One of these indicators lights to indicate the current airflow pressure level setting ("L" (Low), "M" (Medium) or "H" (High)).

(6) Emergency lamp indicator

This indicator lights when the emergency lamp (halogen) is in use, and blinks when the emergency lamp (halogen) is disconnected or not mounted.

(7) Brightness level indicators

These indicators display the current brightness level.

(8) Auto/manual indicator

This indicator displays the brightness selector setting ("auto" or "manual").

(9) Lamp ON/OFF indicator

This indicator lights when the examination lamp (xenon lamp bulb) lights.

(10) Lamp hour meter

This indicator displays the total working hours of the examination (xenon) lamp.

1-13

ISSUE1

Specification CLV-180

5-4 Rear and side panels

1-14

ISSUE1

Specification CLV-180

(1) Fuse box

The fuses protect the light source from electrical surges.

(2) SYSTEM terminal

The terminal accepts connection from an external unit.

(3) CV terminal

This terminal is the receptacle for the light source cable to connect the light source to the EVIS EXERA II video system center.

(4) Lamp ignition selector

This selector is set to select automatic or manual ignition of the examination lamp. When the automatic ignition is selected, turning ON the light source lights the examination lamp simultaneously. When the manual ignition is selected, pushing the lamp button on the control panel lights the examination lamp.

(5) Foot switch terminal

This connector is the receptacle for the foot switch (MAJ-1391) for use in the naked-eye PDD observation.

(6) Water container holder

This holder is used for the installation of the water container.

(7) Potential equalization terminal

For safety, this terminal is connected to a potential equalization terminal of the other equipment connected with the light source, and the electric potential of their equipment are made the same.

(8) Lamp cover

This cover has to be removed to replace the examination lamp.

1-15

ISSUE1

Troubleshooting CLV-180

Chapter 2: Troubleshooting

2-1

ISSUE1

Troubleshooting CLV-180

1. Contents

Symptom

Failure Mode

2-1

No power

Power input failure

2-2

Xenon Lamp not lit

Lamp failure

2-3

Lamp going out

Lamp going out

2-4

Scope not mounting

Scope connection failure

2-5

Incapability of discriminating GI, SP, or high-brightness

Scope discrimination failure

 

SP

 

 

<GI Scope>

 

 

AIR button or Transillumination button cannot be

 

 

selected when GI Scope is connected.

 

 

<SP>

 

 

AIR button, Transillumination button, and

 

 

high-brightness button can be selected when SP

 

 

Scope is connected.

 

 

<High-brightness SP>

 

 

High-brightness button cannot be selected when

 

 

high-brightness Scope and Light Guide are

 

 

connected.

 

2-6

Cooling fan not working

Cooling fan malfunction

2-7

Exit light from Light Source when Scope is removed

Shield failure at scope removal

2-8

Pump not working when Pump ON switch is pushed, or

Pump malfunction

 

no change in air supply volume

 

2-9

Turret Board not rotating when Special Light

Filter changing malfunction

 

Observation Scope is connected and Mode switch is

 

 

pushed

 

2-10

No light from Scope

Exit light failure

2-11

View area is dark or excessively light

Manual Brightness failure

2-12

View area is dark or excessively light

Automatic Brightness failure

2-13

No change in LED display when switch is pushed

Panel malfunction

2-14

Emergency Lamp is not automatically lit when Xenon

Emergency Lamp malfunction

 

Lamp is put out

 

2-15

Values set before shutdown are not saved when power

Back up malfunction

 

shuts down

 

2-16

Time not changing to "0" when Lamp Life Meter Reset

Lamp Life Meter Reset malfunction

 

switch is pushed

 

2-17

No change in light quantity when Trans Illumination

Transillumination malfunction

 

switch is pushed

 

2-2

ISSUE1

Troubleshooting CLV-180

2. Troubleshooting CLV-180

2-1 Power input failure

Incapable to input power 2-1-1 Block diagram

2-1-2 Estimated location of failure

No

Estimated failure location

 

Inspection method

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Input voltage

1

Verify the voltage of wall socket.

 

 

 

 

 

 

* Conformity with commercial power voltage standard.

 

 

 

 

 

 

 

 

 

 

 

Tester

 

 

Power Cable

1

Connect the Power Cable to wall socket and verify voltage at inlet side.

 

 

 

* Conformity with commercial power voltage standard.

 

 

 

 

 

 

 

 

 

 

 

Tester

 

 

Body Inlet Plug

1

Confirm that the inlet plug is not bend or distorted.

 

 

 

 

 

 

 

 

Visual inspection

 

 

Fuse

1

Confirm that the fuse was not burned out.

 

 

 

 

 

 

 

 

 

 

 

 

 

Tester

 

 

Power Switch

1

Confirm that the Power Switch inside the equipment is pushed when

 

 

 

the Main Switch on the Front Panel is pushed.

 

 

 

 

 

 

 

Visual

Inspection

 

 

 

2

Confirm conductivity of the Power Switch when power

is ON.

 

 

 

 

 

 

 

 

 

 

 

Tester

 

 

Interlock Switch

1

Confirm that the switch is pushed by the Lamp Door.

 

 

 

 

 

 

 

Visual

Inspection

 

 

 

2

Confirm conductivity of the Interlock Switch when power is ON.

 

 

 

 

 

 

 

 

Tester

 

 

 

 

 

 

 

Interlock Jig

 

 

Flat Cable

1

Verify connection of the Flat Cable

 

 

 

 

 

 

 

Main Board (J12) <--> Front Panel

Visual

Inspection

 

 

 

 

 

 

 

 

2-3

ISSUE1

 

 

 

Troubleshooting

CLV-180

 

 

 

 

 

Main Board

1

Confirm that the input voltage to Main Board J16 is correct.

 

Converter

 

J16 (Main Board) <- -> Converter)

Tester

 

 

 

 

 

 

 

 

 

1pin : DC +15V±0.75V

 

 

 

2pin : DC +12V±0.6V

 

 

 

3pin : DC + 5V±0.25V

 

 

 

4pin : GND

 

 

 

5pin : DC -15V±0.75V

 

 

 

YES Replace Main Board

 

 

 

NO Replace Converter

 

Power Switch LED

1

Confirm that the Power Switch LED is lit:

 

 

 

NO Replace the Power Switch LED

 

Front Panel Unit

1

Confirm that the Front Panel of LED is lit:

 

 

 

NO Replace the Front Panel

2-2 Lamp failure

Xenon Lamp of light source not lit 2-2-1 Block diagram

2-2-2 Estimated location of failure

 

No

Estimated failure location

 

Inspection method

 

 

 

 

 

 

 

 

 

Power input failure

 

Refer to 2-1

 

 

 

Emergency Lamp Malfunction

 

Refer to 2-16

 

 

 

Panel malfunction

 

Refer to 2-13

 

 

 

Thermo Switch /

1

Status of Thermo Switch when buzzer is sounding

 

 

 

Main Board

 

Thermo Switch

 

 

 

 

 

Less than 85

 

 

 

 

 

Conductive

 

 

 

 

 

85 or higher

 

 

 

 

 

Non-conductive

 

 

 

 

 

Within standard -> Replace Main Board

 

 

 

 

 

Out of standard -> Replace Thermo Switch

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ISSUE1

2-4

 

 

 

 

 

 

 

 

 

 

 

 

Troubleshooting

CLV-180

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Main Board

1

Confirm that the signals in the Main Board are correct:

 

 

 

Main Board J6/2Pin

 

 

 

 

 

 

 

 

 

HIGH : XENON Lamp lit

*HIGH : approx. 5 V

 

 

 

LOW : XENON Lamp not lit

 

 

 

 

 

 

 

 

 

H -> L when the Ignition Button is pushed ON to turn on Lamp

 

 

 

L -> H when the Ignition Button is pushed OFF to turn off Lamp

 

 

 

(Long push)

 

 

 

 

 

 

 

 

 

NO Replace Main Board

 

 

 

 

 

 

 

Surrounding the Lamp

1

Verify Lamp attachment.

 

 

 

 

 

 

 

 

 

Confirm that Lamp is attached properly.

 

 

 

Confirm that heat compound is applied properly.

 

 

 

Confirm that the area surrounding the Xenon lamp contacts is clean.

 

 

 

* Cleaning is required if the discoloration is found.

 

 

2

Check the area surrounding the Lamp House.

 

 

 

No unnecessary objects (e.g. metal chip)

 

 

 

No leakage of high voltage pulse

 

 

 

Check connection with Terminal F and Terminal R

 

 

 

Confirm there is enough distance between Round Terminal and Lamp

 

 

 

House

 

 

 

 

 

 

 

Xenon Lamp /

1

Clattering noise at ignition

 

 

 

 

 

 

 

Igniter

 

YES Replace Xenon Lamp

 

 

 

NO Igniter

 

 

 

 

 

 

 

Converter

1

Xenon Lamp not lit even after procedures 1-7

 

 

 

YES Replace Converter

 

 

 

 

 

 

2-3 Lamp malfunction

 

 

 

 

 

 

 

 

Lamp going out

 

 

 

 

 

 

 

 

2-3-1 Estimated location of failure

 

 

 

 

 

 

 

 

No

Estimated location failure

 

Inspection method

 

 

 

 

 

 

Log

1

Verify frequency of Lamp failure errors.

 

 

 

* Refer to EVIS communication Checker for verification method.

 

 

 

 

 

 

Evis communication checker

 

 

 

 

 

 

CLV-180 Communication Cable

 

 

Power input failure

 

Refer to 2-1

 

 

 

 

 

 

 

Emergency Lamp malfunction

 

Refer to 2-16

 

 

 

 

 

 

 

Cooling Malfunction

 

Refer to 2-6

 

 

 

 

 

 

 

Lamp failure

 

Refer to 2-2

 

 

 

 

 

 

2-4 Scope connection failure

 

 

 

 

 

 

 

 

Scope not mounting

 

 

 

 

 

 

 

 

2-4-1 Estimated location of failure

 

 

 

 

 

 

 

 

No

Estimated location failure

 

Inspection Metod

 

 

 

 

 

Scope

1

Confirm that Scope is a compatible one.

 

Scope Socket

1

Confirm that the Scope Socket is free of abnormalities.

 

 

 

* Insert Heat Cover Positioning Jig to Scope Socket and confirm that

 

 

 

it's not caught inside.

 

 

 

 

 

 

 

 

 

 

 

 

 

Heat Cover Positioning Jig

 

2-5

ISSUE1

Troubleshooting CLV-180

2-5 High-brightness malfunction

Not changing to high-brightness function mode 2-5-1 Block diagram

2-5-2 Estimated location of malfunction

No

Estimated location failure

 

Inspection Method

 

 

 

 

 

 

 

High-brightness scope

1

Confirm that scope is compatible for high-brightness.

 

Panel malfunction

 

Refer to 2-13

 

 

Scope Sensor /

1

Confirm following status of J15 connector on Main Board when

 

Main Board

 

high-brightness scope is connected:

 

 

 

Main Board J15

Tester

 

 

 

 

 

 

 

 

 

When high-brightness scope is connected

 

 

 

Pin 2: HIGH

 

 

 

Pin 5: LOW

 

 

 

Pin 8: LOW

 

 

 

Pin 9: GND

 

 

 

* HIGH: approx. 5 V

 

 

 

YES Replace Main Board

 

 

 

NO Replace Scope Sensor

2-6

ISSUE1

Loading...
+ 56 hidden pages