Welch Allyn CP 100 User Manual

0 (0)

CP 10012-Lead Resting

Electrocardiograph

Directions for Use

ii

Welch Allyn CP 100 Electrocardiograph

Copyright 2008, Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate, in any form, this manual or any part thereof without permission from Welch Allyn.

Caution: Federal US law restricts sale of the device identified in this manual to, or on the order of, a licensed physician.

Welch Allyn assumes no responsibility for any injury, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or indications for use published in this manual.

Welch Allyn is a registered trademark of Welch Allyn, Inc., and CP 100 and CardioPerfect are trademarks of Welch Allyn, Inc.

SD is a trademark of Toshiba.

Software in this product is copyright Welch Allyn or its vendors. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated within this instrument as intended in the operation of the product in which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title and ownership of the software remains with Welch Allyn or its vendors.

For information about any Welch Allyn product, please call Welch Allyn Technical Support:

USA

1

800 535 6663

Australia

+ 61 29 638 3000

 

+ 1 315 685 4560

 

 

Canada

1

800 561 8797

China

+ 86 216 327 9631

European Call Center

+ 353 46 906 7790

France

+ 331 6009 3366

Germany

+ 49 747 792 7186

Japan

+ 81 33 219 0071

Latin America

+ 1 305 669 9003

Netherlands

+ 31 15 750 5000

Singapore

+ 65 6419 8100

South Africa

+ 27 11 777 7555

United Kingdom

+ 44 207 365 6780

Sweden

+ 46 85 853 6551

Reorder Number (multi-language CD): 401150

Mat. Number (manual only): 708794, Ver: F

Welch Allyn

Welch Allyn LTD.

 

 

 

4341 State Street Road, PO Box 220

Navan Business Park

 

 

 

 

 

 

Skaneateles Falls, NY 13153-0220

Dublin Road

 

Navan, County Meath, Republic of Ireland

 

Tel.: 353-46-90-67700

 

Fax: 353-46-90-67755

www.welchallyn.com

 

 

 

 

Printed in USA

 

 

 

 

iii

Contents

1 - Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Product Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Standard Features & Benefits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Controls, Indicators, and Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 About the Main Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Moving Through the Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 About the Patient Cable and Leads. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Using the Electrocardiograph Safely. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

General Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 General Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Getting Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

2 - Setting Up the Electrocardiograph . . . . . . . . . . . . . . . . . . . . . . . . . 19

Inspecting the Electrocardiograph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

Connecting the Patient Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

Loading the Thermal Chart Paper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

Powering the Electrocardiograph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Verifying Proper Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

Connecting an External USB Printer (Optional). . . . . . . . . . . . . . . . . . . . . . . . . . 24

3 - Reviewing the System Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

“System Settings” Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Reviewing the Device Configuration Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Reviewing the Device Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Transferring a Configuration to Another Electrocardiograph . . . . . . . . . . . . . . . . 30

iv

Contents

Welch Allyn CP 100 Electrocardiograph

4 - Reviewing the ECG Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

“ECG Settings” Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 About Auto ECG Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Reviewing the Format Settings for Auto Reports. . . . . . . . . . . . . . . . . . . . . 35 Reviewing the Interpretation and Copy Settings for Auto Reports. . . . . . . . 36 Reviewing the Patient Data Fields Available. . . . . . . . . . . . . . . . . . . . . . . . . 37 Reviewing the Auto Report Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Reviewing the Miscellaneous ECG Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Turning the Augmented Pediatric Lead Set On and Off . . . . . . . . . . . . . . . . . . . 41

5 - Performing ECG Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

Connecting the Leads to the Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 Recording an Auto ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48 Recording a Normal Auto ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Recording a Stat Auto ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Recording a Rhythm ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 Adjusting the ECG Waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

6 - Maintaining the Electrocardiograph . . . . . . . . . . . . . . . . . . . . . . . . 57

Inspecting the Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58

Cleaning the Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58

Testing the Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59

Recharging a Fully Discharged Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

Replacing the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

Replacing the Battery (DC) Fuse. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

Replacing the AC Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

Storing the Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65

Discarding the Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65

7 - Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67

Problem-Solving Suggestions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71

Service Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72

A - Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73

B - EMC Guidance and Manufacturer’s Declarations . . . . . . . . . . . . . 75

Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79

Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81

1

1 Introduction

About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 Product Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 Indications for Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 Standard Features & Benefits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6 Controls, Indicators, and Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7 About the Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Moving Through the Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 About the Patient Cable and Leads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Using the Electrocardiograph Safely . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Getting Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

2

Chapter 1 Introduction

Welch Allyn CP 100 Electrocardiograph

About This Manual

This manual is written for clinical professionals with a working knowledge of medical procedures and terminology as required for monitoring cardiac patients.

Before using the CP 100 electrocardiograph for clinical applications—or before setting up, configuring, troubleshooting, or servicing the electrocardiograph—you must read and understand this manual and all other information accompanying the electrocardiograph and related options or accessories.

Product Overview

The Welch Allyn CP 100 electrocardiograph features a full alphanumeric keyboard, an LCD display, full-size user-programmable reports, and the ability to operate on either battery or AC power.

ECG tests sent to a memory card or removable USB storage device are compatible with the Welch Allyn CardioPerfect™ workstation, which in turn can connect with other electronic patient-information systems, such as billing and medical records.

For details, see the following sections:

“Standard Features & Benefits” on page 3

“Options” on page 5

“Specifications” on page 73

Intended Use

The Welch Allyn electrocardiography and spirometry products (subject devices) are intended for use by trained operators in health facilities. The subject devices provide the following diagnostic functions:

Acquiring and printing ECG waveforms using ECG front end modules (patient cables) and associated accessories that provide signal acquisition for up to twelve (12) leads of patient ECG waveforms through surface electrodes adhered to the body.

Using optional algorithms to generate measurements, data presentations, graphical presentations and interpretative statements on an advisory basis. These are presented for review and interpretation by the clinician based upon knowledge of the patient, the result of physical examination, the ECG tracings and other clinical findings.

Directions for Use

Chapter 1 Introduction

3

Indications for Use

The electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function.

The 12-lead ECG interpretive algorithm provides a computer-generated analysis of potential patient cardiac abnormalities which must be confirmed by a physician with other relevant clinical information.

Standard Features & Benefits

Full alphanumeric keypad

Enter patient information quickly and easily.

LCD display

Enter data and program the system easily.

Unlimited storage on removable media

Use SD™ memory cards or removable USB storage devices to save as many ECG records as you like. (Media are not included.) If both an SD memory card and a USB storage device are connected, files are sent to the SD card. For media specifications, see “Standard connectivity” on page 74.

Battery operation

Use the electrocardiograph almost anywhere. On battery power, you can print up to 100 ECGs continuously before needing to recharge.

User-definable ECG report format

Customize the report format for efficient reporting.

Removable leads for ECG patient cable

Replace leads individually if needed.

Compatibility with CardioPerfect workstation software

Store and manage data electronically by transferring records to a Welch Allyn CardioPerfect workstation via an SD memory card or removable USB storage device (media not included).

Compatibility with external printer

You can connect an external printer. For details, see “Connecting an External USB Printer (Optional)” on page 24.

4

Chapter 1 Introduction

Welch Allyn CP 100 Electrocardiograph

Pacemaker detection

If the software detects the possible presence of a pacemaker, it asks you whether the patient has a pacemaker. If you say no, interpretation (if purchased) is included in the report. If you say yes, interpretation is not included in the report, and a message indicating that a pacemaker was detected is displayed on the report.

Support for augmented pediatric lead set

The augmented pediatric lead set, an alternate placement of the precordial leads on pediatric patients, is easier to use on the small chests of infants and young children.

It provides optimal vectors for early development of the heart for the change of right-to- left ventricular dominance. For electrode placement locations, see Figure 28 on page 46.

Directions for Use

Chapter 1 Introduction

5

Options

These options are available both for initial purchases and for upgrades.

Automatic ECG interpretation

The optional MEANS interpretation algorithm, developed by the University of Rotterdam in the Netherlands, provides automatic analysis of ECG tests. For more information, see the MEANS Physicians' Manual or the PEDMEANS Physicians’ Manual on the CD that came with your electrocardiograph. The MEANS algorithm is used for adult patients 18 years and older. The PEDMEANS algorithm is used for pediatric patients from 1 day through 17 years old.

Carts

Two specially designed carts are available for convenient transport and use of the electrocardiograph, as shown here with the optional cable arm and shelf.

Figure 1. Office Cart

Figure 2. Hospital Cart

Cable arm and

Cable arm and

shelf (optional)

shelf (optional)

Free-motion

Durable, high-

plastic wheels

quality, rubber

 

wheels with locks

6

Chapter 1 Introduction

Welch Allyn CP 100 Electrocardiograph

Accessories

To order accessories, call Welch Allyn. For phone numbers, see page ii.

Item

Customer Order Number

Quantity

 

 

 

Resting tab electrodes

45008-0000

1000

Resting tab electrode adaptors

58581-0000

10

Thermal chart paper (1 case = 5 pads, 200 sheets each)

94018-0000

1 case

Welch cup

RE-ELEC-CUP

1

Limb lead clamps, IEC

RE-ELEC-CLP

4

Limb lead clamps, AHA

401432

4

 

 

 

Patient cable (Figure 11 on page 12)

 

 

AHA

400293

1

IEC

400294

1

IEC, vacuum adapter

100920

1

Vacuum system

VAC-DT100-CP

 

 

 

 

Lead wires (10 wires per set)

 

 

AHA banana

401129

1 set

IEC banana

401122

1 set

AHA pinch

401123

1 set

IEC pinch

401124

1 set

 

 

 

Battery (Figure 39 on page 61)

100660

1

Dust cover

701586

1

 

 

 

Carts

 

 

Utility cart

08265-0000

1

• Office cart (Figure 1 on page 5)

401393

1

• Hospital cart (Figure 2 on page 5)

401394

1

• Cable arm & shelf option (page 5)

401161

1

 

 

 

Interpretation upgrade option

100623

1

 

 

 

Product information

 

 

• Electrode placement wall poster

71300-0000

1

• CP 100 12-Lead Resting Electrocardiograph Directions

708794

1

 

for Use

 

 

• CP 100 product information multi-language CD

401150

1

 

 

 

 

Directions for Use

Chapter 1 Introduction

7

Controls, Indicators, and Connectors

This section describes the controls, indicators, and connectors that are part of the electrocardiograph.

Figure 3. Top

1 ! 2@ 3 # 4

$ 5% 6+ 7 - 8* 9 ( 0 )

Keyboard

See Figure 1 on page 9.

Figure 4. Back

AC power inlet

Equipotential stud

AC fuses

Welch Allyn CP 100 User Manual

8

Chapter 1 Introduction

Welch Allyn CP 100 Electrocardiograph

Figure 5. Right Side

Com port B (USB)

SD memory card

Not

Reset

 

slot

functional

button

Figure 6. Front

Com port A (for patient cable)

Figure 7. Left Side

Paper tray latch

Directions for Use

Chapter 1 Introduction

9

Table 1. Keyboard

A B C D

N

1 ! 2@ 3 # 4

$ 5% 6+ 7 - 8* 9 ( 0 )

M

E

L

 

 

 

 

 

K

 

 

 

 

 

 

J

I

H

G

F

Key

 

 

Function

 

 

A.

On/Off

 

See “Powering the Electrocardiograph” on page 22.

 

B.

Backspace

 

Deletes the character to the left of the cursor.

 

 

C.

Menu

 

See “About the Main Menu” on page 10.

 

 

D.

Auto ECG

 

Begins Auto ECGs, normal and stat. See “Recording an Auto ECG” on page 48.

E.

Rhythm ECG

 

Begins a Rhythm ECG. See “Recording a Rhythm ECG” on page 54.

F.

Navigation arrows

See “Moving Through the Menus” on page 11.

 

 

G.

Enter

 

See “Moving Through the Menus” on page 11.

 

 

H.

Stop/Cancel

 

Stops any current activity. See “Moving Through the Menus” on page 11.

I.

Space

 

Enters a space.

 

 

J.

Control

 

(Applicable only for keyboards with characters above certain keys.) To enter a character

 

 

 

above a key, hold the Ctrl key while pressing that key. To capitalize the character, hold

 

 

 

Shift + Ctrl while pressing the key.

 

 

K.

Shift

 

Capitalizes letters. Also enters symbols (%, #, etc.). To enter a symbol, hold the

 

 

 

Shift key while pressing a number key.

 

 

L.

Help

 

See “Getting Help” on page 18.

 

 

M.

Tab

 

Moves through the data-entry fields.

 

 

N.

Green LED

 

Lights up when the electrocardiograph is connected to AC power.

10

Chapter 1 Introduction

Welch Allyn CP 100 Electrocardiograph

About the Main Menu

The main menu appears when you press the Menu key .

Figure 8. Main Menu

9:17AM Oct 16 08

Main Menu

1 ECG Settings

2 System Settings

0 Exit

Submenu

Purpose

Procedure

 

 

 

ECG Settings

Review or change ECG settings: Auto

 

Report format, Rhythm Report format, and

 

so on.

System Settings

Review or change system settings: device

 

configuration, device info, user setup, and

 

so on.

See “Reviewing the ECG Settings” on page 31.

See “Reviewing the System Settings” on page 25.

Directions for Use

Chapter 1 Introduction

11

Moving Through the Menus

Figure 9. Standard Menu

9:17AM Oct 16 08

Edit Auto Report

1Format

2Interp Settings

3Patient Data

0 Previous Menu

Figure 10. Parent Menu With Submenu

9:17AM Oct 16 08

Format

1 Lead Arrangement

2 Rhythm Lead 1

3 Rhythm Lead 2

4 Rhythm Lead 3

5 Extended Measurements

6 Average Cycles

0 Previous Menu

3x4

+1R

3x4

+3R

6x2

 

12x1

6x2

50 mm/s

6x2

10 Ext.

2x6+1R

Desired Actions

Keys to Press

To move up or down a list

or

 

To open a standard menu (Figure 9)

or

or item’s number or letter

To move from parent menu to submenu on same screen (Figure 10)

To perform an action

To accept data

To check or uncheck a checkbox

To return to parent menu from submenu on same screen (Figure 10)

(To select the highlighted submenu item.)

 

or

(To make no change.)

To move back through the menus

or zero key

 

To move through data-entry fields

or

 

To return to the Ready screen from a standard menu

 

 

(Figure 9)

 

 

12

Chapter 1 Introduction

Welch Allyn CP 100 Electrocardiograph

About the Patient Cable and Leads

The patient cable processes the patient’s ECG data and transmits it to the electrocardiograph. To make handling convenient, the ten leads are arranged to point toward the appropriate parts of the body. The cable rake, which slides easily, prevents the chest leads from tangling.

Figure 11. Patient Cable and Leads

 

Chest leads

Left arm lead

 

 

 

 

Cable

Left leg lead

 

rake

Electrocardiograph connector

Right arm lead

Right leg lead

Directions for Use

Chapter 1 Introduction

13

Symbols

The symbols illustrated on the following pages may appear on the electrocardiograph, on the packaging, on the shipping container, or in this manual.

Documentation Symbols

WARNING Indicates conditions or practices that could lead to illness, injury, or death.

Caution In this manual, indicates conditions or practices that could damage the equipment or other property.

Caution On the product, means “Consult accompanying documentation.”

 

 

Shipping, Storing, and Environment Symbols

 

This end up

Keep dry

 

Fragile

Relative humidity limit

+49°C

Temperature limits

Altitude limits

 

-20°C

Certification Symbols

Meets essential requirements of European Medical Device Directive 93/42/EEC

Complies with applicable U.S. and Canadian medical safety standards

 

 

European Regulatory Manager

Australian registered importer

 

 

 

N344

 

 

 

 

14

Chapter 1 Introduction

Welch Allyn CP 100 Electrocardiograph

 

Operation Symbols

 

For details on the keys, see Figure 1 on page 9.

On/standby (off)

Help

Navigation arrows

Backspace

Enter

Shift

Auto ECG

Menu

Rhythm ECG

Stop/Cancel

SD memory card slot

Com port A

 

(for patient cable)

Alternating current

Battery charge level

Battery is charging.

Battery is charged.

Dangerous voltage

Direct current

AC fuse replacement information

Ground equipotential

T2.0A/250V

Do not reuse.

Do not dispose of this product as unsorted municipal waste. Prepare this product for reuse or separate collection as specified by Directive 2002/96/EC of the European Parliament and the Council of the European Union on Waste Electronic and Electrical Equipment (WEEE). If this product is contaminated, this directive does not apply. For more specific disposal information, see www.welchallyn.com/weee, or contact Welch Allyn Customer Service at +44 207 365 6780.

Sealed lead-acid battery

Recycle.

Defibrillation-proof Type CF applied parts. (While the electrocardiograph is safety-rated “CF” for direct cardiac contact, it is not intended to be connected directly to the patient’s heart. Only surface contact with the patient’s skin is intended.)

Com port B (USB)

Directions for Use

Chapter 1 Introduction

15

Using the Electrocardiograph Safely

Before using or servicing the electrocardiograph, you must read and understand the following safety-related information.

General Warnings

The following warning statements apply to electrocardiograph use in general. Warning statements that apply specifically to particular procedures, such as connecting the patient cable or performing an ECG test, appear in the corresponding sections of the manual.

Warning statements indicate conditions or practices that could lead to illness, injury, or death.

Warnings Related to the Environment

WARNING To ensure patient and device safety, leave 5 feet (1.5 meters) of open area around the patient.

WARNING To avoid a possible explosion, do not use the electrocardiograph in the presence of flammable anesthetics: mixtures containing air, oxygen, or nitrous oxide.

WARNING When transporting the electrocardiograph on a cart, tuck the patient cable away from the wheels so that it does not present a hazard.

Warnings Related to Accessories and Other Equipment

WARNING For operator and patient safety, peripheral equipment and accessories that can come in direct patient contact must be in compliance with all appropriate safety, EMC, and regulatory requirements. See “EMC Guidance and Manufacturer’s Declarations” on page 75.

WARNING All signal input and output (I/O) connectors are intended for connection of only devices complying with IEC 60601-1, or other IEC standards (for example, IEC 60950), as appropriate to the device. Connecting additional devices to the electrocardiograph might increase chassis or patient leakage currents. To maintain operator and patient safety, consider the requirements of IEC 60601-1-1. Measure the leakage currents to confirm that no electric shock hazard exists. In the case of a USB printer, the printer (non-medical electrical equipment) shall be situated outside the patient environment (reference IEC 60601-1-1). The printer used should be approved to the appropriate safety standard for non-medical electrical equipment (IEC 60950, or its national variants), and use of an isolation transformer is recommended. If there is a requirement for the printer to be situated within the patient environment it is the responsibility of the user to ensure that the system provides a level of safety in compliance with IEC 60601-1 and 60601-1-1.

WARNING The electrocardiograph has not been designed for use with highfrequency (HF) surgical equipment and does not protect against hazards to the patient.

16

Chapter 1 Introduction

Welch Allyn CP 100 Electrocardiograph

Warnings Related to Using the Electrocardiograph

WARNING This device captures and presents data reflecting a patient’s physiological condition. When reviewed by a trained physician or clinician, this data can be useful in determining a diagnosis. However, the data should not be used as a sole means for determining a patient’s diagnosis.

WARNING To avoid serious injury or death, take these precautions during patient defibrillation:

Avoid contact with the electrocardiograph, patient cable, and patient.

Verify that the patient leads are properly connected. See “Connecting the Patient Cable” on page 20.

Place defibrillator paddles properly in relation to electrodes.

After defibrillation, pull each patient lead out of the patient cable and inspect the tips for charring (black carbon marks). If there is any charring, the patient cable and individual leads must be replaced. If there is no charring, fully reinsert the leads into the patient cable. (Charring can occur only if a lead is not fully inserted into the patient cable before defibrillation.)

WARNING To prevent the spread of infection, take these precautions:

Dispose of single-use components (for example, electrodes) after using them once.

Regularly clean and disinfect all components that come in contact with patients. See “Cleaning the Equipment” on page 58.

Avoid ECG testing for patients with open, infectious sores.

WARNING Avoid positioning any leads or cables so that they could easily trip someone or become wrapped around a patient’s neck.

WARNING Satisfactory maintenance procedures must be implemented, or equipment failure and health hazards may result.

WARNING Only qualified service personnel should attempt to repair the electrocardiograph. In case of a malfunction, call Technical Support and precisely describe the problem. For phone numbers, see page ii.

Directions for Use

Chapter 1 Introduction

17

General Cautions

The following caution statements apply to electrocardiograph use in general. Caution statements that apply specifically to particular procedures, such as connecting the patient cable or performing an ECG test, appear in the corresponding sections of the manual.

Caution statements indicate conditions or practices that could damage the equipment or other property.

Caution When removing the electrocardiograph from storage, allow it to thermally stabilize to surrounding environmental conditions before using it.

Caution To prevent possible damage to the keypad, do not use sharp or hard objects to press keys. Only use fingertips.

Caution Do not expose the patient cable to strong ultra-violet radiation.

Caution Do not pull or stretch the patient cable. Doing so could result in mechanical or electrical failures. Form the patient cable into a loose loop before storing.

Caution Avoid positioning the patient cable where it might get pinched or stepped on. If the cable’s impedance is altered, measurements might no longer be accurate, and repair might be necessary.

Caution Using the equipotential terminal for anything but grounding purposes may contribute to damage of the device.

Caution Use only parts and accessories supplied with the device and available through Welch Allyn. The use of accessories other than those specified may result in degraded performance of this device.

Caution Portable and mobile RF communications equipment can affect the performance of the electrocardiograph.

Caution The electrocardiograph meets the Class A requirements of IEC 60601- 1-2:2000 regarding incidental emission of radio frequency interference. As such it is suitable for use in commercial grade electrical environments. If the electrocardiograph is used in residential grade electrical environments and you experience incidental interference with other equipment that uses radio frequency signals to operate, minimize the interference as described under “EMC Guidance and Manufacturer’s Declarations” on page 75.

Caution Other medical equipment—including but not limited to defibrillators, ultrasound machines, pacemakers, and other stimulators—may be used simultaneously with the electrocardiograph. However, such devices may disturb the electrocardiograph signal.

Caution The power cord must be disconnected from AC power before cleaning, maintaining, or servicing.

18

Chapter 1 Introduction

Welch Allyn CP 100 Electrocardiograph

Getting Help

You can get help with the electrocardiograph in a variety of ways beyond this manual.

Press the Help key from the Ready screen or Lead Status screen for a list of topics available to print.

Review the other information that came with the electrocardiograph. For list, see “Product information” on page 6.

Contact Welch Allyn. For phone numbers, see page ii.

19

2 Setting Up the Electrocardiograph

Inspecting the Electrocardiograph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20

Connecting the Patient Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20

Loading the Thermal Chart Paper. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21

Powering the Electrocardiograph. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22

Verifying Proper Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23

Connecting an External USB Printer (Optional) . . . . . . . . . . . . . . . . . . . . . . .24

20

Chapter 2 Setting Up the Electrocardiograph

Welch Allyn CP 100 Electrocardiograph

Inspecting the Electrocardiograph

1.Look for obvious signs of shipping damage. If you find any damage, contact Technical Support. For phone numbers, see page ii.

2.Verify that you have received all appropriate options and accessories. See “Options” on page 5 and “Accessories” on page 6.

Connecting the Patient Cable

WARNING Conductive parts of the patient cable, electrodes and associated connections of defibrillation-proof Type CF applied parts, including the neutral conductor of the patient cable and electrode, should not come into contact with other conductive parts, including earth ground.

WARNING To avoid injury to the patient or damage to the device, never plug patient leads into any other device or wall outlet.

1.Insert all leads into their proper positions, as labeled on the connectors.

Insert connectors fully so that no part of the metal ring remains exposed.

For example, see Figure 12. (To see the whole patient cable with all leads inserted, see Figure 11 on page 12.)

WARNING Failure to insert all connectors fully may result in a loss of energy being delivered to the patient during defibrillation and damage to the patient cable itself. For other warnings related to defibrillation, see page 16.

2.Plug the patient cable into the port on the front of the electrocardiograph. See Figure 13.

Figure 12. Inserting the Leads

Figure 13. Plugging in the Connector

Metal ring

Directions for Use

Chapter 2 Setting Up the Electrocardiograph

21

Loading the Thermal Chart Paper

1.Squeeze the latch. Pull the paper door to the left. See Figure 14. If any paper remains in the tray, remove it.

2.Remove the outer packaging, including the cardboard bottom, from a new pack of paper. Pull the top sheet back so that the paper’s grid side faces up and the Welch Allyn name is on the bottom of the paper.

3.Slide the paper into the tray. See Figure 15.

If humidity is high, remove up to 10 sheets so that the paper fits properly.

4.Lay the top sheet over the paper door. Push the door to the right until it clicks. See Figure 16.

Figure 14. Opening the Paper Door

Figure 15. Loading the Paper

Latch

Figure 16. Closing the Paper Door

Tear bar

Paper tray

Tips for handling thermal paper:

Store in a cool, dry, dark place.

Avoid exposure to bright light or UV sources.

Avoid exposure to solvents, adhesives, or cleaning fluids.

Do not store with vinyls, plastics, or shrink wraps.

22

Chapter 2 Setting Up the Electrocardiograph

Welch Allyn CP 100 Electrocardiograph

Powering the Electrocardiograph

The electrocardiograph can run on AC or battery power.

WARNING To ensure that electrical safety is maintained when using AC power, the device must be plugged into a hospital-grade outlet.

WARNING Where the integrity of external protective earth conductor arrangement is in doubt, use battery power.

Caution Medical electrical equipment needs special precautions regarding EMC and must be installed and used according to the information provided in “EMC Guidance and Manufacturer’s Declarations” on page 75.

To Connect to AC Power

Plug one end of the power cord into the electrocardiograph’s AC power inlet. Plug the other end into an AC outlet. The green LED on the keyboard lights up, indicating that power is connected. See Figure 17.

To Keep the Battery Charged

Leave the electrocardiograph connected to AC power whenever possible. Battery charge status is indicated on the screen by an icon: . Whenever the battery is charging and is not yet fully charged, this icon appears: . For maximum battery performance, as often as possible leave the electrocardiograph plugged in until you see the “fully charged” icon: .

When the charge gets low, the icon flashes. When the charge gets too low to operate, a warning message appears and the electrocardiograph beeps every 15 seconds for 1 minute, then it turns off.

For more, see “Recharging a Fully Discharged Battery” on page 60.

Figure 17. AC Power Inlet and Green LED

AC power inlet

On/Off key

1 ! 2@ 3 # 4 $ 5% 6+ 7 - 8* 9 ( 0 )

Green LED

Directions for Use

Chapter 2 Setting Up the Electrocardiograph

23

To Turn the Electrocardiograph On

Press .

To Turn the Electrocardiograph Off

Press and hold.

Note If Power-Save is enabled, the electrocardiograph turns off automatically after several idle minutes. To learn how to enable or disable Power-Save, see “Reviewing the Device Configuration Settings” on page 27.

Verifying Proper Operation

Once your electrocardiograph is set up, verify proper operation by using an ECG simulator to acquire and print a standard 12-lead ECG of known amplitude. See Step 2 on page 59.

Note As part of your initial set-up, you may want to adjust the display contrast. To learn how, see “Reviewing the Device Configuration Settings” on page 27.

You may also want to change other software settings, as described in the following chapters:

“Reviewing the System Settings” on page 25

“Reviewing the ECG Settings” on page 31

Loading...
+ 63 hidden pages