All information set forth in this document, all rights to such information, any and all inventions disclosed herein and any patents
that might be granted by employing the materials, methods, techniques or apparatus described herein are the exclusive property of
Metronics Incorporated, Bedford, New Hampshire.
No part of this document may be reproduced, stored in a retrieval
system, or transmitted in any form or by any means, electronic,
mechanical, photocopying, recording, or otherwise, without the
prior permission of Metronics Incorporated. The information contained herein is designed only for use with the Quadra-Chek 100
Digital Readout. Metronics Incorporated is not responsible for
any use of this information as applied to any other apparatus.
Disclaimer
Trademarks
The information contained in this document is subject to change
without notice. Metronics Incorporated assumes no responsibility or liability for any errors or inaccuracies contained herein, or
for incidental or consequential damage in connection with the
furnishing, performance, or use of this guide.
Metronics Inc. shall not be liable to the purchaser of this product
or third parties for damages, losses, costs, or expenses incurred by
the purchaser or third parties as a result of: accident, misuse, or
abuse of this product or unauthorized modifications, repairs, or
alterations to this product, or failure to strictly comply with
Metronics Incorporated’s operating and maintenance instructions.
Metronics, Quadra-Chek, Quadra-Chek 100, and QC100 are registered trademarks of Metronics Incorporated.
Other product names used herein are for identification purposes
only and may be trademarks of their respective owners. Metronics
Incorporated disclaims any and all rights to those marks.
Page 5
Safety and
Maintenance
Considerations
General safety precautions must be followed when operating the
system. Failure to observe these precautions could result in damage to the equipment, or injury to personnel. It is understood
that safety rules within individual companies vary. If a conflict
exists between the material contained in this guide and the rules
of a company using this system, the more stringent rules should
take precedence.
Although Metronics strongly recommends that you read all parts
of this guide, it is essential for reliable and safe operation to read
the first portions up to and including Chapter 2: Operation, prior
to operating the QC100.
Location and Mounting
Cleaning
Electrical
The QC100 must be placed or mounted on a stable, reliable surface. If the QC100 should fall, it could become seriously damaged
and more importantly could cause injuries to the user. For mounting instructions, refer to the Getting Started section of Chapter 2:
Operation, for the mounting bolt hole pattern.
Use only a cloth dampened with water and a mild detergent for
cleaning the exterior surfaces. Never use abrasive cleaners, and
never use strong detergents or solvents. Only dampen the cloth,
do not use a cleaning cloth that is dripping wet.
Do not allow the power cord to be located such that it can be
walked on or create a tripping hazard.
WW
ARNINGARNING
W
ARNING
WW
ARNINGARNING
Unplug the QC100 from the electrical outlet before cleaning.Unplug the QC100 from the electrical outlet before cleaning.
Unplug the QC100 from the electrical outlet before cleaning.
Unplug the QC100 from the electrical outlet before cleaning.Unplug the QC100 from the electrical outlet before cleaning.
WW
ARNINGARNING
W
ARNING
WW
ARNINGARNING
The QC100 is equipped with a 3-wire power plug that includesThe QC100 is equipped with a 3-wire power plug that includes
The QC100 is equipped with a 3-wire power plug that includes
The QC100 is equipped with a 3-wire power plug that includesThe QC100 is equipped with a 3-wire power plug that includes
a separate ground connection. Always connect the powera separate ground connection. Always connect the power
a separate ground connection. Always connect the power
a separate ground connection. Always connect the powera separate ground connection. Always connect the power
plug to a 3-wire grounded outlet. The use of 2-wire powerplug to a 3-wire grounded outlet. The use of 2-wire power
plug to a 3-wire grounded outlet. The use of 2-wire power
plug to a 3-wire grounded outlet. The use of 2-wire powerplug to a 3-wire grounded outlet. The use of 2-wire power
plug adapters or any other connection accessories that removeplug adapters or any other connection accessories that remove
plug adapters or any other connection accessories that remove
plug adapters or any other connection accessories that removeplug adapters or any other connection accessories that remove
the third grounded connection create a safety hazard andthe third grounded connection create a safety hazard and
the third grounded connection create a safety hazard and
the third grounded connection create a safety hazard andthe third grounded connection create a safety hazard and
should not be permitted. If a 3-wire grounded outlet is notshould not be permitted. If a 3-wire grounded outlet is not
should not be permitted. If a 3-wire grounded outlet is not
should not be permitted. If a 3-wire grounded outlet is notshould not be permitted. If a 3-wire grounded outlet is not
available, ask your electrician to provide one.available, ask your electrician to provide one.
available, ask your electrician to provide one.
available, ask your electrician to provide one.available, ask your electrician to provide one.
Page 6
General maintenanceUnplug the QC100 from the wall outlet and seek the assistance of
a qualified service technician if:
• The power cord is frayed or damaged or the power plug is
damaged
• Liquid is spilled or splashed onto the enclosure
• The QC100 has been dropped or the exterior has been damaged
• The QC100 exhibits degraded performance or indicates a
need for service some other way
Page 7
Backup battery
charging
The QC100 utilizes an internal battery to maintain system configuration settings during power interruptions. A battery circuit
charges this backup battery during periods of normal operation.
When the QC100 is used on a daily basis, the backup battery will
be adequately charged to maintain configuration settings during
power outages. However, over extended periods of nonuse, the
battery might lose the charge necessary to maintain configuration settings. To keep the backup battery charged over extended
periods of nonuse, apply power to the QC100 and leave it turned
on for a minimum of 2 days each month.
FCC compliance
statement
FCC Rule NP15R Rev. 23, June, 1989
This equipment has been tested and found to comply with the
limits for a Class A digital device, pursuant to Part 15 of the FCC
Rules. These limits are designed to provide reasonable protection
against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses, and
can radiate radio frequency energy and, if not installed and used
in accordance with the instructions in this guide, may cause harmful
interference to radio communication. Operation of this equipment in a residential area is likely to cause harmful interference, in
which case the user will be required to correct the interference at
his own expense.
Shielded cables must be used with this unit to ensure compliance
with Class A FCC limits. The connection of a nonshielded equipment interface cable to this equipment will invalidate the FCC Certification of this device and may cause interference levels which
exceed the limits established by the FCC for this equipment. It is
the responsibility of the user to obtain and use a shielded equipment interface cable with this device. Do not leave cables connected to unused interfaces. Changes or modifications not expressly approved by the manufacturer could void the user’s authority to operate the equipment.
FF
or Canadian Users:or Canadian Users:
F
or Canadian Users:
FF
or Canadian Users:or Canadian Users:
This Class “A” digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe “A” respecte toutes les exigences
du Reglement sur le matériel brouilleur du Canada.
The material in this guide is divided into 5 chapters ranging from
introductory material and basic operation to more advanced topics including system configuration. If you are an advanced user of
digital readouts, feel free to skim the material contained in the
first two chapters of this guide, but do not begin to operate the
readout unless you’re confident that you understand the essential
information contained in these chapters.
After an initial review, this guide will become an invaluable reference source that should be used routinely to answer questions
and ensure smooth continued operation of the readout. The comprehensive table of contents and index facilitate rapid access to
specific information contained within the guide.
This user’s guide is intended for operators, technicians, supervisors and dealer representatives.
NOTENOTE
NOTE
NOTENOTE
While system configuration and set up instructions areWhile system configuration and set up instructions are
While system configuration and set up instructions are
While system configuration and set up instructions areWhile system configuration and set up instructions are
provided in provided in
provided in
provided in provided in
intended only for technically qualified supervisors and dealerintended only for technically qualified supervisors and dealer
intended only for technically qualified supervisors and dealer
intended only for technically qualified supervisors and dealerintended only for technically qualified supervisors and dealer
representatives. Operators should not attempt to alter therepresentatives. Operators should not attempt to alter the
representatives. Operators should not attempt to alter the
representatives. Operators should not attempt to alter therepresentatives. Operators should not attempt to alter the
configuration of the QC100.configuration of the QC100.
configuration of the QC100.
configuration of the QC100.configuration of the QC100.
Chapter 1: Overview
An overview of QC100 system including specifications
Chapter 4: System SetupChapter 4: System Setup
Chapter 4: System Setup
Chapter 4: System SetupChapter 4: System Setup
, this information is, this information is
, this information is
, this information is, this information is
Chapter 2: Operation
Detailed descriptions of controls and indicators, and instructions for measuring, constructing and creating features
Chapter 3: Communications
Serial (RS232) and parallel communication with printers and
IBM-compatible personal computers
i
Page 10
Introduction
Chapter 4: System setup
System configuration instructions for technically qualified supervisors and dealer representatives
Chapter 5: Options
Descriptions of QC100 series options
Conventions
Basic terms
Measurement polarity
Measurement resolution in
the guide
The terms
refer to the Quadra-Chek 100 digital readout. The term
tortor
tor is used to refer to any device intended to measure parts in 2
tortor
dimensions. The term
portion of measuring machines.
It is assumed in all discussions of measurements that:
• Moving crosshairs to the right• Moving crosshairs to the right
• Moving crosshairs to the right causes measure counts to
• Moving crosshairs to the right• Moving crosshairs to the right
• Moving crosshairs up • Moving crosshairs up
• Moving crosshairs up causes measure counts to increase in a
• This count direction can easily be reversed if necessary.
• •
In most cases the display resolution shown on screens in this guide
will be one ten-thousandths of an inch (0.0001) or two microns
metric (0.002). The display resolutions shown in this manual are
not meant to suggest resolutions that should be set on your system for the instrument being used, they are only examples.
Quadra-Chek 100Quadra-Chek 100
Quadra-Chek 100,
Quadra-Chek 100Quadra-Chek 100
increase in a
positive directionpositive direction
positive direction; 1 to 2 to 3 and so on.
positive directionpositive direction
to increase in a
3 degrees and so on.
Reversing the count direction is discussed later in Chapter 4:
System Setup.
positive directionpositive direction
positive direction; 1 to 2 to 3 and so on.
positive directionpositive direction
positive directionpositive direction
positive direction; 1 degree to 2 degrees to
positive directionpositive direction
QC100QC100
QC100 and
QC100QC100
crosshairscrosshairs
crosshairs is used to refer to the targeting
crosshairscrosshairs
systemsystem
system used in this guide
systemsystem
compara-compara-
compara-
compara-compara-
ii
Page 11
Quadra-Chek® QC100
Icons
Type styles
Accuracy and precision
This guide uses the following icons to highlight information:
The raised hand icon indicates a warning regarding a situationThe raised hand icon indicates a warning regarding a situation
The raised hand icon indicates a warning regarding a situation
The raised hand icon indicates a warning regarding a situationThe raised hand icon indicates a warning regarding a situation
or condition that could lead to personal injury or death. Yor condition that could lead to personal injury or death. Y
or condition that could lead to personal injury or death. Y
or condition that could lead to personal injury or death. Yor condition that could lead to personal injury or death. Y
should not proceed until you read and thoroughly understandshould not proceed until you read and thoroughly understand
should not proceed until you read and thoroughly understand
should not proceed until you read and thoroughly understandshould not proceed until you read and thoroughly understand
the warning message. Wthe warning message. W
the warning message. W
the warning message. Wthe warning message. W
type.type.
type.
type.type.
The exclamation point icon indicates a caution regarding aThe exclamation point icon indicates a caution regarding a
The exclamation point icon indicates a caution regarding a
The exclamation point icon indicates a caution regarding aThe exclamation point icon indicates a caution regarding a
situation or condition that could lead to equipmentsituation or condition that could lead to equipment
situation or condition that could lead to equipment
situation or condition that could lead to equipmentsituation or condition that could lead to equipment
malfunction or damage. Ymalfunction or damage. Y
malfunction or damage. Y
malfunction or damage. Ymalfunction or damage. Y
read and fully understand the caution message. Cautionread and fully understand the caution message. Caution
read and fully understand the caution message. Caution
read and fully understand the caution message. Cautionread and fully understand the caution message. Caution
messages are shown in bold type.messages are shown in bold type.
messages are shown in bold type.
messages are shown in bold type.messages are shown in bold type.
The note icon indicates additional or supplementaryThe note icon indicates additional or supplementary
The note icon indicates additional or supplementary
The note icon indicates additional or supplementaryThe note icon indicates additional or supplementary
information about an activity or concept. Notes are shown ininformation about an activity or concept. Notes are shown in
information about an activity or concept. Notes are shown in
information about an activity or concept. Notes are shown ininformation about an activity or concept. Notes are shown in
bold type.bold type.
bold type.
bold type.bold type.
WW
arnings, cautions, notes and the titles of front panel keys orarnings, cautions, notes and the titles of front panel keys or
W
arnings, cautions, notes and the titles of front panel keys or
WW
arnings, cautions, notes and the titles of front panel keys orarnings, cautions, notes and the titles of front panel keys or
system commands and menus shown in instructions are shownsystem commands and menus shown in instructions are shown
system commands and menus shown in instructions are shown
system commands and menus shown in instructions are shownsystem commands and menus shown in instructions are shown
in bold type.in bold type.
in bold type.
in bold type.in bold type.
The QC100 is capable of great accuracy and precision and may
surpass the measurement-accuracy capability of the human operator. The system is also capable of displaying higher resolution
and more significant figures than is often needed. These facts
may be important in your interpretation and use of the measurement data provided by the system.
arning messages are shown in boldarning messages are shown in bold
arning messages are shown in bold
arning messages are shown in boldarning messages are shown in bold
ou should not proceed until youou should not proceed until you
ou should not proceed until you
ou should not proceed until youou should not proceed until you
ouou
ou
ouou
System configuration and
setup
Metronics, Incorporated assumes that all QC100 systems will be
configured and setup for the end-user by dealer or OEM representatives. This setup includes calibration according to the user’s
optical comparator or similar device, and configuration of error
compensation if necessary.
When the end-user installs a new comparator, it will probably be
necessary to perform configuration and setup again. In that case,
technically qualified supervisory personnel can refer to Chapter 4:
System Setup, or can contact their dealer representatives for assistance.
iii
Page 12
Introduction
WW
ARNINGARNING
W
ARNING
WW
ARNINGARNING
There is never any reason to open the QC100 enclosure. ThereThere is never any reason to open the QC100 enclosure. There
There is never any reason to open the QC100 enclosure. There
There is never any reason to open the QC100 enclosure. ThereThere is never any reason to open the QC100 enclosure. There
are no userare no user
are no user
are no userare no user
with any electronic instrument, there is a danger of electricalwith any electronic instrument, there is a danger of electrical
with any electronic instrument, there is a danger of electrical
with any electronic instrument, there is a danger of electricalwith any electronic instrument, there is a danger of electrical
shock and damage to the instrument if the enclosure is opened.shock and damage to the instrument if the enclosure is opened.
shock and damage to the instrument if the enclosure is opened.
shock and damage to the instrument if the enclosure is opened.shock and damage to the instrument if the enclosure is opened.
-serviceable components or assemblies inside. As-serviceable components or assemblies inside. As
-serviceable components or assemblies inside. As
-serviceable components or assemblies inside. As-serviceable components or assemblies inside. As
The Quadra-Chek 100® is an advanced digital readout system for
performing 2, 3 and 4 axis measurements at very high levels of
precision and accuracy. The QC100 can be used with optical comparators, toolmaker’s microscopes, touch-probes and video measurement systems to perform precision measurements as part of
in-line production activities or final quality inspection.
The intuitive interface will be familiar to users of the Quadra-Chek
1000® and other Metronics products. Operators will find the
QC100® easy to understand and use thanks to a large LCD display
and front panel keys clearly marked with familiar function and
mode symbols.
The large LCD displays alphanumeric and graphic information for
the current measurement, measurement features and measurement data clearly on one screen, eliminating the need to page or
scroll for information.
Panel keys are provided for entering numeric data, zeroing
axes, turning the LCD on or off
and sending data to a printer or
computer.
Softkeys are also provided that
work in conjunction with changing measurement-specific functions displayed on the LCD
screen.
Two large FastTrack keys located
over the LCD can be programmed as hotkeys that initiate frequently used system
functions and can be quickly
1-1
Page 18
Chapter 1Overview
LCD screen
Softkeys
FastT rack
pressed without looking at the front panel. In addition, most keys
on the front panel can be programmed as hotkeys to initiate commonly used functions.
Zero axes
Send data
Numeric keypad
Command keys
LCD ON/OFF
IRDA port
Cursor keys
(menu scrolling)
1-2
All front panel keys provide tactile sensory feedback, and many
key-press operations can be configured by supervisors or dealer
representatives to generate an audible sound.
Speaker and external speaker
jack outputs are provided that
can be adjusted for quiet or
noisy environments. Ear phones
can be plugged into the external speaker jack to facilitate silent operation in quiet environments.
Speaker Jack
Phone Jack
Tilt adjust
Page 19
Input/Output connectors
Measurements viewed on the front panel LCD can be transmitted
to an IBM-compatible personal computer over a standard serial
port connection, or to a printer over a parallel or serial port. An
infrared (IRDA) port is also included on the front panel for communication with computers and other infrared-compatible devices.
Quadra-Chek® QC100
The compact ergonomic design
and adjustable-tilt front panel of
the QC100 allow users to locate
and mount the instrument in a
wide variety of environments to
accommodate nearly any viewing requirement. The tilt front
panel can be adjusted and secured in any convenient position. Rubber feet on the bottom
prevent slipping when the system is not permanently bolted
to a worksurface. Bolt holes are
provided in the mounting stand
that match the pattern of the
QC 1000.
Available options and
software version
An optional foot switch and remote keypad are available for
control and data entry when the
user is not close to the front
panel.
You system’s available options and the software version are listed
in the Language screen of the Setup menu. Refer to Chapter 4:
Setup for details.
1-3
Page 20
Chapter 1Overview
Specifications
Input VInput V
Input V
Input VInput V
FF
use:use:
F
use:1.6 Amp 250V Slow Blow 5X20mm
FF
use:use:
Input FInput F
Input F
Input FInput F
CACA
CA
CACA
FF
or continued protection against fire, replace only with a fuseor continued protection against fire, replace only with a fuse
F
or continued protection against fire, replace only with a fuse
FF
or continued protection against fire, replace only with a fuseor continued protection against fire, replace only with a fuse
of the specified voltage and current ratings.of the specified voltage and current ratings.
of the specified voltage and current ratings.
of the specified voltage and current ratings.of the specified voltage and current ratings.
Dimensions:Dimensions:
Enclosure (WxHxD):11.5” X 7.5” X 2.75”
Base (WxHxD):10” X 2” X 7.8”
Enclosure weight:3.5 lbs.
Base weight:7 lbs.
LCD:LCD:
LCD:
LCD:LCD:
Size/color:6” black and white
Display digit size:0.5”
oltage Range:oltage Range:
oltage Range: 85 VAC to 264 VAC. 1.0 Amp maximum
oltage Range:oltage Range:
(Auto switchable)
requency:requency:
requency: 43Hz to 63Hz
requency:requency:
UTIONUTION
UTION
UTIONUTION
non-condensing
1-4
ENC tests:ENC tests:
ENC tests:
ENC tests:ENC tests:
EN61326:1998EN61326:1998
EN61326:1998
EN61326:1998EN61326:1998
EMC for electrical equipment for measurement, control and laboratory use
EN61010EN61010
EN61010
EN61010EN61010
Safety requirements for electrical equipment for measurement,
control and laboratory use
WW
ARNING:ARNING:
W
ARNING:
WW
ARNING:ARNING:
TT
o reduce the risk of electrical shock, never remove the covero reduce the risk of electrical shock, never remove the cover
T
o reduce the risk of electrical shock, never remove the cover
TT
o reduce the risk of electrical shock, never remove the covero reduce the risk of electrical shock, never remove the cover
or open the enclosure. There are no useror open the enclosure. There are no user
or open the enclosure. There are no user
or open the enclosure. There are no useror open the enclosure. There are no user
inside. Refer all service requirements to qualified serviceinside. Refer all service requirements to qualified service
inside. Refer all service requirements to qualified service
inside. Refer all service requirements to qualified serviceinside. Refer all service requirements to qualified service
personnel.personnel.
personnel.
personnel.personnel.
-servicable parts-servicable parts
-servicable parts
-servicable parts-servicable parts
Page 21
Chapter 2
Operation
This chapter describes how to operate the QC100. Instructions
will show typical LCD screen contents. This chapter includes:
• Measuring with the QC100page 2-2
• Probing and targetingpage 2-2
• Datumpage 2-2
• Front panel controls and LCD screenspage 2-3
• Front panel keyspage 2-3
• LCD menuspage 2-8
• Preparing for operationpage 2-9
• Performing measurementspage 2-15
The contents of LCD screens will reflect a system using X and Y
axis encoders configured with factory default settings. Your system may also include Z and/or Q axis encoders and might have
slightly different defaults and operating parameters, however the
instructions in this chapter are generic and will apply equally well
to all systems.
2-1
Page 22
Chapter 2Operation
Measuring with the
QC100
Probing and targeting
Datum
The QC100 performs very precise dimensional measurements. Part
measurement is a simple process that includes positioning and
securing the part for measurement, establishing a datum and probing the part by targeting data points. The measurement result is
displayed on the DRO screen and can be printed or sent to a computer.
Once a datum has been established, points can be
measured by
ometry using crosshairs or by sensing the part with a touch-probe.
Crosshairs are targeted over a point by moving the part in the
stage. When a touch-probe is used, the probe is placed in contact with the part and indicates the part surface displacement as
the part is moved. In either case, displacement information is
displayed continuously on the DRO screen.
NOTENOTE
NOTE
NOTENOTE
This guide will refer to the process of a measuring point asThis guide will refer to the process of a measuring point as
This guide will refer to the process of a measuring point as
This guide will refer to the process of a measuring point asThis guide will refer to the process of a measuring point as
targeting a point targeting a point
targeting a point
targeting a point targeting a point
method and hardware used.method and hardware used.
method and hardware used.
method and hardware used.method and hardware used.
The locations of probed or targeted points are always referred to
an absolute datum point or origin. The absolute datum coordinates are 0,0. The absolute datum can be set to any physical
point on or off the part being measured by pressing Zero axes
keys on the front panel. Setting the absolute datum and orienting
the axes establishes a reference frame within which a second (incremental) datum can be defined for incremental measurements.
targeting
or or
or
or or
the desired physical point on the part ge-
probing a featureprobing a feature
probing a feature
probing a featureprobing a feature
independent of the independent of the
independent of the
independent of the independent of the
probed
and
2-2
The datums can be zeroed using the Zero axes keys or can be set
to non-zero values that are entered from the numeric keypad using the Preset function.
Presetting a datum is useful when measurements are made from a
known part coordinate such as a point specified by a part drawing.
Page 23
Quadra-Chek® QC100
Front panel controls
and LCD screens
Softkeys
The QC100 front panel controls will be explained in detail. Explanations will include the typical LCD screen images associated with
each control. Please read this portion of the guide carefully as
subsequent operating instructions depend on a familiarity with
user controls and LCD screen content.
The softkeys initiate functions that appear above the keys at the
bottom of the LCD screen. As different system functions are used,
the supporting softkey screen commands change in support.
Softkey commandsSoftkey commands
Softkey selections presented at the bottom of LCD screens provide
alternative viewing, measurement and configuration functions that
FastT rack
Zero axes
LCD screen
Softkeys
Cursor keys
(menu scrolling)
Send data
Numeric keypad
Command keys
LCD ON/OFF
IRDA port
2-3
Page 24
Chapter 2Operation
are unique to specific screens. Softkeys can be configured to provide a variety of commonly used functions by supervisors or dealer
representatives.
FastT rack
Zero axes
Send data
LCD screen
Softkeys
Command keys
Numeric keypad
Command keys
LCD ON/OFF
IRDA port
Cursor keys
(menu scrolling)
The command keys are used to enter individual data points, finish
a data entry session, cancel (or delete) the last data entry, or quit
the current activity and return to the DRO screen.
EnterEnter
Enter
EnterEnter
Enters data for a measurement. Information highlighted on the
LCD is stored as measurement or configuration data. This information includes probed location coordinates (each time a point is
targeted), data from stored features or alphanumeric data for a
configuration data field.
2-4
FinishFinish
Finish
FinishFinish
Completes the part probing or data entry process for a measurement.
Page 25
Quadra-Chek® QC100
CancelCancel
Cancel
CancelCancel
Erases the last data value entered into the system from part probing or from the numeric keypad. The Cancel key can also be used
to delete any highlighted feature from the feature list.
QuitQuit
Quit
QuitQuit
Aborts the current task and returns to the DRO screen.
Cursor keys
Zero axes keys
FastTrack keys
Numeric keypad
Send data
LCD ON/OFF
The 4 cursor keys (also referred to as cursor arrows) are used to
scroll lists, and to navigate menus.
The top, middle and bottom zero axes keys are used to zero the
current position of the X, Y, Z and Q axes respectively. The axes are
zeroed to set the absolute or incremental datums.
Programmed as hotkeys to initiate frequently used functions.
The keypad is used for numeric data entry. The decimal point key
and +/- key can also be used to reduce or increase the LCD screen
contrast if a decimal point or +/- input are not expected by the
system as part of a data entry process. The screen contrast setting
will be saved when the system is powered down and used as the
default contrast setting the next time the system is turned on.
Transmits measurement result information for the displayed feature to a computer over the serial port, to a printer over the parallel or serial port, or to any infrared-compatible device over the
infrared (IRDA) port.
Toggles the LCD display on and off without removing power from
the instrument. Pressing this switch while in the DRO screen also
re-locks the setup functions if a supervisor password was entered
earlier to unlock them.
IRDA port
Infrared transmission window for communication with infraredcompatible devices.
2-5
Page 26
Chapter 2Operation
Softkey menus
Setup menu
Extra menu
An important collection of configuration, measurement and
display functions can be made
available by activating the
Softkey MenuSoftkey Menu
Softkey Menu functions. These
Softkey MenuSoftkey Menu
functions are contained in the
displayed by pressing the
Setup menuSetup menu
The
Setup menu is used only by supervisors and dealer represen-
Setup menuSetup menu
tatives to configure the advanced measurement, display and communication parameters of the QC100 system. Use of the Setup
menu is discussed for supervisors and dealer representatives later
in Chapter 4: Setup.
NOTENOTE
NOTE
NOTENOTE
Operators should not attempt to use the Setup menu to changeOperators should not attempt to use the Setup menu to change
Operators should not attempt to use the Setup menu to change
Operators should not attempt to use the Setup menu to changeOperators should not attempt to use the Setup menu to change
system configurations.system configurations.
system configurations.
system configurations.system configurations.
Extra menuExtra menu
The
Extra menu provides access
Extra menuExtra menu
to functions that control measuring, communication and display parameters. The
menumenu
menu is displayed by pressing
menumenu
Extra softkeyExtra softkey
the
Extra softkey.
Extra softkeyExtra softkey
SetupSetup
Setup and
SetupSetup
Menu softkeyMenu softkey
Menu softkey.
Menu softkeyMenu softkey
ExtraExtra
Extra
ExtraExtra
ExtraExtra
menusmenus
Extra
menus which are
ExtraExtra
menusmenus
2-6
DMS/DDDMS/DD
DMS/DD
DMS/DDDMS/DD
Toggles the display of angles
between degrees, minutes and
seconds and decimal degrees.
Zero 2Zero 2
Zero 2
Zero 2Zero 2
Zeroes the absolute and incremental datums.
NOTENOTE
NOTE
NOTENOTE
The Extra menu includes functions that zero all axes when theThe Extra menu includes functions that zero all axes when the
The Extra menu includes functions that zero all axes when the
The Extra menu includes functions that zero all axes when theThe Extra menu includes functions that zero all axes when the
QC100 system includes a Z and/or Q axis.QC100 system includes a Z and/or Q axis.
QC100 system includes a Z and/or Q axis.
QC100 system includes a Z and/or Q axis.QC100 system includes a Z and/or Q axis.
Send XSend X
Send X
Send XSend X
Sends the current X axis data displayed on the DRO screen to a
printer or computer.
Page 27
Quadra-Chek® QC100
Send YSend Y
Send Y
Send YSend Y
Sends the current Y axis data displayed on the DRO screen to a
printer or computer.
Send 2Send 2
Send 2
Send 2Send 2
Sends the current X and Y axes data displayed on the DRO screen
to a printer or computer.
NOTENOTE
NOTE
NOTENOTE
The Extra menu includes functions that send data from allThe Extra menu includes functions that send data from all
The Extra menu includes functions that send data from all
The Extra menu includes functions that send data from allThe Extra menu includes functions that send data from all
axes when the QC100 system includes a Z and/or Q axis.axes when the QC100 system includes a Z and/or Q axis.
axes when the QC100 system includes a Z and/or Q axis.
axes when the QC100 system includes a Z and/or Q axis.axes when the QC100 system includes a Z and/or Q axis.
Clear (CL) datumsClear (CL) datums
Clear (CL) datums
Clear (CL) datumsClear (CL) datums
Resets all absolute and incremental datums to machine coordinates.
Min MaxMin Max
Min Max
Min MaxMin Max
Configures an axis to accumulate and retain minimum and maximum measured values until the Finish key is pressed. This is useful for applications such as touch-probe measurements of runout.
Repeat (Rpt) presetRepeat (Rpt) preset
Repeat (Rpt) preset
Repeat (Rpt) presetRepeat (Rpt) preset
Resets datums to the last preset values.
MCMC
SS
MC
S
MCMC
SS
Allows the user to clear encoder datums, cross reference marks
and re-establish machine coordinates without cycling the system
power.
2-7
Page 28
Chapter 2Operation
LCD screens
A variety of LCD screens are displayed during the course of normal
operation. These screens are nearly self-explanatory because they
use similar presentations of information and data.
PP
owerower
-up screen-up screen
P
ower
-up screen
PP
owerower
-up screen-up screen
The power-up screen is presented each time the system is
turned on. Pressing any key advances to the DRO screen.
DRO screenDRO screen
DRO screen
DRO screenDRO screen
The DRO screen shows the current linear or angular positions
of axes, the current unit of mea-
MMMM
sure (
MM or
MMMM
tum point reference for measurements (
for incremental).
Pressing the
toggles the unit of linear measure between metric and English. Pressing the
toggles the datum between absolute and incremental.
InchInch
Inch) and the da-
InchInch
AbsAbs
Abs for absolute,
AbsAbs
MM/Inch softkeyMM/Inch softkey
MM/Inch softkey
MM/Inch softkeyMM/Inch softkey
IncInc
Inc
IncInc
Abs/Inc softkeyAbs/Inc softkey
Abs/Inc softkey
Abs/Inc softkeyAbs/Inc softkey
2-8
When the
Setup menu, pressing it toggles the displayed value for that axis
between a radius (by halving a diameter) and a diameter (by doubling a radius).
Softkeys are also provided for accessing the
presettingpresetting
presetting the axis datums to non-zero values.
presettingpresetting
Dia/Rad softkeyDia/Rad softkey
Dia/Rad softkey function is activated for an axis in the
Dia/Rad softkeyDia/Rad softkey
menumenu
menu functions and
menumenu
Page 29
Quadra-Chek® QC100
Preparing for
operation
Safety and other remindersSafety and other reminders
Safety and other reminders
Safety and other remindersSafety and other reminders
First-time users should thoroughly understand the preceding portions of this chapter before operating the QC100. The operating
instructions assume a basic familiarity with the process of measuring features, QC100 controls and indicators and the information
presented on LCD screens.
NOTENOTE
NOTE
NOTENOTE
It is assumed that operators have a basic understanding ofIt is assumed that operators have a basic understanding of
It is assumed that operators have a basic understanding of
It is assumed that operators have a basic understanding ofIt is assumed that operators have a basic understanding of
part fixturing and the geometric probing techniques requiredpart fixturing and the geometric probing techniques required
part fixturing and the geometric probing techniques required
part fixturing and the geometric probing techniques requiredpart fixturing and the geometric probing techniques required
for their coordinate measuring devices.for their coordinate measuring devices.
for their coordinate measuring devices.
for their coordinate measuring devices.for their coordinate measuring devices.
Location and mountingLocation and mounting
Location and mounting
Location and mountingLocation and mounting
Verify that the QC100 is mounted on a stable reliable surface. If it
should fall, it could become seriously damaged or cause injuries
to the user.
PP
ower cord and plugower cord and plug
P
ower cord and plug
PP
ower cord and plugower cord and plug
The power cord should not be located where it can be walked on
or create a tripping hazard. Always connect the 3-wire power
plug to a 3-wire grounded outlet. Use of a 2-wire power plug
adapter or any other connection accessory that removes the third
grounded connection creates a safety hazard and should not be
permitted.
PP
ower surge suppressorower surge suppressor
P
ower surge suppressor
PP
ower surge suppressorower surge suppressor
A high-quality power surge suppressor should be used to limit the
amplitude of potentially damaging power line transients caused
by the operation of heavy electrical machinery or lightning.
LiquidsLiquids
Liquids
LiquidsLiquids
Do not spill or splash liquids on the QC100 enclosure.
Configuration and system setupConfiguration and system setup
Configuration and system setup
Configuration and system setupConfiguration and system setup
System configuration and setup should be performed only by technically qualified supervisors or dealer representatives. Operators
should not attempt to alter the configuration of the QC100.
2-9
Page 30
Chapter 2Operation
Getting started
Very little preparation is required before operating the QC100.
Connections to the system should be checked and power applied.
Checking connectionsChecking connections
Checking connections
Checking connectionsChecking connections
Perform a routine inspection of all connections to the digital readout system. Connections should be clean and tight. Cables should
be located away from moving objects and should not create tripping hazards.
Cables connected to the parallel and serial (RS232) output ports
should be shielded and should be terminated and firmly connected
on both ends.
CACA
UTIONUTION
CA
UTION
CACA
UTIONUTION
Never connect electrical cables to the system when the powerNever connect electrical cables to the system when the power
Never connect electrical cables to the system when the power
Never connect electrical cables to the system when the powerNever connect electrical cables to the system when the power
is on.is on.
is on.
is on.is on.
Applying powerApplying power
Applying power
Applying powerApplying power
To apply power, press the power
switch located at the back of the
enclosure to ON (1). The system will initialize and display the
power-up screen.
2-10
Press any key on the front panel
to complete the system initialization. The DRO screen will be
displayed showing the current
positions of encoders on each
axis.
NOTENOTE
NOTE
NOTENOTE
If the encoders used with your system include reference marks,If the encoders used with your system include reference marks,
If the encoders used with your system include reference marks,
If the encoders used with your system include reference marks,If the encoders used with your system include reference marks,
you may be instructed to cross the reference marks on eachyou may be instructed to cross the reference marks on each
you may be instructed to cross the reference marks on each
you may be instructed to cross the reference marks on eachyou may be instructed to cross the reference marks on each
axis before the DRO screen is displayed.axis before the DRO screen is displayed.
axis before the DRO screen is displayed.
axis before the DRO screen is displayed.axis before the DRO screen is displayed.
Page 31
Quadra-Chek® QC100
Getting ready to measure
The first steps in any measurement session will include adjusting
the LCD viewing, setting measurement and display parameters,
and setting the datums.
Adjusting LCD tiltAdjusting LCD tilt
Adjusting LCD tilt
Adjusting LCD tiltAdjusting LCD tilt
The tilt angle of the QC100 enclosure can usually be adjusted for
optimum viewing by simply swiveling the enclosure on its base to
the desired position. If the enclosure angle is locked in position,
loosen the two allen screws at the base of the enclosure, move
the enclosure to the desired angle and tighten the allen screws so
that the enclosure is held firmly in position but can be adjusted
again without loosening the allen screws.
Adjusting LCD contrastAdjusting LCD contrast
Adjusting LCD contrast
Adjusting LCD contrastAdjusting LCD contrast
The LCD contrast can be adjusted for optimum viewing from the
DRO screen. Press the decimal point of the numeric keypad to
decrease contrast or the +/- key to increase contrast.
Setting measurement and display parametersSetting measurement and display parameters
Setting measurement and display parameters
Setting measurement and display parametersSetting measurement and display parameters
The measure and display parameters that you establish prior to a
measurement session will depend on the measurement that you
are making and the way you would like to have data and graphics
displayed on the LCD screen.
Unit of linear measureUnit of linear measure
Unit of linear measure
Unit of linear measureUnit of linear measure
Linear measurements can be displayed in mm or inches. Select the
desired unit of measure if necessary by pressing the
softkeysoftkey
softkey. The unit of measure will be displayed at the lower left
softkeysoftkey
corner of the LCD. This selection can be changed at any time.
MM/InchMM/Inch
MM/Inch
MM/InchMM/Inch
Unit of angular measureUnit of angular measure
Unit of angular measure
Unit of angular measureUnit of angular measure
Angular measurements can be
displayed in degrees, minutes
and seconds (
mal degrees (
desired unit of measure if necessary by pressing the
softkeysoftkey
softkey, the
softkeysoftkey
Press the
The result will be displayed on the DRO screen. This
selection can be changed at any time.
DMSDMS
DMS) or in deci-
DMSDMS
DDDD
DD). Select the
DDDD
MenuMenu
Menu
MenuMenu
Extra softkeyExtra softkey
Extra softkey, then selecting
Extra softkeyExtra softkey
Finish keyFinish key
Finish key to toggle between
Finish keyFinish key
DMS/DDDMS/DD
DMS/DD.
DMS/DDDMS/DD
DMSDMS
DMS and
DMSDMS
DDDD
DD.
DDDD
2-11
Page 32
Chapter 2Operation
DatumDatum
Datum
DatumDatum
Select the absolute datum (Abs) if necessary by pressing the
Inc softkeyInc softkey
Inc softkey. The selected datum will be displayed at the lower left
Inc softkeyInc softkey
corner of the LCD. This selection can be changed at any time.
Zeroing datumsZeroing datums
Zeroing datums
Zeroing datumsZeroing datums
The system simultaneously uses 2 reference datums for measurements. One is fixed (absolute), while the other can be redefined
(incremental) during the measurement session to make incremental measurements. Reference points for absolute and incremental
measurements can be set by zeroing the
(incremental) datums.
Absolute datumAbsolute datum
Absolute datum
Absolute datumAbsolute datum
Absolute measurements are those referred to the absolute datum.
The absolute datum is defined as the origin (0,0), and should be
set at the beginning of each measurement session. To set the
absolute datum:
Step 1
Move the stage to target the
point you wish to define as the
origin, or absolute datum point,
and press the
if necessary to select
Abs/Inc softkeyAbs/Inc softkey
Abs/Inc softkey
Abs/Inc softkeyAbs/Inc softkey
AbsAbs
Abs in the lower left corner of the LCD screen.
AbsAbs
AbsAbs
Abs (absolute) and
AbsAbs
Abs/Abs/
Abs/
Abs/Abs/
IncInc
Inc
IncInc
2-12
Step 2
Press the
zero-out the measurement axes
on the DRO screen.
Incremental datumIncremental datum
Incremental datum
Incremental datumIncremental datum
An incremental measurement is
one that is made between any
reference point and some other
point. Incremental measurements are facilitated in the QC100 by the incremental datum which
can be zeroed at any point on or off the part without changing
the absolute datum.
Zero axes keysZero axes keys
Zero axes keys to
Zero axes keysZero axes keys
Page 33
Quadra-Chek® QC100
Presetting datumsPresetting datums
Presetting datums
Presetting datumsPresetting datums
Datums can be set to non-zero values that match known part coordinates using the numeric keypad and
Presetting datums is useful when measurements must be made
from a part coordinate specified on a part drawing. To preset
datums to a known part coordinate:
Step 1
Select the datums you wish to
preset (
Step 2
Move the stage to target the
part coordinate that you wish to
use as a measurement datum
and press the
You will be instructed to select
the axes to preset.
Step 3
Select the axes you wish to preset by pressing the desired zero
axis softkey(s). An axis will be
underlined.
AbsAbs
Abs or
AbsAbs
IncInc
Inc).
IncInc
Preset softkeyPreset softkey
Preset softkey.
Preset softkeyPreset softkey
PresetPreset
Preset function.
PresetPreset
Step 4
Enter the preset coordinate for
the datum using the numeric
keypad.
2-13
Page 34
Chapter 2Operation
STEP 5
Preset other axes if desired and
press the
The DRO screen will display the
new preset coordinates.
Finish keyFinish key
Finish key.
Finish keyFinish key
2-14
Page 35
Quadra-Chek® QC100
Performing measurements
Linear and angular measure-Linear and angular measure-
Linear and angular measure-
Linear and angular measure-Linear and angular measurementsments
ments
mentsments
This section will describe how to perform measurements with the
QC100. The QC100 can perform simple linear or angular displacement measurements such as a distance between points or an angle
of rotation, or can store values to measure cumulative minimum
and maximum displacements for touch-probe or crosshair measurements such as shaft runout.
NOTENOTE
NOTE
NOTENOTE
The instructions assume that the QC100 system is properlyThe instructions assume that the QC100 system is properly
The instructions assume that the QC100 system is properly
The instructions assume that the QC100 system is properlyThe instructions assume that the QC100 system is properly
connected and that any necessary measurement or displayconnected and that any necessary measurement or display
connected and that any necessary measurement or display
connected and that any necessary measurement or displayconnected and that any necessary measurement or display
adjustments have been made to meet the requirements ofadjustments have been made to meet the requirements of
adjustments have been made to meet the requirements of
adjustments have been made to meet the requirements ofadjustments have been made to meet the requirements of
the userthe user
the user
the userthe user
where the current positions of axis encoders are displayed.where the current positions of axis encoders are displayed.
where the current positions of axis encoders are displayed.
where the current positions of axis encoders are displayed.where the current positions of axis encoders are displayed.
NOTENOTE
NOTE
NOTENOTE
VV
alues on screens shown in the instructions are only examples,alues on screens shown in the instructions are only examples,
V
alues on screens shown in the instructions are only examples,
VV
alues on screens shown in the instructions are only examples,alues on screens shown in the instructions are only examples,
not actual measurements.not actual measurements.
not actual measurements.
not actual measurements.not actual measurements.
To perform linear or angular measurements:
Step 1
Move the stage to target the start point for the measurement and
zero or preset the axes datums using the
softkeysoftkey
softkey.
softkeysoftkey
. Each set of instructions will begin at the DRO screen. Each set of instructions will begin at the DRO screen
. Each set of instructions will begin at the DRO screen
. Each set of instructions will begin at the DRO screen. Each set of instructions will begin at the DRO screen
Zero axis keysZero axis keys
Zero axis keys or
Zero axis keysZero axis keys
PresetPreset
Preset
PresetPreset
Step 2
Move the stage to target the point to be measured or rotate to
the angle to be measured. The measurement results will be displayed on the DRO screen and can be printed or sent to a computer by pressing the
measurement.
Send data keySend data key
Send data key. Repeat for each axis of
Send data keySend data key
2-15
Page 36
Chapter 2Operation
Min Max measurementsTo perform minimum and maximum measurements:
Step 1
Move the stage to target the start point for the measurement and
zero or preset the axes datums using the
softkeysoftkey
softkey.
softkeysoftkey
Step 2
Press the
press the
Extra menuExtra menu
Extra menu will be displayed.
Extra menuExtra menu
Step 3
Scroll to highlight
ing the
press the
Menu softkeyMenu softkey
Menu softkey, then
Menu softkeyMenu softkey
Extra softkeyExtra softkey
Extra softkey. The
Extra softkeyExtra softkey
Min MaxMin Max
Min Max us-
Min MaxMin Max
cursor arrow keyscursor arrow keys
cursor arrow keys, and
cursor arrow keyscursor arrow keys
Finish keyFinish key
Finish key.
Finish keyFinish key
Zero axis keysZero axis keys
Zero axis keys or
Zero axis keysZero axis keys
PresetPreset
Preset
PresetPreset
2-16
You will be instructed to select
a measurement axis.
Page 37
Quadra-Chek® QC100
Step 4
Press the
the desired measurement axis.
The
will be displayed.
Perform the required measurements. The minimum and maximum measurement values will
automatically be accumulated
and retained for the selected
axis.
The measurement result will be
displayed on the
screenscreen
screen and can be printed or
screenscreen
sent to a computer by pressing the
FinishFinish
Finish key to end the Min Max measurement session and return to
FinishFinish
the DRO screen.
Zero axis softkeyZero axis softkey
Zero axis softkey of
Zero axis softkeyZero axis softkey
Min Max measure screenMin Max measure screen
Min Max measure screen
Min Max measure screenMin Max measure screen
Min MaxMin Max
Min Max
Min MaxMin Max
Send data keySend data key
Send data key. Press the
Send data keySend data key
2-17
Page 38
Chapter 2Operation
2-18
Page 39
Chapter 3
Communications
This chapter explains how to establish the interface between a
printer and a QC100. There are many manufacturers of printers,
and unfortunately not all printers are compatible. It would be impossible to support all of them.
NOTENOTE
NOTE
NOTENOTE
It is the users responsibility to check printer compatibilityIt is the users responsibility to check printer compatibility
It is the users responsibility to check printer compatibility
It is the users responsibility to check printer compatibilityIt is the users responsibility to check printer compatibility
bling and system configuration accordinglybling and system configuration accordingly
bling and system configuration accordingly
bling and system configuration accordinglybling and system configuration accordingly
..
.
..
, ca-, ca-
, ca-
, ca-, ca-
RS232 port
Default settings
RS232 pin designations
The RS232 port must be configured to communicate with your
printer or computer. The following is a list of factory default settings. Refer to the
default settings.
• Baud Rate: 9600
• Parity:No Parity
• Stop Bits: 2
• Data Bits: 8
Ports
section of Chapter 4: Setup to change
3-1
Page 40
Chapter 3Communications
Print section of Setup menu:Print section of Setup menu:
Print section of Setup menu:
Print section of Setup menu:Print section of Setup menu:
Line terminator: Select carriage return or Carriage
return/Line feed to terminate each line of data.
Print axis labels: Include (or suppress) axis labels in data.
Print axis units:Include (or suppress) axis units of measure in
data.
PP
orts section of Setup menu:orts section of Setup menu:
P
orts section of Setup menu:
PP
orts section of Setup menu:orts section of Setup menu:
• Baud Rate: 9600
• Parity:No Parity
• Stop Bits: 2
• Data Bits: 8
Print formats
Angular formats
Angle format (DMS)
In the following, the "v" means to ignore the character it is substituted for i.e., X, Y, Z, r, d, <, T, R, A, etc.
"d" means digit
"-" is the negative sign
"s" means space
"c" means carriage return
For positive outputs, the space character will be substituted for
the negative sign "-". The leading digits may be spaces.
When a QC100 is shipped from the factory, each output line is
terminated by a carriage return (ASCII 13).
The factory default is the DMS format.
In DMS (degree, minute, second) notation, the printout is as follows:
Character
Position:1234567890123456789
Content:v v s - ddd. dd. d. sssssc
Example 1:<14 5 . 0 0 . 0 1
Example 2:A- 5 . 2 3 . 0 0
3-2
Page 41
Quadra-Chek® QC100
Angle format (DD)
Linear formats
Linear format (Metric)
Linear format (English)
In DD (decimal degrees) notation, the printout is as follows:
Character
Position:1234567890123456789
Content:v v s s - ddd. dddssssssc
Example 1:<14 5 . 0 0 1
Example 2:A- 5 . 3 0 3
Units (in or mm) are always at position 15 and 16. The factory
default Metric resolution is 1 micron. The factory default English
resolution is 0.0001 inch.
For Metric values with a display resolution of 1 micron, the format
is:
Character
Position:1234567890123456789
Content:v s s - dddd. ddds s mms s c
Example 1:r324. 051mm
Example 2:x- 5 . 3 0 3mm
For English values with a display resolution of .0001 inch, the format is:
Measurement Formats
Character
Position:1234567890123456789
Content:v s s - ddd. dddds s i ns s c
Example 1:r2 . 0 2 1 3i n
Example 2:x- 1 5 . 1 4 4 2i n
The QC100 is able to print measurements in formats specified in
the Print section of the Setup menu. These include:
backspace
horiz. tab
line feed
vert. tab
form feed
carr. ret.
so
si
dle
dcl
dc2
dc3
dc4
nak
syn
etb
can
em
su b
esc
fs
gs
rs
us
sp ace
!
"
#
$
%
&
'
(
)
*
+
'
U
V
W
X
Y
Z
[
\
]
^
_
`
a
b
c
d
e
f
g
h
i
j
k
l
m
n
o
p
q
r
s
t
u
v
w
x
y
z
3-4
Page 43
Chapter 4
Setup
Configuring the QC100
The Setup menu
The operating parameters of the QC100 must be configured prior
to using the system the first time, or any time part measurement,
reporting, or data communication requirements change. System
configurations are entered into the system using a few navigation
and selection keys located of the front panel. The process is facilitated by a simple Setup menu structure of configuration functions that can be quickly navigated on the large LCD display.
All operating parameters of the QC100 can be configured using
functions in the Setup menu. These parameters include:
• Language shown on the LCD display
• Display format of measurement data
• Encoder configuration
• Hotkey configuration
• Print configuration for reports
• Ports configuration for serial and parallel communication
• Supervisor password and program lock
• Squareness angle
• LEC (Linear Error Correction)
• SLEC (Segmented Linear Error Correction) configuration
• NLEC (Non-Linear Error Correction)
• Scale factor
• Misc (Miscellaneous) parameters such as key delay
Parameter configurations will be retained by the system until they
are changed again in the Setup menu in a new session.
4-1
Page 44
Chapter 4Setup
Accessing the Setup menu
Using the Setup menu
Example of configuring
Display parameters
Setup menu Setup menu
The
Setup menu can be dis-
Setup menu Setup menu
played from the DRO screen by
pressing the
then the
bottom left of the LCD.
Pressing the
display the contents of the
Setup menuSetup menu
Setup menu on the left portion
Setup menuSetup menu
of the screen with the
guage guage
guage selection highlighted.
guage guage
Selecting functions in the Setup
menu is a straightforward process of scrolling through and
highlighting Setup functions
using the
ciated operating parameters displayed on the right portion of the
screen and can be configured or changed using the
softkeyssoftkeys
softkeys or the
softkeyssoftkeys
figuring a Display parameter demonstrates using the
In this example the
DD DD
in
DD (decimal degrees) units of angular measure.
DD DD
Step 1
Press the
Setup menu.Setup menu.
the
Setup menu.
Setup menu.Setup menu.
Menu softkeyMenu softkey
Menu softkey and
Menu softkeyMenu softkey
Setup softkeySetup softkey
Setup softkey at the
Setup softkeySetup softkey
Setup softkeySetup softkey
Setup softkey will
Setup softkeySetup softkey
Lan-Lan-
Lan-
Lan-Lan-
cursor keyscursor keys
cursor keys. Once a setup function is highlighted, asso-
cursor keyscursor keys
cursor keyscursor keys
cursor keys,
cursor keyscursor keys
numeric keypadnumeric keypad
numeric keypad. The following example of con-
numeric keypadnumeric keypad
Setup menuSetup menu
Setup menu.
Setup menuSetup menu
Display Display
Display parameters will be configured to start
Display Display
Menu softkeyMenu softkey
Menu softkey, then press the
Menu softkeyMenu softkey
Setup softkeySetup softkey
Setup softkey to display
Setup softkeySetup softkey
4-2
Step 3
Press the
access the
function.
down cursor keydown cursor key
down cursor key to
down cursor keydown cursor key
Display Setup menuDisplay Setup menu
Display Setup menu
Display Setup menuDisplay Setup menu
Page 45
Quadra-Chek® QC200
Step 4
Press the
highlight parameter values on
the right side of the screen, then
press the
highlight the
parameterparameter
parameter.
parameterparameter
Alternative selections for the
highlighted parameter will become available over the softkeys
at the bottom of the screen. Pressing the softkey under the desired selection will change the highlighted parameter.
Step 5
To set the
rameterrameter
rameter to decimal degrees,
rameterrameter
press the
tom of the LCD screen. The
Startup Angular parameterStartup Angular parameter
Startup Angular parameter
Startup Angular parameterStartup Angular parameter
will be changed to
LCD display.
right cursor keyright cursor key
right cursor key to
right cursor keyright cursor key
down cursor keydown cursor key
down cursor key to
down cursor keydown cursor key
Startup AngularStartup Angular
Startup Angular
Startup AngularStartup Angular
Startup Angular pa-Startup Angular pa-
Startup Angular pa-
Startup Angular pa-Startup Angular pa-
DD softkeyDD softkey
DD softkey at the bot-
DD softkeyDD softkey
DDDD
DD on the
DDDD
In this manner, any Setup menu
function can be selected, and
any configurable parameter changed to one of the alternatives
presented over softkeys across the bottom of the LCD screen.
Configurable parameters on the right portion of the screen can be
scrolled by pressing the
edly pressing the
Once the highlighted parameter has been changed, pressing the
left cursor keyleft cursor key
left cursor key or the
left cursor keyleft cursor key
turn to the list of
screen. Pressing the
menumenu
menu and return to display the
menumenu
NOTENOTE
NOTE
NOTENOTE
A limited collection of setup functions are available underA limited collection of setup functions are available under
A limited collection of setup functions are available under
A limited collection of setup functions are available underA limited collection of setup functions are available under
normal conditions. The password must be entered into thenormal conditions. The password must be entered into the
normal conditions. The password must be entered into the
normal conditions. The password must be entered into thenormal conditions. The password must be entered into the
Supervisor setup function to make all setup functions availableSupervisor setup function to make all setup functions available
Supervisor setup function to make all setup functions available
Supervisor setup function to make all setup functions availableSupervisor setup function to make all setup functions available
for configuration. Entering the password for configuration isfor configuration. Entering the password for configuration is
for configuration. Entering the password for configuration is
for configuration. Entering the password for configuration isfor configuration. Entering the password for configuration is
explained in the explained in the
explained in the
explained in the explained in the
that followthat follow
that follow
that followthat follow
Enter keyEnter key
Enter key.
Enter keyEnter key
Setup menu functionsSetup menu functions
Setup menu functions on the left portion of the
Setup menu functionsSetup menu functions
..
.
..
up or down cursor arrowup or down cursor arrow
up or down cursor arrow or by repeat-
up or down cursor arrowup or down cursor arrow
Finish keyFinish key
Finish key will save a new setting and re-
Finish keyFinish key
Finish keyFinish key
Finish key a second time will exit the
Finish keyFinish key
DRO screenDRO screen
DRO screen.
DRO screenDRO screen
SupervisorSupervisor
Supervisor
SupervisorSupervisor
portion of the Setup Instructions portion of the Setup Instructions
portion of the Setup Instructions
portion of the Setup Instructions portion of the Setup Instructions
SetupSetup
Setup
SetupSetup
4-3
Page 46
Chapter 4Setup
Setup instructions
Accessing the Setup Menu
Use of control keys
The methods of selecting and configuring system parameters are
explained previously in the
strated in
selections available in the
changed if necessary the first time the QC100 system is used. After the first use, system parameters should be changed as necessary whenever part measurement, or data communication requirements change.
Access the
pressing the
the bottom right of the LCD, and
then pressing the
at the bottom left of the LCD.
The
be displayed on the left portion
of the screen with the Language
selection highlighted.
The Enter, Finish, Cancel and
Quit keys have slightly different functions during setup activities.
Example of configuring Display parameters
Setup menuSetup menu
Setup menu by
Setup menuSetup menu
Menu softkeyMenu softkey
Menu softkey at
Menu softkeyMenu softkey
Setup softkeySetup softkey
Setup softkey
Setup softkeySetup softkey
Setup menuSetup menu
Setup menu contents will
Setup menuSetup menu
Using the Setup menu
Setup menuSetup menu
Setup menu should be evaluated and
Setup menuSetup menu
and are demon-
. All of the
Enter key
Finish key
Cancel key
Quit key
4-4
Pressing the
and advances to the next setup parameter. Pressing the
at the bottom of a list advances to the top of the list.
Pressing the
category and returns to the
screen.
Pressing the
from the front panel keypad. Pressing the
last (left most) digit in a highlighted data field. Pressing the
cel keycel key
cel key in an alpha field, or a field changed by softkey selections
cel keycel key
has no effect.
Pressing the
category and returns to the
Enter keyEnter key
Enter key stores the highlighted value or selection
Enter keyEnter key
Enter keyEnter key
Enter key
Enter keyEnter key
Finish keyFinish key
Finish key stores all parameters for the current setup
Finish keyFinish key
Setup menuSetup menu
Setup menu on the left portion of the
Setup menuSetup menu
Cancel keyCancel key
Cancel key is only active when entering numeric data
Cancel keyCancel key
Cancel keyCancel key
Cancel key erases the
Cancel keyCancel key
Can-Can-
Can-
Can-Can-
Quit keyQuit key
Quit key stores all parameters for the current setup
Quit keyQuit key
Setup menuSetup menu
Setup menu.
Setup menuSetup menu
Page 47
Quadra-Chek® QC200
Entering the Supervisor
Password for configuration
Language
Only a limited subset of setup
functions will be available until
the supervisor password is entered for configuration. To enter the
scroll through the
to
Press the
highlight the
blockblock
block.
blockblock
Use the
the front panel to enter the
pervisor passwordpervisor password
pervisor password, then press
pervisor passwordpervisor password
the
Scroll through the
with the
to highlight the
lection.
supervisor passwordsupervisor password
supervisor password,
supervisor passwordsupervisor password
Setup menuSetup menu
Setup menu
Setup menuSetup menu
SupervisorSupervisor
Supervisor.
SupervisorSupervisor
left cursor arrowleft cursor arrow
left cursor arrow to
left cursor arrowleft cursor arrow
password entrypassword entry
password entry
password entrypassword entry
numeric keyboardnumeric keyboard
numeric keyboard on
numeric keyboardnumeric keyboard
Enter keyEnter key
Enter key.
Enter keyEnter key
Setup menuSetup menu
Setup menu
Setup menuSetup menu
up or down cursor keyup or down cursor key
up or down cursor key
up or down cursor keyup or down cursor key
LanguageLanguage
Language se-
LanguageLanguage
su-su-
su-
su-su-
Software version and
available options
Press the
select items in the right portion
of the screen. Press the
down cursor arrowdown cursor arrow
down cursor arrow or the
down cursor arrowdown cursor arrow
ter buttonter button
ter button to scroll through the
ter buttonter button
list of languages available for selection. Highlight the desired
language and press the
softkeysoftkey
softkey to select the language.
softkeysoftkey
Press the
Finish buttonFinish button
Finish button to save the lan-
Finish buttonFinish button
guage selection and return to
the
Your system’s
in the Language screen.
right cursor keyright cursor key
right cursor key to
right cursor keyright cursor key
left cursor arrowleft cursor arrow
left cursor arrow or
left cursor arrowleft cursor arrow
Setup menuSetup menu
Setup menu.
Setup menuSetup menu
software versionsoftware version
software version and available
software versionsoftware version
up orup or
up or
up orup or
En-En-
En-
En-En-
YY
eses
Y
es
YY
eses
optionsoptions
options are listed
optionsoptions
4-5
Page 48
Chapter 4Setup
Display
Startup Linear/Angular
Scroll through the
to highlight the
tiontion
tion.
tiontion
Press the
select items in the right portion
of the screen.
Display parameters are divided
into measurement display parameters and measure resolution.
Set the measurement display
parameters as required. Highlighting a parameter causes selection alternatives to be displayed over the softkeys across the bottom of the screen.
Startup LinearStartup Linear
Startup Linear and
Startup LinearStartup Linear
display of linear and angular measurements. Linear measurements
can be displayed in English or metric units of measure, and angular measurements can be displayed in degrees, minutes and seconds or in decimal degrees. Enter the desired settings and scroll
to highlight
right cursor keyright cursor key
right cursor key to
right cursor keyright cursor key
Setup menuSetup menu
Setup menu
Setup menuSetup menu
Display func-Display func-
Display func-
Display func-Display func-
RadixRadix
Radix.
RadixRadix
Startup AngularStartup Angular
Startup Angular settings define the default
Startup AngularStartup Angular
Radix
4-6
Radix settings are available for
English (a dot: 1.000) or European (a comma: 1,000) presentations. Enter the desired parameter setting and scroll to
Current AngularCurrent Angular
Current Angular.
Current AngularCurrent Angular
Page 49
Quadra-Chek® QC200
Current Angular
Display Resolution (Disp Res)
Current AngularCurrent Angular
Current Angular displays of
Current AngularCurrent Angular
measured angles can be set to
degrees, minutes and seconds
or to decimal degrees. This setting defines the presentation of
measured angles until power to
the system is turned off. When
power is reapplied to the system
Startup AngularStartup Angular
the
Startup Angular setting will
Startup AngularStartup Angular
be used to define the display of
measured angles. Enter the desired parameter setting and scroll
AxisAxis
to
Axis.
AxisAxis
Select an encoder axis and set
the numeric values of the display resolution parameters to
match the encoder and measurement instrument being used
by scrolling to a parameter and
entering numeric values using
the front panel keypad. Press
Enter keyEnter key
the
Enter key to complete the
Enter keyEnter key
entry of each new value. Press
left cursor arrowleft cursor arrow
the
left cursor arrow or
left cursor arrowleft cursor arrow
buttonbutton
button to save the
buttonbutton
rametersrameters
rameters and return to the
rametersrameters
Setup menuSetup menu
Setup menu.
Setup menuSetup menu
FinishFinish
Finish
FinishFinish
Display pa-Display pa-
Display pa-
Display pa-Display pa-
4-7
Page 50
Chapter 4Setup
Encoders
Axis
Scroll through the
to highlight the
tion.
Press the
select items in the right portion
of the screen.
Encoder parameters include the
Axis being configured, selection
of Rad/Dia presentation, encoder resolution, Teach (resolution)
function, type of encoder, reference marks, machine zero offset,
reversed state, interpolation value and units of measure. Most
encoder inputs are metric and the Units will be set to MM.
The encoders used with your
system will be shown above the
softkeys across the bottom of
the screen. The example shown
here is a 4 axis system.
Press the
you wish to configure (X axis in
this example) and press the
down cursor arrowdown cursor arrow
down cursor arrow to high-
down cursor arrowdown cursor arrow
light the Rad/Dia parameter.
right cursor keyright cursor key
right cursor key to
right cursor keyright cursor key
softkeysoftkey
softkey for the axis
softkeysoftkey
Setup menuSetup menu
Setup menu
Setup menuSetup menu
EncoderEncoder
Encoder selec-
EncoderEncoder
Rad/Dia Type
4-8
YY
eses
Press the
the user to present a measurement on the selected axis as either a radius or diameter. When
Rad/DiaRad/Dia
the
Rad/Dia function is active,
Rad/DiaRad/Dia
RadRad
a
Rad or
RadRad
added to the bottom of the
DRO screenDRO screen
DRO screen.
DRO screenDRO screen
softkeysoftkey
Y
es
softkey to allow
YY
eses
softkeysoftkey
DiaDia
softkeysoftkey
Dia
softkey will be
DiaDia
softkeysoftkey
Page 51
Quadra-Chek® QC200
Res
Teach amount/Scale Cnts
Enter the resolution of the selected axis encoder using the
numeric keypadnumeric keypad
numeric keypad on the front
numeric keypadnumeric keypad
panel. Press the
to store the data and advance
the display to highlight
AmtAmt
Amt.
AmtAmt
If the encoder resolution is unknown, it can be determined
using the
and a measurement standard.
To determine the encoder resolution:
Step 1
Mount the standard into the
stage and move the stage to
target the first side of the known distance (or angle).
Step 2
Set the
sary.
TT
T
TT
Units Units
Units data field to the encoder’s units of measure if neces-
Units Units
Enter buttonEnter button
Enter button
Enter buttonEnter button
TT
eacheach
T
each
TT
eacheach
each Amteach Amt
each Amt function
each Amteach Amt
Step 3
Enter the standard’s known dis-
TT
tance (or angle) into the
AmtAmt
Amt data field and press the
AmtAmt
TT
each softkeyeach softkey
T
each softkey.
TT
each softkeyeach softkey
eacheach
T
each
TT
eacheach
4-9
Page 52
Chapter 4Setup
Step 4
Press the
the
Step 5
Move the stage to target the
other side of the standard’s
known distance (or angle), press
the
Finish keyFinish key
Finish key to complete the
Finish keyFinish key
Teach process.
Reset softkeyReset softkey
Reset softkey to zero
Reset softkeyReset softkey
Scale cntsScale cnts
Scale cnts.
Scale cntsScale cnts
Enter keyEnter key
Enter key, then press the
Enter keyEnter key
Type
Ref Marks
4-10
Press the
Press the
of the display that corresponds
to the type of encoder used for
the selected axis and press the
down cursor arrowdown cursor arrow
down cursor arrow or the
down cursor arrowdown cursor arrow
ter buttonter button
ter button to advance the dis-
ter buttonter button
play and highlight
Encoder reference marks are
used by the system for Segmented Linear Error Correction
(SLEC) or Non-Linear Error Correction (NLEC) to establish the
machine coordinate system.
Upon system startup the QC100
may be configured to prompt
the user to move each axis encoder across a reference mark prior to measurement activities.
right and down cursor arrow keysright and down cursor arrow keys
right and down cursor arrow keys to highlight
right and down cursor arrow keysright and down cursor arrow keys
softkeysoftkey
softkey at the bottom
softkeysoftkey
En-En-
En-
En-En-
Ref MarksRef Marks
Ref Marks.
Ref MarksRef Marks
TT
ypeype
T
ype.
TT
ypeype
Page 53
Quadra-Chek® QC200
Mach 0 offset
Reversed
Press the
coder on the selected axis and press the
Enter keyEnter key
Enter key to advance the display and highlight
Enter keyEnter key
Enter the machine zero offset
for the highlighted encoder into
the
press the
or the
the display and highlight
versedversed
versed.
versedversed
The normal convention for the display of the polarity of measurements is:
• Motion to the right is positive (X axis)
• Motion up is positive (Y axis)
• Motion up is positive (Z axis)
• Clockwise rotation is positive (Q axis)
The settings for any axis can be
reversed to accomodate the encoder requirements of the measuring system being used. Press
the
polarity for the selected axis, or
press the
the polarity of the encoder input. Press the
rowrow
row or
rowrow
display and highlight
softkeysoftkey
softkey for the type of reference mark used by the en-
softkeysoftkey
down cursor arrowdown cursor arrow
down cursor arrow or
down cursor arrowdown cursor arrow
Mach 0 offsetMach 0 offset
Mach 0 offset.
Mach 0 offsetMach 0 offset
Mach 0 offset Mach 0 offset
Mach 0 offset data field and
Mach 0 offset Mach 0 offset
down cursor arrowdown cursor arrow
down cursor arrow
down cursor arrowdown cursor arrow
Enter buttonEnter button
Enter button to advance
Enter buttonEnter button
Re-Re-
Re-
Re-Re-
No softkeyNo softkey
No softkey to retain the
No softkeyNo softkey
YY
es softkeyes softkey
Y
es softkey to reverse
YY
es softkeyes softkey
down cursor ardown cursor ar
down cursor ar
down cursor ardown cursor ar
Enter keyEnter key
Enter key to advance the
Enter keyEnter key
InterpolationInterpolation
Interpolation.
InterpolationInterpolation
--
-
--
Interpolation
Press the
valuevalue
value required for the encoder
valuevalue
being used on the selected axis
and press the
rowrow
row or
rowrow
display and highlight
softkey interpolationsoftkey interpolation
softkey interpolation
softkey interpolationsoftkey interpolation
down cursor ardown cursor ar
down cursor ar
down cursor ardown cursor ar
Enter keyEnter key
Enter key to advance the
Enter keyEnter key
UnitsUnits
Units.
UnitsUnits
--
-
--
4-11
Page 54
Chapter 4Setup
Units
Hot Keys
Press the softkey Units type that
corresponds to the encoder being used on the selected axis and
scroll back to
configure other axes encoders if
necessary.
Once all encoder axes have been
configured, press the
sor arrowsor arrow
sor arrow or
sor arrowsor arrow
turn to the
Front panel keys, remote keypad keys, and foot switch keys can be
defined as
measurement functions. Front panel keys, keypad keys and foot
switch keys are mapped to system functions in the
tion of the
other system keys include:
• Unit keys• Unit keys
• Unit keys
• Unit keys• Unit keys
The 10 number keys (0-9) of the front panel numeric keypad
• Remote keys• Remote keys
• Remote keys
• Remote keys• Remote keys
The 10 number keys (0-9) of the remote keypad
• F• F
oot switchoot switch
• F
oot switch
• F• F
oot switchoot switch
The 2 switches on the optional foot switch assembly
• Wide keys• Wide keys
• Wide keys
• Wide keys• Wide keys
The 2 FastTrack selector keys located above the LCD
AxisAxis
Axis to select and
AxisAxis
left curleft cur
left cur
left curleft cur
Finish keyFinish key
Finish key to re-
Finish keyFinish key
Setup menuSetup menu
Setup menu.
Setup menuSetup menu
Hot KHot K
eyseys
Hot K
eys that are used to activate QC100 display and
Hot KHot K
eyseys
Setup menuSetup menu
Setup menu. Keys that can be mapped to functions or
Setup menuSetup menu
--
-
--
Hot KHot K
eyseys
Hot K
eys por-
Hot KHot K
eyseys
4-12
NOTENOTE
NOTE
NOTENOTE
Once a key is mapped to a function or another key and becomesOnce a key is mapped to a function or another key and becomes
Once a key is mapped to a function or another key and becomes
Once a key is mapped to a function or another key and becomesOnce a key is mapped to a function or another key and becomes
an active Hot Kan active Hot K
an active Hot K
an active Hot Kan active Hot K
pressed only within the correct context. Fpressed only within the correct context. F
pressed only within the correct context. F
pressed only within the correct context. Fpressed only within the correct context. F
system is being used to measure a dimension and expects thesystem is being used to measure a dimension and expects the
system is being used to measure a dimension and expects the
system is being used to measure a dimension and expects thesystem is being used to measure a dimension and expects the
Enter key to be used to signify data input, it will only interpretEnter key to be used to signify data input, it will only interpret
Enter key to be used to signify data input, it will only interpret
Enter key to be used to signify data input, it will only interpretEnter key to be used to signify data input, it will only interpret
the Enter key for data input and will not invoke any Hot Kthe Enter key for data input and will not invoke any Hot K
the Enter key for data input and will not invoke any Hot K
the Enter key for data input and will not invoke any Hot Kthe Enter key for data input and will not invoke any Hot K
function that may have been mapped to it.function that may have been mapped to it.
function that may have been mapped to it.
function that may have been mapped to it.function that may have been mapped to it.
eyey
, it will invoke the designated function when, it will invoke the designated function when
ey
, it will invoke the designated function when
eyey
, it will invoke the designated function when, it will invoke the designated function when
or example, if theor example, if the
or example, if the
or example, if theor example, if the
eyey
ey
eyey
Page 55
Using the Cursor arrows, scroll
through the
highlight the
Setup menuSetup menu
Setup menu to
Setup menuSetup menu
Hot KHot K
eyseys
Hot K
eys function.
Hot KHot K
eyseys
Quadra-Chek® QC200
Defining Hot Key functions
Press the
keykey
key to select items in the right
keykey
portion of the screen.
The method used to define
key functionskey functions
key functions is identical for
key functionskey functions
each of the four key types (Unit, Remote, Foot, and Wide). A key
type is selected using a softkey at the bottom of the LCD; Unit,
Remote, Foot, or Wide.
After a key type is selected, the specific key is selected from the list
available by scrolling with the
Unit:Unit:
Unit: Key 1 to key 10
Unit:Unit:
Remote:Remote:
Remote: Key 1 to key 10
Remote:Remote:
right cursor arrowright cursor arrow
right cursor arrow
right cursor arrowright cursor arrow
hothot
hot
hothot
cursor arrowscursor arrows
cursor arrows.
cursor arrowscursor arrows
FF
ootoot
F
oot: Key 1 or key 2
FF
ootoot
4-13
Page 56
Chapter 4Setup
WideWide
Wide: Key 1 or key 2
WideWide
After the specific key is selected,
the key assignment type is selected using a softkey at the bottom of the LCD.
None
Key
Special
No assignment will be made,
this will not be a hot key
A key will be assigned by pressing one of the front panel keys
A QC100 system function will be assigned by scrolling
and selecting from a list. Functions in the special list
include:
DMS/DDDMS/DD
DMS/DD
DMS/DDDMS/DD
Toggles the display of angles between degrees, minutes and seconds and decimal degrees.
Zero 2Zero 2
Zero 2
Zero 2Zero 2
Zeroes the absolute and incremental datums.
NOTENOTE
NOTE
NOTENOTE
The Extra menu includes functions that zero all axes when theThe Extra menu includes functions that zero all axes when the
The Extra menu includes functions that zero all axes when the
The Extra menu includes functions that zero all axes when theThe Extra menu includes functions that zero all axes when the
QC100 system includes a Z and/or Q axis.QC100 system includes a Z and/or Q axis.
QC100 system includes a Z and/or Q axis.
QC100 system includes a Z and/or Q axis.QC100 system includes a Z and/or Q axis.
Send XSend X
Send X
Send XSend X
Sends the current X axis data displayed on the DRO screen to a
printer or computer.
4-14
Send YSend Y
Send Y
Send YSend Y
Sends the current Y axis data displayed on the DRO screen to a
printer or computer.
Page 57
Quadra-Chek® QC200
Send 2Send 2
Send 2
Send 2Send 2
Sends the current X and Y axes data displayed on the DRO screen
to a printer or computer.
NOTENOTE
NOTE
NOTENOTE
The Extra menu includes functions that send data from allThe Extra menu includes functions that send data from all
The Extra menu includes functions that send data from all
The Extra menu includes functions that send data from allThe Extra menu includes functions that send data from all
axes when the QC100 system includes a Z and/or Q axis.axes when the QC100 system includes a Z and/or Q axis.
axes when the QC100 system includes a Z and/or Q axis.
axes when the QC100 system includes a Z and/or Q axis.axes when the QC100 system includes a Z and/or Q axis.
Clear (CL) datumsClear (CL) datums
Clear (CL) datums
Clear (CL) datumsClear (CL) datums
Resets all absolute and incremental datums to machine coordinates.
Min MaxMin Max
Min Max
Min MaxMin Max
Configures an axis to accumulate and retain minimum and maximum measured values until the Finish key is pressed. This is useful for applications such as touch-probe measurements of runout.
Repeat (Rpt) presetRepeat (Rpt) preset
Repeat (Rpt) preset
Repeat (Rpt) presetRepeat (Rpt) preset
Resets datums to the last preset values.
MCMC
SS
MC
S
MCMC
SS
Allows the user to clear datums, cross reference marks and reestablish machine coordinates without cycling the system power.
Example of assigning hot
keys
An example of assigning hot keys is provided below and demonstrates how to make two different types of hot key assignments
to the numeric keypad. The methods illustrated in this example
can be applied directly to the assignment of Hot Keys to the Unit,
Remote, Foot and Wide keys.
Press the
keykey
key to select items in the right
keykey
portion of the Hot Keys screen
and press the
the bottom of the LCD if necessary.
right cursor arrowright cursor arrow
right cursor arrow
right cursor arrowright cursor arrow
Unit softkeyUnit softkey
Unit softkey at
Unit softkeyUnit softkey
4-15
Page 58
Chapter 4Setup
Use the
rowrow
row to scroll the list of 10 Unit
rowrow
Keys and highlight the desired
key.
in this example.
Press the
bottom of the LCD if no assignment is required or to remove
an existing key assignment.
Press the
tom of the LCD to assign a front
panel key. You will be instructed
to press the key that you wish
to assign.
Press the desired front panel key
CancelCancel
(
Cancel in this example). The
CancelCancel
new assignment will appear on
the LCD.
up or down cursor ar-up or down cursor ar-
up or down cursor ar-
up or down cursor ar-up or down cursor ar-
Unit key number 9Unit key number 9
Unit key number 9 is used
Unit key number 9Unit key number 9
None softkeyNone softkey
None softkey at the
None softkeyNone softkey
KK
ey softkeyey softkey
K
ey softkey at the bot-
KK
ey softkeyey softkey
4-16
Press the
bottom of the LCD to assign a
QC100 function. Use the
down cursor arrowdown cursor arrow
down cursor arrow to scroll the
down cursor arrowdown cursor arrow
list of functions and highlight
the desired function. Press the
Enter keyEnter key
Enter key to assign the func-
Enter keyEnter key
tion.
Special softkeySpecial softkey
Special softkey at the
Special softkeySpecial softkey
up orup or
up or
up orup or
Page 59
The new assignment will appear
on the LCD.
Quadra-Chek® QC200
Print
Line terminator
Print axis labels
Scroll through the
to highlight the
Press the
select items in the right portion
of the screen.
Select a carriage return (
carriage return/line feed (
to terminate each line of data
sent to a printer or computer.
Press the
or the
Print Axis LabelsPrint Axis Labels
Print Axis Labels.
Print Axis LabelsPrint Axis Labels
Select
in the data sent to a printer or
computer. Press the
sor arrowsor arrow
sor arrow or the
sor arrowsor arrow
advance to
right cursor arrowright cursor arrow
right cursor arrow to
right cursor arrowright cursor arrow
down cursor arrowdown cursor arrow
down cursor arrow
down cursor arrowdown cursor arrow
Enter keyEnter key
Enter key to advance to
Enter keyEnter key
YY
eses
Y
es to include axis labels
YY
eses
Print Axis UnitsPrint Axis Units
Print Axis Units.
Print Axis UnitsPrint Axis Units
Setup menuSetup menu
Setup menu
Setup menuSetup menu
Print selectionPrint selection
Print selection.
Print selectionPrint selection
CRCR
CR) or
CRCR
CR/LFCR/LF
CR/LF)
CR/LFCR/LF
down curdown cur
down cur
down curdown cur
Enter keyEnter key
Enter key to
Enter keyEnter key
--
-
--
4-17
Page 60
Chapter 4Setup
Print Axis Units
YY
Select
of measure in the data sent to a
printer or computer. Press the
Finish keyFinish key
Finish key to return to the
Finish keyFinish key
Setup menu.Setup menu.
Setup menu.
Setup menu.Setup menu.
eses
Y
es to include axis units
YY
eses
Ports
Baud
Scroll through the
to highlight the
Serial (RS232) and parallel communications are configured in
the screen.
Press the
select items in the right portion
of the screen.
The RS232 baud rate can be
decremented or incremented to
the desired value by pressing the
Dec or Inc softkeysDec or Inc softkeys
Dec or Inc softkeys.
Dec or Inc softkeysDec or Inc softkeys
right cursor arrowright cursor arrow
right cursor arrow to
right cursor arrowright cursor arrow
Setup menuSetup menu
Setup menu
Setup menuSetup menu
PP
orts selectionorts selection
P
orts selection.
PP
orts selectionorts selection
4-18
The available baud rates are
1200, 2400, 4800, 9600,
19,200, 33,400, 56,000, and
115,000.
Press the
press the
and highlight
Dec or Inc softkeysDec or Inc softkeys
Dec or Inc softkeys to select the required baud rate and
Dec or Inc softkeysDec or Inc softkeys
down cursor arrowdown cursor arrow
down cursor arrow or
down cursor arrowdown cursor arrow
WW
ord Lenord Len
W
ord Len.
WW
ord Lenord Len
Enter keyEnter key
Enter key advance the display
Enter keyEnter key
Page 61
Quadra-Chek® QC200
Word Length (Word Len)
Stop Bits
Parity
Select the required number of
bits per word by pressing the de-
softkeysoftkey
sired
softkey and press the
softkeysoftkey
down cursor arrowdown cursor arrow
down cursor arrow or the
down cursor arrowdown cursor arrow
ter keyter key
ter key to advance the display
ter keyter key
and highlight
Select the required number of
stop bits by pressing the desired
softkeysoftkey
softkey and press the
softkeysoftkey
cursor arrowcursor arrow
cursor arrow or the
cursor arrowcursor arrow
to advance the display and high-
PP
arityarity
light
P
arity.
PP
arityarity
Select the required parity format
by pressing the desired
and press the
rowrow
row or the
rowrow
the display and highlight
shakeshake
shake.
shakeshake
Stop BitsStop Bits
Stop Bits.
Stop BitsStop Bits
down cursor ardown cursor ar
down cursor ar
down cursor ardown cursor ar
Enter keyEnter key
Enter key to advance
Enter keyEnter key
En-En-
En-
En-En-
downdown
down
downdown
Enter keyEnter key
Enter key
Enter keyEnter key
softkeysoftkey
softkey
softkeysoftkey
Hand-Hand-
Hand-
Hand-Hand-
--
-
--
Handshake
Select the required handshaking
protocol by pressing the desired
softkeysoftkey
softkey and press the
softkeysoftkey
cursor arrowcursor arrow
cursor arrow or the
cursor arrowcursor arrow
to advance the display and high-
DataData
light
Data.
DataData
downdown
down
downdown
Enter keyEnter key
Enter key
Enter keyEnter key
4-19
Page 62
Chapter 4Setup
Data
Supervisor
Measurement data can be sent to the serial or parallel port for
printing.
Press the
the
Scroll through the
to highlight the
lectionlection
lection. The
lectionlection
wordword
word can be entered into the
wordword
PP
assword data fieldassword data field
P
assword data field to enable
PP
assword data fieldassword data field
setup functions, and a startup
encoder zero can be specified.
None softkeyNone softkey
None softkey if no data transmission is desired. Press
None softkeyNone softkey
PP
osition softkeyosition softkey
P
osition softkey to print the data displayed on the LCD.
PP
osition softkeyosition softkey
Setup menuSetup menu
Setup menu
Setup menuSetup menu
Supervisor se-Supervisor se-
Supervisor se-
Supervisor se-Supervisor se-
Supervisor pass-Supervisor pass-
Supervisor pass-
Supervisor pass-Supervisor pass-
Password
Startup zero
4-20
Press the
and enter the
wordword
word using the
wordword
padpad
pad. Press the
padpad
keykey
key when the password is com-
keykey
plete. Press the
arrowarrow
arrow or
arrowarrow
Startup zeroStartup zero
to
Startup zero.
Startup zeroStartup zero
Press the
encoder reference mark crossings upon system startup. Press
the
left cursor arrow or
buttonbutton
button to return to the
buttonbutton
menumenu
menu.
menumenu
right cursor arrowright cursor arrow
right cursor arrow
right cursor arrowright cursor arrow
Supervisor pass-Supervisor pass-
Supervisor pass-
Supervisor pass-Supervisor pass-
numeric key-numeric key-
numeric key-
numeric key-numeric key-
EnterEnter
Enter or
EnterEnter
down cursordown cursor
down cursor
down cursordown cursor
Enter keyEnter key
Enter key to advance
Enter keyEnter key
YY
es softkeyes softkey
Y
es softkey to require
YY
es softkeyes softkey
left cursor arrowleft cursor arrow
left cursor arrowleft cursor arrow
FinishFinish
Finish
FinishFinish
FinishFinish
Finish
FinishFinish
SetupSetup
Setup
SetupSetup
Page 63
Quadra-Chek® QC200
SquarenessScroll through the
squareness correction coefficient is used to compensate for small
machine errors that can exist between the X and Y measurement
axes. The intersection angle between two perfectly square lines is
measured on a standard and entered into the
Enter the intersection angle between two square lines measured on the standard into the
Angle data fieldAngle data field
Angle data field using the key-
Angle data fieldAngle data field
board on the front panel. Press
left cursor arrowleft cursor arrow
the
left cursor arrow or
left cursor arrowleft cursor arrow
buttonbutton
button to return to the
buttonbutton
menumenu
menu.
menumenu
Angle data fieldAngle data field
Angle data field.
Angle data fieldAngle data field
Setup menuSetup menu
Setup menu to highlight
Setup menuSetup menu
FinishFinish
Finish
FinishFinish
SetupSetup
Setup
SetupSetup
SquarenessSquareness
Squareness. The
SquarenessSquareness
4-21
Page 64
Chapter 4Setup
Error CorrectionLinear, segmented linear and optional non-linear error correction
methods can be used with the QC100. Linear and segmented
linear error corrections are provided with every QC100 system and
non-linear error correction is provided as an option. Each method
compensates for encoder and machine travel variations using error correction coefficients developed by comparing actual measurements of a standard to the standard’s nominal values.
Linear error correction (LEC) uses one correction coefficient for
each axis to compensate for variations along the axis. A standard
is measured and the standard’s nominal and measured values are
specified during the setup process. The resulting correction coefficient is then applied to total measured displacements. For example,
a linear error correction coefficient of 0.0002 “/inch applied to a
6” measurement along the X axis would produce a measurement
result of 6.0012”.
Segmented linear error correction (SLEC) divides each axis into up
to 7 segments. Each segment uses one error correction coefficient to compensate for variations within the segment. Each axis
also includes a machine zero offset to refer the position of the
standard’s datum to the machine zero reference. The standard’s
nominal and measured segment values must be specified during
the setup process. The machine zero offset for each axis must
also be specified. When SLEC is used, the stage must be moved to
cross a reference mark/hard-stop on each axis upon system startup
to initialize SLEC.
4-22
Non-linear error correction (NLEC) uses a standard that divides
the entire measurement area into a grid of up to 30 X 30 cells.
Each cell uses error correction coefficient for each axis to compensate for variations within a cell. The standard’s nominal and measured X and Y values must be specified for the positions of each
cell during the setup process. The machine zero offset to the grid
datum must also be specified. When NLEC is used, the stage must
have a repeatablemachine zero defined on startup to initialize
NLEC.
Page 65
Quadra-Chek® QC200
Linear Error Correction (LEC)
Scroll through the
to highlight
rection (LEC)rection (LEC)
rection (LEC). Linear error cor-
rection (LEC)rection (LEC)
rection coefficients are used to
compensate for overall encoder
and machine travel variations. A
standard is measured, and the
measured value is compared to
the known value of the standard
to generate the correction coefficient. Correction coefficients are generated for each axis and
used in subsequent measurements to provide linear compensation for encoder variations.
Press the
select items in the right portion
of the screen.
Enter the nominal value of the
standard for each axis into the
appropriate
fieldfield
field using the
fieldfield
on the front panel. Use the
down cursor arrowdown cursor arrow
down cursor arrow to scroll to
down cursor arrowdown cursor arrow
the desired data field.
right cursor arrowright cursor arrow
right cursor arrow to
right cursor arrowright cursor arrow
Setup menuSetup menu
Setup menu
Setup menuSetup menu
Linear Error CorLinear Error Cor
Linear Error Cor
Linear Error CorLinear Error Cor
Standard dataStandard data
Standard data
Standard dataStandard data
numeric keypadnumeric keypad
numeric keypad
numeric keypadnumeric keypad
--
-
--
Segmented Linear Error
Correction (SLEC)
Measure the standard on each axis and enter the measured value
into the appropriate
padpad
pad on the front panel. Use the
padpad
the desired data field.
When all the nominal and observed data is entered, press the
cursor arrowcursor arrow
cursor arrow or
cursor arrowcursor arrow
Scroll through the
to highlight
Error Correction (SLEC)Error Correction (SLEC)
Error Correction (SLEC). Linear
Error Correction (SLEC)Error Correction (SLEC)
error correction coefficients are
applied to segments of each axis
to compensate for encoder and
machine travel variations. The
standard is measured on each
axis, and the measured values
Segmented LinearSegmented Linear
Segmented Linear
Segmented LinearSegmented Linear
Observed data fieldObserved data field
Observed data field using the
Observed data fieldObserved data field
down cursor arrowdown cursor arrow
down cursor arrow to scroll to
down cursor arrowdown cursor arrow
Finish buttonFinish button
Finish button to return to the
Finish buttonFinish button
Setup menuSetup menu
Setup menu
Setup menuSetup menu
numeric key-numeric key-
numeric key-
numeric key-numeric key-
Setup menuSetup menu
Setup menu.
Setup menuSetup menu
leftleft
left
leftleft
4-23
Page 66
Chapter 4Setup
are compared to the nominal values to generate correction coefficients. Correction coefficients are generated for each segment
that deviates from the nominal value. The correction coefficients
are then used to provide linear compensation for only the segments requiring it.
Mount the standard on the stage and zero axes on the datum of
the standard. Measure and graph the
ment values as shown below.
that combine to form a straight line.that combine to form a straight line.
that combine to form a straight line.
that combine to form a straight line.that combine to form a straight line.
NOTENOTE
NOTE
NOTENOTE
The numbers used in this example are only meant to illustrateThe numbers used in this example are only meant to illustrate
The numbers used in this example are only meant to illustrate
The numbers used in this example are only meant to illustrateThe numbers used in this example are only meant to illustrate
the process and are not intended to suggest actual values.the process and are not intended to suggest actual values.
the process and are not intended to suggest actual values.
the process and are not intended to suggest actual values.the process and are not intended to suggest actual values.
A segment is any group of pointsA segment is any group of points
A segment is any group of points
A segment is any group of pointsA segment is any group of points
nominal
Select the desired axis and enter the nominal and actual
points of each segmentpoints of each segment
points of each segment
points of each segmentpoints of each segment
the SLEC
servedserved
served data fields as shown be-
servedserved
low. A maximum of 7 segments
(0-6) can be entered.
and
actual
StandardStandard
Standard and
StandardStandard
seg-
endend
end
endend
into
Ob-Ob-
Ob-
Ob-Ob-
4-24
Enter the offset from the Machine zero to the standard’s datum for the current axis into the
MZ Offset data fieldMZ Offset data field
MZ Offset data field.
MZ Offset data fieldMZ Offset data field
Press the
axis.
When all the data are entered for all axes, press the
arrowarrow
arrow or
arrowarrow
ON softkeyON softkey
ON softkey to use SLEC compensation for the selected
ON softkeyON softkey
Finish buttonFinish button
Finish button to return to the
Finish buttonFinish button
Setup menuSetup menu
Setup menu.
Setup menuSetup menu
left cursorleft cursor
left cursor
left cursorleft cursor
Page 67
Quadra-Chek® QC200
Non-Linear Error
Correction (NLEC)
Scroll through the
to highlight
Correction (NLEC)Correction (NLEC)
Non-Linear Error CorrectionNon-Linear Error Correction
(NLEC) is provided only as a(NLEC) is provided only as a
(NLEC) is provided only as a
(NLEC) is provided only as a(NLEC) is provided only as a
purchased option, and is notpurchased option, and is not
purchased option, and is not
purchased option, and is notpurchased option, and is not
included in all systems.included in all systems.
included in all systems.
included in all systems.included in all systems.
Error correction coefficients are applied to each cell of a grid that
covers the measuring area to compensate for encoder and machine travel variations.
The cells of a standard grid are measured at each corner. The measured values are compared to the nominal values of the standard
to generate correction coefficients for each cell of the grid. The
correction coefficients are then used to provide error compensation for the entire measuring area.
The 5 X 4 grid in the following example contains 12 cells. Each
cell corner is an X,Y position for data entry starting in the lower
left corner at 1,1. The lower left data entry position 1,1 corresponds to the datum for the grid (0,0).
Setup menuSetup menu
Setup menu
Setup menuSetup menu
Non-Linear ErrorNon-Linear Error
Non-Linear Error
Non-Linear ErrorNon-Linear Error
The offset from the Machine zero to the standard grid’s datum
must also be entered as data.
4-25
Page 68
Chapter 4Setup
Mount the standard grid and zero the axies on the standard’s
datum in the lower left corner. Measure and chart each position
(cell corner) as shown below. Actual measured values, or the difference between nominal and the measured values (Error) can be
charted. Actual measured values are charted in this example for
only the first 2 columns of cells.
NOTENOTE
NOTE
NOTENOTE
The numbers used in this example are only used to illustrateThe numbers used in this example are only used to illustrate
The numbers used in this example are only used to illustrate
The numbers used in this example are only used to illustrateThe numbers used in this example are only used to illustrate
the process and are not intended to suggest actual values. Inthe process and are not intended to suggest actual values. In
the process and are not intended to suggest actual values. In
the process and are not intended to suggest actual values. Inthe process and are not intended to suggest actual values. In
this example the grid is divided into 1 inch cellsthis example the grid is divided into 1 inch cells
this example the grid is divided into 1 inch cells
this example the grid is divided into 1 inch cellsthis example the grid is divided into 1 inch cells
PP
PP
ositionosition
P
osition
PP
ositionosition
1,1X=0X=0
NominalNominal
Nominal
NominalNominal
Y=0Y=0
ActualActual
Actual
ActualActual
ositionosition
P
osition
PP
ositionosition
2,1X=1X=0.9999
NominalNominal
Nominal
NominalNominal
Y=0Y=0.0001
ActualActual
Actual
ActualActual
1,2X=0X=0.0001
Y=1Y=1.0000
1,3X=0X=0.0001
Y=2Y=1.9998
1,4X=0X=0.0000
Y=3Y=3.0001
Entering grid size:
Scroll to the
data fieldsdata fields
data fields and enter the size
data fieldsdata fields
X and Y Grid SizeX and Y Grid Size
X and Y Grid Size
X and Y Grid SizeX and Y Grid Size
of X and Y grid. In this example
the grid consists of 12 cells that
form a matrix of data points (5)
X by (4) Y.
Entering grid cell size:
Scroll to the
data fieldsdata fields
data fields and enter the di-
data fieldsdata fields
X and Y Cell SizeX and Y Cell Size
X and Y Cell Size
X and Y Cell SizeX and Y Cell Size
mension of the grid cell. In this
example the cell size is 1 inch.
2,2X=1X=1.0000
Y=1Y=1.0001
2,3X=1X=1.0001
Y=2Y=2.0000
2,4X=1X=1.0000
Y=3Y=3.0001
4-26
Page 69
Entering machine zero offsets:
Select the
data fields data fields
data fields and enter the X and
data fields data fields
Y offsets from the machine reference to the datum of the standard grid. In this example the
Machine 0 values were 0.50 and
0.48 inches.
Entering the data type:
Scroll to the
tual data fieldtual data field
tual data field and select the
tual data fieldtual data field
data type for cell dimension
measurements. The type can be
the
tual)tual)
tual), or the
tual)tual)
tween the nominal and actualtween the nominal and actual
tween the nominal and actual
tween the nominal and actualtween the nominal and actual
measurement (Error)measurement (Error)
measurement (Error). In this
measurement (Error)measurement (Error)
example Actual is selected.
Inputting measurement data:
Scroll to the
data fields (Pdata fields (P
data fields (P
data fields (Pdata fields (P
sary increment or decrement the
contents using the
softkeyssoftkeys
softkeys to select the
softkeyssoftkeys
sition 1,1sition 1,1
sition 1,1. This corresponds to
sition 1,1sition 1,1
the datum (0,0) for the standard
grid.
X and Y Machine 0X and Y Machine 0
X and Y Machine 0
X and Y Machine 0X and Y Machine 0
Nominal Error/Ac-Nominal Error/Ac-
Nominal Error/Ac-
Nominal Error/Ac-Nominal Error/Ac-
actual measurement (Ac-actual measurement (Ac-
actual measurement (Ac-
actual measurement (Ac-actual measurement (Ac-
difference be-difference be-
difference be-
difference be-difference be-
X and Y positionX and Y position
X and Y position
X and Y positionX and Y position
os)os)
os) and if neces-
os)os)
Inc or DecInc or Dec
Inc or Dec
Inc or DecInc or Dec
grid po-grid po-
grid po-
grid po-grid po-
Quadra-Chek® QC200
NOTENOTE
NOTE
NOTENOTE
Grid position 1,1 is the datum for grid measurements, thereforeGrid position 1,1 is the datum for grid measurements, therefore
Grid position 1,1 is the datum for grid measurements, therefore
Grid position 1,1 is the datum for grid measurements, thereforeGrid position 1,1 is the datum for grid measurements, therefore
the Actual or Error values for X and Y must always be zero.the Actual or Error values for X and Y must always be zero.
the Actual or Error values for X and Y must always be zero.
the Actual or Error values for X and Y must always be zero.the Actual or Error values for X and Y must always be zero.
4-27
Page 70
Chapter 4Setup
Input the chart’s Actual (or Error) data by incrementing the
Y position (PY position (P
Y position (P
Y position (PY position (P
tual (or Error) data fieldstual (or Error) data fields
tual (or Error) data fields and entering the
tual (or Error) data fieldstual (or Error) data fields
selected position. The following example enters data for the first
few positions of the chart.
NOTENOTE
NOTE
NOTENOTE
When a position is selected, the Nominal and anticipated ActualWhen a position is selected, the Nominal and anticipated Actual
When a position is selected, the Nominal and anticipated Actual
When a position is selected, the Nominal and anticipated ActualWhen a position is selected, the Nominal and anticipated Actual
(or Error) values will be displayed. The Actual values are(or Error) values will be displayed. The Actual values are
(or Error) values will be displayed. The Actual values are
(or Error) values will be displayed. The Actual values are(or Error) values will be displayed. The Actual values are
changed if necessary by simply entering data over the displayedchanged if necessary by simply entering data over the displayed
changed if necessary by simply entering data over the displayed
changed if necessary by simply entering data over the displayedchanged if necessary by simply entering data over the displayed
value.value.
value.
value.value.
Scroll to the
crement it to 2 by pressing the
Increment (Inc) softkeyIncrement (Inc) softkey
Increment (Inc) softkey.
Increment (Inc) softkeyIncrement (Inc) softkey
Scroll to the
data fieldsdata fields
data fields and enter the
data fieldsdata fields
Y measurement dataY measurement data
Y measurement data for the
Y measurement dataY measurement data
1,2 position from the chart.
os) valuesos) values
os) values to select a position, scrolling to the
os) valuesos) values
X and Y data X and Y data
X and Y data for the
X and Y data X and Y data
Y PY P
os field os field
Y P
os field and in-
Y PY P
os field os field
Actual X and YActual X and Y
Actual X and Y
Actual X and YActual X and Y
X andX and
X and
X andX and
X andX and
X and
X andX and
Ac-Ac-
Ac-
Ac-Ac-
4-28
Y PY P
Scroll back to the
and increment it to 3 by pressing the
Increment (Inc) softkeyIncrement (Inc) softkey
Increment (Inc) softkey.
Increment (Inc) softkeyIncrement (Inc) softkey
Y P
Y PY P
os fieldos field
os field
os fieldos field
Page 71
Quadra-Chek® QC200
Scroll to the
fieldsfields
fields and enter the
fieldsfields
measurement data measurement data
measurement data for the 1,3
measurement data measurement data
position.
Scroll back to the
and increment it to 4 by pressing the
Scroll to the
data fields data fields
data fields and enter the
data fields data fields
Y measurement dataY measurement data
Y measurement data for the
Y measurement dataY measurement data
1,4 position.
Actual X and Y dataActual X and Y data
Actual X and Y data
Actual X and Y dataActual X and Y data
X and YX and Y
X and Y
X and YX and Y
Y PY P
os fieldos field
Y P
os field
Y PY P
os fieldos field
Increment (Inc) softkeyIncrement (Inc) softkey
Increment (Inc) softkey.
Increment (Inc) softkeyIncrement (Inc) softkey
Actual X and YActual X and Y
Actual X and Y
Actual X and YActual X and Y
X andX and
X and
X andX and
X PX P
Scroll back to the
and increment it to 2 by pressing the
Scroll to the
rement it back to 1 using the
Decrement softkeyDecrement softkey
Decrement softkey.
Decrement softkeyDecrement softkey
Increment (Inc) softkeyIncrement (Inc) softkey
Increment (Inc) softkey.
Increment (Inc) softkeyIncrement (Inc) softkey
Y PY P
os fieldos field
Y P
os field and Dec-
Y PY P
os fieldos field
X P
X PX P
os fieldos field
os field
os fieldos field
4-29
Page 72
Chapter 4Setup
Scale factor
Scroll to the
fields fields
fields and enter the
fields fields
measurement datameasurement data
measurement data for the 2,1
measurement datameasurement data
position.
Continue incrementing X and Y
positions and entering measurement data using this method
until all the data has been entered.
When all the grid data has been
entered, scroll to the
Off data field Off data field
Off data field and press the
Off data field Off data field
softkeysoftkey
softkey to activate NLEC error
softkeysoftkey
correction.
When all the data are entered
for all axes, press the
sor arrowsor arrow
sor arrow or
sor arrowsor arrow
return to the
Scroll through the
to highlight the
function.
Actual X and Y dataActual X and Y data
Actual X and Y data
Actual X and Y dataActual X and Y data
X and YX and Y
X and Y
X and YX and Y
NLEC On/NLEC On/
NLEC On/
NLEC On/NLEC On/
OnOn
On
OnOn
left curleft cur
left cur
left curleft cur
Finish buttonFinish button
Finish button to
Finish buttonFinish button
Setup menuSetup menu
Setup menu.
Setup menuSetup menu
Setup menuSetup menu
Setup menu
Setup menuSetup menu
Scale factorScale factor
Scale factor
Scale factorScale factor
--
-
--
Active
4-30
The values measured on all axes
will be scaled by a user-defined
multiplier when the
tor functiontor function
tor function is active. This func-
tor functiontor function
tion is useful when measuring
parts that shrink or expand under actual-use conditions.
YY
Press the
the
Press the
or Enter key to highlight the
Multiplier data fieldMultiplier data field
Multiplier data field.
Multiplier data fieldMultiplier data field
eses
Y
es softkey to activate
YY
eses
Scale factor functionScale factor function
Scale factor function.
Scale factor functionScale factor function
down cursor arrowdown cursor arrow
down cursor arrow
down cursor arrowdown cursor arrow
Scale fac-Scale fac-
Scale fac-
Scale fac-Scale fac-
Page 73
Quadra-Chek® QC200
Multiplier
User settable
Misc
Enter the
the
the numeric
surements will be multiplied by
this value before they are displayed on the DRO screen.
Press the
the user to enter the
menumenu
menu and set a new
menumenu
value for the
out entering the Supervisor
password.
Scroll through the
to highlight the
The auto-repeat interval for certain keys that repeat when they
are pressed and held and the
volume of the speaker deep can
be configured.
Scale factorScale factor
Scale factor value into
Scale factorScale factor
Multiplier data fieldMultiplier data field
Multiplier data field using
Multiplier data fieldMultiplier data field
keypadkeypad
keypad. All mea-
keypadkeypad
YY
eses
Y
es softkey to allow
YY
eses
SetupSetup
Setup
SetupSetup
MultiplierMultiplier
Multiplier
MultiplierMultiplier
Scale factorScale factor
Scale factor with-
Scale factorScale factor
Setup menuSetup menu
Setup menu
Setup menuSetup menu
Misc functionMisc function
Misc function.
Misc functionMisc function
Key delay
Volume
When certain front panel keys
(such as number keys) are
pressed and held, they will repeat automatically. Enter small val-
KK
ues into the
values for slow repeating.
Enter values from 0 to 10 into the
for quiet beeps or large values for loud beeps.
ey delay data fieldey delay data field
K
ey delay data field for rapid repeating or large
KK
ey delay data fieldey delay data field
VV
olume data field;olume data field;
V
olume data field; small values
VV
olume data field;olume data field;
4-31
Page 74
Chapter 4Setup
4-32
Page 75
Chapter 5
QC100 Series Options
This chapter describes the QC100 options. These options are available at the time of equipment purchase or any time afterward.
The installation of some QC100 options may require the system to
be returned to Metronics for reconfiguration of the software and
hardware. For additional information, please refer to the specific
option description.
Remote keypad
Programmable
footswitch
Non-linear error
correction (NLEC)
The Remote Keypad enables the operator to execute hot key functions or simple key presses from up to nine feet away from the
QC100. The keypad is shown on the next page. Please refer to
Chapter 4: Setup for a detailed description of hotkeys.
The Programmable Footswitch has two keys that can be assigned
a variety of hotkey functions. The footswitch is shown on the next
page. Please refer to Chapter 4: Setup for a detailed description
of hotkeys.
Operation of the QC100 is unchanged when using the keypad or
footswitch. Keys can be assigned to perform user-defined functions using hotkeys.
NLEC is a program that automatically corrects for most non-linear
measurement errors. Please refer to Chapter 4: Setup for a detailed description of NLEC.
5-1
Page 76
Chapter 5Options
QC100 series option
configurations
The QC100 option configurations are shown in the chart below:
NOTENOTE
NOTE
NOTENOTE
A single A single
A single
A single A single
footswitch options to the QC100. The keypad and program-footswitch options to the QC100. The keypad and program-
footswitch options to the QC100. The keypad and program-
footswitch options to the QC100. The keypad and program-footswitch options to the QC100. The keypad and programmable footswith can not be used simultaneouslymable footswith can not be used simultaneously
mable footswith can not be used simultaneously
mable footswith can not be used simultaneouslymable footswith can not be used simultaneously
To install and use the keypad or footswitch, simply plug the phone
plug into the phone jack in the left side of the QC100.
phone jack is provided for attaching the keypad andphone jack is provided for attaching the keypad and
phone jack is provided for attaching the keypad and
phone jack is provided for attaching the keypad andphone jack is provided for attaching the keypad and
..
.
..
5-2
Page 77
Index
AA
A
AA
absolute datum 2-2
Accessing the Setup menu 4-2
Active 4-30
Adjusting LCD contrast 2-11
Adjusting LCD tilt 2-11
Altitude 1-4
Angle format (DD) 3-3
Angle format (DMS) 3-2
Angular formats 3-2
angular measure 2-11
Applying power 2-10
ASCII Codes 3-2, 3-4
assigning hotkeys 4-15
Axis 4-8
axis encoders 2-1
axis labels 4-17
Axis Units 4-18