Embedded Power for Business-Critical Continuity
LPS170-M Series
Medical
175 Watts
Total Power: |
100 - 175 Watts |
Input Voltage: |
85-264 VAC |
|
120-300 VDC |
# of Outputs: |
Single |
Special Features
•Medical safety approvals
•Active power factor correction
•IEC EN61000-3-2 compliance
•Wide Range Adjustable output Remote sense on main output
•Single wire current sharing
•Power fail and remote inhibit
•Built-in EMI filter
•Low output ripple
•Overvoltage protection
•Overload protection
•Thermal overload protection
•DC power good
•5 V standby output
•12 V Aux output
•Optional cover (-C suffix)
Safety
VDE |
0750/EN60601-1 (IEC601) |
UL |
UL2601 |
CSA |
CSA 22.2 No. 601.1 |
CE |
Mark (LVD) |
Rev. 08.07.07
LPS170-M Series
1 of 3
Electrical Specifications
Input
Input range |
85-264 VAC; 120-300 VDC |
||
Frequency |
47-67 Hz |
||
Inrush current |
38 A max, cold start @ 25°C |
||
Efficiency |
75% typical at full load |
||
EMI filter |
FCC Class B conducted |
||
|
|
|
CISPR 22 Class B conducted |
|
|
|
EN55022 Class B conducted |
|
|
|
VDE 0878 PT3 Class B conducted |
Power Factor |
0.99 typical |
||
Safety ground |
<250 μA @ 50/60 Hz, 264 VAC inputS |
||
leakage current |
|
||
Output |
|
||
Maximum power |
110 W convection (75 W with cover) |
||
|
|
|
175 W with 30 CFM forced air |
|
|
|
(130 W with cover) |
Adjustment range |
2:1 wide ratio minimum |
||
Standby outputs |
5 V @ 2 A regulated ±5% |
||
Hold-up time |
20 ms @175 W load at nominal line |
||
Overload protection |
Short circuit protection on all outputs. |
||
|
|
|
Case overload protected @ 110-145% above peak rating |
Overvoltage protection |
10% to 40% above nominal output |
||
Aux output |
12 V @ 1 A -5 %, +10% |
||
|
|
|
|
|
|
|
|
|
|
|
|