Philips Avalon FM20, Avalon FM30 Service manual

4.5 (24)

Obstetrical Care

S E R V I C E G U I D E

Avalon Fetal Monitor

FM20

FM30

F E T A L M O N I T O R I N G

Printed in Germany 07/05

*M2703-9000A*

Part Number M2703-9000A

4512 610 10021

S

Table of Contents

1

Introduction

1

 

Who Should Read This Guide

1

 

What to Do Next

1

 

Repair Strategy

2

 

Manufacturer’s Information

2

 

Passwords

3

 

Warnings and Cautions

3

2

Site Preparation

5

 

 

 

 

Introduction

5

 

Site Planning

5

 

Roles and Responsibilities

5

 

Site Preparation Responsibilities

5

 

Procedures for Local Staff

6

 

Procedures for Philips Personnel

7

 

Site Requirements

7

 

Space Requirements

7

 

Environmental Requirements

7

 

Safety Requirements (Customer or Philips)

8

 

Electrical Requirements (Customer or Philips)

8

 

Connecting Non-Medical Devices

8

 

Cabling Options and Requirements for Connection to OB TraceVue

9

 

Mounting Options

9

 

Input Devices

10

3

Installation Instructions

11

 

 

 

 

Initial Inspection

11

 

Mechanical Inspection

11

 

Electrical Inspection

11

 

Claims for Damage

11

 

Repackaging for Shipment or Storage

12

 

Mounting Instructions

12

 

Line Voltage Selection

12

 

Connecting the Monitor to AC Mains

12

 

Connecting the Monitor to Non-Medical Devices

13

 

Checking and Setting Line Frequency

13

 

Checking/Setting Paper Scale

13

 

Checking/Setting Paper Speed

13

 

PS/2 Keyboard/Mouse

14

i

4

Theory of Operation

15

 

Monitor Hardware Overview

15

 

Power Supply

16

 

Connector Block

16

 

Bus Master Board

16

 

Main CPU Board

16

 

Fetal Recorder (Thermal Printer Unit)

16

 

Recorder Adapter Board

17

 

Thermal Line Printhead (TLPH)

17

 

Paper Sensor

17

 

Stepper Motor

17

 

LCD Display and Touchscreen

17

 

Noninvasive Blood Pressure Assembly

17

 

SpO2 Assembly

17

 

Input/Output Interface Boards

18

 

Transducer Hardware Overview

18

 

Transducer Types

19

 

Functional Description of the Transducer CPU

19

 

CPU (Micro Controller)

19

 

Analog-to-Digital Converter

19

 

Communication Transceiver (CAN Bus Driver)

19

 

EEPROM

19

 

Toco Transducer Frontend

19

 

Ultrasound Transducer Frontend

20

 

Toco+ Transducer Frontends

20

 

Toco Frontend

20

 

IUP Frontend

20

 

ECG Frontend

20

5

Interfaces

21

 

 

 

 

LAN / RS232 Interface

21

 

Dual PS/2 Interface

21

6

Testing and Maintenance

23

 

 

 

 

Recommended Frequency

23

 

When to Perform Test Blocks

24

 

Test Reporting

25

 

How to Carry Out the Test Blocks

25

 

Preventive Maintenance Procedures

26

 

Other Regular Tests

26

 

Visual Check

26

 

Fetal Recorder Maintenance

26

 

Testing Transducers and Patient Modules

26

 

Ultrasound Transducer Electrical Check

26

 

Toco Transducer Electrical Check

28

ii

Testing the Patient Module (M2738A)/Toco+ Transducer (M2735A): DECG Mode

28

Testing the Patient Module (M2738A)/Toco+ Transducer (M2735A): MECG Mode

29

Testing the Patient Module (M2738A)/Toco+ Transducer (M2735A): IUP Mode

30

Performance Assurance Tests

31

Noninvasive Blood Pressure Performance Tests

31

Accuracy Test

31

Leakage Test

32

Linearity Test

32

Valve Test

32

SpO2 Performance Test

32

Safety Tests

33

Warnings, Cautions, and Safety Precautions

33

Electrical Isolation Diagram

34

Safety Test Procedures

34

S(1): Sum of Functional Earth and Enclosure Leakage Current Test

34

S(2): Patient Leakage Current - Single Fault Condition (SFC), Mains on Applied Part

35

System Test

36

What is a Medical Electrical System?

36

General Requirements for a System

36

System Example

36

Touchscreen Calibration

37

Disabling/Enabling Touch Operation

38

Checking the Fetal Recorder Offset

39

Setting the Fetal Recorder Offset

39

Fetal Recorder Selftest Report

40

7 Troubleshooting

41

 

 

Who Should Perform Repairs

41

Replacement Level Supported

41

Hardware Revision Check

41

Software Revision Check

42

Obtaining Replacement Parts

42

Troubleshooting Guide

42

Checks for Obvious Problems

42

Checks Before Opening the Instrument

42

Checks with the Instrument Switched On, AC connected

43

Individual Parameter INOPs

43

Initial Instrument Boot Phase

44

Troubleshooting Tables

44

How to Use the Troubleshooting Tables

44

Boot Phase Failures

45

Screen is Blank

46

Touchscreen Not Functioning

46

General Monitor INOP Messages

47

Keyboard/Mouse Not Functioning

48

Alarm Tones

48

iii

 

Alarm Behavior

48

 

Fetal Recorder

48

 

LAN / RS232

51

 

Transducers

52

 

Status Log

53

 

Troubleshooting with the Support Tool

54

 

Troubleshooting the Individual Measurements or Applications

54

8

Disassembly and Reassembly

55

 

 

 

 

Introduction

55

 

Serial Numbers

56

 

Removing the Top Cover Assembly

57

 

Removing the Display Assembly

59

 

Replacing the Display Assembly

62

 

Recorder Disassembly

64

 

Removing the Drawer Assembly

64

 

Replacing the Drawer Assembly

66

 

Removing the Recorder Chassis

68

 

Removing the Thermal Line Printhead (TLPH)

72

 

Replacing the TLPH

73

 

Replacing the Recorder Chassis

74

 

Removing the Paper Sensor Assembly

75

 

Replacing the Paper Sensor Assembly

75

 

Removing the Recorder Adapter Board

75

 

Replacing the Recorder Adapter Board

77

 

Removing the Stepper Motor

78

 

Replacing the Stepper Motor

79

 

Replacing the Top Cover Assembly

80

 

Removing the Power Supply Assembly

81

 

Replacing the Power Supply Assembly

82

 

Removing the Noninvasive Blood Pressure Assembly

82

 

Replacing the Noninvasive Blood Pressure Assembly

84

 

Removing the SpO2 Assembly

85

 

Replacing the SpO2 Assembly

86

 

Removing the Interface Boards

87

 

Removing the Main CPU Board

88

 

Replacing the Main CPU Board

89

 

Exchanging the Loudspeaker

89

 

Exchanging the Transducer Cable

90

 

Exchanging the Transducer Belt Button

92

9

Parts

95

 

 

 

 

Monitor

95

 

Transducers

96

 

Patient Modules

97

 

Mounting Hardware

97

iv

Assembly and Kit Contents

98

Bottom Housing Assembly (M2703-64101)

98

Power Supply Assembly (M2703-60001)

99

Top Cover Assembly (M2703-60052)

99

Top Cover Housing (M2703-64102)

100

Stepper Motor Assembly (M2703-60004)

100

Paper Sensor Assembly (M2703-60003)

100

Drawer Assembly (M2703-64651)

100

Display Assembly (M2703-64503)

101

Transducer Cable Assembly (M2735-64201)

101

Small Parts Kit - Top (M2735-64202)

102

Small Parts Kit - Bottom (M2703-64203)

102

Belt Button Kit (M2703-64204)

102

10 Upgrades

105

 

 

11 Understanding Configuration

107

 

 

What is Configuration Mode?

107

Understanding Settings

108

Entering and Leaving Configuration Mode

108

Storing Changes in the User Defaults

109

Loading the Factory Default

109

Loading the User Defaults

110

Loading Configurations Using the Support Tool

110

About Configuration Files (.cfg)

111

Selecting the Correct Configuration

111

12 Configuration Settings Appendix

113

 

 

Documenting Monitor Configurations

113

Using the Configuration Tables

113

Configuration Table Example

114

Understanding Configuration Implications

114

Measurement-Related Settings

115

Color Configuration

115

Configuring FHR (Ultrasound)

115

FHR Configuration Implications

115

Configuring Toco

116

Configuring IUP

116

Configuring DFHR (DECG)

116

DFHR Configuration Implications

116

Configuring MHR (ECG)

117

ECG Configuration Implications

117

Configuring Pulse

118

Configuring SpO2

119

SpO2 Configuration Implications

119

Configuring Noninvasive Blood Pressure (NBP)

119

v

NBP Configuration Implications

120

Monitor-Related Settings

121

Configuring Alarms

121

Alarm Settings Configuration Implications

121

Configuring the NST Timer

122

NST Timer Configuration Implications

122

Configuring Fetal Recorder Settings

122

Recorder Configuration Implications

123

Configuring User Interface Settings

123

User Interface Configuration Implications

124

Hardware Settings

124

Global Settings

125

Global Settings Configuration Implications

125

vi

1

Introduction

This Service Guide contains technical details for the Avalon FM20 and FM30 Fetal/Maternal Monitors. It provides a technical foundation to support effective troubleshooting and repair. It is not a comprehensive, in-depth explanation of the product architecture or technical implementation. It offers enough information on the functions and operations of the monitoring systems so that engineers who repair them are better able to understand how they work. It covers the physiological measurements and the monitor hardware that acquires and displays them.

The Avalon FM20/FM30 Fetal Monitor Service Guide supplements the maintenance and troubleshooting procedures, carried out by the operator, that are described in the Instructions for Use. Refer to the Instructions for Use for maintenance and troubleshooting procedures that may be performed during normal operation.

Only qualified service personnel should attempt to install the system, disassemble the monitor, remove or replace any internal assemblies, or replace the transducer cable or belt buttons.

Who Should Read This Guide

This guide is for biomedical engineers or technicians responsible for troubleshooting, repairing, and maintaining Philips’ Avalon fetal monitors.

You must:

understand English

be familiar with standard medical equipment installation procedures

be familiar with current conventional technical terms as used throughout this guide

What to Do Next

Familiarize yourself with the contents of this guide and the Instructions for Use before attempting to service or repair the system.

1

1 Introduction

Repair Strategy

Repair Strategy

The Service Support Tool software helps you to determine whether a fault is a hardware or software problem. The main replaceable parts are:

unit exchange for the transducers

replacement of

the top cover assembly

the bottom housing

the power supply assembly

the display assembly

the recorder adapter board

the paper drawer assembly

the paper sensor assembly

the stepper motor assembly

the SpO2 assembly

the noninvasive blood pressure assembly

the main CPU board

the bus master board

the socket connector block

the transducer cable

the transducer belt button

See Chapter 9, “Parts” for part numbers, and Chapter 8, “Disassembly and Reassembly” for repair details.

Repair or replacement of individual components on the boards is not supported, and should never be attempted.

For tests that you are required to perform after repairs, refer to “When to Perform Test Blocks” on page 24.

Manufacturer’s Information

© Copyright 2003 - 2005. Koninklijke Philips Electronics N.V.

All Rights Reserved.

Philips Medizin Systeme Böblingen GmbH

Hewlett-Packard-Str. 2

71034 Böblingen, Germany

2

Passwords

1 Introduction

Passwords

In order to access different modes within the monitor a password may be required. The passwords are listed below.

Monitoring Mode: No password required

Configuration Mode: 71034

Demo Mode: 14432

Service Mode: 1345

Refer to Chapter 11, “Understanding Configuration” before making any changes to the monitor configuration.

Warnings and Cautions

In this guide:

A warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to observe a warning may result in death or serious injury to the user or patient.

A caution alerts you where special care is necessary for the safe and effective use of the product. Failure to observe a caution may result in minor or moderate personal injury or damage to the product or other property, and possibly in a remote risk of more serious injury.

3

1 Introduction

Warnings and Cautions

4

2

Site Preparation

Introduction

This section describes the procedures you should follow to plan and prepare a site for an Avalon FM20/FM30 fetal monitor installation.

Site planning.

Roles and responsibilities for local and Philips personnel.

Site Planning

The careful planning of the site for the FM20/FM30 monitor is essential for its safe and efficient operation. A consulting schedule should be established between the Customer and Philips Sales and Support Representatives, to ensure that all preparations are completed when the system is delivered.

The site planning phases prior to equipment installation are:

Location: Planning the location of the various system components.

Environment: Confirming and correcting, as necessary, the environment of the proposed installation site(s).

System Capabilities: Explaining the possibilities for system expansion.

Mounting: Referencing the mounting hardware information website for the listing of suitable mounting hardware recommended for use with the various system components, and all details on the available mounts and accessories.

Cabling: Identifying the requirements for the cabling, conduiting and faceplates for connecting the various system components.

Roles and Responsibilities

This section describes the procedures necessary to prepare a site for a system installation. The procedures are grouped into two parts: procedures that local staff or contractors are responsible for, and procedures that Philips personnel are responsible for.

Site Preparation Responsibilities

Local Staff

Ensure that all safety, environmental and power requirements are met.

Provide power outlets.

Prepare mounts, and consult Philips for detailed mounting requirements.

5

2 Site Preparation

Introduction

• Pull cables, install conduit, install wallboxes.

Philips Personnel

Provide the customer with the safety, environmental and power requirements.

Assemble mounts, as necessary.

Provide requirements for cabling.

Procedures for Local Staff

The following tasks must be completed before the procedures for Philips personnel may be started.

Providing Power Outlets

Provide a power outlet in the vicinity (1 m or 3 ft) or any peripheral equipment.

WARNING Only the power cables provided with the system may be used. For reasons of safety, power (mains) extension cables or adapters shall not be used.

Preparing Mounts

Where ceiling, wall, or shelf mounts are required for mounting the equipment, the customer is responsible for the following:

Providing and installing all hardware which is required to install the mounting hardware supplied by Philips as detailed in the installation notes.

Making sure that all ceilings, walls, and mounting rails that supports mounting hardware are suitable for their proposed load.

WARNING It is the customer's responsibility to have the attachment of the mounting hardware to the ceiling, wall, or mounting rail and the construction of the ceiling, wall, or mounting rail evaluated for structural integrity and compliance with all local, state and any other required codes by a registered, professional, structural and/or mechanical engineer.

Although considerable effort has been made to ensure the safety of the ceiling mount installation and or mounting guidelines, it is to be understood that the installation itself is beyond the control of Philips Medical Systems. Accordingly, Philips Medical Systems will not be responsible for the failure of any such installation.

Providing Conduit

Providing conduit and/or trunking of a sufficient cross-sectional area for the planned cables and possible future expansion (for additional components or systems).

Providing and/or installing suitable wall boxes to accommodate the faceplates.

Pulling Cables

WARNING NEVER run power cables through the same conduit or trunking used for system cables.

• Installing Wall Boxes

6

Site Requirements

2 Site Preparation

It is the customer's responsibility to provide and install wallboxes to house faceplates. The customer must notify the Philips installation coordinator of which size is to be used.

Procedures for Philips Personnel

Before you begin the procedures in the installation sections, ensure that the customer has completed all necessary preparations outlined in the previous section, “Procedures for Local Staff.”

Site Requirements

The site requirements are listed in this section.

Space Requirements

The situating of the monitor should be planned such that the nursing staff are able to monitor the patient with relative ease, with all patient connectors and controls readily available and the displays clearly visible. The location should also allow access to service personnel without excessive disruption and should have sufficient clearance all round to allow air circulation.

Dimensions and weight:

Monitor:

Size (W x H x D): 335 x 286 x 133 mm (13.2 x 11.3 x 5.2 in)

Weight; 5.1 kg (11.2 lb)

Transducer:

Size (diameter): 83 mm (3.27 in)

Weight (without cable): 190g (6.7 oz.)

Environmental Requirements

The environment where the FM20/FM30 monitor will be used should be reasonably free from vibration, dust and corrosive or explosive gases. The ambient operating and storage conditions for the FM20/FM30 monitor must be observed. If these conditions are not met, the accuracy of the system will be affected and damage can occur.

.

Monitor (M2702A/M2703A)

Temperature Range

Operating

0°C to 45°C (32°F to 113°F)

 

Storage

-20°C to 60°C (-4°F to 140°F)

Humidity Range

Operating

<95% relative humidity @ 40°C/104°F

 

Storage

<90% relative humidity @ 60°C/140°F

Altitude Range

Operating

-500 to 3000 m/-1640 to 9840 ft.

 

Storage

-500 to 3000 m/-1640 to 9840 ft.

 

 

 

Transducers (M2734A/M2735A/M2736A)

Temperature Range

Operating

0°C to 40°C (32°F to 104°F)

 

Storage

-40°C to 60°C (-40°F to 140°F)

 

 

 

7

2 Site Preparation

Site Requirements

Transducers (M2734A/M2735A/M2736A)

Humidity Range

Operating

<95% relative humidity @ 40°C/104°F

 

Storage

<90% relative humidity @ 60°C/140°F

Altitude Range

Operating

-500 to 3000 m/-1640 to 9840 ft.

 

Storage

-500 to 3000 m/-1640 to 9840 ft.

 

 

 

SpO2 Sensors

Operating Temperature Range

0°C to 37°C (32°F to 98.6°F)

 

 

Safety Requirements (Customer or Philips)

The monitor is an electrical Class II device in which the protection against electric shock does not rely on basic insulation and a protective earth conductor but on double and/or reinforced insulation.

WARNING • Always use the supplied power cord with the earthed mains plug to connect the monitor to an earthed AC mains socket. Never adapt the mains plug from the power supply to fit an unearthed AC mains socket.

The protective earth conductor is required for EMC purposes. It has no protective function against electric shock! The protection against electric shock in this device is provided by double and/or reinforced insulation.

Do not use additional AC mains extension cords or multiple portable socket-outlets. If a multiple portable socket-outlet without an approved isolation transformer is used, the interruption of its protective earthing may result in enclosure leakage currents equal to the sum of the individual earth leakage currents, so exceeding allowable limits.

Electrical Requirements (Customer or Philips)

Line Voltage Connection

The FM20/FM30 monitor uses < 60 W.

Line Voltage: the FM20/FM30 monitor may be operated on ac line voltage ranges of 100 to 240V (50/60 Hz).

Connecting Non-Medical Devices

The standard IEC-60601-1-1 applies to any combination of devices, where at least one is a medical device. Therefore IEC-60601-1-1 must still be met after all devices are connected.

8

Site Requirements 2 Site Preparation

WARNING • Do not use a device in the patient vicinity if it does not comply with IEC-60601-1. The whole installation, including devices outside of the patient vicinity, must comply with IEC-60601-1-1. Any non-medical device, including a PC running an OB TraceVue system, placed and operated in the patient’s vicinity must be powered via an approved isolation transformer that ensures mechanical fixing of the power cords and covering of any unused power outlets.

• Do not connect any devices that are not supported as part of a system.

Whenever you combine equipment to form a system, for example, connecting the monitor to an OB TraceVue system, perform a system test according to IEC 60601-1-1 (see “System Test” on page 36).

Figure 1 Equipment Location in the Patient Vicinity

Cabling Options and Requirements for Connection to OB TraceVue

For cabling options and requirements for connection to an OB TraceVue system, refer to the OB

TraceVue Site Preparation Guide and the OB TraceVue Service Guide.

Mounting Options

See “Mounting Hardware” on page 97 for a list of mounting options. Refer to “Mounting Instructions” on page 12, or contact your local Philips representative for advice on mounting the monitor.

9

2 Site Preparation

Site Requirements

Input Devices

The following tables describes the input devices which can be connected to the monitor via the optional PS/2 interface.

Product Option

Part Number

12NC Part

Description

Number

 

Number

 

 

 

 

 

M8024A #A01

862454

9898 031 24741

Slimline Keyboard with integrated Trackball

M8024A #B01

M4046-60104

4512 610 00661

Optical Mouse USB / PS/2

M8024A #C01

M4046-60103

4512 610 00651

Wired Track Ball USB / PS2

M8024A #C02

M4046-60105

4512 610 00671

Wireless Track Ball

M8024A #C03

M4046-60106

4512 610 00681

Wired off table Track Mouse

 

 

 

 

10

3

Installation Instructions

The information contained in this chapter, in addition to that given in the Instructions for Use, should enable the monitor to be installed ready for use (the preparation and planning should be adhered to as specified in the “Site Preparation” chapter). Safety checks and inspection procedures for mounts are explained in the “Testing and Maintenance” chapter, and configuration of the system is explained in the “Configuration” chapter.

Please keep the packing materials until you have completed the initial inspection, in case there is a defect on arrival.

Initial Inspection

Inspect the delivery on arrival.

Mechanical Inspection

Open the shipping container(s) and examine each part of the instrument for visible damage, such as broken connectors or controls, or scratches on the equipment surfaces. If the shipping carton/container is undamaged, check the cushioning material and note any signs of severe stress as an indication of rough handling in transit. This may be necessary to support claims for hidden damage that may only become apparent during subsequent testing.

Electrical Inspection

The instrument has undergone extensive testing prior to shipment. Safety testing at installation is not required (except in situations where devices are interconnected forming a system, see “Connecting Non-Medical Devices” on page 143). An extensive self check may be performed. This recommendation does not supersede local requirements.

All tests are described in the “Testing and Maintenance” chapter of this manual.

Claims for Damage

When the equipment is received, if physical damage is evident or if the monitor does not meet the specified operational requirements of the patient safety checks or the extended self check, notify the carrier and the nearest Philips Sales/Support Office at once. Philips will arrange for immediate repair or replacement of the instrument without waiting for the claim settlement by the carrier.

11

3 Installation Instructions

Repackaging for Shipment or Storage

Repackaging for Shipment or Storage

If the instrument is to be shipped to a Philips Sales/Support Office, securely attach a label showing the name and address of the owner, the instrument model and serial numbers, and the repair required (or symptoms of the fault). If available and reusable, the original Philips packaging should be used to provide adequate protection during transit. If the original Philips packaging is not available or reusable please contact the Philips Sales/Support Office who will provide information about adequate packaging materials and methods.

Mounting Instructions

Every type of compatible mounting solution is delivered with a complete set of mounting hardware and instructions. Refer to the Site prep chapter for a list of mounting options. Refer to the documentation delivered with the mounting hardware for instructions on assembling mounts.

WARNING It is the customer's responsibility to have the attachment of the mounting hardware to the ceiling, wall, or mounting rail and the construction of the ceiling, wall, or mounting rail evaluated for structural integrity and compliance with all local, state and any other required codes by a registered, professional, structural and/or mechanical engineer.

Ensure that this commitment has been met before assembling mounts.

Line Voltage Selection

You do not need to set the line voltage, as this is done automatically by the power supply. The monitor has a wide-range power supply that allows you to operate the monitor from an AC (alternating current) power source of 100 V to 240 V (± 10%) and 50 to 60 Hz (± 5%).

Connecting the Monitor to AC Mains

The monitor is an electrical Class II device in which the protection against electric shock does not rely on basic insulation and a protective earth conductor but on double and/or reinforced insulation.

WARNING • Always use the supplied power cord with the earthed mains plug to connect the monitor to an earthed AC mains socket. Never adapt the mains plug from the power supply to fit an unearthed AC mains socket.

The protective earth conductor is required for EMC purposes. It has no protective function against electric shock! The protection against electric shock in this device is provided by double and/or reinforced insulation.

Do not use AC mains extension cords or multiple portable socket-outlets.

12

Connecting the Monitor to Non-Medical Devices

3 Installation Instructions

Connecting the Monitor to Non-Medical Devices

Connect the monitor to an obstetrical surveillance system, such as OB TraceVue, via the optional system interface. For cabling requirements, refer to “Cabling Options and Requirements for Connection to OB TraceVue” on page 9. For safety-related information, refer to “Connecting NonMedical Devices” on page 8, and “System Test” on page 36.

Checking and Setting Line Frequency

Before using the monitor, check that the line frequency setting is correct for your location, and change the setting if necessary in Configuration Mode.

WARNING An incorrect line frequency setting can affect the ECG filter, and disturb the ECG measurement. Ensure the line frequency setting is correct.

To set the line frequency:

1Enter the Main Setup menu.

2Select Global Settings.

3Select Line Frequency and select 50Hz or 60Hz from the pop-up list.

Checking/Setting Paper Scale

You can check the paper Scale Type (US for paper with a scale of 30-240, or Europe for paper with a scale of 50-210) in the Fetal Recorder menu. In Monitoring Mode, you can see this setting (grayed out), but you cannot change it. It can be changed in Configuration Mode.

1Enter the Main Setup menu by selecting the SmartKey .

2Select Fetal Recorder.

3Check the current setting for Scale Type. If it is not appropriate, change it in the Fetal Recorder menu in Configuration Mode:

Select Scale Type to toggle between US and Europe.

Checking/Setting Paper Speed

You can choose a paper speed of 1, 2, or 3 centimeters per minute (cm/min). The default setting is 3 cm/min.

As a change in paper speed results in a change in the appearance of a FHR trace, you are advised to ensure ALL monitors in your institution are set to the same speed.

To set the paper speed:

1Enter the Main Setup menu using the SmartKey .

2Select Fetal Recorder.

3In the Recorder menu, you can see the current speed setting. Select Recorder Speed.

4Select the desired speed from the given choices: 1, 2 or 3 cm/min.

13

3 Installation Instructions

Checking and Setting Line Frequency

PS/2 Keyboard/Mouse

Switch off the monitor before connecting any PS/2 compatible device.

Connect the PS/2 connector to the PS/2 Interface board in the monitor at the slot indicated by the appropriate symbol.

The default keyboard language setting for all initial configurations is “US”. However, the monitor will not automatically select the best matching language for the keyboard depending on the language of its software. This means that there is no such choice as “Automatic” for the keyboard language.

To configure the keyboard language manually, go to Service Mode, select Main Setup -> Hardware -> Keyboard and then select the proper language. Please note that this setting does not clone.

14

4

Theory of Operation

This chapter describes the functional operation of the monitor and the transducers. It incorporates features of the mechanical design, indicating the physical relationship of the assemblies and components.

Monitor Hardware Overview

Connector Block

Power

 

Recorder

Thermal

Main CPU

Adapter

Printhead

Supply

 

 

 

Paper

 

 

 

Sensor

 

 

 

Stepper

 

 

 

Motor

Bus Master

 

 

 

 

 

 

Touch

 

 

 

Screen

 

 

Display

Display

 

 

Adapter

 

 

Panel

 

 

 

 

 

 

Backlight

 

 

 

Converter

Fetal Recorder

Display Assembly

SpO2

IF Board 1

 

 

 

 

 

 

 

 

NIBP

 

 

 

 

 

IF Board 2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

optional boards

standard boards

15

4 Theory of Operation

Power Supply

The monitor consists of nine main functional components:

Power supply M2703-60001

Connector Block (1253-8415)

Bus Master Board (M2703-66460)

Main CPU Board (M2703-66450)

Fetal Recorder (Thermal Printer Unit)

Display Assembly (M2703-64503)

Noninvasive Blood Pressure Board (optional, M2703-64502)

SpO2 Board (optional, M2703-66453)

Input /Output Interface Boards (optional):

LAN / RS232 (M2703-67501)

Dual PS/2 (M8086-67501)

Power Supply

The power supply is a wide-range input switching unit, with an output of 24V. It is located in the bottom housing assembly.

Connector Block

Any compatible fetal transducer, patient module or remote event marker can be connected in any order to the monitor via the sockets on the Connector Block. The Connector Block is located on the Bus Master Board, and is exchangeable.

Bus Master Board

The signals from the transducers or sensors are conveyed from the sensor sockets on the Connector Carrier Board (M2703-66421) to the Bus Master Board (M2703-66420).

The Bus Master Board is responsible for transducer detection, communicates with the connected transducers via a CAN bus, and communicates parameter data to the Main CPU Board via a serial link for further processing and display.

Main CPU Board

The Main CPU Board controls the monitor’s human interface, and is responsible for the final processing of data from the Bus Master Board. It sends this data to the TFT display, and to the thermal printer unit for recording traces and other patient data. It also controls the optional LAN/RS232 and PS/2 interface boards.

Fetal Recorder (Thermal Printer Unit)

The fetal recorder is located in the Top Cover Assembly. The recorder consists of the following major parts:

16

LCD Display and Touchscreen

4 Theory of Operation

Recorder Adapter Board

Thermal Line Printhead (TLPH)

Paper Sensor

Stepper Motor

Recorder Adapter Board

Recorder signals are handled by the Recorder Adapter Board (M2703-66430), connected to the Main CPU Board. Video signals to the display are also wired through this board, and connection to the Display Adapter Board is made via a silver-colored 50-pin ribbon cable.

The recorder unit, including the TLPH, is connected to the Recorder Adapter Board via a white 50-pin, ribbon cable. The stepper motor and the paper sensor are also connected to the Recorder Adapter Board. The Recorder Adapter Board is connected to the Main CPU Board via a 154-pin connector.

Thermal Line Printhead (TLPH)

The TLPH is located on its own holder in the recorder chassis.

Paper Sensor

The paper sensor hardware consists of a reflective light sensor that detects the black marks on the trace paper, and paper-out. It is attached to the RFI Bracket, and connected to the Recorder Adapter Board via a removable cable connector.

Stepper Motor

The stepper motor is a bipolar motor controlled by a micro-stepping motor driver on the Recorder Adapter Board. The motor is located on the recorder chassis and is connected to the Recorder Adapter Board via a removable cable connector.

LCD Display and Touchscreen

The LCD Display Assembly consists of a four-wire resistive touchscreen, a 6.5” TFT panel, and a backlight inverter, all connected to the Display Adapter Board (M2703-66440) and fitted into the display housing.

The board is connected to the Recorder Adapter Board (M2703-66430) via a 50-pin ribbon cable. The green power LED is incorporated into the Display Adapter Board.

Noninvasive Blood Pressure Assembly

The optional Noninvasive Blood Pressure Assembly (M2703-64602) is located in the front lefthand corner of the bottom housing assembly. It is connected via a serial link to the Main CPU Board.

SpO2 Assembly

The optional SpO2 Assembly (M2703-64603) is physically located on the Bus Master Board, but sends data directly to the Main CPU Board via a serial link.

17

Philips Avalon FM20, Avalon FM30 Service manual

4 Theory of Operation

Input/Output Interface Boards

Input/Output Interface Boards

There are two optional interface boards available:

LAN/RS232 Interface Board, used for connecting to a PC running the Support Tool and to a surveillance and documentation system such OB TraceVue.

PS/2 Interface Board, used for connecting an external keyboard or mouse.

The interface boards plug into the two interface slots on the underside of the device, and are controlled by the Main CPU Board.

Transducer Hardware Overview

 

EEPROM

Transducer

 

 

cable

Analog/ Digital conversion

Master Clock recovery

 

A

 

 

Analog Signal

 

Communication Transceiver

 

 

D

Signal

 

 

 

 

processor

 

Timing and Mode Control

 

 

 

Power and

 

 

Identification

 

Power supply

 

 

and reset

 

 

 

Transducer CPU board

 

Ultrasound frontend board

 

or alternatively

 

TOCO/ECG frontend board

1

18

Transducer Types

4 Theory of Operation

Transducer Types

Transducers consist of the following types:

US

Toco

Toco+ (includes DECG/MECG and IUP capability)

They all share the same power supply, and analog-to-digital conversion circuitries. The processor software is also the same for all transducers. The frontends, however, are specific to each transducer type.

Functional Description of the Transducer CPU

The CPU section of the transducers is made up of the following main functional blocks:

CPU (micro controller)

Analog-to-Digital Converter

Communication Transceiver (CAN bus driver)

EEPROM

CPU (Micro Controller)

A single-chip processor is used to control the transducer, generate the frontend control signals, control the analog-to-digital signal conversion, and to perform the signal processing.

Analog-to-Digital Converter

Analog-to-digital (A/D) signal conversion is carried out by the 16-bit AD converter. Digital signals are directly communicated from the A/D converter to the CPU.

Communication Transceiver (CAN Bus Driver)

The communications transceiver (CAN bus driver) communicates directly with the transducer CPU, and allows the transducer to communicate with the Bus Master Board via the CAN bus.

EEPROM

The serial EEPROM stores all non-volatile data required to operate the transducer (for example, calibration and correction factors for frontend gains and offsets, country-specific information, serial numbers and error logs).

Toco Transducer Frontend

Uterine activity is measured by evaluating the hardness of the mother’s abdomen with a pressure sensitive resistor bridge (strain gauge sensor element). The strain gauge sensor element requires an excitation voltage and its differential output signal is proportional to the pressure applied to it. A DC excitation voltage is used, and the resulting output signal is fed directly to an A/D signal converter before being sent to the processor.

19

4 Theory of Operation

Ultrasound Transducer Frontend

Ultrasound Transducer Frontend

The ultrasound frontend is a pulsed Doppler system with a 1.0 MHz ultrasound frequency, and a pulse repetition rate of 3 kHz. Seven ultrasound crystals are used as transmitter and receiver.

Toco+ Transducer Frontends

Several parameter frontends are combined on one board. In addition to the Toco frontend, additional supported parameters are DECG, MECG and IUP.

A seven-pin ‘D-type’ socket carries all parameter related inputs and outputs. An external mode resistor, connected to one of the pins, automatically detects which mode to set when an adapter cable is plugged in (whether it is DECG, MECG, or IUP).

Toco Frontend

See “Toco Transducer Frontend” on page 19.

IUP Frontend

Intrauterine pressure (IUP) is measured via a piezo resistive bridge with AC excitation connected to the RA / LA input pins of the ECG amplifier. A/D conversion of the IUP signal is done by the 16-bit A/D converter.

ECG Frontend

The ECG frontend measures both DECG and MECG, using a 3-lead system (RA, LA and reference electrode). The ECG mode is automatically detected when an adapter cable is attached. Input lines are ESD protected.

20

5

Interfaces

There are two interface boards available as options for the Avalon fetal monitors:

LAN / RS232 system interface

Dual PS/2 interface

Dual PS/2 system

interface

LAN / RS232 system

interface

The interfaces are “plug-and-play” boards, and fit into dedicated slots on the underside of the monitor. See “Removing the Interface Boards” on page 87 for details of how to remove and fit the boards.

LAN / RS232 Interface

The LAN / RS232 system interface has two fully isolated ports:

The LAN connection can be used for connecting the monitor to PC for configuration or upgrade using the Support Tool, and for future system expansion.

The RS232 connection can be used for connecting the monitor to an obstetrical information and surveillance system, such as OB TraceVue.

Dual PS/2 Interface

This interface provides two PS/2 ports to enable the monitor to be connected to off-the-shelf, “plug-and- play” input devices:

Mouse: any specified PS/2 mouse or trackball may be used for navigation and data entry.

Computer keyboard: a PS/2 computer keyboard can be used for data entry instead of the on-screen pop-up keyboard.

21

5 Interfaces

Dual PS/2 Interface

22

Loading...
+ 110 hidden pages