CH-9101 Herisau/Switzerland
E-Mail info@metrohm.com
Internet www.metrohm.com
Instuctions for Use
8.830.1003
CH-9101 Herisau/Switzerland
E-Mail info@metrohm.com
Internet www.metrohm.com
830 IC Interface
5.830.0020 Program
Instructions for Use
8.830.1003 07.2003 / pkl
Teachware
Metrohm AG
Oberdorfstrasse 68
CH-9101 Herisau
teachware@metrohm.com
st
Edition 2003
1
These instructions are protected by copyright. All rights reserved.
Although all the information given in these instructions has been checked with great care, errors
cannot be entirely excluded. Should you notice any mistakes please inform the author at the
address given above.
Fig. 11: Connection of 819, 820, 818 and Triathlon ............................................. 18
Fig. 12: Connection of 819, 820, 818 and 766 ..................................................... 19
Fig. 13: Connection of 819, 820, 818 and 791 ..................................................... 20
Fig. 14: Connection of 819, 820, 818 and 761 ..................................................... 21
Fig. 15: Connection of 817, 812 and 818 (MIC 8) ................................................ 22
Fig. 16: Connection to 819, 820, 818 and 816 ..................................................... 23
Fig. 17: Connection to 819, 820, 2x818 and 828 (MIC 10)................................... 24
Fig. 18: Connection to 819, 820, 818 and 837 ..................................................... 25
830 IC Interface / 8.830.1003 Instructions for Use
II
1.1 Instrument description
1 Introduction
1.1 Instrument description
The 830IC Interface provides the connection between the PC and external IC or HPLC peripheral instruments. Up to 16 instruments including 4 detectors can be connected to the 830 IC Interface and controlled
by means of the «IC Net» PC software. The 830 IC Interface can also
record and convert the analog signals from a maximum of 4 channels
(two at once) which are processed at very high resolution.
Fig. 1 shows an overview of all the Metrohm instruments which can be
connected to the 830 IC Interface and are described in these ‘Instructions for Use’. It is also possible to connect Bischoff instruments;
please refer to the «IC Net» on-line program help as well as the relevant
Bischoff instruction manuals.
761 Compact IC
766/788 IC Sample Processor
817 Bioscan
791 IC VA Detector
Triathlon Autosampler
819 IC Detector
820 IC Separation Center
818 IC Pump
833 IC Liquid Handling Unit
830 IC Interface
812 IC Valve Unit
816 IC Eluent Selector
828 IC Dual Suppressor
837 IC Inline Degasser
Fig. 1: Connection possibilities at 830 IC Interface
830 IC Interface / 8.830.1003 Instructions for Use
1
1 Introduction
1.2 Parts and controls
In this section you will find the numbers and designations of the parts
and controls of the 830 IC Interface. The numbering applies throughout the instructions for use, i.e. bold numbers in the text (e.g.
to the parts and controls illustrated here.
1
5
) refer
Fig. 2: Front of 830 IC Interface
Mains pilot lamp
1
Lights up when instrument is switched
on
2
830 IC Interface / 8.830.1003 Instructions for Use
1.2Parts and controls
2 3 4 5
Fig. 3: Rear of 830 IC Interface
Earthing socket
2
Mains switch
3
For switching the instrument on/off.
Serial number
4
6
7 8 9 10 11 12 13 14
Remote input/output lines connec-
9
tion (COM)
RS232 interfaces
10
System 2
Mains connection plug
5
Mains connection see section 2.2
PC connection
6
RS232 interface
System 1
Analog signal connection
7
Remote input/output lines connec-
8
tion (RUN)
Analog signal connection
11
Remote input/output lines connec-
12
tion (RUN)
Remote input/output lines connec-
13
tion (COM)
RS232 interfaces
14
830 IC Interface / 8.830.1003 Instructions for Use
3
1 Introduction
1.3 Information on the Instructions for Use
Please read through these Instructions for Use carefully before you put
the 830 IC Interface into operation. The Instructions for Use contain
information and warnings to which the user must pay attention in order
to assure safe operation of the instrument.
1.3.1 Organization
These 8.830.1003Instructions for Use for the 830 IC Interface provide a comprehensive overview of the installation, startup procedure,
operation and technical specifications of this instrument. The Instructions for Use are organized as follows:
Section 1 Introduction
Section 2 Installation
General description of instrument, parts and controls and safety notes
Mains connection, connection to PC,
connection of external instruments
Section 3 Operation
Operation via «IC Net»
Section 4 Appendix
Technical data, standard equipment, options, warranty, declarations of conformity, index
To find the required information on the instruments you will find it an
advantage to use either the Table of contents or the Index at the
back.
4
830 IC Interface / 8.830.1003 Instructions for Use
1.3Information on the Instructions for Use
1.3.2 Notation and pictograms
The following notations and pictograms (symbols) are used in these Instructions for Use:
Range Menu item, parameter or entry
value
in «IC Net» program
SYSTEM STATE Program window
in «IC Net» program
<OK> Button
in «IC Net» program
4 Part or control of 830
Danger/Warning
This symbol indicates a possible
risk of death or injury to the user
and possible damage to the
instrument or its components by
electricity.
Danger/Warning
This symbol indicates a possible
risk of death or injury to the user
and possible damage to the
instrument or its components.
Attention
This symbol indicates important
information that you should read
before continuing.
Information
This symbol indicates additional
information and tips which may be
of particular use to you.
830 IC Interface / 8.830.1003 Instructions for Use
5
1 Introduction
1.4 Safety notes
While electrical safety in the handling of the 830 IC Interface is assured
in the context of the specifications IEC 1010-1 (protection class 1, degree of protection IP40), the following points should be noted:
• Mains connection
The mains connection must be effected in accordance with the
instructions in section 2.2.
• Opening the instrument
Inside the instrument there are no parts which must be set or adjusted
by the user.
If the 830 IC Interface is connected to the power supply, the instrument must not be opened nor must parts be removed from it, otherwise there is a danger of coming into contact with components which
are live. Hence, always disconnect the instrument from all voltage
sources before you open it and ensure that the mains cable is
5
disconnected from mains connection
!
• Protection against static charges
Electronic components are sensitive to static charging and can be
destroyed by discharges. Before you touch any of the components
inside the 830 IC Interface, you should earth yourself and any tools
you are using by touching an earthed object (e.g. housing of the
instrument or a radiator) to eliminate any static charges which exist.
6
830 IC Interface / 8.830.1003 Instructions for Use
2.1Setting up the instrument
2 Installation
2.1 Setting up the instrument
2.1.1 Packaging
The 830 IC Interface is supplied together with the separately packed
accessories in special packagings containing shock-absorbing foam
linings designed to provide excellent protection. The instrument itself is
packed in an evacuated polyethylene bag to prevent the ingress of
dust. Please store all these special packagings as only they assure
transport of the instrument free from damage.
2.1.2 Check
After receipt, immediately check whether the shipment is complete and
has arrived without damage (compare with delivery note and list of
accessories in section 4.2). In the case of transport damage, see
instructions in section 4.5.1 "Warranty".
2.1.3 Location
Position the instrument in the laboratory at a location convenient for operation, free from vibrations and protected against a corrosive atmosphere and contamination by chemicals.
2.1.4 Arrangement of the instruments
The modular IC instruments can be piled up in any order. The best
place for the 830 IC Interface is on top of a modular IC system.
The 830 IC Interface should always be placed above components
carrying liquids so that any leaks which may occur in the tubing or
connections cannot cause damage by leakage of liquids (e.g. acid).
830 IC Interface / 8.830.1003 Instructions for Use
7
2 Installation
2.2 Mains connection
2.2.1 Mains voltage and fuses
The 830 Interface has a power supply which automatically adjusts itself
to the existing mains voltage (100…240 V) and frequency (50…60 Hz).
It is equipped with an electronic overload protection device and also
has two fuses; however, these should only be exchanged by Metrohm
service technicians.
2.2.2 Mains cable
The instrument is supplied with one of three mains cables
• 6.2122.020 with plug SEV 12 (Switzerland, …)
• 6.2122.040 with plug CEE(7), VII (Germany, …)
• 6.2133.070 with plug NEMA 5-15 (USA, …)
which are three-cored and fitted with a plug with an earthing pin. If a different plug has to be fitted, the yellow/green lead (IEC standard) must
be connected to the earthing socket 2 (protection class 1).
Any break in the earthing inside or outside the instrument can make it
a hazard!
Plug the mains cable into mains connection plug 5 at the 830 IC Interface (see Fig. 3).
2.2.3 Switching the instrument on/off
The 830 IC Interface is switched on and off using mains switch 3 (see
Fig. 3). When the instrument is switched on a start-up check routine is
running the mains pilot lamp 1 flashes twice slowly and remains on.
8
830 IC Interface / 8.830.1003 Instructions for Use
2.3Connection to the PC
2.3 Connection to the PC
2.3.1 Connecting cable
Always switch off 830 IC Interface and PC before you connect the two
instruments with the 6.2134.100 Cable
Connect the PC connection 6 at the 830 IC Interface to one of the serial
COM ports at the PC using the 6.2134.100 Cable (9 pin/9 pin). If only a
25-pin COM interface is available on the PC then the 6.2125.110
Adapter cable or a commercially available adapter must be used.
6.2134.100
830
Fig. 4: Connection of 830 IC Interface to PC
2.3.2 Software installation
The PC program «IC Net 2.3» is required for the operation of the 830 IC
Interface; this is contained on the 6.6034.033 CD included in the accessories. This program runs under Windows 2000 and Windows XP operating systems and is installed according to section 1.4.2 of the «IC Net»
Instructions for Use.
830 IC Interface / 8.830.1003 Instructions for Use
9
2 Installation
2.4 Connection of external instruments
2.4.1 General information
Before an external instrument is connected to the 830 IC Interface, the
3
830 IC Interface must always be switched off using mains switch
Each system of the two versions of the 830 IC Interface has four RS232
interfaces 10 or 14 for connection of a maximum of 8 external instruments, an 7 or 11 interface for analog signals from a maximum of 2 detectors and 8 remote output lines 8/9 or 12/13 for controlling external
instruments by making contacts or impulses (see Fig. 3). Information
about the technical data of these interfaces is given in section 4.1.
For installation and startup of external instruments proceed as follows:
1 Switch off all instruments
!
• Switch off 830 IC Interface and all external instruments using
the mains switch.
2 Connect instruments
• Connect the instruments to the 830 IC Interface according to
the connection diagrams shown in sections 2.4.2 to 2.4.15 by
using the cables named in the diagrams. Other instrument
combinations can be set up by using these examples as a
guide.
3 Switch on all instruments
• Switch on 830 IC Interface and all external instruments using
the mains switch.
4 Instrument settings
Operation with the 830 IC Interface requires that the following
settings must be made:
• Advanced IC instruments:
818 IC Pump
819 IC Detector
820 IC Separation Center
833 IC Liquid Handling Unit
These instruments are already pre-configured, no additional
settings must be made.
Also all other Metrohm IC instruments can be operated with the
830 IC Interface, for the settings of the instruments please refer
to the respective instructions for use.
• 732 IC Detector:
>CONFIG/RS settings
handshake: SWchar
10
830 IC Interface / 8.830.1003 Instructions for Use
Loading...
+ 35 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.