KROHNE IFC-110 User Manual

4 (1)
KROHNE IFC-110 User Manual

© KROHNE 09/2003

Supplementary

Installation and

Operating Instructions

HART Communicator 275

Asset Management Solutions (AMS)

Process Device Manager (PDM)

IFC 110

KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 D-47058 Duisburg

1/15

Tel.: 0203-301 0 Fax: 0203-301 389 · e-mail: krohne@krohne.de

 

Supplementary Installation and Operating Instructions IFC110 HART

HC275, AMS, PDM

1

General Information

3

2

IDs and Revision numbers

5

3

Implementation Peculiarities

6

4

HART Communicator 275 (HC275)

6

 

4.1

Installation

6

 

4.2

Operating

6

5

Asset Management Solutions (AMS)

6

 

5.1

Installation

6

 

5.2

Operating

6

6

Process Device Manager (PDM)

7

 

6.1

Installation

7

 

6.2

Operating

7

KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 D-47058 Duisburg

2/15

Tel.: 0203-301 0 Fax: 0203-301 389 · e-mail: krohne@krohne.de

 

Supplementary Installation and Operating Instructions IFC110 HART

HC275, AMS, PDM

1 General Information

The IFC110 is a "four-wire" transmitter with 4...20mA current output and HARTâ capability. It is available with active current output or active/passive current output.

General characteristics of the IFC110 HARTâ interface:

·Multidrop Mode is supported

·Burst Mode is not supported

Electrical connection: Refer to sections 2.1, 2.2, 2.6 of the “Installation and Operating Instructions Electromagnetic flowmeters ALTOFLUX IFC 110 F” (KROHNE).

There are two ways of using the HARTâ communication:

a) As a point-to-point connection between the IFC110 and the HARTâ master equipment.

Point-to-Point Analog/Digital Mode

Primary Master

Terminal I

Terminal I

HART

 

Analog

Modem

 

 

 

 

 

HART

4 mA

20

 

 

 

³ 250 O

4...20 mA

IFC110

Addr. 0

Secondary Master

KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 D-47058 Duisburg

3/15

Tel.: 0203-301 0 Fax: 0203-301 389 · e-mail: krohne@krohne.de

 

Supplementary Installation and Operating Instructions IFC110 HART

HC275, AMS, PDM

b) As a multipoint connection (multidrop) with up to 15 devices (IFC110 or other HARTâ equipment).

Multidrop Mode

Primary Master

HART

Modem

 

HART

 

 

Power

³ 250 O

 

 

 

 

 

Supply

 

 

 

 

 

 

Terminal I

 

4 mA

4 mA

Terminal I

 

 

. . .

 

 

Secondary Master

Up to 15

 

 

 

Slaves

 

 

 

. . .

 

 

IFC110

 

IFC110

 

Addr. > 0

 

Addr. > 0

 

(passive Current Output)

 

(passive Current Output)

KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 D-47058 Duisburg

4/15

Tel.: 0203-301 0 Fax: 0203-301 389 · e-mail: krohne@krohne.de

 

Supplementary Installation and Operating Instructions IFC110 HART

HC275, AMS, PDM

If the IFC110 is equipped with a continously active current output a 'third wire' is needed to properly connect it together with two-wire loop powered devices in the same network.

Multidrop Mode (‘three-wire’)

(Connecting two-wire and four-wire devices in the same network)

Primary Master

 

HART

 

 

 

4 mA

 

4 mA

HART

 

two-wire

 

Modem

 

loop powered

 

 

 

devices

 

 

 

. . .

 

 

Addr. > 0

 

Addr. > 0

Power

HART

 

Terminal I

Supply

 

 

 

 

4 mA

4 mA

Terminal I

 

 

 

 

four-wire

 

 

 

active source

 

 

³ 250 O

devices

 

 

 

 

Secondary Master

 

. . .

 

 

IFC110

IFC110

 

 

 

Addr. > 0

 

 

 

Addr. > 0

 

 

 

2 IDs and Revision Numbers

The HART Device Descriptions described in this document have the following IDs and revision numbers:

Manufacturer ID:

69 (0x45)

Device Type:

243 (0xF3)

Device Revision:

2

DD Revision:

1

HART Universal Revision:

5

HC 275 OS Revision:

³ 4.9

AMS Version:

³ 5.0

PDM Version:

³ 5.1

For information about Transmitter Revisions and related Device Descriptions refer to the KROHNE HART Device List.

KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 D-47058 Duisburg

5/15

Tel.: 0203-301 0 Fax: 0203-301 389 · e-mail: krohne@krohne.de

 

Loading...