SYSTEM REPORT:
[SIEMENS MRI 0.35T, MODEL: Magnetom C]
Probo Medical Ltd |
The images used in the report are for reference |
Mount Court |
only and may show objects that are not included |
Edison Close |
in the offered system configuration. Please refer to |
Gloucester, GL2 2FN |
the sales order or contact Seller for clarification. |
www.probomedical.co.uk |
|
QF24 QA InspectionReport– MRI |
V1, November2021 |
SYSTEM LOCATION |
32, Orchard Rd FY8 1PF Black Pool |
PROJECT NO. |
|
PERFORMED BY |
Jasvir Singh |
DATE |
07.12.2022 |
SYSTEM INFORMATION |
|
Manufacturer and model |
SIEMENS, MODEL: Magnetom C |
System model no. |
10019418 |
System serial no. |
29113 |
System YOM |
2006 |
Gradient type |
Concerto 0.35 |
Number of Channels |
4 |
MAGNET INFORMATION |
|
Magnet YOM |
2006 |
Magnet model no. |
08640463 |
Magnet serial no. |
29113 |
MAGNET INFORMATION |
|
Current helium level |
N/A |
Magnet pressure |
N/A |
Shield temperature |
N/A |
COOLING SYSTEM INFORMATION |
|
Compressor model no. |
N/A |
Compressor YOM |
N/A |
Compressor running hours |
NA |
Compressor pressure level |
N/A |
Last absorber replacement |
N/A |
Cold-head model no. |
N/A |
Cold-head YOM |
NA |
Last cold-head replacement |
NA |
PATIENT TABLE INFORMATION |
|
Table. model no. |
08882800 |
Table serial no. |
NA |
Table YOM |
2006 |
Extra trolley available? |
YES, Model: 04758434 |
Positioning pads incl.? |
YES |
CONSOLE AND WORKSTATION(S) INFORMATION |
|
Console model no. |
Host PC Model: 10143230, Imager: 10143232 |
Console serial no. |
1412, 1857 |
Console software version |
A 30 |
Console software options |
Picture attached |
ACCESSORIES, MANUAL HARD-COPIES AND SOFTWARE DISCS |
|
Software and manuals (list) |
Manuals available but software CDs not available. |
Probo Medical Ltd |
The images used in the report are for reference |
Mount Court |
only and may show objects that are not included |
Edison Close |
in the offered system configuration. Please refer to |
Gloucester, GL2 2FN |
the sales order or contact Seller for clarification. |
www.probomedical.co.uk |
|
QF24 QA InspectionReport– MRI |
V1, November2021 |
Any additional items/accessories? |
NA |
MAINTENANCE |
|
Any major part replacements? |
NA |
Is system logbook present? |
NA |
COMMENTS |
|
Functional condition |
|
Cosmetic condition |
Good |
|
|
|
COILS |
|
|
|
|
Coil name & channels |
Model no. |
Serial no. |
Test result |
|
1. Wrist Array Coil |
10045324 |
1039 |
N/A |
|
2. |
Neck Array large |
08635562 |
2175 |
N/A |
3. |
Neck Array medium |
08635679 |
1605 |
N/A |
4. |
Body spine array XXL |
08635505 |
1118 |
N/A |
5. |
Body spine array Large |
08635521 |
1087 |
N/A |
6. |
Body spine array medium |
08635539 |
1543 |
N/A |
7. |
Body spine lower coil |
07100121 |
2019 |
N/A |
8. |
Shoulder array coil |
08635620 |
1776 |
N/A |
9. |
Extremity array S |
08635596 |
1221 |
N/A |
10. Extremity array L |
08635588 |
NA |
N/A |
|
11. Head Array |
08635554 |
1839 |
N/A |
|
PHANTOMS |
|
|
|
|
Model name |
Amount |
Model no. |
Serial no. |
|
All phantoms available (picture |
|
|
|
|
attached) |
|
|
|
|
QA TESTING |
|
|
|
|
[List of tests performed + results] |
QA Could not performed because system is not working order. |
Probo Medical Ltd |
The images used in the report are for reference |
Mount Court |
only and may show objects that are not included |
Edison Close |
in the offered system configuration. Please refer to |
Gloucester, GL2 2FN |
the sales order or contact Seller for clarification. |
www.probomedical.co.uk |
|
QF24 QA InspectionReport– MRI |
V1, November2021 |