Aneroid Menu
Präzision in Ihren Händen. Precision at your fingertips.
94-01-130 Stand: Nov. 2001
Sercice Manual / Service Anleitung
+
Ersatzteilkatalog / Spare Part Catalogue
2001 / 2002
Speidel + Keller GmbH & Co. KG Zollerstrasse 2-4
D - 72417 Jungingen Tel.: (++49) 7477-9271-0
Fax.: (++49) 7477-9271-90 e-mail: info@speidel-keller.de homepage: www.speidel-keller.de
Seite / Page 2
SERVICE MANUAL
WELCH ALLYN / SPEIDEL + KELLER
Dear S+K Servicepartner,
We are very happy to welcome you as a Service partner of Speidel + Keller.
The company Speidel + Keller is continually trying to offer optimal conditions and terms, so that, a smooth servicing of our units is guaranteed.
The Speidel + Keller blood pressure units distinguish themselves through very high quality. This is a warranty that S+K product possess a very long lifetime.
Even though the best units naturaly need regulary inspections so that a orderly run can be accomplished. Further it is a neccessarity in reference to the requirements of the MPG
and the EC-medical product guidelings, service the medical products on a regular base as well as carring out measurement inspections.
Because of the above mentioned reasons we have issued these repair instruction guides for all our units.
These instructions shall escort you to carry out an orderly repair.
Mangement Service
Seite / Page 3
Table of contents |
/ Inhaltsverzeichnis: |
English |
Deutschsprachig |
|
|
Page: |
Seite: |
Preface / Vorwort |
|
2 |
22 |
Table of contents / Inhaltsverzeichnis |
3 |
Entfällt |
|
Failure search mechanical blood pressure units |
4 |
23 |
|
Failure search mechanical blood pressure units |
5 |
24 |
|
Failure search mercury blood pressure units |
6 |
25 |
|
Instruction of dismantling and re-assembling |
|
|
|
Reparaturanweisung Demontage und Wiedermontage |
|
|
|
Primus 50 Metal |
Art. No. 50-01-000 |
7 |
26 |
Primus 68 Metal |
Art. No. 50-04-000 |
7 |
26 |
Manuell 50 Metal |
Art. No. 50-16-000 |
7 |
26 |
Manuell 68 Metal |
Art. No. 50-17-000 |
7 |
26 |
Primus 50 KC |
Art.No. 50-81-000 |
8 |
27 |
Primus 68 KC |
Art. No. 50-84-000 |
8 |
27 |
Manuell 50 KC |
Art. No. 50-96-000 |
8 |
27 |
Manuell 68 KC |
Art.No. 50-97-000 |
8 |
27 |
Primus Stabil 3 |
Art.No. 50-25-000 |
9 |
28 |
Primus Stabil yellow |
Art.No. 50-49-000 |
9 |
28 |
Primus Stabil red |
Art.No. 50-59-000 |
9 |
28 |
Primus Stabil blue |
Art.No. 50-50-000 |
9 |
28 |
Primus Stabil green |
Art.No. 50-45-000 |
9 |
28 |
Manuell Stabil 3 |
Art.No. 50-63-000 |
9 |
28 |
ISO Stabil 5 |
Art.No. 50-71-000 |
10 |
29 |
Disytest |
Art.No. 52-09-000 |
11 |
30 |
Maxi Stabil 3 |
Art.No. 59-32-000 |
12 |
31 |
Fixomed |
Art. No. 50-91-000 |
13 |
32 |
Hand 2003 |
Art.No. 50-03-000 |
14 |
33 |
Oscillophon |
Art.No. 51-05-000 |
15 |
34 |
Instruction for end evaluation /measurements – |
|
|
|
Anweisung für die Endprüfung |
16 |
35 |
|
Tool list for blood pressure units - Werkzeugliste |
17 |
36 |
|
Failure search for Stehoscopes - Fehlersuche Stehoskope |
|
|
|
Stethoscope |
Art.No. 84-10-000 |
18 |
37 |
Planoscope 40 mm |
Art.No. 83-11-080 |
18 |
37 |
Planoscope 30 mm |
Art.No. 83-12-080 |
18 |
37 |
Duophon 40 mm |
Art.No. 82-11-080 |
18 |
37 |
Duophon 30 mm |
Art.No. 82-15-080 |
18 |
37 |
Professionell |
Art.No. 86-01-080 |
18 |
37 |
Harvey Elite |
Art.No. 87-01-080 |
18 |
37 |
Failure search for hand fans - Fehlersuche Handventile |
|
|
|
Hand fan |
Art.No. 17-01-000 |
19 |
38 |
Hand fan |
Art.No.17-02-000 |
19 |
38 |
Hand fan |
Art.No. 17-08-000 |
19 |
38 |
Hand fan |
Art.No. 17-10-000 |
19 |
38 |
Failure search for special medical manometers - |
|
|
|
Fehlersuche Spezial Manometer |
|
|
|
Pressure inf. Cuff |
Art.No. 51-08-000 |
20 |
39 |
Sanipress unit |
Art.No. 50-09-000 |
20 |
39 |
Tournipress |
Art.No. 53-05-000 |
21 |
40 |
Seite / Page 4
Repair instruction for Speidel + Keller blood pressure units
Failure search mechanical blood pressure units
--------------------------------------------------------------------------------------------------------------------------------------
Units / Products concerned: |
|
|
------------------------------------ |
|
|
- Primus |
50/68 1 and 2 tube KC and Metal |
Art. No. 50-01-000/ 50-04-000/ 50-81-000/ 50-84-000 |
- Manuell 50/68 1 and 2 tube KC and Metal |
Art. No. 50-16-000/ 50-17-000/ 50-96-000/ 50-97-000 |
|
- Primus Stabil 3 and Primus Persona |
Art No. 50-25-000, 50-49-000 /50-59-000/ 50-50-000/ |
|
included the colour serie |
50-45-000 |
|
- Manuell |
Stabil 3 |
Art.No. 50-63-000 |
- Disytest |
|
Art. No. 52-09-000 |
- ISO Stabil 5 |
Art. No. 50-71-000 |
|
- Fixomed |
|
Art. No. 50-91-000 |
- Hand 2001 und 2003 |
Art. No. 54-03-000, 54-07-000 |
|
- Oscillophon |
Art. No. 51-05-000 |
|
- Minimus |
|
Art. No. 51-02-000 |
- Maxi Stabil 3 |
Art. No. 59-32-000 |
|
1. Failure search - Locating missing parts |
|
|
--------------------------------------------------------- |
|
|
- Cuff - Glass - Scale - Valve parts - Ball - |
Screws - Washer - other Accessories |
2.Search on cause and reason for failure Possible causes/- reasons:
------------------------------------------------------
-Leaking membran/ base plate measurement inspection
-Needle outside the tolerance on the zeropoint
-Hanging needle / jerky movements Propably bent movement axle movement spring tangled in the segment
-Glass / Scale fallen out, broken bent, scratched etc.
(dropped unit)
Neccessary repair
Upcoming repair:
----------------------------------------------------------------------
Replace complete base plate and movement +
Replace needle + measurement inspection + eventually replace complete base plate and movement
Replace complete base plate with movement + carry out measurement inspection
Insert glass and scale, carry out |
measurement inspection |
- Leaking valve part, tube valve version |
Replace valve + leakage test |
- Leak on the cuff |
Replace rubber bag + leakage test |
- Leak on tube connector/other connector pieces/ |
Replace all defect parts + carry out leakage test |
part valve |
|
- Hart turning release screws |
Replace release screw + leakage test |
- Defect trigger / bezel |
Replace trigger + test bezel on function |
- Leak on valve part, trigger version |
Replace complete valve group + leakage test |
- No air relase or air admittance possible |
Clean air inlet and outlet, possible replacement of the |
release valve + test on fuction |
|
2.Search on cause and reason for failure Possible causes/- reasons:
---------------------------------------------------------
-Not possible to inflate the ball / defect Possible damaged check valve /
or rip in material (also hand valve)
-Defect valve body / screw / trigger version body + measurement inspection
-Missing washer
-Expired calibration time /
Seite / Page 5
Neccessary repair
Upcoming repair:
---------------------------------------------------------------------
Assemble new ball + test on function
Take out the complete body and re-assemble it with new
Replace washer + leakage test
Carry out measurement inspection + attach test label Prepare documentation (see instruction on measurement inspection
Instructions on measurement inspection on all blood pressure units - please see attachment Measurement inspections.
Instructions on work procedures for dismantling and re-assembling (see attached description)
Seite / Page 6
Repair instruction for Speidel + Keller blood pressure units
Mercury units
Units |
Products concerned: |
|
- |
Miniatur 300 B |
Art. No. 65-11-000 |
- |
Miniatur 300 S |
Art. No. 65-13-000 |
- |
Mercuro 300 |
Art. No. 74-01-000 |
1. |
Failure search - Locating missing parts |
|
|
- Cuffs - Glass tube - Filter caps - Washers - |
Cover - Ball valve - other accessories |
2. |
Search on cause and reason for failure |
Neccessary repair |
|
Possible causes/- reasons: |
Upcoming repair: |
------------------------------------------------- |
-------------------------------------------- |
|
- |
Broken / dirty glass tube |
Replace or clean glass tube + measurement inspection |
- |
Defect mercury container / ripped |
Assemble new mercury container + insert mercury + carry out |
|
|
measurement inspection. |
- |
Bezel does not shift, no mercury inlet |
Assemble bezel and container + pur in new mercury |
|
possible |
+ measurement inspection |
- |
Mercury outflow |
Evaluate the leak place, replace defect parts + fill up |
|
mercury + carry out measurement inspection |
(Caution: Contact with mercury is |
|
|
|
|
|
hazardous to your health) |
- |
Defect /leak tube connector |
Insert new washer and assemble tube connector + refill |
|
mercury + carry out measurement inspection |
|
- |
Defect filter or filter cap/ leaking |
Insert new filter and filter cap + refill mercury + carry out |
|
loosen, etc. |
+ measurement inspection |
- |
Defect housing parts / scale |
Replace parts + carry out function test. |
- |
Expired calibration time |
Carry out measurement inspection + attach new test label with |
|
|
information on the year. |
If you need more information about the assembling combinations of the single parts please use the
exploding pictures from the following Spare Part Catalogue Page 28 –30.
Instructions on measurement inspection on all blood pressure units - please see attachment Measurement inspections.
Instructions on work procedures for dismantling and re-assembling ( see attached description
Seite / Page 7
Repair instruction for Speidel + Keller blood pressure units |
Rev./ Index: 001B |
||||
Unit type: |
Primus 50 metal |
Art.No. 50-01-000 |
/ Manuell 50 metal |
Art.No. 50-16-000 |
|
------------- |
Primus 68 metal |
Art.Nr. 50-04-000 / |
Manuell 68 metal |
Art.No. 50-17-000 |
|
Dismantle: |
Instruction of complete dismantling and re-assembling of the unit |
||||
1.1 |
unscrew glass holder 03/43-901/ 03-44-901. |
|
|||
---------------- |
1.2 |
loosen glass 03-44-902 or glass remains with screwdriver (caution: adhered) |
1.3loosen pointer 18-50-923 / 18-87-906 with slip device resp. tweezers
1.4remove scale 20-64-151 (caution: adhered)
1.5unscrew the complete base plate with movement 01-21-000 on hand of the base plate tool.
1.6unscrew cup screws 50-01-905 with screwdriver.
remove the seal 50-01-901 and cup 50-01-904 / 50-04-901 at 68mm
1.7loosen the safety pin 19-40-905 with screwdriver and remove the release screw 19-40-901. Remove the single parts spring 19-40-904 and washer 19-40-903.
1.8remove the ball 17-31-902 and span ring 17-31-911 from the body and loosen valve screw 15-41-000 with special tool.
Point 1.8 not ablicaple for units manufactured since October 1999. loosen span ring15-40-903 with a tweezers, remove seal 15-40-902 and valve tube 15-40-907 with special tool
1.9loosen spoon screws 50-01-902 with screwdriver
Re-assemble: |
2.1 |
screw on the scoop 50-01-902 with screwdriver. |
----------------- |
2.2 |
screw on new cup with screwdriver onto the valve body 50-01-900 |
|
2.3 |
insert washer 19-40-903 and spring into valve body. |
|
|
screw on the release screw and push the clip ring back in. |
|
2.4 |
For units from October 99 assemble new valve 15-40-907, seal 15-40-902 and new circlip |
|
15-40-903 with special tool. |
|
|
2.5 |
Insert washer 50-01-901 and complete base plate with movement with spec. base plate |
|
tool. |
|
|
2.6 |
Put unit on calibration device and put on the special S+K test scale |
|
2.7 |
Put on needle 18-50-923 / 18-87-906 and adjust with the 0-line |
|
2.8 |
Go to pressure level 300mmHg and safe. Drive pressure level 50, 100, 150, 200, 250, |
|
300, |
350 mmHg upand downwards, test tolerance +/- 3mmHg / and if neccessary re- |
|
calibrate with a special calibration pen. |
|
|
2.9 |
Set needle on the 0-position and impress with special trigger. |
|
2.10 |
Insert new glass 03-44-904 and glue it on. (Glue Sicomet 9010) |
|
2.11 |
screw on the glass ring 03-43-901/ 03-44-901. |
End-evaluation / |
|
Carry out end-evaluation according to need of repair (see repair instruction |
Function / |
|
screw valve version,) |
Leakage test: |
|
If operating on the unit or the mechanism a complete end-evaluation has to be carried |
|
|
out, including measurement inspection (see realization of mesurement technical |
|
|
inspections after MPG.) |
|
- |
If unsure, if the unit was dropped, we suggest to carry out a measurement technical |
|
|
inspection. |
Remarks: The neccessary repair expenditure has to be established and carried out. The needed
upcoming procedure can be taken out of the complete insctruction.
If you need more information about the assembling combinations of the single parts please use the
exploding pictures from the following Spare Part Catalogue Page 3, 5, 7, 9
Seite / Page 8
Repair instruction for Speidel + Keller Blood pressure unit |
Rev./ Index: 001B |
|||
unit type: |
- Primus 50 |
KC Art.No. 50-81-000 / Manuell 50 KC Art.No. 50-96-000 |
||
------------- |
- Primus 68 |
KC Art.No. 50-84-000 / Manuell 68 KC Art.No. 50-97-000 |
||
|
Instruction of dismantling and re-assembling the unit |
|
||
Dismantle: |
1.1 |
clip out glass with a screwdriver on the bezel |
|
|
--------------- |
1.2 |
loosen glass 03-44-902 or glass remains with screwdriver (Caution: adhered) |
1.3loosen needle 18-50-923/ 18-87-906 with slip device resp. tweezers
1.4remove scale 20-64-151 (caution: adhered)
1.5unscrew the complete base plate with movement 01-21-000 on hand of the base plate tool.
1.6unscrew cup screws 50-01-905 with screwdriver.
remove the seal 50-01-901 and cup 50-01-904 / 50-04-901 at 68mm
1.7loosen the safety pin 19-40-905 with screwdriver and remove the release screw 19-40-901. Remove the single parts spring 19-40-904 and washer 19-40-903.
1.8remove the ball 17-31-902 and span ring 17-31-911 from the body and loosen valve screw 15-41-000 with special tool
Point 1.8 not ablicaple for units manufactured since October 1999. loosen span ring15-40-903 with a tweezers, remove seal 15-40-902 and valve tube 15-40-907 with special tool
1.9loosen spoon screws 50-01-902 with screwdriver
Re-assemble: |
2.1 |
screw on the scoop 50-01-902 with screwdriver. |
----------------- |
2.2 |
screw on new cup with screwdriver onto the valve body 50-01-900 |
|
2.3 |
insert washer 19-40-903 and spring into valve body. |
|
|
screw on the release screw and push the clip ring back in. |
|
2.4 |
For units from October 99 assemble new valve 15-40-907, seal 15-40-902 and new circlip |
|
15-40-903 with special tool. |
|
|
2.5 |
Insert washer 50-01-901 and complete base plate with movement with spec. base plate |
|
tool. |
|
|
2.6 |
Put unit on calibration device and put on the special S+K test scale |
|
2.7 |
Put on needle 18-50-923 and adjust with the 0-line |
|
2.8 |
Go to pressure level 300mmHg and safe. Drive pressure level 50, 100, 150, 200, 250, |
|
300, |
350 mmHg upand downwards, test tolerance +/- 3mmHg / and if neccessary re- |
|
calibrate with a special calibration pen. |
|
|
2.9 |
Set needle on the 0-position and impress with special trigger. |
|
2.10 |
Insert new glass 03-44-904 and glue it on. (Glue Sicomet 9010) |
|
2.11 |
clip the glass back in to the bezels. |
End-evaluation / |
|
Carry out end-evaluation according to need of repair (see repair instruction |
Function / |
|
screw valve version, page 1) |
Leakage test: |
|
If operating on the unit or the mechanism a complete end-evaluation has to be carried |
|
|
out, including measurement inspection (see realization of mesurement technical |
|
|
inspections after MPG.) |
|
- |
If unsure, if the unit was dopped, we suggest to carry out a measurement technical |
|
|
inspection. |
Remarks: The neccessary repair expenditure has to be established and carried out. The needed
upcoming procedure can be taken out of the complete instructions.
If you need more information about the assembling combinations for single parts please use the exploding
pictures from the following Spare Part Catalogue Page 4, 6, 8, 10
Seite / Page 9
Repair instruction for Speidel + Keller Blood pressure units |
Rev./ Index: 001B |
||
Model: |
- Primus Stabil 3 Art.No. 50-25-000, 50-49-000, 50-59-000, 50-50-000, 50-45-000 |
||
---------- |
- Manuell Stabil 3 Art.No. 50-63-000 |
|
|
|
Instruction of dismantling and re-assembling the unit |
|
|
Dismantle: |
1.1 |
turn out the bezel 50-25-908 clock wise. |
|
--------------- |
1.2 |
clip out the glass 03-44-904 with a screwdriver on the bezel (Caution: adhered) |
1.3loosen needle 18-50-921 with slip device resp. tweezers
1.4remove scale 20-64-161 (caution: adhered)
1.5unscrew the complete base plate with movement 01-21-000 on hand of the base plate tool.
1.6unscrew cup screws 50-01-905 with screwdriver. remove the seal 50-01-901 and cup 50-01-907:
1.7loosen the safety pin 19-40-905 with screwdriver and remove the release screw 19-40-901. Remove the single parts spring 19-40-904 and washer 19-40-903.
1.8remove the ball 17-31-902 and span ring 17-31-911 from the body and loosen valve screw 15-41-000 with special tool
Point 1.8 not ablicaple for units manufactured since October 1999. loosen span ring15-40-903 with a tweezers, remove seal 15-40-902 and valve tube 15-40-907 with special tool
1.9loosen spoon screws 50-01-902 with screwdriver .
Re-assemble: |
2.1 |
screw on the scoop 50-01-902 with screwdriver. |
------------ |
2.2 |
screw on new cup with screwdriver onto the valve body 50-01-900 |
|
2.3 |
insert washer 19-40-903 and spring into valve body. |
|
|
screw on the release screw and push the clip ring back in. |
|
2.4 |
For units from October 99 assemble new valve 15-40-907, seal 15-40-902 and new circlip |
|
15-40-903 with special tool. |
|
|
2.5 |
Insert washer 50-01-901 and complete base plate with movement with spec. base plate |
|
tool. |
|
|
2.6 |
Put unit on calibration device and put on the special S+K test scale |
|
2.7 |
Put on needle 18-50-921 and adjust with the 0-line |
|
2.8 |
Go to pressure level 300mmHg and safe. Drive pressure level 50, 100, 150, 200, 250, |
|
300, |
350 mmHg upand downwards, test tolerance +/- 3mmHg / and if neccessary re- |
|
calibrate with a special calibration pen. |
|
|
2.9 |
Set needle on the 0-position and impress with special trigger. |
|
2.10 |
Insert new glass 03-44-904 and glue it on. (Glue Sicomet 9010) |
|
2.11 |
bring on the bezel 50-25-908 (double noses) and turn it glockwise until stop. |
End-evaluation / |
|
Carry out end-evaluation according to need of repair (see repair instruction |
Function / |
|
screw valve version, page 1) |
Leakage test: |
|
If operating on the unit or the mechanism a complete end-evaluation has to be carried |
|
|
out, including measurement inspection (see realization of mesurement technical |
|
|
inspections after MPG.) |
|
- |
If unsure, if the unit was dopped, we suggest to carry out a measurement technical |
|
|
inspection. |
Remarks: |
|
The neccessary repair expenditure has to be established and carried out. The needed |
upcoming procedure can be taken out of the complete instructions.
If you need more information about the assembling combinations for single parts please use the exploding pictures from the following Spare Part Catalogue Page 13 - 18.
Seite / Page 10
Repair instruction for Speidel + Keller Blood pressure units |
Rev./ Index: 001B |
||
Model: |
- ISO Stabil 5 Art. No. 50-71-000 |
|
|
------------- |
- Stabil 3 (predecessor. model Stabil 5) |
|
|
|
Instruction of dismantling and re-assembling the unit |
|
|
Dismantle: |
1.1 |
turn out the bezel 50-25-908 clock wise. |
|
--------------- |
1.2 |
clip out the glass 03-44-904 with a screwdriver on the bezel (Caution: adhered) |
1.3loosen needle 18-50-921 with slip device resp. tweezers
1.4remove scale 20-64-161 (caution: adhered)
1.5unscrew the complete base plate with movement 01-21-000 on hand of the base plate tool.
1.6unscrew cup screws 50-01-905 with screwdriver and then remove the cup 50-25-905
1.7loosen the closing ring o n the ball 17-31-00 0 and then take it off the valve
1.8release trigger 50-71-812 from valve body and take out pin 50-61-813
1.9unscrew check valve 15-40-000 with special tool. For units manufactured since October 1999.
loosen span ring15-40-903 with a tweezers, remove seal 15-40-902 and valve tube 15-40-907 with special tool
Re-assemble: |
2.1 |
Insert pin 50-61-813 though drilling o n the valve body. |
------------ |
|
build in feather 54-31-904 on valve body and then insert trigger into valve body |
|
|
push trigger pin with special pin forward and release trigger. |
|
|
Carry out functional test ( should run easy without creaking noise up to the stop) |
|
2.2 |
Screw in new valve screw 15-41-000 and screw the ball 17-31-000 back on. |
|
|
Turn on check valve as safety device. |
|
2.3 |
Put the cup 50-25-905 onto the valve body and screw it in. |
|
2.4 |
Insert washer 50-01-901, then complete ground plate with movement and screw |
|
|
it in with the special base plate tool. |
|
|
For units manufactured since October 1999. |
|
|
loosen span ring15-40-903 with a tweezers, remove seal 15-40-902 and |
|
|
valve tube 15-40-907 with special tool. |
|
|
Assemble ball tube 17-31-911 and ball 17-31-902 |
|
2.5 |
Put unit into the calibration device and put on the special S+K test scale. |
|
2.6 |
Bring on trigger 18-50-921 and adjust it to the zero-line. |
|
2.7 |
Start pressure 300 mmHg and adjust. |
|
|
Pressure levels 50, 100, 150, 200, 250, 300 , 350 mmHg drive upand downwards |
|
|
and test tolerance +/- 3mmHg / if necessary re-adjust with special calibration pin. |
|
2.8 |
Set new trigger to zero-position and impress with a special printer. |
|
2.9 |
Eventually assemble (engage) cub with a new spacer ring 50-25-912. |
|
2.10 |
clip and clue in new glass 03-44-904 (Sicomet 9010). |
|
2.11 |
Put on bezel 50-25-908 ( 2 nose whiskers left wing) and turn it clockwise until stop. |
End-evaluation / Carry out end-evaluation according to need of repair
Function /
Leakage test: If operating on the unit or the mechanism a complete end-evaluation has to be carried out, including measurement inspection (see realization of mesurement technical inspections after MPG.)
-If unsure, if the unit was dopped, we suggest to carry out a measurement technical inspection.
Remarks: The neccessary repair expenditure has to be established and carried out. The needed
upcoming procedure can be taken out of the complete instructions.
Seite / Page 11
Repair instruction for Speidel + Keller Blood pressure units |
Rev./ Index: 001B |
||
Model: |
- Disytest / Disytest HomeArt.No. 52-09-000 |
|
|
------------- |
|
|
|
|
Instruction of dismantling and re-assembling the unit |
|
|
Dismantle: |
1.1 |
Enter with screwdriver and clip out glass 52-03-903. |
|
--------------- |
1.2 |
Loosen needle 18-50-924 / 18-50-923 (Persona) with the slip device resp. tweezers |
1.3Clip out the scale 20-51-106 / 20-51-103 (Persona) as you have down with the glass.
1.4Unscrew the complete base plate with movement 01-21-000 on hand of the base plate tool, take of the washer 50-01-901.
1.5Remove ball 17-52-000 and span ring 17-01-902.
1.6Turn off the check valve 15-40-000 with special tool. For units manufactured since October 1999.
loosen span ring15-40-903 with tweezers, remove seal 15-40-902 and valve tube 15-40-907 with special tool.
1.7Release trigger 52-03-902 on the valve body, with special tool (inside, six-cant wrench) turn off trigger release valve 12-12-900, take out spring 19-12-904.
Re-assemble: |
2.1 |
Insert spring 19-12-904 in the valve body , insert trigger release valve 19-12-900 |
----------------- |
|
and tighten it with the special six-cant wrench, press in trigger 52-03-902 |
|
|
to the grid - Caution: test trigger function - should have easy play and easy handling |
|
|
between the trigger release valves. |
|
2.2 |
Screw in new check valve 15-41-000, push in span ring 17-01-902 and ball 17-52-000. |
|
|
For units manufactured since October 1999. |
|
|
Assemble new valve 15-40-907, seal 15-40-902 and circlip 15-40-903 with special tool. |
|
|
Assemble ball tube 17-31-911 and Ball 17-31-902 |
|
2.3 |
Insert washer 50-01-901 and screw in the complete base plate including movement with |
|
|
the special base plate tool. |
|
2.4 |
Put the unit up onto the calibration device and insert the special S+K test scale „Disytest“. |
|
2.5 |
Put on needle 18-50-924 and adjust to the zero-line. |
|
2.6 |
Bring pressure to 300 mmHg and adjust Pressure levels 50, 100, 150, 200, 250, 300 , |
|
|
350 mmHg run up and downwards, test on tolerance +/- 3mmHg , if necessary adjust |
|
|
with special calibration pin. |
|
2.7 |
Set new pointer on zero-position and press on with special trigger. |
|
2.8 |
Clip in new glass 52-03-903. |
End-evaluation / Carry out end-evaluation according to need of repair
Function /
Leakage test: If operating on the unit or the mechanism a complete end-evaluation has to be carried out, including measurement inspection (see realization of mesurement technical inspections after MPG.)
-If unsure, if the unit was dropped, we suggest to carry out a measurement technical inspection.
Remarks: The neccessary repair expenditure has to be established and carried out. The needed
upcoming procedure can be taken out of the complete instructions.
If you need more information about the assembling combinations for single parts please use the exploding
picture from the following Spare Part Catalogue Page 11.
Seite / Page 12
Repair instruction for Speidel + Keller Blood pressure unit |
Rev./ Index: 001B |
|||
Model: |
- Maxi Stabil 3 |
Art.No. 59-30-000 |
|
|
------------ |
|
|
|
|
|
Instruction of dismantling and re-assembling the unit |
|
||
Dismantle: |
1.1 |
Turn bezel 59-30-901 anticlockwise and remove the housing |
|
|
--------------- |
|
(Caution bezel may be adhered - dissolve). |
|
|
|
1.1 |
Enter the groove with a scrwedriver and clip out the glass 59-30-902. |
||
|
1.2 |
Loosen needle 18-87-908 with the slip device resp. tweezers |
|
|
|
1.3 |
Clip out the scala 20-51-159 with a screwdriver (as done with the Glas). |
||
|
1.4 |
Unscrew the complete base plate with movement 01-10-000 with |
the base plate |
|
|
|
tool, take of the washer 50-01-915. |
|
|
|
1.5 |
Unscrew the metal tube connector 16-40-000 and knurled nut 59-30-922 |
||
|
1.6 |
CAUTION: |
No further dismantling possible, because the air release connector is |
|
|
|
clued. A dismantling of the co nnector could damage the housing. If this connector is |
||
|
|
damaged the complete housing 59-30-750 has to be exchanged. |
|
|
Re-assemble: |
2.1 |
Screw in the metal tube connector 16-40-000 and knurled nut 59-30-922 |
||
----------------- |
2.2 |
Insert washer 50-01-901 and screw the complete base plate inclusive movement back in |
||
|
|
with the special base plate tool. |
|
|
|
2.3 |
Put the unit on the calbration device and bring on the special S+K test scale „Maxi. |
||
|
2.4 |
Put on the needle 18-87-908 and adjust on the 0-line. |
|
|
|
2.5 |
Bring pressure to 300 mmHg and adjust Pressure levels 50, 100, 150, 200, 250, 300 , |
||
|
|
350 mmHg run up and downwards, test on tolerance +/- 3mmHg , if necessary adjust |
||
|
|
with special calibration pin. |
|
|
|
2.6 |
Impress new scala (prescribed groove). |
|
|
|
2.7 |
Set the new pointer on zero-position and impress with special press. |
||
|
2.8 |
Clip in new glass 59-30-902. |
|
End-evaluation / Carry out end-evaluation according to need of repair
Function /
Leakage test: If operating on the unit or the mechanism a complete end-evaluation has to be out, including measurement inspection (see realization of mesurement technical inspections after MPG.)
-If unsure, if the unit was dopped, we suggest to carry out a measurement technical inspection.
Remarks: The neccessary repair expenditure has to be established and carried out. The needed upcoming procedure can be taken out of the complete instructions.
If you need more information about the assembling combinations for single parts please use the exploding
pictures from the following Spare Part Catalogue Page 19 and 20.
Seite / Page 13
Repair instruction for Speidel + Keller Blood pressure units |
Rev./ Index: 001B |
|||
Model: |
- Fixomed |
Art. No. 50-91-000 |
|
|
------------- |
|
|
|
|
|
Instruction of dismantling and re-assembling the unit |
|
||
Dismantle: |
1.1 |
Unscrew the safety ring 03-43-901. |
|
|
--------------- |
1.2 |
Loosen glass 03-42-902 or glass remains with screwdriver (Caution: adhered) |
1.3Loosen needle 18-87-905 with the slip device resp. tweezers.
1.4Remove the scala 20-57-122 (Caution: adhered)
1.5Unscrew the complete base plate and movement 01-21-000 with base plate tool.
1.6Unscrew the flat head screws 50-01-905 with screwdriver. Remove washer 50-01-901 and cup 50-04-901 (68mm).
1.7Pull off the ball 17-31-902 and span ring 17-31-901 from the body and unscrew the check valve 15-41-000 with special wrench
For units manufactured since October 1999.
Loosen span ring15-40-903 with tweezers, remove seal 15-40-902 and valve tube 15-40-907 with special tool.
1.8Loosen spoon 50-01-902 with screwdriver.
1.9A dismantling of the release valve should not be carried out, because the re-assembling is extremely difficult.
If the release valve is damanged, the complete valve body has to be replaced 50-85-000.
Re-assemble: |
2.1 |
Screw on the spoon 50-01-902 with screwdriver. |
------------ |
2.2 |
Screw on the new cups with screwdriver on the valve body 50-85-900. |
|
2.3 |
Screw in new check valve 15-41-000, and assemble new span ring 17-31-901 and Ball |
|
|
17-31-902. |
|
|
For units manufactured since October 1999. |
|
|
Assemble new valve 15-40-907, seal 15-40-902 and circlip 15-40-903 with special tool. |
|
|
Assemble ball tube 17-31-911 and Ball 17-31-902 |
|
2.4 |
Insert washer 50-01-901 and screw on the complete base plate with special base plate tool |
|
2.5 |
Put unit into the calbration device and bring on the special test scale "Fixomed" |
|
2.6 |
Put on needle 18-87-905 and adjust it to the zero-line. |
|
2.7 |
Bring pressure to 300 mmHg and adjust Pressure levels 50, 100, 150, 200, 250, 300 , |
|
|
350 mmHg run up and downwards, test on tolerance +/- 3mmHg , if necessary adjust |
|
|
with special calibration pin. |
|
2.8 |
Insert scala 20-57-122 in cup. |
|
2.9 |
Set needle on zero-position and impress with special press. |
|
2.10 |
Insert new glass 03-42-902. |
|
2.11 |
Screw on new safety glass 03-43-901 and tighten it. |
End-evaluation / Carry out end-evaluation according to need of repair
Function /
Leakage test: If operating on the unit or the mechanism a complete end-evaluation has to be carried out, including measurement inspection (see realization of mesurement technical inspections after MPG.)
-If unsure, if the unit was dopped, we suggest to carry out a measurement technical inspection.
Remarks: The neccessary repair expenditure has to be established and carried out. The needed
upcoming procedure can be taken out of the complete instructions.
Seite / Page 14
Repair instruction for Speidel + Keller Blood pressure units |
Rev./ Index: 001B |
|||
Model: |
- Hand 2003 |
Art. No. 50-03-000 |
|
|
------------- |
|
|
|
|
|
Instruction of dismantling and re-assembling the unit |
|
||
Dismantle: |
1.1 |
Unscrew safety glass 03-43-901. |
|
|
--------------- |
1.2 |
Loosen glass 03-42-902 or glass remains with screwdriver (Caution: adhered) |
1.3Loosen needle 18-87-906 with the slip device resp. tweezers.
1.4Remove scala 20-57-940 (Caution: adhered)
1.5Unscrew complete base plate incl. movement 01-21-000 with special base plate tool and remove washer 50-01-901.
1.6Loosen and unscrew the span ring ot the ball 17-31-000. Unscrew check valve 15-40-000 with special tool.
1.7Loosen spoons 50-01-902 with screwdriver.
1.8A dismantling of the release valve should not be carried out, because the re-assembling is extremely difficult.
If the release valve is damanged, the complete valve body has to be replaced 50-85-000.
Re-assemble: Goes the same for the unit Hand Version 2001 (50 mm)
----------------- 2.1 Screw the new check valve 15-40-000 in to the complete valve part 54-07-700 put up the new washer 54-31-910 and screw the Ball 17-31-000 on to the housing.
2.2Insert washer 50-01-901 and screw on the complete base plate with special base plate tool
2.3Put unit into the calbration device and bring on the special test scale 20-57-940
2.4Put on needle 18-87-906 and adjust it to the zero-line.
2.7Bring pressure to 300 mmHg and adjust Pressure levels 50, 100, 150, 200, 250, 300 , 350 mmHg run up and downwards, test on tolerance +/- 3mmHg , if necessary adjust with special calibration pin.
2.8Insert scala 20-57-940 in cup.
2.9Set needle on zero-position and impress with special press.
2.10Insert new glass 03-42-902.
2.11Screw on new safety glass 03-43-901 and tighten it.
End-evaluation / Carry out end-evaluation according to need of repair
Function /
Leakage test: If operating on the unit or the mechanism a complete end-evaluation has to be carried out, including measurement inspection (see realization of mesurement technical inspections after MPG.)
-If unsure, if the unit was dopped, we suggest to carry out a measurement technical inspection.
Remarks: The neccessary repair expenditure has to be established and carried out. The needed
upcoming procedure can be taken out of the complete instructions.
If you need more information about the assembling combinations for single parts please use the exploding
picture from the following Spare Part Catalogue Page 23.
Seite / Page 15
Repair instruction for Speidel + Keller Blood pressure units |
Rev./ Index: 001B |
||
Model: |
- Oscillophon Art. No. 51-05-000 |
|
|
------------- |
|
|
|
|
Instruction of complete dismantling and re-assembling of the unit |
||
Dismantle: |
1.1 |
Dismantle unit from the cuff with screwdriver |
|
-------------- |
1.2 |
Unscrew safety ring 03-44-901. |
|
1.3Loosen glass 03-44-902 or glass remains with screwdriver (Caution: adhered)
1.4Loosen needle 18-50-923 with the slip device resp. tweezers.
1.5Remove scala 20-57-940 (Caution: adhered)
1.6Loosen three screws with screwdriver and complete base plate with movement 01-22-000 plus take out three washers 51-01-901.
Assemble:
------------- 2.1 Insert the three washer 51-01-901 and screw the three screws 01-21-901 back in with a
screwdriver and tighten the complete base plate with the movement
2.2Put unit into the calbration device and bring on the special test scale 20-64-151
2.3Put on needle 18-50-923 and adjust it to the zero-line.
2.4Bring pressure to 300 mmHg and adjust Pressure levels 50, 100, 150, 200, 250, 300 , 350 mmHg run up and downwards, test on tolerance +/- 3mmHg , if necessary adjust with special calibration pin.
2.5Insert scala 20-64-151 in cup. (note: inlay groove)
2.6Set needle on zero-position and impress with special press.
2.7Insert new glass 03-44-902.
2.8Screw on new safety glass 03-44-901 and tighten it.
End-evaluation / Carry out end-evaluation according to need of repair
Function /
Leakage test: If operating on the unit or the mechanism a complete end-evaluation has to be carried out, including measurement inspection (see realization of mesurement technic inspections after MPG.)
-If unsure, if the unit was dopped, we suggest to carry out a measurement technical inspection.
Remarks: The neccessary repair expenditure has to be established and carried out. The needed
upcoming procedure can be taken out of the complete instructions.
If you need more information about the assembling combinations for single parts please use the exploding
picture from the following Spare Part Catalogue Page 12.
Seite / Page 16
Instruction for end evaluation including measurement inspections
1. Visual test:
-Check complete unit (chrome damages, etc).
-Check if pointer position is at zero range
2. Measurement test
-Connected unit to the certified pressure gauge class 0,1 (max. deviation 0,8 mmHg).
-Accuracy statistical pressure / Hysterisis on pressure values 50,100,150,200,250,300, mmHg test up-and downwards Tolerance = < 3 mmHg on all test points / Hysterisis = < 4 mmHg on all test points. - start pressure decrease 300 mmHg /seal unit system and test leakage. Pressure decrease should stay
at = < 4 mmHg/min.
3. Function test
-Connect unit with complete cuff onto the test tube, girth appox. 35 cm and pump up to 300 mmHg /test cuff for leakage, then open air release valve / should turn easy / test pointer run and test speed outlet. During the speed air release the time for pressure decrease from 260 mmHg to 15 mmHg should not exceed 10 Sec.
Run of the release screws / test printer - trigger grid (ie. Disytest)
4. |
Test Documentation |
- |
Documentation as by S+K requested test report, entry of unit type, S/N, established deviations, leakage, test |
|
date, tester, user, remarks. |
|
These documentation has to be made available for S+K. |
5. |
Following function |
|
Fasten new test label with year when test was carried out. |
6. |
Important remarks: |
- |
Measurement technical controls with confirmation of a test label can only be carried out on CE - marked |
|
blood pressure meters. Older units with PTB-Registration should be submitted to the notified body after |
|
inspection for calibration as done in the past. |
Seite / Page 17
Tools: |
Description |
Article/ Order No. |
Complete tool set |
|
94-01-000 |
Base plate wrench metal |
for movement assembling (all types) |
94-01-101 |
Test scale 50 mm metal |
for unit calibration |
94-01-103 |
Test scale 68 mm metal |
for unit calibration |
94-01-104 |
Test scale Disy |
for unit calibration |
94-01-105 |
Calibration stick |
for unit calibration |
94-01-106 |
Needle presser |
for needle assembling |
94-01-107 |
Wrench SW3,5 |
for release assembling |
94-01-108 |
Wrench SW6,5 |
for valve assembling |
94-01-109 |
Wrench SW6 (inside) |
for valve part assembling |
94-01-110 |
Square wrench for seal |
Seal dismantling (old units) |
94-01-111 |
Test scale 50 mm KC |
for unit calibration |
94-01-112 |
Test scale 68mm KC |
for unit calibration |
94-01-113 |
Tweezcers size AA |
for various operations |
94-01-115 |
Test scale for Stabil |
for unit calibration |
94-01-116 |
Test scale Maxi |
for unit calibration |
94-01-117 |
Calibration stand |
for unit calibration |
94-01-118 |
Neddle disconnector |
for needle dismantling |
94-01-119 |
Screw driver (fine) 1,5 |
assembling tool for various units |
94-01-120 |
Screw driver(fine) 2,0 |
assembling tool for various units |
94-01-121 |
Pinch pliers |
for ball sleve assemble / pinching |
94-01-123 |
Spare part list + repair instructions |
for all blood pressure units |
94-01-130 |
Ordering of single tools is possible. Please direct your orders to a S+K Sales department.
Other necessary tools:
----------------------------
Test gauge |
Measurement sector at least 0-300 mmHg |
+ Certificate |
with evidence Accuracy= < 0,8 mmHg |
Spanblock |
for dismantling the seal |
Adjustment plate |
for adjusting defect parts |
self purchase nec. / on request also from S+K
self purchase nec.
self purchase nec.
Seite / Page 18
Repair instruction for Speidel + Keller blood pressure units "Accessories for blood pressure units"
Failure search mechanical blood pressure units
Units / Products concerned: |
|
|
- Membranstetoscope for self mesurement |
Art. No. 84-10-000 (integrated in cuff) |
|
- Planoskop 40 mm |
Art. No. 83-11-080 |
|
- Planoskop 30 mm (child) |
Art. No. 83-12-080 |
|
- Duophon |
40 mm |
Art. No. 82-11-080 |
- Duophon |
30 mm (child) |
Art. No. 82-15-080 |
- Harvey Elite |
all models |
|
- Professional |
all models |
1. |
Failure search - Locating missing parts |
|
----------------------------------------------------- |
|
|
- |
Missing membran - Safety ring - Ear tips - |
Membran holder - Y-Tube |
2. |
Search on cause and reason for failure |
Neccessary repair |
----------------------------------------------------- |
|
----------------------------------------- |
- |
Defect / broken membran |
Built in new membran + carry out function test |
|
(can occur on all models) |
|
- |
Bent / defect ear piece |
Fit new ear piece + funtion test |
- |
Defect / plugged / missing |
Assemble new ear tips + function test |
|
(can occur on all models) |
|
- |
Defect / missing membran holder |
Assemble new membran holder + function test |
- |
Y-tube porous / defect. Ripped material, etc. |
Assemble new Y-tube + function test |
|
(all models) |
|
- |
Tube connector turns / can not be |
Insert tube connector + function test |
|
assembled (only on membran stethoscope |
|
|
for self measurement) |
|
-Not possible to switch over the Duophon shaft bent.
Dismantle and re-assemble (only Duophon) eventually assemble new shaft + function test
Instructions for function test / Membranstethoscope integrated in cuff "Self measuring"
1.Apply cuff on upper arm
2.Connect tube connector on S+K blood pressure unit.
3.Insert ear tips into the auditory canal
4.Inflate cuff (approx. 30 mmHg over expected blood pressure)
5.Release air and carry out the complete measurement.
6.Korotoff sounds should be well heared from the start of the Systole till the Diastole.
Planoskop / Professional
1.Apply cuff on upper arm
2.Push the Planoskop und er the Cuff (Membran part skin contact)
3.Insert ear tips into the auditory canal
4.Inflate cuff (approx. 30 mmHg over expected blood pressure)
5.Release air and carry out the complete measurement.
6.Korotoff sounds should be well heared from the start of the Systole till the Diastole.
Duophon / Harvey Elite
1.Switch Duophon several times o ver - should be easy going and engage on half turn. Duophon shaft hole visible - Duophon out of order
Duophon shaft hole not visible - turned to membran - Duo phon on stand by.
Seite / Page 19
Repair instruction for Speidel + Keller blood pressure units
Failure search accessories hand fan
Units / Products concerned:
---------------------------------------
- |
Hand fan |
Art. No. 17-01-000 |
- |
Hand fan |
Art. No. 17-02-000 |
- |
Hand fan |
Art. No. 17-08-000 |
- |
Hand fan |
Art. No. 17-10-000 |
1. |
Failure search - Locating missing parts |
----------------------------------------------------------------
- Trigger - Ball - Ring / Release screw - Valve part
2. Search on cause and reason for failure
------------------------------------------------------
-Not possible to orderly inflate ( defect ball valve)
-Leakage in System
tube connector / release screw
-Worn out release screw
-Defect / missing / broken trigger
Instruction for function test:
All hand valves:
Neccessary repair
----------------------------------------
Inspect ball valve - replace if necessary + carry out function test
Assemble new valve check screw / release screw + carry out function test
Assemble new release screw + function test
Assemble new trigger + function test
-Connect hand fan with connector tube onto a cuff
-Cuff around test tube (Girth approx. 35 cm)
-Interclose test gauge 0-600 mmHg
-Pump in pressure air till approx. 500 mmHg
-Wait for air balance
-Test leakage =< 4 mmHg / min
-Operate release screw and trigger, then carry out function test (Should be easy to operate )
-Test speed air release (the air release with pressed valve should be 260 - 15 mmHg =< 10 seconds).
If you need more information about the assembling combinations for single parts please use the exploding
pictures from the following Spare Part Catalogue Page 24 - 27.
Seite / Page 20
Repair instructions for Speidel + Keller
----------------------------
|
Units / Products concerned: |
|
- |
Pressure infusion cuff |
Art. No. 51-08-000 |
- |
Pressure unit Sanipress |
Art. No. 50-09-000 |
1.Failure Search - Locating missing parts
-Cuff - Glass - Scale - Valve parts - Ball valve - Screws - Washers - other Accessories
2.Search on cause and reason for failure Possible causes/- reasons:
-Leaking membran/ base plate
-Needle outside the tolerance on the zeropoint
-Hanging needle / jerky movements Probably bent movement axle movement spring tangled in the segment
-Glass / Scale fallen out, broken bent, scratched etc.
(dropped unit)
Neccessary repair
Upcoming repair:
Replace complete base plate and movement + measurement inspection
Replace needle + measurement inspection + eventually replace complete base plate and movement
Replace complete base plate with movement + carry out measurement inspection
Insert glass and scale, carry out measurement inspection
- |
Leaking valve part, tube valve version |
Replace valve + leakage test |
- Leak on the cuff |
Replace rubber bag + leakage test |
|
- Leak on tube connector/other connector pieces/ |
Replace all defect parts + carry out leakage test |
|
|
part valve |
|
- Hart turning release screws |
Replace release screw + leakage test |
|
- Defect trigger / bezel |
Replace trigger + test bezel on function |
|
- Leak on valve part, trigger version |
Replace complete valve group + leakage test |
|
- Defect air pump / not possible to inflate |
Replace air pump + leakage test |
|
- No air inlet or release possible |
Clean inand outlets. Replace release screw/ valve if |
|
|
|
necessary + function test |
- No speed air release possible |
Clean inand outlets. Replace release screw/ valve if |
|
|
|
necessary + function test |
- Ball + valve not pumping / d efect |
Assemble new ball + function test |
|
|
possible defect check valve / |
|
|
or rip in material |
|
- |
Defect valve body |
Completely dismantle and reassemble new body + |
|
|
measurement inspection. |
- Missing washer |
Replace washer + leakage test |
Seite / Page 21
Repairs instruction for Speidel + Keller - Medical Special units.
Units / Products concerned:
-Pressure unit Tournipress Art.No. 53-05-000
1.Failure Search - Locating missing parts
-Cuff - Glass - Scale - Valve parts - Ball valve - Screws - Washers - other Accessories
2.Search on cause and reason for failure Possible causes/- reasons:
-Pressure not adjustable / Defect pressure regulator / regulator head
-Aeration not possible, aeration -function valve hangs / defect
-Pressure does not show on display at end pressure 600
Pressure display does not react if the main operation gets cut off.
-Defect connector to supplier / defect air coupling
-Not possible to switch over from distal
to proximal and reverse / defect switch valve ventil defekt
-Timer does not function / no reaction
-Leak membran/base plate from Manometer
-Needle outside the toleranc on the zero point
-Needle hangs / jerky movements Probably bent movement axle/ movement spring tangled up in Segment
-Glass/ Scale fallen out, broken
bent, scratched, etc (dropped unit)
-Leak on cuff
-Leakage on tube connector connector pieces / valve bodies
-No air inlet / release possible
-Accessory ball + valve not possible to inflate / probably defect check valve or rip in material
Neccessary repair
Upcoming repair:
Replace pressure regulator with regulator head , re-adjust pressure, carry out function + leakage test
Assemble new aeration valve + carry out function + leakage test
Assemble new pressure dislay + carry out function + and leakage test
Built in new shutter coupling + carry out function + leakage test
Built in new switch valve + carry out function test
Replace battery and solder new electric connectors + carry out function test.
Replace complete base plate + movement carry out measurement inspection.
Replace needle + measurement inspection - eventually replace base plate including movement
Replace complete base plate and movement + measurement inspection
Insert new glass + scale + carry out measurement
Replace rubber bag + leakage test
Replace all defect parts + carry out leakage test
Clean air inlets and outlets. Possibly replace release screw / valve + carry out function test.
Assemble new ball + function test
- Missing washers |
Replace washers + leakage test |
Seite / Page 22
SERVICE ANLEITUNG
WELCH ALLYN / SPEIDEL +KELLER
Sehr geehrter S+K Servicepartner,
wir freuen uns Sie als Reparaturpartner von Speidel + Keller begrüßen zu dürfen.
Die Fa. Speidel + Keller ist stets bemüht Ihren Servicepartnern optimale Bedingungen zu schaffen, damit eine reibungsloser Service der Produkte gewährleistet ist.
Speidel +Keller Blutdruckmeßgeräte zeichnen sich durch eine qualitativ hochwertige Verarbeitung aus. Dies ist ein Garant dafür, daß S+K Produkte eine sehr lange Lebensdauer besitzen.
Natürlich bedarf auch das beste Gerät einer regelmäßígen Wartung damit stetig ordnungsgemäß und reibungslos damit gearbeitet werden kann.
Zudem ist es eine Notwendigkeit, im Hinblick auf die Bestimmungen des MPG
und der EGMedizinprodukterichtlinien, regelmäßige Wartungen und meßtechnische Kontrollen an Medizinprodukten durchzuführen.
Aus den obengenannten Gründen haben wir Reparaturanweisungen für alle unsere Geräte erstellt.
Diese Anweisungen sollen Ihnen bei der Durchführung einer ordnungsgemäßen Reparatur begleitend helfen.
Seite / Page 23
Fehlersuche Mechanische Blutdruckmeßgeräte
--------------------------------------------------------------------------------------------------------------------------------------
|
Betroffene Geräte / Produkte: |
|
||
----------------------------------------- |
|
|||
- |
Primus |
50/68 |
1 u. 2 Schlauchv. KC und Metall |
Art. Nr. 50-01-000/ 50-04-000/ 50-81-000/ 50-84-000 |
- |
Manuell 50/68 |
1 u. 2 Schlauchv. KC und Metall |
Art. Nr. 50-16-000/ 50-17-000/ 50-96-000/ 50-97-000 |
|
- |
Primus Stabil 3 und Stabil 3 Home |
Art Nr. 50-25-000/ 50-49-000/ 50-59-000/ 50-50-000 |
||
|
und Farbserie gelb,blau,rot,blau |
50-45-000 |
||
- |
Manuell |
Stabil 3 |
Art.Nr. 50-63-000 |
|
- |
Disytest / |
Disytest Home |
Art. Nr. 52-09-000 |
|
- |
ISO Stabil 5 |
|
Art. Nr. 50-71-000 |
|
- |
Fixomed |
|
|
Art. Nr. 50-91-000 |
- |
Hand 2003 |
|
Art. Nr. 54-03-000 |
|
- |
Oscillophon |
|
Art. N . 51-05-000 |
|
- |
Minimus |
|
|
Art. Nr. 51-02-000 |
- |
Maxi Stabil 3 |
|
Art. Nr. 59-32-000 |
|
1. |
Fehlersuche - Feststellung von fehlenden Teilen |
|
||
----------------------------------------------------------------- |
|
|||
|
-Manschette -Glas -Skala -Ventilteile -Ball |
- Schrauben -Dichtungen -sonstiges Zubehör |
2.Fehlersuche n. Fehlerart und Ursache Mögliche Fehlerarten- /ursachen:
-----------------------------------------------------
-Membrane/Grundplatte undicht
-Zeiger außerh. Toleranz am Nullpunkt
-Zeiger hängt / ruckartige Bewegungen Vermutlich Werkachse verbogen / Werkfeder im Segment verwickelt.
-Glas/ Skala herausgefallen, gebrochen verbogen , zerkratzt etc.
(Gerät gefallen)
-Ventilteil undicht Schraubventilversion
-Undichtigkeiten an der Manschette
-Undichtigkeiten an Schlauchverbind. Anschlußstücken/ Ventilkörper
-Auslaßschrauben schwergängig
-Drücker defekt / Rastung defekt
-Ventilteil undicht Drückerversion
-Kein Lufteinlaß/ Auslaß möglich
Notwendiger Reparaturaufwand
Durchzuführende Reparatur:
------------------------------------------------------------
Komplette Grundplatte m. Werk auswechseln + meßt.Kontrolle
Zeiger wechseln + meßtechnische Kontrolle - eventuell komplette Grundplatte mit Werk wechseln
Zeiger und komplette Grundplatte mit Werk wechseln + meßtechnische Kontrolle
Neues Glas + Skala einsetzen + meßtechnische Kontrolle durchführen
Ventil auswechseln + Dichtheitsprüfung
Gummibeutel wechseln + Dichtheitsprüfung
Komplettaustausch der defekten Teile + Dichtheitsprüfung vornehmen
Auslaßschraube wecheln + Dichtheitsprüfung
Drücker wechseln + Funktionsp rüfung der Rastung
Komplette Ventilbaugruppe auswechseln +
Dichtheitsprüfung
Lufteinlaß/ Ausgänge reinigen, Event. Auslaßschraube Ventil wechseln + Funktionsprüfung
2. Fehlersuche n. Fehlerart und Ursache
Mögliche Fehlerarten- /ursachen:
------------------------------------------------------
-Ball nicht pumpbar / defekt Vermutlich Rückschlagventil defekt / oder Materialriß (auch Handventile)
-Ventilkörper / Schraub/ Drückerversion defekt
-Fehlende Dichtungen
-Kalibrierzeitraum abgelaufen / überschritten
Seite / Page 24
Notwendiger Reparaturaufwand
Durchzuführende Reparatur:
--------------------------------------------
Neuer Ball montieren + Funktionsprüfung
Komplettausbau und Wiedereinbau neuer Körper + meßtechnischer Kontrolle
Dichtungen ersetzen + Dichtheitsprüfung
Meßtechnische Kontr olle durchführen + Prüfmarke anbringen , Dokumentation erstellen (siehe Anweisung meßtechnische Kontrollen
Beschreibung zur Durchführung der meßtechnischen Kontrolle - siehe Blatt meßtechnische
Kontrollen.
Beschreibung der Arbeitsfolge bei Demontage und Wiedermontage siehe spezielle
Einzelbeschreibung
Seite / Page 25
Reparaturanweisung für Speidel + Keller Blutdruckmeßgeräte
Quecksilbergeräte: |
Betroffene Geräte / Produkte: |
|
- |
Miniatur 300 B |
Art. Nr. 65-11-000 |
- |
Miniatur 300 S |
Art. Nr. 65-13-000 |
- |
Mercuro 300 |
Art. Nr. 74-01-000 |
1. |
Fehlersuche - Feststellung von fehlenden Teilen |
|
------------------------------------------------------------------- |
||
|
- Manschette, - Glasrohr, - Filterkappen, |
- Dichtungen, - Deckel, - Ballventil, - sonstiges Zubehör |
2. |
Fehlersuche n. Fehlerart und Ursache: |
Notwendiger Reparaturaufwand: |
|
Mögliche Fehlerarten- / Ursachen: |
Durchzuführende Reparatur: |
----------------------------------------------------- |
|
---------------------------------------------- |
- |
Glasrohr gebrochen / verschmutzt |
Glasrohr wechseln bzw. reinigen + meßtechn. Kontrolle |
- |
Quecksilberbehälter defekt / Riß |
Neuer Quecksilberbehälter montieren + Quecksilber |
|
|
einfüllen + meßtechnischer Kontrolle durchführen. |
- |
Rasthebel läßt sich nicht umlegen |
Neuer Rasthebel und Behälter montieren + Quecksilber |
|
kein Quecksilbereinfließen möglich |
einfüllen + meßtechnische Ko ntrolle durchführen |
- |
Quecksilberausfluß |
Untersuchung der undichten Stelle , Austausch des d efekten |
|
(Vorsicht Berührung mit Quecksilber |
Teiles + Quecksiber nachfüllen + meßtechn. Kontrolle |
|
ist gesundheitsschädlich) |
durchf. |
- |
Schlauchanschluß defekt/ undicht |
Neue Dichtung und Schlauchanschlußmontage + |
|
|
Quecksilber nachfüllen + meßtechn. Kontrolle durchführen. |
- |
Filter oder Filterkappe defekt / undicht |
Neuer Filter und Filterkappe einsetzen + Quecksilber |
|
losgelöst etc. |
nachfüllen + meßtechn. Kontrolle durchführen |
- |
Gehäuseteile / Skala etc. defekt |
Teile ersetzen + Funktionsprüfung durchführen. |
- |
Kalibrierungszeitraum überschritten |
Meßtechnische Kontr olle durchführen + neuer Prüfkleber |
|
|
mit Jahresangabe anbringen. |
Beschreibung zur Durchführung der meßtechnischen Kontrolle - siehe Blatt meßtechnische
Kontrollen.
Beschreibung der Arbeitsfolge bei Demontage und Wiedermontage siehe spezielle
Einzelbeschreibung.
Weitere Details zur Demontage und Wiedermontage sind aus den Bildern im Anhang ersichtlich.
Hier kann die Zusammenbauansicht entnommen werden siehe Ersatzteilkatalog Seite 28 - 30.
Seite / Page 26
Reparaturanweisung für Speidel + Keller Blutdruckmeßgerät |
Rev./ Index: 001B |
|||
Gerätetyp: |
Primus 50 Metall Art.Nr. 50-01-000 / |
Manuell 50 Metall Art.Nr. 50-16-000 |
||
|
Primus 68 Metall Art.Nr. 50-04-000 / |
Manuell 68 Metall Art.Nr. 50-17-000 |
||
Demontage: |
Beschreibung einer kompletten Demontage und Wiedermontage des Gerätes |
|||
1.1 |
Glashaltering 03/43-901/ 03-44-901 abschrauben. |
|
||
---------------- |
1.2 |
Glas 03-44-902 oder Glasreste mit Schraubendreher lösen (Vorsicht verklebt) |
1.3Zeiger 18-50-923 / 18-87-906 mit Abhebevorrichtung bzw. Pinzette lösen.
1.4Skala 20-64-151 abnehmen (Vorsicht verklebt)
1.5Komplette Grundplatte mit Werk 01-21-000 mit Grundplattenschlüsse ausschrauben.
1.6Becherschrauben 50-01-905 mit Schraubendreher ausschrauben. Dichtung 50-01-901 und Becher 50-01-904 / 50-04-901 bei 68mm entnehmen.
1.7Sprengring 19-40-905 mit Schraubendreher lösen und Auslaßschraube 19-40-901 ausdrehen. Einzelteile Feder 19-40-904 und Unterlegscheiben 19-40-903 entnehmen.
1.8Ball 17-31-902 und Klemmring 17-31-911vom Körper abziehen und Ventilsitzschraube 15-41-000 mit Spezialschlüssel ausschrauben. Für Geräte ab Oktober 1999 entfällt 1.8
Sprengring 15-40-903 mit Pinzette lösen, Dichtung 15-40-902 und Ventilschlauch 15-40-907 mit Spezialwerkzeug auslösen.
1.9Löffelschrauben 50-01-902 mit Schraubendreher lösen
Montage: |
2.1 |
Löffel 50-01-902 mit Schraubendreher anschrauben. |
------------- |
2.2 |
Neuer Becher mit Schraubendreher auf Ventilkörper 50-01-900 |
|
|
aufschrauben. |
|
2.3 |
Unterlegscheiben 19-40-903 und Feder in Ventilkörper einlegen |
|
|
Auslauslaßschraube einschrauben und Sprengring eindrücken. |
|
2.4 |
Für Geräte ab Oktober 1999 Neues Ventil 15-40-907, Dichtung 15-40-902 |
|
|
und Sprengring 15-40-903 mit Spezialwerkzeug montieren. |
|
|
Ballhülse 17-31-911 und Ball 17-31-902 montieren. |
|
2.5 |
Dichtung 50-01-901 einlegen und komplette Grundplatte |
|
|
mit Werk mit spez. Grundplattenschlüssel einschrauben |
|
2.6 |
Gerät in Justiervorrichtung legen u. spezielle S+K Prüfskala aufsetzen |
|
2.7 |
Zeiger 18-50-923 / 18-87-906 aufsetzen und mit 0 -Strich abgleichen |
|
2.8 |
Druckstufe 300 mmHg anfahren und e instellen. |
|
|
Druckstufen 50, 100, 150, 200, 250, 300 , 350 mmHg auf und abwärts |
|
|
anfahren und Toleranz prüfen +/- 3mmHg / gegebenenfalls mit speziellem |
|
|
Justierstift nachjustieren. |
|
2.9 |
Zeiger auf 0- Stellung setzen und mit speziellem Aufdrücker aufdrücken |
|
2.10 |
Neues Glas 03-44-904 einsetzen und einkleben (Kleber Sicomet 9010) |
|
2.11 |
Glashaltering 03-43-901/ 03-44-901 aufschrauben. |
Endprüfung / |
- |
Endprüfung je nach Reparaturaufwand durchführen ( siehe |
Funktion / |
|
Reparaturanweisung Schraub ventilversion S. 1) |
Dichtprüfung: |
- |
Beim Eingriff ins Gerät oder ins Meßwerk ist auf alle Fälle |
----------------- |
|
eine komplette Endpr üfung mit meßtechnischer Kontrolle |
|
|
durchzuführen (siehe Durchführung meßtechn. Kontrolle |
|
|
nach MPG ). |
|
- |
Bei Unsicherheiten ob Gerät gefallen ist wird die Durchführung |
Bemerkungen: |
|
einer meßtechnischen Kontrolle empfohlen. |
|
Die jeweils notwendige Reparaturaufwand ist festzustellen und |
|
------------------- |
|
durchzuführen. Die notwendig werdende Arbeitsfolge ist aus |
|
|
dieser Gesamtbeschreibung zu entnehmen |