Tektronix TDS3052B, TDS3064B, TDS3054B, TDS3044B, TDS3034B Service Manual

...
4.8 (4)

Service Manual

TDS3000B Series

Digital Phosphor Oscilloscopes

071-0972-04

This document supports firmware version 3.00 and above for TDS3000B Series instruments only.

Warning

The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.

Copyright © Tektronix, Inc. All rights reserved.

Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.

Printed in the U.S.A.

Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077

TEKTRONIX, TEK, TEKPROBE, and Tek Secure are registered trademarks of Tektronix, Inc.

DPX, WaveAlert, OpenChoice, and e*Scope are trademarks of Tektronix, Inc.

Table of Contents

General Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

vii

Service Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

xi

Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

xiii

Contacting Tektronix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

xiv

Specifications

 

Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-1

Operating Information

 

Connecting Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-2

Installing an Application Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-7

Installing a Communication Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-8

Front-Panel Menus and Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-9

Front-Panel Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-18

Rear-Panel Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-19

Communication Module Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-20

Probe Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-21

Self Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-22

Transporting the Oscilloscope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-23

Theory of Operation

 

Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-1

Battery Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-2

Main Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-2

Display Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-3

Front-Panel Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-3

Disk Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-3

Performance Verification

 

Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-2

Performance Verification Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-5

Adjustment Procedures

 

Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-1

Overview of the Adjustment Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-2

Signal Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-3

Factory Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-6

TDS3000B Series Service Manual

i

Table of Contents

Maintenance

Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-1

Preventing ESD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-1

Inspection and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-2

Removal and Installation Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-5

Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-39

Troubleshooting Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-40

Unpacking and Repacking Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-43

Options

 

Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7-1

Electrical Parts List

 

Electrical Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8-1

Diagrams

 

Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

9-1

Mechanical Parts List

 

Parts Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10-1

Using the Replaceable Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10-2

ii

TDS3000B Series Service Manual

Table of Contents

List of Figures

Figure 3-1: TDS3000B series block diagram . . . . . . . . . . . . . . . . . . .

3-1

Figure 6-1: Pulling the hub pin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-8

Figure 6-2: Releasing the hub assembly . . . . . . . . . . . . . . . . . . . . . . .

6-9

Figure 6-3: Reinstalling a hub cover . . . . . . . . . . . . . . . . . . . . . . . . . .

6-11

Figure 6-4: Communication module guide . . . . . . . . . . . . . . . . . . . . .

6-12

Figure 6-5: Main board disk drive flex cable connector J800 . . . . .

6-13

Figure 6-6: Removing the bezel from a new disk drive . . . . . . . . . . .

6-14

Figure 6-7: Rear chassis cables to the main board . . . . . . . . . . . . . .

6-16

Figure 6-8: Installing the fan, line filter, and line power cable . . . . .

6-18

Figure 6-9: Installing the AC line filter and shield . . . . . . . . . . . . . .

6-19

Figure 6-10: Installing the line filter and line power cable . . . . . . . .

6-20

Figure 6-11: Removing the J500 connector clip . . . . . . . . . . . . . . . . .

6-22

Figure 6-12: Disconnect main board cable connectors . . . . . . . . . . .

6-23

Figure 6-13: Main Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-24

Figure 6-14: Install the J500 connector clip . . . . . . . . . . . . . . . . . . . .

6-26

Figure 6-15: Front chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-27

Figure 6-16: Display module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-28

Figure 6-17: Front-panel cable routing . . . . . . . . . . . . . . . . . . . . . . . .

6-29

Figure 6-18: Disconnect back light cables . . . . . . . . . . . . . . . . . . . . .

6-30

Figure 6-19: Remove inverter board . . . . . . . . . . . . . . . . . . . . . . . . . .

6-31

Figure 6-20: Remove display from display chassis . . . . . . . . . . . . . .

6-32

Figure 6-21: Remove back light tube . . . . . . . . . . . . . . . . . . . . . . . . .

6-33

Figure 6-22: Installing display glass shield . . . . . . . . . . . . . . . . . . . . .

6-34

Figure 6-23: Remove front-panel knobs . . . . . . . . . . . . . . . . . . . . . . .

6-35

Figure 6-24: Remove front-panel board . . . . . . . . . . . . . . . . . . . . . . .

6-36

Figure 6-25: Remove front-panel keypads . . . . . . . . . . . . . . . . . . . . .

6-37

Figure 6-26: Checking front-panel lock nuts . . . . . . . . . . . . . . . . . . .

6-38

Figure 6-27: Primary troubleshooting procedure . . . . . . . . . . . . . . .

6-40

Figure 6-28: AC power supply troubleshooting procedure . . . . . . . .

6-41

Figure 6-29: Module isolation troubleshooting procedure . . . . . . . .

6-42

TDS3000B Series Service Manual

iii

Table of Contents

Figure 10-1: Exploded view, front case half . . . . . . . . . . . . . . . . . . . .

10-4

Figure 10-2: Exploded view, LCD display . . . . . . . . . . . . . . . . . . . . .

10-6

Figure 10-3: Exploded view, main board and disk drive . . . . . . . . .

10-8

Figure 10-4: Exploded view, circuit board assembly and fan . . . . . .

10-10

Figure 10-5: Exploded view, rear case half . . . . . . . . . . . . . . . . . . . . .

10-13

iv

TDS3000B Series Service Manual

Table of Contents

List of Tables

Table 1-1: Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-1

Table 6-1: Internal inspection check list . . . . . . . . . . . . . . . . . . . . . .

6-3

Table 6-2: Removal and installation procedures . . . . . . . . . . . . . . . .

6-6

Table 7-1: TDS3000B Series options . . . . . . . . . . . . . . . . . . . . . . . . .

7-1

TDS3000B Series Service Manual

v

Table of Contents

vi

TDS3000B Series Service Manual

General Safety Summary

 

Review the following safety precautions to avoid injury and prevent damage to

 

this product or any products connected to it. To avoid potential hazards, use this

 

product only as specified.

To Avoid Fire or Personal

Use Proper Power Cord. Use only the power cord specified for this product and

Injury

certified for the country of use.

 

Connect and Disconnect Properly. Connect the probe output to the measurement

 

instrument before connecting the probe to the circuit under test. Disconnect the

 

probe input and the probe ground from the circuit under test before disconnecting

 

the probe from the measurement instrument.

 

Ground the Product. When operating with AC power, this product is grounded

 

through the grounding conductor of the power cord. To avoid electric shock, the

 

grounding conductor must be connected to earth ground. Before making

 

connections to the input or output terminals of the product, ensure that the

 

product is properly grounded.

 

When operating with battery power, this product must still be grounded. To

 

prevent electric shock, always connect a grounding wire between the ground

 

terminal on the rear panel and earth ground.

 

Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings

 

and markings on the product. Consult the product manual for further ratings

 

information before making connections to the product.

 

Connect the ground lead of the probe to earth ground only.

TDS3000B Series Service Manual

vii

General Safety Summary

Replace Batteries Properly. Replace batteries only with the proper type and rating specified.

Recharge Batteries Properly. Recharge batteries for the recommended charge cycle only.

Do Not Operate Without Covers. Do not operate this product with covers or panels removed.

Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present.

Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel.

Do Not Operate in Wet/Damp Conditions.

Do Not Operate in an Explosive Atmosphere.

Keep Product Surfaces Clean and Dry.

Provide Proper Ventilation. Refer to the manual’s installation instructions for details on installing the product so it has proper ventilation.

Safety Terms and Symbols Terms in this Manual. These terms may appear in this manual:

WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.

CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.

Terms on the Product. These terms may appear on the product:

DANGER indicates an injury hazard immediately accessible as you read the marking.

WARNING indicates an injury hazard not immediately accessible as you read the marking.

CAUTION indicates a hazard to property including the product.

viii

TDS3000B Series Service Manual

General Safety Summary

Symbols on the Product. These symbols may appear on the product:

WARNING

Protective Ground

CAUTION

Battery

High Voltage

(Earth) Terminal

Refer to Manual

Information

Ethernet Port

Chassis Ground

 

 

Preventing Electrostatic Damage

CAUTION. Electrostatic discharge (ESD) can damage components in the oscilloscope and its accessories. To prevent ESD, observe these precautions when directed to do so.

Use a Ground Strap. Wear a grounded antistatic wrist strap to discharge the static voltage from your body while installing or removing sensitive components.

Use a Safe Work Area. Do not use any devices capable of generating or holding a static charge in the work area where you install or remove sensitive components. Avoid handling sensitive components in areas that have a floor or benchtop surface capable of generating a static charge.

Handle Components Carefully. Do not slide sensitive components over any surface. Do not touch exposed connector pins. Handle sensitive components as little as possible.

Transport and Store Carefully. Transport and store sensitive components in a static-protected bag or container.

TDS3000B Series Service Manual

ix

General Safety Summary

Product End-of-Life Observe the following guidelines when recycling an instrument or component:

Handling

Mercury Notification. This oscilloscope uses an LCD backlight lamp that contains mercury. Disposal may be regulated due to environmental considerations. Please contact your local authorities or the Electronics Industries Alliance (www.eiae.org) for disposal or recycling information.

Battery Recycling. This product may contain a Nickel Cadmium (NiCd) or lithium ion (Li-ion) rechargeable battery, which must be recycled or disposed of properly. Please properly dispose or recycle the battery according to local government regulations.

Transportation and The TDS3BATB lithium ion rechargeable battery pack contains less than 8 Lithium Ion Batteries grams of lithium, with individual cells containing less than 1.5 grams of lithium,

as measured by International Civil Aviation Organization (ICAO) standards. Consult your air carrier for applicability and determination of any special lithium ion battery transportation requirements.

x

TDS3000B Series Service Manual

Service Safety Summary

Only qualified personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary before performing any service procedures.

Do Not Service Alone. Do not perform internal service or adjustments of this product unless another person capable of rendering first aid and resuscitation is present.

Disconnect Power. To avoid electric shock, switch off the instrument power, then disconnect the power cord from the mains power.

Use Care When Servicing With Power On. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.

To avoid electric shock, do not touch exposed connections.

TDS3000B Series Service Manual

xi

Service Safety Summary

xii

TDS3000B Series Service Manual

Preface

This Service manual provides information to troubleshoot, disassemble, and repair the following TDS3000B Series Digital Phosphor Oscilloscopes instrument to the module level:

HTDS3012B

HTDS3014B

HTDS3024B

HTDS3032B

HTDS3034B

HTDS3044B

HTDS3052B

HTDS3054B

HTDS3064B

TDS3000B Series Service Manual

xiii

Preface

Contacting Tektronix

Phone

1-800-833-9200*

Address

Tektronix, Inc.

 

Department or name (if known)

 

14200 SW Karl Braun Drive

 

P.O. Box 500

 

Beaverton, OR 97077

 

USA

Web site

www.tektronix.com

Sales support

1-800-833-9200, select option 1*

Service support

1-800-833-9200, select option 2*

Technical support

www.tektronix.com/support

 

1-800-833-9200, select option 3*

 

6:00 a.m. - 5:00 p.m. Pacific Standard time

*This phone number is toll free in North America. After office hours, please leave a voice mail message.

Outside North America, contact a Tektronix sales office or distributor; see the Tektronix web site for a list of offices.

xiv

TDS3000B Series Service Manual

Specifications

Specifications

This appendix contains specifications for the TDS3000B Series oscilloscopes. All specifications are guaranteed unless noted as “typical.” Typical specifications are provided for your convenience but are not guaranteed. Specifications that are marked with the n symbol are checked in Performance Verification.

All specifications apply to all TDS3000B models unless noted otherwise.

To meet specifications, two conditions must first be met:

HThe oscilloscope must have been operating continuously for twenty minutes within the operating temperature range specified.

HYou must perform the Compensate Signal Path operation described on page 2-22 prior to evaluating specifications. If the operating temperature changes by more than 10 °C, you must perform the Compensate Signal Path operation again.

Table 1- 1: Specifications

Acquisition

Acquisition modes

Sample (Normal), Peak detect, Envelope, and Average

 

 

 

Single Sequence

Acquisition mode

Acquisition stops after

 

 

 

 

Sample, Peak Detect

One acquisition, all channels

 

 

simultaneously

 

 

 

 

Average, Envelope

N acquisitions, all channels

 

 

simultaneously, N is settable from

 

 

2 to 256 (or ∞ for Envelope)

TDS3000B Series Service Manual

1- 1

Specifications

Table 1- 1: Specifications (cont.)

Inputs

Input coupling

DC, AC, or GND

 

 

Channel input remains terminated when using GND coupling.

Input impedance,

1 MΩ ±1% in parallel with 13 pF ±2 pF, TekProbe compatible

DC coupled

50 Ω ±1%; VSWR ≤ 1.5:1 from DC to 500 MHz, typical

 

Maximum voltage at

Overvoltage category

Maximum voltage

input BNC (1 MΩ)

 

 

CAT I Environment

150 VRMS (400 Vpk)

 

 

(refer to page 1--11)

 

 

 

 

 

CAT II Environment

100 VRMS (400 Vpk)

 

(refer to page 1--11)

 

 

 

 

 

For steady-state sinusoidal waveforms, derate at 20 dB/decade above

 

200 kHz to 13 Vpk at 3 MHz and above.

Maximum voltage at

5 VRMS with peaks ≤ ±30 V

 

input BNC (50 Ω)

 

 

 

 

Maximum floating

0 V from chassis (BNC) ground to earth ground, or

voltage

 

 

 

30 VRMS (42 Vpk) only under these conditions: no signal voltages

 

>30 VRMS (>42 Vpk), all common leads connected to the same voltage,

 

no grounded peripherals attached

 

 

 

Channel-to-channel

Measured on one channel, with test signal applied to another channel,

crosstalk, typical

and with the same scale and coupling settings on each channel

 

 

 

 

Frequency range

Crosstalk

 

 

 

 

≤ 100 MHz

≥ 100:1

 

 

 

 

≤ 200 MHz

≥ 50:1

 

 

 

 

≤ 300 MHz

≥ 50:1

 

 

 

 

≤ 400 MHz

≥ 30:1

 

≤ 500 MHz

≥ 30:1

 

 

 

 

≤ 600 MHz

≥ 30:1

 

 

 

Differential delay,

100 ps between any two channels with the same scale and coupling

typical

settings

 

 

 

 

1- 2

TDS3000B Series Service Manual

Specifications

Table 1- 1: Specifications (cont.)

Vertical

 

 

 

 

 

 

 

 

 

 

 

 

 

Number of channels

TDS30x2B

 

TDS30x4B

 

 

 

 

 

 

 

 

 

2 plus external trigger input

4 plus external trigger input

 

 

 

 

 

Digitizers

9-bit resolution, separate digitizers for each channel sample

 

 

simultaneously

 

 

 

 

 

 

 

 

 

 

 

SCALE Range

1 MΩ

 

50 Ω

 

(at BNC)

 

 

 

 

 

 

1 mV/div to 10 V/div

1mV/div to 1 V/div

 

 

 

 

Fine SCALE

Adjustable with ≥ 1% resolution

 

 

 

 

 

 

 

 

 

 

Polarity

Normal and Invert

 

 

 

 

 

 

 

 

 

 

 

Position range

±5 divisions

 

 

 

 

 

 

 

 

n Analog

Bandwidth limit set to Full, operating ambient ≤30 °C, derate 1%/°C

bandwidth, 50 Ω

above 30 °C

 

 

 

 

(also typical at 1 MΩ

 

 

 

 

 

 

Scale range

 

5 mV/div to

2 mV/div to

 

1 mV/div to

with standard probe)

 

 

1 V/div

4.98 mV/div

 

1.99 mV/div

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TDS301xB

 

100 MHz

100 MHz

 

90 MHz

 

 

 

 

 

 

 

 

 

 

TDS302xB

 

200 MHz

100 MHz

 

90 MHz

 

 

TDS303xB

 

300 MHz

250 MHz

 

150 MHz

 

 

 

 

 

 

 

 

 

 

TDS304xB

 

400 MHz

250 MHz

 

150 MHz

 

 

 

 

 

 

 

 

 

 

TDS305xB

 

500 MHz

300 MHz

 

175 MHz

 

 

 

 

 

 

 

 

 

 

TDS306xB

 

600 MHz1

300 MHz

 

175 MHz

 

Calculated rise

TDS301xB

 

3.5 ns

 

 

time, typical

 

 

 

 

 

 

 

TDS302xB

 

1.75 ns

 

 

 

 

 

 

 

 

 

 

 

 

 

TDS303xB

 

1.2 ns

 

 

 

 

 

 

 

 

 

TDS304xB

 

0.9 ns

 

 

 

 

 

 

 

 

 

TDS305xB

 

0.7 ns

 

 

 

 

 

 

 

 

 

TDS306xB

 

0.6 ns

 

 

 

 

 

 

 

 

 

1

500 MHz at 5 mV/div; 600 MHz at 10 mV/div and greater.

 

 

 

TDS3000B Series Service Manual

1- 3

Specifications

Table 1- 1: Specifications (cont.)

Vertical

Analog bandwidth

Selectable between 20 MHz, 150 MHz (not available on TDS301xB), or

limit, typical

Full

 

 

 

 

 

 

 

 

Lower frequency limit,

7 Hz for 1 MΩ, reduced by a factor of ten when using a 10X passive

AC coupled, typical

probe; 140 kHz for 50 Ω

 

 

 

 

 

Peak detect or

Minimum width of pulse with amplitude of ≥2 div to capture 50% or

Envelope pulse

greater amplitude

 

 

 

 

 

response, typical

 

 

 

 

 

 

Sample rates ≤125 MS/s

 

 

Sample rates ≥250 MS/s

 

 

 

 

 

 

 

 

 

 

 

 

1 ns

 

 

1/sample rate

 

 

 

DC gain accuracy

± 2%, derated at 0.025%/°C for temperatures below +18 °C and above

 

 

+30 °C, in Sample or Average acquisition mode

 

 

 

 

 

DC measurement

Measurement type

 

 

DC Accuracy (in volts)

accuracy

 

 

 

 

 

 

 

 

 

 

 

Sample acquisition

Absolute measurement of any

± [0.022 ¢ | reading -- (offset --

 

mode, typical

waveform point

 

 

position) | + offset accuracy +

 

 

 

 

 

0.15 div + 0.6 mV]

 

 

 

 

 

 

Delta voltage between any two

± [0.022 ¢ | reading | + 0.15 div

 

 

points on a waveform

 

 

+ 1.2 mV]

 

 

 

 

 

n Average

Absolute measurement of any

± [0.022 ¢ | reading -- (offset --

 

acquisition mode

waveform point

 

 

position) | + offset accuracy

 

(≥16 averages)

 

 

 

+ 0.1 div]

 

 

 

 

 

 

Delta voltage between two points

± [0.022 ¢ | reading | + 0.05 div]

 

 

on a waveform

 

 

 

 

 

 

 

 

 

 

Offset range

Scale range

 

 

Offset range

 

 

 

 

 

 

 

 

1 mV/div to 9.95 mV/div

 

 

±100 mV

 

 

10 mV/div to 99.5 mV/div

 

 

±1 V

 

 

 

 

 

 

 

 

 

 

 

100 mV/div to 995 mV/div

 

 

±10 V

 

 

 

 

 

 

 

 

 

1V/div to 10 V/div

 

 

±100 V

 

 

 

 

 

Offset accuracy, all

± [0.005 ¢ | offset -- position | + 0.1 div]

 

 

ranges

 

 

 

 

 

 

 

 

Note: Convert both the constant offset and position terms to volts by

 

 

multiplying by the volts/div setting

 

 

 

2

0.02 term (gain component) derated at 0.00025/

˚

C above 30

˚

C

 

 

 

1- 4

TDS3000B Series Service Manual

Tektronix TDS3052B, TDS3064B, TDS3054B, TDS3044B, TDS3034B Service Manual

Specifications

Table 1- 1: Specifications (cont.)

Horizontal

 

 

 

 

 

 

 

 

 

Acquisition

Normal

 

Fast trigger

(horizontal) resolution

(10,000 point record)

(500 point record)

 

 

 

 

 

Maximum

700 wfms/s

 

3,600 wfms/s

acq rate, typical

 

 

 

 

 

 

 

 

 

Sample rate range

TDS301xB

TDS302xB,

TDS304xB--

 

 

 

TDS303xB

TDS306xB

 

 

 

 

 

 

Normal

100 S/s to

100 S/s to

100 S/s to

 

1.25 GS/s

2.5 GS/s

5 GS/s

 

 

 

 

 

 

 

 

Fast trigger

5 S/s to

5 S/s to

5 S/s to

 

 

1.25 GS/s

2.5 GS/s

5 GS/s

 

 

 

 

 

 

Seconds/division

4 ns/div to

2 ns/div to

1 ns/div to

 

range

10 s/div

10 s/div

10 s/div

 

 

 

 

 

 

n Sample rate and

±20 ppm over any ≥1 ms time interval

delay time accuracy

 

 

 

 

Trigger

 

 

 

 

 

 

 

 

 

External trigger input,

TDS30x2B:

 

 

 

typical

1 MΩ in parallel with 17 pF, TekProbe compatible

 

TDS30x4B:

 

 

 

 

1 MΩ in parallel with 52 pF, not TekProbe compatible

External trigger

Overvoltage category

Maximum voltage

maximum voltage

 

 

 

 

CAT I Environment

150 VRMS (400 Vpk)

 

 

(refer to page 1--11)

 

 

 

 

 

 

CAT II Environment

100 VRMS (400 Vpk)

 

(refer to page 1--11)

 

 

 

 

 

 

For steady-state sinusoidal waveforms, derate at 20 dB/decade above

 

200 kHz to 13 Vpk at 3 MHz and above

External trigger maxi-

0 V from chassis (BNC) ground to earth ground

mum floating voltage

 

 

 

 

 

30 VRMS (42 Vpk) only under these conditions: no signal voltages

 

>30 VRMS (>42 Vpk), all common leads connected to the same voltage,

 

no grounded peripherals attached

 

 

 

 

 

 

n Edge trigger

Source

 

Sensitivity

sensitivity

 

 

 

 

Any channel, DC coupled

≤0.6 div from DC to 50 MHz,

 

 

 

 

increasing to 1 div at oscilloscope

 

 

 

bandwidth

Edge trigger

Source

 

Sensitivity

sensitivity, typical

 

 

 

 

External trigger

 

200 mV from DC to 50 MHz,

 

 

 

 

 

increasing to 750 mV at 300 MHz

 

 

 

 

External/10 trigger

500 mV from DC to 50 MHz,

 

 

 

increasing to 3 V at 300 MHz

 

 

 

 

 

TDS3000B Series Service Manual

1- 5

Specifications

Table 1- 1: Specifications (cont.)

Trigger

 

Any channel, NOISE REJ coupled

3.5 times the DC-coupled limits

 

 

 

 

Any channel, HF REJ coupled

1.5 times the DC-coupled limit

 

 

from DC to 30 kHz, attenuates

 

 

signals above 30 kHz

 

Any channel, LF REJ coupled

1.5 times the DC-coupled limits

 

 

for frequencies above 80 kHz,

 

 

attenuates signals below 80 kHz

Trigger level range

Source

Sensitivity

 

 

 

 

Any channel

±8 divisions from center of

 

 

screen, ±8 divisions from 0 V if LF

 

 

REJ trigger coupled

 

External trigger

±800 mV

 

 

 

 

External/10 trigger

±8 V

 

 

 

 

Line

Fixed at the midlevel of the AC

 

 

line

 

 

 

SET LEVEL TO 50%,

Operates with input signals ≥45 Hz

 

typical

 

 

 

 

 

Trigger level

Source

Sensitivity

accuracy, typical

 

 

Any channel

±0.2 divisions

 

 

 

 

 

External trigger

±20 mV

 

External/10 trigger

±200 mV

 

 

 

 

Line

N/A

 

 

 

Trigger holdoff range

250.8 ns to 10 s

 

 

 

Video trigger

Triggers on negative sync of NTSC, PAL, or SECAM signal

sensitivity, typical

Source

Sensitivity

 

 

 

 

 

Any channel

0.6 to 2.5 divisions of video sync

 

 

tip

 

 

 

 

External trigger

150 mV to 625 mV of video sync

 

 

tip

 

 

 

 

External/10 trigger

1.5 V to 6.25 V of video sync tip

 

 

 

1- 6

TDS3000B Series Service Manual

Specifications

Table 1- 1: Specifications (cont.)

Trigger

 

 

 

 

 

B Trigger

Trigger After Time

Trigger After B Events

 

 

 

Range

13.2 ns to 50 s

1 event to 9,999,999 events

 

 

 

Minimum time

5ns from the end of the time

5 ns between the A trigger event

between arm and

period and the B trigger event

and the first B trigger event

trigger, typical

 

 

 

 

 

Minimum Pulse

B event width, 2 ns

Width, typical

 

 

 

 

 

Maximum Fre-

B event frequency, 250 MHz

quency, typical

 

 

Display

 

 

 

 

Display screen

6.5 in (165 mm) diagonal color liquid crystal

 

 

 

Display resolution

640 horizontal by 480 vertical pixels

 

Backlight intensity,

200 cd/m2

 

typical

 

 

 

 

 

Display color

Up to 16 colors, fixed palette

 

External display filter

Scratch-resistant tempered glass

 

I/O ports

 

 

 

 

 

Ethernet port

RJ-45 female connector 10BaseT

 

 

 

Parallel printer port

DB-25 female connector, Centronics compatible

GPIB interface

Available with optional accessory TDS3GV

 

 

RS-232 interface

DB-9 male connector, available with optional accessory TDS3GV

 

 

VGA signal output

DB-15 female connector, 31.6 kHz sync rate, EIA RS-343A compliant,

 

available with optional accessory TDS3GV

 

 

 

Probe compensator

5.0 V into ≥1 MΩ load,

 

output, typical

frequency = 1 kHz

 

Miscellaneous

 

 

 

 

Nonvolatile memory

Typical retention time ≥ 5 years for front-panel settings, unlimited for

 

saved waveforms and setups

 

 

 

Floppy disk

3.5 in, DOS format, 720 KB or 1.44 MB compatible

 

 

Internal clock

Provides date/time stamp for stored data and the current time and date

 

to the display, if enabled

 

 

 

 

TDS3000B Series Service Manual

1- 7

Specifications

Table 1- 1: Specifications (cont.)

Power sources

AC line power

Operates the oscilloscope and charges the optional internal battery

 

 

Source voltage

90 VRMS to 250 VRMS, continuous range

Source frequency

47 Hz to 440 Hz

Power consump-

75 W maximum

tion

 

 

 

Battery power

Optional accessory TDS3BATB, rechargeable battery pack

Operating time,

3 hours, depending on operating conditions

typical

 

 

 

Battery charge

30 hours in the oscilloscope, 5 hours using the optional external

time, typical

charger TDS3CHG

 

 

Line fuse

Internal, not user replaceable

Environmental

 

 

 

Temperature

Operating range (no disk installed): +5 °C to +50 °C

 

Nonoperating range (no disk installed): --20 °C to +60 °C

 

Typical operating range for floppy disks: +10 °C to +50 °C

 

 

Humidity

Operating range (no disk installed): 20% to 80% RH below 32 °C,

 

derate upper limit to 21% RH at 50 °C

 

Nonoperating range (no disk installed): 5% to 90% RH below 41 °C,

 

derate upper limit to 30% RH at 60 °C

 

Typical operating range for floppy disks: 20% to 80% RH below 32 °C,

 

derate upper limit to 21% RH at 50 °C

 

 

Pollution Degree

Pollution Degree 2: Typical home or office environment.

1- 8

TDS3000B Series Service Manual

Specifications

Table 1- 1: Specifications (cont.)

Environmental

Altitude

Operating limit: 3000 m

 

Nonoperating limit: 15,000 m

Random vibration

Operating: 0.31 gRMS from 5 Hz to 500 Hz, 10 minutes on each axis

 

Nonoperating: 2.46 gRMS from 5 Hz to 500 Hz, 10 minutes on each

 

axis

 

 

Drop resistance,

Survives a 152 mm (6 in) drop onto concrete with only cosmetic

typical

damage

Mechanical

 

 

 

Size

Height: 176 mm (6.9 in), 229 mm (9.0 in) including handle

 

Width: 375 mm (14.75 in)

 

Depth: 149 mm (5.9 in)

Weight

Oscilloscope only: 3.2 kg (7.0 lbs)

 

With accessories and carry case: 4.1 kg (9.0 lbs)

 

When packaged for domestic shipment: 5.5 kg (12.0 lbs)

 

Optional TDS3BATB battery pack: 0.85 kg (1.9 lbs)

 

 

TDS3000B Series Service Manual

1- 9

Specifications

Table 1- 1: Specifications (cont.)

EMC certifications and compliances

EMC Compliance:

Meets the intent of Directive 89/336/EEC for Electromagnetic

European Union

Compatibility. Compliance was demonstrated to the following

 

specifications as listed in the Official Journal of the European

 

Communities:

 

EN 61326

 

EMC requirements for Class A electrical equipment for

 

measurement, control, and laboratory use 1,2

 

IEC 61000-4-2

 

Electrostatic discharge immunity (Performance criterion B)

 

IEC 61000-4-3

 

RF electromagnetic field immunity (Performance

 

criterion B)3

 

IEC 61000-4-4

 

Electrical fast transient/burst immunity (Performance

 

criterion B)

 

IEC 61000-4-5

 

Power line surge immunity (Performance criterion B)

 

IEC 61000-4-6

 

Conducted RF immunity (Performance criterion B)4

 

IEC 61000-4-11

 

Voltage dips and interruptions immunity (Performance

 

criterion B)

 

EN 61000-3-2

 

AC power line harmonic emissions

 

EN 61000-3-3

 

Voltage changes, fluctuations and flicker

 

 

1Emissions that exceed the levels required by this standard may occur when this equipment is connected to a test object.

2To ensure compliance to the standards listed above, attach only high quality shielded cables to this instrument. High quality shielded cables typically are braid and foil types that have low impedance connections to shielded connectors at both ends.

3The increase in trace noise while subjected to a test field (3 V/m over the frequency range 80 MHz to 1 GHz, with 80% amplitude modulation at 1 kHz) is not to exceed 8 major divisions peak-to-peak. Ambient conducted fields may induce triggering when the trigger threshold is offset less than 4 major divisions from ground reference.

4The increase in trace noise while subjected to a test field (3 V/m over the frequency range 150 kHz to 80 MHz, with 80% amplitude modulation at 1 kHz) is not to exceed

2 major divisions peak-to-peak. Ambient conducted fields may induce triggering when the trigger threshold is offset less than 1 major division from ground reference.

1- 10

TDS3000B Series Service Manual

Specifications

Table 1- 1: Specifications (cont.)

EMC certifications and compliances

EMC Compliance:

Meets the intent of Australian EMC Framework as demonstrated to the

Australia/New Zea-

following specification:

land

AS/NZS 2064.1/2

 

 

 

EMC Compliance:

This product was certified by the GOST ministry of Russia to be in

Russia

compliance with all applicable EMC regulations.

 

 

FCC Compliance:

Emissions comply with FCC Code of Federal Regulations 47, Part 15,

U.S.A.

Subpart B, Class A Limits

Safety certifications and compliances

 

 

EC Declaration of

Compliance was demonstrated to the following specification as listed in

Conformity -- Low

the Official Journal of the European Communities:

Voltage

Low Voltage Directive 73/23/EEC as ammended by 93/68/EEC

(TDS3000B Series,

EN 61010-1/A2:1995

P3010, and P6139A)

 

Safety requirements for electrical equipment for

 

 

measurement, control, and laboratory use

(P3010 and P6139A)

EN 61010-2-031:1995

 

 

Particular requirements for hand-held probe assemblies

 

 

for electrical measurement and test equipment

Approvals

 

 

(TDS3000B Series,

UL3111-1 -- Standard for electrical measuring and test equipment

P3010, and P6139A)

CAN/CSA C22.2 No. 1010.1 -- Safety requirements for electrical

 

equipment for measurement, control and laboratory use

(P3010 and P6139A)

EN 61010-2-031:1995 -- Particular requirements for hand-held probe

 

assemblies for electrical measurement and test equipment

 

 

Installation/

Terminals on this product may have different installation category

Overvoltage Category

designations. The installation categories are:

Descriptions

 

 

 

CAT III

Distribution-level mains (usually permanently

 

 

connected). Equipment at this level is typically in a

 

 

fixed industrial location

 

CAT II

Local-level mains (wall sockets). Equipment at this

 

 

level includes appliances, portable tools, and similar

 

 

products. Equipment is usually cord-connected

 

CAT I

Secondary (signal level) or battery operated circuits of

 

 

electronic equipment

 

 

 

TDS3000B Series Service Manual

1- 11

Specifications

1- 12

TDS3000B Series Service Manual