CH-9101 Herisau/Switzerland
E-Mail info@metrohm.com
Internet www.metrohm.com
Titrando
Installation Instructions
8.840.1133 06.2006 / jb
Teachware
Metrohm AG
Oberdorfstrasse 68
CH-9101 Herisau
teachware@metrohm.com
These instructions are protected by copyright. All rights reserved.
Although all the information given in these Instructions has been checked with great care, errors
cannot be entirely excluded. Should you notice any mistakes please inform the author at the
address given above.
Fig. 19: Connecting the 854 iConnect .................................................................................... 29
Fig. 20: Recommended arrangement of magnetic stirring bar (1), electrode (2) and
buret tip (3) ...................................................................................................................30
Fig. 21: Drawing of the KF titration cell 6.5609.000................................................................ 31
Fig. 22: Arrangement of transport tip, buret tip and draw-off tip ........................................... 32
Titrando Installation Instructions III
Contents
IVTitrando Installation Instructions
1 Introduction
1 Introduction
These installation instructions provide you with a comprehensive overview of the installation and specifications of the Titrando system. The
Titrando is the centerpiece of the modular Titrando system. Operation is
carried out either via a Touch Control with touch-sensitive screen
("stand-alone" titrator) or from a computer via the USB connection and
using the PC software PC Control or tiamo.
A Titrando system can integrate several, different devices. With PC
Control/Touch Control, up to three control devices (Titrandos, Dosing
Interfaces, USB Sample Processors, etc.) can be controlled via USB
connection. The tiamo software allows to expand the system with practically any number of control devices. The control devices have three to
four MSB connections (MSB = Metrohm Serial Bus), to which peripheral devices can be connected. These are mainly dosing drives
(700/800 Dosinos and 685/805 Dosimats), stirrers and titration stands,
etc.
Figure 1 shows you the flexibility of the Titrando system. On the left a Titrando is shown with external dosing devices operated by the Touch
Control. To the right of it you can see an automation system consisting
of a USB Sample Processor, a Titrando with internal dosing drive and a
Dosimat operated by the PC Control software.
Fig. 1: The Titrando system
Titrando Installation Instructions 1
1.1 Instrument description
Additional information about the Titrando system can be found on the
Internet under www.titrando.com
Information about specific applications can be found in our "Applica-tion Bulletins" and "Application Notes"; these can be obtained free
of charge from your local Metrohm agency or downloaded from the
Internet under www.metrohm.com.
jects of titration techniques and electrodes are also available.
.
Various monographs on the sub-
1.1 Instrument description
The Titrandos differ mainly from the kind of dosing drive. They are
equipped either with an internal dosing drive for an exchange unit (type
806 or previous) or for the use of external dosing devices (700 and 800
Dosinos with 807 dosing units).
Common features of the Titrando:
• A connection for Touch Control or for a computer with PC software
PC Control or tiamo.
• Four MSB connections (Metrohm Serial Bus) each controlling one
dosing device (Dosimat with exchange unit or Dosino with dosing
unit), a stirrer or titration stand and a remote box.
• One or two measuring interfaces. A measuring interface consists of
a high-impedance measuring input for pH, redox or ISE sensors, an
input for a separate reference electrode, a measuring input for temperature sensors (Pt1000 or NTC) and a measuring input for polarized electrodes.
• Two USB connections allow to connect, for example, a printer, keyboard, barcode reader or additional control devices (USB Sample
Processor, Titrando, Dosing Interface, etc.).
2Titrando Installation Instructions
1 Introduction
1.2 Information about the Installation Instructions
Attention!
Please study these Installation Instructions carefully before you start to
use the Titrando. The instructions contain information and warnings
that must be observed by the user in order to guarantee the safe use
of the instrument. Please keep these instructions near the instrument
so that they are always to hand when required.
1.2.1 Organization
These Installation Instructions for the Titrandos provide you with a comprehensive overview of the installation, startup, troubleshooting and
technical specifications of the instruments.
The Installation Instructions are arranged as follows:
Introduction
General description of the instrument, operating elements and safety information
Installation
Installation of the instrument, connection of the peripheral devices and
accessories
Troubleshooting
Description of possible errors and how to remedy them
Appendix
Technical data, standard equipment, optional accessories, warranty
and declaration of conformity
Index
In order to find the information you require about the Titrando you
should either use the Table of contents or the Index.
Titrando Installation Instructions 3
1.2 Information about the Installation Instructions
1.2.2 Notation and pictograms
The following notation and pictograms are used in these Instructions:
9
[Continue]
Control element, instrument element
see illustrations in Section 1.3
Instruction
Carry out the instructions step by step.
Button
on the user interface
Danger
This symbol indicates a possible risk of
death or injury to the user and possible
damage to the instrument or its components by electricity.
Warning
This symbol indicates a possible risk of
damage to the instrument or its components if the given information is not properly
observed.
Attention
This symbol indicates important information
that you should read before continuing.
Information
This symbol indicates additional information
and tips that may be particularly useful.
4Titrando Installation Instructions
1 Introduction
1.3 Parts and controls
Fig. 2: Front view of a Titrando with internal dosing drive
1 Guide openings
for centering the exchange unit
2 Push rod
of the dosing drive
3 Contact pins
for the data chip
4 Coupling
for switching the flat cock
5 "On" LED
Lights up when the Titrando is connected to the mains supply and a controller (Touch Control or computer) is
connected and switched on.
6 "Status" LED
Shows the current status of the internal
dosing drive (see Section 2.4.2).
Titrando Installation Instructions 5
1.3 Parts and controls
Fig. 3: Front view of a Titrando for the use of external dosing devices
7 Bottle holder
with holding clips, for two reagent bottles
8 "On" LED
Lights up when the Titrando is connected to the mains supply and a controller (Touch Control or computer) is
connected and switched on.
6Titrando Installation Instructions
1 Introduction
Fig. 4: Rear view of the Titrando
9 USB connections USB 1 and USB 2
USB ports (type A) for connection of
printer, keyboard, barcode reader, additional Titrandos, USB Sample Processor, etc.
10 Controller connection
Connection for Touch Control or PC
with installed PC software
11 MSB connections MSB 1 to MSB 4
Metrohm Serial Bus
Connection for external dosing device,
stirrer or remote box
Connection of pH, redox or ISE sensors with built-in or separate reference
electrode
17 Connection for temperature sen-
sors (Temp.)
Pt1000 or NTC
18 Measuring input for polarized elec-
trodes (Pol.)
e. g. double Pt electrodes
Titrando Installation Instructions 7
1.4 Safety notes
1.4 Safety notes
Warning!
This instrument should only be used in accordance with the information given in these installation instructions.
1.4.1 Electrical safety
Please observe the following guidelines:
• Only qualified Metrohm technicians should carry out service work
on electronic components.
• Do not open the Titrando housing. This could destroy the Titrando.
Inside the housing there are no components that the user can service or exchange.
Electrical safety when handling the Titrando is guaranteed within the
framework of the IEC 61010 Standard. The following point must be observed:
Protection against electrostatic charges
Warning!
Electronic components are sensitive to electrostatic charges and can
be destroyed by a discharge. Always remove the mains cable from
the power supply socket before setting up or breaking electrical connections on the rear panel of the instrument.
8Titrando Installation Instructions
2 Installation
2 Installation
This section describes what you should pay attention to when unpacking and setting up the Titrando. It also informs you about how a complete titration system – from a simple system with stirrer and printer up
to a complicated system with additional dosing devices, sample
changer and balance – is assembled.
The following illustration provides an overview of the peripheral devices
that can be connected to a Titrando:
Fig. 5: Titrando – Peripheral devices
Titrando Installation Instructions 9
2.1 Overview
2.1 Overview
The following flow diagram provides an overview of the installation of a
simple titration system with stirrer, external dosing device, printer and
balance. More detailed information can be found in the given sections.
Setup Section 2.2
Touch Control/Computer connection Section 2.3
Stirrer/Titration stand connection Section 2.4.1
Dosing device connection Section 2.4.2 and 2.4.3
Printer connection Section 2.5.1
Balance connection Section 2.5.2
10Titrando Installation Instructions
2 Installation
2.2 Instrument setup
2.2.1 Packaging
The Titrando and the separately packed accessories are delivered in
special packaging that provides excellent protection. Please store this
packaging in a safe place as no other packaging can guarantee the
safe transport of the instrument.
2.2.2 Checks
Please check whether the delivery is complete and undamaged immediately on receipt (compare with delivery note and list of accessories
given in Section 4.2). If transport damage is evident please refer to the
information given in Section 4.4.1.
2.2.3 Location
The Titrando was developed for internal laboratory use; it should not be
used in explosion-endangered locations.
Place the instrument on a suitable vibration-free laboratory bench, protected as much as possible from corrosive atmospheres and contact
with chemicals.
Choose a location where the temperature is usually between +5 °C and
+45 °C. The instrument should be protected against excessive variations in temperature and direct sunlight.
2.3 Controller connection
Two different methods of controlling the Titrando are available:
• The Touch Control with contact-sensitive screen forms a “stand-
alone” titrator together with the Titrando.
• A computer can be used to control the Titrando by using the PC
software PC Control or tiamo.
Attention!
Make sure that the mains cable has been removed from the power
supply socket before setting up or breaking connections between the
instruments.
Titrando Installation Instructions 11
2.3 Controller connection
2.3.1 Touch Control connection
Connect the plug of the Touch Control connection cable to the con-
troller socket.
Fig. 6: Titrando – Touch Control
Note!
The plug is fitted with a "pull-out protection device" that prevents the
cable from being pulled out accidentally. When you wish to insert or
remove the plug you must first pull back the outer plug sleeves
(marked with arrows).
Connect all peripheral devices (see Section 2.4 and 2.5) before you
switch on the Touch Control.
Power for the Touch Control is provided by the Titrando.
Connect the Titrando to the mains supply and switch on the Touch
Control.
After switch-on automatic system tests are carried out on both the Titrando and the Touch Control. The "On" LED on the Titrando lights up
when the system test is finished and the instrument is ready for use.
Attention!
The Touch Control has to be shut down properly with the ON/OFF
switch at the back of the instrument before the power supply is interrupted. Otherwise data can be lost. Since the power for the Touch
Control is supplied by the Titrando, never disconnect the Titrando
from the mains connection (e. g. by switching it off via a mains distributor), before you have switched off the Touch Control.
If you do not want to place the Touch Control directly alongside the Titrando then you can extend the connection between the Titrando and
Touch Control with the 6.2151.010 Cable. The connection length must
not exceed 5 m.
12Titrando Installation Instructions
2 Installation
2.3.2 Computer connection
Install the PC Control or tiamo software on your computer. Quit
the program if you have started it after the installation.
Connect all peripheral devices (see Section 2.4 and 2.5) before you
connect the Titrando to the mains supply.
Connect the Titrando to the mains supply. The LED "On" on the Ti-
trando will not yet light up!
Connect the Titrando to a USB connection (type A) on your com-
puter with the 6.2151.000 cable (see the instruction manual for your
computer).
Windows 2000: The Titrando will be recognized and the driver will be
installed automatically. Windows XP: The Titrando will be recognized
and the wizard for the installation of the driver will be started automatically. Select the option "Install the software automatically" and click on
[Next]. End the wizard with [Finish].
PC Control
Connect the USB dongle ("authorization plug") supplied with the full
version of the PC Control software to any USB socket (type A) of the
computer or Titrando.
Windows 2000: The USB dongle will be recognized and the driver will
be installed automatically. Windows XP: The USB dongle will be recognized and the wizard for the installation of the driver will be started
automatically. Select the option "Install the software automatically" and
click on [Next]. End the wizard with [Finish].
Fig. 7: Titrando – Computer
Note!
The plug for connecting to the Titrando is fitted with a "pull-out protection device" that prevents the cable from being pulled out accidentally.
When you wish to insert or remove the plug you must first pull back
the outer plug sleeves (marked with arrows).
You can extend the connection with a commercially available USB extension cable (type A/m – type A/f). The length of the connection should
not exceed 5 m. If you require a longer connection then you must use a
commercially available USB signal amplifier. Up to five USB signal am-
Titrando Installation Instructions 13
2.4 Device connection at the MSB
plifiers can be connected in series; this allows a maximum extension of
25 m.
Start the PC Control or tiamo software.
The Titrando will be recognized automatically. When the PC Control or
tiamo software is started a system test will be carried out automatically
on the Titrando. The LED "On" on the Titrando lights up when the system test is finished and the instrument is ready for use.
2.4 Device connection at the MSB
The following devices can be connected via each of the four MSB connections (Metrohm Serial Bus): one stirrer or titration stand, one Dosimat or Dosino dosing device and one remote box. The stirrers and
remote boxes each have an MSB output so that the devices can be
switched sequentially ("daisy chain"). The following illustration provides
an overview of the devices that can be connected to an MSB.
Fig. 8: Overview of MSB connections
The MSB 1 of the Titrando with internal dosing drive is occupied by the
internal dosing drive. This means that only a stirrer and a remote box
can be connected to MSB 1.
First connect all peripheral devices and then connect the Titrando to the
mains supply. MSB connections can be extended with the 6.2151.010
cable. The maximum length of the connection is 15 m.
The Titrando automatically recognizes which device has been connected to which MSB connection. The Touch Control or the PC software
(PC Control or tiamo) shows the connected devices in the device manager or the configuration dialog respectively. All devices connected to
the MSB are operated by the Touch Control or PC Control/tiamo.
14Titrando Installation Instructions
2 Installation
Attention!
If you are operating the Titrando with the Touch Control then make
sure that the Touch Control is switched off while you are setting up or
breaking connections between the instruments. If you are operating
the Titrando with the PC software then pull out the plug from the
power supply socket before setting up or breaking MSB connections.
2.4.1 Connecting stirrers and titration stands
You can use the 801 Magnetic Stirrer, the 803 Titration Stand (stirring from below) or the 804 Titration Stand with 802 Rod Stirrer (stirring from above).
Connect the appropriate stirrer as follows:
Fig. 9: Titrando – Stirrer
You can connect a maximum of one 801 Stirrer, one 804 Titration Stand
with 802 Rod Stirrer or one 803 Titration Stand to each MSB socket. The
801 Stirrer as well as the 804 and 803 Titration Stands has a MSB connection to which an additional device, e. g. a dosing device, can be
connected. We recommend that the stirrer is connected to MSB 1, as
this corresponds to the default setting in the methods.
The stand and the stand support are included with the stirrer or titration
stand. Fasten the stand support to the base of the Titrando with the four
screws supplied. Decide whether you want to attach the stirrer to the
right or left of the Titrando. The assembly of the support rod and the
stirrer or titration stand is described in the Instructions for Use of the
801 Stirrer and the 804 Titration Stand or the 803 Titration Stand.
Titrando Installation Instructions 15
Loading...
+ 49 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.