Panasonic MPR-721R, MPR-1411R Datasheet

5 (1)
Panasonic MPR-721R, MPR-1411R Datasheet

MPR-1411/R

MPR-721/R

Large Capacity

Laboratory

Refrigerators

Features:

Energy-saving storage solutions for uniformity, reliability and environmentally safe operation

Panasonic refrigeration system designed for precise and stable reagent, media, pharmaceutical and biologics storage

Wide settable temperature range of 2°C to 23°C

Panasonic… the new name for SANYO

www.panasonic.com/biomedical

MPR-1411/R

MPR-721/R

MPR Series: Large Capacity Laboratory Refrigerators

Stable and reliable refrigerated environment for exacting laboratory requirements. Ideal temperature environment for clinical research, pharmaceutical and industrial use. Adjustable shelves and wide range setpoint for varying laboratory applications including chromatography and pharmaceutical storage.

Temperature Stability

Panasonic’s temperature control system, with microprocessor reliably maintains cabinet temperature at the set level and is unaffected by outside ambient temperature. Forced air circulation ensures that the cabinet temperature returns to the set point quickly after door openings and remains uniform throughout the cabinet.

Wide Settable Temperature Range

With a temperature range of 2°C to 23°C, Panasonic’s MPR Series refrigerators are ideally suited for

many tests that require a stable, cool temperature or general cold storage.

Efficient and Powerful

Refrigeration System

For frequent door openings, Panasonic’s MPR Series refrigerators are equipped with powerful, hermetically sealed compressors.These purpose-built compressors ensure superior pulldown characteristics and precise temperature control.

MPR-1411

MPR-721R

Left: MPR-1411 with 8 adjustable wire shelves and effective capacity of 48.2 cu. ft. Right: MPR-721R with 5 pull out drawers and effective capacity of 23.7 cu. ft. Hybrid Designs of both shelves and drawers are available

Laboratory Refrigerators

 

Volume

Exterior Dimensions

Door

Shelves

Drawers

Voltage, Power

Model Number

(cu.ft.)

(w x f-b x h)

(glass)

(adjustable wire)

(solid roll-out)

Connection

 

 

 

 

 

 

 

MPR-721-PA

24.2

30.3" x 35.7" x 77.0"

single, swinging

4

115V

NEMA 5-15

 

 

 

 

 

 

 

 

 

 

 

 

 

MPR-721R-PA

23.7

30.3" x 35.7" x 77.0"

single, swinging

5

115V

NEMA 5-15

 

 

 

 

 

 

 

 

 

 

 

 

 

MPR-1411-PA

48.2

56.7" x 35.7" x 76.8"

double, swinging

8

115V

NEMA 5-15

 

 

 

 

 

 

 

 

 

 

 

 

 

MPR-1411R-PA

48.0

56.7" x 35.7" x 76.8"

double, swinging

10

115V

NEMA 5-15

 

 

 

 

 

 

 

 

 

 

 

 

 

2

Large Capacity Laboratory Refrigerators

www.panasonic.com/biomedical

 

 

 

MPR Series Features and Benefits

Alarm & Safety Features

Feature

Trigger

Type of Alarm/Response

 

 

 

High Temperature Alarm

2°C to 14°C (selectable) above setpoint

Flashing LED & audible alarm after 15 minutes

 

 

 

Low Temperature Alarm

2°C to 14°C (selectable) below setpoint & 0°C or lower

Flashing LED & audible alarm

 

 

 

Temperature Lock

Lock ON

Set temperature and alarm points

cannot be altered

 

 

 

 

 

Memory Backup

Power failure

Memory of settings

 

 

 

Door Ajar

Door not closed

Door ajar lamp ON, adjustable

audible alarm from 1 to 15 minutes

 

 

 

 

 

Self Diagnostics

Sensor failure and fan or compressor abnormality

Error code shown

 

 

 

Remote Alarm Contacts

N.O. and N.C. contact available for hook up

to an auxiliary alarm system

 

 

 

 

 

Temperature Recorder

Optional, Model #MTR-C954, 7-day 6” circular chart

or touch panel digital temperature recorder

 

 

 

 

 

Cycle Defrost Function

In the 4°C range, frost buildup on the evaporator is inevitable.This can affect the performance of the refrigeration system. Panasonic has solved this problem with a cycle defrost and evaporator temperature sensor system. The system design ensures that defrost occurs only when necessary and not based on time like other competitive units.This system runs automatically so there is no need to turn off power for defrosting.Temperature rise during defrost is minimal with no temperature spikes.The evaporation heater also doubles as protection against drops in cabinet temperature caused by a low ambient temperature.

Cycle Defrost Function Data

 

 

chamber air

 

 

 

4°C

 

 

 

 

 

0°C

 

 

 

 

 

 

surface of

 

 

 

 

 

evaporator

 

 

 

 

Comp-ON

Heater-ON

Comp-ON

Heater-ON

Comp-ON

(Time)

Panasonic Cycle Defrost

40

 

 

 

 

35

AmbientTemp.

 

30

 

 

 

 

 

20

 

 

 

 

10

InsideTemp.

 

 

 

 

 

5

 

 

 

 

0

8

16

24

(Time)

Others-Auto Defrost

Refrigerator

40

 

 

 

 

35

AmbientTemp.

 

30

 

 

 

 

 

20

Temp. Rise During

 

Defrost Cycle

 

 

 

10

 

 

 

 

5

 

 

 

 

0

8

16

24

(Time)

3

Loading...