Metrohm 832KF User Manual

CH-9101 Herisau/Switzerland E-Mail info@metrohm.com Internet www.metrohm.com
832 KF Thermoprep
8.832.1023 10.2004/ dm
Teachware Metrohm AG Oberdorfstr. 68 CH-9101 Herisau teachware@metrohm.com
These instructions are protected by copyright. All rights reserved.
Although all the information given in these instructions has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please inform the author at the address given above.
© Metrohm Ltd. 2002 Printed in Switzerland
II 832 KF Thermoprep

Table of contents

1 Introduction 1
1.1 Description of the instrument ..........................................................1
1.2 Information about these Instructions for Use.................................2
1.3 Parts and controls ............................................................................3
1.3.1 Individual parts and standard accessories ..............................................3
1.3.2 Rear view...................................................................................................4
1.3.3 Temperature controller..............................................................................5
1.3.4 Sample holder for standard vials ..............................................................5
2 Safety information 6
2.1 General..............................................................................................6
2.2 Electrical safety ................................................................................6
3 Installation 7
3.1 Instrument setup...............................................................................8
3.1.1 Packaging .................................................................................................8
3.1.2 Checks ......................................................................................................8
3.1.3 Location.....................................................................................................8
3.2 Mains connection .............................................................................8
3.3 Assembling the 832 KF Thermoprep............................................ 10
3.3.1 Assembling the needles..........................................................................10
3.3.2 Attaching the drying bottles ....................................................................11
3.3.3 Attaching the dust filter ...........................................................................11
3.4 Assembling the coulometric cell .................................................. 12
3.5 Assembling the titration vessel for volumetric KF titrators ........ 12
4 Operation 13
4.1 Inserting and removing sample vials ........................................... 13
4.2 Needle penetration ........................................................................ 13
4.3 The pump ....................................................................................... 14
4.4 The temperature controller ........................................................... 15
4.4.1 Two set temperatures .............................................................................15
4.4.2 Setting a temperature..............................................................................15
4.4.3 Self-optimization of the control parameters............................................16
4.4.4 Key lock function.....................................................................................16
4.5 Working procedure........................................................................ 17
4.5.1 Preparing the 832 KF Thermoprep/756/831 Coulometer system ..........18
4.5.2 Determination with blank value subtraction............................................18
4.5.3 Determination without blank value subtraction.......................................19
4.6 Practical information ..................................................................... 20
5 Maintenance information 21
6 GLP validation 21
832 KF Thermoprep III
7 Troubleshooting 22
7.1 Problems ........................................................................................ 22
8 Annex 23
8.1 Technical data................................................................................ 23
8.1.1 Oven........................................................................................................23
8.1.2 Air pump..................................................................................................23
8.1.3 Temperature controller............................................................................23
8.1.4 Power supply ..........................................................................................23
8.1.5 Safety specifications ...............................................................................23
8.1.6 Electromagnetic compatibility (EMC) .....................................................24
8.1.7 Ambient temperature ..............................................................................24
8.1.8 Dimensions and materials ......................................................................24
8.2 Standard equipment ...................................................................... 25
8.2.1 832 KF Thermoprep................................................................................25
8.3 Optional accessories..................................................................... 27
8.3.1 Sample vials............................................................................................27
8.4 Warranty and conformity............................................................... 28
8.4.1 Warranty ..................................................................................................28
8.4.2 EU Declaration of Conformity for 832 KF Thermoprep ..........................29
8.4.3 Declaration of Conformity .......................................................................30
Index 31

List of illustrations

Fig. 1 Front view ...................................................................................................................3
Fig. 2 Rear view ....................................................................................................................4
Fig. 3 Temperature controller ...............................................................................................5
Fig. 4 Sample holder .......................................................................................................... 5
Fig. 5: 832 KF Thermoprep – peripheral devices ..................................................................7
Fig. 6 Fuse holder.................................................................................................................9
Fig. 7 Needle assembly ...................................................................................................... 10
Fig. 8 Guide head with needles..........................................................................................10
Fig. 9 Coulometer cell......................................................................................................... 12
Fig. 10 Volumetric KF cell ..................................................................................................... 12
Fig. 11 Crimping pliers.......................................................................................................... 13
Fig. 12 Loosening the guide head........................................................................................13
Fig. 13 Needle in lowered position ....................................................................................... 14
IV 832 KF Thermoprep
1.1 Description of the instrument

1 Introduction

Thanks to its reliability and simple handling, the Metrohm 832 KF Ther­moprep is a very versatile instrument. It has been designed for use in production and laboratories and covers a wide range of applications. The 832 KF Thermoprep can be used wherever it is necessary to heat up a solid or liquid sample and/or to drive off water by the application of heat. When combined with a coulometric or volumetric KF titrator, the 832 KF Thermoprep forms the ideal analytical system for determining the water content of samples which contain interfering components or which only release their moisture slowly. Among the decisive advantages of the KF Thermoprep is the reduction of sample preparation to a minimum. By using hermetically sealed sample vessels (headspace vials) the samples can be filled directly on site. The PTFE-coated septa guarantee a constant and true water con­tent, even when the sample have been allowed to stand for a long time.

1.1 Description of the instrument

The 832 KF Thermoprep has been designed primarily for the determina­tion of moisture by the oven method. The samples are heated in the oven block and release their moisture in the form of water vapor; this is transferred to a measuring cell by a carrier gas. The carrier gas stream is produced by a built-in air pump. An inlet valve is also provided for the use of nitrogen or other inert gases. The determination of moisture in the measuring cell can be carried out either coulometrically or volumet­rically according to the Karl Fischer method. The 832 KF Thermoprep is characterized by its compact construction; the oven heating block, air pump, temperature controller, flowmeter and flow controller as well as the drying bottles are already built in.
832 KF Thermoprep, Introduction 1
1.2 Information about these Instructions for Use

1.2 Information about these Instructions for Use

Please read through these Instructions for Use before you start to use the 832 KF Thermoprep. The following notations and pictograms are used in these instructions:
35
Operating element
The operating elements are explained on pages 3ff.
Danger This symbol indicates a possible risk of death or injury to the user and possible damage to the instrument or its compo­nents by electricity.
Danger/Warning This symbol indicates a possible risk of death or injury to the user and possible damage to the instrument or its compo­nents.
Attention This symbol indicates important informa­tion. Read the information provided before you continue.
2 832 KF Thermoprep, Introduction
1.3 Parts and controls

1.3 Parts and controls

1.3.1 Individual parts and standard accessories

1
2
3
11
5
6
8
9
10
4
7
Fig. 1 Front view
12
13
14
15
16
17
18
Guide head
1
Transfer tubing 6.1805.460
2
Spacer 4.832.4190
3
Penetration and outlet needle
4
6.2816.070 and 6.2816.080 Dosing tip 6.1543.060
5
Septum stopper 6.2730.050
6
Sample holder
7
Pump switch
8
Temperature selection switch
9
832 KF Thermoprep, Introduction 3
10
11
12
13
14
15
16
17
18
Temperature controller
Clamping lever
Guide rod
Inlet tubing 6.1805.020
Air/Nitrogen outlet
Drying bottles 6.1608.050
With 6.1602.140 inset, filled with
6.2811.000 molecular sieve
Bottle holder
Flow indicator
Flow controller
1.3 Parts and controls

1.3.2 Rear view

19
20
21
22
23
WARNING - Fire Hazard -
for conti nued pro tection replac e only with the same ty pe and rating of fuse
100 - 120 V: 4 A (T) 220 - 240 V: 2 A (T) f = 50/60 Hz S = 350 VA
Inlet filt er
Key lock
On
Off
Fig. 2 Rear view
Vorsicht: Heisse Gehäuseteile (üb er 60° C)
Attention: Hot parts (a bove 6 0°C)
Cuidado: Partes calientes (mas de 60°C)
Attention: Parties chaudes (p lus de 60°C)
Type 1.832.0010 Nr.
Made by Metrohm Switzerland
0010/01104
24
25
26
27
28
29
30
Tubing guide
19
Ventilation slots
20
Mains switch
21
Mains connection
22
Fuse holder
23
Adjustment nuts
24
25
26
27
28
29
30
Transfer tubing connection
Guide spring
Needle base
Hot air outlet
Key lock switch
Air/nitrogen outlet
Connection for dust filter (6.2724.010)
4 832 KF Thermoprep, Introduction
1.3 Parts and controls

1.3.3 Temperature controller

31
32
33
34
Actual value display
31
red, 4-place
Set value display
32
green, 4-place
Program key
33
for selecting the parameters
Decrement key
34
for changing values
Fig. 3 Temperature controller
35
36
37
38
Output status display
yellow, for outputs 1 to 3 (K1 and K3: oven heating, K2: fan)
LED for ramp/program function
green, lights if configured
EXIT key
to exit the dialog levels
Increment key
for changing values
35
36
37
38

1.3.4 Sample holder for standard vials

The dimensions of the 6.2063.000 sample holder are optimized for the Metrohm 6.2419.007 standard sample vials (6 mL). This ensures the highest possible heat flow from the oven to the sample. If other sample containers with different dimensions shall be used custom made sample holders can be delivered. For this the precise dimensions (incl. dimensional tolerances) must be known. The measures of non-standard sample containers must vary within the following dimensional limits:
Diameter 10.0 … 32.0 mm
Fig. 4 Sample holder
832 KF Thermoprep, Introduction 5
Depth of immersion 20.0 … 45.0 mm
2.1 General

2 Safety information

Warning! This instrument should only be used in accordance with the information given in these Instructions for Use.

2.1 General

This instrument left our works in perfect condition from the point of view of its operational safety (see Technical data, safety specification). To keep it in this condition and to continue to operate safely the following information must be carefully observed.

2.2 Electrical safety

Please observe the following guidelines:
Only qualified Metrohm personnel should carry out service work on electronic components.
Do not open the instrument housing as this could damage the in­strument. The housing contains no components which could be serviced or exchanged by the user.
Electrical safety when handling the instrument is guaranteed within the scope of Standard IEC 61010. However, please observe the following point:
Protection against electrostatic charges
Warning! Electronic components are sensitive to electrostatic charges and can be destroyed by a discharge. Always remove the mains connection cable from socket 12 before making or breaking electrical connections on the rear panel of the instrument.
Connection to the electricity supply:
This instrument must only be operated at the specified mains voltage.
Repair and maintenance:
If faults or malfunctions occur while using the 832 KF Thermoprep we recommended that you first check that the connection to the control in­strument has been made correctly.
The instrument must not be opened. This is reserved exclusively for authorized service personnel.
6 832 KF Thermoprep, Safety information
2.2 Electrical safety

3 Installation

This section describes the things you should pay attention to when un­packing and starting up the 832 KF Thermoprep. It also tells you how a complete automation system is built up. The following illustration provides an overview of the peripheral devices which can be used with an 832 KF Thermoprep:
PC
PC
800
800
800
800
800
800
800
800
Dosino
Dosino
Dosino
Dosino
Dosino
Dosino
Dosino
Dosino
Titriermittel
KF
Titriermittel
KF
Titriermittel
Titriermittel
KF
KF
835
835
835 Titrando
Titrando
Titrando
836
836
836 Titrando
Titrando
Titrando
Titrandos
USB
PC Control tiamo
KF
KFKF
Touch ControlTouch Control
Fig. 5: 832 KF Thermoprep – peripheral devices
LPT 1
LPT 1
USB
PC Software
COM 1
COM 1
COM 2
COM 2
USB
RS232
Tinet 2.x / VESUV 3.0
832 TP832 TP
832 Thermoprep
RS232
756 KF756 KF
831 KF831 KF
KF Coulometer 831 / 756
Titrierm ittelTitrierm ittel
Titrierm ittelTitrierm ittel
Titrierm ittel
Titrierm ittel
KFD 758
KFD 758
KFP 785
KFP 785
KFP 795KFP 795
KF Titrino 795 / 787 / 758
832 KF Thermoprep, Installation 7
3.1 Instrument setup

3.1 Instrument setup

3.1.1 Packaging

The 832 KF Thermoprep and its specially packed accessories are sup­plied in very protective special packaging. Please store this packaging in a safe place; it is the only way in which the safe transport of the in­strument can be guaranteed.

3.1.2 Checks

Please check that the delivery is complete and undamaged immedi­ately on receipt (compare with delivery note and list of accessories given in Section 8.28.2). If transport damage is evident please refer to the information given in Section 8.4.1 “Warranty”.

3.1.3 Location

The 832 KF Thermoprep has been developed for use indoors and must not be used in explosion-endangered surroundings. Place the instrument on a suitable vibration-free laboratory bench, pro­tected as much as possible from corrosive atmospheres and contami­nation by chemicals. Select a location in which the ambient temperature is normally between +5 °C and +45 °C. The instrument should be protected against exces­sive variations in temperature and direct sunshine.
If an instrument which has been stored under cold conditions is brought into a warm room then the atmospheric humidity may con­dense inside the instrument and form water. In order to avoid damaging the instrument please wait for at least one hour before switching it on.

3.2 Mains connection

Please observe the following rules when connecting the instrument to the electricity supply. If the instrument is operated with an incorrectly set mains voltage and/or an incorrect mains fuse then it represents a fire hazard!
Setting the mains voltage Before you switch on the 832 KF Thermoprep for the first time please check that the mains voltage set on the instrument (see illustration on the following page) corresponds to your local mains voltage. If this is
not the case then you must alter the mains voltage as follows:
8 832 KF Thermoprep, Installation
3.2 Mains connection
Pull out mains cable
Remove the mains cable from the mains supply connection of the 832 KF Thermoprep.
Remove the fuse holder
Use a screwdriver to loosen the fuse holder beside the mains supply connection and remove it completely.
Check the fuse and replace it
Carefully remove the built-in fuse for the intended voltage from the fuse holder and check its specifications (the position of the fuse in the fuse holder is indicated by the white arrow beside the voltage range):
100120 V 4 A (slow blow) Metrohm-No. U.600.0022 220240 V 2 A (slow blow) Metrohm No. U.600.0019
Insert fuse
Exchange the fuse if necessary and replace it in the fuse holder.
Insert fuse holder
Depending on the required mains voltage, insert the fuse holder so that the white arrow of the corresponding voltage range points to the white bar; this is to the right of the fuse holder (see below).
220 – 240 V 100 – 120 V
Mains switch
Mains connection
Fuse holder
100 - 120 V
220 - 240 V
Fig. 6 Fuse holder
220 - 240 V
100 - 120 V
832 KF Thermoprep, Installation 9
3.3 Assembling the 832 KF Thermoprep

3.3 Assembling the 832 KF Thermoprep

3.3.1 Assembling the needles

41
3
39
Screw spacer 3 (4.832.4190, L=58 mm) into distributor 39. For special requirements, e.g. different sample vials, custom-made spacers can be supplied, see p. 27.
Place PTFE distance ring 41 (supplied with the needle) on penetration needle 42 (6.2816.070) and insert this into distributor 39 from above (see left). Pull penetration needle 42 down as far as it will go.
Slide hollow outlet needle 43 over penetration needle 42 from below and fasten it to Luer-lock connector of the spacer 3.
Connect inlet tubing 13 (6.1805.020, 52 cm) to distributor 39 from above and fasten it. This fixes penetration needle 42 in the proper position.
44
13
Fig. 7 Needle assembly
43
12
42
27
Fig. 8 Guide head with needles
To prevent possible injury the height of needle base 27, which protects the tip of penetration needle 42, can be adjusted. Screw adjustment nuts 44 of guide rod 12 until the needle tip is flush with the lower side of needle base 27.
10 832 KF Thermoprep, Installation
3.3 Assembling the 832 KF Thermoprep

3.3.2 Attaching the drying bottles

1. Fill 6.1608.050 drying bottles 15 with 6.2811.000 molecular sieve and place 6.1821.040 outlet tube (with 6.2821.090 filter) in
6.1602.140 screw cap. Screw the screw caps onto the drying bottles.
14
16
When attaching the tubing please observe the following:
To filter the carrier gas before leaving the drying bottle, it must be led through the outlet tube to the outlet connection.
2. Place drying bottles 15 in holder 16 of the oven. Screw the 6.1805.180 Tubing (16 cm) to the inlet tubing connection of the front drying bottle. The carrier gas must enter the drying
15
bottle at the top of the bottle. Connect the other end of the tubing to the air/nitrogen outlet 14 (see drawing).
3. The rear drying bottle is connected to the front bottle using 6.1805.010 Tubing (13 cm). Screw the tubing onto the outlet tube of the front bottle and connect the other end to the inlet connec­tion of the rear drying bottle. The carrier gas must enter the rear drying bottle at the top of the bottle.
4. Screw 6.1805.020 Tubing (52 cm) onto the outlet tube of the rear drying bottle and lead it through the tubing guide on the guide head of the 832 KF Thermoprep. The free end of the tubing is screwed onto the connection provided for it on the guide head.
5. 6.1805.460 Transfer tubing (27 cm) is screwed onto the back of the distributor piece. Attach 6.1543.060 Dosing tip to the other end of the tubing and insert it into the titration vessel.

3.3.3 Attaching the dust filter

Attach 6.2724.010 Dust filter to the 'Inlet filter' connection on the rear panel of the instrument. If nitrogen is to be used as the carrier gas then the nitrogen inlet tubing can be attached directly to the dust filter tubing nozzle.
832 KF Thermoprep, Installation 11
3.4 Assembling the coulometric cell

3.4 Assembling the coulometric cell

Generator electrode
Fig. 9 Coulometer cell
with drying tube
Indicator electrode
6.1543.060 Tip
Screw nipple of
6.2730.030
6.1446.060 Stopper
The coulometer cell is setup for the gas inlet as follows:
6.1543.060 Tip is screwed into the 6.1446.060 Stopper by using the screw nipple (screw nipple of 6.2730.030 with O-ring) and the stopper is inserted into the SGJ 14/15 opening.
If simultaneous aspiration is necessary while gas is being led in then the gas inlet tubing must be inserted into the side inlet. This requires the use of the following parts:
Tip 6.1543.060 Screw cap 6.2701.0
40
Sealing ring A.254.0104
A titration vessel with 2 side-mounted inlets is also available: order number 6.1465.320.

3.5 Assembling the titration vessel for volumetric KF titrators

The titration vessel is adapted for the gas inlet in the following way:
Instead of the 6.2730.020 Septum stopper, the
6.2730.050 Stopper is used with 6.1543.060 Tip. (Nipple and O-ring of 6.2730.030 from the Titrino accessories.)
Fig. 10 Volumetric KF cell
12 832 KF Thermoprep, Installation
4.1 Inserting and removing sample vials

4 Operation

This section describes the basic principles of handling the 832 KF Ther­moprep.

4.1 Inserting and removing sample vials

As the oven of the 832 KF Thermoprep can reach a temperature of 250 °C, vials must only be inserted in or removed from the oven by us­ing crimping pliers or other tools. Hot vials should only be placed on temperature-resistant surfaces.
Fig. 11 Crimping pliers

4.2 Needle penetration

Always lower guide head 1 by using both hands. In this way you can avoid injuring yourself on the needle.
Fig. 12 Loosening the guide head
Clamping lever 11
832 KF Thermoprep, Operation 13
4.3 The pump
Loosen the red clamping lever by pressing it strongly and lower the guide head and needle by using both hands. Penetrate the vial septum with the needle and lower the guide head until it reaches the bottom stop.

4.3 The pump

The 832 KF Thermoprep is equipped with a powerful air pump, with which the released humidity of the sample can be transferred to a titra­tion cell in a constant carrier gas stream.
Atach a 6.2724.010 dust filter to the air inlet 30 on the instrument's rear side. Avoid the intrusion of dust particles. The dust filter must be replaced in a regular equipment service.
If thermal unstable samples, which may decompose at higher tempera­tures are treated, the use of nitrogen as carrier gas is highly recom­mended. Connect the nitrogen supply pipe to the hose olive of the dust filter. The maximum pressure of the nitrogen supply must not exceed 1 bar.
Fig. 13 Needle in lowered position
14 832 KF Thermoprep, Operation
4.4 The temperature controller

4.4 The temperature controller

The best proven dTron 16.1 (JUMO) temperature controller of the 832 KF Thermoprep is very easy to use for routine operation of the instru­ment. Fine adjustment of the control parameters is possible but not normally necessary. Unappropriate parameters of configuration settings may cause faults of the instrument. The default settings can be used without any problems. The temperature controller 10 also has a com­fortable and simple self-optimization function which can be used if nec­essary.

4.4.1 Two set temperatures

The temperature controller of the 832 KF Thermo­prep can store two set temperatures (T1 and T2). You can switch between these two set tempera­tures by using the temperature selection switch.
The oven starts to heat up as soon as the instrument is switched on and is controlled at the set temperature (T1 or T2). If a switch is made from one set temperature to the other one then this will be adopted im­mediately by the temperature controller, i.e. the oven will immediately heat up or cool down until the set temperature has been reached.

4.4.2 Setting a temperature

The desired temperature (T1 or T2) is easily set by using the two arrow keys (increment key 38) and (decrement key 34) of the tempera­ture controller. The set temperature can be altered at any time, regard­less of whether the set temperature has been reached or not. Please observe the green set value display 32. This always refers to the set temperature T1 or T2 which has been selected. Longer pressure on an arrow key changes the set value more quickly. Altered set values are immediately accepted by the temperature con­troller.
If the key lock function (tumbler switch 29 on the rear panel of the hous­ing) is switched on then no settings can be made.
Temperature display
The display indicates the real value and the set value (T1 or T2).
832 KF Thermoprep, Operation 15
4.4 The temperature controller

4.4.3 Self-optimization of the control parameters

Self-optimization should only be used when the set temperature cannot be achieved by the temperature controller or when the actual tempera­ture varies about the set temperature by more than 2…5 °C (overshoot­ing).
Preparation
Set the same conditions as would be used in a real determination.
Set the temperature (T1 or T2) to a value corresponding to a real determination (e.g. 120 °C).
Wait until the oven has reached the set temperature.
Insert a sealed sample vial containing a sample in the oven.
Lower the work head and completely penetrate the sample vial sep-
tum with the needle.
Switch on the pump (or open the nitrogen valve) and set the gas flow to a realistic value which would be used during a proper de­termination.
Self-optimization
Start the self-optimization process with EXIT (keep key 37 pressed
down for at least 2s!).
During the self-optimization process "tunE" blinks in the display 31. When the blinking stops the process is finished. Self-optimization nor­mally takes 1 to 5 minutes. The duration depends on the selected tem­perature, i.e. the higher the temperature the shorter it takes.
Save the control parameters which have been determined by press-
ing the EXIT key 37 for 2 seconds.
The self-optimization process can be canceled by pressing the EXIT key 37 briefly.

4.4.4 Key lock function

The temperature controller settings can be protected against accidental alteration. On the rear panel of the 832 KF Thermoprep the 'Key Lock' tumbler switch 29 can be switched to 'On'. If the key lock function is switched on the 'Temperature' tumbler switch 9 can still be used to select the two set temperatures T1 and T2. It is also possible to switch the pump on and off (key 8). However, no alteration can be made to the temperature controller 10 with the key lock function switched on.
16 832 KF Thermoprep, Operation
4.5 Working procedure

4.5 Working procedure

Warning! As both the sample vials and the oven become very hot the sample vials must be removed from the oven by using the crimping pliers. The sample vials should only be placed on a heat-resistant surface.
In principle two different methods can be used.
For samples with a low water content one (or more) blank values must be determined and subtracted from the water content determined for the sample. Sample vials contain a considerable amount of atmos­pheric humidity, e.g. Metrohm standard vials with a volume of 6 mL can contain 60…150 µg water. While the determination is being carried out atmospheric moisture can also enter the coulometer cell as a result of leaks, etc.
For samples with a high water content blank value determinations are not necessary as the atmospheric humidity contained in the vial is negligible when compared with the moisture contained in the sample. If no blank determinations are carried out then the whole procedure is simplified and takes less time.
Determinations with the 832 KF Thermoprep basically take part in three phases:
1. Preparing / Conditioning the system
2. Possibly determining the blank value
3. Determination of the sample
The procedure for determinations with the 756/831 Coulometer is de­scribed below. A similar procedure can be used with a volumetric KF ti­trator (e.g. 795 KFT Titrino).
Note When using either a coulometer or a volumetric KF titrator it is abso­lutely essential that an extraction time of at least 300 s is observed per determination. More details are given in the Instructions for Use of the particular instrument.
832 KF Thermoprep, Operation 17
4.5 Working procedure

4.5.1 Preparing the 832 KF Thermoprep/756/831 Coulometer system

1. Switch on the 832 KF Thermoprep. The oven heats up automati­cally to set temperature T1 or T2.
2. If necessary use the set value switch to alter set temperature T1 or T2.
3. Place an empty sealed vial in the oven.
4. Lower the needle until the stop is reached. Both hands should be used for safety reasons.
5. Switch on air pump (if N
is used open the nitrogen valve).
2
6. Check the gas flow at the flowmeter 17 and alter it if necessary (optimal range: 40…60 mL/min).
7. Start the 756/831 Coulometer and allow it to be conditioned.
8. Let the oven heat up until the set temperature is reached and the coulometer cell is conditioned.

4.5.2 Determination with blank value subtraction

Determination of blank value(s)
1. Load a method for a blank value determination (e.g. Blank) into the coulometer. The mean value of the blank value determinations should be stored as a common variable. Set the extraction time to 300 s.
2. Start conditioning the coulometer (with the <START> key).
3. Wait until the set temperature has been reached and the coulom­eter cell is conditioned.
4. Start the blank determination at the coulometer with <START>.
5. Raise the needle fully and remove the conditioning vial from the oven with the crimping pliers.
6. Insert an empty sealed vial (blank sample) in the oven using the crimping pliers.
7. Lower the needle until it reaches the stop and pierce the septum cap.
8. After the determination wait for post-conditioning to be completed and, if necessary, carry out further blank value determinations, i.e. repeat the procedure from item 3 onward.
Sample determination
1. Load a method for a sample determination into the coulometer. The mean blank value, stored as a common variable, is to be sub­tracted from the sample result. Set the extraction time to 300 s.
2. Start conditioning the coulometer (with the <START> key).
3. Wait until the set temperature has been reached and the coulom­eter cell is conditioned.
4. Start the determination at the coulometer with <START>.
18 832 KF Thermoprep, Operation
4.5 Working procedure
5. Raise the needle fully and remove the blank sample vial from the oven with the crimping pliers.
6. Insert the sealed sample vial in the oven using the crimping pliers.
7. Lower the needle until it reaches the stop. Enter the sample size at the coulometer and confirm it with <ENTER>.
8. After the determination wait for post-conditioning to be completed and, if necessary, carry out further determinations.

4.5.3 Determination without blank value subtraction

Sample determination
1. Load a method for a sample determination into the coulometer. The
result calculation must not contain any blank value subtraction. Set the extraction time to 300 s. 'Value' must be entered under '>Preselection' | 'Request sample size'.
2. Start conditioning the coulometer (with the <START> key).
3. Wait until the set temperature has been reached and the coulometer
cell is conditioned.
4. Start the determination at the coulometer with the <START> key.
5. Raise the needle fully and remove the conditioning vial from the
oven with the crimping pliers.
6. Insert the sealed sample vial) in the oven using the crimping pliers.
7. Lower the needle until it reaches the stop. Enter the sample size at
the coulometer and confirm it with <ENTER>.
8. After the determination wait for post-conditioning to be completed
and, if necessary, carry out further determinations.
832 KF Thermoprep, Operation 19
4.6 Practical information

4.6 Practical information

Selecting the carrier gas
Nitrogen (N sitive to air or oxygen (decomposition) or releases substances which in­terfere with the KF reaction.
Temperature setting
Select as high a temperature as the sample allows (high temperature = short analysis time). However, the sample must not decompose. It should not release any other oxidizable substances than water. The temperature shown on the temperature controller refers to the tempera­ture in the aluminum heating block and not to the sample temperature. Depending on the size of vial used and the temperature setting, the sample temperature could vary by up to 10%.
Gas flow
The gas flow is set on the flow controller on the front panel of the 832 KF Thermoprep and passes through the drying bottles. The gas flow should be kept as low as possible. With very moist samples in particular care must be taken that large amounts of water do not enter the titration vessel too quickly. The solution in cell must be able to absorb the driven-off moisture. A flow rate of 40…60 mL/min is normally optimal. If the needle is inserted in the sample bottle until the stop is reached then the flow of gas through the sample will be transferred to the titra­tion vessel via the outlet needle. The moisture is then titrated in the titra­tion vessel.
Extraction time
An extraction time of at least 5 min should be set on the connected ti­trator, in order to prevent the titration from being stopped before all the sample has released all its moisture.
Conditioning the system
Before a measurement the system must be conditioned with an empty sealed sample vial (conditioning vessel).
Literature
Hydranal de Haën, 1987
The following Metrohm Application Bulletins can be obtained free of charge:
should be used as the carrier gas if the hot sample is sen-
2)
®
Manual, Water reagents for Karl-Fischer-Titration, Riedel-
No. 109 Karl Fischer water determination with the KF drying oven No. 145 Determination of low water contents in plastics using the KF oven method No. 217 Karl Fischer water determination in pharmaceuticals us­ing the oven method
20 832 KF Thermoprep, Operation
4.6 Practical information

5 Maintenance information

The 832 KF Thermoprep does not need a lot of care. Ensure that it does not become excessively dirty and is not exposed to corrosive atmos­pheres.
Renew used-up molecular sieve in good time. The molecular sieve should be exchanged as soon as increased drift values occur in the Karl Fischer cell.
Rinse the tubing and the distributor occasionally and carefully dry the tubing.
Clean the oven compartment at regular intervals.
Replace the dust filter on the rear panel when necessary.

6 GLP validation

Each instrument produced by Metrohm undergoes a rigorous quality control check before delivery.
GLP (Good Laboratory Practice) requires, among other things, that the precision and correctness of analytical instruments are checked at regular intervals by using SOPs (Standard Operating Procedure).
Recommended literature
Metrohm leaflets "Quality Management with Metrohm", detailed in­formation about the principles and procedures of Good Laboratory Practice
Metrohm Application Bulletin 252/1 "Validation of Metrohm titrators according to GLP/ISO 9001"
832 KF Thermoprep, Maintenance information 21
7.1 Problems
s
s

7 Troubleshooting

7.1 Problems

Problem Possible cause Remedy
Titration results widely scattered.
Drift too high.
Titration time too long.
Molecular sieve in the drying bottles is used up.
Condensate in the outlet tubing.
Gas flow too high. Reduce gas flow
Inhomogeneous sample.
Not all the moisture has been driven out.
Leaky tubing connec­tions.
Molecular sieve in the drying bottles is used up.
Leaky titration vessel O­rings. Leaky septum cap. Replace septum cap.
Inhomogeneous sample.
Replace molecular sieve. (If the
molecular sieve in the front bottle is
till OK then this bottle can be used as the rear bottle. The "freshest“ mo­lecular sieve belongs in the front bot­tle.)
Dry the tubing. Reduce gas flow.
Possibly lower the temperature.
Reduce sample particle size before
weighing.
Set "harder" switch-off conditions at
the titrator: Lower stop drift, increased switch-off delay.
Check tubing and replace if neces-
sary.
Replace molecular sieve. (If the
molecular sieve in the front bottle is
till OK then this bottle can be used as the rear bottle. The "freshest“ mo­lecular sieve belongs in the front bot­tle.)
Replace O-rings.
Reduce sample particle size before
weighing.
22 832 KF Thermoprep, Troubleshooting
8.1 Technical data

8 Annex

In this section you will find the most important technical data of the 832 KF Thermoprep, a list of standard and optional accessories and the warranty and conformity declarations.

8.1 Technical data

8.1.1 Oven

Temperature range Correction range Heating power Heating rate
50…250 °C
-10…+10 °C 200 VA 15 °C/min

8.1.2 Air pump

Delivery rate Adjustment range
0…300 mL/min (0…18 L/h) 0…80 mL/min (0…4.8 L/h)

8.1.3 Temperature controller

Make Temperature control
Display
JUMO dTRON 16.1 Self-optimizing microprocessor controller, can be used as 2-point, 3-point or continuous controller 2x 7-segment digital display of set and actual values

8.1.4 Power supply

Voltage Frequency Power consumption
100...120 V, 220...240 V
50...60 Hz 220 VA

8.1.5 Safety specifications

Construction and testing According to IEC 61010, EN 61010, EN50093, protection
class 3
Safety information The Instructions for Use contains safety information that
must be observed by the user in order to ensure the safe operation of the instrument.
832 KF Thermoprep, Annex 23
8.1 Technical data

8.1.6 Electromagnetic compatibility (EMC)

Emission Standards fulfilled:
- EN/IEC 61326-1
- EN 55022 class B
- CISPR 22 class B
- EN/IEC 61000-3-2 class A
- EN/IEC 61000-3-3
- Namur
Immunity Standards fulfilled:
- EN/IEC 61326-1
- EN/IEC 61000-4-2
- EN/IEC 61000-4-3
- EN/IEC 61000-4-4
- EN/IEC 61000-4-5
- EN/IEC 61000-4-6
- EN/IEC 61000-4-8
- EN/IEC 61000-4-11
- Namur

8.1.7 Ambient temperature

Nominal working range
Storage Transport
+5…+45 °C (at max. 85% relative humidity)
-20 °C…+60 °C
-40 °C…+60 °C

8.1.8 Dimensions and materials

Height Width Depth Weight approx. 8.9 kg (without accessories) Housing material Metal housing, surface-refined
49.5 cm
35.5 cm
22.0 cm
24 832 KF Thermoprep, Annex
8.2 Standard equipment

8.2 Standard equipment

Immediately upon receipt of the instrument please check that the deliv­ery is complete.

8.2.1 832 KF Thermoprep

Order no. 2.832.0020
No. Order no. Description
1 1.832.0020 832 KF Thermoprep
KF oven for processing single samples in sealed vials.
The following accessories are supplied as standard:
1 6.1543.060 ETFE/FEP dosing tip
2 6.1602.145 Drying bottle insert
2 6.1608.050 Clear glass bottle GL 45, 100 mL
1 6.1805.010 FEP tubing, M6, L=13 cm
1 6.1805.020 FEP tubing, M6, L=52 cm
1 6.1805.180 FEP tubing, M6, L=16 cm
1 6.1805.460 FEP tubing, M6, L=27 cm
832 KF Thermoprep, Annex 25
8.2 Standard equipment
2 6.1821.040 Inlet tube for drying bottle, with filter,
L=92 mm
1 6.2049.030 Spacer with Luer-Lock needle
connector, L=58mm
1 6.2063.000 Sample holder for 6.2419.007
sample vials (6 mL)
1 6.2621.110 Crimping pliers
1 6.2724.010 Dust filter, 32 mm diameter
1 6.2730.050 Stopper for dosing tip, M10, with
O-ring
1 6.2739.000 Tubing nipple wrench
1 6.2811.000 Molecular sieve, 250 g
1 6.2816.070 Penetration needle
1 6.2816.080 Outlet needle
100 6.2419.007 Sample vials, 6 mL
100 6.1448.057 Septum seals for sample vial 6.2419.007
1 8.800.1023 Instructions for Use for 832 KF Thermoprep
26 832 KF Thermoprep, Annex
8.3 Optional accessories

8.3 Optional accessories

8.3.1 Sample vials

Order no. Description
6.2419.000 Sample vials , 6 mL, 1000 ea.
6.1448.050 Septum seals for sample vial, 1000 ea.
In order to use other sample vials with different dimensions the sample holder of the 832 KF Thermoprep must be replaced. Please contact your Metrohm distributor.
For further accessories please see Metrohm Accessories Catalog.
832 KF Thermoprep, Annex 27
8.4 Warranty and conformity

8.4 Warranty and conformity

8.4.1 Warranty

The warranty on our products is limited to defects that are traceable to material, construction or manufacturing error which occur within 12 months from the day of delivery. In this case, the defects will be recti­fied in our workshops free of charge. Transport costs are to be paid by the customer.
For day and night operation, the warranty is limited to 6 months.
Glass breakage in the case of electrodes or other parts is not covered by the warranty. Checks which are not a result of material or manufac­turing faults are also charged during the warranty period. For parts of outside manufacture insofar as these constitute an appreciable part of our instrument, the warranty stipulations of the manufacturer in question apply.
With the regard to the guarantee of accuracy, the technical specifica­tions in the instruction manual are authoritative.
Concerning defects in material, construction or design as well as the absence of guaranteed features, the orderer has no rights or claims ex­cept those mentioned above.
If damage of the packaging is evident on receipt of a consignment or if the goods show signs of transport damage after unpacking, the carrier must be informed immediately and a written damage report demanded. lack of an official damage report releases Metrohm from any liability to pay compensation.
If any instruments and parts have to be returned, the original packaging should be used if at all possible. This applies above all to instruments, electrodes, burette cylinders and PTFE pistons. Before embedment in wood shavings or similar material, the parts must be packed in a dust­proof package (for instruments, use of a plastic bag is imperative). If open assemblies are enclosed in the scope of delivery that are sensi­tive to electromagnetic voltages (e.g. data interfaces etc.) these must be returned in the associated original protective packaging (e.g. con­ductive protective bag). (Exception: assemblies with built-in voltage source belong in a non-conductive protective packaging).
No warranty responsibility whatsoever will be accepted by Metrohm for damage which arises as a result of non-compliance with these instruc­tions.
28 832 KF Thermoprep, Annex
8.4 Warranty and conformity

8.4.2 EU Declaration of Conformity for 832 KF Thermoprep

EU Declaration of Conformity

The Metrohm Ltd. company, Herisau, Switzerland hereby certifies, that the instrument:
832 KF Thermoprep
meets the requirements of EU Directives 89/336/EEC and 73/23/EEC.
Source of specifications:
EN 61326-1 Electromagnetic compatibility, basic specification Emitted In-
terference, Interference Immunity
EN 61010-1 Safety requirements for electrical laboratory measurement
and control equipment
Description of the instrument:
Instrument with sample heating block, temperature controller and air pump for processing samples according to the Karl Fischer oven method.
Herisau, May 30, 2002
Dr. J. Frank Ch. Buchmann
Development Manager Production and Quality Assurance Manager
832 KF Thermoprep, Annex 29
8.4 Warranty and conformity

8.4.3 Declaration of Conformity

Certificate of Conformity and System Validation
This is to certify the conformity to the standard specifications for electrical appliances and accessories, as well as to the standard specifications for security and to system validation issued by the manufacturing company.
Name of commodity: 832 KF Thermoprep Manufacturer: Metrohm Ltd., Herisau, Switzerland
Technical specifications: Distribution voltage: 100…120, 220…240 V Frequency: 50…60 Hz
The instrument was manufactured and tested according to the following standards:
Electromagnetic compatibility:
Emission IEC 61326-1, EN 55022 class B, CISPR 22 class B EN/IEC 61000-3-2 class A, EN/IEC 61000-3-3
Immunity EN/IEC 61326-1, EN/IEC 61000-4-2, EN/IEC 61000-4-3, EN/IEC 61000-4-4, EN/IEC 61000-4-5, EN/IEC 61000-4-6, EN/IEC 61000-4-8, EN/IEC 61000-4-11, EN/IEC 61000-4-14
Security specifications
EN/IEC 61010-1, UL 3101-1
It has also been certified by the Swiss Electrotechnical Association (SEV), which is member of the International Certification Body (CB/IEC).
The technical specifications are documented in the instruction manual.
Metrohm Ltd. is holder of the SQS-certificate of the quality system ISO 9001 for quality assurance in design/development, production, installation and servicing.
Herisau, May 30, 2002
Dr. J. Frank Ch. Buchmann
Development Manager Production and Quality Assurance Manager
30 832 KF Thermoprep, Annex
Index

Index

<ENTER> ............................. 19
<START>........................18, 19
A
Accessories ............................26
Actual value display ..................5
Adjustment nuts..................4, 10
Air inlet ....................................14
Air outlet................................3, 4
Air pump .................................14
Ambient temperature ..........8, 24
Assembling .............................10
Atmospheric humidity...............8
B
Blank value..............................18
Bottle holder..............................3
C
Care ........................................21
Carrier gas ........................14, 20
Certificate................................30
Clamping lever....................3, 14
Cleaning..................................21
Condensation .........................22
Conformity ..............................29
Connection ...............................6
Corrosive atmospheres ..........21
Coulometric cell ......................12
Crimping pliers..................13, 17
D
Danger ......................................2
Declaration of conformity........29
Decomposition .......................20
Decrement key....................5, 15
Delivery rate ............................23
Depth of immersion ..................5
Diameter ...................................5
Dimensions...............................5
Dirty......................................... 21
Display ....................................23
Distance ring........................... 10
Distributor ...............................10
Dosing tip..................................3
Drift too high ...........................22
Drying bottle .............................3
Drying bottles..........................11
dTron 16.1 (JUMO).................15
Dust filter.................4, 11, 14, 21
E
Electrical safety.........................6
Electricity supply...................6, 8
Electromagnetic compatibility ....
.....................................24, 30
Electrostatic charges ................6
Emission ...........................24, 30
EU Declaration........................ 29
EXIT key ....................................5
Extraction time ..................17, 20
F
Faults ........................................6
Flow controller ..........................3
Flow indicator............................3
Flowmeter ...............................18
Fuse ..........................................9
Fuse holder...........................4, 9
G
Gas flow..................................20
GLP .........................................21
Good Laboratory Practice ......21
Guide head ...................3, 10, 13
Guide rod..................................3
Guide spring .............................4
H
Handling .................................13
Heating rate ............................23
High water content..................17
Hollow needle .........................10
Hot air outlet .............................4
I
Immunity ...........................24, 30
Increment key .....................5, 15
Information................................2
Inlet filter..................................11
Inlet tubing ..........................3, 10
Inserting sample vials .............13
Installation............................. 7, 8
J
JUMO...................................... 15
K
Karl Fischer.............................20
Key lock function ....................16
Key lock switch .........................4
L
Leaking ...................................22
LED ...........................................5
Literature .................................20
Location ....................................8
Low water content...................17
Luer-lock adapter....................10
M
Mains connection .................4, 8
Mains switch .............................4
Maintenance ...........................21
Malfunctions .............................6
Materials .................................24
Maximum pressure .................14
Molecular sieve.................11, 21
N
Needle assembly ....................10
Needle base..............................4
Needle penetration .................13
Needles...................................10
Nitrogen ............................14, 20
Nitrogen outlet ......................3, 4
Notations...................................2
O
Operation ................................13
Outlet needle ......................3, 10
Output status display................5
Oven .........................................3
Overshooting ..........................16
P
Packaging.................................8
Parts.......................................... 3
Penetration needle..............3, 10
Peripheral devices ....................7
Pictograms................................2
Power consumption................23
Power supply ..........................23
Practical information ...............20
Preparing ................................18
Pressure..................................14
Problems.................................22
Procedure ...............................17
Program key..............................5
Pump switch .............................3
Q
Quality control......................... 21
R
Rear view ..................................4
Relative humidity.....................24
Removing sample vials...........13
Repair........................................6
S
Safety information...............6, 23
Sample containers....................5
Sample determination.............19
Sample holder.......................3, 5
Sample vials......................26, 27
Sealing ring.............................12
Security specifications ............30
Self-optimization .....................16
Septum seals ....................26, 27
Septum stopper ........................3
Service-personnel .....................6
Set temperatures ....................15
Set value display................. 5, 15
Setup.........................................8
832 KF Thermoprep, Index 31
Index
SOP ........................................21
Spacer ....................................10
Standard accessories............... 3
Standard equipment...............25
Standards ............................... 24
System conditioning ...............20
T
T1, T2......................................15
Technical data ........................23
Temperature controller ...3, 5, 15
Display/Control parts................5
Temperature display...............15
Temperature selection switch.....
.......................................3, 15
Temperature setting ...............20
Titration vessel........................12
Tolerances ................................ 5
Transfer tubing..........................3
Transfer tubing connection....... 4
Transport damage..................28
Troubleshooting...................... 22
Tubing guide............................. 4
V
Validation ................................ 21
Ventilation slots......................... 4
Voltage.................................. 8, 9
Volumetric KF cell...................12
W
Warning.....................................2
Warranty.................................. 28
Water content .........................17
32 832 KF Thermoprep, Index
Loading...