Instruction Manual
HI 9043 - HI 9044
Portable Microprocessor
K-Type Thermocouple
Thermometers
|
|
|
|
|
|
|
These Instruments are in |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Compliance with CE |
w w w . h a n n a i n s t . c o m |
Directives |
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Sensors and probes are warranted for a period of six months.
This warranty is limited to repair or replacement free of charge.
Damages due to accidents, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure.
First obtain a Returned Goods Authorization number from the Customer Service department, then return the instrument with the Authorization # included along with shipment costs prepaid. If the repair is not covered by the warranty, you will be notified of the charges.
When shipping any instrument, make sure it is properly packaged for complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
Dear Customer,
Thank you for choosing a Hanna product. This manual will provide you with the necessary information for correct operation. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e- mail us at tech@hannainst.com.
These instruments are in compliance with the directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your Dealer or the nearest Hanna Office.
Note: Save all packing materials until you are sure that the instrument functions correctly. Any defective item must be returned in the original packaging together with the supplied accessories.
GENERAL DESCRIPTION
HI 9043andHI 9044 are microprocessor-based thermometers which allow temperature measurements using K-type thermocouple probes. The non-linearity of the temperature probe is linearized by the microprocessor.
Standard features include dual-level LCD for simultaneously displaying of the highest and lowest measured temperatures, display hold, selectable measurement unit (°C or °F) and resolution which automatically switches from 0.1° to 1° for temperatures above 200°C (400°F).
The meters are also provided with low battery detection and BEPS (Battery Error Preventing System), which turns the unit off when the battery is discharged avoiding erroneous readings caused by low battery level.
HI 9043 comes complete with a 9V battery and instruction manual.
HI 9044 comes complete with HI 766E2 fixed probe, a 9V battery and instruction manual.
FUNCTIONAL DESCRIPTION
1.Liquid Crystal Display
2.Measuring unit selection key
3.HI/LO values reset key
4.HOLD key
5.ON/OFF key
1)Current temperature value
2)HOLD indicator
3)Low Battery indicator
4)Minimum temperature value
5)Maximum temperature value
6)K-type probe indicator
7)Measuring unit, °C or °F
SPECIFICATIONS
|
Range (*) |
-50.0 to 199.9°C / 200 to 1350°C |
|
|
-58.0 to 399.9°F / 400 to 2462°F |
|
|
|
|
Resolution |
0.1°C (up to 199.9°C) / 1°C (outside) |
|
|
0.1°F (up to 399.9°F) / 1°F (outside) |
|
|
|
|
Accuracy |
±0.2% F.S. |
|
(@20°C/68°F) |
excluding probe error |
|
|
|
|
Typical EMC |
±3 °C and ±6 °F |
|
Deviation |
with HI 766 K-Thermocouple probe |
|
|
|
|
Display |
Dual-line LCD |
|
|
|
|
Battery Type |
1 x 9V (IEC 6LR61) battery |
|
Life |
approx. 500 hours of continuous use |
|
|
|
|
Probe |
HI 766 K-Thermocouple series |
|
|
• optional for HI 9043 |
|
|
• HI 766E2 fixed for HI 9044 |
|
|
|
|
Environment |
-10 to 50ºC (14 to 122ºF); |
|
|
max. 95% RH non-condensing |
|
|
|
|
Dimensions |
180 x 83 x 40 mm (7.1x3.3x1.6") |
|
|
|
|
Weight |
226 g (8 oz.) meter only |
(*)Range may be limited by probe. The range of HI 9044 is limited to 900°C/1650°F by the fixed probe.
FACTORY RECALIBRATION
All Hanna thermometers have been accurately pre-cali- brated at the factory. It is generally recommended to have all thermometers recalibrated at least once a year.
For an accurate annual recalibration, contact your dealer or the nearest Hanna Service Center.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.