Instruction Manual
HI 2210 HI 2211
Microprocessor-based
pH/mV/°C
Bench Meters
www.hannainst.com
1
Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using the instruments. This manual will provide you with the necessary information for correct use of the instruments, as well as a precise idea of their versatility.
If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com or view our worldwide contact list at www.hannainst.com.
These instruments are in compliance with directives.
WARRANTY
HI 2210 and HI 2211 are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Electrodes and probes are guaranteed for six months. This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packed for complete protection.
TABLE OF CONTENTS |
|
WARRANTY .................................................................................. |
3 |
PRELIMINARY EXAMINATION ........................................................ |
4 |
GENERAL DESCRIPTION ................................................................ |
4 |
FUNCTIONAL DESCRIPTION .......................................................... |
5 |
SPECIFICATIONS ........................................................................... |
6 |
OPERATIONAL GUIDE ................................................................... |
7 |
pH CALIBRATION ....................................................................... |
10 |
TEMPERATURE CALIBRATION (for technical personnel only) ........... |
13 |
pH BUFFER TEMPERATURE DEPENDENCE ...................................... |
15 |
ELECTRODE CONDITIONING & MAINTENANCE ............................... |
16 |
TROUBLESHOOTING GUIDE ........................................................ |
19 |
TEMPERATURE CORRELATION FOR pH SENSITIVE GLASS ................. |
20 |
ACCESSORIES ............................................................................. |
21 |
2 |
3 |
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer or the nearest Hanna Customer Service Center.
Each instrument is supplied with:
•HI 1131B Glass-body Combination pH Electrode
•HI 7662 Temperature Probe
•HI 76404 Electrode Holder
•pH 4.01 & 7.01 Buffer Solutions, 20 mL each
•HI 7071S Electrolyte Solution
•12VDC Power Adapter
•Instruction Manual
Note: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in the original packing with the supplied accessories.
GENERAL DESCRIPTION
The Hanna HI 2210 and HI 2211 are microprocessor-based pH and temperature bench meters.
HI 2211 can also be used for ion concentration (ISE) and Oxidation Reduction Potential (ORP) in the mV range.
pH measurements are compensated for temperature effect manually or automatically with the HI 7662 temperature probe.
The instrument is equipped with a large easy-to-read LCD which shows the pH (or mV) and temperature simultaneously, together with graphic symbols.
The calibration process is guided step by step through clear indications on the LCD. A stability indicator makes the calibration procedure error-free.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
4
FUNCTIONAL DESCRIPTION
Front Panel
Rear Panel
1)Liquid Crystal Display (LCD).
2)CAL key, to enter or exit/escape calibration mode.
3)CFM key, to confirm different values.
4)°C key and °C keys, to manually increase/decrease temperature or select pH buffer.
5)RANGE key, to select measurement range (HI 2211 only).
6)MEM key, to store a value into memory.
7)MR key, to recall the stored value.
8)Secondary LCD.
9)Primary LCD.
10)ON/OFF switch.
11)Power adapter socket.
12)BNC electrode connector.
13)Temperature probe socket.
14)Electrode reference socket.
5
SPECIFICATIONS
|
–2.00 to 16.00 pH |
|
|
|
|
RANGE |
±399.9 mV (HI 2211 only) |
|
±2000 mV (HI 2211 only) |
||
|
||
|
|
|
|
–9.9 to 120.0 °C |
|
|
|
|
|
0.01 pH |
|
|
|
|
RESOLUTION |
0.1 mV (HI 2211 only) |
|
1 mV (HI 2211 only) |
||
|
||
|
|
|
|
0.1 °C |
|
|
|
|
|
±0.01 pH |
|
|
|
|
ACCURACY |
±0.2 mV (HI 2211 only) |
|
±1 mV (HI 2211 only) |
||
@ 20°C / 68°F |
||
±0.5 °C (0.0 – 100.0 °C) |
||
|
||
|
±1 °C (outside) |
|
|
(excluding probe error) |
|
|
|
|
|
1 or 2 point calibration, 5 buffers |
|
pH Calibration |
available |
|
|
(4.01, 6.86, 7.01, 9.18, 10.01) |
|
|
|
|
Temperature compensation |
Manual or Automatic from: |
|
–9.9 to 120.0 °C (14.2 to 248.0 °F) |
||
|
||
|
|
|
pH Electrode |
HI 1131B (included) |
|
|
|
|
Temperature probe |
HI 7662 (included) |
|
|
|
|
Input impedance |
1012 ohms |
|
Power supply |
12 VDC adapter (included) |
|
|
|
|
Dimensions |
240x182x74 mm (9.4x7.1x2.9”) |
|
|
|
|
Weight |
1.1 Kg (2.5 lb); |
|
kit with holder 2.5 Kg (5.5 lb) |
||
|
||
|
|
|
Environment |
0 to 50 °C (32 to 122 °F) |
|
max. 95% RH non-condensing |
||
|
||
|
|
|
Warranty |
2 years |
|
|
|
6
OPERATIONAL GUIDE
POWER CONNECTION
Plug the 12 VDC adapter into the power supply socket.
Notes: • These instruments use non volatile memory to retain the pH, mV, temperature calibrations and all other settings, even when unplugged.
• Make sure a fuse protects the main line.
ELECTRODE AND PROBE CONNECTIONS
For pH or ORP combination electrode connect to the BNC connector on the back of the instrument.
For electrodes with a separate reference connect the electrode’s BNC to the BNC connector and the reference electrode plug to the reference socket.
For temperature measurements and automatic temperature compensation connect the temperature probe to the appropriate socket.
INSTRUMENT START-UP
•Turn the instrument on by pressing the ON/OFF switch located on the rear panel.
•All LCD segments are displayed while the instrument performs a self test.
pH MEASUREMENTS
Make sure the electrode and the instrument have been calibrated together before taking pH measurements.
•Submerge the electrode and the temperature
probe approximately 4 cm (1½”) into the sample to be tested and stir gently. Allow time for the electrode to stabilize.
•The pH is displayed on the primary LCD and the temperature on the secondary LCD.
pH
C
•If the pH reading is out of range, “----” will be displayed on the LCD.
7
If measurements are taken successively in different samples, it is recommended to rinse the electrode thoroughly with deionized water or tap water and then with some of the next sample in order to prevent cross-contamination.
The pH reading is affected by temperature. In order to measure the pH accurately, the temperature effect must be compensated for. To use the Automatic Temperature Compensation feature, connect and submerge the HI 7662 temperature probe into the sample as close as possible to the electrode and wait for a few seconds.
If the temperature of the sample is known, manual temperature compensation can be performed by disconnecting the temperature probe.
The display will then show the default temperature |
|
of 25 °C or the last recorded temperature reading |
C |
with the “°C” tag blinking. The temperature can now |
|
be adjusted with the ARROW keys (from -9.9 °C to |
|
120.0 °C). |
|
ORP MEASUREMENTS (HI 2211 only)
An optional ORP electrode must be used to perform ORP measurements (see Accessories).
Oxidation-reduction potential (REDOX) measurements provide the quantification of the oxidizing or reducing power of the tested sample.
To correctly perform a redox measurement, the surface of the ORP electrode must be clean and smooth.
•Press RANGE to enter mV range.
•Submerge the tip of the ORP electrode 4 cm
(1½") into the sample to be tested and allow a few seconds for the reading to stabilize.
•The instrument displays the mV reading on the primary LCD and the temperature on the secondary LCD.
mV |
C |
•If the reading is out of range, “----” will be displayed on the LCD.
8
TEMPERATURE MEASUREMENTS
Connect the HI 7662 temperature probe to the TEMP socket and turn the instrument on.
Immerse the temperature probe into the sample and allow the reading on the secondary LCD to stabilize.
MEMORY FUNCTION
Press and hold down MEM. The last reading will appear on the LCD along with “MEM” tag until MEM is released.
Press MR and the previously memorized reading will be displayed with “MEM” tag on the LCD.
9