Siemens Mammomat Novation DR Installation Manual

4.2 (6)

Opdimar

SP

Installation

Installation and Start-Up

on MAMMOMAT NovationDR

 

 

 

 

© Siemens AG 2004

 

The reproduction, transmission or

 

use of this document or its contents

 

is not permitted without express

 

written authority. Offenders will be

 

liable for damages.

All rights,

 

including rights created by patent

 

grant or registration of a utility

 

model _or_ design,_are_ reserved.

 

English

 

Print No.: SPB7-230.814.01.01.02

Doc. Gen. Date:

10.04

Replaces: n.a.

(66 31 944)

 

0 - 2

 

 

Revision

 

 

 

 

Chapter

Page

Revision

 

 

 

 

 

All

All

01

 

 

 

 

 

Document revision level

The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.

Please contact your local Siemens office to order current revision levels.

Disclaimer

The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.

Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.

Opdima

SPB7-230.814.01

Page 2 of 4

Siemens AG

Installation

Rev. 01 10.04

CS PS 24

Medical Solutions

Contents

0 - 3

 

Page

1 _______Prerequisites __________________________________________________ 1 - 1

General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1 Training of customer support engineers . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2 Text emphasis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2 Documents required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3 Meters and appliances required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3 Tools required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3 Additional installation material required . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3 Start-up, DHHS and X-ray decree area of application. . . . . . . . . . . . . . . . . . 1 - 3

Notes on start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3 Checking and recording for the area of application of the X-ray decree

(§16 Germany). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4 Checking and data recording in the USA . . . . . . . . . . . . . . . . . . . . . . 1 - 4

2 _______Protective measures for CCD camera ______________________________ 2 - 1 3 _______Protective measures ____________________________________________ 3 - 1 4 _______Preparatory work _______________________________________________ 4 - 1

Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2 Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2 Biopsy controller, miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2 Optional equipment and update kits . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2

5 _______Installation ____________________________________________________ 5 - 1

General . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . 5 - 1

Sub-assemblies, overview. . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . 5 - 1

Removal of MAMMOMAT NovationDR stand covers

. . . . . . . . . . . . . . . . . . 5 - 2

. . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . .

. 5 - 2

Installation of biopsy controller holders . . . . . . .

. . . . . . . . . . . . . . . . .

. 5 - 3

Connection of cables .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

- 5

. . . . . . . . . . .

.

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6

Cables to CCD camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6

Installation of cable duct.

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

- 8

 

Modification of cable outlet cover . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 8

Installation of workstation . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . 5 - 9

 

Preparing the workstation table (option) . . . . . . . . . . . . . . . . . . . . . . . 5 - 9

 

Connecting the workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 12

 

Strain relieving the cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 17

 

Installation of input/output devices. . . . . . . . . . . . . . . . . . . . . . . . . 5 - 18

 

Installation of monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 19

 

Fine adjusting the monitor settings . . . . . . . . . . . . . . . . . . . . . . . .

5 - 19

 

Connecting the workstation to the network . . . . . . . . . . . . . . . . . . . .

5 - 21

 

 

 

 

 

Siemens AG

SPB7-230.814.01

Page 3 of 4

 

Opdima

Medical Solutions

Rev. 01 10.04

CS PS 24

Installation

0 - 4

Contents

 

Page

Installation of Opdima printer (option)

. . . . . . . . . . . . . . . . . . . . . . .5 - 22

Functional test (workstation table) . .

. . . . . . . . . . . . . . . . . . . . . . .5 - 23

6 ______ Final procedures _______________________________________________6 - 1

General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1 Starting up the MAMMOMAT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2 Starting up the workstation and login. . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2 Starting up the MO unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2 Installation of software on OPDIMA WORKSTATION . . . . . . . . . . . . . . . 6 - 2 Mounting of covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3 Filling in of IVK list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3 Performing of DHHS maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3

7 ______ Changes to previous version _____________________________________7 - 1

Opdima

SPB7-230.814.01

Page 4 of 4

Siemens AG

Installation

Rev. 01 10.04

CS PS 24

Medical Solutions

Prerequisites

1 - 1

General

This installation instruction is valid for MAMMOMAT NovationDR with serial numbers > 1100.

The wall socket for the power supply for the computers of Opdimarmust be connected to the same branch circuit as the MAMMOMAT NovationDR power supply

This Installation and Start-Up Instruction is valid for Opdimarsystems:

- with SUN workstation part No. 83 83 106/ Sun Blade 150 with ASW 3.1 and with serial No.W10000

Siemens AG

SPB7-230.814.01

Page 1 of 4

Opdima

Medical Solutions

Rev. 01 10.04

CS PS 24

Installation

1 - 2

Prerequisites

Training of customer support engineers

Due to the technology used in this equipment, setup, service and maintenance may only be carried out by a customer support engineer who has attended a training workshop or has participated in at least one installation.

Text emphasis

DANGER

WARNING

CAUTION

NOTICE

DANGER indicates when there is an immediate danger that leads to death or serious physical injury.

WARNING indicates a risk of danger that may lead to death or to serious physical injury.

CAUTION used with the safety alert symbol indicates a risk of danger that leads to slight or moderate physical injury and/or damage to property.

NOTICE used without the safety alert symbol indicates a risk of danger that if disregarded leads or may lead to a potential situation which may result in an undesirable result or state other than death, physical injury or damage to property.

NOTE

 

NOTE contains information provided with special emphasis to

 

 

facilitate proper use of the equipment or proper execution of a

 

 

 

 

procedure, i.e. hints, tips.

 

 

 

Opdima

SPB7-230.814.01

Page 2 of 4

Siemens AG

Installation

Rev. 01 10.04

CS PS 24

Medical Solutions

Prerequisites

1 - 3

Documents required

Supplement to the Instructions for Use MAMMOMAT NovationDR - Opdimar (included in the Opdimardelivery)

MAMMOMAT 1000/3000/3000 Nova - OpdimarService Instructions (included in the Opdimardelivery)

MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy Attachment and OpdimarDHHS Maintenance Instructions

(included in the Opdimardelivery)

MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy Attachment and OpdimarMeasurement Certificates

(included in the Opdimardelivery)

MAMMOMAT NovationDR- OpdimarPlanning Guide

MAMMOMAT NovationDR- OpdimarInstallation and Start-Up

MAMMOMAT NovationDR - OpdimarWiring Diagram

MAMMOMAT 3000 Installation and Setting Instructions Update Kit for Stereotactic Biopsy Attachment (where applicable)

Meters and appliances required

Protective ground wire tester (44 15 899 RV090)

Service PC (e.g. Siemens-Nixdorf PCD 3NSX/20 or similar) with connecting cable, PC to generator (part No. 99 00 440 RE999)

Stereo calibration phantom (included in the delivery) (part No.64 30 701)

AEC calibration plexiglass, four plates measuring 150 mm x 150 mm x 19 mm and one plate measuring 150 mm x 150 mm x 9,7 mm, part No. 65 61 232 and 65 61 224 respectively

Resolution phantom with at least 10 line pairs per mm

Tools required

Standard installation tools

Electric drill with 5 mm drill

Additional installation material required

Extension mains cord with at least three outlets, of protective earth type

Start-up, DHHS and X-ray decree area of application

Notes on start-up

Opdimar is adjusted, programmed and tested in the factory, leaving only the connection to the on-site mains voltage, the functional tests and customer dependent adjustments to be performed.

Siemens AG

SPB7-230.814.01

Page 3 of 4

Opdima

Medical Solutions

Rev. 01 10.04

CS PS 24

Installation

1 - 4

Prerequisites

Checking and recording for the area of application of the X-ray decree (§16 Germany)

An X-ray unit can only be put into operation if an acceptance test has been performed according to §16 RoeV. It is recommended that the responsible Customer Service Engineer is present to assist the person responsible for this acceptance test.

Checking and data recording in the USA

Maintenance measurements must be made according to the instructions in MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy Attachment and Opdimar DHHS Maintenance Instructions.

Sections

Required labels

Reproducibility

Automatic exposure control (AEC)

The result must be recorded in the document MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy Attachment and Opdimar Measurement Certificates.

Opdima

SPB7-230.814.01

Page 4 of 4

Siemens AG

Installation

Rev. 01 10.04

CS PS 24

Medical Solutions

Protective measures for CCD camera

2 - 1

CAUTION The CCD camera is very sensitive to mechanical shocks and temperature changes. In the camera shock and temperature sensors are integrated.

Risk of damaging the equipment.

The camera has to be handled with extreme care.

When disconnected from the biopsy controller, the camera shall always be stored in the attaché case delivered with the system. Do not touch the pins in the contacts of the camera.

The camera shall be used within 10230i C.

The camera shall be transported or stored within 0240i C.

Siemens AG

SPB7-230.814.01

Page 1 of 2

Opdima

Medical Solutions

Rev. 01 10.04

CS PS 24

Installation

2 - 2

Protective measures for CCD camera

This page intentionally left blank.

Opdima

SPB7-230.814.01

Page 2 of 2

Siemens AG

Installation

Rev. 01 10.04

CS PS 24

Medical Solutions

Siemens Mammomat Novation DR Installation Manual

Protective measures

3 - 1

It is very important that any intervention in the equipment shall start with disconnecting it from the power supply with the main circuit breaker. To prevent accidental triggering of high voltage and radiation, set the switch S2 (SS) on board D750 to OFF.

CAUTION When switching off the workstation use the power off procedure described in the Supplement to the Instructions for Use MAMMOMAT NovationDR- Opdimar.

Switching off the workstation before the software has been closed down may cause damage to the files on the hard disc.

WARNING If the system is only switched off at the control panel or with the circuit breaker CB S700, there is still voltage present within the generator. See the wiring diagram for MAMMOMAT NovationDR.

Life-threatening electric shock hazard exists.

Disconnect mains cable and comply with the information on this page.

WARNING After shut-down of the system, there may still be about 380 V DC present on the intermediate circuit of the MAMMOMAT generator.

Life-threatening electric shock hazard exists.

A bleader resistor is connected over the output DC capacitors.

The DC voltage will discharge below 60 Volt in 3 minutes.

The voltage level is indicated by LEDs H3 and H4.

CAUTION Observe the currently valid guidelines for handling electronics endangered by electrostatic discharge.

Use ESD-equipment, ground prior to making contact and place the components on a conductive surface.

The boards contain electrostatic highly sensitive components requiring particular care in their handling.

Risk of damaging components.

Siemens AG

SPB7-230.814.01

Page 1 of 2

Opdima

Medical Solutions

Rev. 01 10.04

CS PS 24

Installation

3 - 2

Protective measures

This page intentionally left blank.

Opdima

SPB7-230.814.01

Page 2 of 2

Siemens AG

Installation

Rev. 01 10.04

CS PS 24

Medical Solutions

Preparatory work

4 - 1

Scope of delivery

The following parts are always included in the delivery of your equipment:

Technical Manual comprising information for installation, operation and service

CCD camera with a fixed cable

Biopsy controller with five cables and cable duct

Stereo calibration phantom, 64 30 701 (see page 1-3)

Workstation including:

-power supply cable

-network cable

-KVM switch with cables

-MO unit with one power supply cable and one data cable

-Main unit including CD drive

-In some deliveries, the monitor, keyboard and mouse may be delivered. In such case the KVM switch is not needed.

If ordered, the following is delivered:

Biopsy unit

Compression plate for biopsy examination

Printed circuit board D702

MAMMOMAT software V2.2 or higher

Isolation p.c. board D707 kit

DICOM license key

Workstation table

Printer

Siemens AG

SPB7-230.814.01

Page 1 of 2

Opdima

Medical Solutions

Rev. 01 10.04

CS PS 24

Installation

4 - 2

 

 

Preparatory work

Unpacking

 

General

 

 

 

 

 

 

NOTE

 

Follow the directional marks on the box during transport, storage

 

 

 

and unpacking.

 

 

 

 

 

 

 

Workstation

The workstation is packed in a card board box which also contains technical documentation and CDs for application software and operating system1.

Biopsy controller, miscellaneous

The biopsy controller is packed in a card board box together with cables, miscellaneous installation material.

CCD camera

CAUTION

The CCD camera is very sensitive to mechanical shocks and temperature changes. In the camera shock and temperature sensors are integrated.

Risk of damaging the equipment.

The camera has to be handled with extreme care.

When disconnected from the biopsy controller, the camera shall always be stored in the attaché case delivered with the system. Do not touch the pins in the contacts of the camera.

The CCD camera is packed in an attaché case. Keep the case for future use.

Optional equipment and update kits

If optional equipment and/or update kits are ordered, these are packed in separate boxes.

1.Only needed if a software reinstallation is necessary (see MAMMOMAT 1000/3000 Nova - Opdimar Service Instructions). Store the documentation and software CDs in a proper place for future service use.

Opdima

SPB7-230.814.01

Page 2 of 2

Siemens AG

Installation

Rev. 01 10.04

CS PS 24

Medical Solutions