Hanna Instruments HI 9147 User Manual

4 (1)
Hanna Instruments HI 9147 User Manual

Instruction Manual

HI 9147

Portable, Water-resistant

Dissolved Oxygen Meter

with

Galvanic Probe

and

Battery Recharging System

w w w . h a n n a i n s t . c o m

WARRANTY

HI 9147 meters are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions.

The probes are guaranteed for one year.

This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.

If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure.

If the repair is not covered by the warranty, you will be notified of the charges incurred.

If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.

Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.

Dear Customer,

Thank you for choosing a Hanna product.

Please read carefully this instruction manual before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.

This instrument is in compliance with the CE directives.

PRELIMINARY EXAMINATION

Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If there is any damage, notify your Dealer.

The meter is supplied complete with

• D.O. probe, fixed:

HI 76409/4 with 4 m cable for HI 9147-04 HI 76409/10 with 10 m cable for HI 9147-10

Spare membranes (HI 76409A/P) with O-rings, 5 pcs.

HI 7042S electrolyte solution (30 ml bottle)

HI 710016 shockproof rubber boot

9V Ni-MH rechargeable battery

Screwdriver and instructions.

Note: Conserve all packing material until the instrument has been observed to function correctly. Any defective item must be returned in its original packing.

GENERAL DESCRIPTION

Oxycheck is a water-resistant Dissolved Oxygen meter with manual calibration, automatic temperature compensation (ATC), salinity compensation and direct probe.

The instrument is equipped with an inductive system for battery recharge, with no external contact, to ensure a watertight seal. Dissolved Oxygen is indicated in mg/L (ppm) or in % of saturation. The temperature can be measured in the range from -5 to 50°C. The Dissolved Oxygen readings are automatically compensated for the temperature effects on the oxygen solubility and membrane permeability. Moreover, the salinity compensation feature allows determination of Dissolved Oxygen even in salty waters.

The meter is housed in a rugged water-resistant case for maximum protection in the field as well as in the laboratory.

The D.O. probe is provided with a membrane covering the galvanic sensors and a built-in thermistor for temperature measurement and compensation. The thin permeable membrane isolates the sensor elements from the testing solution, but allows oxygen to enter. Oxygen that passes through the membrane causes a current flow, from which the oxygen concentration is determined.

Two models are available:

HI 9147-04 with HI 76409/4 probe (4 m cable)

HI 9147-10 with HI 76409/10 probe (10 m cable)

FUNCTIONAL DESCRIPTION

1)HI 76409/4 or HI 76409/10 galvanic D.O. probe (fixed)

2)Liquid Crystal Display

3)BL key, turn the backlight ON (as long as it is pressed)

4)°C key, display temperature measurement

5)% key, display O2 measurement in % of saturation

6)% trimmer, for calibration adjustment

7)SAL trimmer, for setting salinity compensation value

8)SAL key, display setting of salinity factor (active only in mg/L measurement mode)

9)mg/L key, display O2 measurement in mg/L (ppm)

10)OFF key, turn the instrument OFF

Note: To turn the instrument ON, press any of the measurement keys (°C, %, mg/L)

SPECIFICATIONS

Range

0.0 to 50.0 mg/L (ppm) O2

 

0 to 600 % O2

 

-5.0 to 50.0°C

Resolution

0.1 mg/L or 1% (O2) / 0.1°C

Accuracy

±1% of reading (O2) / ±0.2°C

Typical EMC Deviation

±2% of reading (O2) / ±1°C

Calibration

Manual, in saturated air

Temperature Compensation

Automatic, from -5 to 50°C (23 to 122°F)

Salinity Compensation

0 to 51 g/L (resolution 1 g/L)

Probe (fixed)

HI 76409/4 with 4 m cable or

 

HI 76409/10 with 10 m cable

Battery Type

1 x 9V Ni-MH, rechargeable

Battery Charge Life

200 hours of use (BL off)

Battery Life

Approx. 5 years

Auto-off

After approx. 6 minutes

Environment

0 to 50°C (32 to 122°F);

 

max 95% RH non-condensing

Dimensions

164 x 76 x 45 mm

 

(6.5 x 3.0 x 1.8")

Weight (with HI 76409/4 probe) 511 g (1.12 lb.)

ACCESSORIES

HI 76409/4* Galvanic D.O. probe with built-in temperature sensor and 4 m cable

HI 76409/10* Galvanic D.O. probe with built-in temperature sensor and 10 m cable

HI 76409-0 Protective sleeve for HI 76409 probes series HI 76409A/P Membrane for HI 76409 probes series, 5 pcs. HI 7042S Electrolyte solution, 30 ml bottle

HI 710015 Shockproof rubber boot, blue color HI 710016 Shockproof rubber boot, orange color HI 731326 Calibration screwdriver (20 pcs)

HI 710040 Battery recharger

* To be substituted by authorized technical personnel only

IST9147R2 05/05