Olympus TJF Q180V User manual

0 (0)

Single use soft brush (MAJ-1888)

INSTRUCTIONS

EVIS EXERA II DUODENOVIDEOSCOPE

OLYMPUS TJF TYPE Q180V

Accessories:

Water resistant cap (MH-553)

Biopsy valve (MB-358)

Air/water valve (MH-438)

Suction cleaning adapter (MH-856)

Channel cleaning brush (BW-20T)

Single use channel cleaning brush (BW-201T)

AW channel cleaning adapter (MH-948)

Single use soft brush (MAJ-1888)

Chain for water-resistant cap (MAJ-1119)

Suction valve (MH-443)

Mouthpiece (MB-142)

Channel plug (MH-944)

Injection tube (MH-946)

Single use channel-opening cleaning brush (MAJ-1339)

Single use combination cleaning brush (BW-412T)

MH-553

MAJ-1119

MB-358

MH-443

MH-438

MB-142

MH-856

MH-944

BW-20T

MH-946

 

MH-948

BW-412T

MAJ-1888

BW-201T

MAJ-1339

Refer to the endoscope’s companion manual, the “OPERATION MANUAL” with your endoscope model listed on the cover, for operation information.

USA: CAUTION: Federal law restricts this device to sale by or on the order of a physician.

Revision History

Note: The Revision History shows the latest changes.

Version

Date

 

Description of Changes

 

 

 

 

RC2409 01

March, 2015

Cover

Single use soft brush (MAJ-1888) information.

 

 

Section 1.4

Updated WARNING statements for MAJ-1888 and

 

 

 

using an AER.

 

 

Sections 2.10, 3.1, and 4.1

Updated MAJ-1888 information.

 

 

Section 4.2

Updated MAJ-1888 and AER information.

 

 

Sections 4.3 and 5.1

Updated MAJ-1888 information.

 

 

Section 5.2

Updated “Aspirate water” section.

 

 

Section 5.3

Updated “Detach the endoscope from the light

 

 

 

source” section.

 

 

Section 5.4

Updated “Equipment needed” chart and CAUTION for

 

 

 

MAJ-1888, Step 5 on page 69, and added “Brush and

 

 

 

flush the forceps elevator recess” section.

 

 

Section 5.5

Updated “Flush all channels and around the forceps

 

 

 

elevator with disinfectant solution” section.

 

 

Section 5.6

Updated “Rinse the endoscope and accessories”

 

 

 

section; the “Alcohol flush” section.

 

 

Chapter 7

Updated WARNING statements for AER.

 

 

 

 

GE8415 09

February,

 

 

 

2015

 

 

 

 

 

 

Contents

Contents

 

Chapter 1

General Policy .........................................................

1

1.1

Instructions......................................................................................

1

1.2

Importance of cleaning, disinfection, and sterilization.....................

2

1.3

Signal words ...................................................................................

2

1.4

Precautions .....................................................................................

3

1.5

Reprocessing before the first use ...................................................

8

1.6

Reprocessing and storage after use ...............................................

9

1.7

Reprocessing before patient procedure..........................................

9

Chapter 2 Function and Inspection of the Accessories for

 

 

Reprocessing...........................................................

10

2.1

Water resistant cap (MH-553).........................................................

10

2.2

Channel plug (MH-944)...................................................................

12

2.3

Injection tube (MH-946) ..................................................................

14

2.4

Channel cleaning brush (BW-20T)..................................................

16

2.5

Suction cleaning adapter (MH-856) ................................................

18

2.6

AW channel cleaning adapter (MH-948).........................................

19

2.7

Single use channel cleaning brush (BW-201T)...............................

20

2.8

Single use channel-opening cleaning brush (MAJ-1339) ...............

22

2.9

Single use combination cleaning brush (BW-412T)........................

24

2.10

Single use soft brush (MAJ-1888)...................................................

25

2.11

Chain for water-resistant cap (MAJ-1119) ......................................

27

Chapter 3 Compatible Reprocessing Methods and

 

 

Chemical Agents ....................................................

28

3.1

Compatibility summary....................................................................

28

3.2

Water (for reprocessing) .................................................................

30

3.3

Detergent solution...........................................................................

31

3.4

Disinfectant solution........................................................................

31

3.5

Rinse water .....................................................................................

31

3.6

Alcohol ............................................................................................

31

3.7

Ethylene oxide gas sterilization.......................................................

32

3.8

Steam sterilization (autoclaving) .....................................................

33

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

i

Contents

Chapter 4 Reprocessing Workflow for the Endoscope

 

and Accessories ......................................................

35

4.1Workflow for manually cleaning and disinfecting the endoscope

and accessories ..............................................................................

36

4.2Workflow for cleaning and disinfecting the endoscope and

accessories using an AER ..............................................................

38

4.3Workflow for manually cleaning and sterilizing the endoscope

 

and accessories ..............................................................................

40

Chapter 5 Reprocessing the Endoscope

 

 

(and related reprocessing accessories) ................

42

5.1

Preparing the equipment for reprocessing......................................

44

5.2

Precleaning the endoscope and accessories..................................

45

5.3

Leakage testing of the endoscope ..................................................

51

5.4

Manually cleaning the endoscope and accessories........................

56

5.5

Manually disinfecting the endoscope and accessories ...................

80

5.6

Rinsing the endoscope and accessories following disinfection.......

86

5.7

Sterilizing the endoscope and accessories.....................................

94

Chapter 6 Reprocessing the Accessories ..............................

96

6.1

Manually cleaning the accessories .................................................

98

6.2

Manually disinfecting the accessories.............................................

101

6.3

Rinsing the accessories following disinfection ................................

102

6.4

Sterilizing the accessories...............................................................

105

Chapter 7 Reprocessing Endoscopes and Accessories

 

 

using an Automated Endoscope Reprocessor.....

106

Chapter 8 Storage and Disposal..............................................

108

8.1

Storing the disinfected endoscope and accessories .......................

109

8.2

Storing the sterilized endoscope and accessories ..........................

111

8.3

Disposal ..........................................................................................

111

ii

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

Chapter 1 General Policy

Chapter 1 General Policy

1.1Instructions

This manual contains the cleaning, disinfection, and sterilization methods recommended by Olympus for the endoscopes and accessories listed on the front cover.

This instruction manual contains essential information on reprocessing endoscopes and accessories safely and effectively.

Before reprocessing, thoroughly review this manual and the manuals for the reprocessing equipment and chemicals that will be used for reprocessing. Reprocess all the devices as instructed.

Note that the complete instruction manual set for the endoscope and accessories consists of this manual and the “OPERATION MANUAL” with your endoscope model listed on the cover. Both manuals accompanied the endoscope at shipment.

Keep this manual and all related manuals in a safe and accessible location (e.g., in the reprocessing area).

If you have any questions or comments about any information in this manual, or if a problem occurs while reprocessing that cannot be solved, contact Olympus.

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

1

Chapter 1 General Policy

1.2Importance of cleaning, disinfection, and sterilization

The medical literature reports incidents of cross-contamination resulting from improper cleaning, disinfection, or sterilization. It is strongly recommended that all individuals engaged in reprocessing closely observe all instructions given in this manual and the manuals for all ancillary equipment, and have a thorough understanding of the following items:

Professional health and safety policies of your hospital

Instruction manuals for the endoscope, accessories, and all the other reprocessing equipment

Structure and handling of endoscope and accessories

Handling of pertinent chemicals

When selecting appropriate methods and conditions for cleaning and disinfection and sterilization, follow the policies at your institution, applicable national laws and standards, and professional society guidelines and recommended practices, in addition to the instructions given in this manual.

1.3Signal words

The following signal words are used throughout this manual:

Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.

Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.

Indicates additional helpful information.

2

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

Chapter 1 General Policy

1.4Precautions

An insufficiently cleaned, disinfected, or sterilized endoscope and/or accessories may pose an infection control risk to the patients and/or operators who contact them.

All disinfection methods (whether performed manually or by an automated endoscope reprocessor), and all sterilization methods (whether performed by ethylene oxide gas or steam) require thorough prior cleaning of the instrument being reprocessed. If the equipment is not adequately cleaned prior to disinfection/sterilization, these processes will be ineffective. Immediately after each patient procedure and before disinfection/sterilization, thoroughly clean the endoscope and the accessories used with the endoscope.

All channels of the endoscope, including the instrument channel and all accessories used with the endoscope during the patient procedure, such as all valves, must be cleaned and high-level disinfected or sterilized after each patient procedure, even if the channels or accessories were not used during the patient procedure. Insufficient cleaning and disinfection or sterilization of these components may pose an infection control risk to patients and/or operators.

Disinfectant solutions are hazardous. After disinfection, rinse all external surfaces and channels of the endoscope and accessories thoroughly with water to remove residual disinfectant solution.

The results of sterilization depend on various factors. These factors include how the equipment was packaged, and the placing and loading of the package in the sterilization device. Verify the sterilization process using biological and/or chemical indicators. Follow the guidelines for sterilization issued by national authorities, professional organizations and infection control professionals, as well as the instruction manual for the sterilization device.

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

3

Chapter 1 General Policy

Establish an internal system of identifying contaminated versus reprocessed endoscopes and accessories to prevent both mix-ups and cross-contamination. Touching a reprocessed endoscope and/or accessories with contaminated gloves or placing them on a contaminated hanger or surface, including letting them touch the floor, will recontaminate them.

Prior to each patient procedure, confirm that the endoscope and accessories have been properly reprocessed and stored. If there are any doubts or questions, reprocess them again before the patient procedure, following the instructions given in this manual.

Perform a leakage test on the endoscope after each precleaning procedure. Do not use the endoscope if a leak is detected. Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions. Use of a leaking endoscope may also pose an infection control risk.

Leakage tester

Figure 1.1

Store alcohol in an airtight container. Alcohol stored in an open container may cause a fire hazard and may result in a loss of efficacy due to evaporation.

4

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

Chapter 1 General Policy

Do not use the AW channel cleaning adapter (MH-948) for patient procedures. It will cause continuous insufflation and could result in patient injury.

AW channel cleaning adapter (MH-948)

Figure 1.2

The accessories listed on the front cover of this manual are consumables, meaning that these accessories cannot be refurbished or repaired and are intended to be replaced once they show any signs of wear. Should any irregularity be observed, use a replacement accessory instead. Using defective accessories may cause equipment malfunction, reduce the efficacy of reprocessing, present a risk to patients and/or operators, or damage the endoscope and/or accessories.

Single-use brushes, such as the single use channel cleaning brush (BW-201T), the single use combination cleaning brush (BW-412T), the single use channel-opening cleaning brush (MAJ-1339), and the single use soft brush (MAJ-1888), are designed for cleaning only one endoscope and its related accessories. Dispose of the single-use brush immediately after use. Using a single-use brush to clean multiple endoscopes and/or accessories may reduce its cleaning efficacy and may damage the brush leading to brush breakage or endoscope and/or accessory damage.

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

5

Chapter 1 General Policy

Patient debris and reprocessing chemicals are hazardous. To guard against contact with dangerous chemicals and potentially infectious material, wear appropriate personal protective equipment during cleaning, disinfection, and sterilization. Such protective equipment should include appropriate eyewear, face mask, cap, moisture-resistant clothing, shoe covers, and chemical-resistant gloves that fit properly and are long enough to prevent skin exposure.

The reprocessing room must be adequately ventilated. Adequate ventilation protects against the buildup of toxic chemical fumes.

Always remove contaminated personal protective equipment before leaving the reprocessing area to prevent contamination from spreading.

Only Olympus-recommended or Olympus-endorsed automated endoscope reprocessors (AERs) have been validated by Olympus. When using an AER that is not recommended by Olympus, the manufacturer of the AER is responsible for validating compatibility of the AER with each Olympus endoscope and accessory.

Before using an AER, confirm that it is capable of reprocessing the endoscope including all channels, the forceps elevator recess, and accessories. Be sure to attach all required connectors. Otherwise, insufficient reprocessing may pose an infection control risk. If you are uncertain as to the ability of your AER to reprocess the endoscope including all channels, the forceps elevator recess, and accessories, contact the manufacturer of the AER for specific instructions and information on compatibility and required connectors. When you use an AER which allows you skip some steps in precleaning and manual cleaning of endoscopes, confirm with the AER manufacturer that such skip is applicable to this endoscope and establish detailed precleaning and manual cleaning procedures of this endoscope according to both instructions of this manual and the AER manufacturer.

Put the forceps elevator in intermediate position of the range of movement by turning the elevator control lever and set it in your AER.

6

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

Chapter 1 General Policy

Instructions provided in this manual are not valid for Olympus devices repaired by a non-Olympus facility. The Olympus recommended reprocessing procedures have not been validated for reprocessing devices repaired by a non-Olympus facility. In the event that your device has been repaired by a non-Olympus facility, contact that repair facility for instructions regarding reprocessing.

Prions, which are the pathogenic agent of the Creutzfeldt-Jakob disease (CJD), cannot be destroyed or inactivated by the cleaning, disinfection, and sterilization methods stated in this instruction manual. When using the endoscope and accessories on patients with CJD or variant Creutzfeldt-Jakob disease (vCJD), be sure to use them for such patients only, or immediately dispose of them after use in an appropriate manner to prevent the usage of exposed devices on other patients. For methods to handle CJD, follow the respective guidelines in your country.

The endoscope and accessories may be damaged by published methods for destroying or inactivating prions. For information on the durability of Olympus equipment against a particular reprocessing method, contact Olympus. In general, Olympus cannot guarantee the effectiveness, safety, and durability of cleaning, disinfection, or sterilization methods not described in this reprocessing manual. If you chose to use a reprocessing method not recommended in this manual, the local institution and/or physicians must assume responsibility for its safety and efficacy. Make sure to carefully inspect each piece of endoscopic equipment for irregularities (damage) prior to each patient procedure. Do not use the equipment if any irregularity is found.

Good quality control practices typically require appropriate documentation. Items such as local SOPs (standard operating procedures), confirmation of operator training, routine testing of the disinfectant’s MEC (minimal effective concentration), confirmation of the disinfectant’s use-life, etc., should be documented as performed.

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

7

Chapter 1 General Policy

When reprocessing the endoscope, confirm that the water resistant cap (MH-553) is securely attached to the electrical connector before immersing the endoscope in reprocessing fluids. If the water resistant cap is not securely attached, the reprocessing fluids could enter the endoscope and damage the endoscope.

Electrical connector

Water resistant

cap (MH-553)

 

Figure 1.3

When aerating or irrigating the endoscope channels, the air or water pressure must not exceed 0.5 MPa (5 kgf/cm2,

71 psig). Higher pressures may cause damage to the endoscope.

Store spare accessories in their original packaging to prevent damage.

To prevent damage, do not apply excessive force to the endoscope and accessories during reprocessing.

Vapors from disinfectant solutions and alcohol may damage electronic devices such as computers.

1.5Reprocessing before the first use

New endoscopes, repaired endoscopes, accessories, and the carrying case for endoscopes are not cleaned, disinfected, or sterilized prior to shipping from Olympus, regardless of whether those instruments are for new purchase, demo or loaner purposes. Reprocess all such endoscopes and accessories received from Olympus according to the instructions given in this manual before storage and before using them in a patient procedure.

8

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

Chapter 1 General Policy

1.6Reprocessing and storage after use

Do not reuse rinse water.

High-level disinfectant solutions are only effective when used according to the disinfectant manufacturer’s instructions. Follow the manufacturer’s instructions regarding activation (if required), concentration, temperature, contact time and use life required to achieve high-level disinfection.

If the disinfectant solution is reused, check its efficacy with a test strip according to the disinfectant manufacturer’s recommendations prior to use.

Do not reuse alcohol.

Alcohol is not a sterilant or high-level disinfectant.

To maintain sterility of equipment following sterilization, use sterile packaging and wraps according to national guidelines.

1.7Reprocessing before patient procedure

Improper storage practices, such as not thoroughly drying external and internal surfaces (lumens) including the forceps elevator recess prior to storage, will lead to an infection control risk.

Improper handling, such as touching a reprocessed endoscope and/or accessories with contaminated gloves, placing a reprocessed device on a contaminated hanger or surface, allowing devices to touch the floor, etc., will recontaminate the device.

Some national or professional guidelines recommend reprocessing endoscopes prior to their first use of the day.

Confirm that the endoscope and accessories have undergone proper reprocessing following their last use and that they have been stored properly. Check the storage period of reprocessed endoscopes, and check for surface contamination (e.g., dust). Check the sterilization expiration date(s) of all items so marked and for tears or breaches in sterile packaging. If there are any doubts or questions concerning whether a device is contaminated, reprocess it again following the instructions given in this manual.

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

9

Chapter 2 Function and Inspection of the Accessories for Reprocessing

Chapter 2 Function and Inspection of

the Accessories for

Reprocessing

Certain accessories are required for reprocessing the endoscope. This chapter describes the function of these accessories. It also describes how to inspect these accessories before using them to reprocess the endoscope.

2.1Water resistant cap (MH-553)

Groove

 

Seals

Venting connector

 

Figure 2.1

10

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

Chapter 2 Function and Inspection of the Accessories for Reprocessing

Function

The water resistant cap is attached to the electrical connector on the endoscope to protect the connector and the endoscope from water penetration during reprocessing. During leakage testing, the leakage tester (MB-155) is attached to the venting connector of the water resistant cap.

Electrical connector

Water resistant

cap (MH-553)

 

 

Chain for water-resistant

 

cap (MAJ-1119)

Figure 2.2

Venting

Maintenance

connector

unit (MU-1)

Leakage tester (MB-155)

Figure 2.3

The water resistant cap must be attached to the electrical connector of the endoscope whenever the endoscope is immersed in reprocessing fluids. It is detached from the connector whenever the endoscope is used for patient procedures, being sterilized by ethylene oxide gas, or stored in an endoscope storage cabinet.

Always use a dry water resistant cap. Any water remaining inside the cap may cause damage to the endoscope.

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

11

Chapter 2 Function and Inspection of the Accessories for Reprocessing

Use the chain for water-resistant cap (MAJ-1119) to connect the water resistant cap to the endoscope. The water resistant cap can remain connected to the endoscope by the chain at all times (including during patient procedures, reprocessing, and storage of the endoscope).

Inspection

1.Confirm that the inside of the cap is dry and free from debris. Wipe with a dry, lint-free cloth if the inside of the cap is wet or if debris is detected.

2.Confirm that the seals inside the cap are free from scratches, cuts, and debris.

3.Check to ensure that the venting connector on the cap is not loose.

2.2Channel plug (MH-944)

Biopsy valve cap

Plug frame

 

 

Suction plug

 

Air/water plug

Figure 2.4

12

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

Chapter 2 Function and Inspection of the Accessories for Reprocessing

Function

The channel plug is used to plug the openings of the instrument channel port and the suction and air/water cylinders of the endoscope whenever the injection tube (MH-946) is used to flush the suction and air/water channels of the endoscope with reprocessing fluids.

Channel plug

Instrument channel port

Figure 2.5

When attached to the endoscope, the channel plug is designed to allow a small amount of fluid to exit from the openings of the endoscope. This enables reprocessing fluids to contact the endoscope openings.

Inspection

Confirm that the suction plug, air/water plug, and the biopsy valve cap of the channel plug are free from cracks, scratches, and debris.

The channel plug does not need to be cleaned, disinfected, or sterilized prior to its first use.

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

13

Chapter 2 Function and Inspection of the Accessories for Reprocessing

2.3Injection tube (MH-946)

Information

Suction channel tube

card

 

Connector plug

 

 

Suction

 

Air pipe port

channel port

 

Air/water

 

 

channel port

 

 

 

 

Air/water channel tube

Suction port

 

Filter tube

 

 

(including the filter mesh)

Filter mesh

Figure 2.6

Function

The injection tube is used to inject reprocessing fluids into the instrument channel, suction channel, and air/water channels of the endoscope. It is also used to flush air through these channels to expel fluids.

Syringe

Air/water channel tube

Suction channel tube

 

Suction port

 

 

Air pipe port

Basin

 

 

Reprocessing

Connector plug

fluids

 

Figure 2.7

Inspection

1.Confirm that all components of the injection tube are free from cracks, scratches, flaws, and debris (see Figure 2.6).

2.Confirm that the filter mesh is in the suction port of the injection tube.

14

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

Chapter 2 Function and Inspection of the Accessories for Reprocessing

3. Attach a clean 30 ml syringe to the suction channel port of the injection tube. With the suction port of the injection tube immersed in the water referred to in Section 3.2, withdraw the syringe plunger and confirm that the water is drawn into the syringe. Depress the plunger and confirm that the water is emitted from the suction channel tube of the injection tube. Confirm that the water is not emitted from the suction port when removing the suction port from the water.

Suction channel

Suction channel

tube

port

 

Syringe

 

Suction port

Figure 2.8

4. Move the syringe to the air/water channel port of the injection tube. With the suction port of the injection tube immersed in the water, withdraw the syringe plunger and confirm that the water is drawn into the syringe. Depress the plunger and confirm that the water is emitted from the air pipe port of the injection tube. Confirm that the water is not emitted from the suction port when removing the suction port from the water.

Suction

Air/water channel port

channel tube

 

 

Syringe

 

Suction port

Air pipe

 

port

 

Figure 2.9

The injection tube does not need to be cleaned, disinfected, or sterilized prior to its first use.

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

15

Chapter 2 Function and Inspection of the Accessories for Reprocessing

2.4Channel cleaning brush (BW-20T)

Metal tip

 

 

Shaft

 

Bristles

 

 

 

 

 

 

Brush head

Figure 2.10

Function

The channel cleaning brush is used to brush the inside of the instrument channel and suction channel of the endoscope, and the interior and openings of the suction valve (MH-443), the air/water valve (MH-438) and the biopsy valve (MB-358).

Suction cylinder

Instrument channel

Suction channel

Figure 2.11

16

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

Chapter 2 Function and Inspection of the Accessories for Reprocessing

Depress

Suction valve

Air/water valve

Biopsy valve (MB-358)

(MH-443)

(MH-438)

 

Figure 2.12

Inspection

1.Confirm that the brush head and the metal tip of the distal end are securely attached. Check the brush head for loose or missing bristles.

2.Check the bristles for damage. If the bristles are crushed, gently straighten them with your gloved fingertips.

3.Check for bends, scratches, and other damage to the shaft.

4.Visually check for debris on the shaft and/or the bristles of the brush head. If there is debris on the brush, immerse the brush in the water referred to in Section 3.2 and clean the brush until no debris is observed on the brush.

The channel cleaning brush does not need to be cleaned, disinfected, or sterilized prior to its first use.

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

17

Chapter 2 Function and Inspection of the Accessories for Reprocessing

2.5Suction cleaning adapter (MH-856)

Connecting end

Weighted end

Figure 2.13

Function

The suction cleaning adapter is used to aspirate reprocessing fluids through the instrument channel port of the endoscope.

Suction cylinder

Connecting end

Suction cleaning

adapter

Instrument

channel port

Weighted end

Suction pump

Figure 2.14

Inspection

Check for debris, cracks, scratches, and other damage.

The suction cleaning adapter does not need to be cleaned, disinfected, or sterilized prior to its first use.

18

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

Chapter 2 Function and Inspection of the Accessories for Reprocessing

2.6AW channel cleaning adapter (MH-948)

 

 

Information

Slider

Button

card

 

One-way valve

 

 

Piston

 

Figure 2.15

Function

During precleaning of the endoscope, the AW channel cleaning adapter is attached to the air/water cylinder of the endoscope. When the button of the adapter is depressed, the water in the water container is fed through the air/ water nozzle of the endoscope to clean the nozzle and air/water channels of the endoscope. Air is continuously fed through the air/water channels when the button is not depressed.

 

AW channel

Depress

 

cleaning adapter

 

 

Air/water

Air channel

 

nozzle

 

 

 

Water

Air/water cylinder

 

container

 

 

 

Figure 2.16

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

19

Chapter 2 Function and Inspection of the Accessories for Reprocessing

Inspection

Check for debris, cracks, scratches, and other damage.

The AW channel cleaning adapter does not need to be cleaned, disinfected, or sterilized prior to its first use.

2.7Single use channel cleaning brush (BW-201T)

Tip

Bristles

Caution sticker

 

 

Brush head

Shaft

Figure 2.17

Function

The single use channel cleaning brush is used to brush the inside of the instrument channel and suction channel of the endoscope and the interior and/or openings of the suction valve (MH-443), the air/water valve (MH-438), and the biopsy valve (MB-358).

Suction cylinder

Instrument channel

Suction channel

Figure 2.18

20

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

Chapter 2 Function and Inspection of the Accessories for Reprocessing

Depress

Suction valve

Air/water valve

Biopsy valve (MB-358)

(MH-443)

(MH-438)

 

Figure 2.19

Inspection

1.Remove the brush from its packaging just prior to use.

2.Confirm that the tip and brush head at the distal end are securely attached. Check the brush head for loose or missing bristles.

3.Check the bristles for any damage. If the bristles are crushed, gently straighten them with your fingertips.

4.Check for bends, scratches, and other damage to the shaft.

The single use channel cleaning brush does not need to be cleaned, disinfected, or sterilized prior to use.

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

21

Chapter 2 Function and Inspection of the Accessories for Reprocessing

2.8Single use channel-opening cleaning brush (MAJ-1339)

Brush head

Handle

Figure 2.20

Function

The single use channel-opening cleaning brush is used to brush the suction cylinder, the instrument channel port, the distal end, the forceps elevator, and the forceps elevator recess.

Suction

 

cylinder

Instrument

 

 

channel port

Figure 2.21

22

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

Chapter 2 Function and Inspection of the Accessories for Reprocessing

Forceps elevator

elevator

(a) Brushing when forceps elevator is lowered.

Forceps elevator recess

(b) Brushing when forceps elevator is raised.

Figure 2.22

Inspection

1.Remove the brush from its packaging just prior to use.

2.Check the brush head for loose or missing bristles.

3.Check the bristles for any damage. If the bristles are crushed, gently straighten them with your fingertips.

4.Check for bends, scratches, and other damage to the shaft.

The single use channel-opening cleaning brush does not need to be cleaned, disinfected, or sterilized prior to use.

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

23

Chapter 2 Function and Inspection of the Accessories for Reprocessing

2.9Single use combination cleaning brush (BW-412T)

Tip

Bristles

Shaft

Handle

Bristles

 

Channel cleaning

 

Channel-opening

 

 

brush part

 

cleaning brush part

 

Figure 2.23

Function

The channel cleaning brush part of the single use combination cleaning brush is used to brush the inside of the instrument channel and suction channel of the endoscope, and the interior and/or openings of the suction valve (MH-443), the air/water valve (MH-438), and the biopsy valve (MB-358). The channel-opening cleaning brush part of the single use combination cleaning brush is used to brush the suction cylinder, the instrument channel port, the distal end, the forceps elevator, and the forceps elevator recess of the endoscope.

Inspection

1.Remove the brush from its packaging just prior to use.

2.Confirm that the channel cleaning brush part and tip at the distal end are securely attached.

3.Check the channel cleaning brush and the channel-opening cleaning brush parts for loose or missing bristles.

4.Check the bristles of the channel cleaning brush and the channel-opening cleaning brush parts for any damage. If the bristles are crushed, gently straighten them with your fingertips.

5.Check for bends, scratches, and other damage to the shaft.

The single use combination cleaning brush does not need to be cleaned, disinfected, or sterilized prior to use.

24

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

Chapter 2 Function and Inspection of the Accessories for Reprocessing

2.10 Single use soft brush (MAJ-1888)

Brush head

Shaft

Handle

 

 

 

Bristles

 

Figure 2.24

Function

The single use soft brush is used to brush around the forceps elevator.

Inspection

1.Check the brush head for loose, missing bristles, and other damage. If the bristles are crushed and/or bent, gently straighten them with your fingers.

2.

3.

Check for bends, scratches, and other damage to the shaft.

Check for debris on the shaft and or in the bristles of the brush head.

The single use soft brush does not need to be cleaned, disinfected, or sterilized prior to use.

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

25

Olympus TJF Q180V User manual

Chapter 2 Function and Inspection of the Accessories for Reprocessing

Forceps elevator

Guidewire-locking groove

Groove

(a) Brushing when forceps elevator is lowered.

Forceps elevator

 

Groove

elevator

Forceps elevator

 

(b) Brushing when forceps elevator is raised.

Figure 2.25

26

EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL

Loading...
+ 88 hidden pages