HEIDENHAIN ND 281 B User Manual v.2

User’s Manual
ND 281
Measure d V alue Displays
English (en) 10/2002
B
Display of actual value and input
(9 decades with algebraic sign)
REF SET START PRINT inch
=
<>
21 MIN ACTL MAX DIFF
HEIDENHAIN
Status display with indicators
Numeric keypad with decimal point
789
456
MOD
123
.
0
CL
2
Key Function
• Set datum
• Transfer input value
• Set display to value from P79 (P80!)
• Leave parameter list
• Select datum
• Page backwards in parameter list
MOD
• Select parameter after switch-on
• Page forward in parameter list
• Start series of measurements
• Switch display for measurement series
• Start measured value output “PRINT”
• Delete entry
• Set display to zero (P80!)
• CL plus MOD: select parameter list
• CL plus number: select parameter
• Delete parameter input and show parameter number
• Algebraic sign
• Reduce parameter value
• Decimal point
• Increase parameter value
1)
Only in linear measurement mode.
Indicator Meaning
REF If the decimal point is also blinking:
Display is waiting for reference mark traversing. If decimal point is not blinking: Reference mark has been traversed—display stores datum points in nonvolatile memory Blinking: display is waiting for ENT or CL to be depressed
inch Position values in inches
1)
1 / 2 Selected datum point
1)
PRINT "Linear measurement”
Blinking: Display is waiting for ENT to
be pressed for data output
“Angular measurement”
Measured value output with MOD key
SET Blinking: Display is waiting for input values < / = / > Sorting and tolerance checking:
measured value smaller than lower limit / within the limits / greater than upper limit
/
MAX / Series of measurements: Minimum /
MIN DIFF / ACTL
1)
maximum / greatest difference (MAX–MIN) / current measured value Blinking: Confirm selection or deselect function
START
1)
Series of measurements is running Blinking: Display is waiting for signal to start series of measurements
3
Items supplied with ND 281 B
ND 281 B Measured value display unit,
benchtop model Encoder input 11 µAPP/1 V
Power cord 3 m (9.9 ft) User's Manual ND 281B
Items Supplied
Adhesive plug-in feet For stacking ND 281B units
PP
This manual is for the ND 281 B measured value display with the following software number or higher:
349 797-04
Id. Nr. 344 996-xx
The software number is indicated on a label onthe
4
rear panel.
Contents Working with the ND Display Units
Encoders 6 Reference Mar ks 7 Switch-On, Tr aversing the Reference Marks 8 Datum Setting 9 Finding Minimum and Maximum Values Sorting and Tolerance Checking 13 Measured Value Output 14 Display Freeze 15 Error Messages 16
1)
Only in linear measurement mode
1)
10
Installation and Specifications
Rear Panel, Accessories 17 Mounting 19 Pow er Connection 20 Linear/Angular Measuring Modes 21 Operating Par ameters 22
List of Operating Parameters 24
Linear Encoders 28 Angle Encoders 33 Multipoint Axis Error Compensation 34 Switching Inputs and Outputs EXT (X41) 38 Locking the Keypad 43 Displaying the Software Version 44 Distance-to-Go Mode 45 RS-232-C/V.24 Data Interface (X31) 46 Input/Output of Parameter and 51
Compensation-V alue Lists Output Format of the Parameter List 53 Output Format of the Compensation-Value Table 57 Remote Operation over the RS-232-C/V.24 60
Data Interface Specifications 63
Dimensions 64
Contents
5
Position Encoders
The ND 281 B display unit is designed for use with photo­electrical linear or angular encoders with sinusoidal signals: primarily for HEIDENHAIN MT length gauges.
When shipped by HEIDENHAIN, the display units are set to the linear measurement mode.
You can switch between the linear and angular modes by
Position Encoders
entering the code number 41 52 63 (see “Linear/Angular Measurement Modes”).
On the back of the display you will find two flange sockets for connecting the encoder: X1 for encoders with 11 µA dal current signals and X2 for 1 V
Before shipping, HEIDENHAIN activates the encoder connec­tion X1 for 11 µA P02 you can activate the encoder input that matches your encoder (see “Operating Parameters”).
sinusoidal current signals. With parameter
PP
sinusoidal voltage signals.
PP
sinusoi-
PP
6
Reference Marks
The MT length gauges have one reference mark. The scales of other photoelectric linear or angular encoders can contain one reference mark or many
distance-coded
reference marks.
If there is an interruption of power, the relationship between the position of the length gauge and the displayed position value is lost. The reference marks on the position encoders and the REF reference mark evaluation feature enable the display unit to quickly reestablish this relationship again when the power is restored.
When a reference mark is crossed over, a signal is generated which identifies that position as a reference point. At the same time, the display unit restores the relationship between length gauge position and display values which you last defined by setting the datum.
To restore the datum on scales with distance-coded reference marks, you only need to traverse a maximum of 20 mm for linear encoders, and 10° or 20° for angle encoders, depending on the model.
Scale in Distance-coded linear encoder reference marks
Reference Marks
Reference mark
Reference marks on linear encoders
7
Switch-On, Traversing the Reference Marks
REF mode
ä
0
ENT...CL
5 , 6 9 7
Switch-On, Traversing the Reference Marks
For automation purposes, crossing over the reference marks and the display ENT ... CL can be disabled with parameter P82.
Turn on the power.
1
(Switch located on rear panel.)
ND 281 B appears for two seconds.
ENT ... CL
REF indicator starts blinking.
Switch-on the reference mark evaluation function.
The position value that was last assigned to the reference mark position is displayed.
REF indicator lights.
Decimal point starts blinking.
Cross over the reference mark.
Move the plunger until the display starts counting and the decimal point stops blink-
Û
ing. The display is now ready for operation.
1)
appears.
Crossing over the reference marks automatically switches the display to REF mode: The last assignment of display values to length gauge positions is stored in nonvolatile memory.
1)
Press the CL key if you choose not to traverse the reference
marks. Note that, in this case, the relationship between length gauge position and display value will be lost if the power is interrupted or if the unit is switched off.
8
Datum Setting
The datum setting procedure assigns a display value to a known position. With the ND 200 series, you can set two separate datum points. There are several ways to set the datum:
Enter a numerical value, or
Transfer a value from an operating parameter
(see P79, P80), or
By external signal
Select datum 1 or 2.
Enter a numerical value (here, 5).
5
You can switch between datums 1 and 2 as desired. Datum 2 can be used, for example, for working with incremental dimen­sions.
When you switch back to datum 1, the display unit resumes display of the encoder's actual position.
SET starts blinking.
Confirm the entered numerical value.
Z
?
? ? ?
?
Without datum setting: unknown assignment of measured values to positions
Datum Setting
After datum setting: Assignment of measured values to positions
9
Finding Minimum and Maximum Values from a Series of Measurements
After a series of measurements has been started, the display transfers the first measured value to the memory for minimum and maximum values. Every 0.55 ms, the display compares the current measured value with the memory contents: A new value is stored if it is greater than the stored maximum value or smaller than the stored minimum value. At the same time, the display calculates and stores the difference DIFF between the current MIN and MAX values.
Display Meaning MIN Minimum value from the series of measurements
MAX Maximum value from the series of measurements DIFF Difference MAX – MIN ACTL Current measured value
Starting the measurement series and selecting the display
Finding Minimum and Maximum Values
You can start the series of measurements either by pressing MOD and selecting the desired displayas described on the following pagesor by external signal over the switching inputs at the D-sub connection EXT (X41, see “Switching Inputs and Outputs).
1)
MAX
ACTL
Series of measurements: The MIN, MAX and DIFF values of an uneven surface
DIFF
MIN
10
When a series of measurements is started, the internal MIN/ MAX/DIFF memory is reset.
1)
Only in the linear measurement mode.
Example: Measurement series for determining eccentricity
e
Starting a series of measurements
Switching between MIN, MAX, DIFF and ACTL displays
Repeatedly
MOD
MAX
Repeatedly
Select the display for a series of meas­urements.
The selected indicator starts blinking (here, MAX).
Confirm selection.
... until the START indicator starts blinking.
MOD
START
Start the series of measurements.
Indicator preselection
Press MOD to start the series of measurements and select the display with the indicators.
Operating parameter P86 allows you to define which indicator is displayed first when MOD is pressed.
It is not possible to switch between the displays as described below if the switching input for external control of the series of measurements (pin 6 on D-sub connection EXT) is active.
As an alternative, you can select the display with operating parameter P21 (see "Operating Parameters").
Repeatedly
MOD
MIN
The display now shows the smallest value measured during the current series of measurements.
Select the new display of a series of measurements.
The selected indicator starts blinking (here, MIN).
Confirm the change.
Finding Minimum and Maximum Values
11
Starting a new series of measurements
Repeatedly
MOD
START
Ending a series of measurements
Repeatedly
MOD
Select the indicator START.
The START indicator starts blinking.
Start a new series of measurements.
Select the active indicator (MIN, ACTL, MAX, DIFF).
The indicator that lit up last starts blinking.
End the series of measurements.
Finding Minimum and Maximum Values
or
Repeatedly
MOD
START
Select the indicator START.
The START indicator starts blinking.
12
End the series of measurements.
Sorting and Tolerance Checking
In the sorting and tolerance checking mode, the display unit compares the displayed value with the programmed upper and lower sorting limits. The sorting and tolerance checking mode is enabled and disabled with operating parameter P17.
Entering sorting limits
Sorting limits are entered in operating parameters P18 and P19 (see operating parameters).
Sorting signals
The indicators and switching outputs at D-sub connection EXT (see section on X41) sort the display value into one of three classes.
Display Meaning
= Measured value is within sorting limits < Measured value is smaller than lower sorting limit > Measured value is greater than upper sorting limit
Operating parameters for sorting and tolerance checking
P17 CLASS Sorting ON/OFF P18 L.CLASS. Lower sorting limit P19 U.CLASS. Upper sorting limit
Part 1
Part 2
Part 3
Example: Upper sorting limit = 26.02 mm
Lower sorting limit = 26.00 mm
Accept
Sorting and Tolerance Checking
Rework
Reject
13
Measured Value Output
For technical information on the RS-232-C/V.24 data interface (X31), information on the data format, etc., see the chapter RS-232-C/V.24 Data Interface (X31).
Measured values can be transmitted over the RS-232-C/V.24 interface (X31), for example to a printer or PC.
There are several ways to start measured value output:
ä
In the linear measurement mode:
Measured Value Output
Press MOD repeatedly until the PRINT indicator starts blinking, then start measured value output with ENT. In the angular measurement mode: Press the MOD key (this feature can be disabled with operating parameter 86).
or
ä
Input the command STX (Ctrl B) over the RXD input of the RS-232-C/V.24 interface (X31);
or
ä
Input a signal for measured data output (Pulse or Contact) at the D-sub connection EXT (X41).
PC
START PRINT
MOD
in.
123
0.–CL
REF21SET
=
<>
HEIDENHAIN
MIN ACTL MAX DIFF
789 456
The RS-232-C/V.24 interface (X31) enables you to connect a printer or a PC to your display unit
14
Display Freeze
With the latch command, the display can be stopped for any period of time. The internal counter remains active. Parameter P23 selects the display freeze mode and offers three settings:
Concurrent display, no display freezethe display value is the current measured value.
Frozen displaydisplay value is frozen and is updated with each signal for measured value output.
Frozen/concurrent displaydisplay remains frozen as long as the latch signal is present; after the signal, the display resumes continuous display of the current meas­ured values.
Position
Latch signal
Frozen display
Display Freeze
Frozen/ concurrent display
15
Error Messages
Display Effect/Cause
RS232 FAST Command for measured value
output followed too quickly by another.
1)
SIGNAL Encoder signal is too weak.
Error Messages
The scale may be contaminated.
DSR.MISSING The connected device has not sent
a DSR signal.
1)
REF. ERR. The spacing of the reference marks
as defined in P43 is not the actual
spacing.
1)
FORMAT ERR. Data format, baud rate, etc. do not
1)
match.
FREQUENCY Input frequency too high for encoder
input. Traversing speed may be
too fast.
1)
MEMORY ERR. Checksum error: Check the datum,
operating parameters and compen­sation values for multipoint axis error compensation. If this error recurs, contact your service
agency!
1)
These errors are important for the attached device. The error signal (pin 19) at D-sub connection EXT is active.
Display Effect/Cause
REC. ERROR Error during reception of param-
eter and compensation-value lists.
Other Error displays
If OVERFLOW appears, the measured value is too large or
1)
too small:
ä
Set a new datum.
or
ä
Traverse back.
If all sorting signals light up, the upper sorting limit is smaller than the lower limit:
ä
Change operating parameters P18 and/or P19.
To clear error messages:
Once you have removed the cause of the error:
ä
Clear the error message with the CL key.
16
Rear Panel
Ports X1, X2, X31 and X41 comply with the recommendations in EN 50178 for separation from line power.
Encoder input X1
HEIDENHAIN flange socket 9-pin Input signals 11 µA Maximum encoder cable length 30 m (98.5 ft) Maximum input frequency 100 kHz
Encoder input X2
HEIDENHAIN flange socket 12-pin Input signals 1 V Maximum encoder cable length 60 m (197 ft) Maximum input frequency 500 kHz
PP
Rear Panel, Accessories
PP
17
Rear Panel
Ports X1, X2, X31 and X41 comply with the recommendations in EN 50178 for separation from line power.
RS-232-C/V.24 data interface (X31)
25-pin D-sub connection (female)
Switching inputs and outputs EXT (X41)
25-pin D-sub connection (male)
Rear Panel, Accessories
Accessories
Connecting elements
Connector (female) 25-pin for D-sub connection X41
Id. Nr. 249 154 ZY
Connector (male) 25-pin for D-sub connection X31
Id. Nr. 245 739 ZY
Data interface cable, 3 m (9.9 ft), 25-pin for D-sub con- complete nection X31, Id. Nr. 274 545-01
Power switch
Data interface
Ground connection
Input X1 for one HEIDENHAIN
encoder
11 µA
PP
D-sub connection
Input X2 for one HEIDENHAIN
encoder
1 V
PP
18
Mounting
You can fasten the ND 281 B from below by using M4 screws (see illustration at right).
ND 281 B display units are stackable. Adhesive plug-in feet (supplied with your unit) prevent the stacked units from being moved out of place.
172 ± 0.2
6.77 ± .008"
Hole positions for mounting the ND display unit
15°
140 ± 0.2
5.51 ± .008"
Mounting
Alternatives of stacking the display units
19
Power Connection
The rear panel of the ND 281 B contains a connecting jack for a power cord with Euro connector (power cord supplied with the delivery).
Minimum cross section of the power cord: 0.75 mm Power supply: 100 Vac to 240 Vac (–15% to +10%)
Power supply: 50 Hz to 60 Hz (± 2 Hz)
Power Connection
A voltage selector is therefore not necessary.
Danger of electrical shock!
Unplug the power cord before opening the housing. Connect the grounding conductor. Do not interrupt the grounding conductor.
Danger to internal components!
Do not engage or disengage any connections while the unit is under power. Use only original replacement fuses.
To increase noise immunity, connect the ground terminal on the rear panel to the central ground point of the machine. (Minimum cross-section: 6 mm
20
2
2
)
Linear/Angular Measuring Modes
You can select the linear measuring mode or angular measuring mode by entering the code number 41 52 63:
Select the user parameter P00 CODE
(see "Operating Parameters").
Enter the code number 41 52 63.Confirm your entry with ENT.With the “.” or “–” key, select the ND LENGTH or
ND ANGLE mode, respectively.
Confirm your selection with ENT.The display unit resets itself.For further procedure, see “Switch-On, Traversing the
Reference Marks.”
Linear/Angular Measuring Modes
21
Operating Parameters
Operating parameters allow you to modify the operating characteristics of your ND display unit and define the evaluation of the encoder signals.
Operating parameters are designated by:
• The letter P,
• A two-digit parameter number, and
• An abbreviation.
Example: P01 INCH
Operating Parameters
The factory settings of the operating parameters are indicated in the parameter list (see “List of Operating Parameters”) in boldface type.
Parameters consist of “user parameters” and “protected operating parameters,” which can only be accessed by entering a code number.
User parameters
User parameters are operating parameters that can be changed without entering the code number:
P00 to P30, P50, P51, P79, P86, P98 The functions of the individual user parameters are detailed in
the list of operating parameters (see “List of Operating Parameters”).
To access a user parameter ... ... after switching on the display:
While ENT ... CL is displayed:
Display first user parameter.
MOD
... during operation:
Together:
Display first user parameter.
MOD
To go directly to a user parameter:
Together:
Press and hold CL while entering the first digit of the parameter number (here, 1).
Enter the second digit of the parameter number (here, 9). The display shows the selected user parameter.
22
Code number for changing protected operating parameters
If you wish to change protected operating parameters, you must first enter the code number 9 51 48:
ä
Select the user parameter P00 CODE.
ä
Enter the code number 9 51 48.
ä
Confirm entry with ENT.
Parameter P30 appears on the display. By paging through the list of operating parameters you can display—and, if necessary, change—each protected operating parameter and, of course, each user parameter.
Once you have entered the code number, the protected operating parameters remain accessible until the display unit is switched off.
Functions for changing the operating parameters
Function Key
Page forward in the list of operating parameters
Page backward in the list of operating parameters
Reduce parameter value
Increase parameter value
Correct entry and display parameter designations
Confirm change or numerical entry, exit the list of operating parameters
A changed parameter is stored as soon as you
• Exit the list of operating parameters
or
• Page forward or backward after the change.
MOD
Operating Parameters
23
List of Operating Parameters
Parameter Settings / Function
P00 CODE Enter a code number:
9 51 48: To change the protected
operating parameters
41 52 63: To select the linear or angular
measurement mode
10 52 96: Multipoint axis error
compensation 24 65 84: To lock the keyboard 66 55 44: To show the software version 24 65 82: Distance-to-go display 48 61 53: Input and output of parameter
and compensation-value lists
P01 Units of measurement
List of Operating Parameters
Display in millimeters MM Display in inches INCH
P02 X1/X2 Select an encoder input
Signals at X1 11 µAPP Signals at X2 1 VPP
P08 DISPL. Display mode
Decimal degrees DEC. DEGREE Deg., minutes, seconds DEG.MIN.SEC.
Parameter Settings / Function
P09 ANGLE Angle display
2)
+/- 180° +/- 180 DEG. +/-
¥
ENDLESS
360° 360 DEG.
P11 SCL Scaling factor
1)
Scaling factor off SCALING OFF Scaling factor on SCALNG. ON
P12 SCL Scaling factor
1)
Enter a numerical value
0.100000 < P12 < 9.999999
1)
Default setting: 1.000000
P17 CLASS Sorting and tolerance checking
Sorting and tol. checking ON CLASS ON Sorting and tol. checking OFF CLASS OFF
P18 L.CLASS Lower limit for sorting P19 U.CLASS Upper limit for sorting
2)
P21 SERIES Display for series of measurements
1)
OFF MIN MAX ACTL DIFF
24
1)
Only in linear measurement mode.
2)
Only in angular measurement mode.
Parameter Settings/ Function
P23 DISPL. Display stop for measured value output
Concurrent display, no display freeze;
the display value is the current actual value DISPL. ACTL.
Frozen display; hold display until next measured value output DISPL. HOLD
Frozen/concurrent display; freeze display as long as Pulse/Contact for measured value output is present DISPL. STOP
P30 DIR Counting direction
Positive counting direction with positive direction of traverse DIRECT. POS
Negative counting direction with positive direction of traverse DIRECT. NEG
1)
P31 S. PER. Signal period
of encoder
0.000 000 01 < P31 < 99 999.9999 Default setting: 10 µm
P33 COUNT Counting mode
1)
0-1-2-3-4-5-6-7-8-9 COUNT 0-1 0-2-4-6-8 COUNT 0-2 0-5 COUNT 0-5
P36 SP/R Signal periods per revolution
2)
1 < P36 < 999 999 Default setting: 36 000
Parameter Settings/ Function
P37 STEP Counting mode
2)
0-1-2-3-4-5-6-7-8-9 COUNT 0-1 0-2-4-6-8 COUNT 0-2 0-5 COUNT 0-5
P38 DEC. Decimal places
3)
1 / 2 / 3 / 4 / 5 / 6 (up to 8 for inch display)
P40 COMP. Select encoder compensation
No compensation COMP. OFF Multipoint: up to 64 compensation points for
linear encoders, up to 72 for angle encoders (fixed spacing of 5 deg.) COMP. MULTI
Linear compensation COMP. LIN
1)
Only in linear measurement mode.
2)
Only in angular measurement mode.
3)
Depends on signal period (P31) and unit of measure (P01)
List of Operating Parameters
1)
25
Parameter Settings / Function
P41 L.COMP. Linear error compensation
– 99 999.9 < P41 < + 99 999.9 [µm/m] Default setting: 0
Example: Find input value for P41
Displayed measuring length .............L
= 620.000 mm
d
Actual length (measured, e.g. with the VM 101 comparator
system from HEIDENHAIN) ..............L
= 619.877 mm
a
Length difference .............................DL = L
Compensation factor k (= P41):
k = DL / Ld = – 123 µm / 0.62 m .......k = – 198.4 [µm/m]
P42 BACKLASH Backlash compensation
List of Operating Parameters
Input range (mm): +9.999 to -9.999 Default setting: 0.000
= no backlash compensation During a change in direction, the mechanical
play between the encoder and table, the so­called backlash, can cause error. Positive backlash: The rotary encoder moves before the table, therefore the display value is too high (positive value input). Negative backlash: The rotary encoder moves after the table, therefore the display value is too low (negative value input).
1)
– Ld = – 123 µm
a
1)
Parameter Settings / Function
P43 REF Reference marks
One reference mark SINGLE REF. Distance-coded with 500 • SP
(SP: signal period) 500 SP Distance-coded with 1000 • SP
(e.g. for HEIDENHAIN LS ...C) 1000 SP Distance-coded with 2000 • SP 2000 SP Distance-coded with 5000 • SP 5000 SP
P44 REF Reference mark evaluation
Evaluate the reference mark REF. ON Do not evaluate the reference
mark REF. OFF
P45 ALARM Encoder monitoring
No monitoring ALARM OFF Monitor the frequency FREQUENCY Monitor contamination CONTAMINAT. Contamination + frequency FRQ. + CONT.
P50 RS232 Baud rate
110 / 150 / 300 / 600 / 1200 / 2400 / 4800 / 9600 / 19 200 / 38 400 baud
26
1)
Only in linear measurement mode.
Parameter Settings / Function
P51 RS232 Additional blank lines
during data output BK LINE 1
0 £ P51 £ÿ99 Default setting: 1
P62 A1 Trigger limit 1 P63 A2 Trigger limit 2 P79 PRESET Value for datum point
Enter numerical value for datum setting over switching input or with ENT key
P80 ENT-CL Set display
No zero reset/Set with CL/ENT CL-ENT OFF
Zero reset with CL
setting disabled with ENT CL......ON
Zero reset with CL and set with ENT to value selected in P79 CL-ENT ON
P82 DISPL.ON Message after switch-on
ENT...CL message MESSAGE ON No message MESSG. OFF
P85 EXT.REF External REF
REF over D-sub port EXT EXT.REF ON No REF over D-sub port EXT EXT.REF OFF
Parameter Settings / Function
P86 MOD In the linear measurement mode
First indicator after pressing MOD
START PRINT MIN ACTL MAX DIFF
In the angular measurement mode PRINT via MOD disabled PRINT OFF
PRINT via MOD enabled PRINT ON
P98 LANGUA. Conversational language
German LANGUAGE DE English LANGUAGE EN French LANGUAGE FR Italian LANGUAGE IT Dutch LANGUAGE NL Spanish LANGUAGE ES Danish LANGUAGE DA Swedish LANGUAGE SV Finnish LANGUAGE FI Czech LANGUAGE CS Polish LANGUAGE PL Hungarian LANGUAGE HU Portuguese LANGUAGE PT
List of Operating Parameters
27
Linear Encoders
The ND 281 B display unit is designed for use together with photoelectrical encoders with sinusoidal signals—11 µA 1V
.
PP
Display step with linear encoders
You can select a specific display step by adapting the
Linear Encoders
following operating parameters:
• Signal period (P31)
• Counting mode (P33)
• Decimal places (P38)
Example
Linear encoder with a signal period of 10 µm
Desired display step................. 0.000 5 mm
Signal period (P31) ................... 10
Counting mode (P33) ............... 5
Decimal places (P38)................ 4
The following tables will help you select the appropriate parameter settings.
PP
or
28
Recommended parameter setting s f or HEIDENHAIN linear encoders w it h 11 µA
Model
P 31 P 43
CT MT xx01 LIP 401A/401R
LF 103/103C LF 401/401C LIF 101/101C LIP 501/501C LIP 101
MT xx 10 Single 0.0005
LS 303/303C LS 603/603C
2
4
20 Single/1000 0.01
Reference
marks
Signal
period
in µm
Single 0.0005 Single
Single/5000 0.001
Single
Millimeters
Display step in mm
0.0002
0.0001
0.00005 Recommd. only for LIP 401
0.00002
0.00001
0.000005
0.0005
0.0002
0.0001
0.00005 Recommd. only for LIP 101
0.00002
0.00001
0.0002
0.0001
0.005
P 33 P 38
5 2 1 5
2 1 5
1 5 2 1 5
2 1
5 2 1
1 5
Count.
mode
4 4 4 5
5 5 6
3 4 4 4 5
5 5
4 4 4
2 3
Decimal
places
Inches
Display step in inches
0.00002
0.00001
0.000005
0.000002
0.000001
0.0000005
0.0000002
0.00005
0.00002
0.00001
0.000005
0.000002
0.000001
0.0000005 1 5
0.00002
0.00001
0.000005
0.0005
0.0002
signals
PP
Count.
mode
P 33 P 38
2 1 5 2
1 5 2
5 2 1 5 2
2 1 5
5 2
5 5 6 6
6 7 7
5 5 5 6 6
6 7
5 5 6
4 4
Decimal
places
Linear Encoders
29
Recommd. parameter settings for HEIDENHAIN linear encoders with 11 µAPP signals (continued)
Model
Linear Encoders
LS 106/106C LS 406/406C LS 706/706C ST 1201
LB 302/302C LIDA 10x/10xC
LB 301/301C 100 Single/1000 0.005
LIM 501 10240 Single 0.1
Reference
marks
Signal
period
in µm
P 31 P 43
20
40 Single/2000
Single/1000
-
Millimet er s
Display step in mm
0.001
0.0005
0.005
0.002
0.001
0.0005 Recommd. only for LB 302
0.0002
0.0001
0.002
0.001
0.01
0.05
P 33 P 38
1 5
5 2 1 5
2 1
5 2 1
1 1 5
Count.
mode
3 4
3 3 3 4
4 4
3 3 3
1 2 2
Inches
Display step in inches
Decimal
places
0.00005
0.00002
0.0002
0.0001
0.00005
0.00002
0.000001
0.0000005 1 5
0.0002
0.0001
0.00005
0.005
0.0005
0.002
P 33 P 38
5 2
2 1 5 2
2 1 5
5 5 2
Count.
mode
Decimal
places
5 5
4 4 5 5
5 6
4 4 5
3 4 3
30
Recommended parameter settings for HEIDENHAIN linear encoders with 1 VPP signals
Model
P 31 P 43
LIP 382 0.128 - 0.000002
MT xx81 LIP 481A/481R
LF 183/183C LF 481/481C LIF 181/181C LIP 581/581C VM 182
LS 186/186C LS 486/486C ST 1281
2 Single
4
20
Reference
marks
Signal
period
in µm
Single/5000 0.001
-
Single/1000 0.001
-
Millimeters
Display step in mm
0.000001 2 1
0.0005
0.0002
0.0001
0.00005 Recommd. only for LIP 481 X
0.00002
0.00001
0.000005
0.0005
0.0002
0.0001
0.00005 Recommd. only for VM 182
0.00002
0.00001
0.0005
P 33 P 38
5 2 1 5
2 1 5
1 5 2 1 5
2 1
1 5
Count.
mode
Decimal
6 6
4 4 4 5
5 5 6
3 4 4 4 5
5 5
3 4
places
Inches
Display step in inches
0.0000001
0.0000000 5
0.00002
0.00001
0.000005
0.000002
0.000001
0.0000005
0.0000002
0.00005
0.00002
0.00001
0.000005
0.000002
0.000001
0.0000005 1 5
0.00005
0.00002 5 2
Count.
mode
P 33 P 38
1 5
2 1 5 2
1 5 2
5 2 1 5 2
Decimal
places
7 8
5 5 6 6
6 7 7
5 5 5 6 6
6 7
5 5
Linear Encoders
31
Recommended parameter settings for HEIDENHAIN linear encoders with 1 VPP signals (continued)
Model
Linear Encoders
LB 382/382C LIDA 18x/18xC
LB 381/381C 100 Single/1000 0.005
Reference
marks
Signal
period
in µm
P 31 P 43
40 Single/2000
Millimeters
Display step in mm
0.005
0.002
0.001
0.0005 Recommd. only for LB 382
0.0002
0.0001
0.002
0.001
P 33 P 38
5 2 1 5
2 1
5 2 1
Count.
mode
3 3 3 4
4 4
3 3 3
Inches
Display step in inches
Decimal
places
0.0002
0.0001
0.00005
0.00002
0.00001
0.000005 1 5
0.0002
0.0001
0.00005
Count.
P 33 P 38
2 1 5 2
2 1 5
mode
Decimal
places
4 4 5 5
5 6
4 4 5
32
Recommended paramet e r settings for HEIDENHAIN angle encoders with 11 µA
Model
Signal
P36
ROD 450 / ROD 456 / ROD 486 / ROD 1080
ROD 250 C / ROD 280 C RON 255 C / RON 285 C
ROD 250 C / ROD 280 C ROD 255 C / RON 285 C ROD 700 C / ROD 780 C RON 705 C / RON 785 C RON 706 C / RON 786 C
RON 905 / 36 000 One Single 0.0001° 1 4 ROD 800 C / ROD 880 C
ROD 806 C / ROD 886 C
Convert decimal degrees into degrees, minutes, seconds
1 degree (1°) = 60 minutes (60'); 1 minute (1') = 60 seconds (60") 1 second (1") »ÿ0.000278°
3600 One Single
9000
18 000
36 000
periods per
revolution
Reference marks
Dist.­coded
Dist.­coded
Dist.­coded
Display step
Count.
P43
0.01°
0.005°
0.001°
500
1000
1000 0.0001° 1 4
0.005°
0.001° 5 1
0.001°
0.0005°
0.0001°
P37 P38
/ 1 VPP signals
PP
mode
1 5 1
1 5 1
Decimal
places
3 3
3 3
3 4 4
Angle Encoders
33
Multipoint Axis Error Compensation
If you want to use the multipoint axis error compensation feature, you must:
Activate the feature with operating parameter 40 (see “Operating Parameters”),
Traverse the reference marks after switching on the display unit,
Enter a compensation value table.
Your machine may have a nonlinear axis error due to factors such as axis sag or leadscrew errors. Such deviations are usually measured with a comparator measuring system (such as the HEIDENHAIN VM 101). In the linear measurement mode: You can make a compensation value table with 64 compensation values. In the angular measurement mode: You can make a compensation value table with 72
Multipoint Axis Error Compensation
compensation points (point spacing: 5 degrees). The datum is the reference point.
You select the compensation table through P00 CODE and by entering the code number 10 52 96 (see "Operating Parameters").
Ascertaining the compensation values
To ascertain the compensation values (e.g. with a VM 101) you must select the compensation table and then press the “–” key to select the REF display.
34
The letter “R” at the left of the display indicates that the displayed position value is given with respect to the reference mark. If “R” blinks, you must traverse the reference mark.
Entries in the compensation value table
• Datum
1)
: Here you enter the point at which the compensation is to begin. This point indicates the absolute distance to the reference point.
Do not change the datum after measuring the axis error and before entering the axis error into the compensation table.
• Spacing of the compensation points
1)
: The spacing of the compensation points is expressed as: Spacing = 2 x [µm]. Enter the value of the exponent
x
in into the compensation value table. Minimum input value: 6 (= 0.064 mm) Maximum input value: 20 (= 1048.576 mm) Example: 900 mm traverse with 15 compensation points
==> 60.000 mm spacing between points. Nearest power of two: 2
16
= 65.536 mm (see “Table for determining the point spacing”) Input value in the table: 16
• Compensation value: You enter the measured compensation value (in milli­meters) for the displayed compensation point. Compensa­tion point 0 always has the value 0 and cannot be changed.
1)
Only in the linear measurement mode
Table for determining the point spacing
Exponent Point spacing
in mm in inches
6 0.064 0.0025“ 7 0.128 0.0050“ 8 0.256 0.0100“ 9 0.512 0.0200“ 10 1.024 0.0403“ 11 2.048 0.0806“ 12 4.096 0.1612“ 13 8.192 0.3225“ 14 16.384 0.6450“ 15 32.768 1.290“ 16 65.536 2.580“ 17 131.072 5.160“ 18 262.144 10.32“ 19 524.288 20.64“ 20 1048.576 41.28“
Multipoint Axis Error Compensation
35
Selecting the compensation table, entering an axis correction
together
with
Select the operating parameters.
MOD
Select P00 CODE.
P00 CODE
Enter the code number 10 52 96, confirm with ENT.
DATUM (shown for approx. two seconds)
MOD
Multipoint Axis Error Compensation
SPACING
1)
Enter the active datum for the error on the axis to be corrected, e.g. 27 mm. Press MOD to select the next input box.
Enter the spacing of the compensation points on the axis to be corrected, for
4 x
MOD
example 210 mm (equals 1.024 mm). Press MOD four times to select COMP. NR. 01. (You cannot enter values in the POS. NR. 00, COMP. NR. 00 and POS. NR. 01 boxes.)
36
COMP. NR. 01
Enter the associated compensation value, e.g. 0.01 mm. Press MOD twice to
2 x
MOD
select COMP. NR. 02. (You cannot enter any values in the POS. NR. 02 box).
COMP. NR. 02
Enter all further compensation points. If you want so select a compensation
2 x
MOD
point directly, press CL and simultaneously enter the desired
1)
compensation point number.
Conclude entry.
1)
Only in the linear measurement mode
1)
For "angle measurement": Datum = reference point
1)
Point separation: 5 degrees
Deleting a compensation value table
together
with
P00 CODE
DATUM
DELETE
Select the operating parameters.
MOD
Select P00 CODE.
Enter the code number 10 52 96, confirm with ENT.
Select the “delete” function.
Confirm with ENT or cancel with CL.
Exit the compensation table mode.
Multipoint Axis Error Compensation
37
Switching Inputs and Outputs EXT (X41)
Inputs at D-sub connection EXT (X41)
Danger to internal components!
Voltage sources for external circuitry must conform to the recommendations in EN 50178 for low­voltage electrical separation. Connect inductive loads only with a quenching diode parallel to the inductance.
Only use shielded cables!
Connect the shield to the connector housing.
Outputs at D-sub connection EXT (X41)
Pin Function
14 Display value is zero 15 Measured value ³ trigger limit A1 (P62) 16 Measured value ³ trigger limit A2 (P63)
Switching Inputs and Outputs EXT (X41)
17 Measured value < lower sorting limit (P18) 18 Measured value > upper sorting limit (P19) 19 Error (see “Error Messages”)
38
1)
Only in the linear measurement mode
Pin Function
1, 10 0 V 2 Reset display to zero, clear error message 3 Set display to the value selected in P79 4 Ignore reference mark signals 5 Start series of measurements
1)
6 Externally select display value for series of
measurements
1)
7 Display MIN value of series of measurements 8 Display MAX value of series of measurements 9 Display difference MAX – MIN
1)
22 Pulse: Transmit measured value 23 Contact: Transmit measured value 25 Enable or disable REF mode
(current REF status is changed) 12, 13, 24 11, 20, 21
Do not assign
Vacant
Special case: Display current measured value ACTL
If you wish to display the current measured value ACTL of a series of measurements, note for inputs 7, 8 and 9: Either none or more than one of these inputs must be active.
1)
1)
Inputs
Outputs
Input signals
Internal pull-up resistor 1 kW, active with low level Trigger by making contact against 0 V or
by low-level signal over TTL logic device Delay for set/zero reset: td £ 2ms Minimum pulse duration for all signals: t
³ 30 ms
min
Signal level of inputs
Status Level
High + 3.9 V £ U £ + 15 V Low – 0.5 V £ U £ + 0.9 V; I £ 6 mA
E 0V
E 0V
t
min
t
min
³ÿ30 ms
Output signals
"Open collector" outputs, active with low level Delay until signal output: td £ 30 ms Signal duration of zero signal, trigger limit A1, A2: t0 ³ 180 ms
Signal level of outputs
Status Level
High U £ + 32 V; I £ 10 µA Low U £ + 0.4 V; I £ 100 mA
I 100 mA
+ UB 32 V
C
B
U
CE
E
Pin 1.10
0 V
Switching Inputs and Outputs EXT (X41)
39
Setting and zero resetting the display
With an external signal, you can set the display to the value selected in parameter P79 (pin 3) or reset each axis to zero (pin 2).
Enabling and disabling REF mode
Operating parameter P85 allows you to activate the input (pin 25) which will be used for setting the display externally to REF mode when the unit is switched on or when the power is restored after an interruption. The next signal deactivates REF mode again (switchover function).
Ignoring reference mark signals
If this input (pin 4) is active, the display will ignore all
Switching Inputs EXT (X41)
reference mark signals. A typical application of this function is for measuring lengths with a rotary encoder and spindle; in this case, a cam switch releases the reference mark signal at a preset position.
40
Externally selecting MIN/MAX Starting a series of measurements Switching the display between MIN/MAX/DIFF/ACTL
You can activate the operating mode for finding minimum and maximum values from a series of measurements with an external signal (pin 6, low-level signal must be present continuously). The setting selected with MOD or operating parameter P21 is disabled. You can switch to MIN/MAX/DIFF/ ACTL display (pins 7, 8, 9, low-level signal must be present continuously) and START (pin 5, Pulse) of a new series of measurements only by external signal over the switching inputs.
1)
1)
Only in linear measurement mode.
Switching signals
As soon as the trigger points defined in parameters are reached, the corresponding outputs (pins 15, 16) are activated. You can set up to two trigger points. The switching point “zero” has a separate output (see “Zero crossover”).
Signals for sorting and tolerance checking
If the sorting limits defined in parameters are exceeded, the corresponding outputs (pins 17, 18) are activated.
Output
Switching point
Measured value < lower sorting limit
Path
Signals Operating parameters Pin
Switching signals P62, switching limit 1 15
P63, switching limit 2 16
Sorting signals P18, lower sorting limit 17
P19, upper sorting limit 18
Zero crossover
The display value “zero” activates the corresponding output (pin 14). Minimum signal duration is 180 ms.
Measured value > upper sorting limit
Lower limit Upper limit
Path
Weg
Path
5
Pin 15
(A1)
t
d
t
v
t
t
d
v
Time curve of signal at pin 15 for trigger limit (A1) = 5 mm , td £
t
t
30 ms
Switching Outputs EXT (X41)
41
Switching signal for errors
The display unit permanently monitors functions such as measuring signal, input frequency, and data output, and displays an error message if it detects an error.
If errors occur that seriously influence measurement or data output, the display unit activates a switching output. This feature allows monitoring of automated processes.
Switching Outputs EXT (X41)
ERROR xx
Error
42
Locking the Keypad
You can lock or release the keypad by entering the code number 24 65 84:
Select the user parameter P00 CODE (see “Operating
Parameters”).
Enter the code number 24 65 84.Confirm the entry with ENT.With the “•” or “–” key, select KEYS ON or
KEYS OFF.
Confirm your selection with ENT. If the keypad is locked, you can only select the datum or
select P00 CODE over the MOD key.
Locking the Keypad
43
Displaying the Software Version
To display the software version of the display unit, enter the code number 66 55 44:
Select the user parameter P00 CODE. ➤ Enter the code number 66 55 44.Confirm your entry with ENT.The display unit shows the software number.With the “–” key you can switch the display to the date of
issue.
To exit the software version display mode, press ENT.
Displaying the Software Version
44
Distance-to-Go Mode
1)
Function of switching outputs A1 and A2
Normally, the display shows the actual position of the en­coder. However, it is often more helpful to display the re­maining distance to an entered nominal position—especially when you are using the display unit for machine tools and au­tomation purposes. You can then position simply by traversing to display value zero.
You can access the distance-to-go display by entering the
code number 24 65 82.
Display Meaning
DELTA ON Distance-to-go display active DELTA OFF Distance-to-go display not active
“Traversing to zero” with distance-to-go display
ä
Select datum point 2.
ä
Enter the nominal position.
Move the axis until the display value is zero.
In the distance-to-go mode, switching outputs A1 (pin 15) and A2 (pin 16) have a different function: they are symmetrical to the display value zero. For example, if a switching point of 10 mm is entered in P62, output A1 switches at both +10 mm and –10 mm. The figure below shows output signal A1 when approaching zero from the negative direction.
t
d1
t
d2
Time curve of a signal for switching limit (A1) = 10 mm, t
£
d1
30 ms, t
d2
£
180 ms
Distance-to-Go Mode
1)
Only in linear measurement mode
45
RS-232-C/V.24 Data Interface (X31)
The RS-232-C/V.24 interface (X31) of your display unit enables you to transmit measured data in ASCII format, for example to a printer or PC.
Connecting cable
You can use a connecting cable with full wiring (figure at upper right) or simplified wiring (below right). A cable with full wiring is available from HEIDENHAIN.
Data transmission cable for RS-232-C/V.24 D-sub (male) 25-pin/D-sub (female) 25-pin
Id. Nr. 274 545-.. On this type of cable, pin 6 and pin 8 are additionally connected over a jumper.
Data transmission cable for RS-232-C/V.24 D-sub (male) 25-pin/D-sub (female) 9-pin
Id. Nr. 368 017-..
RS-232-C/V.24 Data Interface (X31)
Maximum cable length: 20 m (66 ft)
25 pol. 25 pol.
Data transmission cable 25-pin/25-pin (full wiring)
25 pol. 25 pol.
Data transmission cable 25-pin/25-pin (simplified wiring)
46
25 pol. 9 pol.
Data transmission cable 25-pin/9-pin
Pin layout RS-232-C/V.24 (X31)
Data format and control characters
Pin Signal Assignment
1 CHASSIS GND Chassis ground 2 TXD Transmitted data 3 RXD Received data 4 RTS Request to send 5 CTS Clear to send 6 DSR Data set ready 7 SIGN. GND Signal ground 8 to 19
Not assigned
20 DTR Data terminal ready 21 to 25 –
Not assigned
Levels for TXD and RXD
Logic level Voltage level
Active – 3 V to – 15 V Not active + 3V to +15 V
Levels for RTS, CTS, DSR and DTR
Logic level Voltage level
Active + 3 V to + 15 V Not active – 3V to – 15 V
Data format 1 start bit
7 data bits Even parity bit 2 stop bits
Control characters Call measured value: STX (Ctrl B)
Interrupt DC3 (Ctrl S) Continue DC1 (Ctrl Q)
Interrogate error message: ENQ (Ctrl E)
Example: Data sequence during measured value output
Measured value = – 5.23 mm The measured value is within the sorting limits ( = ) and is the current value ( A ) of a series of measurements.
Measured value output
- 5 . 2 3 = A < C R > < L F >
123456 7 8
1
Algebraic sign
2
Numerical value with decimal point (10 characters on the whole, leading zeros are output as blank spaces.) (Angle measurement “min, sec” up to 3 dec. spaces.)
3
Blank space
4
Unit: Blank space = mm; " = inch; ? = fault
5
Sorting status (<, >, =; ? if P18 > P19) or blank space
6
Series of measurements (S = MIN; A = ACTL; G = MAX; D = DIFF) or blank space
7
CR (carriage return)
8
LF (line feed)
RS-232-C/V.24 Data Interface (X31)
47
Operating parameters for measured value output
Parameter Function
P50 RS232 Baud rate P51 RS232 Number of additional blank lines for
measured value output
Display freeze during measured value output
In operating parameter P23, you can specify how the measured value output signal will affect the display unit.
Display freeze during measured value output P23
Concurrent display, no display freeze: The
display value is the current measured value DISPL. ACTL.
Frozen display: Display is stopped
RS-232-C/V.24 Data Interface (X31)
(frozen) and updated by every measured value output signal DISPL. HOLD
Frozen/concurrent display: Display is frozen as long as a measured value output signal is present DISPL. STOP
Measured value output via PRINT function
In the linear measurement mode, press MOD repeatedly until the PRINT indicator starts blinking. Start the measured value output with ENT. In the angular measurement mode, press MOD (this feature can be disabled with operating parameter 86).
Duration of measured value transfer
187 + (11 • number of blank lines)
t
= [s]
D
Indicator preselection (linear measurement mode)
Operating parameter P86 allows you to define which indicator is displayed first when MOD is pressed.
baud rate
48
Measured value output after signal through the “Contact” or “Pulse” inputs
To start measured value output through the EXT interface (X41), you can either:
ä
Close the “Contact” input (pin 23 on X41) against 0 V, for example with a simple switch (make contact);
or
ä
Close the “Pulse” input (pin 22 on X41) against 0 V, for example by triggering the input with a TTL logic device (such as SN74LSxx).
Characteristic times for measured value output
EXT(X41)
Pin 23 Pin 1(0V)
EXT(X41)
Pin 22 Pin 1(0V)
Process Time
Minimum duration of “Contact” signal te ³ 7 ms Minimum duration of “Pulse” signal te ³
1.5 µs Storage delay after “Contact” t1 £ 5 ms Storage delay after “Pulse” t1 £ 1 µs Measured value output after t2 £ Regeneration time t3 ³
57 ms 0
Duration of measured value transfer
187 + (11 • number of blank lines)
= [s]
t
D
baud rate
Triggering the “Contact” and “Pulse” inputs at D-sub connection EXT (X41)
t
e
t
1
t
2
t
D
t
e
t
3
Signal transit times for measured value output after “Pulse” or “Contact”
RS-232-C/V.24 Data Interface (X31)
49
t
1
t
2
Ctrl B
t
D
t
3
Ctrl B
Measured value output with CTRL B
If the display unit receives the control character STX (Ctrl B) over the RS-232-C/V.24 interface, it transmits the current measured value back over the interface. Ctrl B is received over the RXD line of the interface and the measured values are transferred over the TXD line.
The measured values can be received and saved using a terminal program (e.g. HyperTerminal, which is a component of Windows
®
).
The basic program at the upper right shows the fundamental structure of a program for measured value output.
10 L%=18 20 CLS 30 PRINT "V.24/RS-232-C" 40 OPEN "COM1:9600,E,7" AS#1 50 PRINT #1, CHR$ (2); 60 IF INKEY$<>""THEN 130 70 C%=LOC(1) 80 IF C%<L%THEN 60 90 X$=INPUT$(L%,#1) 100 LOCATE 9,1 110 PRINT X$; 120 GOTO 50 130 END
Characteristic times for measured value output
Process Time
Storage after t1 £ 1ms
RS-232-C/V.24 Data Interface (X31)
Measured value output after t2 £ 50 ms Regeneration time t3 ³ 0
These times are prolonged if functions are active (for example, series of measurements with DIFF value display).
Duration of measured value transfer
187 + (11 • number of blank lines)
baud rate
50
tD = [s]
BASIC program for measured value output with “Ctrl B”
Propagation times for measured value output after “Ctrl B”
Input/Output of Parameter and Compensation-Value Lists
Calling the "data transfer" function:
together with
Select the operating parameters.
MOD
Select P00 CODE.
REC. PARAM.
If reqd.
or
The display unit is ready to receive a parameter list over the RS-232-C/V.24 interface. After successful reception of the parameter list, the display unit resets itself and restarts. Continue in the data transfer menu with the decimal point key.
P00 CODE
Enter the code number 48 61 53 and confirm with ENT.
Data transfer function:
TRANSFER
Continue with ENT.
SEND PARAM.
If reqd.
or
Press the ENT key if you want to transmit the parameter list over the RS-232-C/V.24 interface. After transmission, return to the beginning ("TRANSFER") to transmit or receive further lists. Press the decimal point key to continue in the data tranfer menu.
SEND COMP.
If reqd.
or
REC. COMP.
If reqd.
or
With the ENT key, you transmit the compensation value list over the RS-232-C/ V.24 interface. After transmission, return to the beginning ("TRANSFER") to transmit or receive more lists. Continue in the data transfer menu with the decimal point key.
The display unit is ready to receive a parameter list over the RS-232-C/V.24 interface. After successful reception, return to the beginning ("TRANSFER") to transmit or receive more lists. Continue in the data transfer menu with the decimal point key.
Exit the data transfer function.
Input/Output of Parameter and Compensation-Value Lists
51
Note on the input/output of parameter and compensation-value lists
With a terminal program (e.g. HyperTerminal, included with Windows unit over the RS-232-C/V.24 interface as text files and save them on your PC. Each list must be saved as a separate text file. You can then use the terminal program to transmit the text files back to the display unit .
If you wish, you can edit the text files with a text editor and change the parameter values, for example. However, this requires knowledge of the different output formats of the lists (see following pages). When receiving lists, the display unit expects the same list structure as used for the output.
When receiving lists, the display unit first waits for the starting character < * >. The receiving mode ends as soon as the display unit has received the final character < * >.
The lists received are first checked for the type of display unit (line 2 of output list). The receiving display unit accepts only lists prepared by the same type of display unit. Furthermore, the list is checked for completeness. Lists that contain, for example, too many or too few parameters are ignored. If an error occurs, the following error message appears:
REC. ERROR To clear the error message, press the CL key.
®
), you can receive the lists output by the display
If the display unit receives invalid parameter values, it sets the respective operating parameter to the default setting. Example: “P01 INCH = INCH = 3” The value 3 is not allowed. The parameter P01 is set to the default setting “P01 MM = MM = 0”.
52
Input/Output of Parameter and Compensation-Value Lists
Output Format of the Parameter List
1st line
Each parameter output begins with the start character < * > ( HEX: 0x2A) * <CR> <LF>
3 characters
2nd line
Output of the counter designation N D - 2 8 1 B M M <CR> <LF>
13 characters 5 characters 2 characters Model of display unit Unit of meas. End of line
Subsequent lines for the individual parameters: a: Parameters:
Parameter settings can be changed with the MINUS key (e.g.: counting direction positive/counting direction negative etc.) Examples:
P 1 1 S C L = S C A L I N G . O F F = 0 <CR> <LF> 15 characters 3 char. 13 characters 3 char. 6 characters 2 characters
P 5 0 R S 2 3 2 = 3 8 4 0 0 B A U D = 3 8 4 0 0 <CR> <LF> 15 characters 3 char. 13 characters 3 char. 6 characters 2 characters
Parameter designation Text Separator Parameter in plain language Separator Parameter value End of line left-aligned right-aligned block right-aligned block right-aligned
Output Format of the Parameter List
53
b: Parameters:
Parameter settings can be changed by entering a value (e.g.: LINEAR COMP. 13.600 etc.)
P 1 8 L . C L A S S . = + 1 2 0 . 0 0 0 0 <CR> <LF> 15 characters 3 char. 13 characters 2 characters
P 4 1 L . C O M P . = - 1 4 0 0 . 0 <CR> <LF> 15 characters 3 char. 13 characters 2 characters
Parameter designation Text Separator Parameter value End of line left-aligned right-aligned block right-aligned
Last line:
Each parameter list ends with the final character <*> (HEX: 0x2A) * <CR> <LF>
Output Format of the Parameter List
The parameter text is sent in the language you have set and corresponds to the dialog text that is shown on the display unit. The parameter value, not the text, is definitive when the parameters are transferred to the display unit.
54
Parameter list for ND 281 B: "Linear measurement" mode of operation (factory default setting) Parameter list Description
* Start character (*); ND-281 B MM Model of display unit; MM or IN; P01 MM = MM = 0 Unit of measure: MM = 0; INCH = 1; P02 X1/X2 = X1 11 uAPP = 0 Encoder input: X1 11µAPP = 0; X2 1VPP = 1; P11 SCL = SCALING OFF = 0 SCALING OFF = 0; ON = 1; P12 SCL = 1.000000 SCALING = 1.000000; (value input without sign) P17 CLASS. = CLASS. OFF = 0 Sorting: CLASS. OFF = 0; CLASS. ON = 1; P18 L.CLASS. = + 0.0000 Lower limit: L.CLASS = 0; (value input) P19 U.CLASS. = + 0.0000 Upper limit: U.CLASS = 0; (value input) P21 SERIES = DISPL. OFF = 0 SERIES: DISPL.OFF =0; MIN=1; MAX=2; ACTL=3; DIFF=4; P23 DISPL. = DISPL. ACTL. = 0 DISPLAY: ACTUAL = 0; HOLD = 1; STOP = 2; P30 DIR = DIRECT. POS = 0 COUNTING DIRECTION POS = 0; NEG = 1; P31 S. PER. = 10 SIGNAL PERIOD = 10 µm; (value input without sign) P33 STEP = COUNT 0-5 = 5 COUNTING MODE 0-5 = 5; 0-2 = 2; 0-1 = 1; P38 DEC. = DP POS 4 = 4 DECIMAL PLACES 4 (range: 1-8) P40 COMP. = COMP. OFF = 0 COMPENSATION OFF = 0; LIN = 1; MULTI = 2; P41 L.COMP. = + 0.0 LINEAR COMPENSATION= 0 µm/m (value input) P42 BKLASH = + 0.0000 BACKLASH compensation= 0.0000 mm (value input) P43 REF = SINGLE REF. = 0 SINGLE REF. = 0; 500; 1000; 2000; 5000S P; P44 REF = REF. ON = 1 REF.ON = 1; REF. OFF = 0; P45 ALARM = FRQ.+ CONT. = 3 OFF = 0; FRQ. = 1; CONTAMINAT. = 2; FRQ + CONT. = 3; P50 RS232 = 9600 BAUD = 9600 BAUD RATE = 9600; (110-38400) P51 RS232 = BK LINE 1 = 1 LINE FEEDS = 1; (0-99) P62 A1 = + 0.0000 Switching limit 1: A1 = 0; (value input) P63 A2 = + 0.0000 Switching limit 2: A2 = 0; (value input) P79 PRESET = + 0.0000 DATUM SETTING= 0; (value input) P80 ENT-CL = CL-ENT OFF = 0 CL-ENT OFF=0; CL-ON = 1; CL-ENT ON = 2; P82 DISPL.ON = ENT..CL ON = 1 DISPLAY: ENT...CL ON = 1; ENT...CL OFF = 0; P85 EXT.REF = EXT.REF OFF = 0 EXTERNAL REF OFF = 0; EXTERNAL REF ON= 1; P86 MOD = MOD START = 0 MOD key: START= 0; PRINT = 1; MIN = 2; ACTL = 3; MAX = 4; DIFF = 5; P98 LANGUA. = LANGUAGE DE = 1 NATIONAL LANGUAGE: 0 = EN; 1 = DE; 2 = FR; 3 = IT; 4 = NL; 5 = ES; * 6 = DA; 7 = SV; 8 = FI; 9 = CS; 10 = PL; 11 = HU;
12 = PT;
Final character(*);
Output Format of the Parameter List
55
56
Parameter list for ND 281 B: "Angular measurement" mode of operation (factory default setting) Parameter list Description
* Start character (*); ND-281 B DEC Model of display unit; DEC (decimal) or DMS (min-sec); P02 X1/X2 = X1 11 uAPP = 0 Encoder input: X1 11µAPP = 0; X2 1VPP = 1; P08 DISPL. = DEC. DEGREE = 0 Display: DEC.DEGREE = 0; DEG.MIN.SEC = 1; P09 ANGLE = +/-180 DEG. = 0 Angle: +/- 180 DEG. = 0; 360 DEG. = 1; ENDLESS = 2; P17 CLASS = CLASS. OFF = 0 Sorting: CLASS. OFF = 0; CLASS. ON = 1; P18 L.CLASS. = + 0.0000 Lower limit: L.CLASS = 0; (value input) P19 U.CLASS. = + 0.0000 Upper limit: U.CLASS = 0; (value input) P23 DISPL. = DISPL. ACTL. = 0 DISPLAY: ACTUAL = 0; HOLD = 1; STOP = 2; P30 DIR = DIRECT. POS = 0 COUNTING DIRECTION POS = 0; NEG = 1; P36 SP/R = 36000 SIGNAL PERIODS / R = 36000 (value input); P37 STEP = COUNT 0-5 = 5 COUNTING MODE 0-5 = 5; 0-2 = 2; 0-1 = 1; P38 DEC. = DP POS. 4 = 4 DECIMAL PLACES 4 (range: 1-8) P40 COMP. = COMP. OFF = 0 COMPENSATION OFF = 0; LIN = 1; MULTI = 2; P43 REF = SINGLE REF. = 0 SINGLE REF. = 0; 500; 1000; 2000; 5000 SP; P44 REF = REF. ON = 1 REF.ON = 1; REF. OFF = 0; P45 ALARM = FRQ.+ CONT. = 3 OFF = 0; FRQ.= 1; CONTAMINAT.= 2; FRQ+CONT. = 3; P50 RS232 = 9600 BAUD = 9600 BAUD RATE = 9600; (110-38400) P51 RS232 = BK LINE 1 = 1 LINE FEEDS = 1; (0-99)
Output Format of the Parameter List
P62 A1 = + 0.0000 Switching limit 1: A1 = 0; (value input) P63 A2 = + 0.0000 Switching limit 2: A2 = 0; (value input) P79 PRESET = + 0.0000 DATUM SETTING = 0; (value input) P80 ENT-CL = CL-ENT OFF = 0 CL-ENT OFF= 0; CL-ON = 1; CL-ENT ON = 2; P82 DISPL.ON = ENT..CL ON = 1 DISPLAY: ENT...CL ON = 1; ENT...CL OFF = 0; P85 EXT.REF = EXT.REF OFF = 0 EXTERNAL REF OFF = 0; EXTERNAL REF ON = 1; P86 MOD = PRINT OFF = 0 MOD KEY: PRINT OFF = 0; PRINT ON = 1; P98 LANGUA. = LANGUAGE DE = 1 NATIONAL LANGUAGE: 0 = EN; 1 = DE; 2 = FR; * 3 = IT; 4 = NL; 5 = ES;
Final character (*);
6 = DA; 7 = SV; 8 = FI; 9 = CS; 10 = PL; 11= HU; 12 = PT;
Output Format of the Compensation-Value Table
Line: Start
Each compensation-value output begins with the start character < * > ( HEX: 0x2A) * <CR> <LF> 3 characters
Line: Counter model designation
Output of model designation and unit of measure N D - 2 8 1 B M M <CR> <LF> 13 characters 5 characters 2 characters
Model display unit, left-aligned Unit of meas. End of line
Line: Compensation value 0
Output of compensation-value no. 0 C M P . N R . 0 0 = + 0 . 0 0 0 0 <CR> <LF> 13 characters 3 char. 13 characters 2 characters
Compensation-value no., left-aligned Sep. block Compensation value, right-aligned End of line
Output of compensation values 1 - 63
Output of compensation values C M P . N R . 6 3 = + 0 . 0 1 2 3 <CR> <LF> 13 characters 3 char. 13 characters 2 characters
Compensation-value no., left-aligned Sep. block Compensation-value, right-aligned End of line
Last line:
Each compensation-value table ends with the final character <*> (HEX: 0x2A) * <CR> <LF> 3 characters
Output Format of the Compensation-Value Table
57
Compensation-value table for ND 281 B (linear measurement): factory default setting Compensation-value table Description:
* Start character ( * ); ND-281 B MM Model of display unit; unit of measure (MM or IN); SCALING = 14 Spacing = 14 ( range: 6 – 20) DATUM = + 0.0000 Datum point 0 mm (value input) CMP. NR. 00 = + 0.0000 Compensation value 0 = 0.000 mm (compensation value 0 is always 0) CMP. NR. 01 = ---------- Compensation value 1 = no value entered CMP. NR. 02 = ---------- Compensation values 2 – 63: no values entered (axis is not corrected) CMP. NR. 03 = ---------- Compensation-value table is empty CMP. NR. 04 = ---------­CMP. NR. 05 = ---------­CMP. NR. 06 = ---------­CMP. NR. 07 = ---------­. . . . . . . . . . . .
Output Format of the Compensation-Value Table
CMP. NR. 60 = ---------­CMP. NR. 61 = ---------­CMP. NR. 62 = ---------­CMP. NR. 63 = ----------
58
* Final character (*);
Compensation-value table for ND 281 B (angular measurement): Active compensation Compensation-value table Description:
* Start character ( * ); ND-281 B DMS Model of display unit; DEC (decimal) or DMS (deg-min-sec); CMP. NR. 00 = + 0.00.00 Compensation value 0 = 0.0000mm (compensation value 0 is always 0) CMP. NR. 01 = + 0.00.03 Compensation values 1 – 18 are assigned values (value input) CMP. NR. 02 = + 0.00.05 i.e. a rotary encoder is corrected in 5-degree steps from 0 - 90 degrees CMP. NR. 03 = + 0.01.01 Input in deg-min-sec CMP. NR. 04 = + 0.00.43 CMP. NR. 05 = + 0.00.21 CMP. NR. 06 = + 0.00.06 CMP. NR. 07 = - 0.00.04 CMP. NR. 08 = - 0.00.12 CMP. NR. 09 = - 0.00.24 CMP. NR. 10 = - 0.00.44 CMP. NR. 11 = - 0.00.52 CMP. NR. 12 = - 0.00.43 CMP. NR. 13 = - 0.00.35 CMP. NR. 14 = - 0.00.24 CMP. NR. 15 = - 0.00.19 CMP. NR. 16 = - 0.00.13 CMP. NR. 17 = - 0.00.05 CMP. NR. 18 = + 0.00.00 CMP. NR. 19 = ---------- Compensation values 11 – 71: no values entered (memory empty) CMP. NR. 20 = ---------­. . . CMP. NR. 70 = ---------­CMP. NR. 71 = ---------­* Final character (*);
Output Format for the Compensation-Value Table
59
Remote Operation over the RS-232-C/V.24 Data Interface
You can operate the display unit remotely over the RS-232-C/ V.24 data interface. The following commands are available on the ND 281 B: Format: <ESC>TXXXX<CR> Pressed key <ESC>AXXXX<CR> Output of screen contents <ESC>FXXXX<CR> Execute function <ESC>SXXXX<CR> Special function
Sequence of commands Meaning <ESC>T1000<CR> ‘CE+0‘ keys <ESC>T1001<CR> ‘CE+1‘ keys <ESC>T1002<CR> ‘CE+2‘ keys <ESC>T1003<CR> ‘CE+3‘ keys <ESC>T1004<CR> ‘CE+4‘ keys <ESC>T1005<CR> ‘CE+5‘ keys <ESC>T1006<CR> ‘CE+6‘ keys <ESC>T1007<CR> ‘CE+7‘ keys <ESC>T1008<CR> ‘CE+8‘ keys <ESC>T1009<CR> ‘CE+9‘ keys
Remote Operation over the RS-232-C/V.24 Data Interface
60
Sequence of commands Meaning <ESC>T0000<CR> ‘0‘ key <ESC>T0001<CR> ‘1‘ key <ESC>T0002<CR> ‘2‘ key <ESC>T0003<CR> ‘3‘ key <ESC>T0004<CR> ‘4‘ key <ESC>T0005<CR> ‘5‘ key <ESC>T0006<CR> ‘6‘ key <ESC>T0007<CR> ‘7‘ key <ESC>T0008<CR> ‘8‘ key <ESC>T0009<CR> ‘9‘ key <ESC>T0100<CR> ‘CL' key <ESC>T0101<CR> ‘-‘ key <ESC>T0102<CR> ‘.‘ key <ESC>T0104<CR> ‘ENT‘ key <ESC>T0105<CR> ‘MOD‘ key <ESC>T0107<CR> ‘1/2‘ (datum) key
<ESC>A0000<CR> Output of model designation <ESC>A0100<CR> Output of 14-segment display <ESC>A0200<CR> Output of current value <ESC>A0301<CR> Output of error text <ESC>A0400<CR> Output of software number <ESC>A0900<CR> Output of status indicators
<ESC>F0000<CR> REF function <ESC>F0001<CR> Start measurement <ESC>F0002<CR> Print
<ESC>S0000<CR> Counter RESET <ESC>S0001<CR> Lock keypad <ESC>S0002<CR> Release keypad
1)
Only in the linear measurement mode.
1)
Description of RS-232-C/V.24 commands:
The display unit supports the XON-XOFF protocol when executing commands. As soon as the internal character buffer (100 characters) is full, the display unit sends the control character XOFF to the sender. After the buffer has been executed, the display unit sends the control character XON to the sender and is then ready to receive data again.
Pressed key (TXXXX commands)
The display unit acknowledges each identified key command by sending the control character ACK (Acknowledge Control­F). Then the key command becomes effective.
Output of 14-segment display:
The contents displayed are transmitted (also dialogs and error messages).
<STX> - 1 2 3 4 5 . 6 7 8 9 <CR> <LF> String: STX;
Min. 10 to max. 13 characters; CR; LF; (depending on the number of commas and decimal points)
Output of current value:
The current position value (without decimal point, with leading zeros) is transmitted.
The display unit responds to commands that cannot be identified or to invalid commands by sending the control character NAK (No acknowledge Control-U).
Output of model designation:
The model of display unit, software number, and the date of software release are transmitted.
Example: <STX> N D - 2 8 1 B <CR> <LF>
3 4 9 7 9 7 - 0 4 <CR> <LF> 2 0 0 1 - 0 5 - 0 4 <CR> <LF>
String: STX;
10 characters; CR; LF; 10 characters; CR; LF; 10 characters; CR; LF;
<STX> + 1 2 3 4 5 6 7 8 9 <CR> <LF> String: STX;
Algebraic sign; numerical value with 9 characters; CR; LF;
Output of error text:
The displayed error text is transmitted. (Functions only if an error message is displayed.)
<STX> F O R M A T . E R R . <CR> <LF> String: STX;
13 characters; CR; LF;
Remote Operation over the RS-232-C/V.24 Data Interface
61
Remote Operation over the RS-232-C/V.24 Data Interface
62
Output of software number:
The current software number is transmitted. <STX> 3 4 9 7 9 7 - 0 4 <CR> <LF> String: STX;
10 characters; CR; LF;
Output of status indicators:
The condition of the status display is transmitted. Example: 0 = status indicator dark 1 = status indicator glows 2 = status indicator blinks
<STX> 0 1 0 0 1 0 0 1 0 0 1 2 0 0 <CR><LF> a b c d e f g h i j k l m n String: STX;
4 characters; CR; LF;
a = REF (reference point) h = < (sorting) b = datum 1 i = = (sorting) c = datum 2 j = > (sorting) d = SET (datum setting) k = MIN (meas. series) e = START (measurement series) l = ACTL (meas. series) f = PRINT (data output) m = MAX (meas. series) g = inch (inch display) n = DIFF (meas. series)
Execute functions (FXXX commands):
The display unit acknowledges every correctly received command by transmitting the control character ACK (Acknowledge, Control-F). Then it executes the command. It answers unrecognized or invalid commands by sending the control character NAK (No acknowledge Control U).
REF function:
Activate or deactivate REF mode (current REF condition is changed).
Print
Output of the current measured value. The measured value (string) is transmitted as described in the manual (page 47). Same function as calling the measured value with STX (Control B).
Special functions (SXXX commands): Counter RESET:
The software resets the counter and the counter restarts. (Function same as switching the display unit off and on.)
Locking the keypad:
The display unit acknowledges the special function by sending the control character ACK (acknowledge). All keys of the display unit are locked. The counter can be operated only by external RS-232-C/V.24 commands and via X41. The keyboard can be unlocked either by sending the special function
Specifications
Housing ND 281 B
Benchtop design, cast-metal housing (W · H · D) 239 mm · 84.6 mm · 224 mm
Operating temperature 0° to 45° C (32° to 113° F)
Storage temperature –20 °C to 70 °C (–4 °F to 158 °F) Weight Approx. 1.5 kg (3.3 lb) Relative humidity < 75% annual average
< 90% in rare cases
Power supply Primary-clocked power supply
100 Vac to 240 Vac (–15% to +10%) 50 Hz to 60 Hz (± 2 Hz)
Line fuse F 1 A inside the housing Power consumption 8 W (typically) Electromagnetic
compatibility Class B according to EN 55022
Noise immunity As per VDE 0843 Parts 2 and 4,
severity 4
Protection IP40 according to IEC 529 Encoder inputs For linear and angle encoders with
sinusoidal output signals (11 µA
PP
Reference mark evaluation for distance­coded and single reference marks
Input frequency X1 11 µA
PP
: Max. 100 kHz for 30 m cable length X2 1 V
PP
:
Max. 500 kHz for 60 m cable length
Display step Adjustable Datum points Two Functions • Series of measurements
1)
• Sorting and tolerance checking
• Switching and sorting signals
• Set display and reset display to
zero with external signal
• Measured value output
RS-232-C/V.24 Baud rates: Interface 110, 150, 300, 600, 1200, 2400,
4800, 9600, 19 200, 38 400 baud
/1 VPP);
Specifications
1)
Only in linear measurement mode.
63
ND 281 B: Dimensions in mm/inches
1
SET START PRINT2 inch
REF
< = >
MIN ACTL
MAX DIFF
Specifications
X
35
1.43"
64
325
33.5
1.37"
12.8"
172 ± 0.2
6.77" ± .008"
239
9.41"
140 ± 0.2
5.51" ± .008" 224
8.82"
DR. JOHANNES HEIDENHAIN GmbH
Dr.-Johannes-Heidenhain-Straße 5
83301 Traunreut, Germany
{ +49 (86 69 ) 31-0 | +49 (86 69) 50 61
e-mail: info@heidenhain.de
Technical support | +49 (86 69) 31-10 00 Measuring systems { +49 (8669) 31-31 04
e-mail: service.ms-support@heidenhain.de
TNC support { +49 (86 69 ) 31-31 01
e-mail: service.nc-support@heidenhain.de
NC programming { +49 (8669) 31-31 03
e-mail: service.nc-pgm@heidenhain.de
PLC programming { +49 (8669) 31-31 02
e-mail: service.plc@heidenhain.de
Lathe controls { +49 (711) 95 28 03-0
e-mail: service.hsf@heidenhain.de
www.heidenhain.de
HEIDENHAIN (G.B.) Limited
200 London Road, Burgess Hill West Sussex RH15 9RD, Great Britain
{ (0 1444) 2477 11 | (0 1444) 870024
350 267-23 · SW 349 797-04 · 6 · 5/2003 · E · Printed in Germany · Subject to change without notice
65
Loading...