SYSTEM REPORT:

[SIEMENS MRI 0.35T, MODEL: Magnetom C]

Probo Medical Ltd

The images used in the report are for reference

Mount Court

only and may show objects that are not included

Edison Close

in the offered system configuration. Please refer to

Gloucester, GL2 2FN

the sales order or contact Seller for clarification.

www.probomedical.co.uk

 

QF24 QA InspectionReport– MRI

V1, November2021

SYSTEM LOCATION

32, Orchard Rd FY8 1PF Black Pool

PROJECT NO.

 

PERFORMED BY

Jasvir Singh

DATE

07.12.2022

SYSTEM INFORMATION

 

Manufacturer and model

SIEMENS, MODEL: Magnetom C

System model no.

10019418

System serial no.

29113

System YOM

2006

Gradient type

Concerto 0.35

Number of Channels

4

MAGNET INFORMATION

 

Magnet YOM

2006

Magnet model no.

08640463

Magnet serial no.

29113

MAGNET INFORMATION

 

Current helium level

N/A

Magnet pressure

N/A

Shield temperature

N/A

COOLING SYSTEM INFORMATION

 

Compressor model no.

N/A

Compressor YOM

N/A

Compressor running hours

NA

Compressor pressure level

N/A

Last absorber replacement

N/A

Cold-head model no.

N/A

Cold-head YOM

NA

Last cold-head replacement

NA

PATIENT TABLE INFORMATION

 

Table. model no.

08882800

Table serial no.

NA

Table YOM

2006

Extra trolley available?

YES, Model: 04758434

Positioning pads incl.?

YES

CONSOLE AND WORKSTATION(S) INFORMATION

Console model no.

Host PC Model: 10143230, Imager: 10143232

Console serial no.

1412, 1857

Console software version

A 30

Console software options

Picture attached

ACCESSORIES, MANUAL HARD-COPIES AND SOFTWARE DISCS

Software and manuals (list)

Manuals available but software CDs not available.

Probo Medical Ltd

The images used in the report are for reference

Mount Court

only and may show objects that are not included

Edison Close

in the offered system configuration. Please refer to

Gloucester, GL2 2FN

the sales order or contact Seller for clarification.

www.probomedical.co.uk

 

QF24 QA InspectionReport– MRI

V1, November2021

Any additional items/accessories?

NA

MAINTENANCE

 

Any major part replacements?

NA

Is system logbook present?

NA

COMMENTS

 

Functional condition

 

Cosmetic condition

Good

 

 

COILS

 

 

 

Coil name & channels

Model no.

Serial no.

Test result

1. Wrist Array Coil

10045324

1039

N/A

2.

Neck Array large

08635562

2175

N/A

3.

Neck Array medium

08635679

1605

N/A

4.

Body spine array XXL

08635505

1118

N/A

5.

Body spine array Large

08635521

1087

N/A

6.

Body spine array medium

08635539

1543

N/A

7.

Body spine lower coil

07100121

2019

N/A

8.

Shoulder array coil

08635620

1776

N/A

9.

Extremity array S

08635596

1221

N/A

10. Extremity array L

08635588

NA

N/A

11. Head Array

08635554

1839

N/A

PHANTOMS

 

 

 

Model name

Amount

Model no.

Serial no.

All phantoms available (picture

 

 

 

attached)

 

 

 

QA TESTING

 

 

 

[List of tests performed + results]

QA Could not performed because system is not working order.

Probo Medical Ltd

The images used in the report are for reference

Mount Court

only and may show objects that are not included

Edison Close

in the offered system configuration. Please refer to

Gloucester, GL2 2FN

the sales order or contact Seller for clarification.

www.probomedical.co.uk

 

QF24 QA InspectionReport– MRI

V1, November2021