Welch Allyn Blood Pressure Monitor User manual

4 (1)
Welch Allyn Blood Pressure Monitor User manual

Aneroid Menu

Präzision in Ihren Händen. Precision at your fingertips.

94-01-130 Stand: Nov. 2001

Sercice Manual / Service Anleitung

+

Ersatzteilkatalog / Spare Part Catalogue

2001 / 2002

Speidel + Keller GmbH & Co. KG Zollerstrasse 2-4

D - 72417 Jungingen Tel.: (++49) 7477-9271-0

Fax.: (++49) 7477-9271-90 e-mail: info@speidel-keller.de homepage: www.speidel-keller.de

Seite / Page 2

SERVICE MANUAL

WELCH ALLYN / SPEIDEL + KELLER

Preface

Dear S+K Servicepartner,

We are very happy to welcome you as a Service partner of Speidel + Keller.

The company Speidel + Keller is continually trying to offer optimal conditions and terms, so that, a smooth servicing of our units is guaranteed.

The Speidel + Keller blood pressure units distinguish themselves through very high quality. This is a warranty that S+K product possess a very long lifetime.

Even though the best units naturaly need regulary inspections so that a orderly run can be accomplished. Further it is a neccessarity in reference to the requirements of the MPG

and the EC-medical product guidelings, service the medical products on a regular base as well as carring out measurement inspections.

Because of the above mentioned reasons we have issued these repair instruction guides for all our units.

These instructions shall escort you to carry out an orderly repair.

Mangement Service

Seite / Page 3

Table of contents

/ Inhaltsverzeichnis:

English

Deutschsprachig

 

 

Page:

Seite:

Preface / Vorwort

 

2

22

Table of contents / Inhaltsverzeichnis

3

Entfällt

Failure search mechanical blood pressure units

4

23

Failure search mechanical blood pressure units

5

24

Failure search mercury blood pressure units

6

25

Instruction of dismantling and re-assembling

 

 

Reparaturanweisung Demontage und Wiedermontage

 

 

Primus 50 Metal

Art. No. 50-01-000

7

26

Primus 68 Metal

Art. No. 50-04-000

7

26

Manuell 50 Metal

Art. No. 50-16-000

7

26

Manuell 68 Metal

Art. No. 50-17-000

7

26

Primus 50 KC

Art.No. 50-81-000

8

27

Primus 68 KC

Art. No. 50-84-000

8

27

Manuell 50 KC

Art. No. 50-96-000

8

27

Manuell 68 KC

Art.No. 50-97-000

8

27

Primus Stabil 3

Art.No. 50-25-000

9

28

Primus Stabil yellow

Art.No. 50-49-000

9

28

Primus Stabil red

Art.No. 50-59-000

9

28

Primus Stabil blue

Art.No. 50-50-000

9

28

Primus Stabil green

Art.No. 50-45-000

9

28

Manuell Stabil 3

Art.No. 50-63-000

9

28

ISO Stabil 5

Art.No. 50-71-000

10

29

Disytest

Art.No. 52-09-000

11

30

Maxi Stabil 3

Art.No. 59-32-000

12

31

Fixomed

Art. No. 50-91-000

13

32

Hand 2003

Art.No. 50-03-000

14

33

Oscillophon

Art.No. 51-05-000

15

34

Instruction for end evaluation /measurements –

 

 

Anweisung für die Endprüfung

16

35

Tool list for blood pressure units - Werkzeugliste

17

36

Failure search for Stehoscopes - Fehlersuche Stehoskope

 

 

Stethoscope

Art.No. 84-10-000

18

37

Planoscope 40 mm

Art.No. 83-11-080

18

37

Planoscope 30 mm

Art.No. 83-12-080

18

37

Duophon 40 mm

Art.No. 82-11-080

18

37

Duophon 30 mm

Art.No. 82-15-080

18

37

Professionell

Art.No. 86-01-080

18

37

Harvey Elite

Art.No. 87-01-080

18

37

Failure search for hand fans - Fehlersuche Handventile

 

 

Hand fan

Art.No. 17-01-000

19

38

Hand fan

Art.No.17-02-000

19

38

Hand fan

Art.No. 17-08-000

19

38

Hand fan

Art.No. 17-10-000

19

38

Failure search for special medical manometers -

 

 

Fehlersuche Spezial Manometer

 

 

Pressure inf. Cuff

Art.No. 51-08-000

20

39

Sanipress unit

Art.No. 50-09-000

20

39

Tournipress

Art.No. 53-05-000

21

40

Seite / Page 4

Repair instruction for Speidel + Keller blood pressure units

Failure search mechanical blood pressure units

--------------------------------------------------------------------------------------------------------------------------------------

Units / Products concerned:

 

------------------------------------

 

- Primus

50/68 1 and 2 tube KC and Metal

Art. No. 50-01-000/ 50-04-000/ 50-81-000/ 50-84-000

- Manuell 50/68 1 and 2 tube KC and Metal

Art. No. 50-16-000/ 50-17-000/ 50-96-000/ 50-97-000

- Primus Stabil 3 and Primus Persona

Art No. 50-25-000, 50-49-000 /50-59-000/ 50-50-000/

included the colour serie

50-45-000

- Manuell

Stabil 3

Art.No. 50-63-000

- Disytest

 

Art. No. 52-09-000

- ISO Stabil 5

Art. No. 50-71-000

- Fixomed

 

Art. No. 50-91-000

- Hand 2001 und 2003

Art. No. 54-03-000, 54-07-000

- Oscillophon

Art. No. 51-05-000

- Minimus

 

Art. No. 51-02-000

- Maxi Stabil 3

Art. No. 59-32-000

1. Failure search - Locating missing parts

 

---------------------------------------------------------

 

- Cuff - Glass - Scale - Valve parts - Ball -

Screws - Washer - other Accessories

2.Search on cause and reason for failure Possible causes/- reasons:

------------------------------------------------------

-Leaking membran/ base plate measurement inspection

-Needle outside the tolerance on the zeropoint

-Hanging needle / jerky movements Propably bent movement axle movement spring tangled in the segment

-Glass / Scale fallen out, broken bent, scratched etc.

(dropped unit)

Neccessary repair

Upcoming repair:

----------------------------------------------------------------------

Replace complete base plate and movement +

Replace needle + measurement inspection + eventually replace complete base plate and movement

Replace complete base plate with movement + carry out measurement inspection

Insert glass and scale, carry out

measurement inspection

- Leaking valve part, tube valve version

Replace valve + leakage test

- Leak on the cuff

Replace rubber bag + leakage test

- Leak on tube connector/other connector pieces/

Replace all defect parts + carry out leakage test

part valve

 

- Hart turning release screws

Replace release screw + leakage test

- Defect trigger / bezel

Replace trigger + test bezel on function

- Leak on valve part, trigger version

Replace complete valve group + leakage test

- No air relase or air admittance possible

Clean air inlet and outlet, possible replacement of the

release valve + test on fuction

 

2.Search on cause and reason for failure Possible causes/- reasons:

---------------------------------------------------------

-Not possible to inflate the ball / defect Possible damaged check valve /

or rip in material (also hand valve)

-Defect valve body / screw / trigger version body + measurement inspection

-Missing washer

-Expired calibration time /

Seite / Page 5

Neccessary repair

Upcoming repair:

---------------------------------------------------------------------

Assemble new ball + test on function

Take out the complete body and re-assemble it with new

Replace washer + leakage test

Carry out measurement inspection + attach test label Prepare documentation (see instruction on measurement inspection

Instructions on measurement inspection on all blood pressure units - please see attachment Measurement inspections.

Instructions on work procedures for dismantling and re-assembling (see attached description)

Seite / Page 6

Repair instruction for Speidel + Keller blood pressure units

Mercury units

Units

Products concerned:

-

Miniatur 300 B

Art. No. 65-11-000

-

Miniatur 300 S

Art. No. 65-13-000

-

Mercuro 300

Art. No. 74-01-000

1.

Failure search - Locating missing parts

 

 

- Cuffs - Glass tube - Filter caps - Washers -

Cover - Ball valve - other accessories

2.

Search on cause and reason for failure

Neccessary repair

 

Possible causes/- reasons:

Upcoming repair:

-------------------------------------------------

--------------------------------------------

-

Broken / dirty glass tube

Replace or clean glass tube + measurement inspection

-

Defect mercury container / ripped

Assemble new mercury container + insert mercury + carry out

 

 

measurement inspection.

-

Bezel does not shift, no mercury inlet

Assemble bezel and container + pur in new mercury

 

possible

+ measurement inspection

-

Mercury outflow

Evaluate the leak place, replace defect parts + fill up

 

mercury + carry out measurement inspection

(Caution: Contact with mercury is

 

 

 

 

hazardous to your health)

-

Defect /leak tube connector

Insert new washer and assemble tube connector + refill

 

mercury + carry out measurement inspection

 

-

Defect filter or filter cap/ leaking

Insert new filter and filter cap + refill mercury + carry out

 

loosen, etc.

+ measurement inspection

-

Defect housing parts / scale

Replace parts + carry out function test.

-

Expired calibration time

Carry out measurement inspection + attach new test label with

 

 

information on the year.

If you need more information about the assembling combinations of the single parts please use the

exploding pictures from the following Spare Part Catalogue Page 28 –30.

Instructions on measurement inspection on all blood pressure units - please see attachment Measurement inspections.

Instructions on work procedures for dismantling and re-assembling ( see attached description

Seite / Page 7

Repair instruction for Speidel + Keller blood pressure units

Rev./ Index: 001B

Unit type:

Primus 50 metal

Art.No. 50-01-000

/ Manuell 50 metal

Art.No. 50-16-000

-------------

Primus 68 metal

Art.Nr. 50-04-000 /

Manuell 68 metal

Art.No. 50-17-000

Dismantle:

Instruction of complete dismantling and re-assembling of the unit

1.1

unscrew glass holder 03/43-901/ 03-44-901.

 

----------------

1.2

loosen glass 03-44-902 or glass remains with screwdriver (caution: adhered)

1.3loosen pointer 18-50-923 / 18-87-906 with slip device resp. tweezers

1.4remove scale 20-64-151 (caution: adhered)

1.5unscrew the complete base plate with movement 01-21-000 on hand of the base plate tool.

1.6unscrew cup screws 50-01-905 with screwdriver.

remove the seal 50-01-901 and cup 50-01-904 / 50-04-901 at 68mm

1.7loosen the safety pin 19-40-905 with screwdriver and remove the release screw 19-40-901. Remove the single parts spring 19-40-904 and washer 19-40-903.

1.8remove the ball 17-31-902 and span ring 17-31-911 from the body and loosen valve screw 15-41-000 with special tool.

Point 1.8 not ablicaple for units manufactured since October 1999. loosen span ring15-40-903 with a tweezers, remove seal 15-40-902 and valve tube 15-40-907 with special tool

1.9loosen spoon screws 50-01-902 with screwdriver

Re-assemble:

2.1

screw on the scoop 50-01-902 with screwdriver.

-----------------

2.2

screw on new cup with screwdriver onto the valve body 50-01-900

 

2.3

insert washer 19-40-903 and spring into valve body.

 

 

screw on the release screw and push the clip ring back in.

 

2.4

For units from October 99 assemble new valve 15-40-907, seal 15-40-902 and new circlip

 

15-40-903 with special tool.

 

2.5

Insert washer 50-01-901 and complete base plate with movement with spec. base plate

 

tool.

 

 

2.6

Put unit on calibration device and put on the special S+K test scale

 

2.7

Put on needle 18-50-923 / 18-87-906 and adjust with the 0-line

 

2.8

Go to pressure level 300mmHg and safe. Drive pressure level 50, 100, 150, 200, 250,

 

300,

350 mmHg upand downwards, test tolerance +/- 3mmHg / and if neccessary re-

 

calibrate with a special calibration pen.

 

2.9

Set needle on the 0-position and impress with special trigger.

 

2.10

Insert new glass 03-44-904 and glue it on. (Glue Sicomet 9010)

 

2.11

screw on the glass ring 03-43-901/ 03-44-901.

End-evaluation /

 

Carry out end-evaluation according to need of repair (see repair instruction

Function /

 

screw valve version,)

Leakage test:

 

If operating on the unit or the mechanism a complete end-evaluation has to be carried

 

 

out, including measurement inspection (see realization of mesurement technical

 

 

inspections after MPG.)

 

-

If unsure, if the unit was dropped, we suggest to carry out a measurement technical

 

 

inspection.

Remarks: The neccessary repair expenditure has to be established and carried out. The needed

upcoming procedure can be taken out of the complete insctruction.

If you need more information about the assembling combinations of the single parts please use the

exploding pictures from the following Spare Part Catalogue Page 3, 5, 7, 9

Seite / Page 8

Repair instruction for Speidel + Keller Blood pressure unit

Rev./ Index: 001B

unit type:

- Primus 50

KC Art.No. 50-81-000 / Manuell 50 KC Art.No. 50-96-000

-------------

- Primus 68

KC Art.No. 50-84-000 / Manuell 68 KC Art.No. 50-97-000

 

Instruction of dismantling and re-assembling the unit

 

Dismantle:

1.1

clip out glass with a screwdriver on the bezel

 

---------------

1.2

loosen glass 03-44-902 or glass remains with screwdriver (Caution: adhered)

1.3loosen needle 18-50-923/ 18-87-906 with slip device resp. tweezers

1.4remove scale 20-64-151 (caution: adhered)

1.5unscrew the complete base plate with movement 01-21-000 on hand of the base plate tool.

1.6unscrew cup screws 50-01-905 with screwdriver.

remove the seal 50-01-901 and cup 50-01-904 / 50-04-901 at 68mm

1.7loosen the safety pin 19-40-905 with screwdriver and remove the release screw 19-40-901. Remove the single parts spring 19-40-904 and washer 19-40-903.

1.8remove the ball 17-31-902 and span ring 17-31-911 from the body and loosen valve screw 15-41-000 with special tool

Point 1.8 not ablicaple for units manufactured since October 1999. loosen span ring15-40-903 with a tweezers, remove seal 15-40-902 and valve tube 15-40-907 with special tool

1.9loosen spoon screws 50-01-902 with screwdriver

Re-assemble:

2.1

screw on the scoop 50-01-902 with screwdriver.

-----------------

2.2

screw on new cup with screwdriver onto the valve body 50-01-900

 

2.3

insert washer 19-40-903 and spring into valve body.

 

 

screw on the release screw and push the clip ring back in.

 

2.4

For units from October 99 assemble new valve 15-40-907, seal 15-40-902 and new circlip

 

15-40-903 with special tool.

 

2.5

Insert washer 50-01-901 and complete base plate with movement with spec. base plate

 

tool.

 

 

2.6

Put unit on calibration device and put on the special S+K test scale

 

2.7

Put on needle 18-50-923 and adjust with the 0-line

 

2.8

Go to pressure level 300mmHg and safe. Drive pressure level 50, 100, 150, 200, 250,

 

300,

350 mmHg upand downwards, test tolerance +/- 3mmHg / and if neccessary re-

 

calibrate with a special calibration pen.

 

2.9

Set needle on the 0-position and impress with special trigger.

 

2.10

Insert new glass 03-44-904 and glue it on. (Glue Sicomet 9010)

 

2.11

clip the glass back in to the bezels.

End-evaluation /

 

Carry out end-evaluation according to need of repair (see repair instruction

Function /

 

screw valve version, page 1)

Leakage test:

 

If operating on the unit or the mechanism a complete end-evaluation has to be carried

 

 

out, including measurement inspection (see realization of mesurement technical

 

 

inspections after MPG.)

 

-

If unsure, if the unit was dopped, we suggest to carry out a measurement technical

 

 

inspection.

Remarks: The neccessary repair expenditure has to be established and carried out. The needed

upcoming procedure can be taken out of the complete instructions.

If you need more information about the assembling combinations for single parts please use the exploding

pictures from the following Spare Part Catalogue Page 4, 6, 8, 10

Seite / Page 9

Repair instruction for Speidel + Keller Blood pressure units

Rev./ Index: 001B

Model:

- Primus Stabil 3 Art.No. 50-25-000, 50-49-000, 50-59-000, 50-50-000, 50-45-000

----------

- Manuell Stabil 3 Art.No. 50-63-000

 

 

Instruction of dismantling and re-assembling the unit

 

Dismantle:

1.1

turn out the bezel 50-25-908 clock wise.

 

---------------

1.2

clip out the glass 03-44-904 with a screwdriver on the bezel (Caution: adhered)

1.3loosen needle 18-50-921 with slip device resp. tweezers

1.4remove scale 20-64-161 (caution: adhered)

1.5unscrew the complete base plate with movement 01-21-000 on hand of the base plate tool.

1.6unscrew cup screws 50-01-905 with screwdriver. remove the seal 50-01-901 and cup 50-01-907:

1.7loosen the safety pin 19-40-905 with screwdriver and remove the release screw 19-40-901. Remove the single parts spring 19-40-904 and washer 19-40-903.

1.8remove the ball 17-31-902 and span ring 17-31-911 from the body and loosen valve screw 15-41-000 with special tool

Point 1.8 not ablicaple for units manufactured since October 1999. loosen span ring15-40-903 with a tweezers, remove seal 15-40-902 and valve tube 15-40-907 with special tool

1.9loosen spoon screws 50-01-902 with screwdriver .

Re-assemble:

2.1

screw on the scoop 50-01-902 with screwdriver.

------------

2.2

screw on new cup with screwdriver onto the valve body 50-01-900

 

2.3

insert washer 19-40-903 and spring into valve body.

 

 

screw on the release screw and push the clip ring back in.

 

2.4

For units from October 99 assemble new valve 15-40-907, seal 15-40-902 and new circlip

 

15-40-903 with special tool.

 

2.5

Insert washer 50-01-901 and complete base plate with movement with spec. base plate

 

tool.

 

 

2.6

Put unit on calibration device and put on the special S+K test scale

 

2.7

Put on needle 18-50-921 and adjust with the 0-line

 

2.8

Go to pressure level 300mmHg and safe. Drive pressure level 50, 100, 150, 200, 250,

 

300,

350 mmHg upand downwards, test tolerance +/- 3mmHg / and if neccessary re-

 

calibrate with a special calibration pen.

 

2.9

Set needle on the 0-position and impress with special trigger.

 

2.10

Insert new glass 03-44-904 and glue it on. (Glue Sicomet 9010)

 

2.11

bring on the bezel 50-25-908 (double noses) and turn it glockwise until stop.

End-evaluation /

 

Carry out end-evaluation according to need of repair (see repair instruction

Function /

 

screw valve version, page 1)

Leakage test:

 

If operating on the unit or the mechanism a complete end-evaluation has to be carried

 

 

out, including measurement inspection (see realization of mesurement technical

 

 

inspections after MPG.)

 

-

If unsure, if the unit was dopped, we suggest to carry out a measurement technical

 

 

inspection.

Remarks:

 

The neccessary repair expenditure has to be established and carried out. The needed

upcoming procedure can be taken out of the complete instructions.

If you need more information about the assembling combinations for single parts please use the exploding pictures from the following Spare Part Catalogue Page 13 - 18.

Seite / Page 10

Repair instruction for Speidel + Keller Blood pressure units

Rev./ Index: 001B

Model:

- ISO Stabil 5 Art. No. 50-71-000

 

-------------

- Stabil 3 (predecessor. model Stabil 5)

 

 

Instruction of dismantling and re-assembling the unit

 

Dismantle:

1.1

turn out the bezel 50-25-908 clock wise.

 

---------------

1.2

clip out the glass 03-44-904 with a screwdriver on the bezel (Caution: adhered)

1.3loosen needle 18-50-921 with slip device resp. tweezers

1.4remove scale 20-64-161 (caution: adhered)

1.5unscrew the complete base plate with movement 01-21-000 on hand of the base plate tool.

1.6unscrew cup screws 50-01-905 with screwdriver and then remove the cup 50-25-905

1.7loosen the closing ring o n the ball 17-31-00 0 and then take it off the valve

1.8release trigger 50-71-812 from valve body and take out pin 50-61-813

1.9unscrew check valve 15-40-000 with special tool. For units manufactured since October 1999.

loosen span ring15-40-903 with a tweezers, remove seal 15-40-902 and valve tube 15-40-907 with special tool

Re-assemble:

2.1

Insert pin 50-61-813 though drilling o n the valve body.

------------

 

build in feather 54-31-904 on valve body and then insert trigger into valve body

 

 

push trigger pin with special pin forward and release trigger.

 

 

Carry out functional test ( should run easy without creaking noise up to the stop)

 

2.2

Screw in new valve screw 15-41-000 and screw the ball 17-31-000 back on.

 

 

Turn on check valve as safety device.

 

2.3

Put the cup 50-25-905 onto the valve body and screw it in.

 

2.4

Insert washer 50-01-901, then complete ground plate with movement and screw

 

 

it in with the special base plate tool.

 

 

For units manufactured since October 1999.

 

 

loosen span ring15-40-903 with a tweezers, remove seal 15-40-902 and

 

 

valve tube 15-40-907 with special tool.

 

 

Assemble ball tube 17-31-911 and ball 17-31-902

 

2.5

Put unit into the calibration device and put on the special S+K test scale.

 

2.6

Bring on trigger 18-50-921 and adjust it to the zero-line.

 

2.7

Start pressure 300 mmHg and adjust.

 

 

Pressure levels 50, 100, 150, 200, 250, 300 , 350 mmHg drive upand downwards

 

 

and test tolerance +/- 3mmHg / if necessary re-adjust with special calibration pin.

 

2.8

Set new trigger to zero-position and impress with a special printer.

 

2.9

Eventually assemble (engage) cub with a new spacer ring 50-25-912.

 

2.10

clip and clue in new glass 03-44-904 (Sicomet 9010).

 

2.11

Put on bezel 50-25-908 ( 2 nose whiskers left wing) and turn it clockwise until stop.

End-evaluation / Carry out end-evaluation according to need of repair

Function /

Leakage test: If operating on the unit or the mechanism a complete end-evaluation has to be carried out, including measurement inspection (see realization of mesurement technical inspections after MPG.)

-If unsure, if the unit was dopped, we suggest to carry out a measurement technical inspection.

Remarks: The neccessary repair expenditure has to be established and carried out. The needed

upcoming procedure can be taken out of the complete instructions.

Seite / Page 11

Repair instruction for Speidel + Keller Blood pressure units

Rev./ Index: 001B

Model:

- Disytest / Disytest HomeArt.No. 52-09-000

 

-------------

 

 

 

 

Instruction of dismantling and re-assembling the unit

 

Dismantle:

1.1

Enter with screwdriver and clip out glass 52-03-903.

 

---------------

1.2

Loosen needle 18-50-924 / 18-50-923 (Persona) with the slip device resp. tweezers

1.3Clip out the scale 20-51-106 / 20-51-103 (Persona) as you have down with the glass.

1.4Unscrew the complete base plate with movement 01-21-000 on hand of the base plate tool, take of the washer 50-01-901.

1.5Remove ball 17-52-000 and span ring 17-01-902.

1.6Turn off the check valve 15-40-000 with special tool. For units manufactured since October 1999.

loosen span ring15-40-903 with tweezers, remove seal 15-40-902 and valve tube 15-40-907 with special tool.

1.7Release trigger 52-03-902 on the valve body, with special tool (inside, six-cant wrench) turn off trigger release valve 12-12-900, take out spring 19-12-904.

Re-assemble:

2.1

Insert spring 19-12-904 in the valve body , insert trigger release valve 19-12-900

-----------------

 

and tighten it with the special six-cant wrench, press in trigger 52-03-902

 

 

to the grid - Caution: test trigger function - should have easy play and easy handling

 

 

between the trigger release valves.

 

2.2

Screw in new check valve 15-41-000, push in span ring 17-01-902 and ball 17-52-000.

 

 

For units manufactured since October 1999.

 

 

Assemble new valve 15-40-907, seal 15-40-902 and circlip 15-40-903 with special tool.

 

 

Assemble ball tube 17-31-911 and Ball 17-31-902

 

2.3

Insert washer 50-01-901 and screw in the complete base plate including movement with

 

 

the special base plate tool.

 

2.4

Put the unit up onto the calibration device and insert the special S+K test scale „Disytest“.

 

2.5

Put on needle 18-50-924 and adjust to the zero-line.

 

2.6

Bring pressure to 300 mmHg and adjust Pressure levels 50, 100, 150, 200, 250, 300 ,

 

 

350 mmHg run up and downwards, test on tolerance +/- 3mmHg , if necessary adjust

 

 

with special calibration pin.

 

2.7

Set new pointer on zero-position and press on with special trigger.

 

2.8

Clip in new glass 52-03-903.

End-evaluation / Carry out end-evaluation according to need of repair

Function /

Leakage test: If operating on the unit or the mechanism a complete end-evaluation has to be carried out, including measurement inspection (see realization of mesurement technical inspections after MPG.)

-If unsure, if the unit was dropped, we suggest to carry out a measurement technical inspection.

Remarks: The neccessary repair expenditure has to be established and carried out. The needed

upcoming procedure can be taken out of the complete instructions.

If you need more information about the assembling combinations for single parts please use the exploding

picture from the following Spare Part Catalogue Page 11.

Seite / Page 12

Repair instruction for Speidel + Keller Blood pressure unit

Rev./ Index: 001B

Model:

- Maxi Stabil 3

Art.No. 59-30-000

 

------------

 

 

 

 

 

Instruction of dismantling and re-assembling the unit

 

Dismantle:

1.1

Turn bezel 59-30-901 anticlockwise and remove the housing

 

---------------

 

(Caution bezel may be adhered - dissolve).

 

 

1.1

Enter the groove with a scrwedriver and clip out the glass 59-30-902.

 

1.2

Loosen needle 18-87-908 with the slip device resp. tweezers

 

 

1.3

Clip out the scala 20-51-159 with a screwdriver (as done with the Glas).

 

1.4

Unscrew the complete base plate with movement 01-10-000 with

the base plate

 

 

tool, take of the washer 50-01-915.

 

 

1.5

Unscrew the metal tube connector 16-40-000 and knurled nut 59-30-922

 

1.6

CAUTION:

No further dismantling possible, because the air release connector is

 

 

clued. A dismantling of the co nnector could damage the housing. If this connector is

 

 

damaged the complete housing 59-30-750 has to be exchanged.

 

Re-assemble:

2.1

Screw in the metal tube connector 16-40-000 and knurled nut 59-30-922

-----------------

2.2

Insert washer 50-01-901 and screw the complete base plate inclusive movement back in

 

 

with the special base plate tool.

 

 

2.3

Put the unit on the calbration device and bring on the special S+K test scale „Maxi.

 

2.4

Put on the needle 18-87-908 and adjust on the 0-line.

 

 

2.5

Bring pressure to 300 mmHg and adjust Pressure levels 50, 100, 150, 200, 250, 300 ,

 

 

350 mmHg run up and downwards, test on tolerance +/- 3mmHg , if necessary adjust

 

 

with special calibration pin.

 

 

2.6

Impress new scala (prescribed groove).

 

 

2.7

Set the new pointer on zero-position and impress with special press.

 

2.8

Clip in new glass 59-30-902.

 

End-evaluation / Carry out end-evaluation according to need of repair

Function /

Leakage test: If operating on the unit or the mechanism a complete end-evaluation has to be out, including measurement inspection (see realization of mesurement technical inspections after MPG.)

-If unsure, if the unit was dopped, we suggest to carry out a measurement technical inspection.

Remarks: The neccessary repair expenditure has to be established and carried out. The needed upcoming procedure can be taken out of the complete instructions.

If you need more information about the assembling combinations for single parts please use the exploding

pictures from the following Spare Part Catalogue Page 19 and 20.

Seite / Page 13

Repair instruction for Speidel + Keller Blood pressure units

Rev./ Index: 001B

Model:

- Fixomed

Art. No. 50-91-000

 

-------------

 

 

 

 

 

Instruction of dismantling and re-assembling the unit

 

Dismantle:

1.1

Unscrew the safety ring 03-43-901.

 

---------------

1.2

Loosen glass 03-42-902 or glass remains with screwdriver (Caution: adhered)

1.3Loosen needle 18-87-905 with the slip device resp. tweezers.

1.4Remove the scala 20-57-122 (Caution: adhered)

1.5Unscrew the complete base plate and movement 01-21-000 with base plate tool.

1.6Unscrew the flat head screws 50-01-905 with screwdriver. Remove washer 50-01-901 and cup 50-04-901 (68mm).

1.7Pull off the ball 17-31-902 and span ring 17-31-901 from the body and unscrew the check valve 15-41-000 with special wrench

For units manufactured since October 1999.

Loosen span ring15-40-903 with tweezers, remove seal 15-40-902 and valve tube 15-40-907 with special tool.

1.8Loosen spoon 50-01-902 with screwdriver.

1.9A dismantling of the release valve should not be carried out, because the re-assembling is extremely difficult.

If the release valve is damanged, the complete valve body has to be replaced 50-85-000.

Re-assemble:

2.1

Screw on the spoon 50-01-902 with screwdriver.

------------

2.2

Screw on the new cups with screwdriver on the valve body 50-85-900.

 

2.3

Screw in new check valve 15-41-000, and assemble new span ring 17-31-901 and Ball

 

 

17-31-902.

 

 

For units manufactured since October 1999.

 

 

Assemble new valve 15-40-907, seal 15-40-902 and circlip 15-40-903 with special tool.

 

 

Assemble ball tube 17-31-911 and Ball 17-31-902

 

2.4

Insert washer 50-01-901 and screw on the complete base plate with special base plate tool

 

2.5

Put unit into the calbration device and bring on the special test scale "Fixomed"

 

2.6

Put on needle 18-87-905 and adjust it to the zero-line.

 

2.7

Bring pressure to 300 mmHg and adjust Pressure levels 50, 100, 150, 200, 250, 300 ,

 

 

350 mmHg run up and downwards, test on tolerance +/- 3mmHg , if necessary adjust

 

 

with special calibration pin.

 

2.8

Insert scala 20-57-122 in cup.

 

2.9

Set needle on zero-position and impress with special press.

 

2.10

Insert new glass 03-42-902.

 

2.11

Screw on new safety glass 03-43-901 and tighten it.

End-evaluation / Carry out end-evaluation according to need of repair

Function /

Leakage test: If operating on the unit or the mechanism a complete end-evaluation has to be carried out, including measurement inspection (see realization of mesurement technical inspections after MPG.)

-If unsure, if the unit was dopped, we suggest to carry out a measurement technical inspection.

Remarks: The neccessary repair expenditure has to be established and carried out. The needed

upcoming procedure can be taken out of the complete instructions.

Seite / Page 14

Repair instruction for Speidel + Keller Blood pressure units

Rev./ Index: 001B

Model:

- Hand 2003

Art. No. 50-03-000

 

-------------

 

 

 

 

 

Instruction of dismantling and re-assembling the unit

 

Dismantle:

1.1

Unscrew safety glass 03-43-901.

 

---------------

1.2

Loosen glass 03-42-902 or glass remains with screwdriver (Caution: adhered)

1.3Loosen needle 18-87-906 with the slip device resp. tweezers.

1.4Remove scala 20-57-940 (Caution: adhered)

1.5Unscrew complete base plate incl. movement 01-21-000 with special base plate tool and remove washer 50-01-901.

1.6Loosen and unscrew the span ring ot the ball 17-31-000. Unscrew check valve 15-40-000 with special tool.

1.7Loosen spoons 50-01-902 with screwdriver.

1.8A dismantling of the release valve should not be carried out, because the re-assembling is extremely difficult.

If the release valve is damanged, the complete valve body has to be replaced 50-85-000.

Re-assemble: Goes the same for the unit Hand Version 2001 (50 mm)

----------------- 2.1 Screw the new check valve 15-40-000 in to the complete valve part 54-07-700 put up the new washer 54-31-910 and screw the Ball 17-31-000 on to the housing.

2.2Insert washer 50-01-901 and screw on the complete base plate with special base plate tool

2.3Put unit into the calbration device and bring on the special test scale 20-57-940

2.4Put on needle 18-87-906 and adjust it to the zero-line.

2.7Bring pressure to 300 mmHg and adjust Pressure levels 50, 100, 150, 200, 250, 300 , 350 mmHg run up and downwards, test on tolerance +/- 3mmHg , if necessary adjust with special calibration pin.

2.8Insert scala 20-57-940 in cup.

2.9Set needle on zero-position and impress with special press.

2.10Insert new glass 03-42-902.

2.11Screw on new safety glass 03-43-901 and tighten it.

End-evaluation / Carry out end-evaluation according to need of repair

Function /

Leakage test: If operating on the unit or the mechanism a complete end-evaluation has to be carried out, including measurement inspection (see realization of mesurement technical inspections after MPG.)

-If unsure, if the unit was dopped, we suggest to carry out a measurement technical inspection.

Remarks: The neccessary repair expenditure has to be established and carried out. The needed

upcoming procedure can be taken out of the complete instructions.

If you need more information about the assembling combinations for single parts please use the exploding

picture from the following Spare Part Catalogue Page 23.

Seite / Page 15

Repair instruction for Speidel + Keller Blood pressure units

Rev./ Index: 001B

Model:

- Oscillophon Art. No. 51-05-000

 

-------------

 

 

 

 

Instruction of complete dismantling and re-assembling of the unit

Dismantle:

1.1

Dismantle unit from the cuff with screwdriver

 

--------------

1.2

Unscrew safety ring 03-44-901.

 

1.3Loosen glass 03-44-902 or glass remains with screwdriver (Caution: adhered)

1.4Loosen needle 18-50-923 with the slip device resp. tweezers.

1.5Remove scala 20-57-940 (Caution: adhered)

1.6Loosen three screws with screwdriver and complete base plate with movement 01-22-000 plus take out three washers 51-01-901.

Assemble:

------------- 2.1 Insert the three washer 51-01-901 and screw the three screws 01-21-901 back in with a

screwdriver and tighten the complete base plate with the movement

2.2Put unit into the calbration device and bring on the special test scale 20-64-151

2.3Put on needle 18-50-923 and adjust it to the zero-line.

2.4Bring pressure to 300 mmHg and adjust Pressure levels 50, 100, 150, 200, 250, 300 , 350 mmHg run up and downwards, test on tolerance +/- 3mmHg , if necessary adjust with special calibration pin.

2.5Insert scala 20-64-151 in cup. (note: inlay groove)

2.6Set needle on zero-position and impress with special press.

2.7Insert new glass 03-44-902.

2.8Screw on new safety glass 03-44-901 and tighten it.

End-evaluation / Carry out end-evaluation according to need of repair

Function /

Leakage test: If operating on the unit or the mechanism a complete end-evaluation has to be carried out, including measurement inspection (see realization of mesurement technic inspections after MPG.)

-If unsure, if the unit was dopped, we suggest to carry out a measurement technical inspection.

Remarks: The neccessary repair expenditure has to be established and carried out. The needed

upcoming procedure can be taken out of the complete instructions.

If you need more information about the assembling combinations for single parts please use the exploding

picture from the following Spare Part Catalogue Page 12.

Seite / Page 16

Instruction for end evaluation including measurement inspections

1. Visual test:

-Check complete unit (chrome damages, etc).

-Check if pointer position is at zero range

2. Measurement test

-Connected unit to the certified pressure gauge class 0,1 (max. deviation 0,8 mmHg).

-Accuracy statistical pressure / Hysterisis on pressure values 50,100,150,200,250,300, mmHg test up-and downwards Tolerance = < 3 mmHg on all test points / Hysterisis = < 4 mmHg on all test points. - start pressure decrease 300 mmHg /seal unit system and test leakage. Pressure decrease should stay

at = < 4 mmHg/min.

3. Function test

-Connect unit with complete cuff onto the test tube, girth appox. 35 cm and pump up to 300 mmHg /test cuff for leakage, then open air release valve / should turn easy / test pointer run and test speed outlet. During the speed air release the time for pressure decrease from 260 mmHg to 15 mmHg should not exceed 10 Sec.

Run of the release screws / test printer - trigger grid (ie. Disytest)

4.

Test Documentation

-

Documentation as by S+K requested test report, entry of unit type, S/N, established deviations, leakage, test

 

date, tester, user, remarks.

 

These documentation has to be made available for S+K.

5.

Following function

 

Fasten new test label with year when test was carried out.

6.

Important remarks:

-

Measurement technical controls with confirmation of a test label can only be carried out on CE - marked

 

blood pressure meters. Older units with PTB-Registration should be submitted to the notified body after

 

inspection for calibration as done in the past.

Seite / Page 17

Tools:

Description

Article/ Order No.

Complete tool set

 

94-01-000

Base plate wrench metal

for movement assembling (all types)

94-01-101

Test scale 50 mm metal

for unit calibration

94-01-103

Test scale 68 mm metal

for unit calibration

94-01-104

Test scale Disy

for unit calibration

94-01-105

Calibration stick

for unit calibration

94-01-106

Needle presser

for needle assembling

94-01-107

Wrench SW3,5

for release assembling

94-01-108

Wrench SW6,5

for valve assembling

94-01-109

Wrench SW6 (inside)

for valve part assembling

94-01-110

Square wrench for seal

Seal dismantling (old units)

94-01-111

Test scale 50 mm KC

for unit calibration

94-01-112

Test scale 68mm KC

for unit calibration

94-01-113

Tweezcers size AA

for various operations

94-01-115

Test scale for Stabil

for unit calibration

94-01-116

Test scale Maxi

for unit calibration

94-01-117

Calibration stand

for unit calibration

94-01-118

Neddle disconnector

for needle dismantling

94-01-119

Screw driver (fine) 1,5

assembling tool for various units

94-01-120

Screw driver(fine) 2,0

assembling tool for various units

94-01-121

Pinch pliers

for ball sleve assemble / pinching

94-01-123

Spare part list + repair instructions

for all blood pressure units

94-01-130

Ordering of single tools is possible. Please direct your orders to a S+K Sales department.

Other necessary tools:

----------------------------

Test gauge

Measurement sector at least 0-300 mmHg

+ Certificate

with evidence Accuracy= < 0,8 mmHg

Spanblock

for dismantling the seal

Adjustment plate

for adjusting defect parts

self purchase nec. / on request also from S+K

self purchase nec.

self purchase nec.

Seite / Page 18

Repair instruction for Speidel + Keller blood pressure units "Accessories for blood pressure units"

Failure search mechanical blood pressure units

Units / Products concerned:

 

- Membranstetoscope for self mesurement

Art. No. 84-10-000 (integrated in cuff)

- Planoskop 40 mm

Art. No. 83-11-080

- Planoskop 30 mm (child)

Art. No. 83-12-080

- Duophon

40 mm

Art. No. 82-11-080

- Duophon

30 mm (child)

Art. No. 82-15-080

- Harvey Elite

all models

- Professional

all models

1.

Failure search - Locating missing parts

 

-----------------------------------------------------

 

-

Missing membran - Safety ring - Ear tips -

Membran holder - Y-Tube

2.

Search on cause and reason for failure

Neccessary repair

-----------------------------------------------------

 

-----------------------------------------

-

Defect / broken membran

Built in new membran + carry out function test

 

(can occur on all models)

 

-

Bent / defect ear piece

Fit new ear piece + funtion test

-

Defect / plugged / missing

Assemble new ear tips + function test

 

(can occur on all models)

 

-

Defect / missing membran holder

Assemble new membran holder + function test

-

Y-tube porous / defect. Ripped material, etc.

Assemble new Y-tube + function test

 

(all models)

 

-

Tube connector turns / can not be

Insert tube connector + function test

 

assembled (only on membran stethoscope

 

 

for self measurement)

 

-Not possible to switch over the Duophon shaft bent.

Dismantle and re-assemble (only Duophon) eventually assemble new shaft + function test

Instructions for function test / Membranstethoscope integrated in cuff "Self measuring"

1.Apply cuff on upper arm

2.Connect tube connector on S+K blood pressure unit.

3.Insert ear tips into the auditory canal

4.Inflate cuff (approx. 30 mmHg over expected blood pressure)

5.Release air and carry out the complete measurement.

6.Korotoff sounds should be well heared from the start of the Systole till the Diastole.

Planoskop / Professional

1.Apply cuff on upper arm

2.Push the Planoskop und er the Cuff (Membran part skin contact)

3.Insert ear tips into the auditory canal

4.Inflate cuff (approx. 30 mmHg over expected blood pressure)

5.Release air and carry out the complete measurement.

6.Korotoff sounds should be well heared from the start of the Systole till the Diastole.

Duophon / Harvey Elite

1.Switch Duophon several times o ver - should be easy going and engage on half turn. Duophon shaft hole visible - Duophon out of order

Duophon shaft hole not visible - turned to membran - Duo phon on stand by.

Seite / Page 19

Repair instruction for Speidel + Keller blood pressure units

Failure search accessories hand fan

Units / Products concerned:

---------------------------------------

-

Hand fan

Art. No. 17-01-000

-

Hand fan

Art. No. 17-02-000

-

Hand fan

Art. No. 17-08-000

-

Hand fan

Art. No. 17-10-000

1.

Failure search - Locating missing parts

----------------------------------------------------------------

- Trigger - Ball - Ring / Release screw - Valve part

2. Search on cause and reason for failure

------------------------------------------------------

-Not possible to orderly inflate ( defect ball valve)

-Leakage in System

tube connector / release screw

-Worn out release screw

-Defect / missing / broken trigger

Instruction for function test:

All hand valves:

Neccessary repair

----------------------------------------

Inspect ball valve - replace if necessary + carry out function test

Assemble new valve check screw / release screw + carry out function test

Assemble new release screw + function test

Assemble new trigger + function test

-Connect hand fan with connector tube onto a cuff

-Cuff around test tube (Girth approx. 35 cm)

-Interclose test gauge 0-600 mmHg

-Pump in pressure air till approx. 500 mmHg

-Wait for air balance

-Test leakage =< 4 mmHg / min

-Operate release screw and trigger, then carry out function test (Should be easy to operate )

-Test speed air release (the air release with pressed valve should be 260 - 15 mmHg =< 10 seconds).

If you need more information about the assembling combinations for single parts please use the exploding

pictures from the following Spare Part Catalogue Page 24 - 27.

Seite / Page 20

Repair instructions for Speidel + Keller

Medical Special units

----------------------------

 

Units / Products concerned:

 

-

Pressure infusion cuff

Art. No. 51-08-000

-

Pressure unit Sanipress

Art. No. 50-09-000

1.Failure Search - Locating missing parts

-Cuff - Glass - Scale - Valve parts - Ball valve - Screws - Washers - other Accessories

2.Search on cause and reason for failure Possible causes/- reasons:

-Leaking membran/ base plate

-Needle outside the tolerance on the zeropoint

-Hanging needle / jerky movements Probably bent movement axle movement spring tangled in the segment

-Glass / Scale fallen out, broken bent, scratched etc.

(dropped unit)

Neccessary repair

Upcoming repair:

Replace complete base plate and movement + measurement inspection

Replace needle + measurement inspection + eventually replace complete base plate and movement

Replace complete base plate with movement + carry out measurement inspection

Insert glass and scale, carry out measurement inspection

-

Leaking valve part, tube valve version

Replace valve + leakage test

- Leak on the cuff

Replace rubber bag + leakage test

- Leak on tube connector/other connector pieces/

Replace all defect parts + carry out leakage test

 

part valve

 

- Hart turning release screws

Replace release screw + leakage test

- Defect trigger / bezel

Replace trigger + test bezel on function

- Leak on valve part, trigger version

Replace complete valve group + leakage test

- Defect air pump / not possible to inflate

Replace air pump + leakage test

- No air inlet or release possible

Clean inand outlets. Replace release screw/ valve if

 

 

necessary + function test

- No speed air release possible

Clean inand outlets. Replace release screw/ valve if

 

 

necessary + function test

- Ball + valve not pumping / d efect

Assemble new ball + function test

 

possible defect check valve /

 

 

or rip in material

 

-

Defect valve body

Completely dismantle and reassemble new body +

 

 

measurement inspection.

- Missing washer

Replace washer + leakage test

Seite / Page 21

Repairs instruction for Speidel + Keller - Medical Special units.

Units / Products concerned:

-Pressure unit Tournipress Art.No. 53-05-000

1.Failure Search - Locating missing parts

-Cuff - Glass - Scale - Valve parts - Ball valve - Screws - Washers - other Accessories

2.Search on cause and reason for failure Possible causes/- reasons:

-Pressure not adjustable / Defect pressure regulator / regulator head

-Aeration not possible, aeration -function valve hangs / defect

-Pressure does not show on display at end pressure 600

Pressure display does not react if the main operation gets cut off.

-Defect connector to supplier / defect air coupling

-Not possible to switch over from distal

to proximal and reverse / defect switch valve ventil defekt

-Timer does not function / no reaction

-Leak membran/base plate from Manometer

-Needle outside the toleranc on the zero point

-Needle hangs / jerky movements Probably bent movement axle/ movement spring tangled up in Segment

-Glass/ Scale fallen out, broken

bent, scratched, etc (dropped unit)

-Leak on cuff

-Leakage on tube connector connector pieces / valve bodies

-No air inlet / release possible

-Accessory ball + valve not possible to inflate / probably defect check valve or rip in material

Neccessary repair

Upcoming repair:

Replace pressure regulator with regulator head , re-adjust pressure, carry out function + leakage test

Assemble new aeration valve + carry out function + leakage test

Assemble new pressure dislay + carry out function + and leakage test

Built in new shutter coupling + carry out function + leakage test

Built in new switch valve + carry out function test

Replace battery and solder new electric connectors + carry out function test.

Replace complete base plate + movement carry out measurement inspection.

Replace needle + measurement inspection - eventually replace base plate including movement

Replace complete base plate and movement + measurement inspection

Insert new glass + scale + carry out measurement

Replace rubber bag + leakage test

Replace all defect parts + carry out leakage test

Clean air inlets and outlets. Possibly replace release screw / valve + carry out function test.

Assemble new ball + function test

- Missing washers

Replace washers + leakage test

Seite / Page 22

SERVICE ANLEITUNG

WELCH ALLYN / SPEIDEL +KELLER

Vorwort

Sehr geehrter S+K Servicepartner,

wir freuen uns Sie als Reparaturpartner von Speidel + Keller begrüßen zu dürfen.

Die Fa. Speidel + Keller ist stets bemüht Ihren Servicepartnern optimale Bedingungen zu schaffen, damit eine reibungsloser Service der Produkte gewährleistet ist.

Speidel +Keller Blutdruckmeßgeräte zeichnen sich durch eine qualitativ hochwertige Verarbeitung aus. Dies ist ein Garant dafür, daß S+K Produkte eine sehr lange Lebensdauer besitzen.

Natürlich bedarf auch das beste Gerät einer regelmäßígen Wartung damit stetig ordnungsgemäß und reibungslos damit gearbeitet werden kann.

Zudem ist es eine Notwendigkeit, im Hinblick auf die Bestimmungen des MPG

und der EGMedizinprodukterichtlinien, regelmäßige Wartungen und meßtechnische Kontrollen an Medizinprodukten durchzuführen.

Aus den obengenannten Gründen haben wir Reparaturanweisungen für alle unsere Geräte erstellt.

Diese Anweisungen sollen Ihnen bei der Durchführung einer ordnungsgemäßen Reparatur begleitend helfen.

Seite / Page 23

Reparaturanweisung für Speidel + Keller Blutdruckmeßgeräte

Fehlersuche Mechanische Blutdruckmeßgeräte

--------------------------------------------------------------------------------------------------------------------------------------

 

Betroffene Geräte / Produkte:

 

-----------------------------------------

 

-

Primus

50/68

1 u. 2 Schlauchv. KC und Metall

Art. Nr. 50-01-000/ 50-04-000/ 50-81-000/ 50-84-000

-

Manuell 50/68

1 u. 2 Schlauchv. KC und Metall

Art. Nr. 50-16-000/ 50-17-000/ 50-96-000/ 50-97-000

-

Primus Stabil 3 und Stabil 3 Home

Art Nr. 50-25-000/ 50-49-000/ 50-59-000/ 50-50-000

 

und Farbserie gelb,blau,rot,blau

50-45-000

-

Manuell

Stabil 3

Art.Nr. 50-63-000

-

Disytest /

Disytest Home

Art. Nr. 52-09-000

-

ISO Stabil 5

 

Art. Nr. 50-71-000

-

Fixomed

 

 

Art. Nr. 50-91-000

-

Hand 2003

 

Art. Nr. 54-03-000

-

Oscillophon

 

Art. N . 51-05-000

-

Minimus

 

 

Art. Nr. 51-02-000

-

Maxi Stabil 3

 

Art. Nr. 59-32-000

1.

Fehlersuche - Feststellung von fehlenden Teilen

 

-----------------------------------------------------------------

 

 

-Manschette -Glas -Skala -Ventilteile -Ball

- Schrauben -Dichtungen -sonstiges Zubehör

2.Fehlersuche n. Fehlerart und Ursache Mögliche Fehlerarten- /ursachen:

-----------------------------------------------------

-Membrane/Grundplatte undicht

-Zeiger außerh. Toleranz am Nullpunkt

-Zeiger hängt / ruckartige Bewegungen Vermutlich Werkachse verbogen / Werkfeder im Segment verwickelt.

-Glas/ Skala herausgefallen, gebrochen verbogen , zerkratzt etc.

(Gerät gefallen)

-Ventilteil undicht Schraubventilversion

-Undichtigkeiten an der Manschette

-Undichtigkeiten an Schlauchverbind. Anschlußstücken/ Ventilkörper

-Auslaßschrauben schwergängig

-Drücker defekt / Rastung defekt

-Ventilteil undicht Drückerversion

-Kein Lufteinlaß/ Auslaß möglich

Notwendiger Reparaturaufwand

Durchzuführende Reparatur:

------------------------------------------------------------

Komplette Grundplatte m. Werk auswechseln + meßt.Kontrolle

Zeiger wechseln + meßtechnische Kontrolle - eventuell komplette Grundplatte mit Werk wechseln

Zeiger und komplette Grundplatte mit Werk wechseln + meßtechnische Kontrolle

Neues Glas + Skala einsetzen + meßtechnische Kontrolle durchführen

Ventil auswechseln + Dichtheitsprüfung

Gummibeutel wechseln + Dichtheitsprüfung

Komplettaustausch der defekten Teile + Dichtheitsprüfung vornehmen

Auslaßschraube wecheln + Dichtheitsprüfung

Drücker wechseln + Funktionsp rüfung der Rastung

Komplette Ventilbaugruppe auswechseln +

Dichtheitsprüfung

Lufteinlaß/ Ausgänge reinigen, Event. Auslaßschraube Ventil wechseln + Funktionsprüfung

2. Fehlersuche n. Fehlerart und Ursache

Mögliche Fehlerarten- /ursachen:

------------------------------------------------------

-Ball nicht pumpbar / defekt Vermutlich Rückschlagventil defekt / oder Materialriß (auch Handventile)

-Ventilkörper / Schraub/ Drückerversion defekt

-Fehlende Dichtungen

-Kalibrierzeitraum abgelaufen / überschritten

Seite / Page 24

Notwendiger Reparaturaufwand

Durchzuführende Reparatur:

--------------------------------------------

Neuer Ball montieren + Funktionsprüfung

Komplettausbau und Wiedereinbau neuer Körper + meßtechnischer Kontrolle

Dichtungen ersetzen + Dichtheitsprüfung

Meßtechnische Kontr olle durchführen + Prüfmarke anbringen , Dokumentation erstellen (siehe Anweisung meßtechnische Kontrollen

Beschreibung zur Durchführung der meßtechnischen Kontrolle - siehe Blatt meßtechnische

Kontrollen.

Beschreibung der Arbeitsfolge bei Demontage und Wiedermontage siehe spezielle

Einzelbeschreibung

Seite / Page 25

Reparaturanweisung für Speidel + Keller Blutdruckmeßgeräte

Quecksilbergeräte:

Betroffene Geräte / Produkte:

-

Miniatur 300 B

Art. Nr. 65-11-000

-

Miniatur 300 S

Art. Nr. 65-13-000

-

Mercuro 300

Art. Nr. 74-01-000

1.

Fehlersuche - Feststellung von fehlenden Teilen

-------------------------------------------------------------------

 

- Manschette, - Glasrohr, - Filterkappen,

- Dichtungen, - Deckel, - Ballventil, - sonstiges Zubehör

2.

Fehlersuche n. Fehlerart und Ursache:

Notwendiger Reparaturaufwand:

 

Mögliche Fehlerarten- / Ursachen:

Durchzuführende Reparatur:

-----------------------------------------------------

 

----------------------------------------------

-

Glasrohr gebrochen / verschmutzt

Glasrohr wechseln bzw. reinigen + meßtechn. Kontrolle

-

Quecksilberbehälter defekt / Riß

Neuer Quecksilberbehälter montieren + Quecksilber

 

 

einfüllen + meßtechnischer Kontrolle durchführen.

-

Rasthebel läßt sich nicht umlegen

Neuer Rasthebel und Behälter montieren + Quecksilber

 

kein Quecksilbereinfließen möglich

einfüllen + meßtechnische Ko ntrolle durchführen

-

Quecksilberausfluß

Untersuchung der undichten Stelle , Austausch des d efekten

 

(Vorsicht Berührung mit Quecksilber

Teiles + Quecksiber nachfüllen + meßtechn. Kontrolle

 

ist gesundheitsschädlich)

durchf.

-

Schlauchanschluß defekt/ undicht

Neue Dichtung und Schlauchanschlußmontage +

 

 

Quecksilber nachfüllen + meßtechn. Kontrolle durchführen.

-

Filter oder Filterkappe defekt / undicht

Neuer Filter und Filterkappe einsetzen + Quecksilber

 

losgelöst etc.

nachfüllen + meßtechn. Kontrolle durchführen

-

Gehäuseteile / Skala etc. defekt

Teile ersetzen + Funktionsprüfung durchführen.

-

Kalibrierungszeitraum überschritten

Meßtechnische Kontr olle durchführen + neuer Prüfkleber

 

 

mit Jahresangabe anbringen.

Beschreibung zur Durchführung der meßtechnischen Kontrolle - siehe Blatt meßtechnische

Kontrollen.

Beschreibung der Arbeitsfolge bei Demontage und Wiedermontage siehe spezielle

Einzelbeschreibung.

Weitere Details zur Demontage und Wiedermontage sind aus den Bildern im Anhang ersichtlich.

Hier kann die Zusammenbauansicht entnommen werden siehe Ersatzteilkatalog Seite 28 - 30.

Seite / Page 26

Reparaturanweisung für Speidel + Keller Blutdruckmeßgerät

Rev./ Index: 001B

Gerätetyp:

Primus 50 Metall Art.Nr. 50-01-000 /

Manuell 50 Metall Art.Nr. 50-16-000

 

Primus 68 Metall Art.Nr. 50-04-000 /

Manuell 68 Metall Art.Nr. 50-17-000

Demontage:

Beschreibung einer kompletten Demontage und Wiedermontage des Gerätes

1.1

Glashaltering 03/43-901/ 03-44-901 abschrauben.

 

----------------

1.2

Glas 03-44-902 oder Glasreste mit Schraubendreher lösen (Vorsicht verklebt)

1.3Zeiger 18-50-923 / 18-87-906 mit Abhebevorrichtung bzw. Pinzette lösen.

1.4Skala 20-64-151 abnehmen (Vorsicht verklebt)

1.5Komplette Grundplatte mit Werk 01-21-000 mit Grundplattenschlüsse ausschrauben.

1.6Becherschrauben 50-01-905 mit Schraubendreher ausschrauben. Dichtung 50-01-901 und Becher 50-01-904 / 50-04-901 bei 68mm entnehmen.

1.7Sprengring 19-40-905 mit Schraubendreher lösen und Auslaßschraube 19-40-901 ausdrehen. Einzelteile Feder 19-40-904 und Unterlegscheiben 19-40-903 entnehmen.

1.8Ball 17-31-902 und Klemmring 17-31-911vom Körper abziehen und Ventilsitzschraube 15-41-000 mit Spezialschlüssel ausschrauben. Für Geräte ab Oktober 1999 entfällt 1.8

Sprengring 15-40-903 mit Pinzette lösen, Dichtung 15-40-902 und Ventilschlauch 15-40-907 mit Spezialwerkzeug auslösen.

1.9Löffelschrauben 50-01-902 mit Schraubendreher lösen

Montage:

2.1

Löffel 50-01-902 mit Schraubendreher anschrauben.

-------------

2.2

Neuer Becher mit Schraubendreher auf Ventilkörper 50-01-900

 

 

aufschrauben.

 

2.3

Unterlegscheiben 19-40-903 und Feder in Ventilkörper einlegen

 

 

Auslauslaßschraube einschrauben und Sprengring eindrücken.

 

2.4

Für Geräte ab Oktober 1999 Neues Ventil 15-40-907, Dichtung 15-40-902

 

 

und Sprengring 15-40-903 mit Spezialwerkzeug montieren.

 

 

Ballhülse 17-31-911 und Ball 17-31-902 montieren.

 

2.5

Dichtung 50-01-901 einlegen und komplette Grundplatte

 

 

mit Werk mit spez. Grundplattenschlüssel einschrauben

 

2.6

Gerät in Justiervorrichtung legen u. spezielle S+K Prüfskala aufsetzen

 

2.7

Zeiger 18-50-923 / 18-87-906 aufsetzen und mit 0 -Strich abgleichen

 

2.8

Druckstufe 300 mmHg anfahren und e instellen.

 

 

Druckstufen 50, 100, 150, 200, 250, 300 , 350 mmHg auf und abwärts

 

 

anfahren und Toleranz prüfen +/- 3mmHg / gegebenenfalls mit speziellem

 

 

Justierstift nachjustieren.

 

2.9

Zeiger auf 0- Stellung setzen und mit speziellem Aufdrücker aufdrücken

 

2.10

Neues Glas 03-44-904 einsetzen und einkleben (Kleber Sicomet 9010)

 

2.11

Glashaltering 03-43-901/ 03-44-901 aufschrauben.

Endprüfung /

-

Endprüfung je nach Reparaturaufwand durchführen ( siehe

Funktion /

 

Reparaturanweisung Schraub ventilversion S. 1)

Dichtprüfung:

-

Beim Eingriff ins Gerät oder ins Meßwerk ist auf alle Fälle

-----------------

 

eine komplette Endpr üfung mit meßtechnischer Kontrolle

 

 

durchzuführen (siehe Durchführung meßtechn. Kontrolle

 

 

nach MPG ).

 

-

Bei Unsicherheiten ob Gerät gefallen ist wird die Durchführung

Bemerkungen:

 

einer meßtechnischen Kontrolle empfohlen.

 

Die jeweils notwendige Reparaturaufwand ist festzustellen und

-------------------

 

durchzuführen. Die notwendig werdende Arbeitsfolge ist aus

 

 

dieser Gesamtbeschreibung zu entnehmen

Loading...
+ 58 hidden pages