Metrohm 823 User Manual

823 Membrane Pump Unit
Manual
8.823.8001EN
Metrohm AG CH-9100 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com
823 Membrane Pump Unit
8.823.8001EN
Manual
06.2012 fpe
Teachware Metrohm AG CH-9100 Herisau teachware@metrohm.com
This documentation is protected by copyright. All rights reserved.
Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above.
Documentation in additional languages can be found on
http://documents.metrohm.com.
■■■■■■■■■■■■■■■■■■■■■■

Table of contents

1 Introduction 1
1.1 Instrument description ......................................................... 1
1.1.1 Model versions ........................................................................ 1
1.1.2 Connectors .............................................................................. 2
1.1.3 Intended use ........................................................................... 2
1.2 About the documentation ................................................... 2
1.2.1 Symbols and conventions ........................................................ 2
1.3 Safety instructions ................................................................ 3
1.3.1 General notes on safety ........................................................... 3
1.3.2 Electrical safety ........................................................................ 3
1.3.3 Tubing and capillary connections ............................................. 4
1.3.4 Flammable solvents and chemicals ........................................... 5
1.3.5 Recycling and disposal ............................................................. 5
2 Overview of the instrument 6
Table of contents
2.1 Mode of operation of the membrane pump ...................... 6
3 Installation 8
3.1 Setting up the instrument .................................................... 8
3.1.1 Packaging ................................................................................ 8
3.1.2 Checks .................................................................................... 8
3.1.3 Location .................................................................................. 8
3.2 Installing the inlet tubing and outlet tubing ...................... 8
3.3 Connecting the membrane pump ..................................... 10
3.3.1 Connection to a Sample Processor ......................................... 10
3.3.2 Connection to the 731 Relay Box .......................................... 11
4 Operation 13
5 Operation and maintenance 14
5.1 General notes ...................................................................... 14
5.1.1 Care ...................................................................................... 14
5.1.2 Maintenance by Metrohm Service .......................................... 14
5.2 Tubings ................................................................................ 15
5.3 Quality Management and validation with Metrohm ....... 15
6 Troubleshooting 16
6.1 Problems ............................................................................. 16
■■■■■■■■
III
Table of contents
■■■■■■■■■■■■■■■■■■■■■■
7 Technical specifications 17
7.1 Membrane pump ................................................................ 17
7.2 Power supply ...................................................................... 17
7.3 Interfaces and connectors ................................................. 17
7.4 Safety specifications ........................................................... 17
7.5 Electromagnetic compatibility (EMC) ................................ 18
7.6 Ambient temperature ......................................................... 18
7.7 Reference conditions .......................................................... 18
7.8 Dimensions/material ........................................................... 19
8 Conformity and warranty 20
8.1 Quality Management Principles ........................................ 20
8.2 Warranty (Guarantee) ........................................................ 21
9 Accessories 23
9.1 Scope of delivery per model version ................................. 23
9.1.1 823 Membrane Pump Unit .................................................... 23
9.1.2 823 Membrane Pump Unit with aspiration equipment ........... 24
9.1.3 823 Membrane Pump Unit with rinsing equipment ................ 26
9.2 Optional accessories ........................................................... 28
9.2.1 823 Membrane Pump Unit .................................................... 28
9.2.2 823 Membrane Pump Unit with aspiration equipment ........... 28
9.2.3 823 Membrane Pump Unit with rinsing equipment ................ 29
Index 30
■■■■■■■■
IV
■■■■■■■■■■■■■■■■■■■■■■

Table of figures

Figure 1 Front 823 Membrane Pump Unit ....................................................... 6
Figure 2 Mode of operation of the membrane pump ...................................... 6
Figure 3 Connecting the tubings ..................................................................... 9
Figure 4 Screwing the union nuts tightly ......................................................... 9
Figure 5 823 Membrane Pump Unit connected directly to the Sample Pro-
cessor ............................................................................................. 10
Figure 6 823 Membrane Pump Unit via the Remote Box and 731 Relay Box .. 11
Table of figures
■■■■■■■■
V
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction

1.1 Instrument description

The 823 Membrane Pump Unit is a versatile membrane pump. It was designed specifically for the production operation and laboratory environ­ment and enables a wide range of applications. Particularly when used in a Metrohm automation system with Sample Processor and various analysis instruments, it is of indisputable service in processing large sample series in the complete titration area and in various measuring applications or other analytical tasks.
The 823 Membrane Pump Unit uses the membrane pump principle (see Chapter 2.1, page 6) and is suitable for pumping all types of liquid media.
Note

1 Introduction

The 823 Membrane Pump Unit is not suitable for pumping liquids with solids content.
The flow rate of at least 450 mL/min opens up a wide range of applica­tions.

1.1.1 Model versions

The 823 Membrane Pump Unit is available in the following three versions:
2.823.0010
2.823.0020 823 Membrane Pump
2.823.0030 823 Membrane Pump
Each version includes different accessories according to its use (see Chap- ter 9, page 23).
823 Membrane Pump Unit
Unit
Unit
with aspiration equip­ment
with rinsing equipment
■■■■■■■■
1

1.2 About the documentation

1.1.2 Connectors

The 823 Membrane Pump Unit needs 16 - 24 V DC voltage.
Power is supplied to the pump via the direct connection to a Sample Pro­cessor (e.g. 814 or 815) or to the 731 Relay Box (connection via adapter cable 6.2160.010).

1.1.3 Intended use

The 823 Membrane Pump Unit is designed for usage in automated sys­tems in analytical laboratories.
This instrument is suitable for pumping chemicals and flammable samples. Use of the 823 Membrane Pump Unit therefore requires the user to have basic knowledge and experience in handling toxic and caustic substances. Knowledge with respect to the application of the fire prevention measures prescribed for laboratories is also mandatory.
1.2 About the documentation
■■■■■■■■■■■■■■■■■■■■■■
Caution
Please read through this documentation carefully before putting the instrument into operation. The documentation contains information and warnings which the user must follow in order to ensure safe opera­tion of the instrument.

1.2.1 Symbols and conventions

The following symbols and formatting may appear in this documentation:
Method Dialog text, parameter in the software
File New Menu or menu item
Cross-reference to figure legend
The first number refers to the figure number, the sec­ond to the instrument part in the figure.
Instruction step
Carry out these steps in the sequence shown.
■■■■■■■■
2
[Next] Button or key
Warning
This symbol draws attention to a possible life hazard or risk of injury.
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction
Warning
This symbol draws attention to a possible hazard due to electrical current.
Warning
This symbol draws attention to a possible hazard due to heat or hot instrument parts.
Warning
This symbol draws attention to a possible biological hazard.
Caution
This symbol draws attention to a possible damage of instruments or instrument parts.
Note
This symbol marks additional information and tips.

1.3 Safety instructions

1.3.1 General notes on safety

Warning
This instrument may only be operated in accordance with the specifica­tions in this documentation.
This instrument has left the factory in a flawless state in terms of technical safety. To maintain this state and ensure non-hazardous operation of the instrument, the following instructions must be observed carefully.

1.3.2 Electrical safety

The electrical safety when working with the instrument is ensured as part of the international standard IEC 61010.
Warning
Only personnel qualified by Metrohm are authorized to carry out service work on electronic components.
■■■■■■■■
3
1.3 Safety instructions
■■■■■■■■■■■■■■■■■■■■■■
Warning
Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are touched.
There are no parts inside the housing which can be serviced or replaced by the user.
Mains voltage
Warning
An incorrect mains voltage can damage the instrument.
Only operate this instrument with a mains voltage specified for it (see rear panel of the instrument).
Protection against electrostatic charges
Warning
Electronic components are sensitive to electrostatic charges and can be destroyed by discharges.
Do not fail to pull the mains cable out of the mains connection socket before you set up or disconnect electrical plug connections at the rear of the instrument.

1.3.3 Tubing and capillary connections

Caution
Leaks in tubing and capillary connections are a safety risk. Tighten all connections well by hand. Avoid applying excessive force to tubing connections. Damaged tubing ends lead to leakage. Appropriate tools can be used to loosen connections.
Check the connections regularly for leakage. If the instrument is used mainly in unattended operation, then weekly inspections are manda­tory.
■■■■■■■■
4
■■■■■■■■■■■■■■■■■■■■■■

1.3.4 Flammable solvents and chemicals

Warning
All relevant safety measures are to be observed when working with flammable solvents and chemicals.
Set up the instrument in a well-ventilated location (e.g. fume cup-
board).
Keep all sources of flame far from the workplace. Clean up spilled liquids and solids immediately. Follow the safety instructions of the chemical manufacturer.

1.3.5 Recycling and disposal

This product is covered by European Directive 2002/96/EC, WEEE – Waste from Electrical and Electronic Equipment.
The correct disposal of your old equipment will help to prevent negative effects on the environment and public health.
1 Introduction
More details about the disposal of your old equipment can be obtained from your local authorities, from waste disposal companies or from your local dealer.
■■■■■■■■
5

2.1 Mode of operation of the membrane pump

1 2
3
4
5
6

2 Overview of the instrument

Figure 1 Front 823 Membrane Pump Unit
■■■■■■■■■■■■■■■■■■■■■■
Status LED
1
For displaying the operating state.
Membrane pump
3
With arrows to indicate the flow direction.
Outlet connector
5
With union nut for the drainage tubing.
Inlet connector
2
With union nut for the aspiration tubing.
Pressurizing valve
4
Connection cable
6
With M8 plug for connection to a Sample Processor.
2.1 Mode of operation of the membrane pump
Outlet valve
1
■■■■■■■■
6
Figure 2
Mode of operation of the membrane pump
Inlet valve
2
■■■■■■■■■■■■■■■■■■■■■■
2 Overview of the instrument
Pump chamber
3
Eccentric
5
PTFE membrane
4
Principle
The membrane pump is located in the pump head. The membrane made of PTFE 4 is moved up and down by an eccentric 5. This results in a peri­odic volume change of the pump chamber 3. On the downstroke, the medium to be pumped is therefore aspirated through the inlet valve 2. On the upstroke, the membrane forces the medium out of the pump chamber through the outlet valve 1.
The following features characterize the membrane pump:
Self-aspirating, due to the ingenious membrane technology. Dry run safe, i.e., the pump is not damaged if gases are pumped
instead of liquids.
Operational in any position. Small-sized and thus space-saving. High chemical resistance:
As chemically resistant materials such as PTFE, PP and FFPM are used in all parts that come into contact with the media, nearly all neutral and aggressive media may be pumped.
Note
If you use particularly aggressive media, make sure that the above-men­tioned materials are resistant to them prior to using the instrument.
In addition, the membrane pump has a very long lifetime. The only expendable parts are the membrane and the valve plates.
The dosing accuracy of the membrane pump is considerably further increased by a valve at the pump outlet. This valve has the following three functions:
Pressurizing function
When media are pumped into systems with fluctuating pressures, into a vacuum or from an overpressure area, the pressurizing valve ensures a constant backpressure.
Overflow function
The valve prevents an excessive rise in pressure in the system on the pressure side.
Anti-injector function
When the pump is used for dosing into piping systems with high flow rates, the pressurizing valve prevents an injector effect (aspiration of ambient air as a result of the underpressure).
■■■■■■■■
7

3.1 Setting up the instrument

3 Installation

3.1 Setting up the instrument

3.1.1 Packaging

The instrument is supplied in highly protective special packaging together with the separately packed accessories. Keep this packaging, as only this ensures safe transportation of the instrument.

3.1.2 Checks

Immediately after receipt, check whether the shipment has arrived com­plete and without damage by comparing it with the delivery note.

3.1.3 Location

The instrument has been developed for operation indoors and may not be used in explosive environments.
■■■■■■■■■■■■■■■■■■■■■■
Place the instrument in a location of the laboratory which is suitable for operation, free of vibrations, protected from corrosive atmosphere, and contamination by chemicals.
The instrument should be protected against excessive temperature fluctu­ations and direct sunlight.

3.2 Installing the inlet tubing and outlet tubing

As the inlet tubing and outlet tubing, two different tubing types are avail­able. Both tubings are included in the scope of delivery:
PVC tubing (6.1801.120), 2 x 2 m PTFE tubing (6.1812.000), 1 x 4 m
Note
Which of the two tubings to use depends largely on the solvent used. PVC tubing is more flexible and has the advantage of making less noise during the pumping operation. It must be noted that the PVC tubing is not resistant to solvents.
■■■■■■■■
8
Install the inlet tubing and outlet tubing as follows:
Cut the inlet tubing and outlet tubing to the suitable length.
1
■■■■■■■■■■■■■■■■■■■■■■
Connect the tubings to the tubing olives of the pump tubing using
2
union nuts.
Figure 3 Connecting the tubings
Note
To fasten PTFE tubings in place, you have to widen their openings (e.g. using a Phillips screwdriver). You can use a piece of sandpa­per to make the PTFE tubing easier to handle.
3 Installation
Screw the union nuts tightly to ensure a secure tubing connection.
3
Figure 4 Screwing the union nuts tightly
■■■■■■■■
9

3.3 Connecting the membrane pump

1 2
3.3 Connecting the membrane pump
Warning
Follow the instructions listed below for connecting to a control device. The instrument can be damaged if operated with the wrong supply voltage.
Voltage
The 823 Membrane Pump Unit needs 16 - 24 V DC voltage. It can be con­nected directly to Sample Processor models 814, 815, 855 or 864.
When using the 823 Membrane Pump Unit in an automation system with remote control, we recommend using a 731 Relay Box for the power sup­ply.

3.3.1 Connection to a Sample Processor

Plug the M8 plug into the socket on the Sample Processor provided for
this purpose.
■■■■■■■■■■■■■■■■■■■■■■
Sample Processor 814, 815, 855, 864
1
Instruments with Pump Connector.
■■■■■■■■
10
Figure 5
823 Membrane Pump Unit connected directly to the Sample Processor
823 Membrane Pump Unit
2
Membrane pump with M8 plug.
■■■■■■■■■■■■■■■■■■■■■■
1 2
3
4 5
6
Note
The 823 Membrane Pump Unit is not suitable for use in conjunction with the 824 Easy Sample Changer processor.

3.3.2 Connection to the 731 Relay Box

Connect the 823 Membrane Pump Unit with connection via the Remote Box (6.2148.010) and the 731 Relay Box to a remote interface.
Connect the M8 plug of the 772 Pump Unit to the corresponding
socket on the adapter cable.
Plug the adapter cable into low-voltage DC current output DC1 or DC2
of the 731 Relay Box.
3 Installation
Figure 6
Sample Processor 814, 815, 855, 864
1
Connection to remote interface.
Remote cable (6.2125.100)
3
Adapter cable (6.2160.010)
5
823 Membrane Pump Unit via the Remote Box and 731 Relay Box
Remote Box (6.2148.010)
2
731 Relay Box (2.731.0010)
4
823 Membrane Pump Unit
6
■■■■■■■■
11
3.3 Connecting the membrane pump
■■■■■■■■■■■■■■■■■■■■■■
Note
Ensure the correct polarity of the banana plugs. The red plug (positive pole) belongs in the red socket, the black plug (negative pole) belongs in the black socket.
Note
For information on controlling the 823 Membrane Pump Unit using the 731 Relay Box and a suitable control device, refer to the manual of the 731 Relay Box.
■■■■■■■■
12
■■■■■■■■■■■■■■■■■■■■■■
4 Operation

4 Operation

The 823 Membrane Pump Unit does not have its own operating units. It is operated and controlled via the main instrument in the linked system.
■■■■■■■■
13

5.1 General notes

5 Operation and maintenance

5.1 General notes

5.1.1 Care

The 823 Membrane Pump Unit requires appropriate care. Excess contami­nation of the instrument may result in functional disruptions and a reduc­tion in the lifetime of the sturdy mechanics and electronics.
Spilled chemicals and solvents should be removed immediately. Above all, the plug connections on the rear of the instrument (in particular the mains connection socket) should be protected from contamination.
Caution
Although this is extensively prevented by design measures, the mains plug should be unplugged immediately if aggressive media have pene­trated the inside of the instrument, so as to avoid serious damage to the instrument electronics. In such cases, Metrohm Service must be informed.
■■■■■■■■■■■■■■■■■■■■■■

5.1.2 Maintenance by Metrohm Service

Maintenance of the 823 Membrane Pump Unit is best carried out as part of an annual service, which is performed by specialist personnel of the Metrohm company. If working frequently with caustic and corrosive chemicals, a shorter maintenance interval could be necessary.
Metrohm Service offers every form of technical advice for maintenance and service of all Metrohm instruments.
■■■■■■■■
14
■■■■■■■■■■■■■■■■■■■■■■
5 Operation and maintenance

5.2 Tubings

The tubing requires regular inspection in order to work properly:
Do not kink PTFE tubing. Ensure resistance to solvents . Check the tubing regularly for any damage.

5.3 Quality Management and validation with Metrohm

Quality Management
Metrohm offers you comprehensive support in implementing quality man­agement measures for instruments and software. Further information on this can be found in the brochure «Quality Management with Metrohm» available from your local Metrohm agent.
Validation
Please contact your local Metrohm agent for support in validating instru­ments and software. Here you can also obtain validation documentation to provide help for carrying out the Installation Qualification (IQ) and the Operational Qualification (OQ). IQ and OQ are also offered as a service by the Metrohm agents. In addition, various application bulletins are also available on the subject, which also contain Standard Operat- ing Procedures (SOP) for testing analytical measuring instruments for reproducibility and correctness.
Maintenance
Electronic and mechanical functional groups in Metrohm instruments can and should be checked as part of regular maintenance by specialist per­sonnel from Metrohm. Please ask your local Metrohm agent regarding the precise terms and conditions involved in concluding a corresponding maintenance agreement.
Note
You can find information on the subjects of quality management, vali­dation and maintenance as well as an overview of the documents cur­rently available at www.metrohm.com/com/ under Support.
■■■■■■■■
15

6.1 Problems

6 Troubleshooting

6.1 Problems
Problem Cause Remedy
■■■■■■■■■■■■■■■■■■■■■■
The liquid is not aspirated, but just "bubbles" instead.
The tubing connec­tion of the inlet tub­ing and/or outlet tubing is dripping.
The membrane pump does not aspi­rate the solution.
The membrane pump is pumping too little or not at all.
The inlet tubing and the outlet tubing have been switched.
The tubings are not screwed on correctly.
The inlet valve or the outlet valve is blocked.
The PTFE tubings are kinked.
The membrane platelets stick to each other. This may occur if the mem­brane pump has not been in use for an extended period.
Switch the tubings so that the inlet tubing goes into the solution and the outlet tubing goes into the waste canister.
Check the screw connectors.
Contact Metrohm Service at your Metrohm representative.
Straighten the kinked part or use flawless tub­ings.
Spray some water into the inlet valve with a spray bottle, close the valve and allow to react for some minutes.
■■■■■■■■
16
Too much air is trapped between the membrane platelets during aspirating and pumping.
Contact Metrohm Service at your Metrohm representative.
■■■■■■■■■■■■■■■■■■■■■■
7 Technical specifications

7.1 Membrane pump

Flow rate > 450 mL/min

7.2 Power supply

7 Technical specifications

Voltage
Power consump­tion
16 - 24 V DC
650 mA

7.3 Interfaces and connectors

The 823 Membrane Pump Unit does not have any control interfaces. It is operated and controlled via the power supply.

7.4 Safety specifications

This instrument fulfills the following electrical safety requirements:
CE designation in accordance with the EU directives:
2006/95/EC (Low Voltage Directive, LVD) 2004/108/EC (EMC Directive, EMC)
Design and testing EN/IEC/UL 61010-1
CSA-C22.2 No. 61010-1 Protection class III EN/IEC 60529 Protection class IP20
Safety instructions This document contains safety instructions which have to be followed
by the user in order to ensure safe operation of the instrument.
■■■■■■■■
17

7.5 Electromagnetic compatibility (EMC)

7.5 Electromagnetic compatibility (EMC)
Emission
Standards fulfil­led
Immunity
Standards fulfil­led
EN/IEC 61326-1 EN/IEC 61000-6-3 EN 55011 / CISPR 11
EN/IEC 61326-1 EN/IEC 61000-6-2 EN/IEC 61000-4-2 EN/IEC 61000-4-3 EN/IEC 61000-4-4 EN/IEC 61000-4-5 EN/IEC 61000-4-6 EN/IEC 61000-4-11 EN/IEC 61000-4-14 EN/IEC 61000-4-28
■■■■■■■■■■■■■■■■■■■■■■

7.6 Ambient temperature

Nominal function range
Transport and storage
+5 - +45 °C
–40 - +70 °C

7.7 Reference conditions

Ambient tempera­ture
Relative humidity 60%
+25 °C (±3 °C)
■■■■■■■■
18
■■■■■■■■■■■■■■■■■■■■■■

7.8 Dimensions/material

Width 160 mm
Height 70 mm
Depth 200 mm
Weight 1.68 kg (without accessories)
Material Housing: steel sheet, stove-enameled
7 Technical specifications
■■■■■■■■
19

8.1 Quality Management Principles

8 Conformity and warranty

8.1 Quality Management Principles
Metrohm Ltd. holds the ISO 9001:2000 Certificate, registration number 10872-02, issued by SQS (Swiss Association for Quality and Management Systems). Internal and external audits are carried out periodically to assure that the standards defined by Metrohm’s QM Manual are maintained.
The steps involved in the design, manufacture and servicing of instruments are fully documented and the resulting reports are archived for ten years. The development of software for PCs and instruments is also duly docu­mented and the documents and source codes are archived. Both remain the possession of Metrohm. A non-disclosure agreement may be asked to be provided by those requiring access to them.
The implementation of the ISO 9001:2000 quality management system is described in Metrohm’s QM Manual, which comprises detailed instruc­tions on the following fields of activity:
■■■■■■■■■■■■■■■■■■■■■■
Instrument development
The organization of the instrument design, its planning and the intermedi­ate controls are fully documented and traceable. Laboratory testing accompanies all phases of instrument development.
Software development
Software development occurs in terms of the software life cycle. Tests are performed to detect programming errors and to assess the program’s functionality in a laboratory environment.
Components
All components used in the Metrohm instruments have to satisfy the qual­ity standards that are defined and implemented for our products. Suppli­ers of components are audited by Metrohm as the need arises.
Manufacture
The measures put into practice in the production of our instruments guar­antee a constant quality standard. Production planning and manufacturing procedures, maintenance of production means and testing of compo­nents, intermediate and finished products are prescribed.
Customer support and service
Customer support involves all phases of instrument acquisition and use by the customer, i.e. consulting to define the adequate equipment for the analytical problem at hand, delivery of the equipment, user manuals, train-
■■■■■■■■
20
■■■■■■■■■■■■■■■■■■■■■■
ing, after-sales service and processing of customer complaints. The Metrohm service organization is equipped to support customers in imple­menting standards such as GLP, GMP, ISO 900X, in performing Opera­tional Qualification and Performance Verification of the system compo­nents or in carrying out the System Validation for the quantitative determi­nation of a substance in a given matrix.

8.2 Warranty (Guarantee)

Metrohm guarantees that the deliveries and services it provides are free of errors in materials, design or manufacturing.
The general warranty period is 36 months (exclusions below) from the date of delivery or 18 months in the event of continuous operation. The warranty remains valid on the condition that the servicing is provided by a Service Organization authorized by Metrohm at defined intervals and with a defined scope.
The warranty period for anion suppressors is 120 months from the date of delivery or 60 months in the event of continuous operation.
8 Conformity and warranty
The warranty period for IC separation columns is 90 days after start-up.
For third-party components that are recognizable as such, the manufac­turer's warranty regulations apply.
Consumables and materials with limited storage life and glass breakage in the case of electrodes or other glass parts are excluded from the warranty.
Warranty claims cannot be asserted if the customer has failed to meet his payment obligations according to schedule.
During the warranty period, Metrohm undertakes either to replace free of charge or to credit the purchaser for any assemblies or components that can be shown to be faulty. Any transport or customs fees that may apply are the ordering party’s responsibility.
The precondition for this is that the ordering party must use the Return Material Authorization (RMA) to report the faulty part, along with specifi­cation of the article number, the article designation, an adequate error description, the delivery date and (if applicable) the serial number or the chip data, respectively. In addition, the ordering party undertakes to store the faulty part for at least 24 months in accordance with current storage directives (in compliance with ESD guidelines) and to hold it in readiness for onsite inspection or for return shipment to Metrohm. Metrohm reserves the right to invoice the ordering party for these articles, including retroactively, in the event of noncompliance with these pre-conditions.
The original warranty periods for the original part apply to parts that are replaced or repaired under the above-referenced warranties (no extension of the warranty period).
■■■■■■■■
21
8.2 Warranty (Guarantee)
■■■■■■■■■■■■■■■■■■■■■■
Deficiencies arising from circumstances that are not the responsibility of Metrohm, such as improper storage or improper use, etc., are expressly excluded from the warranty.
Metrohm also offers a 120-month spare parts availability guarantee and a 60-month PC software support warranty, calculated from the date on which the product is withdrawn from the market. The content of this war­ranty is the ability of the customer to obtain functioning spare parts or appropriate software support at market prices during the time of the war­ranty period.
If Metrohm AG is unable to meet this obligation due to circumstances beyond the control of Metrohm AG, then the ordering party shall be offered alternative solutions at preferential conditions.
■■■■■■■■
22
■■■■■■■■■■■■■■■■■■■■■■
9 Accessories
Note
Subject to change without notice.

9.1 Scope of delivery per model version

Note
After receiving the instrument, check the shipment to ensure that it is complete. In doing so, check the version (order number) of your 823 Membrane Pump Unit.

9 Accessories

823 Membrane Pump Unit

2.823.0010
Qty.
Order no. Description
1 1.823.0010 Membrane Pump Unit
Membrane pump for direct connection to Sample Processors.
2 6.1801.120 PVC tubing / 4 mm / 6 mm / 2 m
Material: PVC (transparent) Outer diameter (mm): 6 Inner diameter (mm): 4 Length (m): 2
1 6.1812.000 PTFE tubing 4 mm / 6 mm / 4 m
Material: PTFE Outer diameter (mm): 6 Inner diameter (mm): 4
■■■■■■■■
23
9.1 Scope of delivery per model version
Qty. Order no. Description
1 6.1828.000 PVDF connection nipple
For canister 6.1621.000.
Material: PVDF
1 8.823.8001EN Manual 823 Membrane Pump Unit
■■■■■■■■■■■■■■■■■■■■■■

9.1.2 823 Membrane Pump Unit with aspiration equipment

2.823.0020
Qty.
Order no. Description
1 1.823.0010 Membrane Pump Unit
Membrane pump for direct connection to Sample Processors.
1 6.1446.160 Thermometer plug / B-14/15 / oblique insert
For oblique position of inserts in the titration head 6.1458.010.
Material: PTFE Height (mm): 21 Outer diameter (mm): 13 Ground-joint size: B-14/15
1 6.1543.170 Aspiration tip / M8 thread
Aspiration tip for the Sample Processors.
Material: PTFE Length (mm): 198
■■■■■■■■
24
■■■■■■■■■■■■■■■■■■■■■■
Qty. Order no. Description
1 6.1621.000 Canister / 10 L
As rinsing or waste canister in automated systems.
Material: PE Width (mm): 265 Height (mm): 400 Volume (mL): 10000
2 6.1801.120 PVC tubing / 4 mm / 6 mm / 2 m
Material: PVC (transparent) Outer diameter (mm): 6 Inner diameter (mm): 4 Length (m): 2
9 Accessories
1 6.1805.510 PTFE tubing / M8 / 60 cm
With kink protection.
Material: PTFE Inner diameter (mm): 3 Length (mm): 600
1 6.1812.000 PTFE tubing 4 mm / 6 mm / 4 m
Material: PTFE Outer diameter (mm): 6 Inner diameter (mm): 4
■■■■■■■■
25
9.1 Scope of delivery per model version
Qty. Order no. Description
1 6.1820.070 Coupling 4 mm / 6 mm / M8 inner
For extending aspiration tubing in automated systems. Connects the M8 aspiration tubing directly to the PTFE tubing 6.1812.000.
Material: PTFE
1 6.1828.000 PVDF connection nipple
For canister 6.1621.000.
Material: PVDF
1 8.823.8001EN Manual 823 Membrane Pump Unit
■■■■■■■■■■■■■■■■■■■■■■

9.1.3 823 Membrane Pump Unit with rinsing equipment

2.823.0030
Qty.
Order no. Description
1 1.823.0010 Membrane Pump Unit
Membrane pump for direct connection to Sample Processors.
1 6.1621.000 Canister / 10 L
As rinsing or waste canister in automated systems.
Material: PE Width (mm): 265 Height (mm): 400 Volume (mL): 10000
■■■■■■■■
26
■■■■■■■■■■■■■■■■■■■■■■
Qty. Order no. Description
2 6.1801.120 PVC tubing / 4 mm / 6 mm / 2 m
Material: PVC (transparent) Outer diameter (mm): 6 Inner diameter (mm): 4 Length (m): 2
3 6.1805.060 FEP tubing / M6 / 60 cm
Kink-proof and protected against light.
Material: FEP Inner diameter (mm): 2 Length (mm): 600
9 Accessories
1 6.1808.170 Distributor piece for rinsing and aspirating
for Sample Processors
Distributor piece for installation on Sample Processors for rinsing and aspirating when using externally connected pumps.
1 6.1812.000 PTFE tubing 4 mm / 6 mm / 4 m
Material: PTFE Outer diameter (mm): 6 Inner diameter (mm): 4
■■■■■■■■
27

9.2 Optional accessories

Qty. Order no. Description
1 6.1828.000 PVDF connection nipple
For canister 6.1621.000.
Material: PVDF
3 6.2740.020 Spray nozzle
For the fine-spraying of the rinsing solution.
Material: ETFE Outer diameter (mm): 10 Length (mm): 47
■■■■■■■■■■■■■■■■■■■■■■
1 8.823.8001EN Manual 823 Membrane Pump Unit
9.2 Optional accessories

823 Membrane Pump Unit

2.823.0010
Order no.
6.5323.010 VA rinsing equipment for pump units
VA rinsing equipment for automatic rinsing of the measuring container in the VA stand for use with pump units.
8.823.3003EN Declaration of Conformity for 823 Membrane Pump

9.2.2 823 Membrane Pump Unit with aspiration equipment

2.823.0020
Description
Unit
Order no.
8.823.3003EN Declaration of Conformity for 823 Membrane Pump
■■■■■■■■
28
Description
Unit
■■■■■■■■■■■■■■■■■■■■■■

9.2.3 823 Membrane Pump Unit with rinsing equipment

2.823.0030
Order no. Description
8.823.3003EN Declaration of Conformity for 823 Membrane Pump Unit
9 Accessories
■■■■■■■■
29

Index

Index
■■■■■■■■■■■■■■■■■■■■■■
C
Connect
731 Relay Box .................... 11
Membrane pump ............... 10
Sample Processor ............... 10
D
Dimensions .............................. 19
E
Electrostatic charge .................... 4
F
Front of the instrument .............. 6
G
GLP .......................................... 15
Guarantee ................................ 21
I
Instrument description ................ 1
M
Mains voltage ............................. 4
Material .................................... 19
Membrane pump
Flow rate ............................ 17
Mode of operation ............... 6
Model versions ........................... 1
O
Operation ................................. 13
Optional accessories ................. 28
P
Power supply
Power consumption ........... 17
Voltage .............................. 17
Q
Quality Management ................ 15
S
Safety instructions ...................... 3
Scope of delivery ...................... 23
Service ....................................... 3
Service Agreement ................... 15
Standards ........................... 17, 18
T
Tubing
Maintenance ...................... 15
V
Validation ................................. 15
W
Warranty .................................. 21
■■■■■■■■
30
Loading...