Siemens UROSKOP D3 Instruction Manual

5 (1)

UROSKOP D3

SP

Start-up Instructions

with POLYDOROS SX and

FLUOROSPOT H with Supervision

 

 

 

 

© Siemens AG 1998

 

The reproduction, transmission or

 

use of this document or its contents

 

is not permitted without express

 

written authority. Offenders will be

 

liable for damages.

All rights,

 

including rights created by patent

 

grant or registration of a utility

 

model _or_ design,_are_ reserved.

Register 4

English

 

Print No.: RLL5-310.034.04.04.02

Doc. Gen. Date:

12.98

Replaces: RLL5-310.034.04.03.02

0 - 2

 

 

Revision

 

 

 

 

Chapter

Page

Revision

 

 

 

 

 

1

all

04

 

 

 

 

 

2

all

04

 

 

 

 

 

3

all

04

 

 

 

 

 

4

all

04

 

 

 

 

 

5

all

04

 

 

 

 

 

6

all

04

 

 

 

 

 

7

all

04

 

 

 

 

 

8

all

04

 

 

 

 

 

9

all

04

 

 

 

 

 

10

all

04

 

 

 

 

 

11

all

04

 

 

 

 

 

12

all

04

 

 

 

 

 

UROSKOP D3

Register 4

RLL5-310.034.04

Page 2 of 4

Siemens AG

 

 

Rev. 04 12.98

TD PS 24

Medical Engineering

Page

1 _______General information ____________________________________________ 1 - 1

Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1 General information and information on documentation . . . . . . . . . . . . . . . . 1 - 1 Preliminary information on image quality test . . . . . . . . . . . . . . . . . . . . . . 1 - 2 Purpose: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2 Sequence of work: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2 Requirements for start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2 Required Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2 Required measuring instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3

2 _______Start-up preparation ____________________________________________ 2 - 1

Protective measures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1 Measuring the line voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1 Checking the on-site emergency shutdown button . . . . . . . . . . . . . . . . . . . 2 - 1 Checking the phase connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2 Measuring the internal line impedance . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2 Checking unacceptable ground connections . . . . . . . . . . . . . . . . . . . . . . 2 - 3 Switching on the generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3

3 _______Generator functions ____________________________________________ 3 - 1

Checking the emergency stop switches . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1 Checking the control panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1 Function check of the anode starter . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1 Key to 7-segment display on board D95: . . . . . . . . . . . . . . . . . . . . . . 3 - 1

LED status display . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

.

.

. . . 3

- 1

Checking the fluoroscopic and exposure release circuit

. . . . . . . . . . . . . . . . 3 - 2

Fluoroscopic circuit . . . . . . . . . . . . . . . . .

.

. . . . . . . . . .

.

.

. . . 3

- 2

Exposure release circuit . . . . . . . . . . . . . . . .

. . . . . . .

. . . . . .

.

. 3

- 2

Controls and indicator lamps for radiation protection . . . . . . . . . .

. . . . . .

.

. 3 - 3

Door contact for radiation disable function . . . . . . . . . . . . . . . . . . . . . 3 - 3 Indicator lamps for tube assembly selection. . . . . . . . . . . . . . . . . . . . . 3 - 3 Checking the kV/mA values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 3 Checking the mAs values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 4 Start-up of the automatic exposure control . . . . . . . . . . . . . . . . . . . . . . . 3 - 4

4 _______Unit movements________________________________________________ 4 - 1

Preparation and limit values for tabletop lift . . . . . . . . . . . . . . . . . . . . . . . 4 - 1

Lifting movement . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . .

.

. 4

- 2

Oblique projection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

- 3

Tilting movement . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . .

.

. 4

- 4

Longitudinal movement of the tabletop . . . . . . . . . . . . . . . . . . . . . . . . . 4

- 5

Transverse movement of the tabletop . . . . . . . . . . . . . . . . . . . . . . . . . . 4

- 6

Siemens AG

Register 4

RLL5-310.034.04

Page 3 of 6

UROSKOP D3

Medical Engineering

 

Rev. 04 12.98

TD PS 24

 

Page

5 ______ Radiation geometry _____________________________________________5 - 1

Test conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 2 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 3 Test conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 4 Brief description of the test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 4 Visual check of the fluoroscopic field limitation . . . . . . . . . . . . . . . . . . . 5 - 5 Checking the overframing of the fluoroscopic field size displayed on the monitor . 5 - 5 Checking the centering of radiation field and monitor image center . . . . . . . . 5 - 6 Evaluation of overframing in fluoroscopy . . . . . . . . . . . . . . . . . . . . . . 5 - 6 Evaluation of off-center in fluoroscopy . . . . . . . . . . . . . . . . . . . . . . . 5 - 6

6 ______ Checking the tomographic device _________________________________6 - 1

Accuracy of the tomographic height display . . . . . . . . . . . . . . . . . . . .

6 - 2

Resolution and tomographic procedure . . . . . . . . . . . . . . . . . . . . . .

6 - 2

7 ______ Fluoroscopic dose rate __________________________________________7 - 1

General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1

Preliminary remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7

- 1

Basic information on the dose / dose rate measurement for I.I. workstations . . .

7 - 1

Determining the dose/dose rate directly at the I.I. input . . . . . . . . . . . . . .

7 - 1

Formulas for calculating the dose/dose rate in the measurement plane for

 

 

dose/dose rate setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7

- 2

Indirect dose rate test at the customer site . . . . . . . . . . . . . . . . . . . . . . .

7 - 3

Setting when the indirect dose rate is out of tolerance . . . . . . . . . . . . . . . . .

7 - 3

Maximum skin dose rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7

- 4

8 ______ Television system ______________________________________________8 - 1

Function check of ambient light sensor. . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1 B-signal values (without bias light/dark current component) . . . . . . . . . . . . . . 8 - 1 Image artifacts (Para. 5.10 Chapter 5D) . . . . . . . . . . . . . . . . . . . . . . 8 - 2

9 ______ Digital Fluoro Radiography/FLUOROSPOT H ________________________9 - 1

General information and remarks . . . . . . . . . .

. . . . . . . . . . . . . . . . . 9 - 1

Dose/pulse test during start-up (indirect dose check)

. . . . . . . . . . . . . . . . . 9 - 5

BA (video+blanking) signal values / setting the iris diaphragm in DR operation . . . .

9 - 6

Dynamic test . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . .

9 - 7

Edge enhancement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 11 Mean value assessment (GGM= sliding weighted averaging) . . . . . . . . . . .9 - 11 Resolution and minimum contrast . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 12 Checking the DSA system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 14 Checking the hardcopy camera . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 16 Adjusting the hardcopy camera. . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 16

UROSKOP D3

Register 4

RLL5-310.034.04

Page 4 of 6

Siemens AG

 

 

Rev. 04 12.98

TD PS 24

Medical Engineering

Page

Preparing and evaluating the SMPTE test image . . . . . . . . . . . . . . . . . 9 - 16 Image artifacts and transmission interference. . . . . . . . . . . . . . . . . . . 9 - 17 Image artifacts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 18 Definition of the rating numbers:. . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 18 Description of the artifacts: . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 18

10 ______Nominal values _______________________________________________ 10 - 1

I.I. nominal values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1 ADC measuring field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1 MPL system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1 SDM system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1

Fluoroscopic dose rate at the I.I. input

. . . . . . . . .

. . . . .

. . .

. . . . . .

.

10 - 2

VIDEOMED SX, H1X Emulation . . .

. . . . . . . . .

. . . . .

. . .

. . . . . .

.

10 - 3

TV monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 3 TV camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 3

B signal value (interlaced - Bypass) . . . . . . . . . . . . . . . . . . . . . . .

10 - 3

Vignetting (with bias light) (interlaced - Bypass) . . . . . . . . . . . . . . . . .

10 - 4

AGC measuring field sizes in grid units (GU) . . . . . . . . . . . . . . . . . . .

10 - 4

Pedestal and AGC function

 

Pedestal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

.

.

. . 10

- 4

Fixed gain TV unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

.

.

. . 10

- 4

Dynamic test . . . . . . . . . . . .

. . . . . . . . . . . . . .

. . .

. . . . . .

.

10

- 5

Capillary test . . . . . . . . . . . .

. . . . . . . . . . . . . .

. . .

. . . . . .

.

10

- 5

Resolution and minimum contrast of the I.I.-TV system (interl. - Bypass) . . . . 10 - 6

Image artifacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10 - 6

Digital Fluoro Radiography / DFR . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

- 6

FLUOROSPOT H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10

- 6

Dose/pulse at the I.I. input . . . . .

. . . . . . . . . . . . . . . . . . . .

. . .

.

10

- 6

Amplification of DFR system. . . .

. . . . . . . . . . . . . . . . . . . .

. . .

.

10

- 7

Monitor adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 8 BA signal values / Adjustment of iris diaphragm in DR mode . . . . . . . . . . . 10 - 8 Dynamic test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 9

Black level cutoff (noise halo) . . . . . . . . . . . . . . . . . . . . . . . . . .

10 - 10

Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10

- 10

Minimum contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10

- 10

Checking the DSA equipment . . . . . . . . . . . . . . . . . . . . . . . . . .

10 - 11

Image artifacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10

- 11

11 ______Final steps and electrical safety _________________________________ 11 - 1

Final work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 1 Measuring the protective conductor resistance . . . . . . . . . . . . . . . . . . . . 11 - 1

12 ______Changes to previous version ____________________________________ 12 - 1

Siemens AG

Register 4

RLL5-310.034.04

Page 5 of 6

UROSKOP D3

Medical Engineering

 

Rev. 04 12.98

TD PS 24

 

Page

This page intentionally left blank.

UROSKOP D3

Register 4

RLL5-310.034.04

Page 6 of 6

Siemens AG

 

 

Rev. 04 12.98

TD PS 24

Medical Engineering

Safety information

CAUTION

When performing service work and tests, please adhere to the product-specific safety information contained in the document, as well as the general safety information contained in Register 2 of the TI binder.

Tests and adjustments performed with radiation ON are identified with the radiation warning symbol . Radiation protection must be worn during these types of adjustments.

General information and information on documentation

These instructions contain the checks necessary for system startup.The sequence in which they are given must be maintained.

The unit movements necessary for these procedures must be carried out very carefully to avoid damaging the equipment in the event of a malfunction.

Since the unit has been adjusted completely at the factory, only the functioning and factory settings (accuracy of tolerance) are checked; the unit must be adapted to on-site requirements, if necessary.

If any mechanical or electrical malfunctions occur or if tolerances are exceeded (damages caused by transport etc.), perform service work and adjustments according to adjustment instructions RLL5-310.071.01... or the service instructions for the corresponding components.

Measurement results marked” ” must be entered in the UROSKOP test certificate.

For the acceptance test which is required in Germany in accordance with §16 of the Röntgenverordnung (x-ray ordinance) and for the acceptance test required in the USA, the following checks have already been performed in the factory test area and documented in the test certificate:

Visual check of the filter values.

Check of the SID display.

Brightness of the light localizer.

Coincidence of light field and radiation field.

Accuracy of the manual format collimation.

Coincidence of radiation field and film center.

Accuracy of positive beam limitation (PBL) system for cassette formats.

Fluoroscopic field limitation.

Centering of the radiation field and monitor image center.

Function check of the Iontomat measurement fields.

Unit attenuation factor.

Check of the tomographic device.

Siemens AG

Register 4

RLL5-310.034.04

Page 1 of 4

UROSKOP D3

Medical Engineering

 

Rev. 04 12.98

TD PS 24

 

NOTICE

 

Enter the measurement values and the additional values deter-

 

 

mined during system start-up from the test certificate into the

 

 

 

 

acceptance certificate (Germany) or in the acceptance test certifi-

 

 

cate (USA).

 

 

 

Preliminary information on image quality test

Purpose:

The image quality test provides objective data as a basis for image quality and facilitates system start-up.

The test certificate is valid for system acceptance and for future service and maintenance work.

Sequence of work:

Follow the sequence of measurements and checks exactly as describes in order to ensure correct measurements.

Enter the results of each measurement into the test certificates.

Compare the measured values with the values indicated in the list of nominal values.

If the limit values are exceeded, you must determine the cause and correct the error or the erroneous setting before continuing the test. Otherwise the measurements that follow may be incorrect.

Requirements for start-up

The system must be cabled completely.

Switch on high voltage only if radiation is necessary.

Required Documents

Operating instructions

UROSKOP D3

RL5-340.208.01...

 

Wiring diagram

UROSKOP D3

G5403

 

Wiring diagram

POLYDOROS SX X2075

 

Service software

 

 

 

operation

UROSKOP D3

RLL5-310.113.02...

Logbook, Reg. 10

Setting Instructions

POLYDOROS SX RX63-050.032.02...

 

System configuration

POLYDOROS SX RX63-050.034.06...

 

IQ Test certificate

 

RXD0-000.037.01...

Logbook, Reg. 9

Test certificate

UROSKOP D3

RLL5-310.035.01...

Logbook, Reg.3

Test certificate

POLYDOROS SX X2075

Logbook, Reg. 4

UROSKOP D3

Register 4

RLL5-310.034.04

Page 2 of 4

Siemens AG

 

 

Rev. 04 12.98

TD PS 24

Medical Engineering

Required measuring instruments

3-Phase field rotation meter, electronic

97 02 713

Y7933

Internal line impedance meter

84 28 104

RE999

Service PC

 

 

Oscilloscope > 50 MHz, e.g. Tektronix 2232

97 02 234

Y3155

with delayed time base

 

 

Multimeter 8060 A (Fluke)

97 02 101

Y4290

mAs meter

81 60 400

RE999

Water level

28 69 436

RE999

Set of Cu filters

44 06 120

RV090

17 m Cu strips (included in the LOG book)

11 67 663 G5247

cm scale

 

 

Tomographic cube with resolution test

44 06 054

RV 090

Protective conductor meter

44 15 899

RV090

8x magnifying glass

44 14 850

RH090

Precision radiation filter

99 00 598

XE999

Centering cross or lead ruler

96 60 051

RE999 or 28 63 025 RE999

Mavo monitor, direct ordering

97 02 432

Y0526

Resolution test patterns, types 41

28 71 820

RE999

Dynamic test case

37 90 156

X1963

heart contour diaphragm

37 90 172

X1963

Capillary test

37 90 180

X1963

DSA displacement test (with DSA capillary

97 50 019

X1963 (only with DSA)

test)

 

 

Aperture plate for tomography

 

 

 

 

 

 

Successful system start-up can only be ensured by using the

 

CAUTION

 

 

 

 

 

 

above measurement instruments.

 

 

 

 

 

 

 

 

 

 

 

 

 

Siemens AG

Register 4

RLL5-310.034.04

Page 3 of 4

UROSKOP D3

Medical Engineering

 

Rev. 04 12.98

TD PS 24

 

This page intentionally left blank.

UROSKOP D3

Register 4

RLL5-310.034.04

Page 4 of 4

Siemens AG

 

 

Rev. 04 12.98

TD PS 24

Medical Engineering

Protective measures

Observe the protective measures described in Register 1 of the UROSKOP D Logbook.

Prior to performing any service work on the generator, switch it off with the power OFF switch on D200.

In order to deenergize all parts of the system, set the system switch to the OFF position.

CAUTION

With the generator switched off, line voltage is still present at transformer T1 and fuse panel D200 (refer to wiring diagram X2075-11).

After switching off the generator, approx. 600 V DC voltage are still present at the inverter; indicated by LEDs V8 and V9 on board D250 (see X2075-17).

Within approx. 1 1/2 minutes the voltage drops to 0 V; the LEDs go out at approx. 30 V.

To prevent unintentional release of high voltage or radiation, set switch SS on D200 to OFF (no actuation of the inverters).

Remove or insert boards only with the generator switched off. When doing so, observe the ESD guidelines (see TI 219, Info RA.5. RA0-000.012.19... on microfilm).

On D211: switch S1 INT OFF (position 2).

On D200: switch SS OFF.

Measuring the line voltage

Remove fuses F1, F2 and F3 from the fuse board in M16 (generator).

System contactor ON.

Measure the line voltage.

The voltage must correspond to the values indicated in the test certificate (logbook, Reg. 3).

Checking the on-site emergency shutdown button

Actuate the emergency shutdown button installed by the customer. The entire system must be disconnected from the power line.

System contactor OFF.

Siemens AG

Register 4

RLL5-310.034.04

Page 1 of 4

UROSKOP D3

Medical Engineering

 

Rev. 04 12.98

TD PS 24

 

Checking the phase connection

Check the correct phase connection of the power lines at M16.F1, F2, F3 (X2075-11), using the rotating field instrument:

In M16, remove the cover from the fuse holders.

Connect the3-phase field rotation meter to the lower connections of F1, F2, F3 (on the side of the power supply).

System contactor ON.

Perform the measurement.

System contactor OFF.

Disconnect the rotating field instrument.

Measuring the internal line impedance

Measure the internal line impedance at M16. F1. F2. F3 (X2075-11):

-Connect the internal line impedance meter between two phases of F1, F2, F3 (at the lower connections of the fuse holders).

-System contactor ON.

-Perform the measurement.

-System contactor OFF.

Record the measured values in the test certificate (logbook Reg. 3)

Enter the measured values into the ”LINE PARAMETER” module of the system programming.

Important: In order to obtain the full generator power, the internal line impedance measured may not exceed the following values:

Uline

max. Rline withPOLYDOROS

measured

SX 50

SX 80

 

 

 

 

 

 

400 V

0.18 Ohm

0.12 Ohm

 

 

 

 

 

440 V

0.20 Ohm

0.16 Ohm

 

 

 

 

 

480 V

0.24 Ohm

0.20 Ohm

 

 

 

 

 

Internal line impedances of > 250 mOhm reduce power. Refer to RX63050.034.05.01.05... Configuration ”LINE PARAMETER”

Reinsert fuses F1, F2 and F3.

Reattach the cover.

UROSKOP D3

Register 4

RLL5-310.034.04

Page 2 of 4

Siemens AG

 

 

Rev. 04 12.98

TD PS 24

Medical Engineering

Checking unacceptable ground connections

(connection between and 0VL ).

Disconnect the connection lead between the upper and lower grounding bus (in the generator power cabinet).

Using the multimeter, measure the resistance between and 0VL: The measured resistance must be approx. > 50 kΩ.

Reconnect the upper and the lower grounding bus.

Switching on the generator

System contactor ON:

-LED V26 (via fuse F11) must light up on fuse board D200. In case of an error, check the fuses in M16 and on D200 (X2075-11 and 12).

Generator ON:

-On fuse board D200, press key and actuate button .

-The LED’s must light up as follows on the service panel of D211:

The +5 V, ± 15 V, +24 V supply voltages are immediately present.

If the +5 V voltage is missing, the generator is disabled.

In this case, check the fuses on D200 (X2075-12). All LED’s on D200 must light up.

-Following connection of the intermediate circuit (activation of charging contactor and power contactor), LEDs V9 and V8 on board D250 must light up (X2075-17).

 

 

 

 

A DC voltage > 500 V is applied to the intermediate circuit.

 

CAUTION

 

 

 

 

 

 

 

 

 

 

 

 

-Following correct initialization (after approx. 10 sec.), the following LEDs must light on the service panel of D211:

( ) ( ) TU1 (after system programming) GEN-OK,+24 V, +15 V, -15 V, +5 V

If no error has occurred, the 7-segment displays on D210 indicate ”E000”. In case of an error, the three-digit error code is displayed.

- When programing the generator, "ALP" appears on the kV display of the control deck.

Check the primary voltage for the supply voltages (X2075-12) at the coil connections of

the GS contactor (X2075-11) in M16: 230 V ± 10 % at 50 Hz

250 V ± 10 % at 60 Hz

Check the blocking:

On D200: Generator OFF

The generator cannot be switched on again from the deck.

Siemens AG

Register 4

RLL5-310.034.04

Page 3 of 4

UROSKOP D3

Medical Engineering

 

Rev. 04 12.98

TD PS 24

 

This page intentionally left blank.

UROSKOP D3

Register 4

RLL5-310.034.04

Page 4 of 4

Siemens AG

 

 

Rev. 04 12.98

TD PS 24

Medical Engineering

Checking the emergency stop switches

System ON.

Press the three system emergency stop switches consecutively: With the emergency stop switch pressed in,

-error code ’’U01’’ must appear in the area dose product display on the control console;

-all system movements must be blocked.

Checking the control panel

System ON.

Select from 40 kV to the max. kV value on the kV control console by pressing the ± kV keys.

Check the mAs display by pressing the ± mAs keys.

Function check of the anode starter

Generator ON : On D95, display123

Set release button S27 to preparation: the rotating anode boosts. Display 345

Let go of the release button S27 again: the rotating anode should brake. Display 63

NOTICE

 

The rotational speed must be checked with the tube assembly

 

 

warmed up.

 

 

 

 

 

Key to 7-segment display on board D95:

1=

Initialization

4=

Start-up

2=

Init-brakes

5=

Continued run

3=

Stand-by

6=

Brakes

If the ERROR message flashes, please refer to the information contained in the service software.

LED status display

light up depending on the operating condition

Tube assy. 2 selected

*

*

Tube assy. 1 selected

 

 

Siemens AG

Register 4

RLL5-310.034.04

Page 1 of 4

UROSKOP D3

Medical Engineering

 

Rev. 04 12.98

TD PS 24

 

Checking the fluoroscopic and exposure release circuit

Fluoroscopic circuit

On board D200: SS switch OFF.

On D211: FL switch ON.

On D211: the following LED’s must light up (X2075-23):

- TU1/TU2

TU 1/ TU 2 (acc. to selected workstation/tube assembly)

-

 

 

 

small focus

-

 

 

 

large focus

 

 

 

 

 

 

- GEN-OK

Generator enable state

- DL

Fluoroscopy (FL) "on" (FL request)

- EXT-HW

Radiation enable state/safety circuit

- SWR

Start/inverter

 

 

 

 

 

NOTICE

 

SWR "start/inverter" is not illuminated as long as Kathotest

 

 

 

 

(module J05) is programmed.

 

 

 

 

 

 

 

 

The "radiation indicator" lights up only with radiation ON.

 

 

 

 

 

FL switch OFF

SS switch ON

FL ON

Measure the minimum and maximum kV and mA values and enter them into the generator certificate.

Exposure release circuit

With preparation:

Set the release button S27 to preparation .

On D211, the following LEDs must light up:

- TU1/TU2

TU1/TU2 (acc. to selected workstation/tube assembly)

-

 

 

/

 

 

(depending on focus selection)

- GEN-OK

Generator enable state

- ZB

 

 

 

ZB "ON" (exposure preparation (ZB) request)

With exposure release:

Press the release button S27 all the way down.

On D211, the following LED’s must light up

- VH:

Exposure request to unit

- AR:

Exposure enable state from unit

- EXT-HW:

Radiation enable state/safety circuit

UROSKOP D3

Register 4

RLL5-310.034.04

Page 2 of 4

Siemens AG

 

 

Rev. 04 12.98

TD PS 24

Medical Engineering

NOTICE

 

SWR "start/inverter" is not illuminated as long as Kathotest

 

 

(module J05) is programmed.

 

 

 

 

The "radiation indicator" lights up only with radiation on.

 

 

 

Release radiation release button S27.

Controls and indicator lamps for radiation protection

At the customer’s request, the following controls and indicators can be connected in the power cabinet according to the installation instructions RLL5-310.031.03..., page 4-6.

Door contact for radiation disable function

(e.g. at the door of the examination room)

Check:

When the door is open, radiation release must not be possible.

Indicator lamps for tube assembly selection

Check:

For fluoroscopy or exposure preparation, the respective indicator lamp

 

 

associated with the tube assembly must light up.

Checking the kV/mA values

System OFF.

Connect oscilloscope to test points MP.kVact and MP.mAact on service board D211.

System ON.

SS switch ON.

On the control console, select 77 kV, 32 mAs for

 

and

 

focus respectively.

 

 

Release one exposure in each case and check the kV/mA values of test points MP.kVact and MP.mAact with the oscilloscope.

Enter the measured kV values into the test generator test certificate and compare them with the values recorded at the factory. They must be within the specified tolerances.

Siemens AG

Register 4

RLL5-310.034.04

Page 3 of 4

UROSKOP D3

Medical Engineering

 

Rev. 04 12.98

TD PS 24

 

Checking the mAs values

System OFF.

On PC board D220 (over H1), remove the jumper from the mAs jacks.

Connect the mAs meter to the mAs jacks.

System ON.

On the control console, select , 77 kV, 80 mAs.

Release an exposure and check the mAs value on the mAs meter.

Enter the measured value into the generator test certificate.

Start-up of the automatic exposure control

Check the functions of the IONTOMAT P and PLANI-IONTOMAT P automatic exposure

control systems according to the generator setting instructions RX63-050.032....

NOTICE

NOTICE

The installation procedures described in the start-up instructions may be disregarded since both subassemblies have already been installed at the factory.

Since the UROSKOP has no selection for the film-screen keys H-D-U, identical values must be programmed for H-D-U in the IONTOMAT.

Checking the reproducibility of the automatic exposure control

Attach 2.1 mm Cu to the multileaf collimator.

Set 77 kV, 100%, large focus.

§16 Select Iontomat.

Release 5 exposures consecutively.

Enter the measured dose values into the generator test certificate.

UROSKOP D3

Register 4

RLL5-310.034.04

Page 4 of 4

Siemens AG

 

 

Rev. 04 12.98

TD PS 24

Medical Engineering

Preparation and limit values for tabletop lift

Connect the service PC to connector X10 of board D1.

System (When the emergency stop switch is actuated,”U01” appears in the display).

If necessary, release the emergency stop switch.

Service PC .

Start the service program

(refer to the service software instructions RLL5-310.113.02...).

Main menu

 

 

Info database

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Unit parameters

 

 

 

 

 

 

Load from unit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Unit state

 

 

 

 

Load from file

 

 

 

 

 

 

 

 

Unit errors

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Calibration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Options

 

 

 

 

 

 

 

 

Download

 

 

 

 

 

 

Unit parameters

 

 

 

 

 

 

 

 

Load from unit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Load from file

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Show

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Modify

 

 

 

 

 

 

Tomography

 

 

 

 

 

 

 

 

 

 

Send to unit

 

 

 

 

 

Memoskop 2K/100

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Write to file

 

 

ceiling height

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Output

 

 

 

 

 

 

 

 

 

 

 

 

 

 

For ceiling height enter the actual room height available

Press return.

Press <F4>. Initial questions with Yes.

Siemens AG

Register 4

RLL5-310.034.04

Page 1 of 6

UROSKOP D3

Medical Engineering

 

Rev. 04 12.98

TD PS 24

 

Lifting movement

Rotating axis

Lifting movement

 

 

 

 

Move into the end positions in gradual increments.

 

CAUTION

 

 

 

 

 

 

 

 

 

 

 

 

Tilt the unit into the 0° position.

Successively move the unit into both end positions.

Measure the distance between the floor or top surface of the installation plate and the top surface of the tabletop or rotating axis for both system positions:

 

Nominal distance between the floor or top surface of the installation plate

 

and top surface of the tabletop

 

and rotating axis

 

 

 

 

 

 

 

 

 

Hu

Ho

 

Au

Ao

 

82 cm ± 5 mm

128 cm ± 5 mm

 

62 cm ± 5 mm

108 cm ± 5 mm

 

 

 

 

 

 

for I.I. 40

86 cm ± 5 mm

128 cm ± 5 mm

 

66 cm ± 5 mm

108 cm ± 5 mm

 

 

 

 

 

 

! minimum distance I.I. floor: 4 cm !

NOTICE

 

Ho = 128 cm, possible only for room heights 2.60 m and above.

 

 

 

Read the values displayed on the PC and compare them with the factory test values on the test certificate.

Initial the comparison as confirmation.

Lowest point: 62 cm to center of rotation with tilted table (+88°).

Correction: UROSKOP D adjustment instructions

UROSKOP D3

Register 4

RLL5-310.034.04

Page 2 of 6

Siemens AG

 

 

Rev. 04 12.98

TD PS 24

Medical Engineering

Oblique projection

 

 

 

 

Move into the end positions in gradual increments.

 

CAUTION

 

 

 

 

 

 

 

 

 

 

 

 

Swivel the tube assembly support arm in each case from the 0° position into both end positions, -15° and +15°.

The 15° values must be indicated on the system control console; at the same time, the safety limit switch must not be pressed.

Read the values displayed on the PC under menu item [Unit state], [Current] and compare them with the factory test values in the test certificate.

Initial the comparison as confirmation.

Correction: UROSKOP D adjustment instructions.

Siemens AG

Register 4

RLL5-310.034.04

Page 3 of 6

UROSKOP D3

Medical Engineering

 

Rev. 04 12.98

TD PS 24

 

Siemens UROSKOP D3 Instruction Manual

Tilting movement

Requirement: The installation plate must be level.

 

 

 

 

Move into the end positions in gradual increments.

 

CAUTION

 

 

 

 

 

 

 

 

 

 

 

 

Tilt the unit in each case from both directions into the 0° position.

Check the 0° position in each case with the precision water level after the automatic stop. Nominal: ± 0.3°

Swivel from the 0° position into the +88° end position; the switching cam must not touch the limit switch.

Check the +88° end position with the precision water level.

Nominal: 88° ± 0.5°

Swivel from the +88° end position into the -15° end position; the switching cam must not touch the limit switch.

Check the -15° end position with the precision water level.

Nominal: - 15° ± 0.5°

Read the values displayed on the PC and compare them with the factory test values in the test certificate.

Initial the comparison as confirmation.

Correction: UROSKOP D adjustment instructions.

UROSKOP D3

Register 4

RLL5-310.034.04

Page 4 of 6

Siemens AG

 

 

Rev. 04 12.98

TD PS 24

Medical Engineering

Wall side

 

 

 

Head end

Physician‘s side

Longitudinal movement of the tabletop

 

 

 

 

Move to the end positions in gradual increments.

 

CAUTION

 

 

 

 

 

 

 

 

 

 

 

 

Move the tabletop successively into end positions A and K.

Measure the distance between the left tabletop edge and the left front of the front guide rail for both table positions.

Nominal: La = 573 mm ± 5 mm

Lk = 573 mm ± 5 mm

Read off the values displayed on the PC and compare them with the factory test values of the test certificate.

Initial the comparison as confirmation.

Correction: UROSKOP D adjustment instructions.

Siemens AG

Register 4

RLL5-310.034.04

Page 5 of 6

UROSKOP D3

Medical Engineering

 

Rev. 04 12.98

TD PS 24

 

Loading...
+ 51 hidden pages