CLINITEK ATLAS® PREVENTIVE MAINTENANCE CHECKLIST

Date:

 

 

FE:

 

 

 

S/N:

 

Customer:

 

 

 

 

City, State:

 

Notes in parentheses "(

)" refer to the paragraph or page number in the Clinitek Atlas or Rack Sample Handler service manuals.

1)CLEAN THE FOLLOWING:

a)

SG Fiber Optic well (see Clinitek Atlas Service Manual step 12 on page 5.7)

b)

Sample Handler and instrument surfaces and covers (see paragraph 5-2 on page 5.1)

c)

Tube Detector window, Home/Tooth Detector window, Waste Bottle Detector and Bar Code window (page 5.3)

d)

Reagent storage module cavity; inspect door gasket (see paragraph 5-5-1 step 5 on page 5.6)

e)

Platen and rinse well (see paragraph 5-5-1 step 2 on page 5.6)

f)

Readhead with cleaning brush (see paragraph 5-4-4 on page 5.4). Do NOT clean channel 14!

g)

Two vertical and horizontal pipette transport steel shafts to remove rust and dirt (see 5-5-1 steps 9 to 10, page 5.7).

h)

Lubricate four shafts with a light coating of light machine oil or synthetic oil (Tri-Flow ). Do not use aerosol spray!

Cavro pump lead screw; lubricate screw with syringe lubricant in supply kit (see paragraph 5-8 on page 5.9)

2)VERIFY PROPER OPERATION OF THE FOLLOWING:

a)

Check that Cavro pump syringe is not worn or damaged. Check all pump valve fittings are intact and tight.

 

Do a Prime Pump while observing pump: no syringe leakage or bubbles may be present

b)

Verify sample tray ID recognition and tube detector operates properly (check with no tubes, an even position,

 

an odd position and with tray full of tubes) (see paragraph 7A-2-3-9 on page 7A.26)

c)

Check bar code reader for proper operation: check both inner and outer tube locations with barcode labels on tubes

d)

Verify set screws are 1.5 mm below surface of pipette pulleys and pulleys don’t slip when pipette moves (page 9.4)

e)

Verify reagent roll format lies flat against platen (see paragraph 8-3 on page 8.7)

f)

Verify reagent pads are centered under readhead channels 1 and 11 (see paragraph 8-3 on page 8.7)

g)

Verify tension sprocket pin tip is at right edge of a hole in reagent roll with roll resting on top of sprocket (page 8.8)

h)

Verify take-up reel set screw is securely fastened to the spur gear shaft (see item 111 on page 10.10)

i)

Verify pipette dispenses on all reagent pads (see paragraph 8-5-2 on page 8.13)

j)

Verify pipette goes to the center of the rinse well (see Figure 8-7 on page 8.20)

k)

Verify pipette correctly detects 3 ml of Cal 1 solution in a tube

l)

Verify output ports P1 and S1 are functional (see paragraph 7A-2-2-9 on page 7A.7)

3)CHECK POWER SUPPLY OUTPUTS:

a)For Switching System International (SSI) Power Supply (part number 95001734 with steel cover, see page 8.23) or Coutant Lambda-Qualidyne Power Supply (part number 95001751, with a blue or black cover, see page 8.24):

+5VDC -0, +0.15 VDC, < 50mV AC ripple: measure at both “+5V” terminals on back of Card Cage, behind display +12VDC -0, +0.25 VDC, < 120mV AC ripple: measure at Power Driver PC terminals "+12VA", “+12VB" and “+12VC" +24VDC ±1 VDC, < 240mV AC ripple. Measure at the Power Driver PCB terminal marked "+24V”.

NOTE: the 24 VDC supply is not adjustable on the steel cover SSI power supply.

b)For Astec Power Supply (part number 95001751 with a bronze cover, see page 8.27 or service bulletin ATLAS-129)

+5VDC -0, +0.15, < 50mV AC ripple: measure at both “+5V” terminals on the back of the Card Cage, behind display +12VDC -0, +0.25, < 120mV AC ripple: measure at Power Driver PCB terminals “+12VA", "+12VB" and "+12VC" +24VDC ±1 VDC, < 240mV AC ripple: measure at the Power Driver PCB terminal marked "+ 24V”

NOTE: the 12 VDC supply is not adjustable on the bronze cover Astec power supply.

4)IF A RACK SAMPLE HANDLER IS USED, DO THE FOLLOWING (using Rack Sample Handler Service Manual):

a)

Verify Transfer Plate (see item 9 on page 10.36 of Rack Sample Handler Service Manual) and racks are clean.

b)

Verify the Reflection Plate (item 2 on page 10.4 of Rack Handler Service Manual) is clean, not scratched or loose.

c)

Verify PH1-7 and PH9 sensor windows are clean (see Figure 8-3 on page 8.4 of Rack Handler Service Manual)

d)

Verify Barcode Reader window is clean (if instrument is using a barcode reader).

e)

Verify waste tubing is draining properly into the waste receptacle.

f)

Clean system with a soft, clean, damp cloth and mild detergent. Follow with a soft, clean, damp cloth. Dry carefully.

g)

Do a With Tube Run in Service Mode (Rack Handler page 7.4) for 30 minutes to verify system works properly.

5)FINAL CHECKOUT

a)

Print out Reflectance Test values and leave them for future comparison (Clinitek Atlas Service Manual page 7A.21)

b)

Turn unit off and on again to return to normal mode

 

 

c)

Complete a successful calibration, then run controls and review the results with customer

d)

Run samples in Analyze operation and observe instrument for proper operation

 

 

e)

Verify that printer and computer outputs are operative (if used).

 

 

 

 

 

 

CUSTOMER SIGNATURE:

 

DATE:

 

Provide copies of this form to the customer and to the appropriate office