Hanna Instruments HI 9635 User Manual

4 (1)
Hanna Instruments HI 9635 User Manual

Instruction Manual

HI 9033 - HI 9034

HI 9039 - HI 9130

HI 933100 - HI 9635

Portable Waterproof

Multi-Range

Conductivity/TDS/

Fertilizer Meters

ins

rumns

 

CONDUCTIVITY

 

mS

 

 

 

 

 

 

°C

HI 933100

 

 

MICROPROCESSOR

 

CONDUCTIVITY

 

 

 

METER

 

ON/OFF

 

 

CAL

CFM

 

 

 

BUF

 

 

 

 

FACTOR

UP

 

 

 

DOWN

 

 

 

 

 

 

 

 

 

 

These Instruments are in

 

 

 

 

 

 

 

 

 

 

 

 

M a n u f a c t u r e r s s i n c e 1 9 7 8

Compliance with the CE Directives

Dear Customer,

Thank you for choosing a Hanna Product. Please read this instruction manual carefully before using the instrument. It will provide you with the necessary information for a correct use of the instrument, as well as a precise idea of its versatility. If you require further technical information, do not hesitate to e-mail us at tech@hannainst.com.

These instruments are in compliance with the directives EN 50081-1 and EN 50082-1.

TABLE OF CONTENTS

 

PRELIMINARY EXAMINATION ........................................................

3

GENERAL DESCRIPTION ................................................................

3

FUNCTIONAL DESCRIPTION OF HI 76302W CONDUCTIVITY PROBE ....

4

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9033 ..........

5

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9034 ..........

6

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9039 ..........

7

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9130 ..........

8

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 933100 ......

9

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9635 ........

10

OPERATIONAL GUIDE .................................................................

12

CONDUCTIVITY/TDS CALIBRATION ...............................................

15

CONDUCTIVITY VERSUS TEMPERATURE CHART .............................

22

TDS VERSUS TEMPERATURE CHART .............................................

23

FERTILIZER CALIBRATION (HI 9039 only) ....................................

24

TEMPERATURE COEFFICIENT (HI 933100 and HI 9635 only) .......

25

TDS FACTOR (HI 9635 only) ......................................................

26

TEMPERATURE CALIBRATION

 

(HI 9039, HI 933100 and HI 9635 only) .................................

27

LCD SYMBOLS GUIDE .................................................................

29

PROBE MAINTENANCE ................................................................

30

BATTERY REPLACEMENT ..............................................................

31

ACCESSORIES .............................................................................

32

WARRANTY ................................................................................

33

CE DECLARATION OF CONFORMITY ..............................................

34

PRELIMINARY EXAMINATION

Remove the instrument from the packing material and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer.

Each meter is supplied complete with:

HI 76302W Conductivity Probe with 1 m (3.3') screened cable

9 V battery (except HI 933100 and HI 9635 which are supplied with 4 x 1.5V AA)

Soft carrying case (HI 9130 only)

Hard carrying case (except for HI 9130)

Note: Save all packing materials until you are sure that the instrument functions correctly. Any defective item must be returned in the original package together with the supplied accessories.

GENERAL DESCRIPTION

HI 9033, HI 9034, HI 9039, HI 9130, HI 933100 and HI 9635 are multirange conductivity/TDS meters with Automatic Temperature Compensation. These meters utilize the latest in four-ring potentiometric technology which has been proven to provide higher accuracy than the common amperometric method.

HI 9033, HI 9130, HI 933100 and HI 9635 have a range of 0 to 199900 µS/cm (µmho/cm); HI 9034 has a range of 0 to 19990 mg/L (ppm). HI 9635 can also take TDS measurements in the 0 to 100,000 mg/L range, and temperature measurements in the 0.0 to 60.0°C range. Thanks to this wide measurement range, these meters can be used to analyze any sample from deionized water to brine.

In particular, HI 9130 with its white casing and easy operation has been specifically designed for the food industry.

HI 9039 can measure conductivity, temperature and two different types of fertilizer concentrations (F1 and F2) for applications in agriculture and horticulture. To measure the fertilizer content, simply standardize the two different fertilizer ranges. To monitor the quality of the water used with the fertilizer, the conductivity must be checked. This also helps in the selection of water to be mixed with the fertilizer.

The temperature of all conductivity/TDS measurements is automati-

3

cally compensated for by the ATC circuitry of the meters. The stainless steel rings of the probe have a large surface for better response and easy cleaning. The detachable probe can be changed without returning the unit to the factory. The probe cable is screened and 1 m (3.3') long.

The meters are constructed with a waterproof, rugged housing to provide the versatility required for both laboratory and field use.

Another feature of the meters is the Battery Error Preventive System (BEPS) which prevents erroneous readings due to low battery voltage.

Moreover, HI 9039, HI 933100 and HI 9635 are equipped with a large LCD that helps the user through operation and calibration with large comprehensive symbols. Besides displaying the selected range, the LCD simultaneously displays the temperature. These meters are microprocessor-based and perform buffer recognition and calibration automatically.

For continuous monitoring HI 933100 and HI 9635 are supplied with built-in 12VDC sockets protected behind a waterproof cover.

The temperature coefficient of HI 933100 and HI 9635 is userselectable from 0.0 to 3.0%. The TDS/conductivity ratio (factor) of HI 9635 can be set to 0.7 (F1) or 0.5 (F2).

 

FUNCTIONAL

DESCRIPTION

OF

H I 76302 W CONDUCTIVITY PROBE

1)

Watertight Shielded

1

 

 

Screened Cable

 

 

2)

4 Stainless Steel Rings

 

 

3)

Air-Release holes

 

 

4)

PVC Protective Sleeve

 

 

 

 

 

3

 

 

2

4

4

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9033

1)Battery Compartment

2)Probe Socket

3) LCD Display

4) "mS" or "µS" Indicator

5)'V' Low Battery Indicator

6)ON/OFF Key

7) 19.99 mS/cm Range

Selector

8) Conductivity Calibration

Trimmer

9) 199.9 µS/cm Range

Selector

10) 1999 µS/cm Range

Selector

11)199.9 mS/cm Range Selector key

Range

0.0 to 199.9 µS/cm

 

0 to 1999 µS/cm

 

0.00 to 19.99 mS/cm

 

0.0 to 199.9 mS/cm

Resolution

0.1 µS/cm

 

1 µS/cm

 

0.01 mS/cm

 

0.1 mS/cm

Accuracy

±1% Full Scale, excluding probe error

(@20°C/68°F)

 

Typical EMC

±1% Full Scale

Deviation

 

Calibration

Manual single point

 

through trimmer on the case

Temperature

Automatic from 10 to 40°C

Compensation

(50 to 104°F) with ß of 2% °C

Probe

HI 76302W with 1 m (3.3') cable

 

(included)

Battery Type/Life

1 x 9V; 100 hours of continuous use

Environment

0 to 50°C (32 to 122°F); 100% RH

Dimensions

196 x 80 x 60 mm (7.7 x 3.1 x 2.4")

Weight

425 g (15 oz.); Kit: 1.4 kg (3.1 lb.)

5

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9034

1) Battery Compartment

2) Probe Socket

3) LCD Display

4) 'V' Low Battery Indicator

5) ON/OFF Key

6) 19.99 g/L Range Selector

7) TDS Calibration Trimmer

8) 1999 mg/L Range Selector

9) 199.9 mg/L Range Selector

Range

0.0 to 199.9 mg/L

 

0 to 1999 mg/L

 

0.00 to 19.99 g/L

Resolution

0.1 mg/L

 

1 mg/L

 

0.01 g/L

Accuracy

±1% Full Scale, excluding probe error

(@20°C/68°F)

 

Typical EMC

±1% Full Scale

Deviation

 

Calibration

Manual single point

 

through trimmer on the case

Temperature

Automatic from 10 to 50°C

Compensation

(50 to 122°F) with ß of 2% °C

Probe

HI 76302W with 1 m (3.3') cable

 

(included)

Battery Type/Life

1 x 9V; 100 hours of continuous use

Environment

0 to 50°C (32 to 122°F); 100% RH

Dimensions

196 x 80 x 60 mm (7.7 x 3.1 x 2.4")

Weight

425 g (15 oz.); Kit: 1.4 kg (3.1 lb.)

6

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9039

1)Battery Compartment

2)Probe Socket

3)LCD Display

4)ON/OFF Key

5)RANGE, to select conductivity or temperature measurement mode or first/second fertilizer

6)'V' Low Battery Indicator

7)CAL, to enter or exit calibration mode

8)CFM, to confirm calibration reading

1

2

3

 

mS

6

 

 

CONDUCTIVITY

V

 

 

°C

 

AGRICARE

4HI 9039 Microprocessor Conductivity Fertilizer meter

7

5

ON/OFF

 

 

8

 

 

 

RANGE CAL

CFM

Range

Conductivity

0 to 9999 µS/cm

 

Temperature

0.0 to 50.0°C

 

Fertilizer

1

0.00 to 9.99 g/L

 

Fertilizer

2

0.00 to 9.99 g/L

 

 

 

Resolution

 

1 µS/cm; 0.1°C; 0.01 g/L

Accuracy

 

 

±1% Full Scale, excluding probe error

(@20°C/68°F)

 

 

Typical EMC

 

±1% Full Scale

Deviation

 

 

Calibration

 

Automatic, single point at 5000 µS/cm

 

 

 

Temperature

 

Automatic from 10 to 50°C

Compensation

 

(50 to 122°F) with ß of 2% °C

Probe

 

 

HI 76302W with 1 m (3.3') cable

 

 

 

(included)

Battery Type/Life

 

1 x 9V; 100 hours of continuous use

Environment

 

0 to 50°C (32 to 122°F); 100% RH

Dimensions

 

196 x 80 x 60 mm (7.7 x 3.1 x 2.4")

Weight

 

 

425 g (15 oz.); Kit: 1.4 kg (3.1 lb.)

7

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9130

1) Battery Compartment

2) Probe Socket

3) "mS" or "µS" Indicator

4) LCD Display

5) 'V' Low Battery Indicator

6)0 to 1999 µS/cm Range Selector

7) 0 to 19.99 mS/cm Range Selector

8) Conductivity Calibration Trimmer

9) ON/OFF Key

10) 0 to 199.9 mS/cm Range Selector

Range

0 to 1,999 µS/cm

 

0.00 to 19.99 mS/cm

 

0.0 to 199.9 mS/cm

Resolution

1 µS/cm

 

0.01 mS/cm

 

0.1 mS/cm

Accuracy

±2% Full Scale, excluding probe error

(@20°C/68°F)

 

Typical EMC

±1% Full Scale

Deviation

 

Calibration

Manual single point

 

through trimmer on the case

 

 

Temperature

Automatic from 10 to 40°C

Compensation

(50 to 104°F) with ß of 2%/°C

Probe

HI 76302W with 1 m (3.3') cable

 

(included)

Battery Type/Life

1 x 9V; 100 hours of continuous use

Environment

0 to 50°C (32 to 122°F); 100% RH

Dimensions

196 x 80 x 60 mm (7.7 x 3.1 x 2.4")

Weight

425 g (15 oz.); Kit: 1.4 kg (3.1 lb.)

8

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 933100

1)

Battery Compartment

1

2)

Probe Socket

 

3)

Liquid Crystal Display

2

4)ON/OFF Key

5)CAL, to enter or exit calibration mode

6)FACTOR, to set the tem-

 

perature coefficient or select

3

 

the temperature mode

 

 

7)

12VDC Power Socket

 

 

8)

'LOBAT'

Low Battery

 

4

9)

CFM, to

confirm the

cali-

5

 

bration

reading

 

10)

 

 

BUF, to select the calibra-

6

tion buffer value

7

 

ins rum n s

 

 

mS

 

CONDUCTIVITY

LO BAT

8

 

°C

HI 933100

 

MICROPROCESSOR

 

CONDUCTIVITY METER

9

 

ON/OFF

10

CAL

CFM

BUF

11

FACTOR UP DOWN

12

11)DOWN, to set the tempera-

ture coefficient or for manual temperature compensation setting

12)UP, to set the temperature coefficient or for manual temperature compensation setting

Range

Conductivity

0.0 to 150.0 µS/cm

 

 

150 to 1500 µS/cm

 

 

1.50 to 15.00 mS/cm

 

 

15.0 to 199.9 mS/cm

 

Temperature

0.0 to 60.0°C

Resolution

0.1 µS/cm; 1 µS/cm; 0.01 mS/cm;

 

 

0.1 mS/cm; 0.1°C

Accuracy

 

±1% Full Scale; ±0.5°C

(@20°C/68°F)

excluding probe error

Typical EMC

±1% Full Scale

Deviation

±0.2 °C

Calibration

Automatic, two points

 

 

at 0 and 1413 or 12880 µS/cm

Temperature

Automatic or manual from 10 to 60°C

Compensation

(50 to 140°F) with ß from 0 to 3%/°C

Probe

 

HI 76302W with 1 m (3.3') cable

 

 

(included)

Battery Type/Life

4 x 1.5V AA; 150 hours of

 

 

continuous use. Power socket for 12VDC

Environment

0 to 50°C (32 to 122°F); 100% RH

Dimensions

196 x 80 x 60 mm (7.7 x 3.1 x 2.2")

Weight

 

425 g (15 oz.); Kit: 1.4 kg (3.1 lb.)

 

 

 

9

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9635

1)Battery Compartment

2)Probe Socket

3)Liquid Crystal Display

4)ON/OFF Key

5)CAL, to enter or exit calibration mode

6)FACTOR, to show temperature, %TC or TDS factor on the secondary display

7)12VDC Power Socket

8)RANGE, to select the measurement mode

9)CFM, to confirm the calibration reading

10)BUF, to select the calibration buffer value

11)DOWN, to set the temperature coefficient or for manual temperature compensation setting

12)UP, to set the temperature coefficient or for manual temperature compensation setting

10

RANGE

Conductivity

0.0 to 150.0 µS/cm

 

 

 

150 to 1500 µS/cm

 

 

 

1.50 to 15.00 mS/cm

 

 

 

15.0 to 199.9 mS/cm

 

 

TDS

0.0 to 150.0 mg/L (ppm)

 

 

 

150 to 1500 mg/L (ppm)

 

 

 

1.50 to 15.00 g/L (ppt)

 

 

 

15.0 to 100.0 g/L (ppt)

 

Temperature

0.0 to 60.0°C

Resolution

Cond.

0.1 µS/cm / 1 µS/cm

 

 

 

0.01 mS/cm / 0.1 mS/cm

 

 

TDS

0.1 mg/L (ppm) / 1 mg/L (ppm)

 

 

 

0.01 g/L (ppt) / 0.1 g/L (ppt)

 

Temperature

0.1°C

Accuracy

 

Cond.

±2% Full Scale

(@20°C/68°F)

TDS

±2% Full Scale

 

Temperature

±0.5°C

 

 

 

excluding probe error

Typical EMC

Cond.

±2% Full Scale

Deviation

TDS

±2% Full Scale

 

Temperature

±0.5°C

Calibration

 

Automatic 1, 2 or 3 point calibration

 

 

 

at 84, 1413 or 12880 µS/cm

 

 

 

Temperature

 

Automatic or manual from 0 to 60°C

Compensation

 

(32 to 140°F) with adjustable ß

 

 

 

from 0.0 to 3.0% /°C and

 

 

 

TDS factor at 0.70 or 0.50.

 

 

 

Compensation referenced to 25°C (77°F)

Probe

 

 

HI 76302W with 1 m (3.3') cable

 

 

 

(included)

Battery Type/Life

4x1.5V AA/150 hours of continuous use

 

 

 

Auto-shut off after 10' of non use

 

 

 

12VDC power socket

Environment

 

0 to 50°C (32 to 122°F); 100% RH

Dimensions

 

196 x 80 x 60 mm (7.7 x 3.1 x 2.4")

Weight

 

 

425 g (15 oz.); Kit: 1.4 kg (3.1 lb.)

11