EasyDiagnost Eleva with Digital RF

Preferred Room Layout

The layout shown below is based upon a typical equipment configuration and should be considered as a general design guideline. Site conditions, application requirements, customer preferences, and/or equipment configuration may significantly impact suite design and equipment layout. It is recommended to request site-specific drawings from a Philips representative early in the design process.

 

 

 

 

 

 

 

19'-0"

 

 

 

 

 

 

 

 

 

 

 

 

[5791mm]

 

 

 

 

 

 

 

 

 

 

 

9'-7"

Table Tilt Axis

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

[2921mm]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

IB

 

 

 

 

 

 

Centerline

5'-0"

[1524mm]

 

 

MS

 

 

 

 

 

0"-18'

[5486mm]

LongitudinalTable

 

CenterlineBuckyWall

6"-3'

[1067mm]

 

 

 

0"-12'

[3658mm]

 

 

 

 

 

 

 

 

 

 

 

IR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TV

 

 

 

 

 

 

 

 

 

DVS

7' (2130mm)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

high wing wall

 

 

 

 

 

 

 

 

 

 

 

 

MU

 

2'-6"

[762mm]

8"-5'

[1727mm]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

QR

EEC

VF

 

 

 

 

 

 

 

 

 

4' [1220mm]

 

 

 

 

 

 

 

 

 

 

 

 

Wide Patient Entry

 

 

 

 

 

 

 

 

 

 

 

 

(Recommended)

 

 

 

 

 

 

Equipment Layout

Recommended Ceiling Height:

9' -

1" (2770mm) (tube arm up) /

 

10'

- 4" (3150mm) (tube arm down)

Ceiling heights (from top of floorplate to bottom of Unistrut) other than recommended may impact equipment functionality; consult with Philips.