Philips IntelliVue MP30 Service manual

4 (7)

IntelliVue MP20/30

Service Guide

IntelliVue Patient Monitor

MP20/MP20Junior/MP30

Patient Monitoring

Part Number M8001-9301F 453564112561

*M8001-9301F*

 

 

Contents

 

1 Introduction

9

 

Who Should Use This Guide

9

 

How to Use This Guide

9

 

Abbreviations

9

 

Responsibility of the Manufacturer

10

 

Passwords

11

 

Warnings and Cautions

11

 

 

2 Theory of Operation

13

 

Monitor Theory of Operation

13

 

 

System Boundaries

14

 

 

Hardware Building Blocks

15

 

 

Data Flow

21

 

 

How does the Support Tool Work with the Monitor

24

 

 

Monitor Software Block Diagram

25

 

 

Block Diagram Legend

26

 

 

3 Testing and Maintenance

31

 

Introduction

31

 

Terminology and Definitions

32

 

Recommended Frequency

33

 

When to Perform Tests

34

 

Testing Sequence

37

 

Visual Inspection

38

 

 

Before Each Use

38

 

 

After Each Service, Maintenance or Repair Event

38

 

 

Power On Test

38

 

Safety Tests

39

 

 

Warnings, Cautions, and Safety Precautions

40

 

 

Safety Test Procedures

41

 

System Test

66

 

 

What is a Medical Electrical System?

66

 

 

General Requirements for a System

67

 

 

System Example

67

 

 

System Installation Requirements

69

 

 

Required Protective Measures at System Installation

70

 

 

System Test Procedure

80

 

Preventive Maintenance Procedures

81

 

 

Noninvasive Blood Pressure Measurement Calibration

81

 

 

Microstream CO2 Calibration

81

 

3

Performance Assurance Tests

81

Basic Performance Assurance Test

82

Full Performance Assurance Test

82

ECG/Resp Performance Test

82

ECG Sync Performance Test

83

SpO2 Performance Test

83

NBP PerformanceTest

84

Invasive Pressure Performance Test

86

Temperature Performance Test

87

M3014A Capnography Extension Performance Tests

87

Microstream CO2 Performance Test

90

BIS Performance Test

95

Nurse Call Relay Performance Test

96

Power Loss Alarm Buzzer Performance Test (only if Multi-Port Nurse Call Connector Board is installed)

98

Docking Station Performance Test

98

IIT Communication Test

98

IntelliVue 802.11 Bedside Adapter Communication Test

99

Reporting of Test Results

101

Carrying Out and Reporting Tests

101

Evaluation of Test Results

104

Other Regular Tests

105

Touchscreen Calibration

105

Disabling/Enabling Touch Operation

106

Printer Test Report

107

Battery Handling, Maintenance and Good Practices

107

About the Battery

108

Checking the Battery Status

109

Battery Status on the Main Screen

110

Battery Status Window

112

Conditioning a Battery

114

Conditioning Batteries

115

After Installation, Testing or Repair

116

 

4 Troubleshooting

119

Introduction

119

How To Use This Section

119

Who Should Perform Repairs

119

Replacement Level Supported

120

Software Revision Check

120

Obtaining Replacement Parts

120

Troubleshooting Guide

120

 

Checks for Obvious Problems

121

 

Checks Before Opening the Instrument

121

 

Troubleshooting Tables

123

4

 

 

Status Log

153

List of Error Codes

154

Troubleshooting with the Support Tool

155

Troubleshooting the Individual Measurements or Applications

155

 

5 Repair and Disassembly

157

Tools required

157

How to find the Correct Procedure

158

Removing Directly Accessible Parts

159

1.1

Removing the Handle

160

1.2

Removing the Measurement Server Holder

160

1.3

Removing the I/O Boards and Interface Board

162

1.4

Removing the Recorder

163

1.5

Removing the Navigation Point Knob

164

1.6

Removing the Quick Release Mount and Knob

164

1.7

Removing the Fix Mount

164

2.0 Opening the Front of the Monitor

166

2.1

Removing the Backlight Tubes

167

2.2

Removing the Backlight Inverter Board

168

2.3

Removing Power On LED board and Silicon Pad

169

2.4

Removing the HIF Board, Silicon Pad and Spacer

170

2.5

Removing the Flex Adapter Cable

171

2.6

Removing the Battery Board

172

2.7

Removing Recorder Board Assembly

173

2.8

Removing the Loudspeaker

174

2.9

Removing the Branding Cover

175

3.0

Separating the Front and Back Half of the Monitor

175

3.1

Removing the LCD Flat Panel Display, Touch Panel and Front Housing

176

3.2

Removing the Main Board

177

 

3.3 Removing the MSL Assembly

179

3.4

Removing the Power Supply

180

3.5

Removing the Frame Housing and Serial Number Plate

180

Multi-Measurement Module (MMS) Disassembly

181

 

Tools required

181

 

Removing the Front Cover

181

 

Removing the Mounting Pin

181

 

Removing the Top Cover

182

 

Removing the DC/DC Board

183

 

Removing the MSL Flex Assembly

183

 

Reassembling the MSL Flex Assembly

184

 

Removing the NBP pump

186

 

Refitting the new NBP Pump

186

 

Refitting the DC/DC board

188

 

Refitting the Cover

188

 

Refitting the Front Cover

189

5

Final Inspection

190

MMS Extensions - Exchanging the Top Cover, MSL Flex Cable and the Dual Link Bar

190

Exchange Procedures

191

Disassembly Procedures for the M3015A MMS Extension (HW Rev. A)

202

Removing the Front Cover

202

Refit Procedures for the MMS Extension

206

Smart Battery Charger LG1480 (M8043A)

207

Cleaning the Air Filter Mats

208

Replacing the Fan

208

IntelliVue Instrument Telemetry (IIT)

210

Docking Station

211

Exchanging the Main Board

211

Exchanging the Flex Cable

214

 

6 Parts

217

MP20/MP30 Parts

218

Multi-Measurement Module (MMS) Parts

223

 

MMS Part Number Overview and Identification

223

 

MMS Firmware Overview

225

 

MMS Part Numbers - Front Bezel for M3001 #A01 & #A03

226

 

MMS Part Numbers - Front Bezel for M3001 #A02

226

 

MMS Part Numbers - Top Cover and MSL Assembly

227

 

MMS Exchange Part Numbers

228

 

MMS Part Numbers - Label Kits

230

 

MMS Part Numbers - NBP Assembly

230

MMS Extension Parts (M3012A, M3014A, M3015A and M3016A)

230

 

MMS Extension Part Numbers - Release Mechanisms

231

 

MMS Extension Part Numbers - Top Cover, Flex Cable and Link Bar

231

 

MMS Extension Part Numbers - Front Bezels

231

 

Exchange Parts List

233

IntelliVue X2 Part Numbers

234

 

BISx Solution Replacable Parts

234

Smart Battery Charger Part Numbers

235

IntelliVue Instrument Telemetry Part Numbers

236

IntelliVue 802.11 Bedside Adapter Part Numbers

236

Docking Station Part Numbers

237

External Display Part Numbers

237

Remote Alarm Device Part Numbers

239

Remote Extension Device Part Numbers

240

 

7 Installation Instructions

241

Installation Checklist

241

Unpacking the Equipment

242

Initial Inspection

242

6

 

 

Mechanical Inspection

242

Electrical Inspection

242

Claims For Damage and Repackaging

243

Mounting Instructions

243

Connecting the Monitor to AC Mains

245

Connections

246

Installing Interface Boards

247

Connection of Devices via the MIB/RS232 Interface

250

Connection of USB Devices

250

Installing the Docking Station

254

Installing Remote Devices

254

Mounting the 15” Remote Display (M8031A)

254

Mounting the 15” Remote Display (M8031B)

255

Mounting the 17” Remote Display (M8033A/B/C)

255

Multi-Measurement Module

257

PS/2 Keyboard/Mouse

266

Philips Clinical Network (Wired)

266

Philips Clinical Network (Wireless)

266

Nurse Call Relay

267

Connections

267

ECG Out Functionality

267

Connections

267

Configuration Tasks

268

Checking Country-Specific Default Settings

268

Setting Altitude, Line Frequency, ECG Cable Colors and Height & Weight Units

269

Setting Altitude and Line Frequency

269

Configuring the Equipment Label

269

Configuring the printer

270

Configuring IP Address, Subnet Mask and Default Gateway

270

Configuration Settings for CSCN Routed Bedside Monitors (RBM)

271

Configuring Routed Bedside Monitors Support

271

Handing Over the Monitor

272

 

8 Site Preparation

275

Introduction

275

 

Site Planning

275

 

Roles & Responsibilities

276

Monitor M8001A and M8002A Site Requirements

278

 

Space Requirements

278

 

Environmental Requirements

278

 

Electrical and Safety Requirements (Customer or Philips)

279

Remote Device Site Requirements

280

 

Connecting Non-Medical Devices

281

 

Multi-Measurement Module (MMS) M3001A and IntelliVue X2 M3002A

281

7

 

Remote Displays (M8031A)

284

 

Remote Displays (M8031B)

285

 

Remote Displays - M8033A

286

 

Remote Displays - M8033B

287

 

Remote Displays - M8033C

288

 

Remote Alarm Devices

290

 

Remote Extension Device

291

 

Local Printer

292

Philips Medical LAN

293

RS232/MIB/LAN Interface

293

Nurse Call Relay Interface

294

ECG Out Interface

294

 

9 Gas Analyzers

297

 

10 Index

299

8

1

Introduction

This Service Guide contains technical details for the IntelliVue MP20, MP20 Junior and MP30 Patient Monitor, the Multi-Measurement Module (MMS), the IntelliVue X2, and the Measurement Server Extensions.

This guide provides a technical foundation to support effective troubleshooting and repair. It is not a comprehensive, in-depth explanation of the product architecture or technical implementation. It offers enough information on the functions and operations of the monitoring systems so that engineers who repair them are better able to understand how they work.

It covers the physiological measurements that the products provide, the Measurement Server that acquires those measurements, and the monitoring system that displays them.

Who Should Use This Guide

This guide is for biomedical engineers or technicians responsible for troubleshooting, repairing, and maintaining Philips’ patient monitoring systems.

How to Use This Guide

This guide is divided into eight sections. Navigate through the table of contents at the left of the screen to select the desired topic. Links to other relevant sections are also provided within the individual topics. In addition, scrolling through the topics with the page up and page down keys is also possible.

Abbreviations

Abbreviations used throughout this guide are:

Name

Abbreviation

IntelliVue MP20/MP30 Patient Monitor

the monitor

Multi-Measurement Module

MMS

9

1 Introduction

Responsibility of the Manufacturer

Name

Abbreviation

Measurement Link

MSL

Medical Information Bus

MIB

Anesthetic Gas Module

AGM

Responsibility of the Manufacturer

Philips only considers itself responsible for any effects on safety, EMC, reliability and performance of the equipment if:

-assembly operations, extensions, re-adjustments, modifications or repairs are carried out by persons authorized by Philips, and

-the electrical installation of the relevant room complies with national standards, and

-the instrument is used in accordance with the instructions for use.

To ensure safety and EMC, use only those Philips parts and accessories specified for use with the monitor. If non-Philips parts are used, Philips is not liable for any damage that these parts may cause to the equipment.

This document contains proprietary information which is protected by copyright. All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws.

Philips Medizin Systeme Böblingen GmbH

Hewlett-Packard Str. 2

71034 Böblingen, Germany

The information contained in this document is subject to change without notice.

Philips makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties or merchantability and fitness for a particular purpose.

Philips shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

10

Passwords

1 Introduction

Passwords

In order to access different modes within the monitor a password may be required. The passwords are listed below.

Monitoring Mode: No password required

Configuration Mode: 71034

Demo Mode: 14432

Service Mode: 1345

Consult the configuration guide before making any changes to the monitor configuration.

Warnings and Cautions

In this guide:

-A warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to observe a warning may result in death or serious injury to the user or patient.

-A caution alerts you where special care is necessary for the safe and effective use of the product. Failure to observe a caution may result in minor or moderate personal injury or damage to the product or other property, and possibly in a remote risk of more serious injury.

11

1 Introduction

Warnings and Cautions

12

2

Theory of Operation

Monitor Theory of Operation

The IntelliVue MP20/MP20Junior/MP30 Patient Monitor:

-displays real-time data

-controls the attached measurement server

-alarms in the case of patient or equipment problems

-offers limited data storage and retrieval (trending)

-interfaces to the Philips Clinical Network and other equipment

A monitor with just a single integrated measurement server can be connected to additional building blocks to form a monitoring system with a large number of measurements, additional interface capabilities and slave display. These elements cooperate as one single integrated real-time measurement system.

NOTE

The following descriptions may vary depending on the monitor option purchased.

13

Philips IntelliVue MP30 Service manual

2 Theory of Operation

Monitor Theory of Operation

System Boundaries

The following diagram discusses specific boundaries within the overall system with respect to their openness and real-time requirements:

Measurement LAN

combines components of one patient monitor; real time requirements across all interconnected elements

Philips Clinical Network (wired LAN)

connects multiple patient monitors, information centers, application servers; closed system, only Philips qualified products (tested and with regulatory approval) are connected, Philips is responsible for guaranteed real-time functionality and performance

Philips Clinical Network (wireless)

like Philips Clinical Network (wired) LAN, however due to current wireless technologies available it has reduced bandwidth, longer latencies, reduced functionality

Hospital LAN, Internet

Standard Network, not under Philips control, no guaranteed service, no real-time requirements

14

Monitor Theory of Operation

2 Theory of Operation

Hardware Building Blocks

The following hardware building blocks make up the monitoring system:

IntelliVue MP20

The MP20 monitor:

-integrates the display and processing unit into a single package

-uses a 10.4” TFT SVGA color display

-uses the Philips Navigation Point as primary input device; computer devices such as mice, trackball, and keyboard can be added optionally

-supports the MMS and MMS extensions.

Building Blocks:

15

2 Theory of Operation

Monitor Theory of Operation

IntelliVue MP30

The MP30 monitor:

-integrates the display and processing unit into a single package

-uses a 10.4” TFT XGA color display

-uses the Touchscreen and Philips Navigation Point as primary input devices. Computer devices such as mice, trackball, and keyboard can be added optionally.

-supports the MMS and MMS extensions

Building Blocks:

Optional Hardware

One slot is provided for one of three available system interface boards. If the monitor is ordered with the wireless LAN option a wireless transmitter is required. For further details regarding the wireless network please refer to the M3185A Philips Clinical Network documentation.

16

Monitor Theory of Operation

2 Theory of Operation

Compatible Devices

M8045A Docking Station

M3001A Multi-Measurement Module (MMS)

M3002A IntelliVue X2

17

2 Theory of Operation

Monitor Theory of Operation

M3012A, M3014A, M3015A, M3016A MMS Extensions

Power Supply

Power Supply Architecture

18

Monitor Theory of Operation

2 Theory of Operation

The AC/DC converter transforms the AC power coming from the power plug into 14 V/80W DC source and isolates the monitoring system from the AC power mains.The 14V is distributed via power bus and supplies power to all the components of the system: The 48V DC power needed for the MMS and MMS Extension is created by an isolating DC/DC converter. The power needed for the backlights is converted to 12V DC by the backlight DC/DC converter. The CPU and the non-isolated I/O boards are supplied with 3.3 V and 5 V DC power. Isolated interface boards require a power of 10V AC. The remote HIF board and the LEDs are supplied with 12V DC power.

CPU Boards

The CPU boards have an MPC852/50 MHz processor in the patient monitor and an MPC860/50MHz in the MMS that provides a number of on-chip, configurable interfaces. An array of fast UARTS with configurable protocol options are implemented in an ASIC (along with other system functions such as independent watchdogs etc.), providing interfacing capabilities to measurement modules and System Interface and I/O boards. The serial interfaces can easily be electrically isolated. The main board contains additional video hardware.

The CPUs provide two LAN interfaces to interconnect CPUs (via the MSL) and to connect to the Philips Clinical Network.

The CPU capabilities are identical. Different loading options are coded on serial EEPROMs to support the automatic configuration of the operating system at boot time.

19

2 Theory of Operation

Monitor Theory of Operation

System Interface and I/O Boards

Interfaces to the monitor are implemented via I/O boards. The location of these boards is restricted by general rules. The I/O slot designations diagram and the I/O matrix which outline the I/O board placement rules can be found in the Installation Instructions section.

The following is a list of Interface (I/O) boards which may be present in your monitor, depending on your purchased configuration:

System Interface boards:

-Video for slave display

-Philips Clinical Network (LAN wired or wireless)

-Basic Alarm Relay (Nurse Call)

-Docking Interface

I/O boards:

-PS/2

-MIB/RS232

-Flexible Nurse Call

-Parallel printer

-USB

-Remote devices (Remote Alarm Device, Remote Extension Device)

-BISx Interface

-IntelliVue 802.11 Bedside Adapter

The specifications for the above listed interfaces can be found in the technical data sheet for the monitor and in the Installation and Specifications chapter of the Instructions for Use.

20

Monitor Theory of Operation

2 Theory of Operation

Data Flow

The following diagram shows how data is passed through the monitoring system. The individual stages of data flow are explained below.

Data Acquisition

Monitoring data (for example patient measurement data in the form of waves, numerics and alerts) is acquired from a variety of sources:

-Measurement Server

The Measurement Server connected to the internal LAN convert patient signals to digital data and applies measurement algorithms to analyze the signals.

-External measurement devices

Data can be also acquired from devices connected to interface boards of the monitor. Software modules dedicated to such specific devices convert the data received from an external device to the format used internally. This applies to the Anesthetic Gas Module.

-Server systems on the Philips Clinical Network

To enable networked applications such as the other bed overview, data can be acquired from server systems attached to the Philips Clinical Network, for example a Philips Information Center

21

2 Theory of Operation

Monitor Theory of Operation

Data Provider System Service

All data that is acquired from measurement servers or external measurement devices is temporarily stored by a dedicated data provider system service. All monitor applications use this central service to access the data in a consistent and synchronized way rather than talking to the interfaces directly.

This service makes the applications independent of the actual type of data acquisition device.

The amount of data stored in the data provider system service varies for the different data types. For example several seconds of wave forms and the full set of current numerical values are temorarily stored in RAM.

Persistent Data Storage System Service

Some applications require storage of data over longer periods of time. They can use the persistent data storage system service. Dependent on the application requirements, this service can store data either in battery backed-up (buffered) memory or in flash memory. The buffered memory will lose its contents if the monitor is without power (not connected to mains) for an extended period of time. The flash memory does not lose its contents.

The trend application for example stores vital signs data in a combination of flash memory and buffered memory, while the system configuration information (profiles) is kept purely in flash memory.

Display and User Interface Service

Applications can use high level commands to display monitoring data or status and command windows on the internal LCD panel. These commands are interpreted by the display manager application. This application controls the dedicated video hardware which includes video memory and a special ASIC.

User input is acquired from a variety of input devices, for example the Navigation Point, the touchscreen or other standard input devices (keyboard, mouse) which may be attached to I/O boards. The system software makes sure that the user input is directed to the application which has the operating focus.

Data Output

The monitoring system is very flexible and customizable regarding its data output devices. Built-in devices (for example LAN, alarm lamps, speaker, video) provide the basic output capabilities.

These capabilities can be enhanced by adding additional I/O boards, as required in the specific end-user setup. The additional I/O boards typically provide data to externally attached devices, for example to printers, RS232 based data collection devices, nurse call systems etc.

The monitor can identify I/O boards by means of a serial EEPROM device that stores type and version information. The operating system detects the I/O boards and automatically connects them with the associated (interface driver) application. For some multi-purpose cards it is necessary to configure the card for a particular purpose first (for example the dual MIB/RS232 card can support external touch display, data import, data export).

22

Monitor Theory of Operation

2 Theory of Operation

Monitor Applications

The monitor applications provide additional system functionality over the basic measurement and monitoring capabilities. This includes for example trending, report generating, event storage or derived measurements.

In general, the monitor applications use the data provider system service to access the measurement data. Application interfaces to the other system services allow the application to visualize data, to store data over extended periods of time or to output data to other devices.

Internal LAN (Measurement Link)

All components of the monitoring system (including measurement servers and CPUs in the monitor) communicate using an IEEE802.3/ Ethernet LAN in the Measurement Link (MSL). This network is used to distribute data between the components, for example:

-Digitized patient signals including wave data, numerical data and status information (typically from the measurement server to a display unit)

-Control data representing user interactions (typically from the display unit to a measurement server)

-Shared data structures, for example representing patient demographical data and global configuration items

The internal LAN allows plug and play configuration of the monitoring system. The system automatically detects plugging or unplugging of measurement servers and configures the system accordingly.

The components on the internal LAN are time-synchronized to keep signal data consistent in the system. Dedicated hardware support for synchronization eliminates any latency of the network driver software.

The integrated LAN provides deterministic bandwidth allocation/reservation mechanisms so that the real-time characteristic of signal data and control data exchange is guaranteed. This applies to the data flow from the measurement server to the monitor (for example measurement signal data) and the data flow from the monitor to a measurement server (for example to feed data to a recorder module).

Integrated communication hubs in the monitor allow flexible cabling options (star topology, daisy chaining of servers).

23

2 Theory of Operation

Monitor Theory of Operation

Philips Clinical Network

The monitoring system may be connected to the Philips Clinical Network, for example to provide central monitoring capabilities or other network services. This connection may be through a normal wired connection or through a wireless connection.

The monitor supports the connection of an external wireless adapter or an internal wireless adapter (#J35). Switching between wired and wireless networks is automatically triggered by the plugging or unplugging of the network cable.

The Philips Clinical Network protocols function very similarly to the protocols used on the internal LAN.

After configuration, the monitoring system sends the digitized patient signals including wave data, numerical data and status information onto the network. Control data representing user interactions can be exchanged between the monitoring system and a central station bi-directionally.

Additional protocols are supported for networked applications, for example for the other bed overview function, which allows viewing of monitoring data from other patients on the network.

For plug and play operation, the monitoring system uses the standard BootP protocol to automatically acquire a network address.

How does the Support Tool Work with the Monitor

The support tool is a Windows application typically installed on the laptop of a customer engineer or a biomedical engineer working in the customer’s own service department.

The purpose of the support tool is to upgrade, configure and diagnose all monitoring components (modules, measurement servers, and monitors) in the system over the network.

The service protocol developed for this purpose uses a raw access to the devices without the need for IP addresses etc. over a standard customer network installation, so that even defective devices can be upgraded as long as the few kBytes of initial boot code are working. The boot code itself can also be upgraded using the same protocol.

The tool allows access to internal service information and to serial numbers. It can be remotecontrolled, for example via a dial-up connection from a response center, provided the proper infrastructure is in place.

For details see the Instructions for Use for the Support Tool.

24

Monitor Theory of Operation

2 Theory of Operation

Monitor Software Block Diagram

shows the functional block diagram for the monitoring system. A legend explaining terms and diagram elements follows. The information below varies depending on the purchased monitor options.

IntelliVue Patient Monitoring System Functional Block Diagram

25

2 Theory of Operation

Monitor Theory of Operation

Block Diagram Legend

Functional Block

Services

Operating System

System Services

Description

The Operating System (OS) provides a layer of isolation between the specific hardware implementation and the application software. The OS performs system checks and allocates resources to ensure safe operation when the system is first started. This includes internal self-tests on several hardware modules and configuration checks for validity of configuration with the operating software. During normal operation, the OS continues to run checks on system integrity. If error conditions are detected the OS will halt monitoring operations and inform the operator about the error condition.

The System Services provide generic common system services. In particular:

They use a real-time clock component to track time. They synchronize to network time sources and verify the accuracy of the system time information. They are also responsible for managing persistent user configuration data for all Measurement Servers and IntelliVue Patient Monitoring System software modules. User configuration data is stored in a non-volatile read/write storage device

Applications

Reports

The Reports Service retrieves current and stored physiological data and status data to format reports for printing paper documentation. The following reports are supported:

-Vital Signs Report

-Graphical Trend Report

-Event Review Report

-Event Episode Report

-ECG Report (12 Lead/Multi-Lead)

-Cardiac Output Report

-Calculations Report (Hemodynamic/Oxygenation/Ventilation)

-Calculations Review Report

-Wedge Report

-Test Report

The Reports service generates report data which can be printed on a local or a central printer.

26

Monitor Theory of Operation

2 Theory of Operation

 

 

 

Functional Block

 

Description

Record

 

The Record Service retrieves current and stored physiological data and status data

 

 

to format a continuous strip recording. A recording can be triggered manually by

 

 

the operator or automatically by an alarm condition. The Record Service uses the

 

 

services of the Recorder Interface to control an M1116B Recorder. The Record

 

 

Service can also send data to a central recorder.

Alarm

 

The Alarm Service contains logic that prioritizes alarm conditions that are

 

 

generated either by the Measurement Servers or by IntelliVue Patient Monitoring

 

 

System software modules. Visual alarm signals (messages) are displayed at the top

 

 

of the IntelliVue Patient Monitoring System display and alarm sounds are

 

 

generated by a loudspeaker. Alarm conditions may be generated when a

 

 

physiological parameter exceeds preselected alarm limits or when a physiological

 

 

parameter or any other software module reports an inoperative status (technical

 

 

alarm, for example, the ECG leads may have fallen off the patient). The Alarm

 

 

service manages the alarm inactivation states, for example suspension of alarms,

 

 

silencing of alarms, and alarm reminder. Alarm signals may also be configured as

 

 

latching (alarm signals are issued until they are acknowledged by the operator,

 

 

even when the alarm condition is no longer true). The Alarm service controls the

 

 

visual alarm signals (alarm lamps).

Trend

 

The Trend service stores the sample values of physiological data and status data

 

 

with a resolution of 12 seconds, 1 minute or 5 minutes for a period of up to 48

 

 

hours. The data is kept in battery buffered read/write storage and flash memory

 

 

devices to be preserved across power failures. The stored data is protected via

 

 

consistency checks and checksums. When a new patient is admitted, the trend

 

 

database erases all data of the previous patient.

HiRes

 

The OxyCRG (Oxygen CardioRespiroGram) service derives a high-resolution

 

 

trend graph from the Beat-to-Beat Heart Rate, SpO2 or tcpO2, and Respiration

 

 

physiological data. The OxyCRG is specialized for neonatal applications, allowing

 

 

the operator to identify sudden drops in Heart Rate (Bradycardia) and SpO2

 

 

(Desaturation), and supporting the operator in visualizing Apnea situations.

ADT

 

The ADT (Admit/Discharge/Transmit) service maintains the patient demographics

 

 

information. The operator may admit a new patient, discharge the old patient and

 

 

enter or modify the patient demographics. The ADT service also supports the

 

 

transport of a patient (trend database) with the M3001A Multi-Measurement

 

 

Module. The ADT service controls the deletion of old patient data, the upload of

 

 

trend data from the M3001A and the switching back of all settings to user

 

 

defaults. It also synchronizes patient information with a central station on the

 

 

network.

Calc Param

 

The Calc Param (Calculated Parameters) service accesses current, stored and

 

 

manually entered physiological data as input to calculation formulas. With these

 

 

formulas, derived hemodynamic, oxygenation and ventilation variables are

 

 

computed. The calculation results, including the input parameters, are stored for

 

 

later review using the Trend service.

 

 

 

27

2 Theory of Operation

Functional Block

Interface Managers

MDSE

Printer

Display & Operator

Interface

Interfaces

LAN

Centronics

Monitor Theory of Operation

Description

The MDSE (Medical Data Service Element) Interface Manager is responsible for the exchange of real-time data between the IntelliVue Patient Monitoring System display unit and the Measurement Servers and Flexible Module Rack as well as between the IntelliVue Patient Monitoring System display unit and other devices attached to the network. MDSE establishes and maintains a data communication link between the devices. It provides configuration information about the remote device to applications in the local device and it allows the exchange of measurement data and status information between the devices.

The Printer Interface Manager provides a high level interface to a printer. It provides means to:

-establish a connection to the printer

-transfer data to the printer

-get status of the printer

-close connection to the printer

The Printer Interface Manager also supervises the connection to the printer and whether the printer accepts data (for example paper out). The Printer Interface Manager notifies the operator in such cases.

The Display and Operator Interface Manager performs the following tasks:

-Screen presentation of real-time and stored physiological measurement data, alarm condition data and status information received from the MDSE interface manager, the Alarm service or other IntelliVue Patient Monitoring System modules

-Screen presentation of operating controls (control windows)

-Processing of operating control commands received from HIF Control interface. The module verifies and interprets the received commands and forwards them to other software modules of the IntelliVue Patient Monitoring System display unit, Measurement Servers or Flexible Module Rack

-Sound generation (issues audible alarm signals and generates audible information signals, for example QRS and SpO2 tones, operator audible feedback)

The LAN interface implements the physical layer of IEEE 802.3. The LAN interface performs Manchester encoding/decoding, receive clock recovery, transmit pulse shaping, jabber, link integrity testing, reverse polarity detection/correction, electrical isolation, and ESD protection. Electronically separated interfaces are used for communication to the Measurement Servers or Flexible Module Rack and to the network.

The Centronics interface implements the standard signaling method for bi-directional parallel peripheral devices according to IEEE 1284-I. The interface is used as a parallel interface to a standard printer with electrical isolation and ESD protection.

28

Monitor Theory of Operation

2 Theory of Operation

 

 

 

Functional Block

 

Description

Display Controller

 

The Display Controller Interface consists of a video controller chip, video RAM

 

 

and the controlling software. The Display Controller interface processes the high

 

 

level display commands (character and graphic generation, wave drawing) and

 

 

translates them into pixels, which are written into the video RAM where the video

 

 

controller chip generates the video synchronization signals and the pixel stream

 

 

for the Color LCD Display.

HIF Control

 

The HIF (Human Interface Control) interface scans the Human Interface devices

 

 

for operator controls (Touch Screen, Speed Point, USB and PS/2 devices), formats

 

 

the collected data and sends it to the display and Operating Interface.

ECG-Out Marker-In

 

The ECG Out/Marker In interface receives the ECG waveform directly from the

 

 

ECG/Resp Arrhythmia ST-Segment physiological algorithm via an RS-422 serial

 

 

interface and converts the digital ECG signal to an analog ECG signal. In addition,

 

 

the ECG Out controller receives from a connected device the marker information

 

 

and forwards this data to the ECG/Resp Arrhythmia ST-Segment physiological

 

 

algorithm. The converted analog signal is used to synchronize a connected device

 

 

to the patient’s ECG

RS-422

 

The serial link RS-422 interface communicates the ECG signal to the ECG

 

 

Output/Marker In of the IntelliVue Patient Monitoring System display unit. The

 

 

interface is a serial, differential, full-duplex link. The interface is ESD protected.

PS/2

 

The PS/2 interface supports the serial protocol of standard PS/2 devices (mouse).

 

 

The PS/2 serial protocol is interpreted by the HIF Control interface.

Nurse Call

 

The Nurse Call board contains 2 connectors. A phone jack type connector and a

 

 

multi-port connector. The phone jack type connector has a single close-on-alarm

 

 

relay. The multi-port connector has three alarm relays which are configurable to

 

 

be open or closed on alarm. In addition, this interface has an audible alert

 

 

capability for loss of AC power.

MIB

 

The MIB interface allows full-duplex, short-haul asynchronous binary

 

 

communication between the monitor and an arbitrary (medical/non-medical)

 

 

device using an eight-pin RJ45 modular connector. Switching between MIB and

 

 

RS232 protocol is possible.

Docking Interface

 

The docking interface provides necessary connections for docking an MP20/30 or

 

 

MP40/50 monitor onto the docking station.

BISx Interface

 

The BISx interface allows the usage of Aspect’s BISx solution with an MP20/30

 

 

monitor.

IIT Interface

 

The IIT interface allows operation of the MP2/X2/MP5/MP20/MP30/MP40/MP50

 

 

monitors with IntelliVue Instrument Telemetry.

 

 

 

29

Loading...
+ 280 hidden pages