Smart Temperature
Transmitters
TX13
RoHS
UK, J, N, E, T, R, S Thermocouples and mV Inputs
UPushbutton Configuration
U56 Preprogrammed Ranges Ready-to-Use
UUser Trim—Null Out Sensor Error
UProgrammable Burnout with Sensor Fail Indicator
UGalvanically Isolated
The TX13 is a “smart” in-head transmitter that accepts thermocouple temperature sensors and converts sensor output over a configured range to a standard industrial (4 to 20 mA) transmission signal.
A simple pushbutton operation allows the user to select TC type, burnout direction, select fixed ranges and trim 4 and 20 mA points. The
in head transmitter incorporates the latest digital technology to ensure accurate drift-free performance.
If required the desired range can be specified at the time of order, removing the need for user configuration. If the range is not specified then the transmitter will be shipped with the default range of
0 to 1000°C (32 to 1832°F) Type K.
Specifications
Isolation: Tested to 250 Vdc
Sensor Burnout: Either up or down scale output
Cold Junction:
Range: -40 to 85°C (-40 to 185°F) Accuracy: ±0.5°C (0.9°F) Tracking: ±0.05°C/°C (0.9°F/°F)
Stability:
Offset: 0.1°C/°C (0.18°F/°F) Span: 0.05°C/°C (0.18°F/°F)
Output
Output Type: 2 wire, 4 to 20 mA current loop
Output Range: 4.0 to 20.0 mA Output Connection: Screw terminal
Maximum Output: 21.5 mA (in high burnout condition)
Minimum Output: 3.8 mA (in low burnout condition)
Accuracy: (mA output/2000) or 5 uA (whichever is the greater)
Loop Voltage Effect: 0.2 uA/V |
|
TX13 shown |
|
||||||
Thermal Drift: 1 uA/°C |
|
|
|||||||
Maximum Output Load: |
|
actual size. |
|
||||||
|
|
|
|||||||
[(Vsupply-10)/20] kΩ |
|
|
|
||||||
(Example: 700 Ω @ 24V) |
|
|
|
||||||
Update Time: 500 mS |
|
|
|
||||||
Response Time: 1 second |
|
Approvals |
|
||||||
Start Up Time: 4 seconds |
|
EMC - BS EN 61326: 1998: Electrical |
|
||||||
(1 out < 4 mA during start up) |
|
equipment for measurement control and |
|
||||||
Warm-Up Time: 1 minute to full accuracy |
laboratory use |
|
|||||||
Power Supply: 10 to 30 Vdc |
|
ANNEX A: Immunity test requirements for |
|
||||||
Environmental |
|
equipment intended for use in industrial |
|
||||||
|
locations |
|
|||||||
Ambient Operating Range: |
|
ANNEX F: Test configurations, operational |
|
||||||
-40 to 85°C (-40 to 185°F) |
|
conditions and performance criteria for |
|
||||||
Ambient Storage Temperature: |
|
transducers with integrated or remote signal |
|
||||||
-50 to 90°C (-58 to 194°F) |
|
conditioning |
|
||||||
Ambient Humidity Range: |
|
IEC 61000-4-2: Electrostatic discharge |
|
||||||
10 to 90% RH non-condensing |
|
IEC 61000-4-3: EM field |
|
||||||
Physical |
|
|
|
IEC 61000-4-4: Transient burst (output) |
|
||||
Dimensions: 43 Dia x 21 mm H |
|
IEC 61000-4-5: Surge (output) |
|
||||||
(1.7 x 0.8") |
|
|
|
Note: Sensor input wires to be less than |
|
||||
Weight: 31 g (0.07 lb) (encapsulated) |
|
3 m (10') to comply. |
|
||||||
Specifications @ 20°C–Input |
|
|
|
||||||
Sensor |
Range °C (°F) |
|
Accuracy |
|
|||||
|
|
|
|
|
-200 to 1370 |
|
±0.5°C (0.9°F) or ±0.1% whichever |
|
|
|
|
|
|
|
(-328 to 2498) |
|
the greater |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
-100 to 1200 |
|
±0.5°C (0.9°F) or ±0.1% whichever |
|
|
|
|
|
|
|
(-148 to 2192) |
|
the greater |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
-100 to 1000 |
|
±0.5°C (0.9°F) or ±0.1% whichever |
|
|
|
|
|
|
|
(-148 to 1832) |
|
the greater |
|
|
|
|
|
|
|
|
|
|||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
-180 to 1300 |
|
±0.5°C (0.9°F) or ±0.1% whichever |
|
|
|
|
|
|
|
(-292 to 2372) |
|
the greater |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
-100 to 400 |
|
±0.5°C (0.9°F) or ±0.1% whichever |
|
|
|
|
|
|
|
(-148 to 752) |
|
the greater |
|
|
|
|
|
|
|
|
|
|||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
-10 to 1760 |
|
±0.5°C (0.9°F) or ±0.1% whichever |
|
|
|
|
|
|
|
(14 to 3200) |
|
the greater over range 800 to 1600 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
-10 to 1760 |
|
±0.5°C (0.9°F) or ±0.1% whichever |
N |
|
|
|
|
|
|
(14 to 3200) |
|
the greater over range 800 to 1600 |
||
mV |
-10 to 70 |
|
±0.02% of full scale |
|
|||||
(14 to 158) |
|
|
|
|
|||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
To Order Visit omega.com/tx13 for Pricing and Details |
|
||||||||
Model No. |
Description |
|
|||||||
TX13 |
Thermocouple/mV smart transmitter |
|
|||||||
TX2-100 |
2-conductor shielded cables, 30 m (100') |
|
|||||||
PSR-24S |
Regulated power supply, 24 Vdc, 400 mA, |
|
|||||||
|
|
|
|
|
screw terminal |
|
|||
PSR-24L |
Regulated power supply, 24 Vdc, 400 mA, UL, |
|
|||||||
|
|
|
|
|
stripped leads |
|
PSR-24L-230 Regulated power supply, 24 Vdc, 400 mA, stripped leads, 230 Vac input, CE
TX-SCALED Scaling charge for factory set up of range (specify temperature range)
Comes complete with operator’s manual.
Ordering Example: TX13, thermocouple smart transmitter.
N-36