Teachware
Metrohm AG
CH-9101 Herisau
teachware@metrohm.com
This documentation is protected by copyright. All rights reserved.
Although all the information given in this documentation has been
checked with great care, errors cannot be entirely excluded. Should you
notice any mistakes please send us your comments using the address
given above.
Documentation in additional languages can be found on
http://products.metrohm.com under Literature/Technical documenta-
Fig. 4: Filter unit PEEK (6.2821.120) ......................................................... 10
Fig. 5: Connection to injection valve......................................................... 11
Fig. 6: Connection of 818 IC Pump to 830 IC Interface ............................ 13
Fig. 7: Connection of 818 IC Pump to PC ................................................ 14
Fig. 8: Components of the pump head .................................................... 26
Fig. 9: Replacement of the piston seal 34................................................ 26
Fig. 10: Components of inlet valve 35 and outlet valve 36......................... 29
818 IC Pump / 8.818.1023 Instructions for Use
II
1.1 Instrument description
1 Introduction
1.1 Instrument description
The 818 IC Pump is a serial dual piston pump especially developed for
ion chromatography and optimized for use with Metrohm ion
chromatographic instruments (e.g. 819 IC Detector and 820 IC
Separation Center). The 818 IC Pump operates with minimal residual
pulsation and has an excellent flow constancy. All wet parts of the
pump are metal free.
The 818 IC Pump is completely remote controlled by the Metohm
software «IC Net» (version 2.3 or higher). All functions are fast and
easily accessible via software and can be clearly arranged at the
monitor. To assure optimum operational reliability, both lower and
upper pressure limit values can be set. If such a limit is reached, the
pump switches itself off automatically. An additional purge valve allows
rapid removal of air from the pump system.
818 IC Pump / 8.818.1023 Instructions for Use
1
1 Introduction
1.2 Parts and controls
In this section you will find the numbers and designations of the parts
and controls of the 818 IC Pump. The numbering applies throughout
the instructions for use, i.e. bold numbers in the text (e.g.
the parts and controls illustrated here.
1 2345
11
) refer to
Fig. 1: Front of 818 IC Pump
Mains pilot lamp
1
Lights up when instrument is switched
on
Purge valve
2
Connection capillary
3
to injection valve of IC system.
Connection capillary
4
in pump head (intergal)
6789
Outlet capillary
5
of Purge-Ventil.
Connection capillary
6
between pump head and purge valve.
Aspirating capillary
7
Pump head 6.2824.100
8
Fastening screws
9
for pump head
2
818 IC Pump / 8.818.1023 Instructions for Use
1.2Parts and controls
Fig. 2: Rear of 818 IC Pump
Fuse cover
10
Changing the fuses, see section 2.6.2.Mains switch
For switching instrument on/off:
I = ON 0 = OFFMains connector
For mains connection see section 2.6.
1110
12
13 14 15 16 17 18
Recorder output
14
For recording the pressure,
1 mV = 1 MPa
Switch "Time Constant"
15
Time constant for the recorder output:
HIGH≈ 1 s; LOW≈ 20 ms
Mains voltage selector
11
Voltage selection see section 2.6.1.
Serial number
12
Control interface
13
details see section 5.2.
Switch for external control
16
Selector to choose the interface for
remot control of the pump
RS 232 / Current Loop
RS 232 interface
17
Current Loop interface
18
(Details on request)
818 IC Pump / 8.818.1023 Instructions for Use
3
1 Introduction
1.3 Information about the Instructions for Use
Please read through these Instructions for Use carefully before
operating the 818 IC Pump. The Instructions for Use contain
information and warnings to which the user must pay attention in order
to assure safe operation of the instrument.
1.3.1 Organization
These 8.818.1023 Instructions for Use for the 818 IC Pump provide a
comprehensive overview of the installation, startup procedure,
operation, fault rectification and technical specifications of this
instrument. The Instructions for Use are organized as follows:
Section 1 Introduction
Description of the instrument, parts and controls,
safety notes
Section 2 Installation
Mounting the pump head, electrical connections,
tubing connections, mains connection
Section 3 Operation
Operation via «IC Net»
Section 4 Notes – Maintenance – Faults
Practical notes, maintenance, fault rectification
Section 5 Interfaces
Description of the interfaces
Section 6 Appendix
Technical data, standard equipment, options,
validation, warranty, declaration of conformity, index
To find the information you require about the instrument please use
either the Table of contents or the Index at the back.
4
818 IC Pump / 8.818.1023 Instructions for Use
1.3Information about the Instructions for Use
1.3.2 Notation and pictograms
The following notations and pictograms (symbols) are used in these
Instructions for Use:
Range Menu item, parameter or entry
value
in «IC Net» program
SYSTEM STATE Program window
in «IC Net» program
<OK> Button
in «IC Net» program
10 Part or control of 818
Danger/Warning
This symbol indicates a possible
risk of death or injury to the user
and possible damage to the
instrument or its components by
electricity.
Danger/Warning
This symbol indicates a possible
risk of death or injury to the user
and possible damage to the
instrument or its components.
Attention
This symbol indicates important
information that you should read
before continuing.
Information
This symbol indicates additional
information and tips which may be
of particular use to you.
818 IC Pump / 8.818.1023 Instructions for Use
5
1 Introduction
1.4 Safety notes
1.4.1 Electrical safety
While electrical safety in the handling of the 818 IC Pump is assured in
the context of the specifications IEC 1010-1 (protection class 1, degree
of protection IP40), the following points should be noted:
• Mains connection
Set the mains voltage and check the mains fuse and mains
connection in accordance with the instructions in section 2.6.
• Opening the instrument
Inside the instrument there are no parts which must be set or adjusted
by the user.
If the 818 IC Pump is connected to the power supply, the instrument
must not be opened nor must parts be removed from it, otherwise
there is a danger of coming into contact with components which are
live. Hence, always disconnect the instrument from all voltage sources
before you open it and ensure that the mains cable is disconnected from mains connection 10 !
• Protection against static charges
Electronic components are sensitive to static charging and can be
destroyed by discharges. Before you touch any of the components
inside the 818 IC Pump, you should earth yourself and any tools you
are using by touching an earthed object (e.g. housing of the
instrument or a radiator) to eliminate any static charges which exist.
1.4.2 General safety rules
• Handling of solvents
Check the pump tubing and all input and output leads periodically for
possible leaks. Follow the relevant instructions regarding the handling
of flammable and/or toxic solvents and their disposal.
6
818 IC Pump / 8.818.1023 Instructions for Use
2.1Setting up the instrument
2 Installation
2.1 Setting up the instrument
2.1.1 Packaging
The 818 IC Pump is supplied together with the separately packed
accessories in special packagings containing shock-absorbing foam
linings designed to provide excellent protection. The actual instrument
is packed in an evacuated polyethylene bag to prevent the ingress of
dust. Please store all these special packagings as only they assure
transport of the instruments free from damage.
2.1.2 Check
After receipt, immediately check whether the shipment is complete and
has arrived without damage (compare with delivery note and list of
accessories in section 6.2). In the case of transport damage, see
instructions in section 6.5.1 "Warranty".
2.1.3 Location
Position the 818 IC Pump in the laboratory at a location convenient for
operation, free from vibrations and protected against a corrosive
atmosphere and contamination by chemicals.
To avoid disturbing temperature influences on the insulated column
compartment, the pump and eluent reservoir must be protected
against direct sunlight.
2.1.4 Arrangement of the instruments
The 818 IC Pump should always be at the very bottom of a IC system.
The 818 IC Pump should always be placed at the bottom of the
instrument stack so that any leaks which may occur in the pump
tubing or connections cannot cause damage to the other instruments
by leakage liquids.
818 IC Pump / 8.818.1023 Instructions for Use
7
2 Installation
2.2 Mounting the pump head
1 Remove the transportation lock
To ensure that the pump drive is not damaged during transport
a black plastic block is mounted in place of the pump head on
the right side of the 818 IC Pump front panel. This transport
safeguard block has to be removed after undoing the four
screws 9.
To avoid any damage to the pump head, the protective block must be
remounted if the pump has to be moved any great distance.
2 Mounting the pump head
The Pump Head8 (6.2824.100) with integrated piston units is
unpacked and fixed in place of the transport safeguard block
with the four hexagon screws9as shown in Fig. 2 .
To ensure that the pump head is not positioned wrongly, the holes at
the rear for the clamping bolts have different depths, i.e. 1 clamping
bolt is longer than the rest. The deepest hole must naturally
accommodate the longest bolt. If this is not the case, the pump will
not function properly.
3 Connecting capillary to pump head
The connection capillary6 fastened to purge valve2 is
screwed onto the pump head8 as shown in Fig. 2.
The pressure range of the 818 IC Pump with the 6.28.24.100 PEEK
…
pump head isf 0
30.0 MPa (0…300 bar)..
8
818 IC Pump / 8.818.1023 Instructions for Use
2.3Attaching the accessories
2.3 Attaching the accessories
2.3.1 Capillaries
Some of the connections under high pressure between the 818 IC
Pump and the IC system must be set up by the user, see section 2.4.
For that the 6.1831.010 PEEK capillary (i.d. = 0.25 mm, o.d. =
length = 3 m) supplied in the accessories of the 818 IC Pump can be
used.
Capillaries fitted with new connectors must have a perfectly flat cut
surface. For PEEK capillaries it is best to use the 6.2621.080 Capillary tubing cutter
1
/16",
Fig. 3: Connect capillaries
PEEK capillary (6.1831.010)
19
20 19
Compression fitting (6.2744.010)
20
For the connection of 6.1831.010 PEEK capillaries or of the 6.1834.000
PTFE aspirating tubing, the supplied 6.2744.010 PEEK Compression fittings are used. It is best to proceed as follows:
1 Mount compression fitting
Slide a compression fitting 20 (6.2744.010) over the end of the
capillary 19 to be fastened as shown in Fig. 3.
2 Insert capillary in connection
Push capillary end in the corresponding connection as far as it
will go (to avoid dead volume).
3 Tighten compression fitting
Tighten compression fitting 20 by hand (never use tools).
2.3.2 Pulsation dampener
To protect the column material against pressure drops caused by the
injector, the use of a pulsation dampener connected between the pump
and the injection valve of the 820 IC Separation Center is
recommended. The optional 6.2620.150 Pulsation dampener MF is
very well suited to this purpose (see section 6.3).
The metal-free 6.2620.150 Pulsation dampener is supplied fully
assembled and has two connections for capillaries, for which two
818 IC Pump / 8.818.1023 Instructions for Use
9
2 Installation
6.2744.010 PEEK compression fittings can be used. The flow direction
is arbitrary. The pulsation dampener is positioned in the interior of the
820 IC Separation Center on the base below the injection valve (see
819/820 Instructions for Use).
The pulsation dampener is filled with isopropanol when new. Rinse
your IC system carefully after the first installation of a new pulsation
dampener.
2.3.3 Filter unit PEEK
The 6.2821.120 Filter unit PEEK 24 supplied (see Fig. 4) serves to
avoid contamination of the piston seals by abrasive particles and can
be used in the pressure range 0…30 MPa (0…300 bar). The filter unit
consists of the housing 21, the filter 22 and the connector 23 to be
screwed into the housing 21. For the connection of capillaries 19, PEEK
compression fittings 20 (6.2744.010) must be used.
New filter 22 are available as an option with the ordering number
6.2821.110 (10 pieces).
For the connection of the filter unit, please note the flow direction
arrow printed on the housing.
The filter unit is filled with isopropanol when new. Rinse your IC system
carefully after the first installation of a new filter unit.
20 19 20 232221 19
Fig. 4: Filter unit PEEK (6.2821.120)
PEEK capillary (6.1831.010)
19
Compression fitting (6.2744.010)
20
Filter housing
21
Filter 6.2821.130
22
Filter connector
23
10
818 IC Pump / 8.818.1023 Instructions for Use
2.4Connecting the tubing
2.4 Connecting the tubing
2.4.1 Connection to injection valve
It is recommended to use 6.1831.010 PEEK capillaries, a 6.2620.150
Pulsation dampener (see section 2.3.2), and a Filter unit PEEK (see
section 2.3.3) to connect the 818 IC Pump to the injection valve of the
820 IC Separation Center. Proceed as follows:
25 243
Fig. 5: Connection to injection valve
Connection capillary
3
to 820 IC Separation Center.
PEEK capillary (6.1831.010)
818
Filter unit PEEK (6.2821.120)
24
Pulsation dampener (6.2620.150)
25
820
1 Connection to 818 IC Pump
• Cut connection capillary 3 (6.1831.010 PEEK capillary) to the
required length and equip it with compression fittings.
• Attach connection capillary 3 to the upper connection of the
purge valve (see Fig. 1).
2 Connection of filter unit PEEK
• Attach the other end of connection capillary 3 at the filter
housing 21 of the filter unit 24 (see Fig. 4).
• Connect a PEEK capillary 6.1831.010 cut to the required
length and equipped with connectors to the filter connector
23 of the filter unit 24.
3 Installation of the capillary in the IC Separation
Center
4 Connection of the pulsation dampener
5 Connection to the injection valve
818 IC Pump / 8.818.1023 Instructions for Use
• Procedure, see 819/820 Instructions for Use.
• Procedure, see section 2.3.2.
• Procedure, see section 2.3.3.
11
Loading...
+ 39 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.