Metrohm 900 User Manual

5 (2)
Metrohm 900 User Manual

900 Touch Control

Manual

8.900.8015EN / 2014-03-10

Metrohm AG CH-9100 Herisau Switzerland

Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com

900 Touch Control

Program version 5.900.0030

Manual

8.900.8015EN / 2014-03-10

ebe

Teachware Metrohm AG CH-9100 Herisau

teachware@metrohm.com

This documentation is protected by copyright. All rights reserved.

Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above.

Documentation in additional languages can be found on

http://documents.metrohm.com.

Melody for the BEEP command: excerpt from "En Altfrentsche", with kind permission of the Laseyer Quartett, Appenzell.

 

Table of contents

Table of contents

1 Introduction

1

1.1

Instrument description .........................................................

1

1.2

Titration and measuring modes ..........................................

1

1.3

About the documentation ...................................................

3

1.3.1

Symbols and conventions ........................................................

3

2 Safety instructions

5

2.1

General notes on safety .......................................................

5

2.2

Electrical safety .....................................................................

5

2.3

Tubing and capillary connections .......................................

6

2.4

Flammable solvents and chemicals .....................................

6

2.5

Recycling and disposal .........................................................

7

3

Overview of the instrument

8

4

Installation

 

10

 

4.1 Setting up the instrument ..................................................

10

 

4.1.1

Packaging ..............................................................................

10

 

4.1.2

Checks ..................................................................................

10

 

4.1.3

Location ................................................................................

10

5 Titrations

 

11

5.1

Dynamic equivalence point titration (DET) .......................

11

5.2

Monotonic equivalence point titration (MET) ..................

11

5.3

Endpoint titration (SET) .....................................................

12

5.4

Water determination according to Karl Fischer (KFT) .....

12

5.5

Endpoint titration with constant maintenance of the

 

 

measured value (STAT) ......................................................

13

6 Operation

 

15

6.1

Switching the instrument on and off ...............................

15

6.2

Fundamentals of operation ...............................................

17

6.2.1

Touch-sensitive screen ...........................................................

17

6.2.2

Display elements and controls ................................................

17

6.2.3

Status display ........................................................................

18

6.2.4

Entering text and numbers .....................................................

19

900 Touch Control

III

Table of contents

7 System settings

22

7.1

General system settings .....................................................

22

7.1.1

Selecting the dialog language ................................................

22

7.1.2

Setting the date, time and local time .....................................

23

7.2

System-specific dialog options ..........................................

24

7.3

User administration ............................................................

28

7.3.1

Editing the user configuration ................................................

29

7.3.2

Creating an identification profile ............................................

33

7.3.3

Defining login options ...........................................................

33

7.3.4

Password options ..................................................................

35

7.3.5

Modification options .............................................................

37

7.3.6

Reasons .................................................................................

38

7.3.7

Audit Trail ..............................................................................

39

7.4

Measured value display .....................................................

40

7.5

Acoustic signals ..................................................................

40

8 Titrants

 

41

8.1

Adding a new titrant ..........................................................

42

8.2

Editing titrant data .............................................................

43

8.3

Monitoring the working life ..............................................

45

8.4

Dosing unit ..........................................................................

46

8.4.1

Parameters for preparing (PREP) and emptying (EMPTY) .........

47

8.4.2

Tubing parameters .................................................................

48

8.4.3

Shift direction of the valve disk ..............................................

51

8.5

Exchange unit .....................................................................

52

8.5.1

Parameters for the preparation (PREP) ....................................

53

8.5.2

Tubing parameters .................................................................

54

8.6

GLP test for exchange unit and dosing unit ....................

56

8.7

Titer determination options and data ..............................

58

8.7.1

Titer validity ...........................................................................

58

8.7.2

Properties of the previous titer determinations .......................

59

9 Reagents

 

61

9.1

Editing reagent data ..........................................................

62

9.2

Reagent monitoring ...........................................................

62

10 Sensors

 

66

10.1

Adding a new sensor .........................................................

67

10.2

Editing the sensor data ......................................................

68

10.3

Monitoring the working life ..............................................

69

IV

900 Touch Control

 

 

Table of contents

10.4

Calibration data (for pHand ISE electrodes and con-

 

ductivity measuring cells only) ..........................................

70

10.4.1 Properties of the previous calibrations ....................................

72

10.5

Limit values for the calibration data .................................

75

10.6

Monitoring the calibration interval (only for pH and

 

ion-selective electrodes and conductivity measuring

 

cells) .....................................................................................

77

11 Device manager

79

11.1

Adding a new device ..........................................................

80

11.2

Configuring the instrument ...............................................

80

11.3

Touch Control .....................................................................

81

11.3.1

E-mail ....................................................................................

82

11.3.2

PC/LIMS report ......................................................................

83

11.3.3

Shared memory .....................................................................

84

11.3.4

TCP/IP settings .......................................................................

86

11.4

Metrohm control devices ...................................................

87

11.4.1 Properties – Control device ....................................................

88

11.4.2 Properties – Measuring input .................................................

88

11.4.3 Properties – MSB connector ...................................................

90

11.4.4 Properties – Peripheral devices ...............................................

91

11.5

USB Sample Processor .......................................................

91

11.5.1 Properties – Sample Processor ...............................................

92

11.5.2

Properties – Tower ................................................................

93

11.5.3 Properties – Swing Head ........................................................

94

11.6

Sample racks .......................................................................

99

11.6.1

Editing rack data ..................................................................

101

11.6.2

Rack adjustment ..................................................................

107

11.7

Printer ................................................................................

108

11.7.1

PDF settings .........................................................................

109

11.7.2

Network printer ...................................................................

110

11.7.3

More options .......................................................................

112

11.8

Balance ..............................................................................

112

11.9

USB/RS-232 adapter .........................................................

114

11.10

PC keyboard ......................................................................

115

11.11

Barcode reader .................................................................

117

12 File manager

119

12.1

Managing files ..................................................................

119

12.1.1

Copying a file ......................................................................

121

12.1.2

Renaming a file ...................................................................

121

12.1.3

File properties ......................................................................

122

12.2

External storage medium ................................................

123

900 Touch Control

V

Table of contents

 

12.3

Creating backups / Restoring data .................................

125

12.3.1

Restoring data .....................................................................

125

13 GLP manager

127

13.1

Automatic system test .....................................................

128

13.2

Test tools ..........................................................................

128

13.3

GLP tests for measurement and titration ......................

129

13.3.1

Parameter description ..........................................................

130

13.4

System validation .............................................................

133

13.4.1

Parameter description ..........................................................

134

13.5

System monitoring ...........................................................

137

13.5.1

Service interval ....................................................................

137

13.5.2

Backup interval ....................................................................

138

14 Common variables

139

14.1

Editing common variables ...............................................

140

14.2

Properties of common variables .....................................

141

14.3

Monitoring validity ...........................................................

142

14.4

Assigning a result automatically to a common varia-

 

 

ble ......................................................................................

143

15 Templates

 

145

15.1

Sample data ......................................................................

145

15.1.1

Sample identification list ......................................................

146

15.1.2

Sample assignment table .....................................................

147

15.2

Custom result templates ..................................................

149

15.2.1

Editing result templates .......................................................

150

15.3

Input lines .........................................................................

153

15.3.1 Editing the input signal ........................................................

154

15.4

Output lines ......................................................................

155

15.4.1 Editing the output signal ......................................................

157

15.5

Custom calibration buffers ..............................................

158

15.5.1

Defining calibration buffers ..................................................

159

15.6

Report header ...................................................................

160

15.7

Custom electrode type .....................................................

161

15.7.1 Limit values for the electrode rating .....................................

162

16 Methods

 

165

16.1

Creating a new method ...................................................

165

16.2

Saving a method ...............................................................

166

16.3

Loading a method ............................................................

167

VI

900 Touch Control

 

 

Table of contents

 

16.4

Editing a method ..............................................................

168

 

16.4.1

Inserting a command ...........................................................

169

 

16.5

Method options ................................................................

170

 

16.5.1

Start options ........................................................................

171

 

16.5.2

Stop options ........................................................................

173

 

16.5.3

Sample data ........................................................................

174

 

16.5.4

Method properties ...............................................................

178

 

16.5.5

Note ....................................................................................

180

 

16.5.6 Saving a determination automatically ...................................

180

17

Electronic signatures

182

 

17.1

Signing methods/determinations electronically ............

182

 

17.2

Deleting electronic signatures .........................................

183

18

Control

 

184

19

Favorites

 

188

 

19.1

Creating favorites .............................................................

189

 

19.1.1

Editing favorites ...................................................................

189

20

Sample data

192

 

20.1

Entering sample data in the main dialog .......................

192

 

20.2

Requesting sample data at the start of the determina-

 

 

tion ....................................................................................

193

21

Sample table

195

 

21.1

General ..............................................................................

195

 

21.2

Edit the sample data ........................................................

198

 

21.3

Properties ..........................................................................

200

22

Determination run

203

 

22.1

Carrying out a single determination ...............................

203

 

22.2

Performing a sample series .............................................

204

 

22.3

Canceling determinations manually ...............................

205

23

Live modifications

206

23.1Editing the sample data of the running determina-

tion ....................................................................................

206

23.2Editing the sample table while a determination is run-

 

ning ....................................................................................

207

23.3

Live display .......................................................................

208

23.4

Live parameters ................................................................

210

900 Touch Control

VII

Table of contents

 

24 Results and more determination data

212

24.1

More determination data ................................................

213

24.1.1

Details .................................................................................

214

24.2

Messages ...........................................................................

216

24.3

Local common variables ..................................................

217

24.4

Determination properties ................................................

217

24.5

Loading a determination .................................................

220

24.5.1

Determination list ................................................................

221

24.6

Saving a determination ....................................................

225

24.7

Curves ................................................................................

225

24.8

Recalculation and reevaluation .......................................

227

25 Statistics

 

229

25.1

Displaying details for a result ..........................................

231

25.2

Deleting statistical data ...................................................

232

25.3

Adding a determination to a determination series .......

232

26 Result table

 

234

26.1

Properties ..........................................................................

235

26.2

Saving the result table .....................................................

239

26.3

Loading the result table ...................................................

239

27 Printing

 

240

27.1

General report options .....................................................

242

27.2

Settings of the individual reports ...................................

243

27.3

List of all printable reports ..............................................

244

28 Manual control

248

28.1

Opening and closing the manual control .......................

249

28.2

Measuring .........................................................................

250

28.2.1

Parameter description ..........................................................

251

28.3

Dosing ...............................................................................

254

28.3.1

Continuous dosing ..............................................................

256

28.3.2

Dosing fixed volumes ...........................................................

258

28.3.3

Preparing .............................................................................

260

28.3.4

Emptying .............................................................................

261

28.3.5

Filling ..................................................................................

262

28.3.6

Replacing reagent ................................................................

262

28.4

Stirring ...............................................................................

263

VIII

900 Touch Control

 

 

 

Table of contents

28.5 Conditioning a titration cell ............................................

264

28.5.1

 

Parameter description ..........................................................

265

28.6

Remote ..............................................................................

265

28.7

USB Sample Processor .....................................................

266

28.7.1

 

Moving the lift .....................................................................

268

28.7.2

Moving to a rack position ....................................................

270

28.7.3

 

External positions ................................................................

272

29 Parameters

 

 

276

29.1 Dynamic equivalence point titrations (DET) ...................

276

29.1.1

 

Start conditions ...................................................................

276

29.1.2

 

Titration parameters ............................................................

279

29.1.3

 

Stop conditions ...................................................................

283

29.1.4

 

Potentiometric evaluation ....................................................

284

29.1.5

 

Control device .....................................................................

289

29.1.6

 

Sensor .................................................................................

290

29.1.7

 

Dosing device ......................................................................

291

29.1.8

 

Stirrer ..................................................................................

292

29.2 Monotonic equivalence point titrations (MET) ..............

293

29.2.1

 

Start conditions ...................................................................

293

29.2.2

 

Titration parameters ............................................................

296

29.2.3

 

Stop conditions ...................................................................

299

29.2.4

 

Potentiometric evaluation ....................................................

300

29.2.5

 

Control device .....................................................................

305

29.2.6

 

Sensor .................................................................................

306

29.2.7

 

Dosing device ......................................................................

307

29.2.8

 

Stirrer ..................................................................................

308

29.3

Endpoint titrations (SET) ..................................................

309

29.3.1

 

Start conditions ...................................................................

309

29.3.2

 

Control parameters ..............................................................

310

29.3.3

 

Titration parameters ............................................................

314

29.3.4

 

Stop conditions ...................................................................

315

29.3.5

 

Conditioning .......................................................................

316

29.3.6

 

Cell ......................................................................................

318

29.3.7

 

Control device .....................................................................

318

29.3.8

 

Sensor .................................................................................

318

29.3.9

 

Dosing device ......................................................................

320

29.3.10

Stirrer ..................................................................................

321

29.4 Volumetric Karl Fischer titrations (KFT) .........................

321

29.4.1

 

Start conditions ...................................................................

321

29.4.2

 

Control parameters ..............................................................

323

29.4.3

 

Titration parameters ............................................................

326

29.4.4

 

Stop conditions ...................................................................

327

29.4.5

 

Conditioning .......................................................................

328

29.4.6

 

Cell ......................................................................................

330

29.4.7

 

Control device .....................................................................

330

29.4.8

 

Sensor .................................................................................

331

900 Touch Control

IX

Table of contents

 

 

 

29.4.9

 

Dosing device ......................................................................

 

332

29.4.10

Stirrer ..................................................................................

 

333

29.5 Coulometric Karl Fischer titrations (KFC) .......................

334

29.5.1

 

Control parameters ..............................................................

 

334

29.5.2

 

Titration parameters ............................................................

 

336

29.5.3

 

Stop conditions ...................................................................

 

337

29.5.4

 

Conditioning .......................................................................

 

337

29.5.5

 

Cell ......................................................................................

 

340

29.5.6

 

Control device .....................................................................

 

341

29.5.7

 

Sensor .................................................................................

 

341

29.5.8

 

Stirrer ..................................................................................

 

342

29.6 Bromine index determination (BRC) ...............................

 

343

29.6.1

 

Control parameters ..............................................................

 

343

29.6.2

 

Titration parameters ............................................................

 

346

29.6.3

 

Stop conditions ...................................................................

 

347

29.6.4

 

Conditioning .......................................................................

 

347

29.6.5

 

Cell ......................................................................................

 

349

29.6.6

 

Control device .....................................................................

 

350

29.6.7

 

Sensor .................................................................................

 

350

29.6.8

 

Stirrer ..................................................................................

 

351

29.7

STAT titrations (STAT) .....................................................

 

352

29.7.1

 

Start conditions ...................................................................

 

352

29.7.2

 

Control parameters ..............................................................

 

354

29.7.3

 

Titration parameters ............................................................

 

356

29.7.4

 

Stop conditions ...................................................................

 

358

29.7.5

 

Monitoring ..........................................................................

 

359

29.7.6

 

Control device .....................................................................

 

366

29.7.7

 

Sensor .................................................................................

 

366

29.7.8

 

Dosing device ......................................................................

 

367

29.7.9

 

Stirrer ..................................................................................

 

370

29.8

Manual titrations (MAT) ..................................................

 

371

29.8.1

 

Dosing parameters ..............................................................

 

371

29.8.2

 

Control device .....................................................................

 

372

29.8.3

 

Sensor .................................................................................

 

372

29.8.4

 

Dosing device ......................................................................

 

373

29.8.5

 

Stirrer ..................................................................................

 

376

29.9

Measurements (MEAS) .....................................................

 

377

29.9.1

 

Measuring parameters .........................................................

 

377

29.9.2

 

Standard addition ................................................................

 

381

29.9.3

 

Control device .....................................................................

 

385

29.9.4

 

Sensor .................................................................................

 

385

29.9.5

 

Stirrer ..................................................................................

 

387

29.10

Calibration of sensors (CAL) ............................................

 

388

29.10.1

Calibration parameters (CAL pH) ..........................................

 

388

29.10.2

Calibration parameters (CAL Conc) ......................................

 

391

29.10.3

Calibration parameters (CAL Cond) ......................................

 

394

29.10.4

Control device .....................................................................

 

397

X

900 Touch Control

 

Table of contents

29.10.5

Sensor .................................................................................

397

29.10.6

Stirrer ..................................................................................

398

29.11 Electrode test for pH electrodes (ELT) ............................

399

29.11.1

Electrode test parameters ....................................................

400

29.11.2

Control device .....................................................................

401

29.11.3

Sensor .................................................................................

402

29.11.4

Stirrer ..................................................................................

403

29.11.5

Further information .............................................................

403

29.12 Evaluations (EVAL) ...........................................................

405

29.12.1 Fixed endpoint evaluation (EVAL FIX-EP) ..............................

407

29.12.2 pK value and half neutralization potential evaluation (EVAL

 

 

pK/HNP) ..............................................................................

409

29.12.3 Minimum and maximum evaluation (EVAL MIN/MAX) ..........

410

29.12.4 Break point evaluation (EVAL BREAK) ...................................

411

29.12.5 Rate evaluation (EVAL RATE) ................................................

414

29.13 Calculations .......................................................................

415

29.13.1

Calculations (CALC) .............................................................

415

29.13.2

Calculations (CALC LIVE) ......................................................

422

29.13.3

Formula editor .....................................................................

424

29.14 Reports (REPORT) .............................................................

430

29.14.1

General report options .........................................................

430

29.14.2 Settings of the individual reports ..........................................

431

29.14.3

List of reports ......................................................................

432

29.15 Dosing and Liquid Handling ............................................

435

29.15.1 Preparing an exchange or dosing unit (PREP) ........................

435

29.15.2 Emptying a dosing unit (EMPTY) ..........................................

436

29.15.3 Dosing a specified volume (ADD) .........................................

437

29.15.4

Liquid Handling (LQH) ..........................................................

442

29.15.5

Monitored dosing (DOS) ......................................................

445

29.16 Communication ................................................................

462

29.16.1 Scanning remote lines (SCAN) ..............................................

463

29.16.2 Setting remote lines (CTRL) ..................................................

464

29.16.3 Scanning the RS-232 interface (SCAN RS) .............................

466

29.16.4 Defining RS-232 commands (CONTROL RS) ..........................

467

29.17 Automation .......................................................................

467

29.17.1 Rotating sample rack (MOVE) ..............................................

467

29.17.2 Moving the lift (LIFT) ............................................................

470

29.17.3

Controlling pumps (PUMP) ...................................................

471

29.17.4 Resetting the rack (RACK) ....................................................

472

29.17.5 Defining the sample variable (SAMPLE) ................................

472

29.17.6 Creating a subsequence (SUBSEQ) .......................................

473

29.18 Miscellaneous commands ................................................

476

29.18.1 Controlling a stirrer (STIR) ....................................................

476

29.18.2 Pausing the method run (WAIT) ...........................................

477

29.18.3

Scan data (REQUEST) ...........................................................

477

29.18.4 Defining an acoustic signal (BEEP) ........................................

478

900 Touch Control

XI

Table of contents

 

 

29.18.5

Signing a determination (SIGN) ............................................

479

29.18.6

Canceling the method run (END) .........................................

479

30 Operation and maintenance

480

30.1

System initialization .........................................................

480

30.2

Quality management and qualification with

 

 

Metrohm ...........................................................................

481

31 Troubleshooting

482

31.1

Editing methods ...............................................................

482

31.2

Sample table .....................................................................

482

31.3

Results/Statistics ...............................................................

482

31.4

Printing ..............................................................................

483

31.5

Manual control .................................................................

483

31.6

Miscellaneous ...................................................................

484

31.7

SET titration ......................................................................

484

31.8

Karl Fischer titration ........................................................

485

31.9

STAT titration ...................................................................

487

32 Appendix

 

 

488

32.1

Dosing unit ........................................................................

488

32.1.1

Maximum dosing and filling rate ..........................................

488

32.1.2

Default parameters for preparing (PREP) and emptying

 

 

 

(EMPTY) ...............................................................................

488

32.2

Exchange unit ...................................................................

489

32.2.1

Maximum dosing and filling rate ..........................................

489

32.2.2

Default parameters for preparing (PREP) ..............................

489

32.3

Stirring rate .......................................................................

489

32.4

Balance ..............................................................................

490

32.5

Result variables as parameter setting ............................

491

32.6

Electrode calibration with USB Sample Processors .......

492

32.7

Stored buffer series for pH calibration ...........................

494

32.8

Using AuditTrailViewer ....................................................

498

32.8.1

 

Installing AuditTrailViewer ...................................................

498

32.8.2

Opening the Audit Trail .......................................................

499

32.8.3

Contents of the Audit Trail ...................................................

500

32.8.4

Filtering the Audit Trail .........................................................

502

32.8.5

Exporting the Audit Trail ......................................................

502

32.9

Diagnosis ...........................................................................

503

32.9.1

 

LCD test ..............................................................................

503

32.9.2

 

Temperature monitoring ......................................................

504

XII

900 Touch Control

 

 

Table of contents

32.9.3 Formatting an external storage medium ...............................

504

32.9.4 Removing an external storage medium ................................

505

32.9.5 Adjusting the touch screen ..................................................

505

32.9.6 Testing the touch screen ......................................................

506

32.9.7 Software update (loading program versions and language

 

files) ....................................................................................

507

32.9.8

Simulating titration curves ...................................................

510

32.9.9

Service .................................................................................

511

32.10

Arithmetic algorithms in the Titrando ............................

512

33 Technical specifications

514

33.1

Touch screen .....................................................................

514

33.2

Interfaces ..........................................................................

515

33.3

Power supply ....................................................................

515

33.4

Safety specifications ........................................................

515

33.5

Electromagnetic compatibility (EMC) .............................

516

33.6

Ambient temperature ......................................................

516

33.7

Dimensions ........................................................................

516

34 Warranty (guarantee)

517

35 Accessories

 

519

Index

 

521

900 Touch Control

XIII

Table of figures

 

Table of figures

 

Figure 1

Front 900 Touch Control ...................................................................

8

Figure 2

Rear 900 Touch Control ....................................................................

9

Figure 3

Reagent dosing for DET ...................................................................

11

Figure 4

Reagent dosing for MET ..................................................................

11

Figure 5

Reagent dosing for SET ...................................................................

12

Figure 6

Reagent dosing for KFT ...................................................................

13

Figure 7

Reagent dosing for STAT .................................................................

14

Figure 8

Dosing unit – port assignment .........................................................

51

Figure 9

Exchange unit – tubing connections ................................................

56

Figure 10

Directory structure on the external storage medium ......................

124

Figure 11

Live display "Preparing the dosing unit" .........................................

261

Figure 12

Live display "Preparing the exchange unit" .....................................

261

Figure 13

Equivalence point recognition and equivalence point numbering within

 

windows .......................................................................................

288

Figure 14

Tubbs method for determining the equivalence point ....................

289

Figure 15

Equivalence point recognition and equivalence point numbering within

 

windows .......................................................................................

304

Figure 16

Action "Exit method" or "Cancel command" ..................................

363

Figure 17

Action "Pause" ..............................................................................

363

Figure 18

Action "Wait" ................................................................................

364

Figure 19

Evaluation of pK value / half neutralization potential ......................

409

Figure 20

Evaluation of minimum and maximum ...........................................

410

Figure 21

Evaluation of a break point ............................................................

411

Figure 22

Dosing criterion "Volume/Dosing rate" ..........................................

446

Figure 23

Dosing criterion "Volume/Dosing time" ..........................................

446

Figure 24

Dosing criterion "Dosing rate/Dosing time" ....................................

447

Figure 25

Action "Exit method" or "Cancel command" ..................................

455

Figure 26

Action "Pause" ..............................................................................

455

Figure 27

Action "Wait" ................................................................................

455

Figure 28

Rotational speed depending on stirring rate ..................................

490

XIV

900 Touch Control

 

1 Introduction

1 Introduction

1.1Instrument description

The 900 Touch Control is an operating unit with a touch-sensitive screen. You can operate the following control devices with the 900 Touch Control:

Titrandos

USB Sample Processors

846 Dosing Interface

867 pH Module

856 Conductivity Module

With the 900 Touch Control, you can conveniently manage all of your titrants, sensors, methods, etc. For example, you can connect a USB flash drive to the USB connector as an external storage medium. You can use this additional memory not only to store your methods and determinations, but also to create a backup together with all of the data and settings of your system. Thanks to the integrated Ethernet connector, the device can be connected to your network, thus offering you the following options, among others:

Saving data to a PC within the network

Printing reports on a network printer

Sending displayed messages as e-mails

1.2Titration and measuring modes

The 900 Touch Control supports the following titration and measuring modes. Whether a particular mode is available or not depends on the type of the control device connected.

DET

Dynamic equivalence point titration. The reagent is added in variable volume steps.

Measuring modes:

pH (pH measurement)

U (potentiometric voltage measurement)

Ipol (voltametric measurement with selectable polarization current)

Upol (amperometric measurement with selectable polarization voltage)

900 Touch Control

1

1.2 Titration and measuring modes

 

MET

Monotonic equivalence point titration. The reagent is added in constant volume steps.

Measuring modes:

pH (pH measurement)

U (potentiometric voltage measurement)

Ipol (voltametric measurement with selectable polarization current)

Upol (amperometric measurement with selectable polarization voltage)

SET

Endpoint titration at one or two specified endpoints. Measuring modes:

pH (pH measurement)

U (potentiometric voltage measurement)

Ipol (voltametric measurement with selectable polarization current)

Upol (amperometric measurement with selectable polarization voltage)

STAT

Endpoint titration with constant maintenance of the measured value. Measuring modes:

pH (pH measurement)

U (potentiometric voltage measurement)

KFT

Volumetric water content determination according to Karl Fischer. Measuring modes:

Ipol (voltametric measurement with selectable polarization current)

Upol (amperometric measurement with selectable polarization voltage)

KFC

Coulometric water content determination according to Karl Fischer. Measuring mode:

Ipol (voltametric measurement with selectable polarization current)

BRC

Coulometric bromine index determination. Determining the amount of double bonds in e.g. mineral oils.

Measuring mode:

Ipol (voltametric measurement with selectable polarization current)

2

900 Touch Control

 

1 Introduction

MEAS

Measuring modes:

pH (pH measurement)

U (potentiometric voltage measurement)

Ipol (voltametric measurement with selectable polarization current)

Upol (amperometric measurement with selectable polarization voltage)

T (temperature measurement)

Conc (Concentration measurement with or without standard addition)

Cond (conductivity measurement)

CAL

Electrode calibration. Measuring mode:

pH (calibration of pH electrodes)

Conc (calibration of ion-selective electrodes)

Cond (calibration of conductivity measuring cells)

ELT (Electrode test for pH electrodes)

1.3About the documentation

CAUTION

Please read through this documentation carefully before putting the instrument into operation. The documentation contains information and warnings which the user must follow in order to ensure safe operation of the instrument.

1.3.1Symbols and conventions

The following symbols and formatting may appear in this documentation:

 

Cross-reference to figure legend

 

The first number refers to the figure number, the sec-

 

ond to the instrument part in the figure.

 

 

 

Instruction step

 

Carry out these steps in the sequence shown.

 

 

Method

Dialog text, parameter in the software

 

 

File New

Menu or menu item

 

 

[Next]

Button or key

 

 

900 Touch Control

3

1.3 About the documentation

 

WARNING

This symbol draws attention to a possible life-threat- ening hazard or risk of injury.

WARNING

This symbol draws attention to a possible hazard due to electrical current.

WARNING

This symbol draws attention to a possible hazard due to heat or hot instrument parts.

WARNING

This symbol draws attention to a possible biological hazard.

CAUTION

This symbol draws attention to possible damage to instruments or instrument parts.

NOTE

This symbol highlights additional information and tips.

4

900 Touch Control

 

2 Safety instructions

2 Safety instructions

2.1General notes on safety

WARNING

This instrument may only be operated in accordance with the specifications in this documentation.

This instrument has left the factory in a flawless state in terms of technical safety. To maintain this state and ensure non-hazardous operation of the instrument, the following instructions must be observed carefully.

2.2Electrical safety

The electrical safety when working with the instrument is ensured as part of the international standard IEC 61010.

WARNING

Only personnel qualified by Metrohm are authorized to carry out service work on electronic components.

WARNING

Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are touched.

There are no parts inside the housing which can be serviced or replaced by the user.

Mains voltage

WARNING

An incorrect mains voltage can damage the instrument.

Only operate this instrument with a mains voltage specified for it (see rear panel of the instrument).

900 Touch Control

5

2.3 Tubing and capillary connections

 

Protection against electrostatic charges

WARNING

Electronic components are sensitive to electrostatic charges and can be destroyed by discharges.

Do not fail to pull the mains cable out of the mains connection socket before you set up or disconnect electrical plug connections at the rear of the instrument.

2.3Tubing and capillary connections

CAUTION

Leaks in tubing and capillary connections are a safety risk. Tighten all connections well by hand. Avoid applying excessive force to tubing connections. Damaged tubing ends lead to leakage. Appropriate tools can be used to loosen connections.

Check the connections regularly for leakage. If the instrument is used mainly in unattended operation, then weekly inspections are mandatory.

2.4Flammable solvents and chemicals

WARNING

All relevant safety measures are to be observed when working with flammable solvents and chemicals.

Set up the instrument in a well-ventilated location (e.g. fume cupboard).

Keep all sources of flame far from the workplace.

Clean up spilled liquids and solids immediately.

Follow the safety instructions of the chemical manufacturer.

6

900 Touch Control

 

2 Safety instructions

2.5Recycling and disposal

This product is covered by European Directive 2002/96/EC, WEEE – Waste from Electrical and Electronic Equipment.

The correct disposal of your old equipment will help to prevent negative effects on the environment and public health.

More details about the disposal of your old equipment can be obtained from your local authorities, from waste disposal companies or from your local dealer.

900 Touch Control

7

3 Overview of the instrument

1

 

2

3

4

5

6

7

8

 

Figure 1 Front 900 Touch Control

 

 

 

 

 

 

 

 

 

1

Display

 

 

 

2

Fixed key [Home]

 

Touch screen.

 

 

 

 

Opens the main dialog.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3

Fixed key [Back]

 

 

 

4

Fixed key [Help]

 

Opens the next-higher dialog page.

 

 

 

 

Opens the online help for the dialog dis-

 

 

 

 

 

 

played.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5

Fixed key [Print]

 

 

 

6

Fixed key [Manual]

 

Opens the print dialog.

 

 

 

 

Opens the manual control.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

7

Fixed key [STOP]

 

 

 

8

Fixed key [START]

 

Cancels the running determination.

 

 

 

 

Starts a determination.

 

 

 

 

 

 

 

 

 

8

900 Touch Control

 

3 Overview of the instrument

1

2

3

4

5

Figure 2 Rear 900 Touch Control

1Ethernet connector (RJ-45)

For connecting to a network.

2USB connector (type A)

For connecting printers, USB sticks, USB hubs, etc.

3Power switch

Switches the instrument on/off.

4Type plate

Contains the serial number.

5Connection cable

For the connection of the Touch Control to a control device (socket "Controller").

900 Touch Control

9

4.1 Setting up the instrument

 

4 Installation

The connection between the Touch Control and the control device (Titrando, etc.) is described in the respective manual, as is the configuration of the titration system with its peripheral devices, e.g. stirrers and dosing devices.

4.1Setting up the instrument

4.1.1Packaging

The instrument is supplied in highly protective special packaging together with the separately packed accessories. Keep this packaging, as only this ensures safe transportation of the instrument.

4.1.2Checks

Immediately after receipt, check whether the shipment has arrived complete and without damage by comparing it with the delivery note.

4.1.3Location

The instrument has been developed for operation indoors and may not be used in explosive environments.

Place the instrument in a location of the laboratory which is suitable for operation, free of vibrations, protected from corrosive atmosphere, and contamination by chemicals.

The instrument should be protected against excessive temperature fluctuations and direct sunlight.

10

900 Touch Control

 

5 Titrations

5 Titrations

5.1Dynamic equivalence point titration (DET)

Dynamic equivalence point titration is a titration mode for all standard titrations. The reagent is added in variable volume steps. The volume increments vary as a function of the slope of the curve. An attempt is made to achieve constant measured value alterations with each dosing. The optimal volume for dosing is determined from the measured value alterations of the previous dosings. Measured value acceptance is driftcontrolled (equilibrium titration) or after a waiting time. Equivalence points are evaluated automatically.

U/mV

V/mL

Figure 3 Reagent dosing for DET

5.2Monotonic equivalence point titration (MET)

Monotonic equivalence point titration is a titration mode for titrations with relatively high signal fluctuations or suddenly occurring potential jumps and for slow titrations or slow-response electrodes. The reagent is added in constant volume steps. Measured value acceptance is drift-con- trolled (equilibrium titration) or after a waiting time. Equivalence points are evaluated automatically.

U/mV

V/mL

Figure 4 Reagent dosing for MET

900 Touch Control

11

5.3 Endpoint titration (SET)

 

5.3Endpoint titration (SET)

Endpoint titration is a titration mode for rapid routine determinations to a preset endpoint (e.g. titrations in accordance with special norms) and titrations for which reagent overflow must be avoided. The titration termination at the endpoint takes place either drift-controlled or after a waiting period. The volume dosed until the endpoint is used for calculating the content of the sample.

U/mV

EP

Control range

V/mL

t/s

Initial

Continuous

V/mL

dosing

dosing

 

Figure 5 Reagent dosing for SET

5.4Water determination according to Karl Fischer (KFT)

Karl Fischer Titration is a method for volumetric water determination. Conditioning is carried out automatically both before and after the actual titration. The reagent dosing is controlled in such a way that a predefined endpoint is reached as quickly and as accurately as possible. The volume steps and the rate of reagent dosing are regulated by the difference between the current measured value and the predefined endpoint. This means that titration is performed more slowly in the control range and that smaller volumes are added. The titration is stopped at the endpoint either drift-controlled or after a waiting time. The volume dosed until the endpoint is used to calculate the water content of the sample.

12

900 Touch Control

 

5 Titrations

U/mV

KFT Ipol

Control range

EP

V/mL

t/s

V/mL

Initial Continuous dosing dosing

I/µA

KFT Upol

EP

Control range

V/mL

Figure 6 Reagent dosing for KFT

5.5Endpoint titration with constant maintenance of the measured value (STAT)

The pH STAT titration is the one most frequently utilized. Reagent is added until a preset measured value (control point) has been reached. This control point is kept stable by continuously titrating off the substance set free by the reaction with the reagent. The STAT titration keeps the control point constant until the preset stop criterion has been reached. The STAT titration is applied, for example, in analysis of enzymes. The dosing rate resulting from the the control point being kept stable provides insight into the activity of an enzyme.

900 Touch Control

13

5.5 Endpoint titration with constant maintenance of the measured value (STAT)

U/mV

Control range

V/mL

Initial

Incremental dosing

dosing

 

Figure 7 Reagent dosing for STAT

Control point

V/mL

t/s

14

900 Touch Control

Loading...
+ 511 hidden pages