INFICON Micro GC Fusion Gas Analyzer User Manual

0 (0)

PRE-INSTALLATION CHECKLIST

PN 074-615-P1A

 

 

Micro GC Fusion

Pre-Installation Checlist

Site Preparation

Proper site preparation prior to installing Micro GC Fusion is very important. Contact INFICON (www.inficon.com) with any questions regarding site preparation.

NOTE: Checkboxes () identify hardware and actions that must be addressed when installing Micro GC Fusion.

List of Required Tools

1 1/8" wrench recommended, or adjustable large wrench (for attaching regulator to gas tank)

Teflon® tape for regulator connections

NOTE: The following tools are included in the optional Install Kit (PN 952-021-G1).

Open-ended wrench, 1/4 in. and 5/16 in.

Open-ended wrench, 7/16 in. and 9/16 in.

Open-ended wrench, 3/8 in. and 7/16 in.

Hex driver, 2.5 mm ball-point long shank

Tubing cutter

Carrier Gas

. . . .Carrier gas (e.g., helium, argon) purity of 99.999%, or minimum 99.995% when the appropriate carrier gas trap is installed. Contact INFICON for carrier gas trap assistance.

NOTE: A pressurized [>16,547 kPa (>2,400 psi)] 50 liter tank of carrier gas will last approximately one year or more.

. . . .Dual Stage Regulator for pressurized carrier gas tank with outlet pressure regulated to 400-427 kPa (58-62 psi). A Swagelok 1/8 in. fitting is preferred on the regulator outlet to connect the carrier gas feed tube to Micro GC Fusion.

NOTE: In North America, use the CGA 580 inlet connection (PN 952-416-P1) for helium, nitrogen and argon carrier gas. Use the CGA 350 inlet connection (PN 952-415-P1) for hydrogen carrier gas. For installations outside of North America, use the appropriate inlet connections as standard for the region.

CAUTION

Exceeding 427 kPa (62 psi) carrier gas pressure may result in instrument damage.

1 of 5

INFICON Micro GC Fusion Gas Analyzer User Manual

PRE-INSTALLATION CHECKLIST

PN 074-615-P1A

 

 

Micro GC Fusion

Pre-Installation Checlist

. . . .Moisture/hydrocarbon trap (PN G2870A-01) required for the carrier gas when using Molsieve or Alumina columns.

CAUTION

To protect Micro GC Fusion and to enhance performance, carrier gas traps should be used to purify the carrier gas before introduction into Micro GC Fusion.

A hydrocarbon/moisture trap (PN G2870A-01) is usually sufficient.

. . . .Copper or stainless steel tubing (pre-cleaned) 1/8 in. OD for carrier gas connections between the tank regulator output and Micro GC Fusion carrier gas inlet. Micro GC Fusion has up to two 1/8 in. Swagelok male carrier gas inlet connections.

NOTE: 3 m (10 ft.) of pre-cleaned GC grade copper tubing is included with Micro GC Fusion. Additional copper tubing is available in the installation kit (PN 952-021-G1), or can be ordered in 15 m (50 ft.) lengths

(PN 5180-4196).

NOTE: Do not clean copper or stainless steel tubing with solvent. NOTE: Do not use Teflon® or Tygon® tubing for carrier gas.

. . . .Assortment of male and female Swagelok fittings for regulator, tubing, etc., including nuts and front and back ferrules. Typical size is 1/8 in. but adapters for 1/4 in. to 1/8 in. may be needed if 1/4 in. tubing is used.

. . . .1/8 in. tee fitting and 1/8 in. ball valve, depending on connection requirements.

Calibration Gas

. . . .Supply of calibration gas specific to the analysis from a commercial gas supplier. NOTE: Calibration gas must be available at the time of installation.

NOTE: Calibration gas must closely resemble the composition of the actual process gas.

. . . .A regulator, if necessary, to reduce calibration gas pressure below

69 kPa (10 psi) at the Micro GC Fusion inlets. The regulator must be clean, grease-free and non-venting.

NOTE: Checkout gas mixtures are only used for instrument installation and start up.

NOTE: Checkout gas mixtures supplied by INFICON are not calibration gases.

2 of 5

Loading...
+ 3 hidden pages