Philips M3535A, M3536A User manual

0 (0)
NBP and CO
2
Calibration Instructions
PHILIPS
M3535A/M3536A HeartStart MRx
ii
Notice
About This Edition
Printed in the USA
Publication number 453564067081.
The information in this document applies to the
HeartStart MRx, product version indicated below. This
information is subject to change without notice.
Philips shall not be liable for errors contained herein or
for incidental or consequential damages in connection
with the furnishing, performance, or use of this material.
Edition History
Copyright
Copyright © 2007
Koninklijke Philips Electronics N.V.
All rights are reserved. Permission is granted to copy and
distribute this document for your organization’s internal
educational use. Reproduction and/or distribution
outside your organization in whole or in part is
prohibited without the prior written consent of the
copyright holder.
Windows Media Player® is a registered trademark of
Microsoft Corporation.
Use of supplies or accessories other than those
recommended by Philips may compromise product
performance.
Medical Device Directive
The HeartStart MRx complies with the requirements of
the Medical Device Directive 93/42/EEC and carries the
0123
mark accordingly.
Manufacturer:
Philips Medical Systems
3000 Minuteman Road
Andover, MA USA 01810-1099
(978) 687-1501
Authorized EU-representative:
Philips Medizin Systeme Böblingen GmbH
Hewlett Packard Str. 2
71034 Böblingen
Germany
U.S. FCC and Industry Canada Radio Compliance:
Contains FCC ID: PQC-WMTS-MODULE
Operation of this equipment requires the prior coordination with a
frequency coordinator designated by the FCC for the Wireless Medical
Telemetry Service. This device complies with Part 15 of the FCC rules
and RSS-210 of Industry Canada. Operation is subject to the
following conditions:
This device may not cause harmful interference.
This device must accept any interference received, including
interference that may cause undesired operation.
Any changes or modifications to this equipment not expressly
approved by Philips Medical Systems may cause harmful radio
frequency interference and void your authority to operate this
equipment.
Canada EMC: ICES-001
China: After-Sales Service: Beijing MEHECO-PHILIPS Medical
Equipment Service Center.
After-Sales Service Address: No. 208, 2nd District, Wang Jing Li Ze
Zhong Yuan, Chao Yang District, Beijing.
Postal code: 100102.
Telephone: 010-64392415.
Registration number: SFDA(I)20043211207.
Product Standard number: YZB/USA 52-21.
Pub. Number Ed. S/W Version Print Date
453564067081
1 9.00 and earlier Sept., 2007
iii
Conventions Used in This Manual
These upgrade instructions contain the following conventions:
WARNING Warning statements describe conditions or actions that can result in personal injury or loss of life.
CAUTION Caution statements describe conditions or actions that can result in damage to the equipment or loss of
data.
NOTE Notes contain additional information on usage.
TIP Tips provide hands-on insight into servicing this product.
Text represents messages that appear on the display
[Softkey] represents softkey labels that appear on the display above or below the button
to which they correspond.
On-line viewing only
* Hypertext represents hypertext links, which will display as blue; click on the blue link to go
to that destination, then click on the blue destination to return.
Abbreviations
Name Abbreviation
HeartStart MRx Monitor/Defibrillator monitor/defibrillator
device
Carbon dioxide CO
2
End-tidal carbon dioxide EtCO
2
Non-invasive Blood Pressure NBP
iv

Introduction

1
Introduction
CAUTION These instructions are intended for users trained in the safe and proper HeartStart MRx operation.
HeartStart MRx NBP and EtCO
2
calibration is part of the HeartStart MRx maintenance routine that
also includes operational checks, lithium ion battery maintenance, etc. described in HeartStart MRx
Instructions for Use. NBP and EtCO
2
calibration is the only HeartStart MRx maintenance procedure
not covered in the Instructions for Use.
This document covers the following calibration procedures:
Yearly calibration (or every 10,000 cycles, whichever is sooner) of the Non-invasive Blood Pressure
(NBP) module
Yearly calibration (or every 4000 hours, whichever is sooner) of the End-tidal Carbon Dioxide
(EtCO
2
) module.
Use these quick links to access the calibration procedures:
* NBP Module Calibration
* EtCO
2
Module Calibration

HeartStart MRx Calibration Overview

Consider reviewing the “HeartStart MRx Calibration” instructional video included in your calibration
kit before proceeding with the calibration. To watch the DVD on your PC, you may need to install the
latest version of Windows Media Player
®
available for free download from Windows Media download
center
http://www.microsoft.com/windows/windowsmedia/download/.
The video is also available online at
http://theonlinelearningcenter.com/schtml/mrx/calibration/.
Perform calibration when prompted by the
CO2 Calibration Overdue and NBP Calibration
Overdue
inops. If a calibration is overdue, then the HeartStart MRx Op. Check fails with the code
Fail/D (Versions B.05 and above) or Fail/NC (Versions prior to B.05).
Regardless of your configuration settings, millimeters of mercury are the unit of measure for pressure in
the HeartStart MRx calibration. Use the conversion formulae from Ta b le 1 if necessary:

Accessing Service Mode

The calibration activities are accessed through the Service Mode. To access the Service Mode:
1 From the Main Menu select Other, then Service.
2 Enter the “27689” password when prompted by the device.
Table 1 Units of Pressure Conversion
1 unit = __ mmHg
1 kPa
=
7.5 mmHg
1 mb
=
.75 mmHg
1 psi
=
51.7 mmHg
1 atm.
=
760 mmHg
Example:
1 inHg
=
25.4 mmHg 30 inHg = 30 × 25.4 mmHg = 762 mmHg