Sirona Heliodent Plus Installation manual

5 (2)

kÉï=~ë=çÑW== MVKOMNR

ebiflabkqmirp

fелн~дд~нбзе=fелнкмЕнбзел

 

 

 

 

 

bеЦдблЬ

 

 

PLUS

 

0.08S

 

HELIODENT

 

 

 

 

7mA DC

 

 

 

70kV

60kV

 

 

 

0.08S

 

PLUS

 

 

 

 

7mA DC

 

HELIODENT

 

70kV

60kV

 

 

=

 

Sirona Dental Systems GmbH

Table of contents

Installation Instructions HELIODENTPLUS

 

Table of contents

1

About these Installation Instructions........................................................................

6

 

1.1

Scope

...........................................................................................................

6

 

1.2

Round ......................................................and angular support arm system

6

 

1.3

Other documentation ......................................................................required

6

 

1.4

Structure ............................................................................of the document

7

 

 

1.4.1 ....................................................Identification of the danger levels

7

 

 

1.4.2 .............................................................Formats and symbols used

7

2

General safety ......................................................................................information

8

3

Installation options...................................................................................................

10

4

Delivery and transport .............................................................................................

13

 

4.1

Transport ..............................................................and operating conditions

13

 

4.2

Checking ...................................................................................the delivery

13

5

Required materials ..................................................................................................

15

 

5.1

Required .......................................................................installation material

15

 

5.2

Required .........................................................tools and auxiliary materials

16

6

Installation ...............................................................................................................

 

18

 

6.1

Installing .............................................................................the wall adapter

18

 

 

6.1.1 ......................................................................................

Preparation

18

 

 

6.1.2 .................................................

Installation (concealed installation)

20

 

 

6.1.3 ......................................................

Installation (surface installation)

21

 

6.2

Installing ........................................the round support arm and scissor arm

22

 

 

6.2.1 ................................................................Installing the support arm

22

 

 

6.2.2 .................................................................Installing the scissor arm

23

 

6.3

Installing ......................................the angular support arm and scissor arm

26

 

 

6.3.1 ...................................................Installing the angular support arm

26

 

 

6.3.2 ....................................................Installing the angular scissor arm

27

 

6.4

Preparing ..................................................................the protective housing

29

 

6.5

Installing .......................................the remote control and the remote timer

30

 

 

6.5.1 ................................................................

Wall - mounted installation

30

 

 

6.5.2 ..............................................................Modifying the remote timer

32

 

6.6

Installing .............................................................the manual release holder

34

 

 

6.6.1 .....................................................................Installation on the wall

34

 

 

6.6.2 ............................................Installation on the cover (wall module)

35

 

 

6.6.3 ................Installation on the protective cover of the remote control

36

bеЦдблЬ

62 14 972 D 3507

3

D 3507.031.01.13.02 09.2015

Table of contents

Sirona Dental Systems GmbH

 

Installation Instructions HELIODENTPLUS

7

8

9

Electrical connection ................................................................................................

 

37

7.1

Connecting the arm cable.............................................................................

 

37

7.2

Connecting the grounding strap....................................................................

38

7.3

Connecting the control panel ........................................................................

39

 

7.3.1

Direct connection (standard)............................................................

39

 

7.3.2

Connection to the remote timer........................................................

40

7.4

Connecting the release and the door contact ...............................................

41

 

7.4.1

General ............................................................................................

 

41

 

7.4.2

Overview..........................................................................................

 

41

 

 

7.4.2.1 Removing the EMC plate ..................................................

44

 

7.4.3

Installation option

1..........................................................................

44

 

7.4.4

Installation option

2..........................................................................

45

 

7.4.5

Installation option

3..........................................................................

47

 

7.4.6

Installation option

4..........................................................................

48

 

7.4.7

Installation option

5..........................................................................

50

 

7.4.8

Installation option

6..........................................................................

51

 

7.4.9

Installation option

6.1.......................................................................

54

 

7.4.10

Installation option

7..........................................................................

57

 

7.4.11

Installation option

8..........................................................................

58

7.5

Completion....................................................................................................

 

59

7.6

Preparing the power connection ...................................................................

59

Starting up................................................................................................................

 

 

61

8.1

Functional test...............................................................................................

 

61

 

8.1.1

Overview..........................................................................................

 

61

 

8.1.2

Checking the protective ground wires..............................................

62

 

8.1.3

Checking the unit leakage current....................................................

64

 

8.1.4

Checking the X-ray tube assembly ..................................................

65

 

8.1.5

Checking the tube current................................................................

65

 

8.1.6

Checking exposure time and high voltage kV..................................

67

 

8.1.7

Checking the deadman function.......................................................

68

8.2

Attaching the housing covers........................................................................

69

 

8.2.1

Support arm cover............................................................................

69

 

8.2.2

Attach the EMC additional plate for the wall adapter to the power

70

 

 

supply

 

 

 

8.2.3

Wall module .....................................................................................

 

71

 

8.2.4

Remote control.................................................................................

 

72

 

8.2.5

Remote timer....................................................................................

 

73

8.3

Individual unit adaptation ..............................................................................

 

74

Final work.................................................................................................................

 

 

75

4

62 14 972 D 3507

D 3507.031.01.13.02 09.2015

Sirona Dental Systems GmbH

Table of contents

Installation Instructions HELIODENTPLUS

 

10

Appendix..................................................................................................................

 

76

 

10.1

Selection of the release button.....................................................................

76

 

10.2

Service routines............................................................................................

77

 

 

10.2.1

Operation.........................................................................................

77

 

 

10.2.2

Service routine S15.........................................................................

78

 

 

10.2.3 Adjustment following increased AL filter value................................

79

 

 

 

10.2.3.1 Adapting radiation times...................................................

79

 

10.3

Replacing old units.......................................................................................

80

 

 

10.3.1

Overview .........................................................................................

80

 

 

10.3.2 Replacement of horizontally mounted old units with an adapter

80

 

 

 

plate

 

 

 

10.3.3 Replacement of vertically mounted old units...................................

83

 

10.4

Installing the mobile stand............................................................................

86

bеЦдблЬ

62 14 972 D 3507

5

D 3507.031.01.13.02 09.2015

1 About these Installation Instructions

Sirona Dental Systems GmbH

1.1 Scope

Installation Instructions HELIODENTPLUS

1About these Installation Instructions

1.1Scope

These Installation Instructions describe the installation of the HELIODENTPLUS extraoral X-ray system. They are intended for use exclusivelybytrainedandauthorizeddistributorsandservicetechnicians.

1.2Round and angular support arm system

The HELIODENT Plus is available with a round and angular support arm system.

In the images which do not exclusively demonstrate the angular support arm system, the round support arm system is shown.

1.3Other documentation required

In addition to these installation instructions you will require the following documentation:

Drilling template

HELIODENTPLUS drilling template: Order No. 62 15 003

Wiring diagrams

HELIODENTPLUS Wiring References: Order No. 62 15 086

Service Manual

HELIODENTPLUS Service Manual: Order No. 62 15 102

6

62 14 972 D 3507

D 3507.031.01.13.02 09.2015

Sirona Dental Systems GmbH

1 About these Installation Instructions

Installation Instructions HELIODENTPLUS

1.4 Structure of the document

1.4Structure of the document

1.4.1Identification of the danger levels

To prevent personal injury and material damage, please observe the warning and safety information provided in the present operating instructions. Such information is highlighted as follows:

DANGER

An imminent danger that could result in serious bodily injury or death.

WARNING

A possibly dangerous situation that could result in serious bodily injury or death.

CAUTION

A possibly dangerous situation that could result in slight bodily injury.

NOTICE

A possibly harmful situation which could lead to damage of the product or an object in its environment.

IMPORTANT

Application instructions and other important information.

Tip: Information on making work easier.

1.4.2Formats and symbols used

The formats and symbols used in this document have the following meaning:

Prerequisite

Requests you to do something.

1.First action step

2.Second action step

or

Alternative action

 

Result

 

 

Individual action step

 

 

 

See "Formats and symbols

Identifies a reference to another text

used [→7]"

passage and specifies its page

 

 

number.

 

 

 

List

Designates a list.

 

 

"Command / menu item"

Indicates commands, menu items or

 

 

quotations.

bеЦдблЬ

62 14 972 D 3507

7

D 3507.031.01.13.02 09.2015

2 General safety information

Sirona Dental Systems GmbH

 

Installation Instructions HELIODENTPLUS

2 General safety information

DANGER

Potentially lethal shock hazard:

ItisessentialthatyouswitchthelinevoltageOFFatthemainswitch of the building installation before beginning installation of the unit!

DANGER

Potentially lethal shock hazard!

Fixed connection!

Installingamainspluginsteadofthespecifiedfixedconnectioninfringes international medical regulatory actions and is prohibited. In case of error, this puts patients, users, and other parties seriously at risk.

WARNING

Radiation protection

The valid radiation protection regulations and measures must be observed. The statutory radiation protection equipment must be used.

In case of malfunctions, cancel the exposure immediately by letting go of the exposure release button.

CAUTION

Risk of injury!

The installation must be carried out in accordance with the requirements stated in our Installation Instructions.

Installation may be carried out only by personnel specifically authorized by SIRONA.

CAUTION

Electromagnetic compatibility:

The unit should not be operated in the immediate vicinity of other devices (see also "Installation Requirements, Chapter on EMC“).

If this proves to be unavoidable, the unit should be monitored to ensure that it is operating properly.

CAUTION

Electrical components of the unit can be destroyed.

Prior to opening the unit

Please comply with the usual precautionary measures for handling printed circuit boards (ESD).

Make sure you touch a ground point to discharge yourself prior to touching the components.

Use an ESD wrist band and connect it to the protective ground wire.

8

62 14 972 D 3507

D 3507.031.01.13.02 09.2015

Sirona Dental Systems GmbH

2 General safety information

Installation Instructions HELIODENTPLUS

 

CAUTION

Gaps appear between the internal hinges when moving the angular support arm.

Fingers may be crushed in these gaps.

Ensure that you never place your fingers in the gaps between the hinges, neither during operation nor for cleaning purposes.

WARNING

Modifications to the product

Modifications which may affect the safety of the operators or third parties are prohibited by law!

CAUTION

No additional components on the support arm system

It is not permissible to fix additional components to the support arm system.

bеЦдблЬ

62 14 972 D 3507

9

D 3507.031.01.13.02 09.2015

3 Installation options

Sirona Dental Systems GmbH

 

Installation Instructions HELIODENTPLUS

1

S3

2

S9

S3

L9

3

S4

S3

L6

3 Installation options

Designations for release buttons and door contact

Manual release S3

Coiled cable

Release key on the control membrane S4

Directly connected to control board DX4

Remote control release key S9

Integrated in remote control housing

Door contact (safety circuit) S7

Installation option 1

Release in the treatment room without remote control

Release

Manual release S3

Installation option 2

Release in the treatment room with remote control

Release

Manual release S3 or

Remote control release key S9

Installation option 3

Release in the treatment room withRemote Timer

Release

Manual release S3 or

Release key on the control membrane S4

NOTICE

Length of cable supplied for Remote Timer approx. 10 meters (393") (must not be extended).

Conduit int. dia. at least 12 mm (1/2").

10

62 14 972 D 3507

D 3507.031.01.13.02 09.2015

Sirona Dental Systems GmbH

3 Installation options

Installation Instructions HELIODENTPLUS

 

4

S9

S3

L9

Installation option 4

Release outside of the X-ray room with remote control

Release

Manual release S3 or

Remote control release key S9

NOTICE

Installation prerequisites

Use of the remote control is permissible only if the yellow X-Ray LED is visible to the operating personnel during radiation release.

5

S4

S3

L6

Installation option 5

Release outside of the X-ray room withRemote Timer

Release

Manual release S3 or

Release key on the control membrane S4

NOTICE

Length of cable supplied for Remote Timer approx. 10 meters (393") (must not be extended).

Conduit int. dia. at least 12 mm (1/2").

6

 

 

S9

S7

S3

 

L9

Installation option 6

Release outside of the X-ray room with remote control, door contact safety circuit

Door contact

Door contact S7 wired to the wall adapter

Release

Manual release S3 or

Remote control release key S9

NOTICE

Installation requirement

Use of the remote control is permissible only if the yellow X-Ray LED is visible to the operating personnel during radiation release.

bеЦдблЬ

62 14 972 D 3507

11

D 3507.031.01.13.02 09.2015

3 Installation options

Sirona Dental Systems GmbH

 

Installation Instructions HELIODENTPLUS

6.1

 

 

S9

S7

S3

 

L9

Installation option 6.1

Release outside of the X-ray room with remote control, door contact safety circuit

Door contact

Door contact S7 wired to the remote control housing

Release

Manual release S3 or

Remote control release key S9

NOTICE

Installation requirement

Use of the remote control is permissible only if the yellow X-Ray LED is visible to the operating personnel during radiation release.

7

S4

S7

L6

8

 

S7

S3

 

 

L6

Installation option 7

Release outside of the X-ray room with Remote Timer, door contact safety circuit

Door contact

Door contact S7

Release

Release key on the control membrane S4

Installation option 8

Release outside of the X-ray room with Remote Timer, door contact safety circuit

Door contact

Door contact S7 wired to Remote Timer

Release

Manual release S3

12

62 14 972 D 3507

D 3507.031.01.13.02 09.2015

Sirona Dental Systems GmbH

4 Delivery and transport

Installation Instructions HELIODENTPLUS

4.1 Transport and operating conditions

4 Delivery and transport

4.1Transport and operating conditions

Transport and operating conditions:

Transport and storage

-40°C – +70°C (-40°F – 158°F)

temperature:

 

Air humidity:

10% – 95%

Operating conditions as

Ambient temperature +10 °C – +40 °C

specified in IEC 60601-1:

(50 °F – 104 °F)

 

Relative humidity: 30% – 75%

Operating altitude:

≤ 3000 m

4.2Checking the delivery

All products from Sirona are carefully checked prior to shipment. Please perform an incoming inspection immediately after delivery.

1.Check the delivery note to ensure that the consignment is complete.

2.Check whether the product shows any visible signs of damage.

NOTICE

Damage during transport

If the product was damaged during transport, please contact your carrying agent.

If return shipment is required, please use the original packaging for shipment.

HELIODENTPlus Packaging, bottom part

bеЦдблЬ

62 14 972 D 3507

13

D 3507.031.01.13.02 09.2015

 

4 Delivery and transport

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sirona Dental Systems GmbH

 

4.2 Checking the delivery

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Installation Instructions HELIODENTPLUS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HELIODENTPlus Packaging, top part

14

62 14 972 D 3507

D 3507.031.01.13.02 09.2015

Sirona Dental Systems GmbH

5 Required materials

Installation Instructions HELIODENTPLUS

5.1 Required installation material

5 Required materials

5.1Required installation material

CAUTION

Observe wall properties

In installation situations, the technician is responsible for the assessmentof wall properties andselecting themethodof attachingthe unit to the wall.

N

L

 

 

 

 

K

 

O

Tensile force per screw

bеЦдблЬ

K

 

M

 

 

3600 N (800 lbf) if L ≤ 700mm (27 1/2“)

 

 

4200N(950 lbf) if L ≤ 950mm (37 3/8“)

 

L

Length of support arm

 

M

Mounting plate (supplied)

 

N

Anchor plate

 

O

Threaded bolt M8

 

Thepermissibletensileforceoftheselectedattachmentmustatleast equal the tensile force listed above.

Matching wood screws for wooden beams are included in delivery.

For all other wall structures, special wall anchors must be purchased from a selected dealer. The wall anchors and screws should be identical for every attachment point.

Alternatively, an anchor plate can be used as a counter bearing. In this case, M8 threaded rods of the appropriate length for the wall (thickness of the wall + 2 x mounting plate thickness + attachment material) are required.

62 14 972 D 3507

15

D 3507.031.01.13.02 09.2015

5 Required materials

Sirona Dental Systems GmbH

5.2 Required tools and auxiliary materials

Installation Instructions HELIODENTPLUS

5.2Required tools and auxiliary materials

● Drill or drill hammer, depending on the ground

● Masonry drill bit 10 mm (3/8") ● Masonry drill bit 6 mm (1/4")

Wood drill bit 6 mm (1/4")

Metal drill bit 4.3 mm (3/16")

Hammer

● Center punch

● Spirit level

● Scissors

16

62 14 972 D 3507

D 3507.031.01.13.02 09.2015

Sirona Dental Systems GmbH

5 Required materials

Installation Instructions HELIODENTPLUS

5.2 Required tools and auxiliary materials

 

● Screwdriver, Torx®, sizes T10, T15 and T20

● Slotted screwdriver, sizes 2.5mm and 3.5 mm (3/32" and 9/64")

● Allen wrench size 13mm, 4mm

● Socket wrench or ratchet wrench (with extension), size 13 mm

● Retaining ring pliers for outside rings (shafts)

● Drift punch

● Multimeter or ammeter (battery-operated)

Test unit for device leakage current measurement:

For measurements, Sirona recommends an automatic tester (example illustration) which complies with standard IEC 62353. If you do not use an automatic tester, please pay attention to the specifications in the standard IEC 62353.

Power source for protective ground wire test Technical data:

No-load voltage max. 6V

Short-circuit current at least 5A - max. 25A

bеЦдблЬ

62 14 972 D 3507

17

D 3507.031.01.13.02 09.2015

6 Installation

Sirona Dental Systems GmbH

6.1 Installing the wall adapter

Installation Instructions HELIODENTPLUS

6 Installation

6.1

Installing the wall adapter

 

For HELIODENT PLUS on the mobile stand, see Installing the mobile

 

stand [→86]

6.1.1

Preparation

 

Attachment points

These installation instructions distinguish between wall-mounted installation (masonry) and installation on wooden beams.

Wall-mounted installation: 3 attachment points

Positions A1, A3 and A4

Wooden beam installation: 2 attachment points

Positions A2 and A4

NOTICE

Covering up old installation sites

Possibilities for using or covering up the installation sites of old units are described in the section "Replacing old units" [→80].

Getting started

DANGER

Shock hazard!

Be sure to switch off the line power supply before connecting the line voltage!

18

62 14 972 D 3507

D 3507.031.01.13.02 09.2015

Sirona Dental Systems GmbH

6 Installation

Installation Instructions HELIODENTPLUS

6.1 Installing the wall adapter

 

 

 

 

 

 

1.

For concealed installation only: Cut out cable bushings (B) and (F) in

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

the drilling template.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2.

Pull the cable through the cable bushings (B) and (F).

 

 

 

 

 

 

 

 

 

 

 

 

3.

Align the drilling template using a spirit level and attach it to the wall.

A1

 

 

A3

4.

Center punch the fastening points and drill according to the type of

 

 

 

fastening used.

 

 

A2

 

 

 

 

 

 

 

 

- When securing with an anchor plate, drill into wall with a 8mm drill

F

 

 

 

 

 

bit.

 

 

 

 

 

 

 

- If installing with wall plugs, drill according the the wall plugs used

 

 

 

 

 

 

 

(max. 10mm) and insert the wall plugs.

 

 

 

 

 

 

 

- In case of a wooden wall, drill holes at least 80 mm deep with a

 

 

 

 

 

 

 

6mm wood drill bit.

 

 

 

 

 

 

5.

Remove the drilling template.

A4 B

bеЦдблЬ

62 14 972 D 3507

19

D 3507.031.01.13.02 09.2015

6 Installation

Sirona Dental Systems GmbH

6.1 Installing the wall adapter

Installation Instructions HELIODENTPLUS

6.1.2Installation (concealed installation)

1. Remove wall adapter (P1) from the packaging.

 

CAUTION

 

Risk of damage

P1

Make sure that board DX1 is not damaged during installation!

Observe ESD protective measures!

 

F

L9/L6

2.For the power cable: Carefully punch out opening B in the wall adapter with a punch.

3.Position the wall adapter on the prepared installation location.

4.Pull the power cable through the opening (B) in the wall adapter.

5.For remote control / Remote Timeronly (cable L6 or L9): Pull the cable through the opening (F) in the wall adapter.

6.Screw the wall adapter onto the wall loosely to hold it in place.

7.Align the wall adapter with a spirit level and attach the wall adapter securely with suitable screws and washers, depending on the type of attachment.

B

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

20

 

 

 

 

 

 

 

 

 

 

62 14 972 D 3507

 

 

 

 

 

 

 

 

 

 

D 3507.031.01.13.02 09.2015

Sirona Dental Systems GmbH

6 Installation

Installation Instructions HELIODENTPLUS

6.1 Installing the wall adapter

6.1.3Installation (surface installation)

1. Remove wall adapter (P1) from the packaging.

 

CAUTION

 

Risk of damage

P1

Make sure that board DX1 is not damaged during installation!

Observe ESD protective measures!

 

2. Position the wall adapter on the prepared installation location. 3. Screw the wall adapter onto the wall loosely to hold it in place.

4. Align the wall adapter with a spirit level and attach the wall adapter securely with suitable screws and washers, depending on the type of attachment.

5. Insert the power cable into the mounting plate from underneath and secure it with a cap clamp.

L9/L6

For remote control / Remote Timer only (cable L6 or L9):

1.Unscrew the 3 screws from the EMC plate and suspend the EMC plate at the side, see Removing the EMC plate [→44].

2.Insert the cable (L6 or L9) into the mounting plate from underneath.

3.Route the cable toward the top along the side and secure it with a clamp.

bеЦдблЬ

62 14 972 D 3507

21

D 3507.031.01.13.02 09.2015

6 Installation

Sirona Dental Systems GmbH

6.2 Installing the round support arm and scissor arm

Installation Instructions HELIODENTPLUS

6.2

6.2.1

P2

Installing the round support arm and scissor arm

Installing the support arm

1. Remove support arm (P2) from the packaging.

2. Insert the support arm into the bearing (U).

3.After insertion, secure the support arm with a retaining ring (W) to prevent it from getting lifted out.

S

T

NOTICE

Risk of damaging housing parts

 

 

 

 

It is essential that you use only your hands when removing the housing

 

 

shells (A and B).

 

 

The surfaces of the housing shells will be damaged if you use any tools

 

 

such as screwdrivers, etc.

 

 

4.

Unlock the housing (A) by pressing the housing shells together at

 

 

 

position S.

 

 

5.

Remove the housing (A) from the support arm.

 

 

6.

Unlock the housing (B) by pressing the housing shells together at

 

 

 

position T.

 

 

7.

Remove the housing (B) from the support arm.

22

62 14 972 D 3507

D 3507.031.01.13.02 09.2015

Sirona Dental Systems GmbH

6 Installation

Installation Instructions HELIODENTPLUS

6.2 Installing the round support arm and scissor arm

 

6.2.2

 

P3

E

F

V

V

Installing the scissor arm

1.Push the washer (V (inside diameter 20 mm)) over the arm cable (F) (Leave the scissor arm in the package for this step).

2.Place the washer (V) onto the spindle (E).

3.Remove the scissor arm (P3) from the packaging.

4.For transport and for the next installation step, support the scissor arm on your shoulder.

CAUTION

Risk of injury

Do not fold up the scissor arm in the uninstalled state. Carefully transport the scissor arm in the unfolded state!

NOTICE

If the arm is to be laid down in the meantime, put it back in its packaging.

5.Route the arm cable (L1) downwards through the arm and insert the scissorarm.Makesurethatthewasher(V)doesnotfalloffduringthis step.

L1

6. Push the retaining ring (D) through the arm cable

bеЦдблЬ

62 14 972 D 3507

23

D 3507.031.01.13.02 09.2015

6 Installation

Sirona Dental Systems GmbH

6.2 Installing the round support arm and scissor arm

Installation Instructions HELIODENTPLUS

7. Snap the retaining ring into the groove (G).

G

8.

Push the arm cable (L1) through the support arm.

9.

Disconnect the arm cable (L1) and the grounding strap (C) in the

 

accessible area.

10.

Thread the ferrite core (Y) only onto the arm cable (L1) and slide the

C

ferrite core into the support arm.

 

NOTICE

The grounding strap must not be pulled through the ferrite core.

11.Now pull the grounding strap (C) downward through the bearing.

12.Pull the arm cable (L1) downward through the bearing.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

24

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

62 14 972 D 3507

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

D 3507.031.01.13.02 09.2015

Sirona Dental Systems GmbH

6 Installation

Installation Instructions HELIODENTPLUS

6.2 Installing the round support arm and scissor arm

13.Check the support arm with the spirit level. If necessary, loosen the bearing mount (4 nuts) and place the shim (X) underneath.

bеЦдблЬ

62 14 972 D 3507

25

D 3507.031.01.13.02 09.2015

6 Installation

Sirona Dental Systems GmbH

6.3 Installing the angular support arm and scissor arm

Installation Instructions HELIODENTPLUS

6.3Installing the angular support arm and scissor arm

6.3.1Installing the angular support arm

1. Remove support arm (P2) from the packaging.

2. Separate the ends of the threading aid from each other.

 

NOTICE

 

The threading aid must not be removed from the support arm, as

P2

threading the arm cable is much more difficult without the threading aid.

 

3. Set the cover of the arm connection onto the wall module as shown, between the support arm and the wall.

4. Thread the threading aid through the bearing. Position the support arm in the bearing (U).

5. After insertion, secure the support arm with a retaining ring (W) to prevent it from getting lifted out.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

26

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

62 14 972 D 3507

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

D 3507.031.01.13.02 09.2015

Sirona Heliodent Plus Installation manual

Sirona Dental Systems GmbH

6 Installation

Installation Instructions HELIODENTPLUS

6.3 Installing the angular support arm and scissor arm

6.3.2Installing the angular scissor arm

CAUTION

Gaps appear between the internal hinges when moving the angular support arm.

Fingers may be crushed in these gaps.

Ensure that you never place your fingers in the gaps between the hinges, neither during operation nor for cleaning purposes.

P3

E

F

V

 

1.Push the washer V (inside diameter 20 mm)) over the arm cable (F) (Leave the scissor arm in the package for this step).

2.Place the washer (V) onto the spindle (E).

3.Remove the scissor arm (P3) from the packaging.

4.For transport and for the next installation step, support the scissor arm on your shoulder.

CAUTION

Risk of injury

Do not fold up the scissor arm in the uninstalled state. Carefully transport the scissor arm in the unfolded state!

NOTICE

If the arm is to be laid down in the meantime, put it back in its packaging.

bеЦдблЬ

62 14 972 D 3507

27

D 3507.031.01.13.02 09.2015