Hanna Instruments HI 98129 User Manual

4 (1)
Hanna Instruments HI 98129 User Manual

Instruction Manual

HI 98129 - HI 98130

Waterproof pH, EC/TDS & temperature meters

Dear Customer,

Thank you for choosing a Hanna product. This manual will provide you with the necessary information for a correct operation. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.

These instruments are in compliance with the directives.

PRELIMINARY EXAMINATION

 

 

 

 

 

 

 

This Instrument is in

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

w w w . h a n n a i n s t . c o m

Compliance with CE

 

 

 

 

 

 

 

Directives

WARRANTY

HI 98129 and HI 98130 are warranted for one year against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The electrode is warranted for a period of six months. This warranty is limited to repair or replacement free of charge.

Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.

If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.

To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.

All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.

Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.

Remove the instrument from the packing material and examine it carefully. If any damage has occurred during shipment, immediately notify your Dealer or the nearest Hanna Customer Service Center.

The meters are supplied with:

4 x 1.5V batteries

HI 73127 pH electrode

HI 73128 Tool to remove the pH electrode

Note: Conserve all packing material until the instrument has been observed to function correctly. Any defective item must be returned in its original packing.

GENERAL DESCRIPTION

HI 98129 and HI 98130 are waterproof pH, EC/TDS and temperature meters. The housing has been completely sealed against humidity and designed to float.

All pH and EC/TDS readings are automatically temperature compensated (ATC), and temperature values can be displayed in °C or °F units.

For EC/TDS readings, the EC/TDS conversion factor (CONV) is selectable by the user, as well as the temperature compensation coefficient β (BETA).

The meters can be calibrated at one or two points for pH (with auto-buffer recognition and against five memorized buffer values), and at one point for EC/ TDS.

Measurements are highly accurate with a unique stability indicator right on the LCD.

A low battery symbol warns the user when the batteries are to be replaced. In addition the Battery Error Prevention System (BEPS) avoids erroneous reading caused by low voltage level by turning the meter off.

The HI 73127 pH electrode, supplied with the meters, is interchangeable and can be easily replaced by the user.

The stainless steel encapsulated temperature sensor facilitates faster and more accurate temperature measurement and compensation.

FUNCTIONAL DESCRIPTION

1.Battery compartment

2.Liquid Crystal Display (LCD)

3.Stability indicator

4.Low battery indicator

5.ON/OFF/MODE button

6.HI 73127 pH electrode

7.Temperature sensor

8.EC/TDS probe

9.SET/HOLD button

10.Secondary LCD

11.Primary LCD

SPECIFICATIONS

Range Temperature:

0.0 to 60.0°C or 32.0 to 140.0°F

HI 98129

pH: 0.00 to 14.00

 

EC: 0 to 3999 µS/cm

 

TDS: 0 to 2000 ppm

HI 98130

pH: 0.00 to 14.00

 

EC: 0.00 to 20.00 mS/cm

 

TDS: 0.00 to 10.00 ppt

Resolution

0.1°C or 0.1°F

HI 98129

0.01 pH; 1 µS/cm ; 1 ppm

HI 98130

0.01 pH, 0.01 mS/cm ; 0.01 ppt

Accuracy (@20°C/68°F)

Temperature

±0.5°C or ±1°F

 

EC/TDS

±2% f.s.

 

pH

±0.01

Typical EMC Deviation

Temperature

±0.5°C or ±1°F

 

pH

±0.02 pH

 

EC/TDS

±2% f.s.

Temp.Compensation

pH: Automatic

 

 

EC/TDS: with β =0.0 to 2.4%/ºC

Environment 0 to 50°C (32 to 122°F); RH 100%

EC/TDS Conversion Factor 0.45 to 1.00 (CONV)

Calibration

pH: at 1 or 2 points with 2 sets of memo

 

rized buffers (pH 4.01/7.01/10.01 or

 

pH 4.01/6.86/9.18)

 

EC/TDS: automatic, at 1 point

EC/TDS Calibration Solutions

HI 98129 HI7031 (1413 µS/cm)

 

HI7032 (1382 ppm; CONV=0.5)

 

HI70442 (1500 ppm; CONV=0.7)

HI 98130

HI7030 (12.88 mS/cm)

 

HI70038 (6.44 ppt; CONV=0.5

 

or 9.02 ppt; CONV=0.7)

Electrode

HI 73127 pH electrode (included)

BatteryType/Life 4 x 1.5V with BEPS / typical 100 hours

Auto-off After 8 min.

Dimensions 163 x 40 x 26 mm (6.4 x 1.6 x 1.0”)

Weight

85 g (3.0 oz)

Recommendations for Users

Before using this product, make sure that it is entirely suitable for the environment in which it is used. Operation of this instrument in residential areas could cause unacceptable interferences to radio and TV equipment.

The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all times. During operation, ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic discharges. Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC performance.

To avoid electrical shock, do not use this instrument when voltages at the measurement surface exceed 24 VAC or 60 VDC. To avoid damages or burns, do not perform any measurement in microwave ovens.

ISTR98129R2 01/02