hot point® Dry Block Probe Calibrators

CL900/950 Series

UCalibrates Thermocouples and RTD’s in Seconds

UHighly Accurate 0.9°C (1.5°F) and Exceptional Stability (0.3°F) for Precision Calibrations

UCompletely Portable for Factory, Laboratory and Instrument Shops

UInterchangeable Test Wells and CL950 BuiltIn 5 Port Well Design Provides Customer Flexibility

UWide Temperature Range 40°F Above Ambient to 900°F (482°C) Covers Most Industrial Applications

UNIST Calibration Available

Optional

Patented

Covered by U.S. and International patents and pending applications.

CL950 hot point® dry block calibrator, shown smaller than actual size with MDSS41-TC benchtop thermometer (see omega.com) calibrating multiple thermocouple probes (see omega.com).

OMEGA’s hot point® calibrators allow calibration of thermocouple and RTD probes quickly and easily with accuracy to 0.9°C (1.5°F). An integral RTD sensor assures high measurement accuracy and repeatability. A built-in microprocessor-based PID digital controller assures exceptional stability. Setpoint and actual temperature are displayed simultaneously. Temperature can be set in one-degree increments from 22°C (40°F) above ambient to 482°C (900°F).

The CL900 calibrator accepts a variety of standard and metric optional inserts which are available in 4 and 6" depths and for probe diameters from 18 to 38". An undrilled insert is also available. The CL900 comes complete with a 14" diameter, 6" deep insert well and insert block tongs. Additional inserts can be

ordered separately as required for other diameter probes. 6" depth inserts are used for probes 7" and longer; 4" depth inserts are used for probes 5" and longer.

The CL950 calibrator features a built-in well with 5 ports sized to accommodate a 116, 18, 316, and

two 14" diameter probes. The multiport design allows for the use of a calibrated or NIST-traceable probe and meter to be used as a standard while calibrating other probes.

The 5 port design also allows for calibration of more than 1 probe at a time, and eliminates the need for separate insert wells.

Caution: To avoid fire hazard or damage to your calibrator, always allow your calibrator to cool down to ambient temperature before returning to storage.