KEYENCE LS-7600 User Manual

Page 1
96M12221
User’s Manual
High-speed, High-precision Digital Micrometer
LS-7600 Series
Page 2

Preface

This instruction manual provides necessary information on the operation and maintenance of the LS-7600 Series along with precautions. Please read the manual carefully and be sure you understand the information provided before attempting to install and operate the LS-7600 Series. Keep this manual handy for future reference. Please make sure that the end users are provided with this manual.

Symbols

The following symbols are used for the list of precautions to ensure safety and to prevent personal injury and/or property damage when using the LS-7600 Series. The following symbols alert you to important messages. Be sure to read these messages carefully.
DANGER
WARNING
CAUTION
Important:
Note:
Tips
Reference:
Failure to follow instructions may lead to death or serious injury.
Failure to follow instructions may lead to injury.
Failure to follow instructions may lead to product damage (product malfunctions, etc.).
Provides important precautions and restrictions on proper opera­tion.
Provides additional information on proper operation.
Provides useful information on proper operation.
Provides reference pages.

General precautions

1. No part of this manual may be reprinted or reproduced in any form or by any means without the prior written permission of KEYENCE CORPORATION.
2. The content of this manual is subject to change without notice.
3. KEYENCE has thoroughly checked and reviewed this manual. Please contact the sales office listed at the end of this manual if you have any questions or comments regarding this manual, or if you find an error.
4. KEYENCE assumes no liability for damages resulting from the use of the information in this manual, item 3 above notwithstanding.
5. KEYENCE will replace any incomplete or incorrectly collated manual.
6. All company names and product names in this manual are registered trademarks or trademarks of their respective owners.
Page 3

Safety Precautions

General precautions
• At startup and during operation, be sure to monitor the functions and performance of the LS-7600 Series.
•Take sufficient safety measures to prevent damage to the human body and/or equip­ment that may be caused if this product should fail to operate properly.
• Do not modify the LS-7600 Series or use it in any way other than as described in the specifications. Its functions and performance are not guaranteed under said conditions.
• When the LS-7600 Series is used in combination with other instruments, its functions and performance may be degraded, depending on the operating conditions and surrounding environment.
• Do not use the LS-7600 Series for the purpose of protecting the human body.
• Do not expose the LS-7600 and peripheral devices to sudden temperature change, as this may cause condensation.
Precautions
WARNING
Operation
• Always use the LS-7601 at 24 VDC, otherwise a fire, electric shock or product failure may result.
• Prepare the applicable AC power supply cable if you use LS­S11.
• Secure the GND line of the power cable of the AC power supply stand to protective ground, otherwise an electric shock or product failure may result.
• Do not disassemble or modify the LS-7600 Series. This may cause fire or electric shock.
When abnormal conditions are encountered
If the following conditions are encountered, immediately turn off the power. Continuing to use the LS-7600 Series under abnormal conditions may cause fire, electric shock or equipment failure. Contact your nearest KEYENCE sales office for repairs.
• When water or foreign matter enters the controller.
• When the LS-7600 Series is dropped or the housing is dam­aged.
• When the controller produces smoke or an abnormal smell.
96M12221
i
Page 4
CAUTION
Usage
Be sure to turn off the power to the LS-7600 Series and any
• connected devices before connecting or disconnecting the cables. Otherwise, the camera and connected devices may be damaged.
• Do not turn off the power while setting a parameter. Otherwise, the settings may be partially or completely lost.
• Do not block the ventilation slots on the LS-7600 Series and peripheral devices. A rise in inner temperature may cause equipment failure.
Proper environment and conditions
To use the LS-7600 Series properly and safely, do not install the LS-7600 Series in locations with the following conditions. Use of this equipment in an improper environment may cause fire, electric shock, or equipment failure.
• Locations with high humidity, a large amount of dust, or poor ventilation
• Locations where the temperature rises excessively due to direct sunlight, etc.
• Locations where corrosive or flammable gas exists
• Locations where the LS-7600 Series is directly subjected to vibration or impact
• This product is a precision instrument. Be sure to take appropriate measures against direct vibration or impact. Otherwise, the optical components in the measurement device may be damaged.
• Locations where water, oil or chemicals may splash the LS­7600 Series
• Locations where static electricity is easily built up
Countermeasures against Noise
The LS-7600 Series may malfunction or fail to operate due to noise generated from power lines or high-tension lines. If the LS­7600 Series is operated close to such noise sources, use a noise filter, wire the cable of the LS-7600 Series in an independent conduit, or locate the LS-7600 Series in an area with proper insulation from noise.
Best Management Practice for Perchlorate Materials – California only
When you sell, manufacture and/or waste the products containing perchlorate material in California, the following statement must appear on the exterior of all outer shipping packages and on consumer packages, based on California Code of Regulations.
“Perchlorate Material – special handling may apply, See www.dtsc.ca.gov/hazardouswaste/perchlorate.”
Note: Alternative option is to indicate this statement on your MSDS or manual accompanying with each product.
This product includes a CR coin type cell containing perchlorate material. If you ship this product (or your product including this product) to California, you must ensure to comply with this regulation.
ii
Page 5
Note: Influence of Ambient Operating Temperature
Ensure that the ambient operating temperature is constant. A change in the ambient operating temperature may result in measurement errors. If the ambient operating temperature changes by 10°C, it will take approximately 60 minutes for the interior temperature distribution of the LS-7600 Series to be uniform.
Warming Up
Do not operate the LS-7600 Series for approximately 30 minutes after it is turned on. The internal circuitry of the LS-7600 Series is not stable immediately after it is turned on. The measured value drifts gradually as a result.
Influence of Dust and Dirt
A measurement error may result due to dirt, dust, water, or oil in any of the following cases.
• Cover glass: Blow off the dirt, dust, water or oil on the cover glass with clean air. If the cover glass is excessively dirty, wipe the glass with a soft cloth moistened with isopropyl alcohol.
• Target surface: Blow off the dirt, dust, water or oil on the sur­face with clean air or wipe the surface.
• Intrusion of dirt, dust, water or oil blown or sprayed into optical sensing area: Install a protection cover or perform air purging.
Influence of Vibration
The measured value of an object may fluctuate if the object is vibrating. If that happens, increase the averaging number. This will ensure highly accurate measurement.
Targets
The measured value of an object may have an error due to the shape and surface condition. Locate a reference object within the measuring area and use the calibration function of the LS-7600 Series to correct the error if that happens.
Influence of Air Flow
The measured value of an object may fluctuate due to a slow flow of air. In that case, the following countermeasures are effective.
• Protect the measuring head with a cover.
• Use a fan to mix the air vigorously between the transmitter and receiver.
Maintenance
Do not wipe the LS-7600 Series with a damp cloth or a cloth moistened with benzine or paint thinner. This may cause the discoloration or deformation of the LS-7600 Series. If the LS­7600 Series is excessively dirty, use a tightly squeezed cloth moistened with thin neutral detergent. Then wipe the LS-7600 Series with a soft dry cloth.
Precautions for LED
The LS-7010M/LS-7030M/LS-7070M Measuring Head uses an LED light source. Abide by the following precautions when using the unit.
CAUTION
• Do not look at the LED light source for a long time. The LED used as the light source is classified into class 1 according to IEC60825-1 standards. Basically the light beam of the LED is safe, but do not look at the LED for a long time.
• Do not disassemble the unit. The unit does not incorporate a mechanism to stop the emission of the LED when the unit is disassembled. Never attempt to disassemble the unit, otherwise you will be exposed to LED light.
iii
Page 6

Precautions for CE Markings

Keyence confirms that the LS-7600 Series meets EC directive requirements. The LS-7600 Series bears CE markings. Keep the following conditions if the LS-7600 Series is used in European countries.
EMC Directives
Keep the following condition so that the LS-7600 Series will satisfy EN61326-1 require­ments.
• The power cable and I/O cable connected to the controller are both less than 30m in length.
Low-voltage Directives
When using the LS-S11 AC Power Supply Stand as a dedicated power supply to the LS­7600 Series, keep the following conditions so that the LS-S11 will satisfy EN61010-1 requirements. The LS-S11 satisfies EN60825-1 (LED Class 1) requirements with no conditions attached.
Environment Conditions
Installation category (Overvoltage category): II (see note 1) Pollution degree: 2 (see note 2)
Operation and Installation Conditions
• The LS-S11 AC Power Stand is a dedicated power supply for the LS-7600 Series. Never use the LS-S11 for other purposes.
• When using the LS-7600 Series in European countries, check that the AC power supply cable satisfies EC directive requirements certified by an authorized third party. Consult your Keyence representative for the AC power supply cable in detail.
• Before turning the LS-7600 Series on, connect the ground terminal of the AC power supply cable to the protective ground terminal of the power supply.
• Be sure to use a time lag fuse rated 250 V, 3.15 A for the replaceable fuse of the LS­S11.
Note 1: Installation categories (overvoltage categories) include impulse overvoltage regu-
Note 2: The pollution degree refers to the level of solid, liquid, or gas contamination that
lations at transient overvoltage levels specified by EC directives. Installation category II indicates a level where power is provided from fixed installations, such as distribution panels.
deteriorates the dielectric strength or surface resistance of the LS-7600 Series. Pollution degree 2 indicates the normal indoor environment (free of non­conductive contamination).
iv
Page 7

About this Manual

The following section provides information on the configuration of each page of this manual along with symbols and terms used in this manual.

Page Configuration and Symbols

Chapter 4 Operation Control
4-4 Automatic Zero Function
The automatic zero function sets measuring values to zero (0.00000) instantly. This func­tion is convenient for zero-point calibration for a variety of target. Furthermore, master calibration is possible by using this function in combination with the offset function.
Tips: Describes information that will deepen the user’s comprehension of the manual along with other useful information.
Caution: Indicates information that, if not heeded, could result in relatively serious or minor injury, damage to the LS-7600 Series, or faulty operation.
Note: This function will not be available while the controller is awaiting the result of
measurement, during which the LCD monitor will display “– – –. – – – –.”)
The automatic zero value will not be lost after the LS-7600 Series is turned
Tips
off. The automatic zero value for each program will be stored.
Automatic zero settings are made through the following functional components.
• Control panel/Remote control console
• External terminal
4
• RS-232C interface
Control Panel/Remote Control Console
If a single measuring value is displayed, an automatic zero setting will be made for the corresponding OUT number only. If both measuring values are displayed, an automatic zero setting will be made for the selected OUT number on the screen.
1
Press the [ZERO] key once. Automatic zero will be set.
To cancel the setting, press the [ZERO] key again.
External I/O Terminal
Automatic zero will be set by short-circuiting the I/O, ZERO, and COM terminals on the rear panel of the controller together. Refer to Section 7 I/O Terminals on page 7-1 for details.
CAUTION
Tips
4-10
• Do not look at the LED light source for a long time. The LED used as the light source is classified into class 1 according to IEC60825-1 standards. Basically the light beam of the LED is safe, but do not look at the LED for a long time
Master Calibration with Offset Function (Automatic Offset Function)
Set the size of a master workpiece to an offset value and make an automatic zero setting while measuring the master workpiece. Then the size of the master workpiece will be displayed as an offset value. Refer to 3-3 Measure­ment of Outer Diameter and Width (with Two Measuring Heads) on page 3-4 for an application example. Refer to Offset on page 5-43 for the offset function in detail.
OUT1
UPPER 1.00000 STANDARD LOWER –1.00000
ZERO
0.00000
10.00000
CAM1
RUN
P – 00
01/02/18 17:50:42
Note: Describes items that need the user’s attention in order to prevent mistakes in the operation of the LS­7600 Series.
The above page sample is for your reference only. It does not coincide with any of
the actual pages of this manual.

Definition of Terms

This manual uses the following terms.
Term Description
LS-7600 Series Refers to the LS-7601 controller and the measuring head as a set. Controller Refers to the LS-7601 controller and remote control console
(accessory).
Measuring head Refers to the LS-7010M/LS-7030M/LS-7070M measuring head.
v
Page 8

Optimum Measurement Methods

Refer to the following methods, and select the optimum measurement method according to the type of target.
Measurement of Outer Diameter
of a Round Bar or Wire
Measurement of Average Outer
Diameter of a Round Bar or Wire
Measurement of Sheet Width
D
Measurement of outer diameter and width (with single measuring head) (Page 3-2)
Measurement of Inner
Diameter of a Disc
Measurement of inner diameter and clearance (Page 3-5)
Measurement of Connector
Pins and IC Pins
Y
X
Measurement of outer diameter and width (with two measuring heads in X and Y directions) (Page 3-3)
Measurement of Outer Diameter
and Eccentricity of Roller
Measurement of outer diameter and eccentricity (Page 3-6)
Measurement of Size
Uniformity
L
Measurement of outer diameter and width (with two measuring heads for larger objects) (Page 3-4)
Measuring Robot Movement
and Positioning of LCD Plate
Measurement of movement and positioning (Page 3-7)
Calculation of Max., Min.,
Average, and Standard Deviation
Values of Measured Data
Measurement of pitch (Page 3-8)
Group processing (Page 3-9)
Statistical processing (Page 3-10)
vi
Page 9

Configuration of Manual

Chapter 1 Getting Started
Provides information on precautions and necessary preparations required.
Chapter 2 Basic Operation of LS-7600 Series
Provides information on the basic operation of the LS-7600 Series, panel keys, and remote control console.
Chapter 3 Easy Setting Guide
Provides information on targets and settings for typical applications respectively.
Chapter 4 Operation Control
Provides information on setting control items on the control panel while the LS-7600 Series is in measurement operation.
Chapter 5 Function Settings
Provides information on the overview of program settings and setting methods.
Chapter 6 Environment Settings
Provides information on basic operation settings.
Chapter 7 I/O Terminals
Provides information on the specifications of the I/O terminals and timing charts.
1
2
3
4
5
6
7
Chapter 8 RS-232C
Provides information on how to connect external devices and the communications function to operate the LS-7600 Series through the
external devices.
Chapter 9 Specifications
Provides the specifications, characteristics, and external dimensions of the LS-7600 Series.
Chapter 10 Troubleshooting
Provides information on troubleshooting and error messages.
Appendix
Provides a list of options and the index of the manual.
8
9
10
Appendix
vii
Page 10

Contents

Safety Precautions ..............................................................................................................i
Precautions for CE Markings ........................................................................................... iv
About this Manual .............................................................................................................. v
Page Configuration and Symbols.................................................................................... v
Definition of Terms .......................................................................................................... v
Optimum Measurement Methods .................................................................................... vi
Configuration of Manual .................................................................................................. vii
Contents........................................................................................................................... viii
Chapter 1 Getting Started
Outline and Features of LS-7600 Series ....................................................................... 1-2
System Applications .......................................................................................................1-4
Package Contents ...........................................................................................................1-5
LS-7601 ....................................................................................................................... 1-5
LS-7010M .................................................................................................................... 1-5
LS-7030M .................................................................................................................... 1-5
LS-7070M .................................................................................................................... 1-6
LS-C✱✱A ..................................................................................................................... 1-6
LS-C✱✱AM ..................................................................................................................1-6
LS-S11 .........................................................................................................................1-7
Functions and Nomenclature......................................................................................... 1-8
Controller ..................................................................................................................... 1-8
Measuring Head........................................................................................................... 1-9
Mounting and Connection ............................................................................................ 1-11
Mounting the Measuring Head ................................................................................... 1-11
Mounting the Controller.............................................................................................. 1-13
Connection .................................................................................................................1-17
Turning AC Power Supply Stand On and Off.............................................................. 1-19
Adjustments to LCD Monitor ....................................................................................... 1-20
Measurement and Setting Overview ........................................................................... 1-21
Resetting to Initial Status .............................................................................................1-22
Chapter 2 Basic Operation of LS-7600 Series
Control Panel Functions and Nomenclature ................................................................ 2-2
Panel Key Functions .................................................................................................... 2-2
Remote Control Console Functions and Nomenclature............................................... 2-3
Names and Functions of the LCD Monitor Screen ...................................................... 2-4
Chapter 3 Easy Setting Guide
Measurement of Outer Diameter and Width (with Single Measuring Head) ..............3-2
Measurement of Outer Diameter and Width (with Two Measuring Heads).................3-3
Measurement of Outer Diameter and Width
(with Two Measuring Heads for Larger Objects) ..........................................................3-4
Measurement of Inner Diameter and Clearance ........................................................... 3-5
Measurement of Outer Diameter and Eccentricity ....................................................... 3-6
Measurement of Movement and Positioning ................................................................3-7
Measurement of Pitch ..................................................................................................... 3-8
Group Processing ........................................................................................................... 3-9
Statistical Processing ...................................................................................................3-10
Chapter 4 Operation Control
Switching Display Screen .............................................................................................. 4-2
Measured Value Display ..............................................................................................4-2
NG RECORD Display .................................................................................................. 4-5
viii
Page 11
Target Viewer Function ................................................................................................... 4-7
Switching Cameras ...................................................................................................... 4-8
Hold Function ..................................................................................................................4-9
Automatic Zero Function ..............................................................................................4-10
Key Lock Function ........................................................................................................ 4-11
Program Changes ......................................................................................................... 4-12
Chapter 5 Function Settings
Flow of Program Settings............................................................................................... 5-2
Default Values and Possible Setting Ranges................................................................ 5-4
Copying Program Setting Details and Initialization .....................................................5-6
Setting Screen .................................................................................................................5-7
Area Settings ...................................................................................................................5-9
Area Selection............................................................................................................ 5-10
Area Display ...............................................................................................................5-15
Focus Monitor ............................................................................................................ 5-16
Area Check ................................................................................................................ 5-17
Difference Check........................................................................................................ 5-18
Function Output Selection.......................................................................................... 5-19
Change of Edge Detection Threshold ........................................................................5-20
Calibration Settings ...................................................................................................... 5-22
Calibration Type Selection ..........................................................................................5-24
Logic Calibration Settings .......................................................................................... 5-24
Two-point Calibration Settings.................................................................................... 5-26
One-point Calibration Settings ................................................................................... 5-27
Initialization of Calibration .......................................................................................... 5-28
Output Settings ............................................................................................................. 5-29
Area Calculation......................................................................................................... 5-30
Number of Averaging Measurements......................................................................... 5-31
Measuring Mode ........................................................................................................ 5-33
Offset ......................................................................................................................... 5-43
Analog Output Settings ..............................................................................................5-44
Comparator Output Setting ........................................................................................ 5-47
Tolerance (LIMIT) Settings ........................................................................................... 5-49
Limit Range ................................................................................................................ 5-50
Limit Settings ............................................................................................................. 5-51
Hold Value Elimination ............................................................................................... 5-52
Option Settings ............................................................................................................. 5-55
Tolerance Mode (LIMITS)........................................................................................... 5-56
Timing Mode .............................................................................................................. 5-56
Display Unit ................................................................................................................5-57
Minimum Display Unit ................................................................................................ 5-58
I/O Mode .................................................................................................................... 5-59
Units, Default Minimum Values, and Setting Ranges .................................................5-60
Graph TIME ................................................................................................................5-62
Statistical Processing ................................................................................................. 5-62
Group Processing ...................................................................................................... 5-63
Application Function................................................................................................... 5-64
Group Processing ......................................................................................................... 5-65
Statistical Processing ...................................................................................................5-66
Application Function .................................................................................................... 5-68
Drill Function ..............................................................................................................5-68
Dual Head Function ................................................................................................... 5-70
ix
Page 12
Chapter 6 Environment Settings
Possible Environment Settings...................................................................................... 6-2
Default Settings............................................................................................................ 6-2
Changing Environment Settings .................................................................................. 6-4
Details of Environment Settings .................................................................................... 6-5
Program ....................................................................................................................... 6-5
RS-232C ...................................................................................................................... 6-7
Head ............................................................................................................................ 6-9
Buzzer ........................................................................................................................6-12
Key Lock .................................................................................................................... 6-13
Display Language ...................................................................................................... 6-13
LCD Monitor Backlight ............................................................................................... 6-14
Date and Time............................................................................................................6-14
Chapter 7 I/O Terminals
Nomenclature and Functions of I/O Terminals .............................................................7-2
Te r minal Block.............................................................................................................. 7-2
Connector I/O............................................................................................................... 7-4
VIDEO Output ..............................................................................................................7-7
Functions ..................................................................................................................... 7-7
Electrical Specifications ............................................................................................... 7-9
Timing Chart ..................................................................................................................7-11
Chapter 8 RS-232C
Specifications .................................................................................................................. 8-2
Communication using Serial Commands .....................................................................8-4
Connecting to a PC/PLC Link Unit ............................................................................... 8-4
Connecting to KV Series .............................................................................................. 8-5
Outline of Command Formats ......................................................................................8-6
Measurement Serial Commands ................................................................................. 8-7
Setup Serial Commands ............................................................................................8-14
Serial Commands to Check the LS Settings .............................................................. 8-22
Batch Commands ...................................................................................................... 8-23
Timing Chart ...............................................................................................................8-24
ASCII Table ................................................................................................................8-25
Communication using an External Synchronous Trigger ......................................... 8-26
Command List ............................................................................................................... 8-32
Chapter 9 Specifications
Specifications .................................................................................................................. 9-2
Controller ..................................................................................................................... 9-2
Measuring Head........................................................................................................... 9-3
Optional Accessories ................................................................................................... 9-3
Characteristics ................................................................................................................ 9-4
Measurement Range and Accuracy ............................................................................. 9-4
Temperature Characteristics ........................................................................................9-5
Response Delay Time.................................................................................................. 9-6
Focus Characteristics...................................................................................................9-6
Mode Specifications ....................................................................................................... 9-7
External Dimensions....................................................................................................... 9-8
Chapter 10 Troubleshooting
Troubleshooting Guide ................................................................................................. 10-2
Error Messages ............................................................................................................. 10-6
x
Page 13
Appendix
List of Optional Accessories........................................................................... Appendix-2
Setting Record Sheet ....................................................................................... Appendix-3
Environment Setting Record Sheet ................................................................ Appendix-5
Index ................................................................................................................. Appendix-6
xi
Page 14
MEMO
xii
Page 15
Chapter 1
Getting Started
This section provides information on the configuration of the LS-7600 Series, precautions, and necessary preparations required before operating the LS-7600 Series. Familiarize yourself with this section before using the LS-7600 Series.
Outline and Features of LS-7600 Series .............................. 1-2
System Applications ............................................................. 1-4
Package Contents ................................................................ 1-5
Functions and Nomenclature ............................................... 1-8
Mounting and Connection .................................................. 1-11
Turning AC Power Supply Stand On and Off ..................... 1-19
Adjustments to LCD Monitor .............................................. 1-20
Measurement and Setting Overview .................................. 1-21
Resetting to Initial Status ................................................... 1-22
1-1
Page 16
Chapter 1 Getting Started

Outline and Features of LS-7600 Series

1
The LS-7600 Series is a high-speed, high-accuracy digital micrometer used for the dimen­sional measurement of objects without coming into contact with the objects. It is such a versatile model that it has a wide range of applications including in-line measurement and offline measurement.
Special Features
High-speed Sampling of 2400 Times per Second
Ensures high-speed sampling that is twice as fast as the sampling speed of conven­tional models. The continuous measurement of extruded products and the in-line measurement of moving workpieces are possible.
Repeat Accuracy of ±0.15
Incorporates the latest optical system, thus ensuring excellent repeat accuracy that is twice as high as the accuracy of conventional models, thus supporting the manufacture of high-precision products.
Connecting Two Measuring Heads for two-channel Simultaneous Measurement
Two measuring heads can be used in combination for the simultaneous measurement of two targets.
Stable Detection of Transparent Objects with Threshold Change
The threshold change function supported by the DE processor makes it possible to detect transparent objects stably.
Incorporates a Target Viewer Function for Measurement Point Monitoring
Incorporates a CMOS image sensor to monitor targets, thus making it possible to check measurement points no matter how complicated in shape the targets are.
Easy-to-use Remote Control Console
A simple, interactive menu is used on the handheld remote control console for ease of operation.
A Variety of Inspection Functions
Incorporates a group comparison function and statistical processing function.
μm
1-2
Page 17
Chapter 1 Getting Started
Principle of Measurement
The high-intensity GaN green LED radiates light, which will be changed into uniform parallel light through the special diffusion unit and collimator lens and emitted to the target in the measuring range. Then the shadow image of the target will appear on the HL-CCD (high-speed linear CCD) through the telecentric optical system. The output incident signal of the HL-CCD will be processed by the DE (digital edge-detection) processor in the controller and CPU. As a result, the dimensions of the target will be displayed and output.
1
HL-CCD for measurement use
CMOS image sensor
Receiver Transmitter
A/D conversion
A/D conversion
Telecentric optical system
DE processor
Controller
High-intensity GaN green LEDBeam splitter
Special diffusion unit
Target
Frame memory
CPU
Collimator lens
TFT LCD monitor
Video output
Control input Control output BCD output
Analog output RS-232C interface
1-3
Page 18
Chapter 1 Getting Started

System Applications

1
The LS-7600 Series has a wide range of applications when used in combination with commercially available peripheral devices.
PC
Control values or measuring values are retrieved over RS­232C.
Indicator Light and Buzzer
Alarms are generated according to the output of discrimination results.
LS-7601
Programmable Logic Controller
Control output and measured values are retrieved and programs for measurement timing control are changed.
1-4
Photoelectric Sensor or Proximity Sensor
Timing input signals are sent whenever targets are detected.
Recorder
The results of measurements are recorded.
Printer
The results of measurements are printed.
Page 19

Package Contents

Chapter 1 Getting Started
Check that nothing is missing from the LS-7600 Series package before use.

LS-7601

Controller Remote Control
Console
Four Panel Mounting Brackets

LS-7010M

Measuring Head
Allen-head bolt
(Five, M3 x 45 with a washer)
Allen-head bolt
(Four, M4 x 20)
1
LS-7600 Series User’s Manual

LS-7030M

Measuring Head
Allen-head bolt
(Six, M5 x 45 with a washer)
Allen-head bolt
(Four, M4 x 20)
1-5
Page 20
Chapter 1 Getting Started

LS-7070M

1
Measuring Head
Allen-head bolt
(Six, M4 x 50 with a washer)
Allen-head bolt
(Four, M5 x 20)
LS-C✱✱A
Extension Cable (Cable between the controller and measuring head)
LS-C3A: 3-m cable
✱✱... LS-C10A: 10-m cable ✱✱... LS-C30A: 30-m cable
Up to two cables are connectable, provided that the total length of the cables does not exceed 40 m.
LS-C✱✱AM
1-6
Camera cable
LS-C3AM: 3-m cable
✱✱... LS-C10AM: 10-m cable ✱✱... LS-C30AM: 30-m cable
Up to two cables are connectable, provided that the total length of the cables does not exceed 40 m.
Page 21
Chapter 1 Getting Started

LS-S11

Stand Unit Base Unit
Screws (Four, M4 x 8)
Keyence ships each package with utmost care and attention. If there should be any
improper or damaged product, contact your Keyence representative.
1
1-7
Page 22
Chapter 1 Getting Started

Functions and Nomenclature

1
The following section provides information on the functions and nomenclature of the controller and measuring head.

Controller

LCD Monitor
Control Panel
(Page 2-2, Section 2 Basic Operation of LS-7600 Series)
Power Supply Indicator
The indicator is lit when power is supplied to the LCD monitor.
Connector I/O
Used for signal I/O (Page 7-2, Section 7 I/O Terminals)
LCD Monitor Adjustment Trimmer
Used to make color, hue, or brightness adjustments to the LCD monitor. (Page 1-20)
Controller (Bottom)
Controller (Rear Side)
Modular Connector
Used to connect the remote control console connector or RS-232C cable connector. (Page 8-2, Section 8 RS-232C)
Terminal Block
Used for signal I/O (Page 7-2, Section 7 I/O Terminals)
1-8
Power Input Terminals (24 VDC)
Provide 24 VDC power supply.
DC IN
Connect the LS-S11 AC Power Supply Cable.
Camera Connector
Connect the camera cable.
Extension Connector
Connect the cable between the controller and the measuring head.
Video Output Connector
Connect the monitor cable.
Page 23

Measuring Head

Chapter 1 Getting Started
LS-7010M
Multi-purpose Mounting Holes
A fixture for targets can be attached.
Receiver (R Head)
Receives light from the transmitter.
Camera Cable
Connected to the controller.
LS-7030M
Multi-purpose Mounting Holes
A fixture for targets can be attached.
Receiver (R Head)
Receives light from the transmitter.
Camera Unit
Extension Cable
Connected to the controller.
Cover Glass
Cover Glass
Transmitter (T Head)
Emits light for measurement.
TR Base
Used to mount the transmitter and receiver.
TR Cables
Used to connect the transmitter
Mounting Holes
Used to mount the measuring head.
Transmitter (T Head)
Emits light for measurement.
TR Base
Used to mount the transmitter and receiver.
TR Cables
Used to connect the transmitter and receiver.
1
Camera Cable
Connected to the controller.
Mounting holes
Used to mount the measuring head.
Extension Cable
Connected to the controller.
1-9
Page 24
1
Chapter 1 Getting Started
LS-7070M
Multi-purpose Mounting Holes
A fixture for targets can be attached.
Receiver (R Head)
Receives light from the transmitter.
Extension Cable
Connected to the controller.
Camera Cable
Connected to the controller.
Cover Glass
TR Base
Used to mount the transmitter and receiver.
TR Cables
Used to connect the transmitter and receiver.
Mounting holes
Used to mount the measuring head.
Transmitter (T Head)
Emits light for measurement.
1-10
Page 25

Mounting and Connection

Chapter 1 Getting Started
Mount the measuring head and the controller, and connect them with the cables.
Mounting the Measuring Head
The measuring head can be mounted with or without the TR base employed, according to the type of targets and operating environment. The following section provides information on how to mount the measuring head with or without the TR base employed, along with mounting conditions required.
Mounting the Controller
Mount the controller to panels, such as control or operation panels, or to the LS-S11 AC Power Supply Stand. The LS-S11 Power Supply Stand is sold separately.
Connection
After the measuring head and the controller are mounted, connect them with the cables.

Mounting the Measuring Head

Mounting Restrictions
The transmitter and receiver must abide by the following restrictions when the measur­ing head is mounted.
Vertical Position Mismatching Angle Mismatching
±1 mm max.
TransmitterReceiver
±0.25° max
TransmitterReceiver
1
±1 mm max.
The values are the same for LS-7010M, LS-7030M, and LS-7070M.
TransmitterReceiver
±0.25° max
±0.25° max
±0.25° max
TransmitterReceiver
TransmitterReceiver
TransmitterReceiver
1-11
Page 26
1
Chapter 1 Getting Started
With TR Base
The transmitter and receiver are mounted to the TR base at the time of shipment. This section provides how to install the transmitter and receiver (LS-7030M in this case) that are mounted on the TR base. LS-7010M and LS-7070M can also be installed in the same way.
Use the four, M4x20 Allen-head bolts, which are provided with the package, to mount the TR base. To mount the TR base with the bolts on the surface of the TR base, use the mounting holes on the surface.
To mount the TR base with the bolts on the back of the TR base, use the 10-mm-deep M4 (10-mm-deep M5 for LS-7070M) mounting holes on the back.
Without TR Base
T
ake the following steps to use the transmitter and receiver (LS-7030M in this case) after dismounting them from the TR base. same way.
LS-7010M and LS-7070M
can also be used in the
1 Loosen the bolts on the back of the
TR base and remove the measuring head.
1-12
2 Mount the measuring head.
Use the side mounting holes on the transmitter and receiver as shown in the illustration. Mount them with the bolts provided with the package.
Page 27
Use the 5-mm-deep M4 bottom mounting holes of the transmitter and receiver as shown in the illustration.
Chapter 1 Getting Started
1
When Using the Dual Head
Top
Function
When using the dual head function, mount
the two measuring heads so that their top
sides face each other.

Mounting the Controller

Mounting Restrictions
Be aware of the following items when mounting the controller.
CAUTION
Note: • For the maintenance, operability, and ventilation of the LS-7600
• Do not install the controller upside down.
• Do not block the ventilation louver. Otherwise, the interior heat may cause the controller to malfunction.
Series, separate the controller as far as possible from periph­eral structures or parts.
• The mounting angle must abide by the following restrictions.
30˚
Side
Display panel
• Check that the controller will not be exposed to the heat radia­tion of peripheral devices.
• Separate the controller as much as possible from devices with arc generation, such as electromagnetic switches or non-fuse breakers.
1-13
Page 28
1
Chapter 1 Getting Started
Mounting to Panel
The controller can be mounted to control panels with the mounting brackets provided with the package. Arrange the following panel mounting dimensions.
207
+1
169
0
+1
124
0
200
Unit: mm
1 Mount the controller from the front
side of the panel.
1-14
2 Attach resin caps to the mounting
brackets.
3 Insert the mounting brackets to the
four mounting holes on the side of the controller respectively.
Mounting holes
Mounting holes
Page 29
4 Secure the mounting brackets with a
Phillips screwdriver.
Chapter 1 Getting Started
1
CAUTION
• Tighten the mounting screws to a maximum torque of 0.4 Nm. Do not tighten up the mounting screws excessively, otherwise the panel or controller casing may be deformed.
• Do not remove the gasket, otherwise the controller will not satisfy IP64 conditions.
Mounting to AC Power Supply Stand
The following section provides information on how to mount the controller to the AC power supply stand (sold separately). The AC power supply stand consists of a stand unit and base unit. Mount the controller to the stand unit first. Connect the power supply cable, extension cable (i.e., the cable between the controller and measuring head), and camera cable. Then mount them to the base unit.
1 Remove the gasket around the
controller.
Gasket
2 Mount the controller to the stand.
Use the two screws provided with the package and secure the controller on the side.
The stand must be located on the groove from which the gasket was removed.
Secure these portions with screws.
1-15
Page 30
1
Chapter 1 Getting Started
3 Connect the cable between the con-
troller and measuring head.
If only a single measuring head is used, connect the cable to the HEAD1 connec­tor.
4 Connect the camera cable to the
controller.
If only a single measuring head is used, connect the cable to the CAMERA1 connector.
5 Connect the power supply cable of
the AC power supply stand to the controller.
Cable between controller and measuring head
Camera cable
6 Mount the stand unit to the base unit.
Mount the stand unit with the two screws provided with the package.
(1) Lean the stand unit in the front
direction and insert the tab.
1-16
(2) Lean the stand unit in the direction
indicated by the arrow. Pay utmost attention so that the connected cable will not be caught by the stand unit.
(3) Secure these portions
with screws.
Page 31
Chapter 1 Getting Started
Tips
The bottom of the AC power supply stand has legs. Erect the legs and use the AC power supply stand if required.

Connection

Connect the measuring head, controller, remote control console, and power supply with the cables.
CAUTION
Make the following five connections.
• Transmitter and receiver: Connect the cables from the transmitter and receiver respec­tively.
• Measuring head and controller: Connect the cable between the controller and measuring head.
• Camera unit and controller: Connect the camera cable.
• Remote control console and controller: Connect the remote control console.
• Power supply cable: Supply 24 VDC to the controller directly or through the AC power supply stand.
Do not supply power to the controller while connecting the cables. If the AC power supply stand is used, turn off the AC power supply stand while connecting the cables. The measur­ing head, controller, or peripheral devices may be damaged if power is supplied while connecting the cables.
1
KZ-U3 24-VDC Power
Supply (Optional)
Power supply cable to the AC power supply stand
Camera cable (LS-C3AM/LS-C10AM /LS-C30AM)
Up to two cables are connectable, provided that the total length of the cables does not exceed 40 m.
LS-7601
Remote control console
Measuring head (LS-7010M/LS-7030M/LS-7070M)
Cable between transmitter and receiver
(OP-42182/OP-42183) (Optional)
Cable between controller and measuring head (LS-C3A/LS-C10A/LS-C30A) Up to two cables are connectable, provided that the total length of the cables does not exceed 40 m.
Measuring head (LS-7010M/LS-7030M/LS-7070M)
1-17
Page 32
1
Chapter 1 Getting Started
Transmitter and Receiver
Connect the cables of the transmitter and receiver respectively.
Connect both connectors so that the notched part of each connector will coincide with each other in position.
Measuring Head and Controller
Connect the cables of the controller and the measuring head.
Connect both connectors so that the notched part of each connector will coincide with each other in position.
Transmitter side
Receiver side
Camera Unit and Controller
Connect the camera cable to the camera unit cable.
Connect both connectors so that the notched part of each connector will coincide with each other in position.
Power Supply Cable
Supply 24 VDC to the controller through I/O terminals 1 and 2 on the rear panel of the controller. Secure the wires to the terminals with a Phillips screwdriver.
Using AC Power Supply Stand (Sold
Separately): Connect the output cable of the AC
power supply stand to the DC IN connector of the controller. Then connect the AC cord provided with the AC power supply stand to the electric outlet.
1-18
24 VDC
Controller side
Controller side
Measuring head side
Measuring head side
Page 33
Chapter 1 Getting Started

Turning AC Power Supply Stand On and Off

This section provides information on how to turn the AC power supply stand on and off.
Turning Power On
1 Turn on the POWER switch on the
rear panel of the AC power supply stand.
Turning Power Off
1 Turn off the POWER switch on the
rear panel of the AC power supply stand.
Adjustments to LCD
1
1-19
Page 34
Chapter 1 Getting Started

Adjustments to LCD Monitor

1
The color, hue, and brightness adjustments to the LCD monitor. The LCD monitor is fac­tory-adjusted. Therefore, make these adjustments only if they are required according to the operating conditions.
Tips
The LCD monitor will be reset to the initial state by setting each trimmer to the center position.
Color Adjustments
Turn the COLOR adjustment trimmer.
The color will be lighter by turning the trimmer counterclockwise, and deeper by turning the trimmer clockwise.
Light
Deep
Hue Adjustments
Turn the HUE adjustment trimmer.
The monitor screen will be more greenish by turning the trimmer counterclockwise, and more reddish by turning the trimmer clock­wise.
Greenish
Reddish
Brightness Adjustments
Turn the BRIGHT adjustment trimmer
The monitor screen will be brighter by turning the trimmer counterclockwise, and darker by turning the trimmer clockwise.
1-20
Bright Dark
Page 35

Measurement and Setting Overview

Chapter 1 Getting Started
The LS-7600 Series has two modes (i.e., the RUN mode and PROGRAM mode). Program settings (measuring condition settings) and environment settings (basic settings) can be made in the LS-7601 in PROGRAM mode.
Power On
RUN mode
Measures according to the set program.
Refer to Section 4 Operation Control on page 4-1.
RUN
PROG
RUN
or the [Slide switch] key operation on the console.
PROG
Program Settings
Settings for measurement are registered as programs. The following settings are possible.
• Tolerance Setting Set the tolerance range of the measured value.
• Area Setting Set the measuring part.
• Compensation Setting Set the compensation detail of errors in measurement.
• Output Setting Set the processing and output details of the measured value.
• Option Setting Set the display unit and other items.
Refer to Section 5 Function Setting on page 5-1.
1
Press the [ESC] key for 2 seconds.
[ESC] key
Environment Settings
Make basic settings for the operation of the LS-7600 Series.
Refer to Section 6 Environment Settings on page 6-1.
The LS-7600 Series will be reset to the initial status by turning the LS-7600 Series on with the [Zero] key pressed. (Page 1-22)
1-21
Page 36
Chapter 1 Getting Started

Resetting to Initial Status

1
The following section provides information on how to reset all the programs and environ­ment settings to factory-set values.
Tips
Refer to Resetting All Programs on page 6-6 if the resetting of only the pro­grams to factory-set values is required.
1 Turn on the LS-7600 Series with the
[Zero] key pressed.
A confirmation message for resetting will appear on the LCD monitor screen.
2 Press the [Left or Right] key and
select Y.
Initialize setting?
(Y/N)
3 Press the [ENT] key.
All the programs and environment settings are reset and the measuring or setting screen will appear.
1-22
Page 37
Chapter 2
Basic Operation of the LS-7600 Series
This section provides information on the basic operation of the LS-7600 Series, panel keys, and remote control console.
Control Panel Functions and Nomenclature ........................ 2-2
Names and Functions of the LCD Monitor Screen............... 2-4
2-1
Page 38
2
Chapter 2 Basic Operation of the LS-7600 Series

Control Panel Functions and Nomenclature

The LS-7600 Series is operated with the panel keys and remote control console. The LCD monitor will display the result of measurement and necessary data for setting.

Panel Key Functions

[SCREEN] key
[STATS] key
Press this key to use the statistical processing function. (Page 5-66, Section 5-10 Group Processing)
PROGRAM No.
Press this key to change the screen display. (Page 4-2, Section 4-1 Switching Display
SCREEN
STATS
ZERO
POWER
RUN
PROG
ESC
PRINT
ENT
HOLD
LS-7601
Screen)
[RUN/PROG] key
Press this key to change the LS­7600 Series between RUN mode and PROG mode.
[PROGRAM No.] key
Press this key to change the program number. (Page 4-12, Section 4-6 Program Changes)
[ESC] key
Press this key to quit or abort the set item. When in the Run mode, this key has the same function as the RESET input terminal on the rear panel.
[Cursor] key
Press this key to move the cursor or make set value changes.
[PRINT] key
Press this key to print out the screen display.
Cursor Keys
Four cursor keys (i.e., the Up, Down, Left, and Right keys) are available. Each key is used for the following purposes.
•To move between set items.
•To input numeric values, such as toler­ance values.
2-2
[ZERO] key
Press this key to use the automatic zero function. The key has the same function as the ZERO input terminal of the rear panel. (Page 4-10, Section 4-4 Automatic Zero Function)
[ENT] key
Press this key to input the set value change.
[HOLD] key
Press this key to hold the measured value. The key has the same function as the TIMING input terminal on the rear panel. (Page 4-9, Section 4-3 Hold Function)
Moving up
Moving rightMoving left
Moving down
Page 39
Chapter 2 Basic Operation of the LS-7600 Series

Remote Control Console Functions and Nomenclature

[ENT] key
Use this key to move the cursor or input the set value change.
[ESCAPE] key
Press this key to quit or abort the set item. When in the Run mode, this key has the same function as the RESET input terminal on the rear panel.
[SCREEN] key
Press this key to change the screen display. (Page 4-2, Section 4-1 Switching Display Screen)
ENT Key
Pressing Key at Center Position
• To select menu items.
• To input entered values.
ENT
ESCAPE
SCREEN HOLD
ZERO
POWER
RUN
PROG
[Mode selection slide switch] key
Slide this key upward or downward to set the LS-7600 Series to RUN mode or PROG mode.
[HOLD] key
Press this key to hold the measured value. The key has the same function as the TIMING input terminal on the rear panel. (Page 4-9, Section 4-3 Hold Function)
[ZERO] key
Press this key to use the automatic zero function. The key has the same function as the ZERO input terminal of the rear panel. (Page 4-10, Section 4-4 Automatic Zero Function)
Press
2
Moving Key in Arrow Directions
Press the key upward, downward, left, or right for either of the following purposes.
• To move between items.
• To input numeric values, such as toler­ance values.
Moving up
Moving rightMoving left
Moving down
2-3
Page 40
2
Chapter 2 Basic Operation of the LS-7600 Series

Names and Functions of the LCD Monitor Screen

This section describes the contents displayed on the LCD monitor screen. For the opera­tion method, refer to Switching Display Screen on page 4-2.
Single Display
Camera number
Application mode display
OUT number display Displays the current OUT number.
Statusdisplay
Target viewer display Displays the image of the measuring target.
Target Viewer
Function on page 4-7
Tolerance display
Displays the preset tolerance values. Tolerance (LIMIT) Settings on page 5-49
display Displays the camera number whose image is on the target viewer.
OUT2
DUAL
HOLDZEROSTATS

23.45670
   UPPER 1.00000 STANDARD23.00000 LOWER –1.00000  
Date and time display
page 6-1
Mode display Displays the current mode (measurement or setting).
CAM1CAM2
RUN
P – 00
GO
01/02/18 17:50:42
Environment Settings on
Page 1-21
Program number display Displays the current program number.
Page 4-12, 5-1
Measured value display Displays the measurement result.
Judgment result display Displays the result judged with the preset limits.
Both Value Display
Target viewer display
Applicationmode display
Measured value display
Tolerance display
Measurement mode display Displays the preset measurement mode.
Page 5-33
Tips
The both-value display screen can display trend graphs or statistical process-
ing. Refer to Switching Display Screen on page 4-2.
2-4
Status display
Camera number OUT number
OUT1HOLDZEROSTATS DL
12.34560
 UP 
1.00000  STD  LO   NORMAL 512  OUT2
23.45670
 UP   STD  LO   NORMAL 512
12.00000
–1.00000
1.00000
23.00000
–1.00000
Averaging number display Displays the preset number of averaging.
Page 5-31
CAM1CAM2
RUN
P−00
AREA1  HEAD1  DIA AREA2  HEAD1  DIA
A12 A22
01/02/18 17:50:42
Mode display
Program number display
F
Area settings display
F
Area Settings on page 5-9
Focus monitor display
Page 5-16
Edge count display Displays the number of all the edges recognized by each area.
Page 5-9
Page 41
Chapter 3
Easy Setting Guide
The LS-7600 Series is used for a variety of measurement applications if the LS-7600 Series is set properly. This section provides information on targets and settings for typical applications respectively.
Measurement of Outer Diameter and Width
(with Single Measuring Head) ............................................. 3-2
Measurement of Outer Diameter and Width
(with Two Measuring Heads) ............................................... 3-3
Measurement of Outer Diameter and Width
(with Two Measuring Heads for Larger Objects) ................. 3-4
Measurement of Inner Diameter and Clearance ................. 3-5
Measurement of Outer Diameter and Eccentricity .............. 3-6
Measurement of Movement and Positioning ....................... 3-7
Measurement of Pitch ......................................................... 3-8
Group Processing ............................................................... 3-9
Statistical Processing ........................................................ 3-10
3-1
Page 42
Chapter 3 Easy Setting Guide

Measurement of Outer Diameter and Width (with Single Measuring Head)

In the following example, a single measuring head is used to measure the outer diameter of an object in one direction. The result of measurement is displayed as OUT1.
3
Set Values
Set item
Area setting
Output
setting
Tolerance setting
AREA
Calibration
Operation
Averaging
RUN mode
Offset
Reference
Area Settings (Page 5-9) Calibration Settings (Page 5-22) Output Settings (Page 5-29) Tolerance Settings (Page 5-49)
D
Measures the outer diameter in one direction.
Set value
AREA 1
Head 1 DIA Make settings if required.
OUT 1 OUT 2 A1: +1 A2: 0
Select the value according to the line speed. Normal No settings are required. Set the upper and lower limit values.
No settings are required.
No settings are required.
AREA 2
3-2
Page 43
Chapter 3 Easy Setting Guide

Measurement of Outer Diameter and Width (with Two Measuring Heads)

In the following example, two measuring heads are used to measure the average outer diameter of an object in two directions. The measurement of the object in two directions ensures high accuracy. The result of measurement is displayed as OUT1.
Y
X
X+Y
D=
2
Measures the outer diam­eter in two directions.
3
Set Values
Set item
Area setting
Calibration
Output
setting
Tolerance setting
AREA
Operation
Averaging
RUN mode
Offset
Reference
Area Settings (Page 5-9) Calibration Settings (Page 5-22) Output Settings (Page 5-29) Tolerance Settings (Page 5-49)
Set value
AREA 1
Head 1 DIA Make settings if required.
OUT 1 OUT 2 A1: +1/2 A2: +1/2
Select the value according to the line speed. Normal No settings are required. Set the upper and lower limit values.
Head 2 DIA
No settings are required.
AREA 2
3-3
Page 44
3
L= (X+Y)
Chapter 3 Easy Setting Guide

Measurement of Outer Diameter and Width (with Two Measuring Heads for Larger Objects)

A sheet cannot be located in the measuring area of a single measuring head if the width or diameter of the sheet is excessively large. In that case, the sheet can be measured with two measuring heads as shown below with the automatic zero function and the master workpiece used. The result of output is displayed as OUT1.
When measuring the width of a sheet
Set Values
Reference
3-4
L
diameter of a round bar
L=  (X+Y)
When measuring the outer
Set item
Option setting
Area setting
Calibration
Output
setting
Tolerance setting
Application mode
AREA
Operation
Averaging
RUN mode
Offset
No settings are required. Make settings if required.
A1: +1 A2: +1
Select the value according to the line speed. Normal Input the size of master workpiece. Set the upper and lower limit values.
AREA 1
OUT 1 OUT 2
Set value
DUAL
AREA 2
No settings are required.
No settings are required.
Automatic Zero Function: Measure the master workpiece and make automatic zero settings. (Page 4-10) Option Settings (Page 5-55) Area Settings (Page 5-9) Calibration Settings (Page 5-22) Output Settings (Page 5-29) Tolerance Settings (Page 5-49)
Page 45
Chapter 3 Easy Setting Guide

Measurement of Inner Diameter and Clearance

In the following example, the inner diameter of the workpiece is measured while the workpiece moves. The maximum value measured is taken as the inner diameter of the workpiece. The result of output is displayed as OUT1.
3
Set Values
Set item
Area setting
Calibration
Output
setting
Tolerance setting
AREA
Operation
Averaging
RUN mode
Offset
Reference
Area Settings (Page 5-9) Calibration Settings (Page 5-22) Output Settings (Page 5-29) Tolerance Settings (Page 5-49)
Set value
AREA 1
Head 1 SEG (+1E, +2E)
Make settings if required.
OUT 1 OUT 2 A1: 1 A2: 0
Select the value according to the line speed.
Peak hold
No settings are required. Set the upper and lower limit values.
No settings are required.
No settings are required.
AREA 2
3-5
Page 46
Chapter 3 Easy Setting Guide

Measurement of Outer Diameter and Eccentricity

In the following example, the outer diameter and eccentricity of the roller is measured while the roller moves. The outer diameter is displayed as OUT1, and the eccentricity is dis­played as OUT2.
3
Set Values
Set item
Area setting
Output
setting
Tolerance setting
AREA
Calibration
Operation
Averaging
RUN mode
Offset
Reference
Area Settings (Page 5-9) Calibration Settings (Page 5-22) Output Settings (Page 5-29) Tolerance Settings (Page 5-49)
Eccentricity Outer diameter
Set value
AREA 1
Head 1 DIA Head 1 T-EDGE
Make settings if required.
OUT 1 OUT 2 A1: 1 A2: 0 1 Average hold
No settings are required. No settings are required. Set the upper and lower limit values.
A1: 0 A2: 1 Select the value according to the rpm. Peak-to-peak hold
AREA 2
3-6
Page 47
Chapter 3 Easy Setting Guide

Measurement of Movement and Positioning

In the following example, the movement or position of the workpiece is measured. The result of output is displayed as OUT1.
3
Set Values
Set item
Area setting
Calibration
Output
setting
Tolerance setting
AREA
Operation
Averaging
RUN mode
Offset
Reference
Area Settings (Page 5-9) Calibration Settings (Page 5-22) Output Settings (Page 5-29) Tolerance Settings (Page 5-49)
Set value
AREA 1
Head 1 SEG.
(0E, +2E)
Make settings if required.
OUT 1 OUT 2 A1: 1 A2: 0
Select the value according to the line speed. Normal No settings are required. Set the upper and lower limit values.
No settings are required.
No settings are required.
AREA 2
3-7
Page 48
3
Chapter 3 Easy Setting Guide

Measurement of Pitch

The single pitch of connectors pins or IC pins is measured by the LS-7600 Series. Further­more, multiple pitch of pins is measured with the program function or RS-232C interface used.
The center pitches of the two portions are measured at a time.
Set Values
Reference
3-8
Set item
Area setting
Calibration
Output
setting
Tolerance setting
AREA
Operation
Averaging
RUN mode
Offset
Head 1 SEG
(+2P, +4P)
Make settings if required.
A1: +1 A2: 0
Select the value according to the line speed.
Sample hold 1/2
No settings are required. Set the upper and lower limit values.
RS-232C (Page 8-1) Area Settings (Page 5-9) Calibration Settings (Page 5-22) Output Settings (Page 5-29) Tolerance Settings (Page 5-49)
Set value
AREA 1
Head 1 SEG
(+4P, +6P)
OUT 1 OUT 2
A1: 0 A2: +1
Select the value according to the line speed.
Sample hold 1/2
No settings are required.
AREA 2
Page 49
Chapter 3 Easy Setting Guide

Group Processing

In the following example, the average, maximum, or minimum value is calculated as criteria from the measuring data on the group of targets. The results of output are displayed as OUT2.
Other example
3
The average measured value of the 12 shafts is used as criteria.
Set Values
Set item
Area setting
Calibration
Output
setting
Tolerance setting
Option setting
AREA
Operation
Averaging
RUN mode
Offset
GROUP
Head 1 DIA Make settings if required.
A1: +1 A2: 0
Select the value according to the line speed.
Sample hold
No settings are required. Set the upper and lower limit values.
12 pcs
AREA 1
OUT 1 OUT 2
Reference
Area Settings (Page 5-9) Calibration Settings (Page 5-22) Output Settings (Page 5-29) Tolerance Settings (Page 5-49) Option Settings (Page 5-55)
The average measured value of portions A through D is used as criteria.
Set value
AREA 2
No settings are required.
Setting impossible
Setting impossible Average Setting impossible
3-9
Page 50
3
Chapter 3 Easy Setting Guide

Statistical Processing

In the following example, the values calculated from the measuring data are: Maximum value, minimum value, average value, maximum and minimum values, standard deviation, N number, HH number, HI number, GO number, LO number, and LL number.
The outer diameters of 96 shafts sampled are measured to calcu­late the values.
Set Values
Set item
Area setting
Output
setting
Tolerance setting
Option setting
AREA
Calibration
Operation
Averaging
RUN mode
Offset
STATS
Reference
Area Settings (Page 5-9) Calibration Settings (Page 5-22) Output Settings (Page 5-29) Tolerance Settings (Page 5-49) Option Settings (Page 5-55)
AREA 1
Head 1 DIA Make settings if required.
OUT 1 OUT 2 A1: +1 A2: 0
Select the value according to the line speed.
Sample hold
No settings are required. Set the upper and lower limit values.
96 pcs
Set value
AREA 2
No settings are required.
No settings are required.
3-10
Page 51
Chapter 4
Operation Control
This section provides information on setting control items on the control panel while the LS-7600 Series is in measurement operation.
Switching Display Screen............................................ 4-2
Target Viewer Function ............................................... 4-7
Hold Function .............................................................. 4-9
Automatic Zero Function ............................................4-10
Key Lock Function......................................................4-11
Program Changes ......................................................4-12
4-1
Page 52
Chapter 4 Operation Control

Switching Display Screen

The following section provides information on how to display an OUT1 or OUT2 measured value or both on the LCD monitor. Furthermore, it provides information on trend graph, statistical processing and NG RECORD display. The LCD monitor is operable through the panel keys as well as the remote control console.

Measured Value Display

The OUT1 or OUT2 measured value or both can be displayed at a time. If both values are displayed at a time, the trend graph or statistical processing of the values can be displayed.
Switching between Single Value Display and Both Display
4
4-2
The RUN mode display will change in sequence whenever the [ENT] key (Up or Down key) is pressed.
OUT1 Single Display
The measured values vary between the set LIMITS.
OUT1
12.34560
UPPER 1.00000 STANDARD LOWER –1.00000
12.00000
CAM1
RUN
P – 00
GO
01/02/18 17:50:42
▲ ▼
OUT2 Single Display
OUT2
23.45670
UPPER 1.00000 STANDARD LOWER –1.00000
23.00000
CAM1
RUN
P – 00
GO
01/02/18 17:50:42
▲ ▼
Both Value Display
OUT1
12.34560
UP
1.00000
STD
12.00000
LO
–1.00000
NORMAL 512
OUT2
23.45670
UP
1.00000
STD
23.00000
LO
–1.00000
NORMAL 512
CAM1
RUN
P – 00
AREA1
HEAD1 DIA
AREA2
HEAD1 DIA
A1 2 A2 2
01/02/18 17:50:42
Page 53
Chapter 4 Operation Control
Switching between Measured value, Trend Graph, and Statistical Processing Display
1 Press the [Up or Down] key and
change the single measured value display to both value display screen.
2 Press the [Up or Down] key and select
the OUT1 (upper part) or OUT2 (lower part) or Both values.
3 The selected item changes as shown
below whenever the [SCREEN] key is pressed.
OUT1 measured value
OUT1 trend graph
OUT1 statistical processing
OUT2 measured value
OUT2 trend graph
OUT2 statistical processing
If either OUT1 or OUT2 is selected, the display for only the selected one will be displayed.
OUT1
12.34560
UP
1.00000
STD
12.00000
LO
–1.00000
NORMAL 512
OUT2
23.45670
UP
1.00000 STD 23.00000 LO
–1.00000
NORMAL 512
▲ ▼
OUT1
12.34560
13.10000
11.90000 1M
OUT2
23.45670
24.10000
22.90000 1M
▲ ▼
OUT1
8.55013
8.57665
MAX
8.53745
MIN
8.55013
AVE DIFF
0.03920
SD
0.014496
N
100
OUT2
STAT S
.
MAX MIN AVE DIFF SD N
CAM1
RUN
P – 00
AREA1
HEAD1 DIA
AREA2
HEAD1 DIA
A1 2 A2 2
01/02/18 17:50:42
CAM1
RUN
P – 00
AREA1
HEAD1 DIA
AREA2
HEAD1 DIA
A1 2 A2 2
01/02/18 17:50:42
CAM1
RUN
P – 00
AREA1
HEAD1 DIA
AREA2
HEAD1 DIA
A1 2 A2 2
01/02/18 17:50:42
4
Note: Statistical processing display will be displayed when settings for statistical
processing are made. For details, refer to Statistical Processing on page 5-66.
4-3
Page 54
Chapter 4 Operation Control
Trend Graph Display
Trend graph can be displayed when both outputs are displayed.
Trend Graph Display
4
OUT1
23.45670
24.10000
22.90000 1M
OUT2
10.00015
11.10000
8.90000 1M
CAM1
RUN
P – 00
AREA1
HEAD1 DIA
AREA2
HEAD1 DIA
A1 2 A2 2
01/02/18 17:50:42
The Y-axis of trend graph is automatically set for the next set value and displayed
according to the tolerance set value.
• Reference value mode: (Reference value –A) to (Reference value +A) A = (|Lower limit value| or |Upper limit value|, which ever is larger) x1.1
• Threshold mode: HH value to LL value
A tolerance position is indicated with a yellow line.
• Reference value mode: Reference value + Upper limit and Reference value + Lower limit
• Threshold mode: HI value and LO value
The X-axis of a trend graph is set as graph TIME data. Refer to Option Settings on
page 5-55 for details.
Setting
1m 10m 30m
1H 2H
6H 12H 24H
Single-screen time
1 min 10 min 30 min
1 hr 2 hr
6 hr 12 hr 24 hr
Number of points
120 points 120 points 120 points 120 points 120 points 120 points 120 points 120 points
Refresh time
0.5 s 5 s
15 s
30 s 1 min 3 min 6 min
12 min
4-4
A line is drawn between the maximum and minimum values of all the measuring
points in the refresh time and displayed as one-plot data in the X-axis direction.
A trend graph has 50 points in the Y-axis direction and 120 points in the X-axis
direction.
The screen is scrolled to the left to continue when a single-screen data is displayed.The data scrolled to the left cannot be returned and displayed again.
Page 55
Chapter 4 Operation Control
Note: Graph data will be cleared in the following cases.
• The LS-7600 Series is turned off.
• The program number is changed, copied, or initialized.
•Graph TIME data is changed.
• The unit or minimum display unit is changed. The measured value will be displayed gray when the LS-7600 Series stands by. If the measured value exceeds the range of the Y-axis of the graph, the value will be displayed as the maximum or minimum value in the Y-axis direc­tion.
Statistical Processing Display
The statistical processing screen can be displayed if both values are selected for display. For details, refer to Statistical Processing on page 5-66.
CAM1
RUN
P – 00
AREA1
HEAD1 DIA
AREA2
HEAD1 DIA
A1 2 A2 2
01/02/18 17:50:42
Statistical Processing
Tips
Statistical processing results can be printed as hard copies for storage.
OUT1
OUT2
8.55013
8.57665
MAX
8.53745
MIN
8.55013
AVE DIFF
0.03920
SD
0.014496
N
100
STATS
.
MAX MIN AVE DIFF SD N

NG RECORD Display

Up to 20 latest errors in measurement data are saved. The error data items are displayed in chronological order (with the year, month, date, hour, minute, and second for each error data item). Each error data item includes the program number, OUT number, measured value, and criterion.
4
Switching NG RECORD Display
1 Press the [SCREEN] key for two seconds while the measured value is dis-
played.
NG RECORD items 1 through 10 are displayed. Press the [Down] key to display items 11 through 20.
4-5
Page 56
Chapter 4 Operation Control
2 Press the [SCREEN] key for 2 s to return to the measured value display
screen. NG RECORD Display
NG RECORD
NO
TIME RESULT
P
1 2 3 4 5 6 7 8 9 10
01/01/01 12:00:00
01/01/01 11:55:00
01/01/01 11:43:00
01/01/01 11:39:00
01/01/01 11:15:00
01/01/01 10:53:00
01/01/01 10:47:00
01/01/01 10:32:00
01/01/01 10:32:00
01/01/01 09:37:00
OUT1 P
OUT1 P
OUT2 P
OUT2 P
OUT2 P
OUT2 P
OUT1 P
OUT1 P
OUT2 P
OUT2
00
8.61150 HI
00
8.60895
1.50140
1.49645
1.49610
1.49630
8.60435
8.60460
1.50270
1.50160
HI HI HI HI HI HI HI HI HI
HI
00
00
00
00
00
00
00
00
4
Tips
The NG RECORD can be printed. The NG RECORD will not be lost after the LS-7600 Series is turned off.
Note: • All NG RECORD data will be lost if the programs are all cleared. Refer to
Initialization of All Program Settings on page 6-6. If the present number of NG RECORD data items held by the LS-7600 Series is 20, the oldest one will be deleted when the next NG RECORD data item is input.
• The display is not automatically updated. To update the display, return to the measured value display and then redisplay the information.
NG Data
The following NG data items will be put on record.
RUN mode NG Data Item
Normal
Automatic peak hold
Automatic bottom hold
Automatic peak-to-peak hold
Peak hold
Bottom hold
Peak-to-peak hold
Average hold
Sample hold 1
Sample hold 2
Self-timing hold
A measured value held by TIMING input, the hold key, or RS-232C will be put on NG record if the measured value is not good (NG). The value will not be put on record unless it is held.
A measured value held by TIMING input, the hold key, or RS-232C will be put on NG record if the measured value is not good (NG).
A value that has been measured and entered will be put on NG record if the value is not good (NG).
4-6
Page 57
Chapter 4 Operation Control

Target Viewer Function

A visual check on the image of each target is possible at a glance through the built-in image sensor camera of the measuring head.
Target
AREA1 Cursor
Displays the set measurement range in AREA1.
Scanning Line
Measures the dimensions of targets arranged in line.
Center of measuring area
Displays the center of the measuring area.
OUT1
12.34560
UPPER 1.00000 STANDARD LOWER –1.00000
Clearance (AREA1)
Outer diameter (AREA2)
Measuring point
CAM1
RUN
P – 00
12.00000
01/02/18 17:50:42
AREA2 Cursor
Displays the set measurement range in AREA2.
Refer to Area Settings on page 5-9 for area settings in detail.
Note: • The display screen will be all black if the camera cable is not connected. The
scanning line, AREA1 cursor, AREA2 cursor, and center of the measuring area will be, however, displayed.
• The display position of the scanning line varies for a few pixels to the left or right with the measuring head.
• The image is refreshed 60 times a second. The cursor is refreshed 10 times a second.
4
4-7
Page 58
Chapter 4 Operation Control

Switching Cameras

The CAMERA1 display on the upper right-hand side of the LCD monitor indicates the image sensor camera of measuring head 1. The CAMERA2 display indicates the image sensor camera of measuring head 2.
4
Single Value Display
Press the [SCREEN] key and select CAMERA1, CAMERA2, or both CAMERA1 and CAMERA2.
Both Value Display
1 Press the [Down] key so that the
screen selection cursor will be off.
2 Press the [SCREEN] key and select
CAMERA1, CAMERA2, or both CAMERA1 and CAMERA2.
OUT1
12.34560
UPPER 1.00000 STANDARD LOWER –1.00000
OUT1
12.34560
UP STD LO
NORMAL 512
OUT2
23.45670
UP STD LO
NORMAL 512
1.00000
12.00000 –1.00000
1.00000
23.00000 –1.00000
12.00000
CAM1 CAM2
RUN
P – 00
01/02/18 17:50:42
CAM1 CAM2
RUN
P – 00
AREA1
HEAD1 DIA
AREA2
HEAD1 DIA
A1 2 A2 2
01/02/18 17:50:42
4-8
CAMERA1 and CAMERA2 can be selected at the same time when two measuring heads are connected. In the simultaneous display of CAMERA1 and CAMERA2, the upper half of the screen shows the CAMERA1 area cursor, and the lower half of the screen shows the CAMERA2 area cursor. The measuring target is displayed in black instead of its actual image.
Displays CAMERA2
Area cursors
Displays CAMERA1
Page 59
Chapter 4 Operation Control

Hold Function

Measured values can be held and displayed. Furthermore, MEA-Mode combinations allow the measurement of a variety of values, such as maximum and minimum values. The types of measured values held vary with the MEA-Mode. Refer to MEA-Mode on page 5-33 for details.
Measured values are held and displayed through the following functional components.
• Control panel
• Remote control console
• External I/O terminals
• RS-232C interface
Control Panel/Remote Control Console
If either the OUT1 or OUT2 measured value is displayed, the displayed value will be held. If both OUT1 and OUT2 measured values are displayed, both values will be held.
4
Press the [HOLD] key.
If both values are displayed, the OUT number selected on the screen will be kept on hold. The display “HOLD” will appear on the screen. To cancel the HOLD status, press the [HOLD] key again.
Tips
By pressing the [ESC] key, the measured value kept on hold will be reset.
Display position
OUT1
HOLD
0.00000
UPPER 1.00000 STANDARD LOWER –1.00000
12.00000
External I/O Terminals
Measured values will be held by short-circuiting the timing input terminal and COM terminal together on the rear panel of the controller. Refer to Section 7 I/O Terminals on page 7-1 for the external input terminals to be used.
RS-232C Interface
Commands sent by external devices through the RS-232C interface make it possible to put measured values on hold. Refer to Section 8 RS-232C on page 8-1 for the com­mands in detail.
CAM1
RUN
P – 00
01/02/18 17:50:42
4-9
Page 60
Chapter 4 Operation Control

Automatic Zero Function

The automatic zero function sets measured values to zero (0.00000) instantly. This function is convenient for zero-point calibration for a variety of targets. Furthermore, master calibra­tion is possible by using this function in combination with the offset function.
4
Note: This function will not be available while the controller is awaiting the result of
Tips
Automatic zero settings are made through the following functional components.
• Control panel/Remote control console
• External terminal
• RS-232C interface
measurement, during which the LCD monitor will display “– – –. – – – –.”)
The automatic zero value will not be lost after the LS-7600 Series is turned off. The automatic zero value for each program will be stored.
Control Panel/Remote Control Console
If a single measured value is displayed, an automatic zero setting will be made for the corresponding OUT number only. If both measured values are displayed, an automatic zero setting will be made for the selected OUT number on the screen.
1 Press the [ZERO] key once. Automatic
zero will be set.
To cancel the setting, press the [ZERO] key again.
OUT1
ZERO
0.00000
UPPER 1.00000 STANDARD LOWER –1.00000
10.00000
CAM1
RUN
P – 00
01/02/18 17:50:42
External I/O Terminal
Automatic zero will be set by short-circuiting the ZERO, and COM terminals on the rear panel of the controller together. Refer to Section 7 I/O Terminals on page 7-1 for details.
Note: Automatic zero cannot be canceled through the external I/O terminals.
RS-232C Interface
Commands sent by external devices through the RS-232C interface make it possible to set or cancel the automatic zero function. Refer to Automatic Zero ON/OFF on page 8­11 for details.
Tips
Master Calibration with Offset Function (Automatic Offset Function)
Set the size of a master workpiece to an offset value and make an automatic zero setting while measuring the master workpiece. Then the size of the master workpiece will be displayed as an offset value. Refer to Measurement of Outer Diameter and Width (with Two Measuring Heads) on page 3-4 for an application example. Refer to Offset on page 5-43 for the offset function in detail.
4-10
Page 61
Chapter 4 Operation Control

Key Lock Function

This function locks the control panel and remote control console, thus disabling key input. As a result, there will be no risk of making setting changes with careless key input.
Key Lock Types
The following types of panel locks are possible with the KEY LOCK settings as environ­ment settings. Refer to Key Lock on page 6-13 for details.
Set the key lock type to ALL: All keys excluding any key for the display change function
Set the key lock type to PART: All keys excluding any key for the display change function
Tips
Key lock settings are made through the following functional components.
• Control panel
• RS-232C interface
The key lock status will not be lost after the LS-7600 Series is turned off or program number changes are made.
will be locked.
and tolerance setting function will be locked. Select this type if you make tolerance changes frequently without changing any other setting.
Control Panel
Press the Left and Right keys on the control panel together for 2 s. The LCD monitor will display “KEYLOCK.”
To cancel the key lock status, repeat the above.
OUT1
12.34560
UPPER 1.00000 STANDARD LOWER –1.00000
Display position
12.00000
CAM1
RUM
P – 00
KEY-LOCK 01/02/18 17:50:42
4
RS-232C Interface
Commands sent by external devices through the RS-232C interface make it possible to set or cancel the key lock function. Refer to Key Lock ON/OFF on page 8-12 for details.
4-11
Page 62
4
Chapter 4 Operation Control

Program Changes

The following section provides information on how to call 16 programs saved in the LS­7600 Series.
Names and Number of Programs
The selected program is displayed on the LCD monitor next to the [PROGRAM No.] key. Programs 0 through 15 (i.e., 16 programs in total) can be registered.
T1
3.45670
PER 1.00000
ANDARD 10.00000
WER –1.00000
Programs are selected through the following functional components.
• Control panel
• External I/O terminals
• RS-232C interface
CAM1
RUN
P – 00
01/02/18
Display position
Note: The operation method varies with the P-SELECT setting as an environmental
setting. Refer to Program on page 6-5 for environment settings in detail.
P-SELECT setting Operation method
PANEL TERMINAL
Control panel/RS-232C interface I/O terminals (P1 through P4)
Control Panel
Press the [PROGRAM No.] key (Up or Down key) on the control panel.
The program numbers will be displayed in sequence by pressing the key repeatedly.
External I/O Terminals
External I/O terminals P1 through P4 on the rear panel of the controller can be used for program selection. Refer to Section 7 I/O Terminals on page 7-1 for external I/O termi­nals to be used and program numbers in detail.
RS-232C Interface
Commands sent by external devices through the RS-232C interface make it possible to select programs. Refer to PROGRAM No. on page 8-12 for the RS-232C interface in detail.
4-12
Page 63
Chapter 5
Function Settings
The LS-7600 Series is available to the measurement of a variety of items by making setting changes. These settings can be registered as programs that can be used whenever required. This section provides information on the overview of program settings and setting methods.
Flow of Program Settings ............................................5-2
Default Values and Possible Setting Ranges .............. 5-4
Copying Program Setting Details and Initialization ..... 5-6
Setting Screen............................................................. 5-7
Area Settings............................................................... 5-9
Calibration Settings ................................................... 5-22
Output Settings ......................................................... 5-29
Tolerance (LIMIT) Setting .......................................... 5-49
Option Settings.......................................................... 5-55
Group Processing ..................................................... 5-65
Statistical Processing ................................................5-66
Application Function .................................................. 5-68
5-1
Page 64
Chapter 5 Function Settings
t
M
T
t

Flow of Program Settings

5
Area Settings
See page 5-9
Measuring head Selection
Measuring head 1 (HEAD1) Measuring head 2 (HEAD2)
Area Selection
DIA T-EDGE B-EDGE SEG
Select the number of edges and mode (E/P)
Optional Settings
Focus monitor
Area check
Difference check
Function output
Edge detection threshold
Averaging number
Difference value
Area 1
Area 2
Calibration Settings
➾See page 5-22
Calibration Settings (Optional)
Calibration mode selection
Logic calibration Two-point calibration One-point calibration
Calibration Value Initialization
Environment Settings
RS-232C Interface
Program function
Head
Buzzer
5-2
See page 6-1
Language
Backlight
Key lock
Date and time
Page 65
Chapter 5 Function Settings
Output Settings
See page 5-29
Optional Settings
Area calculation
Averaging number
Offset (in combination with the automatic zero function)
Abnormal value elimination
Measuring Mode Settings
Tolerance Settings
Output Settings
Analog output settings
Scaling (Reference value/Threshold) Analog update mode Output waiting voltage
Comparator output settings
Strobe output OUT mode Off-delay time
External I/O
ZERO
TIMING
RESET
See page 5-49
5
Analog output
Comparator output
OUT 1
OUT 2
Function output
Option Settings
Tolerance mode
Reference value Threshold
Group discrimination
See page 5-55
Unit
mm μm inch mil
Minimum display unit
Statistical processing
I/O mode
SUB BCD
Application mode
BCD output
5-3
5-3
Page 66
Chapter 5 Function Settings

Default Values and Possible Setting Ranges

The following programs settings are made by default.
Area Settings (Page 5-9)
5
Item
Head number (HEAD)
Area
Area display (A-DISP) Focus monitor (FOCUS) Area check Difference check Function output setting (FUNCTION)
Edge detection threshold (LEVEL)
Averaging number (AVE) Difference value (DIFF)
Default
DIA
OFF
1000
FOCUS
Calibration Settings (Page 5-22)
Item
Calibration mode
T1-A/T1-B/T2-A/T2-B
Default
THEORY
Tolerance Settings (Page 5-49)
Item Default
Threshold
mode
Reference
value mode
HH output range (HH) HI output range (HI) +30.00000 LO output range (LO) 0.00000 LL output range (LL) -10.00000 HOLD-H OFF HOLD-L OFF STANDARD +10.00000 UPPER +1.00000 LOWER -1.00000 HOLD-H OFF HOLD-L OFF
+40.00000
Possible setting range
1, 2
1
Diameter (DIA) Top edge (T-EDGE) Bottom edge (B-EDGE) Segment (SEG) Specified SEG edge number: -127 to +127 SEG mode: Edge (E) or Pitch (P)
ON
ON, OFF 0 to 9
5
OFF (No area check), 2 to 127 1, 2, 4, 8, 16
1
1 to 9999 (in μm, mm, inch, or mil) Focus monitor, area check, or difference check 10 to 99%
50
Possible setting range
Logic, 2-point, or 1-point calibration
-99.99995 to +99.99995mm
Possible setting range
-99.99995 to +99.99995mm
OFF, -99.99995 to +99.99995mm
-99.99995 to +99.99995mm
OFF, -99.99995 to +99.99995mm
5-4
Page 67
Output Settings (Page 5-29)
Item Default
Area calculation (CALC)
Averaging number (AVE)
Measuring mode (MEA-MODE) Normal
Offset
Threshold
Reference value Analog output setting: Analog update mode (A-SET)
Analog output setting: Output waiting voltage (A-CONT)
Comparator output setting: Strobe output time (STOBE-T)
Comparator output setting: Output logic (OUT-SET)
Comparator output setting: Off-delay time (OFF-DLY)
Scaling at 10 V (A-OUT+10) Scaling at –10 V (A-OUT-10)
Scaling (A-SCALE)
OUT1: +1A1+0A2 OUT2: +0A1+1A2
+0, +1, +1/2, -1, -1/2
512
0.00000
+30.00000
0.00000
1000 μm/V
OFF OFF, ON
+10.5V
0.4ms
N.O.
OFF
1, 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024, 2048, 4096 Normal Peak hold (P-HOLD) Bottom hold (B-HOLD) Peak-to-peak hold (D-HOLD) Average hold (AV-HOLD) Automatic peak hold (A-P-HOLD) Automatic bottom hold (A-B-HOLD) Automatic peak-to-peak hold (A-D-HOLD) Sample hold 1 (S-HOLD1) Sample hold 2 (S-HOLD2) Self-timing hold (SELF-T) Self-timing period: 1 to 9999 ms (By default set to 100 ms)
-99.99995 to +99.99995mm
1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, 2000, 5000, 10000 μm/V
+10.5V, Hold, 0V
0.4, 0.8, 1.6, 3.3, 6.6, 13.3, 26.6ms
Normally open (N.O.) Normally closed (N.C.)
OFF, 0.4, 0.8, 1.6, 3.3, 6.6, 13.3, 26.6ms
Chapter 5 Function Settings
Possible setting range
5
Option Settings (Page 5-55)
Item Default
Tolerance mode (LIMITS) DEV Timing mode (T-MODE) ASYNC Unit (D-UNIT) mm
Minimum display unit (D-RES) 0.00005
I/O mode SUB Trend graph mode (GRAPH-T) Statistical processing mode (STATS) Group processing (GROUP) Application mode (APP-MODE)
OUT1 OUT2
OUT1 OUT2
1m
1m OFF OFF OFF OFF
Reference value (DEV) or threshold (TOL) Synchronous (SYNC) or asynchronous (ASYNC) mm(mm), μm(um), UNIT1(inch), UNIT2(mil) mm: 0.00001, 0.00002, 0.00005,
SUB mode (SUB), BCD mode (BCD) 1m, 10m, 30m, 1H, 2H, 6H, 12H, 24H
1m, 10m, 30m, 1H, 2H, 6H, 12H, 24H OFF, EXT, 2 to 9999999 OFF, EXT, 2 to 9999999 OFF, 2 to 255 OFF, DRILL, DUAL
Items in parentheses indicate what displayed on the LCD monitor.Displayed values are based on a minimum display unit of 0.00005 millimeters. Refer
to page 5-58 for other types of minimum display units.
Possible setting range
0.0001, 0.0002, 0.001,
0.01, 0.1
5-5
Page 68
5
Chapter 5 Function Settings

Copying Program Setting Details and Initialization

The LS-7600 Series can be initialized so that the details of programs will return to ones set by default. Furthermore, the LS-7600 Series has a function making it possible to copy the details of any program to different programs.
Copying Program Setting Details
The details of any program can be copied to different programs by using the P-COPY item as an environment setting. Refer to Copying Programs on page 6-5 for details.
Types of Initialization
The following three types of initialization are available.
• Initialization of a single program
• Initialization all programs
• Total initialization of all programs and environment settings
Initialization of Single Program
The details of a single program can be initialized by using the P-COPY item as an environment setting. Refer to Copying Programs on page 6-5 for details.
Initialization of All Programs
The details of all programs can be initialized by using the P-CLEAR item in the environ­ment setting menu. Refer to Resetting All programs on page 6-6.
Tips
If all programs are initialized and returned to ones set by default, the details of all the programs and environment settings will be initialized. Refer to Resetting to Initial Status on page 1-22 for details.
5-6
Page 69
Chapter 5 Function Settings

Setting Screen

The following section provides information on setting items.
Switching to Program Mode
1 Set the [Slide Switch] key to PROG mode. 2 By repeatedly pressing the [Up or Down] key, the cursor moves between LIMIT,
AREA, OUTPUT, CALIB, and OPTION.
3 Locate the cursor to the desired item and press the [ENT] key to enter the
mode. Tolerance Setting Mode
OUT1
12.34560
UPPER STANDARD LOWER
OUT2
23.45670
UPPER STANDARD LOWER
12.00000 —1.00000
23.00000 —1.00000
1.00000
1.00000
CAM1
01/02/18 17:47:30
PROG P — 00
LIMIT
AREA OUTPUT CALIB OPTION
A1 2 A2 2
5
Area Setting Mode
AREA1 HEAD 1
AREA DIA A-DISP ON FOCUS 5 A-TEST OFF
AREA2 HEAD 1
AREA DIA A-DISP ON FOCUS 5 A-TEST OFF
CAM1
PROG P – 00
LIMIT
AREA
OUTPUT CALIB OPTION
A1 2 A2 2
01/02/18 17:50:42
5-7
Page 70
Chapter 5 Function Settings
5
Output Setting Mode
Calibration Setting Mode
Option Setting Mode
OUT1
12.34560
CALC +1A1 +0A2 AVERAGE 512 MEA-MODE OFFSET 0.00000
OUT2
23.45670
CALC +0A1 +1A2 AVERAGE 512 MEA-MODE OFFSET 0.00000
AREA1
12.34560
T1-A T1-B T2-A T2-B
AREA2
23.45670
T1-A T1-B T2-A T2-B
RIMITS DEV
T-MODE ASYNC D-UNIT mm D-RES 0.00005 C-OUT SUB GRAPH-T OUT1 1m OUT2 1m STATS OUT1 1000 OUT2 1000 GROUP OFF APP-MODE
NORMAL
NORMAL
THEORY
5.003000
5.000000
10.006000
10.000000 SET
THEORY
5.003000
5.000000
10.006000
10.000000 SET
OFF
CAM1
P – 00
LIMIT AREA
OUTPUT
CALIB OPTION
A1 2 A2 2
01/02/18 17:50:42
CAM1
P – 00
LIMIT
AREA OUTPUT
CALIB
OPTION
A1 2 A2 2
01/02/18 17:50:42
CAM1
P – 00
LIMIT
AREA OUTPUT CALIB
OPTION
A1 2 A2 2
01/02/18 17:50:42
PROG
PROG
PROG
5-8
Page 71
Chapter 5 Function Settings

Area Settings

Area settings are used to determine the measurements taken from a variety of objects placed in the beam. The LS-7600 Series can use up to two measuring heads to set two measurements (AREA1 and AREA2) for simultaneous measurements.
Area Settings
Area settings include the following items.
• Area: Select the measuring head number together with the DIA, T-EDGE, B-EDGE, or SEG mode.
• Area display: Turn the cursor on the target viewer ON or OFF.
• Focus monitor: A function to check whether the target is within the measuring range. Set the best measuring range (i.e., the distance between the target and transmit­ter) with this function.
• Area check: Set the total number of edges to be measured. If the number of edges detected does not coincide with the set value, the measured value will be invalidated.
• Difference check: If there is a radical change between measured values, the change will be detected and output. Set the value for the change to be detected.
• Function output selection: Set the items to be output from the output terminal on the rear panel.
• Edge detection threshold change: Set the threshold to be detected when measuring highly transparent objects.
Area Setting Procedure
Make area settings while checking the display on the LCD monitor. The following procedure is for area settings through the remote control console.
5
1 Set the [Slide Switch] key to PROG. 2 Press the [Up or Down] key to move
the cursor to the AREA.
3 Press the [ENT] key.
The cursor frame will be displayed for AREA1.
AREA1 HEAD 1
AREA DIA A-DISP ON FOCUS 5 A-TEST OFF
AREA2 HEAD 1
AREA DIA A-DISP ON FOCUS 5 A-TEST OFF
AREA1 HEAD 1
AREA DIA A-DISP ON FOCUS 5 A-TEST OFF
AREA2 HEAD 1
AREA DIA A-DISP ON FOCUS 5 A-TEST OFF
CAM1
PROG
P – 00
LIMIT
AREA
OUTPUT CALIB OPTION
A1 2 A2 2
01/02/18 17:50:42
CAM1
PROG
P – 00
LIMIT
AREA
OUTPUT CALIB OPTION
A1 2 A2 2
01/02/18 17:50:42
5-9
Page 72
5
Chapter 5 Function Settings
4 Press the [Up or Down] key and move the cursor frame to the area to be set.
To set 2, press the [Up or Down] key and move the cursor frame to AREA2.
5 Press the [ENT] key.

Area Selection

Area
When a target is located in the measuring area as shown below, the light into the receiver is partly obstructed, thus creating a shadow. The contour of the shadow is called “edge.” The area is surrounded by the edge line. The LS-7600 Series can measure objects with area designation or edge designation. Area designation is possible in DIA, T-EDGE, or B-EDGE mode. Edge designation is possible in SEG mode.
EDGE
AREA
Target
Transmitter Receiver
DIA (Diameter)
When the LS-7600 Series is in DIA measurement operation, it measures the outer diameters of round bars or transparent
objects. The area between the top of the first shadow and the bottom of the last shadow is measured as shown in the illustrations.
AREA
AREA
EDGE
EDGE EDGE
5-10
Target
Transmitter
Target
Transmitter Receiver
Receiver
This space is measured.
This space is measured.
Page 73
Chapter 5 Function Settings
T-EDGE (Top Edge)/B-EDGE (Bottom Edge)
Select either of these settings for gap measurement or positioning of rollers.
When the LS-7600 Series is in T-EDGE measurement operation, it measures the width of the bright portion between the top of the measuring head and the first shadow in the measuring area.
When the LS-7600 Series is in B-EDGE measurement operation, it measures the width of the bright portion between the bottom of the measuring area and the bottom of the last object in the beam.
5
Target
Transmitter Receiver
Target
Transmitter
Tips
Setting the reference edge within the measuring area enables measurements
Receiver
This space is measured in T-EDGE mode or B-EDGE mode.
This space is measured in T-EDGE mode.
This space is measured in B-EDGE mode.
in higher repeat accuracy.
SEG (Segment)
Select this setting for the measurement of the inner diameters of discs or the measure­ment of pitches. The edge where the measurement of the object in the measuring area starts is called “start edge” and the edge where the measurement terminates is called a “end edge.” The space between these edges specified is measured. Segment measure­ment is performed in E mode or P mode. If objects are measured with the virtual center of the light axis set, use edge 0.
Setting Range
• Edge or pitch numbers can be designated between –127 and 127.
• The duplicated designation of an edge or pitch number will set the measured value to
0.
• If edges or pitches are designated in reverse order, negative measured values will be
obtained.
5-11
Page 74
Chapter 5 Function Settings
E (Edge) Mode
The LS-7600 Series measures objects based on the edges in this mode.
Set the edge numbers to positive values beginning with the top of the measuring head. The following example shows positive edge numbers and measuring portions.
5
Outer diameter 1 Clearance Outer diameter 2
Top
1E 2E
3E 4E
5E 6E
Bottom
Outer diameter 1: Space between 2E and 3E
(+2E to +3E)
Outer diameter 2: Space between 4E and 5E
(+4E to +5E)
Clearance: Space between 3E and 4E
(+3E to +4E)
Set the edge numbers to negative values beginning with the bottom of the measuring head. The following example shows negative edge numbers and measuring portions.
Outer diameter 1 Clearance Outer diameter 2
Top
-6E
-5E
-4E
-3E
-2E
-1E
Bottom
Outer diameter 1: Space between –5E and –4E
(–5E to –4E)
Outer diameter 2: Space between –3E and –2E
(–3E to –2E)
Clearance: Space between –4E and –3E
(–4E to –3E)
5-12
Page 75
Chapter 5 Function Settings
P (Pitch) Mode
The LS-7600 Series measures objects based on the middle point between the edges of the objects in this mode.
Top
1E
1P
Target center pitch
2E 3E 4E 5E 6E
Bottom
2P 3P 4P 5P
In this example as shown, five P-mode edges exist, based on the light and dark regions in the measuring area. Target center pitch: The space between 2P and 4P
5
Edge 0
Edge 0 refers to a method of measurement, in which the center of the optical axis is used as an origin. In this case, set the edge number to 0. Use this method to position targets or measure the movement or deviation of targets.
Top
Edge position of workpiece : Space between+1E and 0 (+1E to 0E)
In the edge 0 method, the E and P modes will not make any difference.
Tips
Center of optical axis
35.5 ± 0.2
1E
Edge 0
2E
Bottom
• Use edge 0 method if higher repetitive accuracy is required from the meas­urement of objects.
• If DUAL is selected in the application function (page 5-68), when no meas­uring target is in the beam axis, the number of edges becomes “0”.
5-13
Page 76
Chapter 5 Function Settings
Setting Procedure
5
1 Select HEAD and press the [ENT] key.
The cursor moves to the setting item.
2 Press the [Up or Down] key and select
the measuring head number.
3 Press the [ENT] key.
The cursor moves to HEAD.
4 Select AREA and press the [ENT] key.
The cursor moves to the setting item.
5 Press the [Up or Down] key and select
the measuring method.
The following items are displayed for setting.
AREA1
HEAD
AREA DIA A-DISP ON FOCUS 5 A-TEST OFF
AREA1
HEAD
AREA DIA A-DISP ON FOCUS 5 A-TEST OFF
AREA1
HEAD
AREA DIA
A-DISP ON FOCUS 5 A-TEST OFF
AREA1
HEAD
AREA DIA A-DISP ON FOCUS 5 A-TEST OFF
1
1
1
1
5-14
Tips
LCD monitor Setting item
DIA T-EDGE B-EDGE B-EDGE (Bottom edge) +2E~+3E SEG (Segment)
The area cursor on the target viewer moves according to the setting item.
DIA (Diameter) T-EDGE (Top edge)
Page 77
SEG (Segment) Setting (1) Press the [Up or Down] key and input
the start edge (or start pitch) number.
Chapter 5 Function Settings
AREA1
HEAD
AREA +2E +3E A-DISP ON FOCUS 5 A-TEST OFF
1
(2) Press the [Left or Right] key and
move the cursor to the end edge (or end pitch) number.
(3) Press the [Up or Down] key and input
the edge pitch (or end pitch) number.
(4) Press the [ENT] key.
The input values are entered and the cursor moves to AREA.
Note: • The duplicated designation of an edge or pitch number will set the meas-
ured value to 0.
• If edges or pitches are designated in reverse order, negative measured values will be obtained.
AREA1
HEAD
AREA +2E +3E A-DISP ON FOCUS 5 A-TEST OFF
AREA1
HEAD
AREA +2E +3E A-DISP ON FOCUS 5 A-TEST OFF
1
1

Area Display

The area cursor on the target viewer can be turned ON or OFF. If the cursors for both AREA1 and AREA2 are turned OFF, no scanning line will be displayed. When both CAMERA 1 and CAMERA 2 are selected in Switching Cameras (page 4-8), the area cursors cannot be hidden.
Setting Method
1 Select A-DISP and press the [ENT]
key.
The cursor moves to the setting item.
AREA1
HEAD
AREA DIA
A-DISP ON
FOCUS 5 A-TEST OFF
1
5
5-15
Page 78
Chapter 5 Function Settings
5
2 Press the [Up or Down] key and select
ON or OFF.
3 Press the [ENT] key.
The input item is entered and the cursor moves to A-DISP.
AREA1
HEAD
AREA DIA A-DISP ON FOCUS 5 A-TEST OFF
1

Focus Monitor

The focus monitor is a function that makes it possible to check whether the target is lo­cated within the range where the highest precision measurement of the target is ensured. The focus monitor range is the distance between the receiver and target in the measuring area. When the target is within the range specified, the area setting on the LCD monitor will display “F.”
FOCUS display is lit
The center of the measuring area
Setting Range
The value is set at 10 levels between 0 and 9. Set the value to 0 if the focus monitor function is not used. Refer to Focus Characteristics on page 9-6 for the set values and focus range in detail.
Focus display is OFF
Tips
Setting Method
1 Select FOCUS and press the [ENT]
key.
The cursor moves to the setting item.
2 Press the [Up or Down] key and input
the set value.
3 Press the [ENT] key.
The input item is entered and the cursor moves to FOCUS.
5-16
The result of focus monitoring can be externally output as function output. Refer to Function Output Selection on page 5-19 for details. The focus monitor displays and outputs the average value for 256 measurements.
AREA1
HEAD
AREA DIA A-DISP ON FOCUS 5 A-TEST OFF
1
Page 79
Chapter 5 Function Settings

Area Check

Set the total number of edges to be measured. If the number of edges detected does not coincide with the set value, the measured value will be invalidated. This function makes it possible to prevent the measurement of abnormal values that may result if dust or oil is stuck on the measuring head. The LCD monitor will display the following data if the meas­ured value is invalidated.
OUT1
UPPER 1.00000 STANDARD 12.00000 LOWER –1.00000
01/02/18 17:50:42
Setting Range
The setting range is between 2 and 127. Set the value to OFF if no area check is made.
Tips
The result of area checking can be external output as function output. Refer to Function Output Selection on page 5-19 for details. If the area checking
function is set, the number of edges displayed on the LCD monitor will be green for OK and red for NG (no good). If the function is not set, the number of edges displayed will be white (in simultaneous display mode only). Edge 0 is not included in the total number of edges.
Setting Method
1 Select A-TEST and press the [ENT]
key.
The cursor moves to the setting item.
2 Press the [Up or Down] key and select
the set value.
If the [ENT] key is pressed continuously, the set value will automatically increase or decrease.
AREA1
HEAD
AREA DIA A-DISP ON FOCUS 5 A-TEST 4
1
3 Press the [ENT] key.
The input item is entered and the cursor moves to A-TEST.
5
5-17
Page 80
Chapter 5 Function Settings

Difference Check

If there is a radical change between measured values, the change will be detected and output..
Setting Range
• Difference average: 1 to 16 times
• Difference value: 1 to 9999 μm
5
Note: The results of difference checks will be output externally, provided that DIFF
function output settings are made properly in advance. Refer to Function Output Selection on page 5-19 for details. The result output of a difference check is a 26.6-ms-long single output signal, which is not displayed on the LCD monitor.
Setting Method
Difference Average (DIFF AVE) Settings
1 Select DIFF AVE and press the [ENT]
key.
The cursor moves to the setting item.
2 Press the [Up or Down] key and select
the set value.
3 Press the [ENT] key.
The input item is entered and the cursor moves to DIFF AVE.
Difference Value (LIMIT) Settings
AREA1 DIFF
AVE 1 LIMIT 1000um FUNCTION LEVEL 50%
A-TEST
1 Select LIMIT and press the [ENT]
key.
The cursor moves to the setting item.
2 Press the [Left or Right] key to move
between digits. Press the [Up or Down] key to input the set value.
3 Press the [ENT] key.
The input value is entered and the cursor moves to LIMIT.
AREA1 DIFF
AVE 1 LIMIT 1000um FUNCTION LEVEL 50%
A-TEST
5-18
Page 81
Chapter 5 Function Settings

Function Output Selection

Select either Focus, Area, or Difference to be output from the output terminal block on the rear panel.
Setting Range
Any one of the following items is selectable for output. Refer to Section 7 I/O Terminals on page 7-1 for the FUNCTION output terminals in details.
• Focus monitor
• Area check
• Difference check
Setting Method
1 Select FUNCTION and press the [ENT]
key.
The cursor moves to the setting item.
2 Press the [Up or Down] key and select
the setting item.
The following items are displayed for setting.
LCD monitor Setting item
FOCUS A-TEST DIFF Difference check
Focus monitor Area check
3 Press the [ENT] key.
The input item is entered and the cursor moves to FUNCTION.
AREA1 DIFF
AVE 1 LIMIT 1000um
FUNCTION
LEVEL 50%
AREA1 DIFF
AVE 1 LIMIT 1000um FUNCTION LEVEL 50%
A-TEST
FOCUS
5
5-19
Page 82
Chapter 5 Function Settings

Change of Edge Detection Threshold

The following section provides information on how to make threshold changes for the measurement of highly transparent objects.
Transparent tape
Optical
reception
Brighter
Measuring range
Light blocked by tape
Optical reception
Brighter
Threshold (50%)
Threshold (80%)
5
Darker
Detection impossible
50% threshold
When measuring a highly transparent target, such as a plastic tape, the amount of light blocked by the tape is minimal. Therefore, the tape may not be detected if the threshold is low.
Darker
Position Position
In that case, increase the threshold level so that the tape will be detected
Detection
possible
80% threshold
Setting Range
The threshold can be set between 10% and 99% in 1% increments.
Note: • Only a single threshold can be set for each measuring head.
• In the measurement using the window edge such as T-EDGE or B-EDGE, the measurement may become unstable with a higher threshold. In such a case, use the reference edge or edge 0.
Optical Reception Waveform Output
While making a threshold setting for the target in the measuring area, an oscilloscope can be used to output the optical reception waveform of the target so that the threshold can be adjusted while monitoring the waveform.
10V
Approx. 6.5 V
Approx.
10 ms
0V
Trigger pulse
Optical reception waveform (OUT1 analog output on terminal 22)
Threshold level (OUT2 analog output on terminal 23)
5-20
Approx. 450 ms
Page 83
Setting method
Chapter 5 Function Settings
1 Select LEVEL and press the [ENT]
key.
The cursor moves to the setting item.
2 Press the [Up or Down] key and input
the set value.
If the [Up or Down] key is pressed continuously, the set value will automati­cally increase or decrease.
3 Press the [ENT] key.
The input value is entered and the cursor moves to LEVEL.
AREA1 DIFF
AVE 1 LIMIT 1000um FUNCTION
LEVEL 50%
AREA1 DIFF
AVE 1 LIMIT 1000um FUNCTION
LEVEL 50%
FOCUS
FOCUS
5
5-21
Page 84
5
Chapter 5 Function Settings

Calibration Settings

The measured value of an target may have a delicate error due to the surface condition and angle of the target. Calibration settings correct for this error. It is possible to make calibration settings for each area independently.
Types of Calibration Settings
Select the type of calibration from the following ones in advance. Refer to Calibration Type Selection on page 5-24. Logic calibration: Logic calibration requires two types of master workpieces that are
Two-point calibration:Two-point calibration requires two types of master workpieces
One-point calibration: One-point calibration requires a single master workpiece. This
Logic Calibration/Two-point Calibration
Prepare target 1 (T1) and target 2 (T2) as two measuring points. The LS-7600 Series performs calibration on these points based on the T1-A and T2-A display values, which are obtained before calibration, and the T2­B and T2-B display values, which are obtained after calibration. In this case, the non-calibrated display values are converted into the calibrated display values with the span value (inclination) and shift value of the workpiece determined by the LS-7600 Series internally.
different to each other in size. Calibration is performed according to pre-calibration and post-calibration values that are input by the user. This type of calibration is used for the measurement of the distance between two workpieces, for example.
that are different to each other in size.
calibration is used for the size measurement of a single type of workpiece.
Display value
T2-B
(T2 after calibration)
(T2 before calibration)
(T1 after calibration)
(T1 before calibration)
T2-A
T1-B
T1-A
T1
(1st point)
Display value after calibration
Display value before calibration
T2
(2nd point)
Ideal straight line
Workpiece size
5-22
One-point Calibration
Prepare target 1 (T1) as a single measuring point. The LS-7600 Series performs calibra­tion on this point based on the display value before calibration and that after calibration. In this case, the non-calibrated display value is converted into the calibrated display value with the span (inclination) and shift value of the workpiece determined by the LS-7600 Series internally. The span value is always set to 1.
Display value
T1-B
(T1 after calibration)
(T1 before calibration)
T1-A
T1
(1st point)
Ideal straight line
Display value after calibration
Display value before calibration
Workpiece size
Page 85
Chapter 5 Function Settings
Procedure for Calibration Settings
Take the following procedure if the remote control console is used.
1 Set the [Slide Switch] key to PROG.
The LCD monitor changes to the PROGRAM SETUP screen from the RUN screen.
2 Press the [Down] key and move the
cursor to CALIB.
3 Press the [ENT] key.
A frame is displayed on the AREA1 side.
Calibration Settings for AREA1
(1) Press the [ENT] key.
Calibration Settings for AREA2
(1) Press the [Down] key and move
the frame to AREA2.
(2) Press the [ENT] key.
AREA1 THEORY
T1-A ✽✽✽✽✽✽✽✽✽ T1-B ✽✽✽✽✽✽✽✽✽ T2-A ✽✽✽✽✽✽✽✽✽ T2-B ✽✽✽✽✽✽✽✽✽
AREA2 THEORY
T1-A ✽✽✽✽✽✽✽✽✽ T1-B ✽✽✽✽✽✽✽✽✽ T2-A ✽✽✽✽✽✽✽✽✽ T2-B ✽✽✽✽✽✽✽✽✽
AREA1
12.34560
T1-A T1-B T2-A
10.006000
T2-B
10.000000 SET
AREA2
23.45670
T1-A T1-B T2-A
10.006000
T2-B
10.000000 SET
AREA1 THEORY
T1-A ✽✽✽✽✽✽✽✽✽ T1-B ✽✽✽✽✽✽✽✽✽ T2-A ✽✽✽✽✽✽✽✽✽ T2-B ✽✽✽✽✽✽✽✽✽
AREA2 THEORY
T1-A ✽✽✽✽✽✽✽✽✽ T1-B ✽✽✽✽✽✽✽✽✽ T2-A ✽✽✽✽✽✽✽✽✽ T2-B ✽✽✽✽✽✽✽✽✽
THEORY
5.003000
5.000000
THEORY
5.003000
5.000000
SET
SET
SET
SET
CAM1
P — 00
LIMIT AREA
OUTPUT
CALIB
OPTION
A1 2 A2 2
01/02/18 17:50:42
CAM1
P – 00
LIMIT
AREA OUTPUT
CALIB
OPTION
A1 2 A2 2
01/02/18 17:50:42
CAM1
P – 00
LIMIT AREA
OUTPUT
CALIB
OPTION
A1 2 A2 2
01/02/18 17:50:42
PROG
PROG
PROG
5
5-23
Page 86
Chapter 5 Function Settings

Calibration Type Selection

The following section provides information on how to set the calibration type for each area. The procedure below is for AREA1 for a typical example.
1 With the AREA 1 box selected, press ENT. T1-A will be lit. Press the [Up] key
and move the cursor to underline the type of calibration. “THEORY”
The item is underlined and displayed.
2 Press the [ENT] key.
The cursor appears on “THEORY”.
5
3 Press the [Up or Down] key and
select the type of calibration for AREA1.
Any one of the following items displayed can be selected.
LCD monitor
Meaning
Logic calibrationTHEORY Two-point calibration2-TARGET One-point calibration1-TARGET Initialization of calibrationRESET
AREA1
T1-A ✽✽✽✽✽✽✽✽✽ T1-B ✽✽✽✽✽✽✽✽✽ T2-A ✽✽✽✽✽✽✽✽✽ T2-B ✽✽✽✽✽✽✽✽✽
4 Press the [ENT] key. 5 To enable the type of calibration that is selected, calibrate the LS-7600 Series
with the right method. For details, refer to the settings for each type of calibra­tion.
Tips

Logic Calibration Settings

Make logic calibration settings as explained below.
Setting Range
Logic calibration does not require master workpieces. Input the non-calibrated values and calibrated values of two measuring points (target 1 and target 2). Logic calibration uses the following values.
Take steps 1 through 4 to select the type of calibration for AREA2 as well. Refer to page 5-28 for the initialization of calibration.
T1-A: Target 1 display value before calibration T1-B: Target 1 display value after calibration T2-A: Target 2 display value before calibration T2-B: Target 2 display value after calibration
THEORY
SET
5-24
Make settings so that the following conditions will be satisfied. T1-A < T2-A and T1-B < T2-B Furthermore, set the calibration span between 0.5 and 2.0.
Page 87
Chapter 5 Function Settings
1 While the cursor is at T1-A, press the [ENT] key and input the T1-A value.
The display “*********” changes to “+00.00000.” The cursor moves to the “+” position and the value can be input.
(1) Press the [Up or Down] key and
select + or – as desired.
(2) Press the [Left or Right] key and
move the cursor between digits to input the value.
(3) Press the [Up or Down] key to
input the value.
(4) Repeat steps (2) and (3) until all
the digits are input.
(5) Press the [ENT] key.
All the digits are entered and the cursor moves to T1-A.
AREA1
5 00300
T1-A + T1-B ✽✽✽✽✽✽✽✽✽ T2-A ✽✽✽✽✽✽✽✽✽ T2-B ✽✽✽✽✽✽✽✽✽
AREA1
5 00300
T1-A + T1-B ✽✽✽✽✽✽✽✽✽ T2-A ✽✽✽✽✽✽✽✽✽ T2-B ✽✽✽✽✽✽✽✽✽
THEORY
00 .00000
SET
THEORY
00 .00000
SET
2 Press the [Up or Down] key and select T1-B, T2-A and T2-B. Take step 2 and
input the value in the same manner.
3 Press the [Down] key and move the cursor to SET. 4 Press the [ENT] key.
All the input values are entered and the AREA selection screen for calibration settings is displayed.
Note: If the calibration conditions are not satisfied, ERR1 will be displayed. Press
Tips
the [ENT] key or [ESCAPE] key to reset the error status. Then repeat the calibration settings again.
When repeating the logic calibration, the previous settings will appear on the screen, which will be different from those in the above setting example.
5
5-25
Page 88
Chapter 5 Function Settings

Two-point Calibration Settings

Make two-point calibration settings as explained below.
Setting Range
Two-point calibration requires two types of master workpieces that are different to each other in size. If the master workpieces are target 1 (T1) and target 2 (T2) respectively, make settings so that the following conditions will be satisfied. T1-A < T2-A Furthermore, set the calibration span between 0.5 and 2.0.
5
Note: In the case of two-point calibration, the number of averaging measurements is
Press the [ENT] key when the cursor is at 2-TARGET.
fixed at 4,096 times.
1 Place the small master workpiece (T1-A) in the beam. 2 While the cursor is at T1-B, press the [ENT] key. Monitor the measured value
and input the T1-B value.
The display “*********” changes to “+00.00000.” The cursor moves to the “+” position and to the value can be input.
(1) Press the [Up or Down] key and
select + or – as desired.
(2) Press the [Left or Right] key and
move the cursor between digits to input the value.
(3) Press the [Up or Down] key to
input the value.
(4) Repeat steps (2) and (3) until all
the digits are input.
(5) Press the [ENT] key.
When the input value is entered, the measured value of the master workpiece is set for the T1-A. Then the cursor moves to T1-B.
AREA1
T1-A T1-B + T2-A ✽✽✽✽✽✽✽✽✽ T2-B ✽✽✽✽✽✽✽✽✽
AREA1
T1-A T1-B + T2-A ✽✽✽✽✽✽✽✽✽ T2-B ✽✽✽✽✽✽✽✽✽
2-TRAGET
5 00300
0 .00000
00 .00000
SET
2-TRAGET
5 00300
0 .00000
05 .00000
SET
5-26
3 Place the large master workpiece (T2-A) in the beam. 4 Press the [Down] key and select T2-B. Then input the value in the same man-
ner in step 2.
5 Press the [Down] key and move the cursor to SET.
Page 89
Chapter 5 Function Settings
6 Press the [ENT] key.
The input value is entered and the screen returns to the AREA selection status of calibration settings.
Note: If the calibration conditions are not satisfied, ERR1 will be displayed. Press
Tips
the [ENT] key or [ESCAPE] key to reset the error status. Then repeat the calibration settings again.
When repeating the two-point calibration, the previous settings will appear on the screen, which will be different from those in the above setting example.

One-point Calibration Settings

Make one-point calibration settings as explained below.
Note: In the case of one-point calibration, the number of averaging measurements is
Press the [ENT] key when the cursor is at 1-TARGET.
fixed at 4,096 times.
1 Place the master workpiece in the beam. 2 While the cursor is at T1-B, press the [ENT] key. Monitor the measured value
and input the T1-B value.
The display “*********” changes to “+00.00000.” The cursor moves to the “+” position and to the value can be input.
(1) Press the [Up or Down] key and
select + or – as desired.
(2) Press the [Left or Right] key and
move the cursor between digits to input the value.
AREA1
T1-A T1-B +
T2-B ✽✽✽✽✽✽✽✽✽
1-TARGET
5 00300
0 .00000
00 .00000
SET
5
(3) Press the [Up or Down] key to
input the value.
(4) Repeat steps (2) and (3) until all
the digits are input.
(5) Press the [ENT] key.
When the input value is entered, the measured value of the master workpiece is set for the T1-A. Then the cursor moves to T1-B.
4 Press the [Down] key and move teh cursor to SET.
AREA1
T1-A T1-B +
T2-B ✽✽✽✽✽✽✽✽✽
1-TARGET
5 00300
0 .00000
05 .00000
SET
5-27
Page 90
Chapter 5 Function Settings
5 Press the [ENT] key.
The input value is entered and the screen returns to the AREA selection status of calibration settings.
5
Note: If the calibration conditions are not satisfied, ERR1 will be displayed. Press
Tips
the [ENT] key or [ESCAPE] key to reset the error status. Then repeat the calibration settings again.
When repeating the one-point calibration, the previous settings will appear on the screen, which will be different from those in the above setting example.

Initialization of Calibration

Take the following procedure to initialize the calibration values you are setting or you have already set (i.e., to return the values to the ones set by default). In this case, the span value is by default set to 1 and the shift value is by default set to 0.
Note: • It is possible to initialize the calibration value of each area independently.
• All calibration values set by default will be logic calibration values.
1 Press the [ENT] key when the cursor is at RESET.
The cursor moves to CANCEL.
Initializing AREA1 Calibration Value (1) Press the [Down] key and move
the cursor to RESET.
(2) Press the [ENT] key.
The value is initialized and the screen returns to the AREA selection status of calibration settings.
AREA1
RESET
CANCEL
RESET
5-28
Canceling Initialization of AREA1 Calibration Value
(1) Press the [Up or Down] key and
move the cursor to CANCEL.
(2) Press the [ENT] key.
The initialization is canceled and the screen returns to the AREA selection status of calibration settings.
AREA1
RESET
CANCEL
RESET
Page 91
Chapter 5 Function Settings

Output Settings

Values measured in AREA1 and AREA2 can be processed in a variety of ways and output as OUT1 and OUT2 signals. This section provides information on the details and methods of output settings.
Possible Output Settings
The following output settings are possible.
• Area calculation: The measured values of two-area points speci­fied by area settings are used for calculation processing.
• Number of averaging measurements: Set the number of measuring times.
• Measuring mode: Measured values are output “as is” or proc­essed as maximum and minimum timing output values.
• Offset: An offset value is a desired value that can be added to or subtracted from measured values.
• Analog output: Analog voltages are output according to the measured values.
• Discrimination output: Set the details of output that turns ON accord­ing to the judgment results of measured values.
Output Settings
Take the following procedure if the remote control console is used.
1 Set the [Slide Switch] key to PROG.
The LCD monitor changes to the PROGRAM SETUP screen from the RUN screen.
5
2 Press the [Down] key and move the
cursor to OUTPUT.
3 Press the [ENT] key.
The cursor frame is displayed on the OUT1 side.
OUT1
CALC +1A1 +0A2 AVE 512
MEA-MODE
OFFSET 0.00000
OUT2
CALC +0A1 +1A2 AVE 512
MEA-MODE
OFFSET 0.00000
OUT1
CALC +1A1 +0A2 AVE 512
MEA-MODE
OFFSET 0.00000
OUT2
CALC +0A1 +1A2 AVE 512
MEA-MODE
OFFSET 0.00000
NORMAL
NORMAL
NORMAL
NORMAL
CAM1
PROG
P – 00
LIMIT AREA
OUTPUT
CALIB
OPTION
A1 2 A2 2
01/02/18 17:50:42
CAM1
PROG
P – 00
LIMIT AREA
OUTPUT
CALIB
OPTION
A1 2 A2 2
01/02/18 17:50:42
5-29
Page 92
5
Chapter 5 Function Settings
Making Output Settings for OUT1 (1) Press the [ENT] key.
Making Output Settings for OUT2 (1) Press the [Down] key to move the
cursor frame to OUT2.
(2) Press the [ENT] key.
OUT1
CALC +1A1 +0A2 AVE 512
MEA-MODE
OFFSET 0.00000
OUT2
CALC +0A1 +1A2 AVE 512
MEA-MODE
OFFSET 0.00000
NORMAL
NORMAL
CAM1
PROG
P – 00
LIMIT AREA
OUTPUT
CALIB
OPTION
A1 2 A2 2
01/02/18 17:50:42

Area Calculation

The measured values from the two memory areas are used for calculation processing.
Setting Range
Select the coefficient from the following. +1, +1/2, 0, –1/2, and –1
Examples of Area Calculation
By multiplying the measured values with the coefficient selected, automatic size calcu­lation will be obtained. Set the coefficient as explained below to obtain the desired output value.
No Calculation is Performed
5-30
A1 A2
+1
0
Obtaining the Outer Diameter from Two Measuring Heads
Set AREA1 for DIA in measuring head 1 and AREA2 for DIA in measuring head 2.
A1 A2
+1/2 +1/2
Page 93
Obtaining the object width from Two Measuring Heads
Set AREA1 for B-EDGE in measuring head 1 and AREA2 for B-EDGE in measuring head 2.
A1 A2
–1 –1
Setting Method
1 Press the [Up or Down] key and move the cursor to CALC. 2 Press the [ENT] key.
The cursor moves to the coefficient for AREA1 (A1).
3 Press the [Up or Down] key and set
the coefficient for AREA1 (A1).
4 Press the [Right] key and move the
cursor to the coefficient for AREA2 (A2).
Chapter 5 Function Settings
OUT1
CALC +1A1 +0A2 AVE 512 MEA-MODE NORMAL OFFSET 0.00000
5
5 Press the [Up or Down] key and set
the coefficient for AREA2 (A2).
6 Press the [ENT] key.
The input data is entered and the cursor returns to CALC.
OUT1
CALC +1A1 +0A2 AVE 512 MEA-MODE NORMAL OFFSET 0.00000

Number of Averaging Measurements

Take the following procedure to determine the number of times to average measured values.
Setting Range
Select the number of averaging measurements from the following. 1, 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024, 2048, and 4096 If you want to average the fluctuation of measured values and obtain stable results, select a comparatively large number. For high-speed measurement, the number se­lected should be comparatively small.
Note:
When the measurement mode is set to Average Hold, set the averaging number to 1.
5-31
Page 94
5
Chapter 5 Function Settings
Averaging Time and Refresh Cycle
The LS-7600 Series averages measured values in a moving averaging method, thus making it possible to output the measured values within a short time. The following table lists the averaging time, the number of moving averaging data items, and the refresh cycle of a target according to the number of averaging measurements.
Number of
averaging
measurements
1 0.42 1 0.42 2 0.83 2 0.83
4 1.67 2 0.83
8 3.33 2 0.83 16 6.67 2 0.83 32 13.33 2 0.83 64 26.67 2 0.83
128 53.33 2 0.83 256 106.67 4 1.67
512 213.33 8 3.33 1024 426.67 16 6.67 2048 853.33 32 13.33 4096 1706.67 64 26.67
Averaging time (ms)
Measurement averaging time does not indicate the measuring response time of the
LS-7600 Series. Refer to Measuring Response Time on page 9-5 for details.
Number of moving
data samples
Refresh cycle (ms)
Example: If the averaging number is eight, the LS-7600 Series will output the data as
Internal data
(2,400 samples/s)
Output data
Setting Method
1 Press the [Up or Down] key and move the cursor to AVE. 2 Press the [ENT] key.
The cursor moves to the setting item.
3 Press the [Up or Down] key to set the
number of averaging measurements.
4 Press the [ENT] key.
The input value is entered and the cursor returns to AVE.
5-32
shown below.
1
2 3 4 5 6 7 8 9
Measurement averaging time
Measurement averaging time
Measurement averaging time
Refresh
10 11 12 13 14
ABC
Refresh
cycle
cycle
Data samples 1 through 8 are averaged, and data A is output
Data samples 3 through 10 are averaged, and data B is output
Data samples 5 through 12 are averaged, and data C is output
OUT1
CALC +1A1 +0A2 AVE 512 MEA-MODE NORMAL OFFSET 0.00000
Page 95
Chapter 5 Function Settings

Measuring Mode

The Measuring mode allows the LS-7600 Series to capture and hold measured values.
Setting Range
The LS-7600 Series has 11 types of Measuring modes. Refer to the following informa­tion for the outline of each mode.
Normal
The measured value is displayed and output as it occurs.
Bottom Hold
The minimum value during the period specified is displayed and output.
Average Hold
The average value during the period specified is displayed and output.
Auto Bottom Hold
The minimum value measured in displayed and output as it occurs.
Sample Hold 1
The instantaneous value at the time specified is displayed and output.
Synchronous input (timing input)
Measured value
Sampling time
Measured value
Sampling time
Minimum value
Minimum value
Average value
Sample Hold 1 value
Peak Hold
The maximum value during the period specified is displayed and output.
Peak-to-peak Hold
The maximum-to­minimum value during the period specified is displayed and output.
Auto Peak Hold
The maximum value measured is displayed and output as it occurs.
Auto Peak-to-peak Hold
The maximum-to­minimum value measured is displayed and output as it occurs.
Sample Hold 2
The value measured during the period specified is displayed and output.
Synchronous input (timing input)
Max. value
Measured value
Sampling time
Measured value
Sampling time
Sampling time
Maximum-to­minimum value
Max. value
Maximum-to­minimum value
Sample hold 2 value
5
Self-timing Hold
The measured value is displayed and output with no external timing input (i.e., timing input). The self-timing period is adjustable between 1 and 9,999 ms.
ST time
Workpiece insertion
Self-timing value
Sampling time
5-33
Page 96
Chapter 5 Function Settings
Application Example of Measuring modes
Select the measurement mode according to the type of target and measuring condition. The following applications are typical examples of measurement in a variety of Measur­ing modes.
Application Applicable RUN mode
5
Continuous measurement of wire, tube, or fiber.
Measurement of maximum value within specified period.
Measurement of minimum value within specified period.
Measurement of maximum-to­minimum value within specified period.
Measurement of average values within specified period.
Measurement of maximum value per rotation.
Measurement of minimum value per rotation.
Measurement of maximum-to­minimum fluctuation or eccentricity per rotation.
Measurement of average value per rotation.
Single workpiece measurement.
Normal
Peak Hold Auto Peak Hold
Bottom Hold Auto Bottom Hold
Peak-to-peak Hold Auto Peak-to-peak Hold
Average Hold
Peak Hold Auto Peak Hold
Bottom Hold Auto Bottom Hold
Peak-to-peak Hold Auto Peak-to-peak Hold
Average Hold
Sample Hold 1 Sample Hold 2 Self-timing Hold
5-34
Page 97
Chapter 5 Function Settings
Timing Chart
The following table lists timing charts for respective measuring modes.
RUN mode Timing chart name Reference page Normal Peak Hold/Bottom Hold/
Peak-to-peak Hold/Average Hold
Auto Peak Hold/Auto Bottom Hold/ Auto Peak-to-peak Hold
Sample Hold 1 Sample Hold 2 Self-timing Hold
Normal
Peak Hold
Auto Peak Hold
Sample Hold 1 Sample Hold 2 Self-timing Hold
5-35
5-36
5-37
5-38 5-39 5-40
Normal
The value is continuously measured, displayed, and output in normal mode.
Measured value
Waiting for judgment
(---------- displayed)
HI set value
LO set value
HI output
GO output
LO output
Judgment awaiting output
(Comparator waiting output)
Timing input
RESET input
RS-232C output
• In addition to the above output signals, HH output and LL output are available though they are omitted from the above chart.
• In the above chart, the ON status of each output means that NPN open collector output is ON, provided that it is normally open type. If it is normally closed type, the signal will be inverted.
• The ON status of each input means that the input is short-circuited to the COM terminal.
• RS-232C output is ON in synchronization with timing input if the D-SEND item as an environment setting is set to S1 or S2 only. Refer to Data Transmission Mode on page 6-9 for environment settings in detail.
• For strobe output in detail, refer to Strobe Output Time on page 5-47.
• For BCD output in detail, refer to BCD Output on pages 7-6 and 7-9.
OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
ON
ON
ON
ON
ON
ON
Measured value in normal mode
Internal measured value
Measured value
RESET processing
5
t
Tips
• If RESET input is ON while timing input is ON, awaiting the result of meas­urement, the LCD monitor will display the result waiting status “---.-----“ until the timing input is turned OFF.
• Comparator waiting output can be OFF-delayed. Refer to OFF-delay Time on page 5-48 for the details of OFF-delays.
• Timing input and reset input can be sent by RS-232C control. Refer to page 8-11 for details.
5-35
Page 98
Chapter 5 Function Settings
Peak Hold
• Peak hold: The maximum value during the sampling period specified by
• Bottom hold: The minimum value during the sampling period specified is
• Peak-to-peak hold: The maximum-to-minimum value during the period specified is
• Average hold: The average value during the sampling period specified is dis-
external timing input is displayed and output as shown below.
displayed and output.
displayed and output.
played and output.
5
Measured value
Waiting for judgment
(---------- displayed)
Internal measured value
Sampling
period
Timing input
RESET input
Comparator output
Judgment awaiting output
(Comparator waiting output)
Strobe output
BCD output
RS-232C output
• Timing input is OFF during sampling periods. The maximum value during the sampling period immediately before timing input will be output when the timing input turns ON.
• In the following cases, the LCD monitor will display the result waiting status “---.-----“ until the first measured value is determined.
The LS-7600 Series is turned on, setting changes are made, program number changes are made, RESET input is ON, or auto zero is input.
• Comparator output turns ON according to the measured value.
• In the above chart, the ON status of each output means that NPN open collector output is ON, provided that it is normally open type. If it is normally closed type, the signal will be inverted.
• The ON status of each input means that the input is short-circuited to the COM terminal.
• RS-232C output turns ON in synchronization with timing input if the D-SEND item as an environment setting is set to S1 or S2 only. Refer to Data Transmission Mode on page 6-9 for environment settings in detail.
• If timing input is ON during reset processing, the LCD monitor will display the result waiting status “---.-----.”
OFF
OFF
OFF
OFF
OFF
ON
ON
ON
ON
ON
Sampling period
Judgment
Measured value Measured value Measured value
Peak hold measured value
Judgment
Measured valueMeasured value
Sampling period
RESET
processing
t
Judgment
Measured value
5-36
Tips
Note:
• If a workpiece is eliminated during a sampling period and it becomes a period without the internal measured value obtained, the maximum value of the rest of the normal measured values will be output.
• Timing input and RESET input can be under RS-232C control. Refer to page 8-11 for details.
• Strobe output time can be changed. Refer to Strobe Output Time on page 5­47 for details.
When the measurement mode is set to Average Hold, set the averaging number to 1.
Page 99
Chapter 5 Function Settings
Auto Peak Hold
• Auto peak hold: The maximum value measured by the LS-7600 Series in
• Auto bottom hold: The minimum value measured by the LS-7600 Series in
• Auto peak-to-peak hold: The maximum-to-minimum value measured by the LS-7600
Waiting for judgment
(---------- displayed)
Timing input
RESET input
Comparator output
Judgment awaiting output
(Comparator waiting output)
• Comparator output turns ON according to the measured value.
• In the above chart, the ON status of each output means that NPN open collector output is ON, provided
• The ON status of each input means that the input is short-circuited to the COM terminal.
• RS-232C output is ON in synchronization with timing input if the D-SEND item as an environment setting is
• For strobe output in detail, refer to Strobe Output Time on page 5-47.
• For BCD output in detail, refer to BCD Output on pages 7-6 and 7-9.
• If timing input is ON during reset processing, the LCD monitor will display the result waiting status “---.-----.”
RS-232C output
that it is normally open type. If it is normally closed type, the signal will be inverted.
set to S1 or S2 only. Refer to Data Transmission Mode on page 6-9 for environment settings in detail.
continuous measurement operation is displayed and output.
continuous measurement operation is displayed and output.
Series in continuous measurement operation is displayed and output .
Measured value
ON
OFF
ON
OFF
ON
OFF
ON
OFF
Auto peak hold measured value
Internal measured value
RESET
processing
JudgmentJudgment
Measured value
t
5
Tips
• If RESET input is ON while timing input is ON, awaiting the result of meas­urement, the LCD monitor will display the result waiting status “---.-----“ until the timing input is turned OFF.
• Timing input and reset input can be under RS-232C control. Refer to page 8-11 for details.
• If a workpiece is missing and there is a period without the internal measured value obtained, the LCD monitor will display the result waiting status “---.----­.” The LS-7600 Series will start measurement with the previous auto peak value maintained when the internal measured value is obtained.
5-37
Page 100
5
Chapter 5 Function Settings
Sample Hold 1
The instantaneous value at the time specified by external timing input is displayed, and the value is output as shown below.
Measured value
Waiting for judgment
(---------- displayed)
Timing input
RESET input
Comparator output
Judgment awaiting output
(Comparator waiting output)
• When timing input turns ON, the internal measured value will be held. Then the value will be output.
• In the following cases, the LCD monitor will display the result waiting status “---.-----“ until the first measured value is determined.
• Comparator output turns ON according to the measured value.
• In the above chart, the ON status of each output means that NPN open collector output is ON, provided that it is normally open type. If it is normally closed type, the signal will be inverted.
• The ON status of each input means that the input is short-circuited to the COM terminal.
• RS-232C output is ON in synchronization with timing input if the D-SEND item as an environment setting is set to S1 or S2 only. Refer to Data Transmission Mode on page 6-9 for environment settings in detail.
• If timing input is ON during reset processing, the LCD monitor will display the result waiting status “---.-----.”
Strobe output
BCD output
RS-232C output
The LS-7600 Series is turned on, setting changes are made, program number changes are made, RESET input is ON, or auto zero is input.
OFF
OFF
OFF
OFF
OFF
ON
ON
ON
ON
ON
Measured value Measured value Measured value
Sample hold measured value
Internal measured value
Judgment
RESET
processing
Judgment Judgment
Measured valueMeasured valueMeasured value
t
5-38
Tips
• Timing input and RESET input can be under RS-232C control. Refer to page 8-11 for details.
• Strobe output time can be changed. Refer to Strobe Output Time on page 5­47 for details.
Loading...