INFICON P3000XL User Manual

4 (2)
TECHNICAL HANDBOOK
kina26e1-p (1209)
Catalog No.
520-001
520-002
520-003
520-004
520-103
520-104
520-105
520-106
from software version V 2.9
Protec P3000(XL)
Helium Sniffer Leak Detector
0-2
Content 0-3
Content
Content 0-3
General Safety Precautions 0-6
1 General Information 1-1
1.1 Introduction 1-1
1.1.1 Intended Use 1-1
1.1.2 Available Configurations 1-2
1.1.3 Technical Data 1-2
1.2 Support by INFICON 1-3
1.2.1 Service Centers 1-5
1.3 Unpacking 1-7
1.3.1 Supplied Equipment 1-7
1.3.2 Accessories 1-8
1.4 Notes on How to Use This Handbook 1-9
1.4.1 Symbols of Vacuum Technology 1-9
1.4.2 Definition of Terms 1-10
1.5 Instrument Views of the Protec P3000 1-11
1.6 Installation 1-12
1.6.1 Set up 1-12
1.6.2 Mechanical Connections 1-13
1.6.3 Electrical Connections 1-16
1.6.4 RS232 Interface 1-17
1.6.5 I/O Port 1-17
2 How the Protec P3000 Works 2-1
2.1 Description of the Functions 2-1
2.2 Description of the Subassemblies 2-1
2.2.1 Backing Pump 2-1
2.2.2 Wise TechnologyTM Sensor 2-1
2.2.3 Valve Holder 2-2
2.2.4 Control Assembly 2-2
2.3 Description of the Displays and User Interfaces 2-2
2.3.1 Main unit display 2-2
2.3.2 SL3000(XL) Sniffer line with probe display 2-4
2.3.3 Built-in PRO-Check reference leak 2-5
3 Operation of the Protec P3000 3-1
3.1 Start-Up 3-1
3.2 Controls on the main display unit 3-2
3.3 Controls on the probe display unit 3-5
3.4 Performing measurements 3-5
3.4.1 Standard Operation Mode 3-6
3.4.2 I•Guide Operating Mode 3-9
3.4.2.1 Starting the I•Guide Mode 3-9
3.4.2.2 Selecting an I•Guide Program 3-10
3.4.2.3 Using an I•Guide Program 3-10
3.4.3 The Info Page 3-13
0-4 Content
3.5 Calibration and Self-Test 3-15
3.5.1 Verifying a calibration (proof function) 3-15
3.5.2 Internal cali bration 3-17
3.5.3 External calibration 3-17
3.6 Standby 3-20
3.7 Shutdown 3-20
3.8 Storage for fast availability as back-up unit 3-20
4 Equipment Settings 4-1
4.1 Menu Structure 4-1
4.2 The Service Menu 4-3
4.3 Selecting gas equivalents and setting trigger values 4-3
4.3.1 Setting gas parameters 4-3
4.3.2 Selecting a set of gas parameters 4-7
4.4 Settings Sub-menu 4-8
4.4.1 Vacuum & Access 4-8
4.4.2 Audio Functions 4-15
4.4.3 Display Settings 4-17
4.4.4 Setting-up / editing an I•Guide Program 4-19
4.4.5 Miscellaneous Settings 4-21
4.5 Interfaces 4-23
4.5.1 Control location 4-23
4.5.2 Recorder outputs 4-23
4.5.3 RS232 Protocol 4-25
4.5.4 Select PLC inputs 4-27
4.5.5 Baud rate & end sign 4-27
4.5.6 PRO-Check (only available in advanced mode) 4-27
4.6 The Info Menu 4-28
4.7 History & Maintenance 4-32
5 Protec P3000 Messages 5-1
5.1 Error Messages and Warnings 5-1
6 Equipment Connections 6-1
6.1 I/O Port (Control Inputs and Outputs) 6-1
6.1.1 Ground connectors 6-2
6.1.2 24V Output 6-2
6.1.3 PLC Inputs 6-2
6.1.4 PLC Outputs 6-3
6.1.4.1 Relay outputs 6-5
6.1.4.2 Recorder Outputs 6-5
6.1.5 How to perform a calibration? 6-6
6.2 RS232 interface 6-7
7 Maintenance 7-1
7.1 Maintenance schedule 7-1
7.2 Exchanging the air filter 7-2
7.3 Exchanging the external fuses 7-4
Content 0-5
7.4 Replacing filters in the sniffer line 7-4
7.4.1 Replacing the felt discs of the capillary filter
(for SL3000 only) 7-5
7.4.2 Replacing the felt discs when using the water
protection tip (for SL3000 only) 7-7
7.4.3 Checking / replacing the sinter filter
(for SL3000 only) 7-7
7.5 Replacing the filter pad of the sniffer tip
(for SL3000XL only) 7-9
7.6 Switching the capillary filter
(for SL3000 sniffer line only) 7-11
7.6.1 Switching from metal to plastic capillary filter 7-11
7.6.2 Switching from plastic to metal capillary filter 7-12
7.7 Replacing the gas reservoir of the
PRO-Check 7-13
8 The gas library 8-1
0-6 General Safety Precautions
General Safety Precautions
Notice Indicates special requirements the user must comply with.
The INFICON Protec P3000 leak detector has been designed for safe and efficient
operation when used properly and in accordance with this Technica l Handb ook. It is
the responsibility of the user to carefully read and strictly observe all safety
precautions described in this chapter and throughout this Technical Handbook. The
Protec P3000 must only be operated in the proper condition and under th e conditions
described in this Technical Handbook. It must be operated and maintained by train ed
personal only. Consult local, state, and national agencies regarding specific
requirements and regulations. Adress any further safety, operation and / or
maintenance questions to our nearest office.
Failure to observe the following precautions could result in serious personal
injury:
Warning
Indicates procedures that must be strictly observed to prevent haza rd s to per sons.
Caution
Indicates procedures that must strictly be observed to prevent damage to or
destruction of the Protec P3000 leak detector.
Warning
Danger of explosion!
To use the Protec P3000 in explosion hazard areas could cause ignition of
flammable mixtures.
The Protec P3000 must only be operated outside of explosion hazard areas.
Warning
Only 3-core mains cables having a protective ground conductor must be used.
Operation of the Protec P3000 with the ground conductor unconnected is not
permissible.
General Safety Precautions 0-7
Warning
Do not stare into the LEDs of the sniffer line intentionally for extended times or at a
close distance as this may cause permanent damage to the eye.
Warning
Danger of electric shock.
Don’t touch voltaged parts with the sniffer tip. Test samples need to be
disconnected from electricity before leak testing.
Warning
For all contacts of the I/O Port a maximum voltage of 60 V DC or 25 V AC must not
be exeeded or reached to ground or ground equipment conductors.
According to the type of in- or outputs lower voltages had to be accepted. For this,
please refer to the information given in the responding chapters.
Warning
For all maintenance on the Protec P3000, the Protec P3000 must be disconnected
from power.
Warning
Before exchanging the air filter the Protec P3000 must be disconnected from
power.
Warning
Before exchanging the fuses the Protec P3000 must be disconnected from power.
0-8 General Safety Precautions
Failure to observe the following precautions could result in damage to the
equipment:
Caution
The Protec P3000 must not be operated while standing in water or when exposed
to drip water. The same applies to all other kinds of liquids.
This Protec P3000 should only be used in rooms.
Caution
Avoid contact of the Protec P3000 with bases, acids and solvents as well as
exposure to extreme climatic conditions.
Caution
Ensure sufficient air cooling (see also Section 1.1.2)
Caution
Before installation remove the transportation lock.
Caution
In order to ensure adequate ventilation of the Protec P3000, a space of at least
20 cm 8) (8 in.) must be kept unobstructed to the sides. The clearance at the rear
must be no less than 10 cm (4 in.). Moreover, the Protec P3000 handles for carrying
the leak detector at the sides of the main unit must not be covered at any time as
these acts as air inlet and outlet. Avoid the presence of heat sources in the vicinity
of the Protec P3000.
Caution
Before connecting the Protec P3000 to the mains you must make sure that the
mains voltage rating of the Protec P3000 coincides with the locally available mains
voltage.
Caution
Do not suck in any liquids.
General Safety Precautions 0-9
Caution
Permissible maximum input voltage PLC 28 V.
Caution
Permissible max. voltage and current for open collector outputs are: 28 V; 50 mA.
Caution
Maximum load rating relay outputs is 60 V DC / 25 V AC and 1 A per relay.
Caution
The air filter should be checked for contamination at least every 6 months and
should be definitely exchanged after 2 years.
0-10 General Safety Precautions
General Information 1-1
1 General Information
The Protec P3000 helium leak detector is supplied ready for operation. However, we
recommend that you carefully read the Technical Handbook to ensure optimum
operating conditions right from the start. This handbook contains important
information on functions, installation, start-up and operation of the Protec P3000.
If not stated otherwise this technical handbook applies to all three configurations of
the Protec P3000 (see Section 1.1.2). Sections specific to one configuration are
marked as "for … only". Sections marked as "for Protec P300 0XL only" always apply
to the Protec P3000XL with the SL3000XL sniffer line (capable of
HIGH FLOW mode).
1.1 Introduction
1.1.1 Intended Use
The Protec P3000 is a helium leak detector for sniffer applications. It may be used to
localise and quantify leaks in test samples if there is helium under an overpressure
within the test sample and when searching the test sample with a sniffer probe from
the outside (sniffer method). The use of this sniffer probe is mandatory for proper
operation and it is available as an accessory (Cat. No. 525-001 to 525-004).
Caution
The Protec P3000 must not be operated while standing in water or when exp osed
to drip water. The same applies to all other kinds of liquids.
This Protec P3000 should only be used in rooms.
Caution
Avoid contact of the Protec P3000 with bases, acids and solvents as well as
exposure to extreme climatic conditions.
Caution
Ensure sufficient air cooling (see also Section 1.1.2)
1-2 General Information
1.1.2 Available Configurations
The Protec P3000 leak detector is available in four different configurations:
The Standard Pr otec
P3000
The standard Protec P3000 should be used for high sensitivity applications. It
requires the SL3000 sniffer line.
The Protec P3000,
RC version
The Protec P3000, RC version is the standard Protec P3000 but with an external
display unit. It requires the SL3000 sniffer line.
The Protec P3000XL
The Protec P3000XL is the
HIGH FLOW version of the Protec P3000. It is able to detect
leaks at a much higher distance from the potential leak at reduced sensitivity but can
be switched back to normal flow at inc reased sensitivity. In order to use both flow
modes (high and low) it requires the use of the SL 3000XL sniffer line. It may also be
operated with the normal SL3000 sniffer line, in this case the high flow mode is
disabled, however.
The Protec P3000XL,
RC version
The Protec P3000XL, RC version is identical to the Protec P3000XL, but with an
external display unit. It requires the SL3000XL sniffer line.
1.1.3 Technical Data
Physical Data
Lowest detectable leak rate
For Protec P3000 1 x 10
-7
mbar l/s
For Protec P3000XL in
LOW FLOW mode 1 x 10
-7
mbar l/s
For Protec P3000XL in
HIGH FLOW mode 1 x 10
-6
mbar l/s
Measurement range
For Protec P3000 5 decades
For Protec P3000XL in
HIGH FLOW mode 4 decades
Helium sensor Wise Technology
TM
Sensor
Sensor response time 450 ms
General Information 1-3
Electrical Data
Other data
1.2 Support by INFICON
INFICON Service
If equipment is returned to INFICON or an authorised INFICON representative
indicate whether the equipment is free of substances damaging to health o r whether
it is contaminated. If it is contaminated also indicate the nature of the hazard.
INFICON must return any equipment without a Declaration of Contamination to the
sender’s address. You will find an appropriate form below.
Gas flow through the capillary
For Protec P3000 260 - 360 sccm*
For Protec P3000XL in
HIGH FLOW mode 2660 - 3500 sccm*
Time until ready for operation approx. 3 min
* Measured at 1 atm (1013 mbar) at sea altitude. Actual flow may vary with increased
altitude and low atmospheric pressure.
Mains voltages and mains frequencies 90 - 127 V, 50 / 60 Hz
(fixed) 115 - 140 V, 60 Hz
187 - 265 V, 50 / 60 Hz
Power consumption 200 VA
Type of protection IP 20
Overvoltage category II
Mains cord 2.5 m
Noise level < 54 dBA
Dimensions (w x h x d) in mm 610 x 370 x 265
Weight 27 kg
Permissible ambient temperature
(during operation)
10 °C to 45 °C
Permissible storage temperature -40 °C to 60 °C
Relative humidity max. 80% for temperature up to +31°C,
decreasing linearly to 50% at +40°C
Contamination level II
(according to IEC 61010 / Part 1:
“normally only non-conductive pollution
may occur. Occasionally, however, a
temporary conductivity caused by
condensation can be tolerated.”)
Max. altitude above sea level 2000m
1-4 General Information
General
We reserve the right to alter the design or any data given in this handbook.
The illustrations are not binding.
Fig. 1-1 Declaration of Contamination
Declaration of Contamination
Legally binding declaration:
I/we hereby declare that the information on this form is complete and accurate and that I/we will assume any further costs that may
arise. The contaminated product will be dispatched in accordance with the applicable regulations.
Organization/company
Address Post code, place
Phone Fax
Email
Name
Date and legally binding signature Company stamp
1) or not containing any amount
of hazardous residues that
exceed the permissible ex-
posure limits
Process related contamination of product:
toxic no
1) yes
caustic no
1) yes
biological hazard no
yes
2)
explosive no
yes
2)
radioactive no
yes
2)
other harmful substances n o
1) yes
The service, repair, and/or disposal of vacuum equipment and components will only be carried out if a correctly completed declaration has
been submitted. Non-completion will result in delay.
This declaration may only be completed (in block letters) and signed by authorized and qualified staff.
Copies:
Original for addressee - 1 copy for accompanying documents - 1 copy for file of sender
Harmful substances, gases and/or by-products
Please list all substances, gases, and by-products which the product may have come into contact with:
Trade/product name
Chemical name
(or symbol)
Precautions associated
with substance
Action if human contact
Description of product
Type
Article Number
Serial Number
Reason for return
Operating fluid(s) used (Must be drained before shipping.)
The product is free of any sub-
stances which are damaging to
health yes
This form can be downloaded
from our website.
2) Products thus contam i-
nated will not be ac-
cepted without written
evidence of decontam i-
nation!
General Information 1-5
1.2.1 Service Centers
Algeria jhj@agramkow.dk Finland jhj@agramkow.dk
Agramkow
Sonderborg
Phone: +45 741 236 36
Fax: +45 744 336 46
Agramkow
Sonderborg
Phone: +45 741 236 36
Fax: +45 744 336 46
Belarus leakdetection.service@inficon.com France Christophe.Zaffanella@oerlikon.com
INFICON GmbH
Cologne
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
OLV France
Orsay
Phone: +33 476 351 584
Fax: +33 476 351 584
Belgium leakdetection.service@inficon.com Germany leakdetection.service@inficon.com
INFICON GmbH
Cologne
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
INFICON GmbH
Cologne
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
Brazil fernandoz@prestvacuo.com.br Hungary adam.lovics@kon-trade.hu
PV Pest Vácuo Ltda.
Santa de Parnaíba
Phone: +55 114 154 4888
Fax: +55 114 154 4888
Kontrade
Budaörs
Phone: +36 23 50 38 80
Fax: +36 23 50 38 96
Bulgaria leakdetection.service@inficon.com India asdash@hotmail.com
INFICON GmbH
Cologne
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
Dashpute
400 064
Phone: +91 22 888 0324
Fax: +91 22 888 0324
Canada reachus@vpcinc.ca Ireland reach.unitedkingdom@inficon.com
Vacuum Products Canada Ltd.
Ontario
Phone: +905.672.7704
Fax: +905.672.2249
INFICON
Blackburn
Phone: +44 1254 678 250
Fax: +44 1254 698 577
Central America infoqro@meisa.com Italy davide.giovanetti@inficon.com
MEISA S.a. de C.V.
Querètaro
Phone: +52 44 22 25 42 80
Fax: +52 44 22 25 41 57
INFICON GmbH
Castelnuovo
Phone: +39 045 6 40 25 56
Fax: +39 045 6 40 24 21
China reach.china@inficon.com Israel urimark@mark-tec.co.il
INFICON LTD
Hong Kong
Phone: +852.2862.8863
Fax: +852.2865.6883
Mark Technologies Ltd.
Kiriat Ono
Phone: +972 35 34 68 22
Fax: +972 35 34 25 89
INFICON LTD
Beijing
Phone: +86.10.6590.0164
Fax: +86.10.6590.0521
Japan reach.japan@inficon.com
INFICON LTD
Guangzhou
Phone: +86.20.8723.6889
Fax: +86.20.8723.6003
INFICON Co. Ltd.
Yokohama
Phone: +81.45.471.3396
Fax: +81.45.471.3387
INFICON LTD
Shanghai
Phone: +86.21.6209.3094
Fax: +86.21.6295.2852
Czech Republic filip.lisec@inficon.com Korea reach.korea@inficon.com
INFICON GmbH
Pilsen
Phone +420 734 331 758
Fax: +420 604 203 037
INFICON Ltd.
Sungnam city
Phone: +82 312 062 890
Fax: +82 312 063 058
Denmark jhj@agramkow.dk INFICON Ltd.
Suwon City
Phone: +82 312 062 890
Fax: +82 312 063 058
Agramkow
Sonderborg
Phone: +45 744 336 36
Fax: +45 744 336 46
INFICON Ltd.
Cheonan City
Phone: +82 312 062 890
Fax: +82 312 063 058
Egypt jhj@agramkow.dk Latvia leakdetection.service@inficon.com
Agramkow
Sonderborg
Phone: +45 741 236 36
Fax: +45 744 336 46
INFICON GmbH
Cologne
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
Estonia leakdetection.service@inficon.com Lithuania leakdetection.service@inficon.com
INFICON GmbH
Cologne
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
INFICON GmbH
Cologne
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
1-6 General Information
Mexico infoqro@meisa.com Spain richard.cunill@leyboldoptics.com
MEISA S.a. de C.V.
Querètaro
Phone: +52 442 225 42 80
Fax: +52 442 225 41 57
Leybold Optics Ibérica
Barcelona
Phone: +34 93 66 60 778
Fax: +34 93 66 64 612
Netherlands leakdetection.service@inficon.com Sweden jhj@agramkow.dk
INFICON GmbH
Cologne
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
Agramkow
Sonderborg
Phone: +45 741 236 36
Fax: +45 744 336 46
Norway jhj@agramkow.dk Syria leakdetection.service@inficon.com
Agramkow
Sonderborg
Phone: +45 741 236 36
Fax: +45 744 336 46
INFICON GmbH
Cologne
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
Poland kamola@vakpol.com Taiwan Susan.Chang@inficon.com
VAK-POL & GAZ Sp. zo.o
Pulawy
Phone: +48 60 23 15 212
Fax: +48 60 23 15 212
INFICON Company Limited
Chupei City, HsinChu Hsien
Phone: +886.3.5525.828
Fax: +886.3.5525.829
Portugal leakdetection.service@inficon.com Tunisia leakdetection.service@inficon.com
INFICON GmbH
Cologne
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
INFICON GmbH
Cologne
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
Republic of South Africa vacuquip@hotmail.com Turkey jhj@agramkow.dk
Vacuquip
Randburg
Phone: +27 73 15 78 355 Agramkow
Sonderborg
Phone: +45 741 236 36
Fax: +45 744 336 46
Russia leakdetection.service@inficon.com Ukraine leakdetection.service@inficon.com
INFICON GmbH
Cologne
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
INFICON GmbH
Cologne
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
Singapore reach.singapore@inficon.com United Kingdom reach.unitedkingdom@inficon.com
INFICON PTE LTD.
Singapur
Phone: +65.890.6250
Fax: +65.890.6266
INFICON
Blackburn
Phone: +44 1254 678 250
Fax: +44 1254 698 577
Slovakia filip.lisec@inficon.com United Arab Emirates leakdetection.service@inficon.com
INFICON GmbH
Pilsen
Phone +420 734 331 758
Fax: +420 604 203 037
INFICON GmbH
Cologne
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
Slovenia medivak@siol.net USA service.usa@inficon.com
Medivac
Ljubljani
Phone: +386 15 63 91 50
Fax: +386 17 22 04 51
Inficon Inc.
East Syracuse, NY
Phone: +1.315.434.1167
Fax: +1.315.434.2551
South America except Brazil infoqro@meisa.com Inficon Inc.
San Jose, CA
Phone: +1.408.361.1200
Fax: +1.408.362.1556
MEISA S.a. de C.V.
Querètaro
Phone: +52 44 22 12 36 15
Fax: +52 44 22 12 19 40
Inficon Inc.
Austin, TX
Phone: +1.512.448.0488
Fax: +1.512.448.0398
General Information 1-7
1.3 Unpacking
Unpack the Protec P3000 leak detector immediately after it has been received even
if it is to be put into operation at some later date. Examine the shipping container for
any external damage. Completely remove all packaging materials.
Notice Retain the shipping container and the packaging materials in the event of
possible complaints concerning any damages.
Check if the Protec P3000 leak detector is complete (see Section 1.3.1) and carefully
subject it to a visual inspection. If any damage is discovered please immediately
inform the forwarding agent and the insurers. If it is required to exchange the
damaged part please contact our or ders department.
1.3.1 Supplied Equipment
The Protec P3000 leak detector is ready for operation. Before installation please
read Section 1.5. Included with the leak detector are the following items:
Protec P3000 (main unit)
Mains cord, 3m long
Set of fuses (3 x 10 pcs.)
Spare air filter
8 mm hexagonal wrench
Documentation
Operating Instructions (kima26)
Technical Handbook Protec P3000 (kina26e1)
Spare Parts List Protec P3000 (kiua26d2)
Interface Description Protec P3000 (kins26e1)
Notice The SL3000 and SL3000XL sniffer lines are available in different
configurations and need to be ordered separately in the desired length. The
SL3000(XL) sniffer line is not part of the Protec P3000 shipment. (see
Section 1.3.2 Accessories)
Notice The PRO-Check reference leak is an accessory (see Section 1.3.2
Accessories) and needs to be ordered seperately
Notice For the RC versions the display unit and the co nnecting cable ar e not part
of the standard Protec P3000 shipment and need to be ordered separate ly
(see Section 1.3.2)
1-8 General Information
1.3.2 Accessories
For Protec P3000
For Protec P3000XL only
For RC versions only
SL3000 Sniffer line for Protec P3000 Cat. No. / Ref. No.
SL3000-3, 3 m length 525-001
SL3000-5, 5 m length 525-002
SL3000-10, 10 m length 525-003
SL3000-15, 15 m length 525-004
Sniffer tips for SL3000
ST 312, 120 mm, rigid 122 13
FT 312, 120 mm, flexible 122 14
FT 200, 200 mm, rigid 122 18
FT 250, 250 mm, flexible 122 66
ST 385, 385 mm, rigid 122 15
FT 385, 385 mm, flexible 122 16
FT 600, 600 mm, flexible 122 09
ST 500, 500 mm, rigid, 45° angled 122 75
Water protection tip for sniffer 122 46
SL3000XL Sniffer line for Protec P3000XL
SL3000XL-3, 3 m length 521-011
SL3000XL-5, 5 m length 521-012
SL3000XL-10, 10 m length 521-013
SL3000XL-15, 15 m length 521-014
Sniffer tips for SL3000XL
ST 312XL, 120 mm, rigid 122 80
FT 312XL, 120 mm, flexible 122 81
ST 385XL, 385 mm, rigid 122 82
FT 385XL, 385 mm, flexible 122 83
FT 250XL, 250 mm, flexible 122 85
External display unit
for benchtop use 551-100
for rack mounting 551-101
Connecting cable for external display unit
5m length 551-102
Connecting cable for external display unit
0.7m length 551-003
General Information 1-9
For all Protec P3000
configurations
1.4 Notes on How to Use This Handbook
Important remarks concerning operational safety and protection are emphasised as
follows:
Notice Indicates special requirements the user must comply with.
The references to diagrams, e.g. (2-1/6) consist of the Section No., Fig. No. and the
Item No. in that order. For example: (2-1/6) means Section 2, Fig. 1 and Item No. 6
(here: mains switch).
1.4.1 Symbols of Vacuum Technology
In the following some important symbols of vacuum technology as used in this
handbook are shown:
Holder for SL3000(XL) 525-006
PRO-Check reference leak for helium 521-001
Calibrated sniffer leaks for helium
S-TL 4, approx.1 x 10
-4
mbar l/s 122 37
S-TL 5, range 10
-5
mbar l/s 122 38
S-TL 6, range 10
-6
mbar l/s 122 39
Warning
Indicates procedures that must be strictly observed to prevent hazards to persons.
Caution
Indicates procedures that must strictly be observed to prevent damage to or
destruction of the Protec P3000 leak detector.
Diaphragm pump Vacuum gauge
1-10 General Information
1.4.2 Definition of Terms
Main menu
This menu is shown first after operating the Menu push-button.
Sub-menus
Comprise all menus which may be accessed from the main menu. Unauthorised
changes to many of these sub-menus may be prevented by a password (see also
Section 4.3.1).
Menu item
A single menu line.
Default condition
Status of the Protec P3000 when supplied from the factory.
Service menu
Comprises the menu lines in the “Service” sub-menu. The service menu is accessed
by scrolling in the basic menu using the navigation push-b uttons (see also Section
3.2).
Autozero
Determination and compensation of the helium background. With this function, the
internal ZERO level of the leak rate signal is determined in order to avoid a readout
of the internal helium background and mistaking it as a properly measured value. If
subsequently negative leak rates are obtained due to this correction , the stored offset
values are changed so that ZERO will be the lowest value which can be obtained. In
this way the values adapt automatically to a decaying background (adaptive
background correction).
Internal background
The existing partial pressure in the measurement system. The level of the internal
background is measured all the time and subtracted from the measured signal.
I•Guide Mode
In the I•Guide Mode different testing plans can be pre-programmed. During testing
the operator is then constantly prompted for the next action and thus guid ed through
the testing plan.
Unit under test
Test object that needs to be leak checked.
Display limit
Limits the measurement data displayed depending on the unit of measurement and
the operator settings.
General Information 1-11
1.5 Instrument Views of the Protec P3000
Fig. 1-2 Instrument views of Protec P3000
Pos. Description Pos. Description
1 Main display 4 Lemo Connector for SL3000 sniffer line
2 Speaker 5 Handle fore carrying the Protec P3000
3 PRO-Check reference leak
1
2
3
4
5
1-12 General Information
1.6 Installation
1.6.1 Set up
How to remove the transportation lock:
The transportation lock is located on the bottom side of the Protec P3000 and
consists of a yellow knurled screws. Please remove this screw before starting-up the
leak detector. The Protec P3000 is supplied ready for operation. Initial start-up is
described in Section 3.1.
Fig. 1-3 Removing the transportation lock before starting
Caution
Before installation remove the transportation lock.
Caution
In order to ensure adequate ventilation of the Protec P3000, a space of at least
20 cm 8) (8 in.) must be kept unobstructed to the sides. The clearance at the rear
must be no less than 10 cm (4 in.). Moreover, the Protec P3000 handles for carrying
the leak detector at the sides of the main unit must not be covered at any time as
these acts as air inlet and outlet. Avoid the presence of heat sources in the vicinity
of the Protec P3000.
General Information 1-13
1.6.2 Mechanical Connections
SL3000(XL) Sniffer line
In order to operate the Protec P3000 it is essential for the SL3000(XL) sniffer line to
be connected. The connection for the sn iffe r line is loca te d at the fro nt of the Prot ec
P3000 left of the PRO-Check reference leak. Insert th e plug into the opening with the
red dot on the plug and the slot in the front cover aligned until the connector engage s.
To disconnect the plug, retract the coupling and remove the probe’s line.
Water protection tip
(optional)
If you intend to perform leak testing on parts that are not completely dry (e.g. due to
condensation after performance testing), we strongly recommend to use a water
protection tip.
To install the water protection tip,
1 screw off the metallic capillary filter at the very end of the sniffer tip and
2 install the water protection tip instead.
Notice Please do not forget to re-install the little rubber se al when switching to the
water protection tip.
Caution
Twisting the sniffer line spoils the cable.
Do not twist the sniffer line.
Fig. 1-4 Installing water protection tip
1-14 General Information
PRO-Check Reference
leak (optional)
Please insert the PRO-Check reference leak into the opening in the housing of the
main unit. Make sure that the Sub-D plug is pr operly connected with the PRO-Ch eck
leak.
Notice When properly inserted, the PRO-Check reference leak will still protrude by
approx. 10 mm.
On first usage of your PRO-Check you need to initialize the use of this reference leak
in the Protec P3000 software.
Please perform the following steps:
1 Insert the PRO-Check into the appropriate opening of the Protec P3000
2 In the software menu go to HISTORY & MAINTENANCE / REPLACE PRO-CHECK.
Notice This menu item is only available when the Protec P3000 is set to
ADVANCED
MODE (see Section 4.4.3, Keyword: User Mode)
3 On the certificate, which is delivered with the PRO-Check, you will find a serial
number and a 12-digit-code. Enter the serial number in the first line of the open
submenu and the 12-digit-code in the sec on d line an d press OK.
Notice The PRO-Check reference leak must be installed in the Protec P3000
when pressing
OK.
Notice PRO-Check Warntime Expire Date (See Section 7.7.5).
For RC versions only
The RC version has no built-in display unit but a connectors plate is mounted instead .
Please connect the external display unit with the 5m connecting table (Cat.-no. 551-
002)
Fig. 1-5 Initializing the PRO-Check reference leak
General Information 1-15
Holder for SL3000
sniffer line (optional)
An optional holder for the SL3000 sniffer line is available as cat.-no. 525-006. The
holder may be installed on the right or left sid e of the main unit (for right- or left
handed operators) as shown in Fig. 1-8.
The installation is described in Fig. 1-8. There are two little slots on the front side at
the very top area of the blue square front of the main unit. Ho ld the holder horizontally
and then insert the two little hooks of the holder into the two slots (either on the right
or the left side). With the hooks still inserted, let the holder flap down. It will
automatically attach to the metal front by the magnet on the backside of the holder.
Now insert the sniffer probe grip into the opening of the holder and let it sink down
until it rests in the holder.
Fig. 1-6 Protec P3000 with external display unit for: (a) benchtop use (left side), (b)
rack mounting (rigth side)
Fig. 1-7 Usage of sniffer line holder
1-16 General Information
1.6.3 Electri cal Conn ectio ns
Notice The local regulations for electrical connections must always be observed
(in Germany VDE 0100). The mains voltage rating for the Protec P3000
can be read off from the name plate left of the power switch. The mains
voltage setting of the Protec P3000 is fixed and can not be changed. A
separate fuse for each of the mains conductors has been inte g ra te d into
the mains socket (Fig. 1-9/6).
The mains voltage is applied to the Protec P3000 via the detachable mains cable
which is supplied with the leak detector. A main power socket is available for this
purpose at the rear of the main unit.
Fig. 1-8 Installation of sniffer line holder
Fig. 1-9 Electrical connections
Pos. Description Pos. Description
1 Headphone port 4 Name plate
2 I/O Port 5 Power switch
3 RS232 interface 6 Power connector
Caution
Before connecting the Protec P3000 to the mains you must make sure that the
mains voltage rating of the Protec P3000 coincides with the locally available mains
voltage.
1
23 456
General Information 1-17
1.6.4 RS232 Interface
The Protec P3000 is equipped with a RS232 interface which is located on the rear
right side of the main unit. This interface is of the DCE type (Data Communications
Equipment) and allows the connection of a PC for monitoring and data logging. The
connection is provided through a commercially available Sub-D plug. For further
information see “Interface Description Protec P3000” (kins26e1).
1.6.5 I/O Port
The I/O port allows communication with and control through external equipment via
PLC and analog data. For details see Section 6.1.
Through this connection some functions of the Protec P3000 can be controlled
externally or measurement data or the status of the Protec P3000 may be
communicated to external equipment.
Through relay changeover contacts the trigger levels as well as the operating mode
(Ready) of the Protec P3000 may be monitored.
Warning
Only 3-core mains cables having a protective ground conductor must be used.
Operation of the Protec P3000 with the ground conductor unconnected is not
permissible.
2-1 How the Protec P3000 Works
2 How the Protec P3000 Works
2.1 Description of the Functions
The Protec P3000 takes in helium through the SL3000 sniffer line, detects the
amount of helium by means of a helium sensitive sensor and converts the sensor
signal into quantitative leak rates.
The Protec P3000 is composed of the following principal subassemblies:
A Wise Technology sensor as helium detector
A valve holder for controlling different operating stati
A diaphragm pump for pumping the gas to the sensor
An inlet system for the gas flow
The correspond ing electrical and electronic subassemblies for supplying power
and for signal conditioning.
The detector operates under gross vacuum conditions, i. e. the operating pressure
at the detector is several 100 mbar. T he under pressure is generated by a diaphragm
pump. The pressure in front of the sensor is measured with a piezo-resistant
pressure gauge and is about 250 mbar under normal operating conditions.
2.2 Description of the Subassemblies
2.2.1 Backing Pump
A diaphragm pump in the Protec P3000 serves as the backing pump. All data and
further information on this pump are given in the Operating Instructions. The backing
pump generates the flow of the gas through the SL3000 sniffer line.
2.2.2 Wise Technology
TM
Sensor
The helium detector (Wise Technology
TM
sensor) consists of a closed glass
container with a measurement device for the precise determination of the pressure
inside the glass housing and a mem brane chip wit h a large nu mber of small quartz
windows. The membrane is permeable only for helium, all other components of air
are retained by the membrane and from the inside of the glass housing. The Quartz
membrane is heated so that the permeation for helium is sufficiently high and fast.
Inside of the glass housing the total pressure is measured precisely. As only helium
can enter the glass housing the total pressure is equal to the partial pressure of
helium. The determined total pressure inside the housing is proportional to the
helium partial pressure outside the sensor.
How the Protec P3000 Works 2-2
2.2.3 Valve Holder
The valve holder carries the valves that control the gas flow to the Wise Technology
sensor. These valves are used to select the sensitivity of the system, to activate a
protection mode against high helium contamination and to set the system into
STANDBY mode. The Protec P3000 software continuously analyses the situtation and
sets the correct valve position via the control unit.
2.2.4 Control Assembly
The control assembly (microprocessor) is the central assembly of the Protec P3000’s
electronics. All other subassemblies are controlled and monitored by this assembly.
The microprocessor which is located here is thus continuously informed about the
status of the entire Protec P3000 and can respond accordingly. In order to accept
commands from the operator and to output measured values and messages, the
control subassembly is linked to the display unit.
2.3 Description of the Displays and User
Interfaces
2.3.1 Main unit display
This subassembly is used to communicate with the operator. It accepts commands
from the 8 keys on both sides of the display and outputs measurement results and
messages via the display.
Fig. 2-1 Main unit display
Pos.Description Pos. Description
1 Menu buttons 1 to 4 2 Menu buttons 5 to 8
2-3 How the Protec P3000 Works
For RC versions only
The RC version has a connectors plate for the external display unit instead of the
built-in main unit display. Two LED's left of the plug provide information about the
status of the Protec P3000, even when the external display unit is disconnected:
The green LED will indicate that the Protec P3000 is in operation (switched on).
The green LED will show continuous green light if an external display unit is
connected and will blink if no external display unit is detected.
The red LED will be blinking in case of an error message, continuous red light
indicates a warning.
If no display unit is connected, error messages or warnings may be acknowledged
by pressing both buttons of the SL3000 sniffer line simultaneously.
The external display unit also offers four buttons:
The START / STOP buttons have no function (the external displa y unit may also
be used with other INFICON leak detectors which need these buttons)
The MENU button will open the software menu.
The ZERO button will set the current background reading to zero. (For details on
the ZERO function see Section 4.4.1)
Fig. 2-2 Conncetor plate with LEDs