INSTRUCTIONS
ENDOSCOPIC FLUSHING PUMP
OFP
CAUTION
Federal (USA) law restricts this device to sale
by or on the order of a physician. |
0086 |
|
IMPORTANT
Please read this manual carefully before attempting to use the Olympus OFP as it contains important information on the proper care, handling and use of the equipment. In addition, read the manuals of any other units (endoscope, light source, etc.) that form part of the system.
The safety and performance of an endoscopic system depends not only on the endoscope but also on the ancillary equipment used with it. Ensure any ancillary equipment is compatible with the endoscope and other equipment used. These instructions should be retained for reference during the life of the product. If you have any questions concerning the material contained in this manual, please contact your Olympus representative or nearest Olympus office.
INTENDED USE
The Olympus OFP is a peristaltic flushing pump intended for use in medical facilities under the direction of a trained physician. The OFP has been designed to be used with Olympus EVIS EXERA gastrointestinal videoscopes having an auxiliary water inlet as part of an Olympus endoscopic imaging system to facilitate washing of gastric and colonic mucosa to remove blood, faeces and other organic matter, enabling improved visualisation, diagnosis and therapy during endoscopic procedures.
Do not use the equipment for any purpose other than its intended application.
WARNING SYMBOLS USED ON DEVICE
Refer to instructions
|
|
SIGNAL WORDS |
WARNING: |
D Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. |
|
CAUTION: |
D |
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. |
|
|
It may also be used to alert against unsafe practices or potential equipment damage. |
NOTE: |
D |
Indicates additional, helpful information. |
CONTENTS |
|
|
|
|
PAGE |
1 |
STANDARD SET & FEATURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . 1 |
2 |
OPERATING PRECAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . 3 |
3 |
INSTRUCTIONS FOR USE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . 4 |
4 |
STEAM STERILIZATION (AUTOCLAVING) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . 9 |
5 |
CARE AND STORAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. 10 |
6 |
MAINTENANCE AND REPAIR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. 10 |
7 |
SPARES & ACCESSORIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. 12 |
8 |
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. 12 |
9 |
TECHNICAL DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. 13 |
10 |
ENVIRONMENTAL PROTECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. 13 |
11 |
TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. 14 |
1 STANDARD SET & FEATURES
1.1Standard set
Unpack the OFP and associated items from the packaging. Refer to Figure 1-1 below and confirm that all items in the standard set are present. Contact your Olympus service centre or nearest Olympus office if any parts are damaged or missing.
FIGURE 1-1
1
2
4
|
5 |
3 |
|
|
|
Contents |
|
Qty |
|
|
|
|
|
|
1 |
Olympus Flushing Pump OFP . . . . . . . . . . . . . . . |
. . . . . . . . . . . . . . . 1 |
2 |
Fluid container (supplied non--sterile) . . . . . . . . . . . |
. . . . . . . . . . . . . . 1 |
3 |
Irrigation tube (supplied non--sterile) . . . . . . . . . . . . |
. . . . . . . . . . . . . . 2 |
4 |
Footswitch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . . . . . . 1 |
5 |
Filter (supplied non--sterile) . . . . . . . . . . . . . . . . . . |
. . . . . . . . . . . . . 10 |
6 |
Power cable (not shown) . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . . . . . . 1 |
7 |
Instructions (not shown) . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . . . . . . 1 |
1 |
FLUSHING PUMP OFP |
1.2 |
Features |
|
|
|
|
|
Autoclavable |
|
|
|
irrigation tube |
|
|
|
Autoclavable 1 litre |
|
|
|
fluid container |
|
ON/OFF |
|
|
|
power switch |
|
|
|
Flow control |
|
Fluid container |
|
|
strap |
|
|
|
Label 1 |
|
|
Fluid container |
|
|
|
|||
|
|
|
||
Label 2 |
|
holder |
||
|
Peristaltic pump |
|||
|
||||
Footswitch |
||||
|
||||
connection port |
|
|||
|
|
Rear panel |
FIGURE 1-2 |
|
|
|
|
Remote control cable connection
IEC power receptacle |
Label 3 |
(with integral fuses) |
Potential equalisation terminal
1.3Labels
|
|
|
|
|
|
Label 1 -- Refer to Instructions |
|
|
|
|
|
|
|
|
Label 2 |
-- Footswitch connection port |
|
|
|
|
|
|
|
|
||
|
|
|
|
|
|
Label 3 |
-- Rating plate |
|
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
||
|
|
|
|
|
|
Label 4 |
-- “USE BY” Indicates the last month during which the device is intended to be used. |
|
|
|
|
|
|
|
|
||
|
2010--06 |
|
||||||
|
|
eg: June 2010 becomes 2010--06 |
|
|||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
(label 4 is affixed to the irrigation tube). |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
FLUSHING PUMP OFP |
2 |
2 OPERATING PRECAUTIONS
2.1Explosion hazard -- never install or use the OFP within the zone of risk of flammable gases.
2.2The OFP should only be used in a medical facility under the direction of a trained physician.
2.3The operator must assess the condition of the patient and use clinical judgement to set the flow rate from the pump to a suitable level to avoid patient trauma. The flow control should always be set to MIN at the start of the procedure and be increased progressively to a level commensurate with the clinical condition of the patient and degree of washing required.
Note that when the flow control is set to MAX, the flow rate through the washing channel of the endoscope will be in excess of that which can be delivered when using a 50ml syringe.
2.4Use only in conjunction with Olympus EVIS EXERA series endoscopes having an auxiliary water inlet.
2.5Only use the Olympus auxiliary water tube (MAJ--855) and irrigation tube supplied to connect the OFP to the endoscope, otherwise patient safety may be compromised or damage to the pump may occur.
2.6Note that the luer lock connector on the MAJ--855 which attaches to the OFP filter includes a one--way valve to prevent backflow -- do NOT use the MAJ--855 without this connector in place, or irrigation tube and filter, container contamination and damage may result.
2.7The irrigation tube and filter MUST be autoclaved before use.
2.8The irrigation tube and filter MUST be replaced daily -- failure to do so may lead to patient infection and a reduction in equipment performance.
2.9Reprocess the MAJ--855 following each use according to the manufacturer’s instructions supplied with the product.
2.10Check the “USE BY” date of the tubing as indicated on the packaging label. Do not use any tubes beyond the specified “USE BY” date.
2.11Never allow the fluid container to run dry, otherwise air will be supplied to the patient.
2.12To prevent free--flow of fluid to the patient, do not open the pump cover when the OFP is connected to an endoscope which is inserted into the patient and ensure the OFP is positioned below patient level.
2.13If the filter becomes blocked or the irrigation tube from the pump to the fluid container is blocked or kinked, the flow rate will be noticeably reduced. Check the tube for kinks or blockages and the filter for contamination, replace if necessary.
2.14Should the OFP be operated with a blocked endoscope or filter, damage to the irrigation tube may occur. In this event, care must be taken when disconnecting the fluid path as the system may be pressurised. To depressurise the system, the pump head should be opened to allow fluid back into the bottle. This must only be carried out with the endoscope removed from the patient. The filter and irrigation tube MUST be replaced before the unit is reused.
2.15Always switch off the pump before fitting or replacing the irrigation tube.
2.16When the irrigation tube has been correctly installed and the pump operated, do not disturb the position of the irrigation tube eg: by opening the pump head, otherwise the irrigation tube must be replaced. Failure to do so may lead to a reduction in equipment performance.
2.17Periodically inspect the MAJ--855 for obvious signs of wear or damage and replace if necessary or as recommended in the manufacturers instructions.
2.18Periodically inspect the fluid container for obvious signs of wear or damage and replace if necessary.
2.19Always empty the fluid container before moving the equipment. Care should be taken to obtain a firm grip underneath the body of the equipment before moving. The fluid container holder must never be used as a hand hold or carrying handle.
2.20Store and use the OFP within the environmental conditions described in Section 8; failure to do so may lead to equipment malfunction or failure.
2.21Only use replacement parts as specified in Section 7. Failure to do so may lead to equipment malfunction or failure.
3 |
FLUSHING PUMP OFP |