Leica M500 OH3 User manual

4.7 (3)

Leica M500 OH3

Assembly instructions

10 711 895 – Version -

Leica M500 OH3 / Ref. 10 711 895 / Version -

 

Overview of chapters

Contents

2

General safety measures

4

Preparations

6

Assembly

8

Disassembly

19

Checklist

19

Starting up

20

Delivery report

21

Leica M500 OH3 / Ref. 10 711 895 / Version -

1

Contents

Page

General safety measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Intended use of instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Directions for the person responsible for the instrument . . 4 Directions for the user . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Symbols in these assembly instructions . . . . . . . . . . . . . . . . . 4 Goods inward check/delivery inspection . . . . . . . . . . . . . . . . 5 Safety directions in these assembly instructions . . . . . . . . . 5

Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Unpacking the transport container . . . . . . . . . . . . . . . . . . . . . 6

Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Mounting the OH3 stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Mounting the microscope carrier . . . . . . . . . . . . . . . . . . . . . 10 Mounting the weights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Mounting the optics carrier . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Mounting handgrips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Cabling the microscope carrier . . . . . . . . . . . . . . . . . . . . . . . 14 Mounting the removable wheel counterweight

and cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Checking and setting the voltage . . . . . . . . . . . . . . . . . . . . . 15 Mounting the fibre-optic cable . . . . . . . . . . . . . . . . . . . . . . . . 17 Mounting lamps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Disassembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Starting up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Balancing out the Leica M500 OH3 . . . . . . . . . . . . . . . . . . . . 20 Checking the mobility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Carrying out functional checks. . . . . . . . . . . . . . . . . . . . . . . . 20 Inspection and acceptance sequence . . . . . . . . . . . . . . . . . 20

Delivery report. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

2

Leica M500 OH3 / Ref. 10 711 895 / Version -

Leica M500 OH3 / Ref. 10 711 895 / Version -

3

General safety measures

These assembly instructions contain important safety instructions as well as information on setting up the instrument.

Before attempting to set up or use the product read carefully through the assembly instructions.

Intended use of instrument

The Leica M500 OH3 surgical microscope system is an optical instrument which uses magnification and illumination to improve the display of objects.

It can be used for observation, documentation as well as in human and veterinary surgery.

The Leica M500 OH3 surgical microscope system may only be used in closed rooms and beneath a strong ceiling.

Directions for the user

If you use accessories from another manufacturer with the Leica M500 OH3 surgical microscope system, make sure that this manufacturer confirms that the combination is safe to use. Follow the instructions in the user manual in question.

Follow the instructions given in this manual.

Follow the instructions given by your employer regarding the organization of work and safety at work.

Modifications may not be carried out to the Leica M500 OH3 surgical microscope system.

Disposal

The relevant national statutory provisions shall apply to the disposal of the products.

Directions for the person responsible for the instrument

Ensure that the surgical microscope is used only by qualified personnel.

Carry out regular inspections to make sure the user is complying with safety requirements.

Brief the user thoroughly and explain the meaning of the hazard signs and safety instructions.

Allocate responsibilities for getting the instrument ready for work, for operating it and for maintaining it. Monitor compliance with this.

Do not use the Leica M500 OH3 surgical microscope system unless it is in perfect condition.

If any faults develop which might put the safety of people at risk, contact your Leica representative immediately or Leica Microsystems (Switzerland) Ltd., CH-9435 Heerbrugg.

The effect of the Leica M500 OH3 surgical microscope on other instruments has been tested as specified in EN 60 601-1-2. The system has passed the tests relating to emissions and immunity. Respect the usual precautionary and safety measures relating to electromagnetic and other forms of radiation.

Any liability for the functioning of the system will always be the liability of the owner or operator of the system should the system have been incorrectly installed by persons not so commissioned by Leica Ltd.

Symbols in these assembly instructions

The symbols used in these assembly instructions have the following meaning:

Warning

This indicates a potentially hazardous situation which could result in death or serious injury.

Caution

This indicates a potentially hazardous situation which could result in minor or moderate injury but possibly considerable material, financial or environmental damage.

Usage information: Information provided to help the user use the product technically correctly and also efficiently.

Action required: This symbol indicates that you need to perform a specific action or series of actions.

4

Leica M500 OH3 / Ref. 10 711 895 / Version -

General safety measures

Goods inward check/delivery inspection

Damaged goods reduce the safety level.

The use of damaged products can cause severe injury.

Check the packaging and goods for transport damage:

The transport container should be undamaged.

The delivery as listed in the order folder should be complete.

The optional accessories as listed in the order folder should be complete.

The user manual should be present.

If damage has occurred, have the safety of the goods inspected by the Leica service workshop before using them.

Safety directions in these assembly instructions

Caution

Damage to microscope if the screws are tightened too excessively!

Avoid the use of excessive force when tightening all screws.

Set the torque wrench to the torque specified. Apply the wrench until this torque is reached (signalled by a clicking noise). Do not tighten further.

Warning

Risk of severe personal injury when the stand moves in an uncontrolled manner!

Lock the foot brakes during the mounting process.

Caution

Risk of injury through falling wheel counterweight!

During the change, ensure that your feet are not under the wheel counterweight.

Caution

Risk of injury by falling wheel counterweight or cover!

During the change, ensure that your feet are not under the wheel counterweight or the cover.

Warning

Risk of injury by movements of the microscope during balancing out!

Do not remain in the immediate vicinity of the microscope during balancing out.

Caution

Damage to the microscope if cables get caught!

Make sure that cables do not get caught between the microscope carrier and the cover and the microscope carrier.

Leica M500 OH3 / Ref. 10 711 895 / Version -

5

Leica M500 OH3 User manual

Preparations

Unpacking the transport container

Remove all the screws (hexagon, width across flats 10) from

 

 

 

the side wooden strips of the carton package.

Contents

Remove the carton package of the transport container

upwards.

 

 

 

 

 

 

Do not remove the packaging material of the stand

 

 

 

1

before it is no longer positioned on the pallet.

Remove the fixing planks (5) at the base of the stand.

 

 

 

 

 

 

 

 

 

 

2

4

3

5

Place the ramp (1) before the pallet and travel the stand down carefully.

Remove the packaging material of the stand.

1Ramp

2Stand OH3

3Crate with fitting parts - Microscope carrier

- Handles

1

 

-Tool set

-Xenon illumination unit

-Diverse fitting parts

4 Crate with weights

Leica M500 OH3 / Ref. 10 711 895 / Version -