900 Touch Control
Manual
8.900.8015EN / 2014-03-10
Metrohm AG CH-9100 Herisau Switzerland
Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com
900 Touch Control
Program version 5.900.0030
Manual
8.900.8015EN / 2014-03-10 |
ebe |
Teachware Metrohm AG CH-9100 Herisau
teachware@metrohm.com
This documentation is protected by copyright. All rights reserved.
Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above.
Documentation in additional languages can be found on
http://documents.metrohm.com.
Melody for the BEEP command: excerpt from "En Altfrentsche", with kind permission of the Laseyer Quartett, Appenzell.
|
Table of contents |
Table of contents
1 Introduction |
1 |
|
1.1 |
Instrument description ......................................................... |
1 |
1.2 |
Titration and measuring modes .......................................... |
1 |
1.3 |
About the documentation ................................................... |
3 |
1.3.1 |
Symbols and conventions ........................................................ |
3 |
2 Safety instructions |
5 |
|
2.1 |
General notes on safety ....................................................... |
5 |
2.2 |
Electrical safety ..................................................................... |
5 |
2.3 |
Tubing and capillary connections ....................................... |
6 |
2.4 |
Flammable solvents and chemicals ..................................... |
6 |
2.5 |
Recycling and disposal ......................................................... |
7 |
3 |
Overview of the instrument |
8 |
|
4 |
Installation |
|
10 |
|
4.1 Setting up the instrument .................................................. |
10 |
|
|
4.1.1 |
Packaging .............................................................................. |
10 |
|
4.1.2 |
Checks .................................................................................. |
10 |
|
4.1.3 |
Location ................................................................................ |
10 |
5 Titrations |
|
11 |
5.1 |
Dynamic equivalence point titration (DET) ....................... |
11 |
5.2 |
Monotonic equivalence point titration (MET) .................. |
11 |
5.3 |
Endpoint titration (SET) ..................................................... |
12 |
5.4 |
Water determination according to Karl Fischer (KFT) ..... |
12 |
5.5 |
Endpoint titration with constant maintenance of the |
|
|
measured value (STAT) ...................................................... |
13 |
6 Operation |
|
15 |
6.1 |
Switching the instrument on and off ............................... |
15 |
6.2 |
Fundamentals of operation ............................................... |
17 |
6.2.1 |
Touch-sensitive screen ........................................................... |
17 |
6.2.2 |
Display elements and controls ................................................ |
17 |
6.2.3 |
Status display ........................................................................ |
18 |
6.2.4 |
Entering text and numbers ..................................................... |
19 |
900 Touch Control |
III |
Table of contents
7 System settings |
22 |
|
7.1 |
General system settings ..................................................... |
22 |
7.1.1 |
Selecting the dialog language ................................................ |
22 |
7.1.2 |
Setting the date, time and local time ..................................... |
23 |
7.2 |
System-specific dialog options .......................................... |
24 |
7.3 |
User administration ............................................................ |
28 |
7.3.1 |
Editing the user configuration ................................................ |
29 |
7.3.2 |
Creating an identification profile ............................................ |
33 |
7.3.3 |
Defining login options ........................................................... |
33 |
7.3.4 |
Password options .................................................................. |
35 |
7.3.5 |
Modification options ............................................................. |
37 |
7.3.6 |
Reasons ................................................................................. |
38 |
7.3.7 |
Audit Trail .............................................................................. |
39 |
7.4 |
Measured value display ..................................................... |
40 |
7.5 |
Acoustic signals .................................................................. |
40 |
8 Titrants |
|
41 |
8.1 |
Adding a new titrant .......................................................... |
42 |
8.2 |
Editing titrant data ............................................................. |
43 |
8.3 |
Monitoring the working life .............................................. |
45 |
8.4 |
Dosing unit .......................................................................... |
46 |
8.4.1 |
Parameters for preparing (PREP) and emptying (EMPTY) ......... |
47 |
8.4.2 |
Tubing parameters ................................................................. |
48 |
8.4.3 |
Shift direction of the valve disk .............................................. |
51 |
8.5 |
Exchange unit ..................................................................... |
52 |
8.5.1 |
Parameters for the preparation (PREP) .................................... |
53 |
8.5.2 |
Tubing parameters ................................................................. |
54 |
8.6 |
GLP test for exchange unit and dosing unit .................... |
56 |
8.7 |
Titer determination options and data .............................. |
58 |
8.7.1 |
Titer validity ........................................................................... |
58 |
8.7.2 |
Properties of the previous titer determinations ....................... |
59 |
9 Reagents |
|
61 |
9.1 |
Editing reagent data .......................................................... |
62 |
9.2 |
Reagent monitoring ........................................................... |
62 |
10 Sensors |
|
66 |
10.1 |
Adding a new sensor ......................................................... |
67 |
10.2 |
Editing the sensor data ...................................................... |
68 |
10.3 |
Monitoring the working life .............................................. |
69 |
IV |
900 Touch Control |
|
|
Table of contents |
10.4 |
Calibration data (for pHand ISE electrodes and con- |
|
|
ductivity measuring cells only) .......................................... |
70 |
10.4.1 Properties of the previous calibrations .................................... |
72 |
|
10.5 |
Limit values for the calibration data ................................. |
75 |
10.6 |
Monitoring the calibration interval (only for pH and |
|
|
ion-selective electrodes and conductivity measuring |
|
|
cells) ..................................................................................... |
77 |
11 Device manager |
79 |
|
11.1 |
Adding a new device .......................................................... |
80 |
11.2 |
Configuring the instrument ............................................... |
80 |
11.3 |
Touch Control ..................................................................... |
81 |
11.3.1 |
E-mail .................................................................................... |
82 |
11.3.2 |
PC/LIMS report ...................................................................... |
83 |
11.3.3 |
Shared memory ..................................................................... |
84 |
11.3.4 |
TCP/IP settings ....................................................................... |
86 |
11.4 |
Metrohm control devices ................................................... |
87 |
11.4.1 Properties – Control device .................................................... |
88 |
|
11.4.2 Properties – Measuring input ................................................. |
88 |
|
11.4.3 Properties – MSB connector ................................................... |
90 |
|
11.4.4 Properties – Peripheral devices ............................................... |
91 |
|
11.5 |
USB Sample Processor ....................................................... |
91 |
11.5.1 Properties – Sample Processor ............................................... |
92 |
|
11.5.2 |
Properties – Tower ................................................................ |
93 |
11.5.3 Properties – Swing Head ........................................................ |
94 |
|
11.6 |
Sample racks ....................................................................... |
99 |
11.6.1 |
Editing rack data .................................................................. |
101 |
11.6.2 |
Rack adjustment .................................................................. |
107 |
11.7 |
Printer ................................................................................ |
108 |
11.7.1 |
PDF settings ......................................................................... |
109 |
11.7.2 |
Network printer ................................................................... |
110 |
11.7.3 |
More options ....................................................................... |
112 |
11.8 |
Balance .............................................................................. |
112 |
11.9 |
USB/RS-232 adapter ......................................................... |
114 |
11.10 |
PC keyboard ...................................................................... |
115 |
11.11 |
Barcode reader ................................................................. |
117 |
12 File manager |
119 |
|
12.1 |
Managing files .................................................................. |
119 |
12.1.1 |
Copying a file ...................................................................... |
121 |
12.1.2 |
Renaming a file ................................................................... |
121 |
12.1.3 |
File properties ...................................................................... |
122 |
12.2 |
External storage medium ................................................ |
123 |
900 Touch Control |
V |
Table of contents |
|
|
12.3 |
Creating backups / Restoring data ................................. |
125 |
12.3.1 |
Restoring data ..................................................................... |
125 |
13 GLP manager |
127 |
|
13.1 |
Automatic system test ..................................................... |
128 |
13.2 |
Test tools .......................................................................... |
128 |
13.3 |
GLP tests for measurement and titration ...................... |
129 |
13.3.1 |
Parameter description .......................................................... |
130 |
13.4 |
System validation ............................................................. |
133 |
13.4.1 |
Parameter description .......................................................... |
134 |
13.5 |
System monitoring ........................................................... |
137 |
13.5.1 |
Service interval .................................................................... |
137 |
13.5.2 |
Backup interval .................................................................... |
138 |
14 Common variables |
139 |
|
14.1 |
Editing common variables ............................................... |
140 |
14.2 |
Properties of common variables ..................................... |
141 |
14.3 |
Monitoring validity ........................................................... |
142 |
14.4 |
Assigning a result automatically to a common varia- |
|
|
ble ...................................................................................... |
143 |
15 Templates |
|
145 |
15.1 |
Sample data ...................................................................... |
145 |
15.1.1 |
Sample identification list ...................................................... |
146 |
15.1.2 |
Sample assignment table ..................................................... |
147 |
15.2 |
Custom result templates .................................................. |
149 |
15.2.1 |
Editing result templates ....................................................... |
150 |
15.3 |
Input lines ......................................................................... |
153 |
15.3.1 Editing the input signal ........................................................ |
154 |
|
15.4 |
Output lines ...................................................................... |
155 |
15.4.1 Editing the output signal ...................................................... |
157 |
|
15.5 |
Custom calibration buffers .............................................. |
158 |
15.5.1 |
Defining calibration buffers .................................................. |
159 |
15.6 |
Report header ................................................................... |
160 |
15.7 |
Custom electrode type ..................................................... |
161 |
15.7.1 Limit values for the electrode rating ..................................... |
162 |
|
16 Methods |
|
165 |
16.1 |
Creating a new method ................................................... |
165 |
16.2 |
Saving a method ............................................................... |
166 |
16.3 |
Loading a method ............................................................ |
167 |
VI |
900 Touch Control |
|
|
Table of contents |
|
|
16.4 |
Editing a method .............................................................. |
168 |
|
16.4.1 |
Inserting a command ........................................................... |
169 |
|
16.5 |
Method options ................................................................ |
170 |
|
16.5.1 |
Start options ........................................................................ |
171 |
|
16.5.2 |
Stop options ........................................................................ |
173 |
|
16.5.3 |
Sample data ........................................................................ |
174 |
|
16.5.4 |
Method properties ............................................................... |
178 |
|
16.5.5 |
Note .................................................................................... |
180 |
|
16.5.6 Saving a determination automatically ................................... |
180 |
|
17 |
Electronic signatures |
182 |
|
|
17.1 |
Signing methods/determinations electronically ............ |
182 |
|
17.2 |
Deleting electronic signatures ......................................... |
183 |
18 |
Control |
|
184 |
19 |
Favorites |
|
188 |
|
19.1 |
Creating favorites ............................................................. |
189 |
|
19.1.1 |
Editing favorites ................................................................... |
189 |
20 |
Sample data |
192 |
|
|
20.1 |
Entering sample data in the main dialog ....................... |
192 |
|
20.2 |
Requesting sample data at the start of the determina- |
|
|
|
tion .................................................................................... |
193 |
21 |
Sample table |
195 |
|
|
21.1 |
General .............................................................................. |
195 |
|
21.2 |
Edit the sample data ........................................................ |
198 |
|
21.3 |
Properties .......................................................................... |
200 |
22 |
Determination run |
203 |
|
|
22.1 |
Carrying out a single determination ............................... |
203 |
|
22.2 |
Performing a sample series ............................................. |
204 |
|
22.3 |
Canceling determinations manually ............................... |
205 |
23 |
Live modifications |
206 |
23.1Editing the sample data of the running determina-
tion .................................................................................... |
206 |
23.2Editing the sample table while a determination is run-
|
ning .................................................................................... |
207 |
23.3 |
Live display ....................................................................... |
208 |
23.4 |
Live parameters ................................................................ |
210 |
900 Touch Control |
VII |
Table of contents |
|
|
24 Results and more determination data |
212 |
|
24.1 |
More determination data ................................................ |
213 |
24.1.1 |
Details ................................................................................. |
214 |
24.2 |
Messages ........................................................................... |
216 |
24.3 |
Local common variables .................................................. |
217 |
24.4 |
Determination properties ................................................ |
217 |
24.5 |
Loading a determination ................................................. |
220 |
24.5.1 |
Determination list ................................................................ |
221 |
24.6 |
Saving a determination .................................................... |
225 |
24.7 |
Curves ................................................................................ |
225 |
24.8 |
Recalculation and reevaluation ....................................... |
227 |
25 Statistics |
|
229 |
25.1 |
Displaying details for a result .......................................... |
231 |
25.2 |
Deleting statistical data ................................................... |
232 |
25.3 |
Adding a determination to a determination series ....... |
232 |
26 Result table |
|
234 |
26.1 |
Properties .......................................................................... |
235 |
26.2 |
Saving the result table ..................................................... |
239 |
26.3 |
Loading the result table ................................................... |
239 |
27 Printing |
|
240 |
27.1 |
General report options ..................................................... |
242 |
27.2 |
Settings of the individual reports ................................... |
243 |
27.3 |
List of all printable reports .............................................. |
244 |
28 Manual control |
248 |
|
28.1 |
Opening and closing the manual control ....................... |
249 |
28.2 |
Measuring ......................................................................... |
250 |
28.2.1 |
Parameter description .......................................................... |
251 |
28.3 |
Dosing ............................................................................... |
254 |
28.3.1 |
Continuous dosing .............................................................. |
256 |
28.3.2 |
Dosing fixed volumes ........................................................... |
258 |
28.3.3 |
Preparing ............................................................................. |
260 |
28.3.4 |
Emptying ............................................................................. |
261 |
28.3.5 |
Filling .................................................................................. |
262 |
28.3.6 |
Replacing reagent ................................................................ |
262 |
28.4 |
Stirring ............................................................................... |
263 |
VIII |
900 Touch Control |
|
|
|
Table of contents |
28.5 Conditioning a titration cell ............................................ |
264 |
||
28.5.1 |
|
Parameter description .......................................................... |
265 |
28.6 |
Remote .............................................................................. |
265 |
|
28.7 |
USB Sample Processor ..................................................... |
266 |
|
28.7.1 |
|
Moving the lift ..................................................................... |
268 |
28.7.2 |
Moving to a rack position .................................................... |
270 |
|
28.7.3 |
|
External positions ................................................................ |
272 |
29 Parameters |
|
|
276 |
29.1 Dynamic equivalence point titrations (DET) ................... |
276 |
||
29.1.1 |
|
Start conditions ................................................................... |
276 |
29.1.2 |
|
Titration parameters ............................................................ |
279 |
29.1.3 |
|
Stop conditions ................................................................... |
283 |
29.1.4 |
|
Potentiometric evaluation .................................................... |
284 |
29.1.5 |
|
Control device ..................................................................... |
289 |
29.1.6 |
|
Sensor ................................................................................. |
290 |
29.1.7 |
|
Dosing device ...................................................................... |
291 |
29.1.8 |
|
Stirrer .................................................................................. |
292 |
29.2 Monotonic equivalence point titrations (MET) .............. |
293 |
||
29.2.1 |
|
Start conditions ................................................................... |
293 |
29.2.2 |
|
Titration parameters ............................................................ |
296 |
29.2.3 |
|
Stop conditions ................................................................... |
299 |
29.2.4 |
|
Potentiometric evaluation .................................................... |
300 |
29.2.5 |
|
Control device ..................................................................... |
305 |
29.2.6 |
|
Sensor ................................................................................. |
306 |
29.2.7 |
|
Dosing device ...................................................................... |
307 |
29.2.8 |
|
Stirrer .................................................................................. |
308 |
29.3 |
Endpoint titrations (SET) .................................................. |
309 |
|
29.3.1 |
|
Start conditions ................................................................... |
309 |
29.3.2 |
|
Control parameters .............................................................. |
310 |
29.3.3 |
|
Titration parameters ............................................................ |
314 |
29.3.4 |
|
Stop conditions ................................................................... |
315 |
29.3.5 |
|
Conditioning ....................................................................... |
316 |
29.3.6 |
|
Cell ...................................................................................... |
318 |
29.3.7 |
|
Control device ..................................................................... |
318 |
29.3.8 |
|
Sensor ................................................................................. |
318 |
29.3.9 |
|
Dosing device ...................................................................... |
320 |
29.3.10 |
Stirrer .................................................................................. |
321 |
|
29.4 Volumetric Karl Fischer titrations (KFT) ......................... |
321 |
||
29.4.1 |
|
Start conditions ................................................................... |
321 |
29.4.2 |
|
Control parameters .............................................................. |
323 |
29.4.3 |
|
Titration parameters ............................................................ |
326 |
29.4.4 |
|
Stop conditions ................................................................... |
327 |
29.4.5 |
|
Conditioning ....................................................................... |
328 |
29.4.6 |
|
Cell ...................................................................................... |
330 |
29.4.7 |
|
Control device ..................................................................... |
330 |
29.4.8 |
|
Sensor ................................................................................. |
331 |
900 Touch Control |
IX |
Table of contents |
|
|
|
|
29.4.9 |
|
Dosing device ...................................................................... |
|
332 |
29.4.10 |
Stirrer .................................................................................. |
|
333 |
|
29.5 Coulometric Karl Fischer titrations (KFC) ....................... |
334 |
|||
29.5.1 |
|
Control parameters .............................................................. |
|
334 |
29.5.2 |
|
Titration parameters ............................................................ |
|
336 |
29.5.3 |
|
Stop conditions ................................................................... |
|
337 |
29.5.4 |
|
Conditioning ....................................................................... |
|
337 |
29.5.5 |
|
Cell ...................................................................................... |
|
340 |
29.5.6 |
|
Control device ..................................................................... |
|
341 |
29.5.7 |
|
Sensor ................................................................................. |
|
341 |
29.5.8 |
|
Stirrer .................................................................................. |
|
342 |
29.6 Bromine index determination (BRC) ............................... |
|
343 |
||
29.6.1 |
|
Control parameters .............................................................. |
|
343 |
29.6.2 |
|
Titration parameters ............................................................ |
|
346 |
29.6.3 |
|
Stop conditions ................................................................... |
|
347 |
29.6.4 |
|
Conditioning ....................................................................... |
|
347 |
29.6.5 |
|
Cell ...................................................................................... |
|
349 |
29.6.6 |
|
Control device ..................................................................... |
|
350 |
29.6.7 |
|
Sensor ................................................................................. |
|
350 |
29.6.8 |
|
Stirrer .................................................................................. |
|
351 |
29.7 |
STAT titrations (STAT) ..................................................... |
|
352 |
|
29.7.1 |
|
Start conditions ................................................................... |
|
352 |
29.7.2 |
|
Control parameters .............................................................. |
|
354 |
29.7.3 |
|
Titration parameters ............................................................ |
|
356 |
29.7.4 |
|
Stop conditions ................................................................... |
|
358 |
29.7.5 |
|
Monitoring .......................................................................... |
|
359 |
29.7.6 |
|
Control device ..................................................................... |
|
366 |
29.7.7 |
|
Sensor ................................................................................. |
|
366 |
29.7.8 |
|
Dosing device ...................................................................... |
|
367 |
29.7.9 |
|
Stirrer .................................................................................. |
|
370 |
29.8 |
Manual titrations (MAT) .................................................. |
|
371 |
|
29.8.1 |
|
Dosing parameters .............................................................. |
|
371 |
29.8.2 |
|
Control device ..................................................................... |
|
372 |
29.8.3 |
|
Sensor ................................................................................. |
|
372 |
29.8.4 |
|
Dosing device ...................................................................... |
|
373 |
29.8.5 |
|
Stirrer .................................................................................. |
|
376 |
29.9 |
Measurements (MEAS) ..................................................... |
|
377 |
|
29.9.1 |
|
Measuring parameters ......................................................... |
|
377 |
29.9.2 |
|
Standard addition ................................................................ |
|
381 |
29.9.3 |
|
Control device ..................................................................... |
|
385 |
29.9.4 |
|
Sensor ................................................................................. |
|
385 |
29.9.5 |
|
Stirrer .................................................................................. |
|
387 |
29.10 |
Calibration of sensors (CAL) ............................................ |
|
388 |
|
29.10.1 |
Calibration parameters (CAL pH) .......................................... |
|
388 |
|
29.10.2 |
Calibration parameters (CAL Conc) ...................................... |
|
391 |
|
29.10.3 |
Calibration parameters (CAL Cond) ...................................... |
|
394 |
|
29.10.4 |
Control device ..................................................................... |
|
397 |
X |
900 Touch Control |
|
Table of contents |
|
29.10.5 |
Sensor ................................................................................. |
397 |
29.10.6 |
Stirrer .................................................................................. |
398 |
29.11 Electrode test for pH electrodes (ELT) ............................ |
399 |
|
29.11.1 |
Electrode test parameters .................................................... |
400 |
29.11.2 |
Control device ..................................................................... |
401 |
29.11.3 |
Sensor ................................................................................. |
402 |
29.11.4 |
Stirrer .................................................................................. |
403 |
29.11.5 |
Further information ............................................................. |
403 |
29.12 Evaluations (EVAL) ........................................................... |
405 |
|
29.12.1 Fixed endpoint evaluation (EVAL FIX-EP) .............................. |
407 |
|
29.12.2 pK value and half neutralization potential evaluation (EVAL |
|
|
|
pK/HNP) .............................................................................. |
409 |
29.12.3 Minimum and maximum evaluation (EVAL MIN/MAX) .......... |
410 |
|
29.12.4 Break point evaluation (EVAL BREAK) ................................... |
411 |
|
29.12.5 Rate evaluation (EVAL RATE) ................................................ |
414 |
|
29.13 Calculations ....................................................................... |
415 |
|
29.13.1 |
Calculations (CALC) ............................................................. |
415 |
29.13.2 |
Calculations (CALC LIVE) ...................................................... |
422 |
29.13.3 |
Formula editor ..................................................................... |
424 |
29.14 Reports (REPORT) ............................................................. |
430 |
|
29.14.1 |
General report options ......................................................... |
430 |
29.14.2 Settings of the individual reports .......................................... |
431 |
|
29.14.3 |
List of reports ...................................................................... |
432 |
29.15 Dosing and Liquid Handling ............................................ |
435 |
|
29.15.1 Preparing an exchange or dosing unit (PREP) ........................ |
435 |
|
29.15.2 Emptying a dosing unit (EMPTY) .......................................... |
436 |
|
29.15.3 Dosing a specified volume (ADD) ......................................... |
437 |
|
29.15.4 |
Liquid Handling (LQH) .......................................................... |
442 |
29.15.5 |
Monitored dosing (DOS) ...................................................... |
445 |
29.16 Communication ................................................................ |
462 |
|
29.16.1 Scanning remote lines (SCAN) .............................................. |
463 |
|
29.16.2 Setting remote lines (CTRL) .................................................. |
464 |
|
29.16.3 Scanning the RS-232 interface (SCAN RS) ............................. |
466 |
|
29.16.4 Defining RS-232 commands (CONTROL RS) .......................... |
467 |
|
29.17 Automation ....................................................................... |
467 |
|
29.17.1 Rotating sample rack (MOVE) .............................................. |
467 |
|
29.17.2 Moving the lift (LIFT) ............................................................ |
470 |
|
29.17.3 |
Controlling pumps (PUMP) ................................................... |
471 |
29.17.4 Resetting the rack (RACK) .................................................... |
472 |
|
29.17.5 Defining the sample variable (SAMPLE) ................................ |
472 |
|
29.17.6 Creating a subsequence (SUBSEQ) ....................................... |
473 |
|
29.18 Miscellaneous commands ................................................ |
476 |
|
29.18.1 Controlling a stirrer (STIR) .................................................... |
476 |
|
29.18.2 Pausing the method run (WAIT) ........................................... |
477 |
|
29.18.3 |
Scan data (REQUEST) ........................................................... |
477 |
29.18.4 Defining an acoustic signal (BEEP) ........................................ |
478 |
900 Touch Control |
XI |
Table of contents |
|
|
|
29.18.5 |
Signing a determination (SIGN) ............................................ |
479 |
|
29.18.6 |
Canceling the method run (END) ......................................... |
479 |
|
30 Operation and maintenance |
480 |
||
30.1 |
System initialization ......................................................... |
480 |
|
30.2 |
Quality management and qualification with |
|
|
|
Metrohm ........................................................................... |
481 |
|
31 Troubleshooting |
482 |
||
31.1 |
Editing methods ............................................................... |
482 |
|
31.2 |
Sample table ..................................................................... |
482 |
|
31.3 |
Results/Statistics ............................................................... |
482 |
|
31.4 |
Printing .............................................................................. |
483 |
|
31.5 |
Manual control ................................................................. |
483 |
|
31.6 |
Miscellaneous ................................................................... |
484 |
|
31.7 |
SET titration ...................................................................... |
484 |
|
31.8 |
Karl Fischer titration ........................................................ |
485 |
|
31.9 |
STAT titration ................................................................... |
487 |
|
32 Appendix |
|
|
488 |
32.1 |
Dosing unit ........................................................................ |
488 |
|
32.1.1 |
Maximum dosing and filling rate .......................................... |
488 |
|
32.1.2 |
Default parameters for preparing (PREP) and emptying |
|
|
|
|
(EMPTY) ............................................................................... |
488 |
32.2 |
Exchange unit ................................................................... |
489 |
|
32.2.1 |
Maximum dosing and filling rate .......................................... |
489 |
|
32.2.2 |
Default parameters for preparing (PREP) .............................. |
489 |
|
32.3 |
Stirring rate ....................................................................... |
489 |
|
32.4 |
Balance .............................................................................. |
490 |
|
32.5 |
Result variables as parameter setting ............................ |
491 |
|
32.6 |
Electrode calibration with USB Sample Processors ....... |
492 |
|
32.7 |
Stored buffer series for pH calibration ........................... |
494 |
|
32.8 |
Using AuditTrailViewer .................................................... |
498 |
|
32.8.1 |
|
Installing AuditTrailViewer ................................................... |
498 |
32.8.2 |
Opening the Audit Trail ....................................................... |
499 |
|
32.8.3 |
Contents of the Audit Trail ................................................... |
500 |
|
32.8.4 |
Filtering the Audit Trail ......................................................... |
502 |
|
32.8.5 |
Exporting the Audit Trail ...................................................... |
502 |
|
32.9 |
Diagnosis ........................................................................... |
503 |
|
32.9.1 |
|
LCD test .............................................................................. |
503 |
32.9.2 |
|
Temperature monitoring ...................................................... |
504 |
XII |
900 Touch Control |
|
|
Table of contents |
32.9.3 Formatting an external storage medium ............................... |
504 |
|
32.9.4 Removing an external storage medium ................................ |
505 |
|
32.9.5 Adjusting the touch screen .................................................. |
505 |
|
32.9.6 Testing the touch screen ...................................................... |
506 |
|
32.9.7 Software update (loading program versions and language |
||
|
files) .................................................................................... |
507 |
32.9.8 |
Simulating titration curves ................................................... |
510 |
32.9.9 |
Service ................................................................................. |
511 |
32.10 |
Arithmetic algorithms in the Titrando ............................ |
512 |
33 Technical specifications |
514 |
|
33.1 |
Touch screen ..................................................................... |
514 |
33.2 |
Interfaces .......................................................................... |
515 |
33.3 |
Power supply .................................................................... |
515 |
33.4 |
Safety specifications ........................................................ |
515 |
33.5 |
Electromagnetic compatibility (EMC) ............................. |
516 |
33.6 |
Ambient temperature ...................................................... |
516 |
33.7 |
Dimensions ........................................................................ |
516 |
34 Warranty (guarantee) |
517 |
|
35 Accessories |
|
519 |
Index |
|
521 |
900 Touch Control |
XIII |
Table of figures |
|
Table of figures |
|
|
Figure 1 |
Front 900 Touch Control ................................................................... |
8 |
Figure 2 |
Rear 900 Touch Control .................................................................... |
9 |
Figure 3 |
Reagent dosing for DET ................................................................... |
11 |
Figure 4 |
Reagent dosing for MET .................................................................. |
11 |
Figure 5 |
Reagent dosing for SET ................................................................... |
12 |
Figure 6 |
Reagent dosing for KFT ................................................................... |
13 |
Figure 7 |
Reagent dosing for STAT ................................................................. |
14 |
Figure 8 |
Dosing unit – port assignment ......................................................... |
51 |
Figure 9 |
Exchange unit – tubing connections ................................................ |
56 |
Figure 10 |
Directory structure on the external storage medium ...................... |
124 |
Figure 11 |
Live display "Preparing the dosing unit" ......................................... |
261 |
Figure 12 |
Live display "Preparing the exchange unit" ..................................... |
261 |
Figure 13 |
Equivalence point recognition and equivalence point numbering within |
|
|
windows ....................................................................................... |
288 |
Figure 14 |
Tubbs method for determining the equivalence point .................... |
289 |
Figure 15 |
Equivalence point recognition and equivalence point numbering within |
|
|
windows ....................................................................................... |
304 |
Figure 16 |
Action "Exit method" or "Cancel command" .................................. |
363 |
Figure 17 |
Action "Pause" .............................................................................. |
363 |
Figure 18 |
Action "Wait" ................................................................................ |
364 |
Figure 19 |
Evaluation of pK value / half neutralization potential ...................... |
409 |
Figure 20 |
Evaluation of minimum and maximum ........................................... |
410 |
Figure 21 |
Evaluation of a break point ............................................................ |
411 |
Figure 22 |
Dosing criterion "Volume/Dosing rate" .......................................... |
446 |
Figure 23 |
Dosing criterion "Volume/Dosing time" .......................................... |
446 |
Figure 24 |
Dosing criterion "Dosing rate/Dosing time" .................................... |
447 |
Figure 25 |
Action "Exit method" or "Cancel command" .................................. |
455 |
Figure 26 |
Action "Pause" .............................................................................. |
455 |
Figure 27 |
Action "Wait" ................................................................................ |
455 |
Figure 28 |
Rotational speed depending on stirring rate .................................. |
490 |
XIV |
900 Touch Control |
|
1 Introduction |
1 Introduction
The 900 Touch Control is an operating unit with a touch-sensitive screen. You can operate the following control devices with the 900 Touch Control:
Titrandos
USB Sample Processors
846 Dosing Interface
867 pH Module
856 Conductivity Module
With the 900 Touch Control, you can conveniently manage all of your titrants, sensors, methods, etc. For example, you can connect a USB flash drive to the USB connector as an external storage medium. You can use this additional memory not only to store your methods and determinations, but also to create a backup together with all of the data and settings of your system. Thanks to the integrated Ethernet connector, the device can be connected to your network, thus offering you the following options, among others:
Saving data to a PC within the network
Printing reports on a network printer
Sending displayed messages as e-mails
The 900 Touch Control supports the following titration and measuring modes. Whether a particular mode is available or not depends on the type of the control device connected.
DET
Dynamic equivalence point titration. The reagent is added in variable volume steps.
Measuring modes:
–pH (pH measurement)
–U (potentiometric voltage measurement)
–Ipol (voltametric measurement with selectable polarization current)
–Upol (amperometric measurement with selectable polarization voltage)
900 Touch Control |
1 |
1.2 Titration and measuring modes |
|
MET
Monotonic equivalence point titration. The reagent is added in constant volume steps.
Measuring modes:
–pH (pH measurement)
–U (potentiometric voltage measurement)
–Ipol (voltametric measurement with selectable polarization current)
–Upol (amperometric measurement with selectable polarization voltage)
SET
Endpoint titration at one or two specified endpoints. Measuring modes:
–pH (pH measurement)
–U (potentiometric voltage measurement)
–Ipol (voltametric measurement with selectable polarization current)
–Upol (amperometric measurement with selectable polarization voltage)
STAT
Endpoint titration with constant maintenance of the measured value. Measuring modes:
–pH (pH measurement)
–U (potentiometric voltage measurement)
KFT
Volumetric water content determination according to Karl Fischer. Measuring modes:
–Ipol (voltametric measurement with selectable polarization current)
–Upol (amperometric measurement with selectable polarization voltage)
KFC
Coulometric water content determination according to Karl Fischer. Measuring mode:
–Ipol (voltametric measurement with selectable polarization current)
BRC
Coulometric bromine index determination. Determining the amount of double bonds in e.g. mineral oils.
Measuring mode:
–Ipol (voltametric measurement with selectable polarization current)
2 |
900 Touch Control |
|
1 Introduction |
MEAS
Measuring modes:
–pH (pH measurement)
–U (potentiometric voltage measurement)
–Ipol (voltametric measurement with selectable polarization current)
–Upol (amperometric measurement with selectable polarization voltage)
–T (temperature measurement)
–Conc (Concentration measurement with or without standard addition)
–Cond (conductivity measurement)
CAL
Electrode calibration. Measuring mode:
–pH (calibration of pH electrodes)
–Conc (calibration of ion-selective electrodes)
–Cond (calibration of conductivity measuring cells)
–ELT (Electrode test for pH electrodes)
CAUTION
Please read through this documentation carefully before putting the instrument into operation. The documentation contains information and warnings which the user must follow in order to ensure safe operation of the instrument.
The following symbols and formatting may appear in this documentation:
|
Cross-reference to figure legend |
|
The first number refers to the figure number, the sec- |
|
ond to the instrument part in the figure. |
|
|
|
Instruction step |
|
Carry out these steps in the sequence shown. |
|
|
Method |
Dialog text, parameter in the software |
|
|
File New |
Menu or menu item |
|
|
[Next] |
Button or key |
|
|
900 Touch Control |
3 |
1.3 About the documentation |
|
WARNING
This symbol draws attention to a possible life-threat- ening hazard or risk of injury.
WARNING
This symbol draws attention to a possible hazard due to electrical current.
WARNING
This symbol draws attention to a possible hazard due to heat or hot instrument parts.
WARNING
This symbol draws attention to a possible biological hazard.
CAUTION
This symbol draws attention to possible damage to instruments or instrument parts.
NOTE
This symbol highlights additional information and tips.
4 |
900 Touch Control |
|
2 Safety instructions |
2 Safety instructions
WARNING
This instrument may only be operated in accordance with the specifications in this documentation.
This instrument has left the factory in a flawless state in terms of technical safety. To maintain this state and ensure non-hazardous operation of the instrument, the following instructions must be observed carefully.
The electrical safety when working with the instrument is ensured as part of the international standard IEC 61010.
WARNING
Only personnel qualified by Metrohm are authorized to carry out service work on electronic components.
WARNING
Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are touched.
There are no parts inside the housing which can be serviced or replaced by the user.
Mains voltage
WARNING
An incorrect mains voltage can damage the instrument.
Only operate this instrument with a mains voltage specified for it (see rear panel of the instrument).
900 Touch Control |
5 |
2.3 Tubing and capillary connections |
|
Protection against electrostatic charges
WARNING
Electronic components are sensitive to electrostatic charges and can be destroyed by discharges.
Do not fail to pull the mains cable out of the mains connection socket before you set up or disconnect electrical plug connections at the rear of the instrument.
2.3Tubing and capillary connections
CAUTION
Leaks in tubing and capillary connections are a safety risk. Tighten all connections well by hand. Avoid applying excessive force to tubing connections. Damaged tubing ends lead to leakage. Appropriate tools can be used to loosen connections.
Check the connections regularly for leakage. If the instrument is used mainly in unattended operation, then weekly inspections are mandatory.
WARNING
All relevant safety measures are to be observed when working with flammable solvents and chemicals.
Set up the instrument in a well-ventilated location (e.g. fume cupboard).
Keep all sources of flame far from the workplace.
Clean up spilled liquids and solids immediately.
Follow the safety instructions of the chemical manufacturer.
6 |
900 Touch Control |
|
2 Safety instructions |
This product is covered by European Directive 2002/96/EC, WEEE – Waste from Electrical and Electronic Equipment.
The correct disposal of your old equipment will help to prevent negative effects on the environment and public health.
More details about the disposal of your old equipment can be obtained from your local authorities, from waste disposal companies or from your local dealer.
900 Touch Control |
7 |
1
|
2 |
3 |
4 |
5 |
6 |
7 |
8 |
|
|
Figure 1 Front 900 Touch Control |
|
|
|||||
|
|
|
|
|
|
|
||
1 |
Display |
|
|
|
2 |
Fixed key [Home] |
||
|
Touch screen. |
|
|
|
|
Opens the main dialog. |
||
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
||
3 |
Fixed key [Back] |
|
|
|
4 |
Fixed key [Help] |
||
|
Opens the next-higher dialog page. |
|
|
|
|
Opens the online help for the dialog dis- |
||
|
|
|
|
|
|
played. |
|
|
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
||
5 |
Fixed key [Print] |
|
|
|
6 |
Fixed key [Manual] |
||
|
Opens the print dialog. |
|
|
|
|
Opens the manual control. |
||
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
||
7 |
Fixed key [STOP] |
|
|
|
8 |
Fixed key [START] |
||
|
Cancels the running determination. |
|
|
|
|
Starts a determination. |
||
|
|
|
|
|
|
|
|
|
8 |
900 Touch Control |
|
3 Overview of the instrument |
1 |
2 |
3 |
4 |
5 |
Figure 2 Rear 900 Touch Control
1Ethernet connector (RJ-45)
For connecting to a network.
2USB connector (type A)
For connecting printers, USB sticks, USB hubs, etc.
3Power switch
Switches the instrument on/off.
4Type plate
Contains the serial number.
5Connection cable
For the connection of the Touch Control to a control device (socket "Controller").
900 Touch Control |
9 |
4.1 Setting up the instrument |
|
The connection between the Touch Control and the control device (Titrando, etc.) is described in the respective manual, as is the configuration of the titration system with its peripheral devices, e.g. stirrers and dosing devices.
4.1Setting up the instrument
The instrument is supplied in highly protective special packaging together with the separately packed accessories. Keep this packaging, as only this ensures safe transportation of the instrument.
Immediately after receipt, check whether the shipment has arrived complete and without damage by comparing it with the delivery note.
The instrument has been developed for operation indoors and may not be used in explosive environments.
Place the instrument in a location of the laboratory which is suitable for operation, free of vibrations, protected from corrosive atmosphere, and contamination by chemicals.
The instrument should be protected against excessive temperature fluctuations and direct sunlight.
10 |
900 Touch Control |
|
5 Titrations |
5 Titrations
Dynamic equivalence point titration is a titration mode for all standard titrations. The reagent is added in variable volume steps. The volume increments vary as a function of the slope of the curve. An attempt is made to achieve constant measured value alterations with each dosing. The optimal volume for dosing is determined from the measured value alterations of the previous dosings. Measured value acceptance is driftcontrolled (equilibrium titration) or after a waiting time. Equivalence points are evaluated automatically.
U/mV
V/mL
Figure 3 Reagent dosing for DET
Monotonic equivalence point titration is a titration mode for titrations with relatively high signal fluctuations or suddenly occurring potential jumps and for slow titrations or slow-response electrodes. The reagent is added in constant volume steps. Measured value acceptance is drift-con- trolled (equilibrium titration) or after a waiting time. Equivalence points are evaluated automatically.
U/mV
V/mL
Figure 4 Reagent dosing for MET
900 Touch Control |
11 |
5.3 Endpoint titration (SET) |
|
5.3Endpoint titration (SET)
Endpoint titration is a titration mode for rapid routine determinations to a preset endpoint (e.g. titrations in accordance with special norms) and titrations for which reagent overflow must be avoided. The titration termination at the endpoint takes place either drift-controlled or after a waiting period. The volume dosed until the endpoint is used for calculating the content of the sample.
U/mV
EP
Control range
V/mL
t/s
Initial |
Continuous |
V/mL |
dosing |
dosing |
|
Figure 5 Reagent dosing for SET
Karl Fischer Titration is a method for volumetric water determination. Conditioning is carried out automatically both before and after the actual titration. The reagent dosing is controlled in such a way that a predefined endpoint is reached as quickly and as accurately as possible. The volume steps and the rate of reagent dosing are regulated by the difference between the current measured value and the predefined endpoint. This means that titration is performed more slowly in the control range and that smaller volumes are added. The titration is stopped at the endpoint either drift-controlled or after a waiting time. The volume dosed until the endpoint is used to calculate the water content of the sample.
12 |
900 Touch Control |
|
5 Titrations |
U/mV
KFT Ipol
Control range
EP
V/mL
t/s
V/mL
Initial Continuous dosing dosing
I/µA
KFT Upol
EP
Control range
V/mL
Figure 6 Reagent dosing for KFT
The pH STAT titration is the one most frequently utilized. Reagent is added until a preset measured value (control point) has been reached. This control point is kept stable by continuously titrating off the substance set free by the reaction with the reagent. The STAT titration keeps the control point constant until the preset stop criterion has been reached. The STAT titration is applied, for example, in analysis of enzymes. The dosing rate resulting from the the control point being kept stable provides insight into the activity of an enzyme.
900 Touch Control |
13 |
5.5 Endpoint titration with constant maintenance of the measured value (STAT)
U/mV
Control range
V/mL
Initial |
Incremental dosing |
dosing |
|
Figure 7 Reagent dosing for STAT
Control point
V/mL
t/s
14 |
900 Touch Control |