Philips M3921A, M3922A, M3923A, M3924A, M3926A User manual

...
0 (0)
*
Philips A1 and A3 Patient Monitors
M3922A M3923A M3924A
A3 Models: M3926A
M3927A M3928A M3929A
Service Guide
*M3921-9000C
Part Number M3921-9000C
Printed in the U.S.A. February 2002
First Edition

About this Manual

Proprietary Information

This document contains proprietary information, which is protected by copyright. All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
Philips Medical Systems 3000 Minuteman Road Andover, MA 01810-1085 (978) 687-1501
Publication number M3921-9000C Printed in USA

Warranty

The information contained in this document is subject to change without notice.
Philips Medical Systems makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties or merchantability and fitness for Philips Medical Systems shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

Copyright

Printing History

Copyright © 2002 Philips Electronics North America Corporation
New editions of this document incorporate all material updated since the previous edition. Update packages may be issued between editions and contain replacement and additional pages to be merged by a revision date at the bottom of the page. Pages that are rearranged due to changes on a previous page are not considered revised.
The documentation printing date and part number indicate its current edition. The printing date changes when a new edition is printed. (Minor corrections and updates that are incorporated at reprint do not cause the date to change.) The document part number changes when extensive technical changes are incorporated.
This edition, M3921-9000C combines both A1 and A3 Service Guides into one book. There is no M3921-9000B book.
First Edition............................................................... February 2002
ii

Philips Software License Terms

ATTENTION: USE OF THE SOFTWARE IS SUBJECT TO THE PHILIPS SOFTWARE LICENSE

TERMS SET FORTH BELOW. USING THE SOFTWARE INDICATES YOUR ACCEPTANCE OF THESE LICENSE TERMS. IF YOU DO NOT ACCEPT THESE LICENSE TERMS, YOU MAY RETURN THE SOFTWARE FOR A FULL REFUND. IF THE SOFTWARE IS BUNDLED WITH ANOTHER PRODUCT, YOU MAY RETURN THE ENTIRE UNUSED PRODUCT FOR A FULL REFUND.
PHILIPS SOFTWARE LICENSE TERMS
The following License Terms govern your use of the accompanying Software unless you have a separate signed agreement with Philips Medical Systems.
License Grant. Philips Medical Systems grants you a license to Use one copy of the Software. "Use" means storing, loading, installing, executing or displaying the Software. You may not modify the Software or disable any licensing or control features of the Software. If the Software is licensed for "concurrent use", you may not allow more than the maximum number of authorized users to Use the Software concurrently.
About this Manual
Ownership. The Software is owned and copyrighted by Philips or its third party suppliers. Your license confers no title to, or ownership in, the Software and is not a sale of any rights in the Software. Philips’ third party suppliers may protect their rights in the event of any violation of these License Terms.
Copies and Adaptations. You may only make copies or adaptations of the Software for archival purposes or when copying or adaptation is an essential step in the authorized Use of the Software. You must reproduce all copyright notices in the original Software on all copies or adaptations. You may not copy the Software onto any public network.
No Disassembly or Decryption. You may not disassemble or decompile the Software unless Philips prior written consent is obtained. In some jurisdictions, Philips consent may not be required for limited disassembly or decompilation. Upon request, you will provide Philips with reasonably detailed information regarding any disassembly or decompilation. You may not decrypt the Software unless decryption is a necessary part of the operation of the Software.
Transfer. Your license will automatically terminate upon any transfer of the Software. Upon transfer, you must deliver the Software, including any copies and related documentation, to the transferee. The transferee must accept these License Terms as a condition to the transfer.
Termination. Philips Medical Systems may terminate your license upon notice for failure to comply with any of these License Terms. Upon termination, you must immediately destroy the Software, together with all copies, adaptations and merged portions in any form.
Export Requirements. You may not export or re-export the Software or any copy or adaptation in violation of any applicable laws or regulations.
U.S. Government Restricted Rights. The Software and any accompanying documentation have been developed entirely at private expense. They are delivered and licensed as
iii
About this Manual
"commercial computer software" as defined in DFARS 252.227-7013 (Oct. 1988), DFARS
252.211-7015 (May 1991) or DFARS 252.227-7014 (Jun. 1995), as a "commercial item" as defined in FAR 2.101(a), or as "Restricted computer software" as defined in FAR 52.227-19 (Jun. 1987)(or any equivalent agency regulation or contract clause), whichever is applicable. You have only those rights provided for such Software and any accompanying documentation by the applicable FAR or DFARS clause or the Philips standard software agreement for the product involved.
iv

Text Conventions

The following conventions for Notes, Cautions, and Warnings are used in this manual.
NOTE
A Note calls attention to an important point in the text.
A Caution calls attention to a condition or possible situation that could damage or destroy the product or the user’s work.
A Warning calls attention to a condition or possible situation that could cause injury to the user and/or patient.
About this Manual
Caution
WarningWarning

Explanation of Symbols

Symbols on products and packaging mean the following:
Battery Gauge icon
Defibrillator-proof type CF equipment
On/Standby button.
Alarm Limits icon.
Adjust Screen Contrast to display (monochrome) or invert the video to (color).
Adjust Heart Rate Tone volume.
Initiate NiBP measurement
v
About this Manual
T
Temperature
NiBP
ECG
SpO
2
SpO
2
Located on recorder. Produces a 20 Second Snapshot
Located on recorder. Produces a Continuous Print
Signal (ECG) Input
Signal (ECG) Output
Data input/output
Alternating Current
Direct Current
Temperature
Humidity
Altitude or atmospheric pressure
vi
About this Manual
Contains parts that may not be put into normal waste disposal but must be recycled or
dealt wi disposed as chemical waste
Fragile, handle with care
Keep dry
Consult instructions for use
Date of manufacture
Refer to User’s Guide
DC Lines
DC Input
Defib Sync.
RS-232 I/O
Class II Equipment
Power supply if for indoor use only
Silence Alarm
Equipotential Ground
vii
About this Manual
Recorder’s front-panel charging LED
Recorder’s front-panel communication LED
Ground Wire
viii
Contents
1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Manual Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Patient Monitor Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Introduction to External Recorder (M3925A) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Related Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2. Routine Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Periodic Safety And Functional Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Environmental Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3. Performance Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Test And Inspection Matrix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Equipment Needed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Performance Tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Battery Charge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Battery Performance Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Power-On Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Hardware and Software Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
SpO
2
Operation with an ECG Simulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Operation with a Respiration Simulator (A3 only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Verification of Pneumatic System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Operation with a Temperature Simulator (A1 only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Operation with a Temperature Simulator (A3 only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
General Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Serial Interface Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Printer Verification (For A1 Optional Printer M3925A) . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Ground Integrity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Electrical Leakage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4. Power-up Defaults Menu And Diagnostic Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Power-up Defaults Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Diagnostic Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 System Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 System A/D Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 NIBP Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Restoring Factory Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Contents ix
5. Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 How To Use This Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Who Should Perform Repairs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Replacement Level Supported . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Obtaining Replacement Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Generating an Error Log Printout (A1 only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Serviceable Hardware Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 Other Error Codes (All Monitors) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 Buttons/Wheels (All Monitors) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Display/Audible Tones (All Monitors) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59 Operational Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
6. Disassembly Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 How To Use This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 Disassembly Procedures (A1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Closed Case Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 Separation of Front and Rear Case Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 Front Case Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70 Rear Case Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 Main PCB Disassembly Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Disassembly Procedures (A3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Closed Case Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79 Separation of Front and Rear Case Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 Front Case Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82 Rear Case Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
7. Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91 Small Parts Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91 Top Level Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92 Front Case Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97 Rear Case Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103 Main PCB Assembly (A1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 Power Supply, Cables, and Miscellaneous Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109 Exchange Unit Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
8. Packing For Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
General Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113 Repacking In Original Carton . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113 Repacking In A Different Carton. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
x Contents
9. RS-232 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .116 Nurse Call . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
10. Training Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .117 Levels of Involvement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Preventative Maintenance Only . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117 Phone Support or Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .117
Training Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Essential Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Optional Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119 Support Strategies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .120
Unit exchange . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Bench repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .120 Theory of Operation and System Architecture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .121
System Overview - A1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Block Diagram - A1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Block Diagram - A1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
System Overview - A3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Block Diagram - A3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Block Diagram - A3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .129 Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130 Monitor Applications and Algorithms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Reference Documents and Other Sources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132 Preventative Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Index
Contents xi
xii Contents

Manual Overview

1

Introduction

This manual contains information for servicing the A1 and A3 patient monitors, subsequently referred to as the monitor throughout this manual. Only where there are differences in service procedures for the two monitor types is each monitor referred to specifically as A1 and A3. Only qualified service personnel should service this product. Before servicing the monitor, read the User Guide carefully for a thorough understanding of operation.
Introduction 1

Patient Monitor Description

Patient Monitor Description
The purpose and function of the patient monitor is to monitor: ECG; heart rate; noninvasive blood pressure (systolic, diastolic, and mean arterial pressures); functional arterial oxygen saturation; respiration rate (A3 only); and temperature for adult and pediatric patients in all hospital areas and hospital-type facilities. It can be used during hospital transport and in mobile, land-based environments, such as ambulances.
The physical and operational characteristics of the monitor are described in the User Guide.
The measurement parameters and features for each model are indicated below.
Model
NIBP SpO
M3921A Yes Yes No No No No Mono
M3922A Yes Yes Yes No No No Mono
M3923A Yes Yes No Yes No No Mono
M3924A Yes Yes Yes Yes No No Mono
M3925A Optional Standalone Recorder for A1
M3926A Yes Yes Yes Yes Yes
M3927A Yes Yes Yes Yes Yes
M3928A Yes Yes Yes Yes Yes
M3929A Yes Yes Yes Yes Yes
Measurement Parameters and Features
Temp ECG Resp Integral
2
A1 Monitor
A3 Monitor
Recorder
No
Yes
No
Yes
Color/ Mono
Mono
Mono
Color
Color
2 Introduction

Introduction to External Recorder (M3925A)

The recorder is an optional, standalone printer designed for use with the A1 patient monitor. The recorder communicates with the monitor using a null­modem cable connected between each device’s RS-232 connector.
The recorder contains an internal battery, which, when fully charged, will operate the recorder for 3 hours (typical, at 25°C, producing fifteen 20-second printouts per hour). The recorder can be connected to AC power using an external power supply. The recorder uses the same type of power supply as the A1 monitor, the PS-120V or PS-240V.
The recorder does not have an On/Off switch. The recorder can sense when it has an established communication link with the monitor. At that time, the green LINKED indicator on the front panel lights, indicating that the recorder is ready for operation. See the recorder’s User Guide for more information regarding its operation.
Introduction to External Recorder (M3925A)

Related Documents

To perform test and troubleshooting procedures and to understand the principles of operation and circuit analysis sections of this manual, you must know how to operate the monitor. Refer to the User Guide to understand the various sensors, ECG lead, blood pressure cuffs, and temperature probes that work with the monitor. Refer also to the user guide and individual directions for use that accompany these accessories.
Introduction 3
Related Documents
4 Introduction

Routine Maintenance

Cleaning

Warning Do not immerse the monitor or its accessories in liquid or
clean with caustic or abrasive cleaners. Do not spray or pour any liquid on the monitor or its accessories.
To clean the monitor, dampen a cloth with a commercial, nonabrasive cleaner and wipe the exterior surfaces lightly. Do not allow any liquids to come in contact with the power connector or switches. Do not allow any liquids to penetrate connectors or openings in the instrument. For cables, sensors and cuffs, follow the cleaning instructions in the directions for use that accompany these accessories.
2

Periodic Safety And Functional Checks

The monitor requires cleaning, battery maintenance and NiBP performance and verification check every two years. The following performance verification tests may be used following repair or during routine maintenance (if required by your local institution).
1. Inspect the exterior of the monitor for damage.
2. Inspect labels for legibility. If the labels are not legible, contact Philips’ Response Center or your local Philips representative.
3. Verify that the NiBP performs properly as described in “Verification of Pneumatic System” on page 21.

Battery

If the monitor has not been used for a long period of time, the battery will need charging. To charge the battery, connect the monitor to an AC outlet, or external DC supply in the case of the A3, as described in “Battery Charge” on page 13 or the “Setup and Use” chapter of the User Guide.
Note Storing the monitor for a long period without charging the battery
can degrade the battery capacity. A complete battery recharge requires 8 hours.
Chapter 2 - Routine Maintenance 5

Environmental Protection

The battery can be recharged while the monitor is in use, in which case, the battery requires 14 hours to be recharged.
The battery can require a full discharge/charge cycle to restore normal capacity.
If the monitor operates1 for less than one hour on battery power before the low battery alarm occurs, the battery should be conditioned.
If the same symptom persists even after the battery is conditioned and indicating a full charge, the battery should be replaced.
Refer to Chapter 6, “Disassembly Guide”.
Environmental Protection
Follow local governing ordinances and recycling plans regarding disposal or recycling batteries and other device components.
1. See “Battery Performance Test” on page 14 for typical battery operating times and conditions.
6 Chapter 2 - Routine Maintenance

Introduction

3

Performance Verification

This section discusses the tests used to verify performance following repairs or during routine maintenance. All tests can be performed without removing the covers of the monitor.
If the monitor fails to perform as specified in any test, repairs must correct the problem before the monitor is returned to the user.
Performance Verification 7

Test And Inspection Matrix

Test And Inspection Matrix
The following test map shows which tests are required in which situations.
Table 1 Test Map
Service Event (When
Performing…)
Installation Visual and Power On Tests
Repairs
A. Unit Exchange Visual and Power On Tests
B. Unit Opened Power On Test
C. NBP Pump replaced Power On Test
D. SpO
E. Front End Connectors replaced Power On Test
Module replaced Power On Test
2
Test Blocks Required (…Complete
these Tests)
Basic Pneumatic Leakage Test (BPL)
Ground Integrity Test
Leakage Current Test
NBP Tests:
Pneumatic Leakage and Inflation Rate
BPL Test
Tests:
SpO
2
Dynamic Operating Range and
LED Excitation
BPL Test
Performance test for the parameter that had the connector replaced.
F. Power Supply replaced (A3 only) Power On, BPL, and Safety tests
G. Component level repair on any
PC board
H. Main PC board replaced Power On Test
All software upgrades Power On
Preventative maintenance Power On and NBP Performance Tests
8 Performance Verification
Power On Test
All Performance Tests
All Safety Tests
All Performance tests except Battery Charge Test and Battery Performance Test
Table 2 Test and Inspection Reporting
Test And Inspection Matrix
1
Test
Visual: Inspect exterior of monitor for damage
Power-On Self-Test Displays Normal Monitoring
Basic Pneumatic Leakage Test (NBP test)
NiBP Accuracy Test 250 mmHg; ±5 mmHg
NiBP Leakage Test After 1 Minute at 250 mmHg,
NiBP Inflation Rate Test
NiBP Overpressure Test
Expected Results
No Visual Damage V:P or V:F
Screen Configuration and emits tones
After 1 minute at approximately 250 mmHg, pressure drops no more than 6 mmHg.
pressure drops no more than 6 mmHg
Monitor Reports NiBP = 280 mmHg in <6 seconds
Overpressure Deflation is activated at a value between 280 mmHg to 330 mmHg
What to Record
PO:P or PO:F
BPL: Px6
Where x6 =Pressure Drop in mmHg
PN:P/X1…X4 (or X5
PN:F/X1…X4 (or X5
Record Pressure in mmHg = X1
Record Pressure Drop in mmHg = X2
Record Time to 280 mmHg in Seconds = X3
Record value in mmHg at which Overpressure Deflation occurs =X4
2
3
)
3
)
NiBP Deflation Rate Test (A3 only)
Safety Performance Tests
Monitor reports NiBP >10 mmHg and <190 mmHg
All Safety Performance Tests are in range of Table 5 to Table 12
Record value in mmHg on monitor at 1 min = X5
S:P or S:F
1. Details of the tests are included later in this chapter.
2. When authorized Philips Medical Systems personnel service the instru­ment. The results are reported back to Philips. The collected data form a database to be used in product development. It is not necessary for hospi­tal personnel to report these results.
3. In the case of the NiBP Deflation Rate Test, applicable to the A3 only.
Performance Verification 9

Equipment Needed

Equipment Needed
The following table lists the equipment required for performance verification.
Table 3 Required Test Equipment
Required For
Equipment Description A1 A3
Digital multimeter (DMM)
Sensor extension cable M4787A
Finger clip sensor M4789A
Oxisensor® II adhesive sensor
ECG cable M3913A M3923A and
ECG electrodes Standard M3923A and
ECG leads M3914A (IEC) or M3915A
NiBP tubing M3918A
NiBP cuff 40401C
Pulse oximeter tester Nellcor Puritan Bennett SRC-2
ECG simulator Dynatech Nevada medSim
Fluke Model 87 or equivalent
D-25
(AAMI)
300 or equivalent
··
··
··
··
M3924A only
M3924A only
M3923A and M3924A only
··
··
··
··
·
·
·
NiBP simulator Bio-Tek BP Pump or
equivalent
Temperature simulator medSim 300 or equivalent M3922A and
Respiration simulator medSim 300 or equivalent
Safety analyzer Bio-Tek 601 Pro or equivalent
Stopwatch Manual or electronic
10 Performance Verification
··
·
M3924A only
··
··

Performance Tests

Battery Charge

Performance Tests
The battery charge and battery performance test should be performed before monitor repairs whenever the battery is suspected as being a source of the problems. All other tests can be used following repairs or during routine maintenance (if required by your local institution). Before performing the battery performance test, ensure that the battery is fully charged. (See “Battery Charge” below).
This section is written using factory-set power-up defaults. If your institution has pre configured custom defaults, those values display.
To fully charge the battery:
1. Connect the monitor to an AC power source using the proper power cord. For the A1, use the PS-120V or PS-240V external power supply and power cord.
2. • For the A1, verify that the EXTERNAL POWER indicator is
lit.
• For the A3, verify that the
BATTERY CHARGING/AC SOURCE
indicator is lit.
3. Charge the battery for at least 8 hours.
The battery can require a complete discharge/charge cycle to restore its normal capacity, depending on its previous usage.
4. To check for a full charge, perform the procedure described in “Battery Performance Test” on page 12.
Performance Verification 11

Battery Performance Test

Battery Performance Test
The the A1 and A3 mono monitors are specified to typically operate on battery power for a minimum of 4 hours, at 25°C, with no printing, and one NiBP measurement every 15 minutes. The A3 color monitor is specified to typically operate on battery power for a minimum of 3 hours, at 25°C, with no printing, and one NiBP measurement every 15 minutes. Before performing this test, ensure that the battery is fully charged (see “Battery Charge”).
1. Connect the Nellcor Puritan Bennett SRC-2 pulse oximeter tester to the monitor via the M4787A sensor cable.
2. Connect the NiBP simulator to the monitor via the M3918A tubing.
3. Set the SRC-2 switches as follows:
SWITCH POSITION
RATE 38
LIGHT LOW
MODULATION LOW
RCAL/MODE RCAL 63/LOCAL
4. Set the NiBP simulator to simulate a pressure setting of 120/80 mmHg and heart rate of 80 bpm.
5. Ensure that the monitor is not connected to AC power.
6. With the monitor turned off, press the ON/STANDBY button and verify that the battery icon appears at the bottom of the display after the power-on self­test is completed. The boxes in the battery icon should all be filled, indicating that the battery is charged.
7. Verify that the monitor is responding to the SpO audible alarm is sounding. Use the wheel to select the SpO permanently silence the SpO
8. Use the wheel to select the NiBP Menu and set the Automatic Measurement Interval to 15 minutes. Exit the menu and press the front panel NiBP button
to manually start the first NiBP measurement. Subsequent NiBP
measurements are taken automatically every 15 minutes.
audible alarm.
2
simulator signal and that the
2
Menu and
2
12 Performance Verification

Power-On Self-Test

Power-On Self-Test
9. The A1 and A3 mono monitors must operate for at least 4 hours before they automatically power down due to low battery condition. The A3 color monitor must operate for at least 3 hours before the monitor automatically powers down due to low battery condition.
10. Verify that the low battery alarm occurs 15-30 minutes before the battery fully discharges.
11. Allow the monitor to operate until it automatically powers down due to low battery condition. Verify that the audible alarm sounds when the monitor automatically shuts down. Press the Alarm Silence button to terminate this audible alarm.
12. If monitor passes this test, immediately recharge the battery. (See “Battery Charge” on page 11).
1. Connect the monitor to an AC power source. For the A1, use the PS-120V or PS-240V power supply and power cord. For the A3, use the proper power cord supplied.
2. • For the A1, verify that the EXTERNAL POWER indicator is
lit.
• For the A3, verify that the
BATTERY CHARGING/AC SOURCE
indicator is lit.
3. Do not connect any input cables to the monitor.
4. Observe the monitor front panel. With the monitor off, press the ON/
STANDBY button . The monitor must perform the following sequence:
a. The A1 emits a beep.
The A3 emits three consecutively higher pitched beeps.
b. A few seconds later, the display backlight illuminates, but the display is
blank.
c. The version numbers of the boot and operational software display in the
lower left corner of the display.
d. A beep signals the end of the power-on self-test. Power-on self-test takes
approximately 10 seconds to complete.
e. Upon successful completion of the power-on self-test, the display is in
normal monitoring screen configuration. No vital-sign numeric values or waveforms are displayed.
Performance Verification 13

Hardware and Software Tests

Hardware and Software Tests
Hardware and software testing includes the following tests applicable to the indicated models in the series:

SpO2 Testing A1, A3 - all models

Operation with an ECG Simulator A1 - M3923A, M3924A
Operation with a Respiration Simulator A3 - all models
Verification of Pneumatic System A1, A3 - all models
Operation with a Temperature Simulator A1 - M3922A, M3924A
Test Applies to Model(s)
A3 - all models
A3 - all models
SpO
Testing
2
Alarms and Alarm
Silence
General Operation A1, A3 - all models
SpO2 testing includes the following tests:

n Alarms and Alarm Silence

n Heart Rate Tone Volume Control
n Dynamic Operating Range
n LED Excitation Test
1. Connect the SRC-2 pulse oximeter tester to the M4787A sensor cable and connect the cable to the monitor.
2. Set SRC-2 as follows:
SWITCH POSITION
RATE 38
LIGHT LOW
MODULATION OFF
RCAL/MODE RCAL 63/LOCAL
3. Press the ON/STANDBY button to turn the monitor on.
4. After the normal power-up sequence, verify that the SpO
14 Performance Verification
% display initially
2
Hardware and Software Tests
indicates zero or is blank. The pulse bar can occasionally indicate a step change as the monitor is in the pulse search mode.
5. Move the modulation switch on the SRC-2 to LOW.
6. Verify the following monitor reaction:
a. The pulse bar begins to track the artificial pulse signal from the SRC-2.
b. Initially, zero is displayed in the SpO
frame, or it is blank.
2
c. After about 10 to 20 seconds, the monitor displays saturation and heart rate
as specified by the tester. Verify that the values are within the following tolerances:
Oxygen Saturation Range 79% to 83%
Heart Rate Range 35 to 41 bpm
d. The audible alarm sounds and both the SpO2% and HEART RATE displays
flash, indicating both parameters have violated the default alarm limits.
e. The heart rate tone is heard. For this test, the user should set the heart rate
tone source to SpO
7. Press the ALARM SILENCE button on the front panel of the monitor.
The audible alarm is temporarily silenced.
8. Verify the following:
a. The audible alarm remains silenced.
b. The slashed bell icon appears in each numeric frame on the display.
c. The SpO
% and HEART RATE displays continue flashing.
2
from the Heart Rate Menu.
2
d. The heart rate tone remains audible.
e. The audible alarm returns in approximately 60 seconds.
Heart Rate Tone
Volume Control
1. Connect the SRC-2 pulse oximeter tester to the M4787A sensor cable and connect the cable to the monitor.
2. Set SRC-2 as follows:
POSITION
SWITCH
A1 A3
RATE 38 38
LIGHT LOW LOW
MODULATION LOW OFF
RCAL/MODE RCAL 63/LOCAL RCAL 63/LOCAL
Performance Verification 15
Hardware and Software Tests
3. Power on the monitor and verify that the SpO2 and heart rate values are
correctly displayed.
4. Press the ALARM SILENCE button on the front panel of the monitor
to temporarily silence the audible alarm.
5. Verify that the heart rate tone source, found in the Heart Rate Menu, is set to
SpO
.
2
6. Press the Heart Rate Tone VOLUME button on the front panel of the
monitor. Within 3 seconds of having pressed the button, rotate the wheel clockwise and verify that the beeping heart rate tone sound level increases.
7. Rotate the wheel counterclockwise and verify that the beeping heart rate tone decreases until it is no longer audible. Rotate the wheel clockwise to return the beep volume to a comfortable level.
3 seconds after the last button-press or rotation of the wheel, function of the wheel reverts to moving the highlight on the display screen.
Dynamic Operating
Range
The following test sequence verifies proper monitor operation over a range of input signals.
1. Connect the pulse oximeter tester to the monitor and turn the monitor on.
2. Place the SRC-2 pulse oximeter tester in the RCAL 63/LOCAL mode.
3. Set the SRC-2 as indicated in below. Verify that the monitor readings are within the indicated tolerances. Allow the monitor several seconds to stabilize the readings.
Table 4 SRC 2 Settings and Monitor Indications
SRC-2 Settings Monitor Indications
RATE LIGHT MODULATION SpO2 Pulse Rate
38 HIGH2 LOW
112 HIGH1 HIGH
201 LOW LOW
201 LOW HIGH 79-83* 195-207*
An * indicates values that produce an alarm. Press the ALARM SILENCE button to temporarily silence the audible alarm.
For the pulse rate setting of 201 bpm, the pulse rate tolerance of 195 to 207 bpm is greater than the ±3 bpm accuracy specification of the monitor, due to the performance characteristics of the SRC-2 tester.
79-83* 35-41*
79-83* 109-115
79-83* 195-207*
4. Turn the monitor off.

LED Excitation Test

16 Performance Verification
This procedure uses normal system components to test circuit operation. A
Hardware and Software Tests
Nellcor Puritan Bennett Oxisensor II adhesive sensor, model D-25, is used to examine LED intensity control. The red LED is used to verify intensity modulation caused by the LED intensity control circuit.
1. Connect a M4787A sensor extension cable to the monitor.
2. Connect a D-25 sensor to the sensor extension cable.
3. Press the ON/STANDBY button to turn the monitor on.
4. Leave the sensor open with the LED and photo detector visible.
5. After the monitor completes its normal power-up sequence, verify that the sensor LED is brightly lit.
6. Slowly move the sensor LED in proximity to the photo detector element of the sensor. Verify, as the LED approaches the optical sensor, that the LED intensity decreases.
7. Open the sensor and notice that the LED intensity increases.
8. Repeat step 6 and the intensity again decreases. This variation is an indication that the micro-processor is in proper control of LED intensity.
9. Turn the monitor off.

Operation with an ECG Simulator

1. With the monitor off, connect the ECG leads to the appropriate jacks on the ECG tester.
2. Connect the leads to the ECG cable.
3. Connect the cable to the ECG input port on the monitor.
4. Set the ECG simulator as follows:
Heart rate: 30 bpm
Amplitude: 1 millivolt
Lead select: II
Wave Type: Normal sinus rhythm
Patient Type: Adult mode
Note
The accuracy of the monitor’s ECG measurements is ±5 bpm. In the procedure below, add the tolerance of the simulator to the acceptable range of readings.
5. Press On/Standby button to turn monitor on.
Performance Verification 17
Hardware and Software Tests
6. After normal power-up sequence, verify the following monitor reactions:
a. After at least five heartbeats, the monitor displays a heart rate of 30 ±5 bpm.
b. The audible alarm sounds and the
indicating heart rate is below the default lower alarm limit.
7. Press the ALARM SILENCE button. Verify that the audible alarm is
silenced.
8. Increase the heart rate setting on the ECG simulator to 240 bpm.
9. After at least five heartbeats, verify that the monitor displays a heart rate of 240 ±5 bpm.
10. Verify that the audible alarm sounds and the indicating that the heart rate is above the default upper alarm limit.
11. Press ALARM SILENCE button to silence alarm.
12. Decrease the heart rate setting on ECG simulator to 120 bpm.
13. After at least five heartbeats, verify that the monitor displays a heart rate of 120 ±5 bpm.
14. Disconnect the LL lead from ECG simulator.
15. Verify that the
displayed in sounds.
16. Reconnect the LL lead to ECG simulator. Verify that the message no longer appears and audible alarm is silenced.
17. Repeat steps 14 through 16 for LA and RA leads.
18. Turn the monitor off.
Leads Off alarm message appears, three dashes are
HEART RATE display, and the low priority audible alarm
HEART RATE display flashes,
HEART RATE display flashes,
Leads Off alarm

Operation with a Respiration Simulator (A3 only)

1. With the monitor off, connect the ECG leads to the appropriate jacks on the respiration simulator.
2. Connect the ECG leads to the ECG cable.
3. Connect the cable to the ECG input port on the monitor.
Note
The accuracy of A3 measurements is ±3 breaths per minute. In the procedure below, add the tolerance of the simulator to the acceptable range of readings.
4. Set the simulator for a respiration rate of 120 breaths per minute.
5. Press the ON/STANDBY button to turn the monitor on.
18 Performance Verification
Loading...
+ 120 hidden pages