Metrohm 812 User Manual

CH-9101 Herisau/Schweiz Internet www.metrohm.com E-Mail info@metrohm.ch
812 Valve Unit
812 Valve Unit
Metrohm
FILL IN JECT
FILL IN JECT
B
8.812.1003 Instructions for Use
16.11.2001 / dö
Table of contents

Table of contents

1 Introduction ..................................................................................1
1.1 Instrument description .................................................................1
1.2 Parts and controls.........................................................................2
1.3 Information on the Instructions for Use.....................................4
1.3.1 Organization .........................................................................4
1.3.2 Notation and pictograms......................................................5
1.4 Safety notes ...................................................................................6
2 Installation.....................................................................................7
2.1 Setting up the instrument ............................................................7
2.1.1 Packaging.............................................................................7
2.1.2 Check ...................................................................................7
2.1.3 Location................................................................................7
2.1.4 Arrangement of the instruments...........................................7
2.2 Mains connection ..........................................................................8
2.3 Electrical connection ....................................................................8
2.3.1 Connection at the 762 IC Interface.......................................8
2.3.2 Connection at 817 Bioscan ..................................................9
2.4 Connections at injection valve..................................................10
2.4.1 General information ............................................................ 10
2.4.2 Attaching standard accessories.........................................10
2.5 Software installation ...................................................................12
3 Operation......................................................................................13
3.1 Manual operation.........................................................................13
3.2 Operation via «IC Net»................................................................14
3.2.1 812 Valve Unit icon.............................................................14
3.2.2 Settings in the "812 Valve Unit" window..............................14
4 Appendix .......................................................................................17
4.1 Technical data..............................................................................17
4.2 Standard equipment....................................................................19
4.3 Optional accessories ..................................................................20
4.4 Warranty and conformity ...........................................................21
4.4.1 Warranty .............................................................................21
4.4.2 EU Declaration of conformity..............................................22
4.4.3 Certificate of conformity and system validation .................23
4.4 Index ..............................................................................................24
812 Valve Unit
I
Table of contents

List of figures

Fig. 1: Front of 812 Valve Unit........................................................................... 2
Fig. 2
: Rear of 812 Valve Unit ........................................................................... 2
Fig. 3
: Connection of 812 Valve Unit at 762 IC Interface...................................... 9
Fig. 4
: Switching of injection valves................................................................. 10
Fig. 5
: Connection at injection valve................................................................ 11
812 Valve Unit
II
1.1 Instrument description

1 Introduction

1.1 Instrument description

The 812 Valve Unit is an instrument with one or two electrically driven ein 6-port injector valves. Two different versions of the 812 Valve Unit are available:
2.812.0010 Valve Unit with 1 valve
2.812.0020 Valve Unit with 2 valves
Valve parts in contact with the eluent are manufactured out of PEEK material. The instrument is equipped with an external power supply so it can be operated as a stand alone unit and is independent of other Metrohm devices.
The 812 Valve Unit can be used for a variety of different applications:
In combination with the 817 Bioscan or the 791 VA Detec-
tor, the 812 Valve Unit is used as the injector for the sys­tems.
With the 812 Valve Unit, a 761 Compact IC can be up-
graded to be a combined anion an cation system.
In combination with the 816 Eluent Selector, the 2-port
version can be used for automatic column switching with an intermediate rinsing step.
The 812 Valve Unit can also be used in a large variety of
sample preparation techniques, like for cutting tech­niques, or switching between different sample streams.
812 Valve Unit
1
1 Introduction

1.2 Parts and controls

812 Valve Unit
3
21 4
FILL INJECT
A
: Front of 812 Valve Unit
Fig. 1
65
Metrohm
FILL INJECT
B
Type: 1.812.00X0
Nr.: XX XXX
External
Power
24VDC/2A
5VDC/0,5A
. d n u o r
G
r
t
g
t
.
r
e
s
c
a
l
t
l
l
l
i F
t
o
e
i
j
n
S
F
I
P
n
I
M
N
M
N
O
U
O
U
R
C
R
732 IC Detector
C
B
978
d n u o r G
10 11 12
. r
t
g
t
.
r
e
s
c
a
l
t
l
l
l
i F
t
o
e
i
j
n
S
F
I
P
n
I
N
M
M
N
U
O
O
U
R
C
C
R
Made by Metrohm Herisau Switzerland
A
732 IC Detector
Fig. 2
: Rear of 812 Valve Unit
812 Valve Unit
2
1.2 Parts and controls
1 Valve A 7 Connection for external supply
connection of power supply unit (5 V, 0.5 A / 24 V, 2 A)
2 "FILL" key for valve A 8 Terminal block for valve B
(without function for 2.812.0010 instrument version) Ground, Fill, Inject: inputs for control of the valve Pos.Fill: output signal on switching of the valve to position "FILL" Integr.Start: output signal on switching of the valve to position "INJECT"
3 "INJECT" key for valve A 9 Connection for 732 IC Detector B
(without function for 2.812.0010 instrument version)
4 "FILL" key for valve B
(without function for 2.812.0010 instrument version)
10 Terminal block for valve A
Ground, Fill, Inject: inputs for control of the valve Pos.Fill: output signal on switching of the valve to position "FILL" Integr.Start: output signal on switching of the valve to position "INJECT"
5 "INJECT" key for valve B
11 Connection for 732 IC Detector A
(without function for 2.812.0010 instrument version)
6 Ventil B
(not available for 2.812.0010
12 Model plate
with serial number
instrument version)
812 Valve Unit
3
1 Introduction

1.3 Information on the Instructions for Use

Please read through these Instructions for Use carefully before you put the 812 Valve Unit into operation. The Instructions for Use contain information and warnings to which the user must pay attention in order to assure safe operation of the instrument.

1.3.1 Organization

These 8.812.1003 Instructions for Use for the 812 Valve Unit provide a comprehensive overview of the installation, startup procedure, opera­tion and technical specifications of this instrument. The Instructions for Use are organized as follows:
Section 1 Introduction
General description of instrument, parts and controls and safety notes
Section 2 Installation
Mains connection, electrical connection, connection of accessories
Section 3 Operation
Manual operation and operation via «IC Net»
Section 4 Appendix
Technical data, standard equipment, options, warranty, declarations of conformity, index
To find the required information on the instruments you will find it an advantage to use either the Table of contents or the Index at the back.
812 Valve Unit
4
1.3 Information on the Instructions for Use

1.3.2 Notation and pictograms

The following notations and pictograms (symbols) are used in these In­structions for Use:
Range Menu item, parameter or entry
value
in «IC Net» program
SYSTEM STATE Program window
in «IC Net» program
<OK> Button
in «IC Net» program
[ FILL ] Switch or key
7 Part or control of 812
Hazard
This symbol draws attention to a possible danger to life or of injury if the associated directions are not followed correctly.
Warning This symbol draws attention to possible damage to instruments or instrument parts if the associated directions are not followed cor­rectly.
Caution This symbol marks important information. First read the associ­ated directions before you con­tinue.
812 Valve Unit
Comment This symbol marks additional information and tips.
5
1 Introduction

1.4 Safety notes

While electrical safety in the handling of the 812 Valve Unit is assured in the context of the specifications IEC 61010-1 (protection class 1), the following points should be noted:
Mains connection
The mains connection must be effected in accordance with the instructions in section 2.2.
Opening the instrument
Inside the instrument there are no parts which must be set or adjusted by the user.
If the 812 Valve Unit is connected to the power supply, the instrument must not be opened nor must parts be removed from it. Hence, before opening the instrument, always ensure that the power supply unit is disconnected from connection 7!
Protection against static charges
Electronic components are sensitive to static charging and can be destroyed by discharges. Before you touch any of the components inside the 812 Valve Unit, you should earth yourself and any tools you are using by touching an earthed object (e.g. housing of the instru­ment or a radiator) to eliminate any static charges which exist.
812 Valve Unit
6

2.1 Setting up the instrument

2 Installation

2.1 Setting up the instrument

2.1.1 Packaging

The 812 Valve Unit is supplied together with the separately packed ac­cessories in special packagings containing shock-absorbing foam lin­ings designed to provide excellent protection. The instrument itself is packed in an evacuated polyethylene bag to prevent the ingress of dust. Please store all these special packagings as only they assure transport of the instrument free from damage.

2.1.2 Check

After receipt, immediately check whether the shipment is complete and has arrived without damage (compare with delivery note and list of ac­cessories in section 4.2). In the case of transport damage, see instruc­tions in section 4.4.1 "Warranty".

2.1.3 Location

Position the instrument in the laboratory at a location convenient for op­eration, free from vibrations and protected against a corrosive atmos­phere and contamination by chemicals.

2.1.4 Arrangement of the instruments

The 812 Valve Unit can be piled up together with other IC instruments (e.g. 732, 733, 709).
812 Valve Unit
7
2 Installation

2.2 Mains connection

The 812 Valve Unit is operated with the 6.2152.000 Power supply unit which automatically adjusts itself to the existing mains voltage (100240 V) and frequency (5060 Hz).
The power supply unit is connected to the mains using one of the fol­lowing mains cables
6.2122.020 with plug SEV 12 (Switzerland, …)
6.2122.040 with plug CEE(7), VII (Germany, …)
6.2133.070 with plug NEMA 5-15 (USA, …)
which are three-cored and fitted with a plug with an earthing pin.
The cable permanently mounted to the power supply unit is plugged into connection 7 at the 812 Valve Unit (see Fig. 2). The operational readiness is shown by the LEDs lighting up on the "FILL" or "INJECT" keys.

2.3 Electrical connection

2.3.1 Connection at the 762 IC Interface

Always switch off 762 IC Interface and 812 Valve Unit before you connect the two instruments.
The connection of the 812 Valve Unit at the 762 IC Interface enables the three following functions:
Control of valve A by the 762 IC Interface
via remote line
Control of valve B by the 762 IC Interface
via remote line
Automatic start of data acquisition at the
762 IC Interface by switching valve A or valve B to the "INJECT" position
For instrument connection proceed as follows (see Fig. 3):
1 Connection for control of valve A
Connect the plugs inscribed with "GND", "INJECT" and "FILL"
of the 6.2128.100 cable to the connections "Ground", "Fill" and "Inject" on the terminal block 10 of the 812 Valve Unit (see Fig. 2).
Connect the plug inscribed with "Inject NO" of the 6.2128.100
cable to the desired remote line 1...7 "Events - Run" at the 762 IC Interface and the plug inscribed with "Inject COM" to the same remote line "Events - COM" (see Fig. 5 Instructions for use 762).
812 Valve Unit
8
2.3 Electrical connection
6.2128.100
6.2128.100
6.2115.070
BA
812762
Fig. 3
: Connection of 812 Valve Unit 812 at 762 IC Interface
Connect the plug inscribed with "LOAD NO" of the
6.2128.100 cable at the desired remote line 1...7 "Events - Run" at the 762 IC Interface and the plug inscribed with "LOAD COM" to the same remote line "Events - COM" (see Fig. 5 Instructions for use 762).
2 Connection for control of valve B
Connect the plugs inscribed with "GND", "INJECT" and "FILL"
of the 6.2128.100 cable to the connections "Ground", "Fill" and "Inject" on the terminal block 8 of the 812 Valve Unit (see Fig. 2).
Connect the plug inscribed with "Inject NO" of the 6.2128.100
cable to the desired remote line 1...7 "Events - Run" at the 762 IC Interface and the plug inscribed with "Inject COM" to the same remote line "Events - COM" (see Fig. 5 Instructions for use 762).
Connect the plug inscribed with "LOAD NO" of the
6.2128.100 cable at the desired remote line 1...7 "Events - Run" at the 762 IC Interface and the plug inscribed with "LOAD COM" to the same remote line "Events - COM" (see Fig. 5 Instructions for use 762).
3 Connection for external start (option)
Connect the 6.2115.070 cable to the connections "Integr.
Start - RUN" and "Integr. Start - COM" on the terminal block
10 (valve A) or 8 (valve B) of the 812 Valve Unit (see Fig. 2).
Connect the other end of the 6.2115.070 cable to the remote
lines "Events - Start - Run" and "Events - Start - Com" at the 762 IC Interface (see Fig. 5 Instructions for use 762).

2.3.2 Connection at 817 Bioscan

The connection of the 812 Valve Unit to the 817 Bioscan is described in detail in the Instructions for Use 817.
812 Valve Unit
9
2 Installation

2.4 Connections at injection valve

2.4.1 General information

The two six-way valves at the 812 Valve Unit each have 6 connections for capillaries which are switched as follows:
Fill
45
36
12
: Switching of injection valves
Fig. 4
Connections 1 and 2 of the valve are normally used for supply of sam­ple solution, connections 4 and 5 for eluent supply, and connections 3 and 6 for sample loop installation. For solution supply in the low pres­sure range (sample feed by use of a syringe or a peristaltic pump), the
6.1803.020 PTFE capillary (i.d. 0.97 mm) is used; in the high pressure range, the 6.1831.010 PEEK capillary (i.d. 0.25 mm) is used. For both capillaries, a 6.2744.010 PEEK compression fitting is used as connec­tor.
Inject
45
36
12

2.4.2 Attaching standard accessories

The standard accessories supplied with the 812 Valve Unit can be used to equip one valve for manual filling of the sample loop. Proceed as fol­lows (see Fig. 5):
1 Connecting the syringe
Cut off a piece of the 6.1803.020 PTFE capillary to the de-
sired length (at best using the optionally available 6.2621.080 tubing cutter) and fit it on both sides with a 6.2744.010 PEEK compression fitting (do not tighten too firmly).
Screw one end of the PTFE capillary 13 to connection 1 of
the injection valve.
Screw the other end of the PTFE capillary 13 to the
6.2744.120 coupling.
Push 6.2816.020 syringe (without needle) as far as it will go
into the connection of the 6.2744.120 coupling.
812 Valve Unit
10
2.4 Connections at injection valve
17 16
Fig. 5
Syringe tubing
13
6.1803.020 PTFE tubing with
6.2744.120 coupling and
6.2816.020 syringe
14 Aspirating tubing
6.1803.020 PTFE tubing for aspirating the sample
15 Sample loop
6.1825.210 PEEK sample loop (20 µL )
45
15
36
21
13 14
: Connection at injection valve
16 Eluent outlet
6.1831.010 PEEK capillary
17 Eluent inlet
6.1831.010 PEEK capillary
2 Connecting the aspirating tubing
Cut off a piece of the 6.1803.020 PTFE capillary to the de-
sired length (at best using the optionally available 6.2621.080 tubing cutter) and fit it on one side with a 6.2744.010 PEEK compression fitting (do not tighten too firmly).
Screw this end of the aspirating tubing 14 to connection 2 of
the injection valve.
3 Connecting the sample loop
Fit the 6.1825.210 sample loop on both sides with a
6.2744.010 compression fitting.
Screw the sample loop 15 (20 µL) to connections 3 and 6 of
the injection valve.
4 Connecting the eluent supply
Screw the PEEK capillary 17 for eluent supply from the high
pressure pump to connection 5 of the injection valve.
5 Connecting the separating column
Screw the PEEK capillary 16 for eluent supply to the separat-
ing column to connection 4 of the injection valve.
812 Valve Unit
11
2 Installation

2.5 Software installation

The PC program «IC Net 2.1» is required for the operation of the 812 Valve Unit by a PC. This program runs under Windows 95, Windows 98, Windows NT and Windows 2000 operating systems and is installed ac­cording to section 1.4.2 of the «IC Net» Instructions for Use.
The installation of the 812 Valve Unit is described in section 6.11 of the «IC Net» Instructions for Use.
812 Valve Unit
12
3.1 Manual operation

3 Operation

3.1 Manual operation

FILL
Switch instrument on/off
The 812 Valve Unit is switched on and off by connecting/dis­connecting the 6.2152.000 power supply unit attached to connection 7 (see section 2.2).
After the instrument has been connected to the mains the LEDs on the "FILL" or "INJECT" keys light up and show that the instrument is ready for use.
Switch to "FILL" position
By pressing the "FILL" key, the injection valve A or B is switched to the "FILL" position.
Lighting up of the LED in the "FILL" key indicates the "FILL" position.
Fill
45
36
12
INJECT
Switch to "INJECT" position
By pressing the "INJECT" key, the injection valve A or B is switched to the "INJECT" position.
Lighting up of the LED in the "INJECT" key indicates the "INJECT" position.
Inject
45
36
12
812 Valve Unit
13
3 Operation
3.2 Operation via «IC Net»
This section describes only the most important points concerning the operation of the 812 Valve Unit. For further details please refer to the «IC Net» Instructions for Use and to the on-line help in the PC pro­gram.

3.2.1 812 Valve Unit icon

2.812.0010 Valve Unit with 1 injector
2.812.0020 Valve Unit with 2 injectors

3.2.2 Settings in the "812 Valve Unit" window

The 812 Valve Unit window for parameter settings is opened by selecting
Open menu option with the right mouse button or by double-clicking
the the 812 icon in the system window. It consists of the three tabs
Program and Links.
Manual
The Manual tab of the 812 Valve Unit window is used for manual opera­tion of the injection valves.
Manual,
812 Valve Unit
14
3.2 Operation via «IC Net»
Valve A
<Fill> Switch valve A to "FILL" position. <Inject> Switch valve A to "INJECT" position.
Valve B
<Fill> Switch valve B to "FILL" position. <Inject> Switch valve B to "INJECT" position.
Time program
On the Program tab of the 812 Valve Unit window a user-defined time program can be entered. This program starts automatically as defined
Start mode window either at the moment the determination is star-
in the
Start with determination) or at the moment the sample is injected
ted (
Start with inject).
(
Time (1st column) Time at which program instruction is applied.
Range: If no time is entered, the program instruction is applied together with the last instruction with time entry.
Instruction (2nd column) Program instruction (see below).
Parameter (3rd column) Parameter for program instruction (see below).
ENABLED Enable program start (a disabled program is
not started).
<Add> Add new program instruction. <Delete> Delete selected program instruction. <Verify> Test the time program (error messages are dis-
played if program is wrong).
0.0 ... 999.9 min
812 Valve Unit
15
3 Operation
List of program instructions
The following program instructions can be added to the time program on the
Instruction Parameter entry Meaning
ValveA fill, inject Switch injection valve A to "inject" or
ValveB fill, inject Switch injection valve B to "inject" or
Program page:
"
fill" position.
"
fill" position.
Links
The Links tab of the 812 Valve Unit window is used for COM port selec­tion and settings (details see IC Net Instructions for Use).
812 Valve Unit
16
4.1 Technical data

4 Appendix

4.1 Technical data

Parts and controls
Keypad
Chemically resistant membrane keypad made of polyester with function keys
Indicators LEDs for display of valve position
Power supply
External External supply via DIN connector:
5 V / 0.5 A 24 V / 2 A (transient, 200300 ms)
6
3
5
7
1
4
2
1 +5 V DC / 0.5 A 2 not assigned 3 +24 V DC / 2 A 4 0 V (digital) * 5 0 V (analog) * 6 Ground 7 Ground
* Pin 4 and Pin 5 must be
connected in the vicinity of the power source ("zero point of star")
Valve interfaces
Connection Function
Inputs
0V
10k
10k
Ground Ground
+5V
Fill
+5V
Inject
Valve "FILL"
Valve is switched to the "FILL" position
Valve "INJECT"
Valve is switched to the "INJECT" position
H
L
H
L
> 2 ms
t
P
> 2 ms
t
P
Outputs
COM
RUN
COM
RUN
Position "Fill"
A pulse is outputted when the valve is switched to the "FILL" position.
Integrator Start
A pulse is outputted when the valve is switched to the "INJECT" position.
H
L
H
L
250 ms
t
P
250 ms
t
P
812 Valve Unit
17
4 Appendix
Safety specifications
Safety notes The Instructions for Use include information and
warnings, which must be heeded by the user to assure safe operation of the instrument.
Electromagnetic compatibility (EMC)
Emitted interference Standards met:
EN 50081-1/2, EN55022 (class B)
Immunity to interference Standards met:
EN50082-1 IEC61000-4-2 (class 3) IEC61000-4-4 (class 4) IEC61000-4-11 IEC61000-4-14 (class 3)
Ambient temperature
Nominal operating range +5+45°C
(at 2080 % atmospheric humidity)
Storage, Transport –40+70°C
Housing
Material of cover Polyurethane rigid foam (PUR) with fire protection
for fire class UL94VO, FCH-free
Material of base Steel, enameled
Dimensions
Width 255 mm
Height 128 mm
Depth 365 mm
Weight (with accessories) 2.812.0010: 4.3 kg
2.812.0020: 5.1 kg
812 Valve Unit
18
4.2 Standard equipment

4.2 Standard equipment

Subject to changes ! All dimensions are given in mm.
The 812 Valve Unit is available in two versions:
2.812.0010 Valve Unit with 1 valve
2.812.0020 Valve Unit with 2 valves
These instruments include the following parts:
Quant. Order No. Description
2.812.0010
2.812.0020
1 - 1.812.0010 Valve Unit with 1 valve
- 1 1.812.0020 Valve Unit with 2 valves
1 1 6.1803.020 PTFE capillary
Length L = 5 m
1 1 6.1825.210 Sample loop aus PEEK (20 µL)
for injection valve; incl. 2 PEEK compression fittings (6.2744.010)
1 1 6.1831.010 PEEK capillary
Length L = 3 m Diameter d = 0.25 mm
812 Valve Unit
1 1 6.2122.0X0 Mains cable
to customer's specifications:
Cable socket
Type IEC 320/C 13 Type SEV 12 (CH…)............................... 6.2122.020
Type IEC 320/C 13 Type CEE (7), VII (D…)........................... 6.2122.040
Type CEE (22), V Type NEMA 5-15 (USA…) ...................... 6.2122.070
1 2 6.2128.100 Connecting cable
For connection 812 – 762/817
1 1 6.2152.000 Power supply unit
1 2 6.2744.010 PEEK compression fitting
For the connection of 6.1831.010 PEEK capillaries or 6.1803.020 PTFE capillaries, set of 5
Cable connector
26
19
4 Appendix
Quant. Order No. Description
2.812.0010
2.812.0020
1 1 6.2744.120 Coupling 1⁄16" – Luer
Coupling between 6.2744.010 PEEK compression fitting and 6.2816.020 Syringe
1 1 6.2816.020 Syringe
made of PP, volume = 10 mL; for manual filling of the sample loop
1 1 8.812.1003 Instructions for Use (English)
for 812 Valve Unit
22
76

4.3 Optional accessories

Order No. Description
6.1825.XXX Sample loop made of PEEK
For injection valve; incl. 2 PEEK compression fittings (6.2744.010)
6.1825.230: volume = 10 µL
6.1825.210: volume = 20 µL
6.1825.220: volume = 100 µL
6.2115.070 Cable Connection cable 812/762 for external start
6.2621.080 Capillary tubing cutter
for 6.1831.010 PEEK capillaries and
6.1803.020 PTFE capillaries
incl. 5 additional blades
55
118
812 Valve Unit
20

4.4 Warranty and conformity

4.4 Warranty and conformity

4.4.1 Warranty

The warranty on our products is limited to defects that are traceable to material, construction or manufacturing error that occur within 12 months from the day of delivery. In this case, the defects will be recti­fied in our workshops free of charge. Transport costs are to be paid by the customer.
For day and night operation, the warranty is limited to 6 months.
Glass breakage in the case of electrodes or other parts is not covered by the warranty. Checks, which are not a result of material or manufac­turing faults, are also charged during the warranty period. For parts of outside manufacture insofar as these constitute an appreciable part of our instrument, the warranty stipulations of the manufacturer in question apply.
With the regard to the guarantee of accuracy, the technical specifica­tions in the instruction manual are authoritative.
Concerning defects in material, construction or design as well as the absence of guaranteed features, the orderer has no rights or claims ex­cept those mentioned above.
If damage of the packaging is evident on receipt of a consignment or if the goods show signs of transport damage after unpacking, the carrier must be informed immediately and a written damage report demanded. Lack of an official damage report releases Metrohm from any liability to pay compensation.
If any instruments and parts have to be returned, the original packaging should be used if at all possible. This applies above all to instruments, electrodes, burette cylinders and PTFE pistons. Before embedment in wood shavings or similar material, the parts must be packed in a dust­proof package (for instruments, use of a plastic bag is imperative). If open assemblies are enclosed in the scope of delivery that are sensi­tive to electromagnetic voltages (e.g. data interfaces etc.) these must be returned in the associated original protective packaging (e.g. con­ductive protective bag). (Exception: assemblies with built-in voltage source belong in a non-conductive protective packaging). For damage, which arises as a result of non-compliance with these instructions, no warranty responsibility whatsoever will be accepted by Metrohm.
812 Valve Unit
21
4 Appendix

4.4.2 EU Declaration of conformity

EU Declaration of Conformity
The METROHM AG company, Herisau, Switzerland hereby certifies, that the instrument:
812 Valve Unit
meets the requirements of EC Directives 89/336/EEC and 73/23/EEC..
Source of the specifications:
EN 50081-1/2 Electromagnetic compatibility, basic specification;
Emitted Interference
EN 50082-1 Electromagnetic compatibility, basic specification;
Interference Immunity
Description of the instrument:
Instrument with 1 or 2 electrically driven six-way valves.
Herisau, April 4, 2001
Dr. J. Frank Ch. Buchmann
Development Manager Production and Quality Assurance Manager
812 Valve Unit
22
4.4 Warranty and conformity

4.4.3 Certificate of conformity and system validation

Certificate of Conformity and System Validation
This is to certify the conformity to the standard specifications for electrical appli­ances and accessories, as well as to the standard specifications for security and to system validation issued by the manufacturing company.
Name of commodity: 812 Valve Unit
Name of manufacturer: Metrohm Ltd., Herisau, Switzerland
Principal technical information: Voltages: 100-240 V Frequency: 50-60 Hz
This Metrohm instrument has been built and has undergone final type testing according to the standards:
EN50081-1/2, EN50082-1, EN55022 (class B), IEC61000-4-2 (class 3), IEC61000-4-4 (class 4), IEC61000-4-11, IEC61000-4-14 (class 3) Electromagnetic compatibility
It has also been certified by the Swiss Electrotechnical Association (SEV), which is member of the International Certification Body (CB/IEC).
The technical specifications are documented in the instruction manual.
Metrohm Ltd. is holder of the SQS-certificate of the quality system ISO 9001 for quality assurance in design/development, production, installation and servicing.
Herisau, April 4, 2001
Dr. J. Frank Ch. Buchmann
Development Manager Production and Quality Assurance Manager
812 Valve Unit
23
4 Appendix

4.5 Index

A
Ambient temperature............................ 18
Appendix............................................... 17
Arrangement of the instruments..............7
Aspirating tubing 14
Figure................................................ 11
Installation......................................... 11
Attaching standard accessories .......... 10
C
Cable (6.2115.070)............................ 9,20
Cable (6.2128.100)......................... 8,9,19
Capillary tubing cutter
(6.2621.080) ................................ 10,20
Caution ....................................................5
Certificate of conformity and
system validation.............................. 23
Check ......................................................7
Comment.................................................5
Conformity ............................................ 22
Connection
Aspirating tubing .............................. 11
Bioscan 817.........................................9
Eluent supply.................................... 11
Sample loop ..................................... 11
Separating column ........................... 11
Syringe.............................................. 10
Connection 7
Connect power supply unit .................8
Figure...................................................3
Connection 9
Figure...................................................3
Connection 11
Figure...................................................3
Connections at injection valve ............. 10
Coupling (6.2744.120).......................... 20
D
Declaration of conformity ..................... 22
Dimensions........................................... 18
E
Earthing ............................................... 6,8
Electrical connection ...............................8
Electrical safety........................................6
Electromagnetic compatibility.............. 18
Eluent inlet 17
Figure................................................ 11
Eluent outlet 16
Figure................................................ 11
EMC ...................................................... 18
Emitted interference ............................. 18
EU Declaration of conformity ............... 22
F
Front ........................................................2
H
Hazard.....................................................5
Housing .................................................18
I
Immunity to interference .......................18
Information on the
Instructions for Use ............................. 4
Installation ...............................................7
Instructions for Use 8.812.1003..............4
Instrument description ............................1
Introduction .............................................1
K
Key 2 (FILL)
Figure ..................................................3
Operation ..........................................13
Key 3 (INJECT)
Figure ..................................................3
Operation ..........................................13
Key 4 (FILL)
Figure ..................................................3
Operation ..........................................13
Key 5 (INJECT)
Figure ..................................................3
Operation ..........................................13
Keypad .................................................. 17
L
Links ......................................................16
List of program instructions ..................16
Location................................................... 7
M
Mains cable ...........................................19
Mains connection
Procedure............................................ 8
Safety notes.........................................6
Manual operation ..................................13
Model plate 12
Figure ..................................................3
N
Notation ...................................................5
O
Opening the instrument ..........................6
Operation...............................................13
Operation via IC Net.............................. 14
Optional accessories ............................20
Organization ............................................4
P
Packaging ...............................................7
Parts and controls ................................... 2
PEEK capillary (6.1831.010) ............10,19
PEEK compression fitting
(6.2744.010).................................10,19
Pictograms .............................................. 5
Power supply ........................................17
Power supply unit..........................8,13,19
Program instructions............................. 16
Protection class....................................... 6
PTFE capillary (6.1803.020)............. 10,19
R
Rear......................................................... 2
S
Safety notes ............................................6
Safety specifications ............................. 18
Sample loop 15
Figure ................................................ 11
Installation ......................................... 11
Ordering designation ........................ 19
Sample loop (6.1825.XXX).................... 20
Setting up the instrument........................ 7
Software installation .............................. 12
Standard equipment ............................. 19
Static charges .........................................6
Storage.................................................. 18
Switch instrument on/off .......................13
Switching of injection valves ................. 10
Syringe (6.2816.020)........................ 10,20
Syringe tubing 13
Figure ................................................ 11
Installation ......................................... 10
T
Technical data....................................... 17
Terminal block 8
Connection of 6. 2115.070 cable .......9
Connection of 6.2128.100 cable ........9
Figure .................................................. 3
Technical data................................... 17
Terminal block 10
Connection of 6. 2115.070 cable .......9
Connection of 6.2128.100 cable ........8
Figure .................................................. 3
Technical data................................... 17
Time program........................................15
Transport................................................. 7
Transport damage ................................21
V
Valve 1 (A)
Figure .................................................. 3
Operation ..........................................13
Valve 6 (B)
Figure .................................................. 3
Operation ..........................................13
Valve A..............................................15,16
Valve B.............................................. 15,16
Valve interfaces ..................................... 17
W
Warning ................................................... 5
812 Valve Unit
24
Loading...