Welch Allyn CP-150 Service manual

4.3 (3)
CP 150 12-lead resting electrocardiograph
Service manual
© 2013 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is permitted to copy this publication, for internal distribution only, from the media provided by Welch Allyn. No other use, reproduction, or distribution of this publication, or any part of it, is permitted without written permission from Welch Allyn. Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual.
Welch Allyn is a registered trademark of Welch Allyn.
Software in this product is copyright 2013 Welch Allyn. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated with this instrument as intended in the operation of the product in which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled, or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title, and ownership of the software remain with Welch Allyn.
Welch Allyn Technical Support:
http://www.welchallyn.com/about/company/locations.htm
105323 (printed copy)
Material Number 719489
DIR 80018104 Ver. B
Welch Allyn, Inc. 4341 State Street Road Skaneateles Falls, NY 13153-0220 USA
www.welchallyn.com

Contents

Symbols ................................................................................................... 1
Safety ....................................................................................................... 5
Overview ................................................................................................ 11
iii
General warnings ................................................................................................. 5
General cautions .................................................................................................. 7
General safety considerations .............................................................................. 8
Electrostatic discharge (ESD) ............................................................................... 9
Purpose and scope ............................................................................................ 11
Technical support services ................................................................................. 11
Service loaners ................................................................................................... 12
Service options .................................................................................................. 12
Repairs ............................................................................................................... 13
Configuration options ......................................................................................... 14
The Welch Allyn Service Tool ............................................................................ 15
Returning products ............................................................................................ 15
Service menu ......................................................................................... 19
Access the Service screen ................................................................................. 19
Performing the full-functional tests .................................................................... 19
Manage tests and other files ............................................................................. 27
Troubleshooting .................................................................................... 29
Symptoms and solutions ................................................................................... 29
Printer ................................................................................................................ 31
User interface .................................................................................................... 33
Post test error messages ................................................................................... 35
Disassembly, repair, and reassembly .................................................. 37
Required tools and equipment ........................................................................... 38
Power down the electrocardiograph .................................................................. 40
Remove the battery ........................................................................................... 40
Separate the top housing assembly from the bottom housing assembly ......... 41
Electrical safety testing ......................................................................... 69
Software updates process .................................................................... 71
Update process .................................................................................................. 71
Export and import of the device configuration ................................................... 72
Contents CP 150 12-lead resting electrocardiograph
iv
Field replaceable units .......................................................................... 75
Housing top assembly ...................................................................................... 75
Housing base assembly ..................................................................................... 79
Printer 8 inch assembly tray ............................................................................... 82
Bottom assembly ............................................................................................... 86
Partners in Care service and support agreements ............................................. 89
Service and repair training .................................................................................. 89
Appendices ............................................................................................ 91
Decontamination and cleaning requirements .................................................... 91
Interconnect diagram ......................................................................................... 92

Symbols

Documentation symbols

1

Power symbols

WARNING The warning statements in this manual identify conditions or practices that could lead to illness, injury, or death.
Consult operating instructions.
Power on/standby
(LED off) No AC power is present The battery is absent or faulty
(LED green) AC power is present, and battery is charged. (LED amber) AC power is present, and battery is charging.
Caution The caution statements in this manual identify conditions or practices that could result in damage to the equipment or other property, or loss of data. This definition applies to both yellow and black and white symbols.
The battery is charged to the indicated level
Alternating current Battery is charging
Dangerous voltage Battery door
Fuse Rechargeable battery
Symbols CP 150 12-lead resting electrocardiograph
2
Protective earth Nonionizing electromagnetic radiation
Equipotential terminal AC input power

Connectivity symbols

USB Ethernet (RJ-45 )

Shipping, storing, and environment symbols

This end up Keep dry
Fragile Relative humidity limit
Temperature limits Atmospheric air-pressure limits
Separate the battery from other disposables for recycling
Separate the device from other disposables for recycling.
See www.welchallyn.com/weee.
Lithium ion battery
Recycle
China RoHS (restriction of hazardous substances) symbols for control of pollution caused by electronic information products. 5-year environment-friendly use period (EFUP) for batteries. 10-year EFUP for the device. For details, see the accompanying documentation.
Service manual Symbols 3

Miscellaneous symbols

Manufacturer Defibrillation-proof Type CF applied parts
Reference/Model number Serial number
Reorder/Catalog number
Intertek ETL listed Do not reuse
Batch code

Electromagnetic compatibility (EMC) information

For information about electromagnetic compatibility (EMC), see the Welch Allyn website.
http://www.welchallyn.com/apps/products/product_category.jsp?catcode=CARDIO
Symbols CP 150 12-lead resting electrocardiograph
4

Safety

5
All users of the device must read and understand all safety information presented in this manual before using or repairing the device.
United States federal law restricts this device to sale, distribution, or use by or on the order of a licensed medical practitioner.

General warnings

Warnings indicate conditions or practices that could lead to illness, injury, or death.

Warnings related to the environment

WARNING The power cord is considered the disconnect device for isolating this equipment from supply mains. Do not position the equipment so that it is difficult to reach or disconnect.
WARNING To avoid a possible explosion, do not use the electrocardiograph in the presence of flammable anesthetics: mixtures containing air, oxygen, or nitrous oxide.
WARNING When transporting the electrocardiograph on a cart, tuck the patient cable away to keep them clear of the wheels and to minimize trip hazards.

Warnings related to accessories and other equipment

WARNING To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
WARNING For operator and patient safety, peripheral equipment and accessories that can come in direct patient contact must be in compliance with all appropriate safety, EMC, and regulatory requirements.
WARNING All signal input and output (I/O) connectors are intended for connection of only devices complying with IEC 60601-1, or other IEC standards (for example, IEC 60950), as appropriate to the device. Connecting additional devices to the electrocardiograph might increase chassis or patient leakage currents.
Safety CP 150 12-lead resting electrocardiograph
6
WARNING The electrocardiograph has not been designed for use with high-frequency (HF) surgical equipment and does not protect against hazards to the patient.
WARNING Defective batteries can damage the electrocardiograph. Visually inspect the battery at least monthly, if the battery shows any signs of damage or cracking, it must be replaced immediately and only with a battery approved by Welch Allyn.
WARNING Improper disposal of batteries may create an explosion or contamination hazard. Never dispose of batteries in refuse containers. Always recycle batteries according to local regulations.
WARNING All signal input and output (SIP/SOP) connectors should not be contacted by patient directly and indirectly through the user during operation.

Warnings related to using the electrocardiograph

WARNING No modification of this equipment is allowed.
WARNING This device captures and presents data reflecting a patient’s
physiological condition. When reviewed by a trained physician or clinician, this data can be useful in determining a diagnosis. However, the data should not be used as a sole means for determining a patient’s diagnosis or prescribing treatment.
WARNING To provide CF protection use only accessories approved by Welch Allyn. Visit www.welchallyn.com. The use of any other accessories can result in inaccurate patient data, can damage the equipment, and can void your product warranty.
WARNING To avoid serious injury or death, take these precautions during patient defibrillation:
Avoid contact with the electrocardiograph, patient cable, and patient.
Verify that the patient leads are properly connected.
Place defibrillator paddles properly in relation to electrodes.
After defibrillation, pull each patient lead out of the patient cable and inspect the tips for charring (black carbon marks). If there is any charring, the patient cable and individual leads must be replaced. If there is no charring, fully reinsert the leads into the patient cable. (Charring can occur only if a lead is not fully inserted into the patient cable before defibrillation.)
WARNING To prevent the spread of infection, take these precautions:
Dispose of single-use components (for example, electrodes) after using them once.
Regularly clean all components that come in contact with patients.
Avoid ECG testing for patients with open, infectious sores.
WARNING Avoid positioning any leads or cables so that they could easily trip someone or become wrapped around a patient’s neck.
Service manual Safety 7
WARNING To ensure safe use of the device, follow the documented maintenance procedures.
WARNING Only qualified service personnel should attempt to repair the electrocardiograph. In case of a malfunction, call Technical Support.
WARNING Do not perform ST segment analysis on the ECG screen display, since these ECG representations are scaled. Make manual measurements of ECG intervals and magnitudes on printed ECG reports only.
WARNING To maintain diagnostic accuracy and to comply with IEC 60601-02-51 and IEC 60601-02-25, do not scale (re size) when sending a saved ECG to an external printer.
WARNING To avoid injury, do not touch the print head immediately after printing. It might be hot.

General cautions

Cautions indicate conditions or practices that could damage the equipment or other property.
CAUTION When removing the electrocardiograph from storage, allow it to thermally stabilize to surrounding environmental conditions before using it.
CAUTION To prevent possible damage, do not use sharp or hard objects to press the touch screen or the buttons. Only use fingertips.
CAUTION Do not expose the patient cable to strong ultraviolet radiation.
CAUTION Do not pull or stretch the patient cable. Doing so could result in
mechanical or electrical failures. Form the patient cable into a loose loop before storing.
CAUTION Avoid positioning the patient cable where it might get pinched, stretched, or stepped on. Otherwise, measurements might no longer be accurate, and repair might be necessary.
CAUTION Using the equipotential terminal for anything but grounding purposes may contribute to damage of the device.
CAUTION Use only parts and accessories, including thermal paper, that were supplied with the device and available through Welch Allyn. The use of accessories other than those specified may result in degraded performance or unsafe use of this device.
CAUTION Portable and mobile RF communications equipment can affect the performance of the electrocardiograph.
CAUTION The electrocardiograph meets the Class A requirements of IEC 60601-1-2 regarding incidental emission of radio frequency interference. As such it is suitable for use in commercial grade electrical environments. If the electrocardiograph is used in residential grade electrical environments and you experience incidental interference with other equipment that uses radio frequency signals to operate, minimize the interference.
Safety CP 150 12-lead resting electrocardiograph
8
CAUTION Other medical equipment—including but not limited to defibrillators, ultrasound machines, pacemakers, and other stimulators— may be used simultaneously with the electrocardiograph. However, such devices may disturb the electrocardiograph signal.
CAUTION The power cord must be disconnected from AC power before cleaning, maintaining, transporting, or servicing.
CAUTION The requirements of AAMI EC11, Section 3.2.7.2, Frequency and Impulse Response, for an impulse triangle waveform may be impacted by up to 5 milliseconds of small amplitude dampened ringing immediately after the impulse when the muscle filter (35 Hz) is turned on or a small amplitude offset when the baseline filter (0.5 Hz) is turned on. These filters, in any combination turned on or off, meet the AAMI requirements. Measurements performed by the optional interpretation algorithm are unaffected by any filter selections.
Note The entire patient cable, up to and including the electrodes are considered
to be an Applied Part.

General safety considerations

To ensure patient safety, use only accessories recommended or supplied by Welch Allyn. Always use accessories according to your facility’s standards and according to the manufacturer’s recommendations and instructions. Always follow the manufacturer’s directions for use.
Welch Allyn recommends that only Welch Allyn service personnel or an authorized repair center perform warranty service. Performing unauthorized service on a device that is within warranty may void the warranty.
Service manual Safety 9

Electrostatic discharge (ESD)

CAUTION Electrostatic discharge (ESD) can damage or destroy electronic components. Handle static-sensitive components only at static-safe workstation.
CAUTION Assume that all electrical and electronic components of the device are static-sensitive.
Electrostatic discharge is a sudden current flowing from a charged object to another object or to ground. Electrostatic charges can accumulate on common items such as foam drinking cups, cellophane tape, synthetic clothing, untreated foam packaging material, and untreated plastic bags and work folders, to name only a few.
Electronic components and assemblies, if not properly protected against ESD, can be permanently damaged or destroyed when near or in contact with electrostatically charged objects. When you handle components or assemblies that are not in protective bags and you are not sure whether they are static-sensitive, assume that they are static­sensitive and handle them accordingly.
Perform all service procedures in a static-protected environment. Always use techniques and equipment designed to protect personnel and equipment from electrostatic discharge.
Remove static-sensitive components and assemblies from their static-shielding bags only at static-safe workstations—a properly grounded table and grounded floor mat— and only when you are wearing a grounded wrist strap (with a resistor of at least 1 megohm in series) or other grounding device.
Use only grounded tools when inserting, adjusting, or removing static-sensitive components and assemblies.
Remove or insert static-sensitive components and assemblies only with device power turned off.
Insert and seal static-sensitive components and assemblies into their original static­shielding bags before removing them from static-protected areas.
Always test your ground strap, bench mat, conductive work surface, and ground cord before removing components and assemblies from their protective bags and before beginning any disassembly or assembly procedures.
Safety CP 150 12-lead resting electrocardiograph
10

Overview

Purpose and scope

The purpose of this manual is to assist with some common troubleshooting scenarios you may encounter and to explain the export, import, update and upgrade process. It is intended for use only by trained and qualified service personnel.
11
Corrective service is supported to the level of field-replaceable units. These include circuit-board assemblies and some subassemblies, case parts, and other parts.
Find instructions for functional testing and performance verification in the Welch Allyn Service Tool help files (http://www.welchallyn.com/promotions/services/ serviceTool.htm.)
This manual applies only to this device. For servicing of any other device, see the service manual for the specific device.
Service work not described in this manual must be performed by qualified service personnel at the factory or at an authorized Welch Allyn service center.

Related documents

When using this manual, refer to the following:
Welch Allyn CP 150 12-lead resting electrocardiograph Directions for use (part number 105313)
Welch Allyn Service Tool Install guide (part number 103820)
Welch Allyn website: www.welchallyn.com
WARNING When performing a service procedure, follow the instructions exactly as presented in this manual. Failure to do so could damage the device, invalidate the product warranty, and lead to serious personal injury.
CAUTION No component-level repair of circuit boards and subassemblies is supported. Use only the repair procedures described in this manual.

Technical support services

Welch Allyn offers the following technical support services:
Telephone support
Overview CP 150 12-lead resting electrocardiograph
12
Loaner equipment
Service Agreements
Service training
Replacement service parts
Product service
For information on any of these services, contact the Welch Allyn Service Center nearest you.

Service loaners

For warranty or non-warranty repairs not covered under a support agreement, loaners are available for a nominal charge, subject to availability. Payment is required prior to shipment for all loaners not covered under a support agreement. The loaner fee can be found on the Welch Allyn loaner price list.
Welch Allyn Service Centers that provide repair service for this product can, on request, loan a device for use while the device is being repaired. Loaned devices are provided free of charge for products repaired while under a support agreement that includes a free loaner provision.
Loaner equipment for the individual component modules is not available.

Service options

Partners in Care service agreements

While product warranties provide basic assurance of Welch Allyn hardware quality, they may not include the full range of services and support you need. Welch Allyn offers premium service and support through our Partners in Care program. Whether you service your own devices and require a minimum of support or rely on us to service your device, Welch Allyn provides a program that will meet your needs. For more information visit our web site at www.welchallyn.com or call your sales representative.

Warranty service

All repairs on products under warranty must be performed or approved by Welch Allyn. Refer all warranty service to Welch Allyn Product Service or another authorized Welch Allyn Service Center. Obtain a Return Material Authorization (RMA) number for all returns to Welch Allyn Product Service.
CAUTION Unauthorized repairs will void the product warranty.

Non-warranty service

Welch Allyn product service and authorized service centers support non-warranty repairs. Contact any Welch Allyn regional service center for pricing and service options.
Welch Allyn offers modular repair parts for sale to support non-warranty service. This service must be performed only by qualified end-user biomedical/clinical engineers using this service manual.
Service manual Overview 13

Repairs

A Welch Allyn Service Center must perform all repairs on products under warranty, unless you have purchased a Welch Allyn support agreement allowing you to service the device while under warranty.
CAUTION Unauthorized repairs will void the product warranty.
Qualified service personnel or a Welch Allyn Service Center should repair products out of warranty.
If you are advised to return a product to Welch Allyn for repair or routine maintenance, schedule the repair with the service center nearest you.
Welch Allyn Technical Support
If you have a problem with the device that you cannot resolve, call the Welch Allyn Technical Support Center nearest you for assistance. A representative will assist you in troubleshooting the problem and will make every effort to solve the problem over the phone, avoiding a potential unnecessary return.
If your product requires warranty or non-warranty repair service, a Welch Allyn Technical Support representative will record all necessary information to issue an RMA number. The support representative will provide you with the address of the Welch Allyn Service Center to send your device to.
An RMA number must be obtained prior to any return. Be sure to note this number on the outside of your shipping box and include a copy of the RMA in the box.
Returns without an RMA number will not be accepted for delivery.
Technical support is available during local business hours.
Overview CP 150 12-lead resting electrocardiograph
14

Configuration options

Model Accessories Language Power cord
CP150 1 - AHA, disposable EN - English 2 - Europe
A - Interpretation 2 - IEC, disposable FR - French 3 - Israel
None 3 - AHA, reusable DE - German 4 - UK
4 - IEC, reusable ES - Spanish 5 - Switzerland
NL - Dutch 66 - Australia
BP - Brazilian Portuguese
PT - Portuguese B - North America
ZH - Simplified Chinese
RU - Russian G – Argentina
NO - Norwegian N – India/UAE
SV - Swedish Z - Brazil
Examples: CP150-1ENB, CP150A-1ENB, CP150A-4DE5
7 - S. Africa
C - China
Service manual Overview 15

The Welch Allyn Service Tool

The Welch Allyn Service Tool is available in the Silver edition. Download from Welch Allyn website.
Clinicians and technical service personnel can use the service tool to manage and maintain supported Welch Allyn products. You can use the service tool to do the following:
Review device information. When connected to the device, the service tool lists installed modules, installed firmware and hardware versions, warranty and repair information, status, and usage history.
Receive notifications when periodic maintenance is needed. The service tool can help you manage and maintain your entire inventory of supported Welch Allyn products. Through the remote service function, the service tool can connect to Welch Allyn Customer Service. With this functionality you can automatically receive firmware updates and feature upgrades for your supported products, including software upgrades for the service tool.
Install updates. The service tool can read the firmware version of the CP 150 and check for available updates or upgrades.
Install upgrades. An interpretation option is available for purchase. The CP 150 basic can be upgraded to a CP 150A (with the Interpretation option) through the Welch Allyn Service Tool.
Create a work list. The work list provides information about service actions— referred to as work orders—that are waiting for you to perform on your maintained devices. Work orders may include periodic calibrations, upgrades, or license installations.
Schedule periodic maintenance. You can use the service tool to set the service interval for each maintained device.
View and save logs. You can download and save log files from the device for analysis to help diagnose and identify reported issues.
Create user accounts. Administrators can create user accounts and set permission levels to control access to the features, allowing one group to perform administrative tasks and another to perform service tasks. Restricting access prevents the service tool from being used to make unauthorized changes on a connected device.
Perform functional verification. The service tool can be used to test the device to ensure that it meets performance specifications.
Recover devices. In the rare case where a device can no longer boot because of corrupted firmware, the service tool can connect the device to Welch Allyn Technical Support to reinstall the firmware.
Extensible. The device accepts new plug-ins to support future Welch Allyn products.

Returning products

When returning a product to Welch Allyn for service, ensure that you have the following information:
Product name, model number, and serial number. This information may be found on the product and serial number labels on the bottom of the device.
A complete return shipping address.
A contact name and phone number.
Overview CP 150 12-lead resting electrocardiograph
16
Any special shipping instructions.
A purchase-order number or credit-card number if the product is not covered by a warranty.
A full description of the problem or service request.
1. Contact Welch Allyn and request an RMA number.
Note Welch Allyn does not accept returned products without an
RMA.
2. Ship the device to Welch Allyn, observing these packing guidelines:
a. Remove from the package the battery, all hoses, connectors, cables, sensors,
power cords, and other ancillary products and equipment, except those items that might be associated with the problem.
Recommendations for returning the Lithium Ion battery
Use ground transportation to return batteries.
If returning multiple batteries, package each battery individually.
Do not consolidate multiple batteries in a single package.
Use packaging provided by Welch Allyn or the battery manufacturer.
Do not pack a defective battery in checked or carry-on baggage if traveling by air.
Packaging
If you return the battery with the device, remove the battery, seal the battery in an antistatic plastic bag, and place the battery in the position reserved for the battery in the original shipping carton near the device.
If you return the battery separately, package the battery in the replacement battery’s plastic bag and shipping box.
If the original shipping carton or replacement battery shipping box is unavailable, consult the manufacturer website for information regarding shipping lithium ion batteries:
www.nexergy.com/lithium-shipping.htm
WARNING Safety risk. Do not ship any battery that has been physically damaged or shows signs of leakage unless you receive specific instructions which meet the requirements for the shipment of Lithium batteries. Dispose of damaged or leaking batteries in an environmentally safe manner consistent with local regulations.
Note In the United States, the applicable regulations can be
found in the Code of Federal Regulations (CFR). Refer to 49 CFR 173.185 for shipping lithium batteries by air or ground. Use 49 CFR 172.102 sections 29, 188, 189, A54, A55, A100, A101, A103, and A104 to find the special provisions for shipping lithium batteries.
b. Clean the device.
Service manual Overview 17
Note To ensure safe receipt of your device by the service
center and to expedite processing and return of the device to you, thoroughly clean all residues from the device before you ship it to Welch Allyn. For decontamination and cleaning requirements, see “Decontamination and cleaning” in the appendices. If a returned device is found to be contaminated with bodily fluids, it will be returned at the owner’s expense. United States federal regulations prohibit the processing of any device contaminated with blood-borne pathogens. Welch Allyn thoroughly cleans all returned devices on receipt, but any device that cannot be adequately cleaned cannot be repaired.
c. Put the device, enclosed in a plastic bag with a packing list, into the original
shipping carton with the original packing materials or into another appropriate shipping carton.
d. Write the Welch Allyn RMA number with the Welch Allyn address on the outside
of the shipping carton.
Overview CP 150 12-lead resting electrocardiograph
18

Service menu

Access the Service screen

Note The service menu ID and password are based on a telephone keypad.
ID: 7378423 (=SERVICE)
PASSWORD: 6676737 (=NORMSER)
19
1. From the ECG home tab, touch the Settings tab. The ECG tab and the vertical ECG
configuration tab appear.
2. Touch the Service tab.
3. Enter 7378423 as the User ID code and touch the OK button.
4. Enter 6676737 as the Password and touch the OK button.
5. Touch the OK button again.
6. When you are done, touch .
The ECG home tab appears.

Performing the full-functional tests

Complete the full-functional CP 150 tests to verify device functionality.
Note If any of the tests fail, contact Technical Support.
The following off-the-shelf tools are required to complete the functional verification tests:
Standard ECG Patient Simulator (example: Model 430B, Medi Cal Instrument, Inc. or equivalent)
USB Cable (Type A (M) connector to mini-B (M) connector)
Standard USB Thumb drive (256M to 64Gb, recommend 2 Gb)
2 RJ45 Ethernet cables (less than 3 M in length)
With the Service tab open, touch the CP 150 button to perform a hardware test.
Service menu CP 150 12-lead resting electrocardiograph
20

Battery test

Note Ensure that the battery is installed and has been charged for a minimum of
four hours before preforming this test.
1. If the Battery remaining field shows 70% or less, replace the battery.

Audio test

2. Touch the (Next) button to perform the next functional test.
1. Touch the
2. Touch the Yes button if you are able to hear the sound, or touch the No button if you are not able to hear the sound.
3. Touch the (Next) button to perform the next functional test.
button to play the audio sample.
Service manual Service menu 21

USB host test

Note Welch Allyn has not validated specific USB flash drives. Use only 64Gb or
less USB flash drives.
1. Insert a USB flash drive into each of the four ports on the back of the device and follow the instructions on the screen to test each of the ports.
2. Touch the USB host test button to test each port from left to right.
3. Touch the (Next) button to perform the next functional test.

USB client test

1. Connect a USB type mini-B connector to the device client USB port and a type A connector to a free USB port on the PC. Verify that the PC is able to recognize CP150 device.
2. Touch the Yes button to confirm that your PC recognized the CP 150 device, or touch the No button if your PC did not recognized the CP 150 device.
3. Touch the (Next) button to perform the next functional test.
22
Service menu CP 150 12-lead resting electrocardiograph

Keypad test

1. Press the with a (check mark) and "fail" is indicated with an X.
button on the device to check the On/Off keypad. "Pass" is indicated
2. Touch the (Next) button to perform the next functional test.
Service manual Service menu 23

ECG lead test

1. Connect a set of known good Patient Cables, with leads, to an ECG simulator.
Note Any yellow dots on the lead-status screen indicate
unattached or poorly attached leads.
2. Verify the lead placement on the screen by observing that all leads appear green.
3. Touch the (Next) button to perform the next functional test.
Service menu CP 150 12-lead resting electrocardiograph
24

Printer pattern test

1. Touch the Print button to print a test page.
2. Use the specifications in the Print verification checklist table and the Printer pattern test and check point graphic to verify the correct operation of the device printer by comparing the pattern printed on the test page.
Print verification checklist
Point Description
Point A
Point B
Point C Verify that the line is straight and without breaks from top to bottom
Point D
Point E Verify the print is readable
Point F Verify line is straight and measures 100 mm +/ -5 mm
Verify the distance of the arc on the circle from the top edge of the paper is 11 mm +/- 3 mm
Verify the distance of the arc on the circle from the left edge of the paper is 50 mm +/- 5 mm
Verify that the thick black line is uniform in blackness without breaks from top to bottom
Service manual Service menu 25
Printer pattern test and check point
3. Touch the Yes button if the test page printed correctly, or touch the No button if the test page did not print correctly.
4. Touch the (Next) button to perform the next functional test.

Printer speed test

1. Touch each of the 10 mm/s, 25 mm/s, and 50 mm/s buttons to print a test page.
2. Use the speed test print page to verify the correct operation of the device printer by comparing the pattern printed on the test page to point F in the Printer speed test and check point graphic. The line should be a straight line that measures 100 mm (+/-5 mm).
Printer speed test and check point
3. Touch the Yes button if the test page printed correctly, or touch the No button if the test page did not print correctly.
4. Touch the (Next) button to perform the next functional test.
26
Service menu CP 150 12-lead resting electrocardiograph

Ping test

1. Use an ethernet cable to connect the device to a known network and setup the CP 150 DHCP. (Refer to the CP 150 DFU for setup instructions.) Under the Host address, enter the IP of a PC connected to the network. Touch the Ping test button to perform the test.

Test results

2. Touch the (Next) button to exit the functional test.
Note If any of the tests fail, contact Technical Support.
Loading...
+ 68 hidden pages