Technical data:

Display:

LCD-Display.

 

 

 

 

Display resolution:

0.1 °C / °F.

 

 

 

 

Displayed temperature range:

20.0 °C - 42.2 °C

(68

.0 °F - 108.0 °F).

 

Work place temperature range:

10.0 °C - 40.0 °C

(5

0.0 °F - 104.0 °F).

 

Work place humidity:

up to 95 %.

 

 

 

 

In the range between:

35.5 °C - 42.0 °C

(95.9 °F

- 107.6 °F).

 

the accuracy of allowable

 

 

 

 

 

tolerance with probe cover is:

± 0.2 °C

(± 0.4 °F

).

 

Outside the range from:

35.5 °C - 42.0 °C

(95.9 °F - 107.6 °F).

 

the accuracy of allowable

 

 

 

 

 

tolerance with probe cover is:

± 0.3 °C

(± 0.5 °F

).

 

Probe tip:

heated up to approximately 37°C.

 

Power supply:

rechargeable Braun battery-pack with

 

 

base-station and smart plug

 

 

 

or 2 1.5 V type AA (LR 6) batteries.

 

Smart plug:

100-240 V - input / 12 V – output.

 

Base-station:

12 V - input / approx. 2.4 V – output.

 

Battery-life time:

3 month (with *) / 1000 measurements.

 

Rechargeable battery lifetime:

3 month / 500 measurements per charge.

 

Rechargeable battery charging:

Thermometer stored in basis-station.

 

Rechargeable battery charging time:

14 hours.

 

 

 

 

Base-station function indications:

1 LED, green for power connection

 

 

1 LED, yellow for charging indication.

Electronic anti-theft *:

for Pro 4000 appliances only and with the base-station in use.

Automatic turn off:

If no button is pressed within 60 seconds, the

 

appliance will turn off automatically.

Manual turn off:

Push

- button for >3 seconds.

Welch Allyn

701627 REV. G

 

Page 1 of 9

Welch

Allyn

Service Documentation

Part Number: 701627

Rev. G

 

 

 

 

 

General hints

 

 

Production code

Near the battery storage

LOT 4042.

 

 

4 = Year / 042 = Day.

Attention:

Do not dispose of empty batteries or defective rechargeable batteries in household

 

waste. Disposal can take place at appropriate collection sites provided in your country

Hint:

Type BF equipment

 

This appliance conforms to the following standards: DIN EN 60601-1

 

<<Medical electrical equipment>>-Part 1: General requirements for safety.

Hint:

This product conforms to the provisions of the EC directive 93/42/EEC

 

(Medical Device Directive).

 

0297