HP 346a schematic

4.5 (4)

Operating and Service Manual

Agilent 346A/B/C Noise Source

(Including Options

001, 002, and 004)

Serial Numbers

This manual applies directly to instruments with serial numbers prefixed 4124A and below.

Instruments with serial prefixes 4124A and above are supplied with the noise source ENR data preloaded on diskette -

see Chapters 3 & 4.

With changes described in Chapter 2 Manual Changes, this manual also applies to the following prefixes.

Agilent 346B - 2015A and 1935A

Manufacturing Part Number: 00346-90139

July 2001 Supersedes: June 2001

© Copyright 2001 Agilent Technologies

Notice

Information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishings, performance, or use of this material. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of Agilent Technologies.

Certification

Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology and/or The United Kingdom National Physics Laboratories, to the extent allowed by the Institution’s calibration facility, and to the calibration facilities of other International Standards Organization members.

Warranty

This Agilent Technologies instrument product is warranted against defects in material and workmanship for a period of one year from date of shipment. During the warranty period, Agilent Technologies will at its option, either repair or replace products which prove to be defective. For warranty service or repair, this product must be returned to a service facility designated by Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges, duties, and taxes for products returned to Aglent

ii

Technologies from another country. Agilent Technologies warrants that its software and firmware designated by Agilent Technologies for use with an instrument will execute its programming instructions when properly installed on that instrument. Agilent Technologies does not warrant that the operation of the instrument, or firmware will be uninterrupted or error free.

Limitation of Warranty

The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance. NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

Exclusive Remedies

THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.

Safety Notices

This guide uses warnings and cautions to denote hazards.

iii

WARNING

CAUTION

A warning calls attention to a procedure, practice or the like, which, if not correctly performed or adhered to, could result in injury or the loss of life. Do not proceed beyond a warning until the indicated conditions are fully understood and met.

A caution calls attention to a procedure, practice or the like, which, if not correctly performed or adhered to, could result in damage or the destruction of part or all of the equipment. Do not proceed beyond a caution until the indicated conditions are fully understood and met.

iv

HP 346a schematic

v

Table of Contents

 

 

Page

1. General Information

 

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . 2

Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . 2

Instruments Covered in This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . 9

Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 10

Model Characteristics of Individual Noise Sources . . . . . . . . . . . . . . . . . . . . . . . . .

11

Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

12

Equipment Available But Not Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

12

Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

12

Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

13

Handling Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

13

Initial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

13

Original Packaging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

14

Mating Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

14

Storage and Shipping Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

14

Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

15

Operating Environment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

16

Recommended Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

16

Operator’s Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

17

Operator’s Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

22

Performance Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

24

Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

25

Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

26

Replaceable Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

27

Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

29

Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

29

Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

29

Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

30

Returning a Noise Source for Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

30

2. Manual Changes

Manual Changes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

Change A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

3. Using the ENR Data Diskette

vii

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Loading the ENR Data from Diskette. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

Saving the ENR Data to the Noise Figure Analyzer’s Internal Memory . . . . . . . . . . 40 Copying the ENR Data to the Noise Figure Analyzer’s Internal Memory . . . . . . . . . 41

4. ENR File Format

 

Format Details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

44

Comment Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

44

Header Fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

44

ENR Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

46

Examples. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

49

A. Caring for Connectors

 

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

52

Connector Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

52

Handling and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

52

Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

53

Obvious Defects and Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

53

Mating Plane Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

53

Precision 7 mm Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

54

Sexed Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

55

Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

57

Compressed Air . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

57

Cleaning Alcohol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

57

Precision 7 mm Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

58

Cleaning Interior Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

59

Drying Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

60

Mechanical Inspection: Connector Gages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

61

Mechanical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

62

Precision 7mm Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

62

Sexed Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

63

50 Ohm Type-N Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

63

75 Ohm Type-N Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

64

Using Connector Gages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

65

Inspecting and Cleaning the Gage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

65

Zeroing the Gage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

65

Making Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

68

Align Connectors Carefully. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

68

To Make a Preliminary Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

69

Final Connection Using a Torque Wrench . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

70

Disconnection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

71

Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

73

Principles of Microwave Connector Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

75

viii

1

General Information

1

NOTE

General Information

Introduction

Introduction

This manual contains operating and service information for the Agilent Technologies model 346A, 346B, and 346C Noise Sources.

Included in the manual is information necessary to operate the noise sources.

Specifications

The “Specifications - Table 1-1” on page 3, page 4, and page 5 are performance standards or limits against which the noise source may be tested. These specifications for the noise source when used with a Noise Figure Analyzer are ONLY valid if the analyzer has been allowed to meet its specified warm up time of 60 minutes.

ENR expanded uncertainty analysis and supplemental characteristics are not specifications but are typical characteristics included as additional information for the user.

Specifications are valid at ambient temperature 23 ±1 oCelsius only.

2

Chapter 1

 

 

 

General Information

 

 

 

Introduction

 

Specifications - Table 1-1

 

 

 

Frequency Range

 

 

 

 

 

 

 

 

 

346A/B

 

10 MHz — 18 GHz

 

 

 

 

 

 

 

346C

 

10 MHz — 26.5 GHz

 

 

 

 

 

 

Operating

 

 

 

Temperature

 

 

 

 

0 oC to 55 oC

 

 

 

Excess Noise

 

 

 

Ratio (ENR)

 

 

 

Range1

 

 

 

 

 

 

 

346A

4.5 - 6.5 dB

 

 

 

 

346B

1416 dB

 

 

 

 

346C

12 - 17 dB

 

 

 

 

 

1.ENR values are given at cardinal frequency points over the frequency range of each noise source. Calibrated values at cardinal frequencies printed on each noise source label and on a separate report supplied with each Agilent 346 Noise Source. Specifications are valid at calibration temperature only.

Chapter 1

3

General Information

Introduction

Specifications - Table 1-1 Continued

Maximum

Standing Wave

Ratio (SWR) and

Reflection

Coefficient (r) for

Source ON/OFF

States

Instrument model

Frequency Range

Maximum

Reflection

 

(GHz)

Standing Wave

Coefficient

 

 

 

Ratio (SWR)

(Rho) (r)

 

 

 

 

 

346A/B1

0.01

- 0.03

<1.3:1

0.13

 

0.03

- 5.0

<1.15:1

0.07

 

 

 

 

 

5.0 - 18.0

<1.25:1

0.11

 

 

 

 

346C

0.01-18.0

<1.25:1

0.11

 

 

 

 

 

 

18.0

- 26.5

<1.35:1

0.15

 

 

 

 

 

1.Maximum change in complex reflection coefficient between source ON and source OFF at all frequencies for 346A only: 0.01.

4

Chapter 1

General Information

Introduction

Specifications - Table 1-1 Continued

Impedance

50 ohm nominal

Maximum reverse power

1 Watt

Power required

28 ±1 V

346A/B

60 mA peak, 30 mA average for source ON

 

 

346C

45 mA

 

 

Connectors1

346

Output

APC-3.5 (male) standard1

 

 

 

346

Input

Bias: BNC (f)

 

 

 

1.Also mates with female SMA connectors. See option information for other connector styles.

Dimensions

140 x 21 x 31mm (5.5 x 0.8 x 1.2 in)

Net Weight

0.1kg (3.5 oz.)

1.For correct connector usage, refer Table A-2 for the torque settings.

Chapter 1

5

General Information

Introduction

Excess Noise Ratio Expanded Uncertainty (U(Y)) - Table 1-2

ENR values are given at cardinal frequency points over the frequency range of each noise source. These ENR values are printed on the noise source label and on a separate printed report supplied with the 346 Noise Source. Included in the printed report is the measurement uncertainty for each cardinal frequency point.

A significant proportion of the expanded uncertainty (U(Y)) is based on the uncertainties provided by the United States National Institute of Standards and Technology (NIST) and/or The United Kingdom National Physics Laboratories (NPL). Agilent therefore reserve the right to change the overall expanded uncertainties based on changes in uncertainty values within the National Standards Institutes.

Uncertainties are valid at ambient temperature 23°C ±1°C (296K) only.

A typical characteristic plot of ENR (U(Y)) versus each cardinal frequency point is shown in Figure 1-1. the uncertainties given are typical - refer to the printed calibration report for the actual uncertainties for your noise source.

Figure 1-1

Characteristic1ENR Plot versus Cardinal Frequency Points

1. Characteristic values are met or bettered by 90% of instruments with 90% confidence.

6

Chapter 1

General Information

Introduction

Excess Noise Ratio Expanded Uncertainty (U(Y)) - Table 1-2

Continued

Characteristic

ENR (U(Y))

Specification

Instrument model

Frequency

ENR Uncertainty (±dB)1

 

(GHz)

 

 

 

 

346A

0.01- 1.5

0.21

 

 

 

 

 

1.5

- 3.0

0.20

 

 

 

 

 

3.0

- 7.0

0.20

 

 

 

 

 

7.0

- 18.0

0.23

 

 

 

346B

0.01- 1.5

0.20

 

 

 

 

 

1.5

- 3.0

0.19

 

 

 

 

 

3.0

- 7.0

0.20

 

 

 

 

 

7.0

- 18.0

0.23

 

 

 

346C

0.01- 1.5

0.22

 

 

 

 

 

1.5

- 3.0

0.19

 

 

 

 

 

3.0

- 7.0

0.20

 

 

 

 

 

7.0

- 18.0

0.28

 

 

 

 

18.0 - 26.5

0.34

 

 

 

 

1.Characteristic values are met or bettered by 90% of instruments with 90% confidence.

Chapter 1

7

General Information

Introduction

Supplemental Characteristics - Table 1-3

Supplemental

Characteristics

ENR variation with

<0.01 dB/°C for 30 MHz to 26.5 GHz

temperature:

 

 

 

ENR variation with

Internal current regulator for <0.02 dB variation

voltage:

for 28 ±1 V

 

 

Switching speed:

 

For repetitive operation (in previous state for less than 5 seconds):

Turn-on: < 20 μs

Turn-off: <80 μs

For single-shot operation (in previous state more than 5 seconds):

Turn-on: < 3 ms

Turn-off: <80 μs

8

Chapter 1

NOTE

General Information

Instruments Covered in This Manual

Instruments Covered in This Manual

The noise sources covered by this manual have a two-part serial number. The first four digits and letter constitute the serial number prefix. The last five digits form a sequential suffix that is unique to each noise source. The prefix is the same for all noise sources of a particular configuration. It will change when a design modification occurs. The contents of this manual apply directly to those instruments having the same serial number prefixes listed under SERIAL NUMBERS on the title page.

A noise source manufactured after the printing of this manual may have a serial number prefix which is not listed on the title page. This unlisted serial number prefix indicates the noise source is different than those documented in this manual. If manual changes are needed, the manual for this newer noise source is accompanied by a Manual Changes supplement. The supplement contains "change information" that explains how to adapt this manual to the newer noise source.

In addition to change information, the supplement may contain information for correcting errors in the manual. The supplement is identified with the manual print date and part number, both of which appear on the manual title page.

For information concerning a serial number prefix that is not listed on the title page in the Manual Change supplement, contact your nearest Agilent Technologies office.

Agilent/Hewlett-Packard 346 Noise Sources which were manufactured before the printing of this manual will have a serial number prefix lower than the one listed on the title page. Manual changes for these older models are found in Chapter 2 “Manual Changes”.

Agilent Technologies are migrating their existing product serial numbers to a new format. All new products use this new format. The format is the first two letters signify the country of manufacture, for example, US representing the USA, the next four numbers,the serial number prefix, and the last five numbers form a sequential suffix that is unique to each product. For example, US123456789, where 1234 is the prefix and 56789 is the suffix. The two letters MY represent Malaysia.

Chapter 1

9

General Information

Description

Description

The noise source produces noise output (power-on) when +28 V is applied. When it is off, there is residual noise due to thermal agitation in the noise source (power-off). These two noise levels are used to measure the gain and added noise of the device under test, and consequently, its noise figure.

The Excess Noise Ratio (ENR) for each noise source has been measured at major frequencies and recorded on a label attached to the noise source (see Figure 1-2). ENR references power-on to the noise power that exists at 290 Kelvin (17 °C ). In addition, a separate calibration sheet showing the complex reflection coefficient in both the on and off states is included with each instrument.

All three models are provided with a BNC female connector for power input. The output connector is a male APC-3.5 on the standard noise sources. Type-N and APC-7 connectors are available as options for the Agilent 346A and 346B Noise Sources. In addition to these general characteristics, certain characteristics apply to the specific models. These characteristics are listed in Table 1-4.

Figure 1-2

A Partial View of Typical Calibration Label

10

Chapter 1

 

 

 

General Information

 

 

 

Description

Table 1-4

Model Characteristics of Individual Noise Sources

Agilent 346A

 

Agilent 346B

Agilent 346C

 

 

 

Reflection coefficient differential

Reflection coefficient

Reflection coefficient

between on and off states is specified

differential between on and off

differential between on and off

to be no greater than 0.01.

states is not specified, but is

states is not specified, but is

 

 

less than 0.1 typically.

less than 0.1 typically.

 

 

 

Specified output noise spectrum

Specified output noise

Specified output noise

from 10 MHz to 18 GHz.

spectrum from 10 MHz to

spectrum from 10 MHz to

 

 

18 GHz.

26.5 GHz.

 

 

 

Nominal ENR is 6 dB over the

Nominal ENR is 15 dB over the

Nominal ENR is 15 dB over the

specified frequency range.

specified frequency range.

specified frequency range.

 

 

 

Not directly usable for

Suitable for Agilent/HP 8970A

Suitable for Agilent/HP 8970A

Agilent/HP 8970A IF attenuator

IF attenuator calibration

IF attenuator calibration

calibration (Special Function

(Special Function 33).

(Special Function 33).

33). 10 dB of gain is required.

 

 

 

 

 

Not directly usable for

Suitable for Agilent/HP 8970B

Suitable for Agilent/HP 8970B

Agilent/HP 8970B IF attenuator

IF attenuator calibration.

IF attenuator calibration.

calibration. Refer to

 

Refer to "Calibration, IF

Refer to "Calibration, IF

"Calibration, IF Attenuators" in

Attenuators" in the Agilent/HP

Attenuators" in the Agilent/HP

the Agilent/HP 8970B Operating

8970B Operating Manual.

8970B Operating Manual.

Manual.

 

 

 

 

 

 

 

Chapter 1

11

General Information

Description

Warranty

The noise sources are warranted and certified as indicated in this manual. Connector damage resulting from improper use is not covered under warranty.

 

Equipment Available But Not Supplied

 

The following equipment is available from Agilent Technologies for use

 

with the noise sources:

 

• 8710-1766: 3/4" Torque Wrench (APC-7)

 

• 5060-0344: 9/16" Torque Wrench (APC-3.5)

 

 

NOTE

Agilent/HP Model 346B Noise Sources with serial prefixes of 2037A and

 

below, have an APC-3.5 connector of a different configuration (See Figure

 

1-3). The 9/16" torque wrench will not fit these older models.

 

 

Figure 1-3

Agilent/HP Noise source with Older APC-3.5 Connector

Options

The Agilent 346A and 346B Noise Sources are available with the following output connector options:

Option 001, Type-N (male)

Option 002, APC-7

Option 004, Type-N (female)

Since the APC-7 and Type-N connectors do not operate up to 26.5 GHz, no output connector options are available for the Agilent 346C.

12

Chapter 1

CAUTION

CAUTION

General Information

Installation

Installation

Handling Precautions

Do not disassemble the noise source. The diode module is static sensitive and can be damaged or the calibration can be altered.

Do not drop the noise source. Dropping can damage the unit or alter the calibration.

Proper connector care is essential. See Operator’s Maintenance in the Operation section of this manual for more information.

Initial Inspection

Inspect the shipping container for damage. Inspect the noise source for mechanical damage incurred in transit. If the shipping container or cushioning material is damaged, it should be kept until the contents of the shipment have been checked for completeness and the noise source has been mechanically and electrically checked. If the contents are incomplete, if there is mechanical damage or a defect, or if the noise source does not work electrically, notify the nearest Agilent Technologies office. If the shipping container is damaged, or the cushioning material shows signs of unusual stress, notify the carrier as well as the Agilent Technologies office. Keep the shipping materials for the carrier’s inspection.

Chapter 1

13

General Information

Installation

Original Packaging

Container and materials identical to those used in factory packaging are available through Agilent Technologies offices. If the noise source is being returned to Agilent Technologies for servicing, attach a tag indicating the name and address of the company, the technical contact person, phone number and extension, the model number, serial number, type of service being requested, and failure symptoms if applicable. Mark the shipping container FRAGILE. In any correspondence, refer to the noise sources by model number and serial number.

Mating Connectors

The noise sources can be mated with other instrumentation having the connectors listed in Table 1-5.

Table 1-5

Connectors That Can Be Mated With the Noise Sources

 

Configuration

Mating Connector

 

 

 

 

 

Input: all units

BNC male1

 

 

Output: standard

APC-3.5 female

 

 

SMA female

 

 

 

 

 

 

Opt.001

Type-N female

1

 

 

 

 

 

 

 

 

Opt.002

APC-7

 

 

 

 

 

 

Opt.004

Type-N male1

 

 

 

 

 

1Must comply with U.S. Military Standard MII,C-39012

Storage and Shipping Environment

The noise sources should be stored in a clean, dry environment. The following environmental limitations apply to both storage and shipment:

Temperature: -55 °C to +75 °C

Humidity: <95% relative

Altitude: <15300 meters (50000 feet)

14

Chapter 1

CAUTION

NOTE

Figure 1-4

General Information

Operation

Operation

This section refers to operation with noise figure meters. For more detailed operating instructions, refer to the operating manual for the noise figure meter used.

Use a dc blocking capacitor to protect the noise source from damage when connected to any system where a dc voltage is present on the output center conductor.

Noise figure measurements of devices (such as amplifiers, mixers, transistors, and receivers) can be made using the noise source with a noise figure meter. Figure 1-4 depicts a simple test setup for a noise figure measurement.

The noise figure meter must have a +28 ±1 V switched supply.

Typical Noise Figure Measurement Test Setup

Chapter 1

15

General Information

Operation

Operating Environment

The operating environment of the noise sources should be within the following limitations:

Temperature: 0 °C to +55 °C

Humidity: <95% relative

Altitude: <4600 metres (15 000 feet)

Recommended Test Equipment

Table 1-6 is a list of equipment that can be used to perform an operational verification check.

Table 1-6

Recommended Test Equipment For Operator’s Checks

Check

 

Instrument

Critical Specification

Recommended Model

No.

 

 

 

 

 

 

 

 

 

1

 

Power Meter and Power

Minimum Sensitivity: 0.1 nW

Agilent/HP E4418A

 

 

Sensor

Frequency Range: 10 MHz to

with

 

 

 

 

 

 

18 GHz

Agilent/HP 8481D, or

 

 

 

 

 

 

 

 

E4412A, or E9300A

 

 

 

 

 

 

 

Power Supply

Voltage: 28.0 ±1.0 V

Agilent/HP 6028A

 

 

 

Current: 100 mA

 

 

 

 

 

 

2

 

Noise Figure Meter

Voltage Output: 28.0 ±1.0 V

Agilent/HP 8970B

 

 

 

Input Noise Figure:<7.4 FdB

or

 

 

 

at 100 MHz

Agilent N8972/3/4/5A

 

 

 

 

 

 

 

 

 

16

Chapter 1

General Information

Operation

Operator’s Check

The operator’s checks in this section should be performed if failure of the noise source is suspected. The checks can be used only to verify that the noise sources are producing a broadband noise spectrum. They cannot be used to check the units against specifications. Only one of the checks is necessary to verify operation. Table 1-6 shows the recommended test equipment used for each check.

Operator’s Check with Power Meter (Check 1)

Step 1. Connect the equipment as shown in Figure 1-5.

Figure 1-5

Operator’s Check Test Setup 1

Step 2. Turn the power supply OFF. Zero the power meter.

Step 3. Turn the power supply ON (+28 V). Measure the power output with the noise source on.

Step 4. Verify that the result of the measurement is within the following limits and hence the noise source is operating correctly:

Agilent 346A Power output = -66 ±4 dBm.

Agilent 346B Power output = -56 ±4 dBm.

Agilent 346C Power output = -56 ±4 dBm.

Chapter 1

17

General Information

Operation

Operator’s Check with Agilent/HP 8970B Noise Figure Meter (Check 2a)

Step 1. Remove any cables from the noise figure meter input. Press PRESET. After 5 seconds, verify the noise figure display shows --FdB and the left display shows 30 MHz.

Step 2. Connect the equipment as shown in Figure 1-6.

Figure 1-6

Operator’s Check Test Setup 2a

Step 3. Enter a tuned frequency of 100 MHz. (See "Fixed Frequency Tuning" in the Agilent/HP 8970B Operating Manual).

Step 4. Enter special function 5.3 to enable the instrument for spot ENR entry. (See "Special Functions" in the Agilent/HP 8970B Operating Manual).

Step 5. Enter the ENR at 100 MHz from the noise source calibration label. (See "Spot ENR, Thot, and TCold, in the Agilent/HP 8970B Operating Manual).

Step 6. Confirm that the noise source is operating correctly.

The noise figure measurement of the noise figure meter will appear in the noise figure display. If the result of the measurement is less than 7.4 FdB, the noise source is operating.

Step 7. Press PRESET to return the instrument to preset conditions.

18

Chapter 1

Loading...
+ 56 hidden pages