Omega Products PHUTX601 Manual

0 (0)

1⁄4 DIN pH, ORP, and Conductivity Controllers and Transmitters

PHUCN601

UFor pH, ORP, Conductivity/Resistivity

U14 DIN, NEMA 4X (IP66), Polycarbonate Enclosure

UUser-Friendly Design with Interior LCD Menu

UEasy to Read with Large, Bright LED Display

UQuick and Easy to Calibrate

UReduces Chemical Consumption

with Cycle-Timed Relays on Controller

UTwo Control Relays with Third Relay for High/Low Alarm on Controller

The PHUCN600 Series multi-parameter controller is a microprocessor-based controller capable of measuring pH, ORP, or conductivity.

There are two displays on the PHUCN600 Series: a bright LED numeric display with bar graph on the outside front panel, and a 2-line, 16-character LCD display on the inside. The LED readout on the outside panel can be seen several yards away. The distinctive, color-coded bar graph will immediately indicate if you are within the process parameters that you set (green), if the control relays are on (yellow), and if you are in alarm condition (red). This makes diagnosing pump and alarm malfunctions easy. All configuration and control functions are performed on the LCD menu on the inside front panel.

A universal mounting kit is included for surface, panel, and pipemount applications. The ¹⁄ DIN enclosure makes panel-mount cutouts and engineering simple.

Complete and versatile, the PHUTX600 is a ¹⁄ DIN two-wire transmitter, which can measure conductivity, pH, or ORP. The PHUTX600 comes complete with a universal mounting kit for surface, panel, and pipe-mount applications. The NEMA 4X (IP66) ¹⁄ DIN enclosure is perfect for stand-alone or panel-mount operation.

PHUCN600 and PHUTX600 Series are packaged in a rugged NEMA 4X (IP66) polycarbonate enclosure, making them ideally suited for heavy-duty applications such as industrial wastewater neutralization, municipal water and wastewater, pulp and paper, and process control.

When the PHUCN600 Series is powered up for the first time,

it will display the meter selection screen where the meter type is selected. This meter selection screen will only be displayed when the PHUCN600 Series is powered up for the first time.

After the user selects a meter type, the PHUCN600 Series will remain set to that meter type until it is changed with the meter selection menu function in the Utilities menu.

To return the PHUCN600 Series to its factory settings, the user re-selects the current meter type from the meter selection menu function. This will override all setpoints and return all settings to the factory settings.

The PHUCN600 Series User’s menu has been divided into five main categories:

Calibration, used to calibrate the PHUCN600 Series with the selected sensor

Utilities, used to manually control or override the outputs

Setup, used to configure the PHUCN600 Series’ many options

Diagnostics, used to troubleshoot any problems

Outputs, used to configure the PHUCN600 Series outputs

D-X

For Sales

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

& Service

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

U.S.A. and Canada

Omega Products PHUTX601 Manual

PHUCN601 Enclosure dimensions: mm (inches)

MOUNTING

BRACKET

91 (3.60)

 

MINIMUM

 

 

 

 

INTERIOR

 

 

 

 

CLEARANCE

91

 

 

 

FOR PANEL

 

 

 

MOUNTING

(3.60)

 

 

 

162 (6.37)

 

 

 

 

 

PANEL CUTOUT

 

 

 

97

 

 

 

 

(3.80)

 

 

 

TOP VIEW

 

 

 

 

 

184

 

 

 

 

(7.23)

7.0

 

 

 

 

64

 

 

145

(0.26)

 

 

(2.5)

 

 

4 PLACES

 

 

(5.70)

 

 

 

 

 

 

PHUCN601

 

 

 

ALARM

3.80

29

5.0

117

RELAY B

(0.20)

 

(97)

(1.15)

(4.63)

RELAY A

4 PLACES

ALARM

 

 

 

 

TEMP

 

 

 

CAL

UP DOWN ENTER

 

 

 

 

 

12 CONDUIT HOLES

 

 

 

 

4 PLACES

 

 

FRONT VIEW

128

10

MOUNTING

 

 

(5.05)

(0.38)

BRACKET

SIDE VIEW

PHUCN601 inside shown smaller than

Mounting

PANEL MOUNT

SURFACE MOUNT

SPECIFICATIONS FOR PHUCN600 AND PHTX600 SERIES

 

pH

ORP

Conductivity

Display, PHUCN600/610 Series

Front Panel: 4 x 7 segment 12.7 mm (12") LED display, 1 LED indicator online, 7 LED bar graph

 

Inside Panel: 2 x 16 alpha-numeric LCD display

 

Display, PHUTX600

2 x 16 alpha numeric LCD display

 

 

Power, PHUCN600/610 Series

120 or 240 Vac 50/60 Hz (less than 12 VA)

 

Power, PHUTX600

4 to 20 mA loop-powered 16 to 32 Vdc

 

Measuring Range

0.01 to 14.00 pH

-1999 to 1999 mV

µS/cm3: 0 to 2.000,

 

-5 to 95°C

-5 to 95°C

0 to 20.00, 0 to 200.0, 0 to 2000,

 

(23 to 203°F)

(23 to 203°F)

mS/cm: 0 to 20.00, 0 to 200.0

 

 

 

MΩ/cm: 0 to 19.99

 

 

 

0 to 100°C (32 to 212°F)

Temperature Compensation

Automatic or manual,

Not required

Automatic or manual;

 

-5 to 95°C (23 to 203°F)

 

User-selectable temperature

 

 

 

compensation slope 0.0 to 10.0%/°C.

 

 

 

0 to 100°C (32 to 212°F).

Temperature Units

°C or °F, selectable

 

 

Temperature Sensor

User selectable: 300 Ω NTC thermistor, 3000 Ω NTC thermistor or Pt. 1000 RTD

Calibration Modes

Auto-calibration

Manual calibration

Dry calibration

 

Manual calibration

Temperature

Sample calibration

 

Temperature calibration

calibration

Temperature calibration

Ambient Conditions

-20 to 60°C (-4 to 140°F), 0 to 90% RH (non-condensing)

 

Menu Access Front Panel

Auto-calibration

Manual calibration

Dry calibration

 

Manual calibration

Temperature calibration

Sample calibration

 

Temperature display

 

Temperature calibration

Menu Access Inside Panel

Full Access to all parameters of operations menu

 

Sensor to Transmitter Distance

Differential Sensor: 914 m (3000')

914 m (3000')

91 mm (300')

 

Combination Sensor: 3.0 m (10')

3.0 m (10')

3.0 m (10')

Relay Outputs

Two Control Relays: 10 A/NO, 5 A/NC @ 240 Vac or 28 Vdc.

 

for PHUCN600/610 Series

Mode: Process control; adjustable parameters: process direction, (rising or falling) on-setpoint,

 

off setpoint, (0 to 100% of full scale), cycle timer (on/off, 0 to 600 seconds), failsafe (on/off).

 

One Alarm Relay: 10 A/NO, 5 A/NC @ 240 Vac or 28 Vdc.

 

 

Mode: high/low alarm; adjustable parameters: low on/low off setpoint (0 to 100% of full scale,

 

low on must be less than low off), high on/high off setpoint (0 to 100% of full scale, high on must

 

be greater than high off).

 

 

Analog Outputs

4 to 20 mA channel 1 isolated output, range expand 0 to 100% of full scale

 

(min segment 10% of full scale), max load 800 V

 

Order Online

 

 

 

omega.com®

 

D-X

Over 100,000 Products Available!

Loading...
+ 2 hidden pages