KEYENCE LS-7000 User Manual

96M12219
User’s Manual
High-speed, High-precision Digital Micrometer
LS-7000 Series

Preface

This instruction manual provides necessary information on the operation and maintenance of the LS-7000 Series along with precautions. Please read the manual carefully and be sure you understand the information provided before attempting to install and operate the LS-7000 Series. Keep this manual handy for future reference. Please make sure that the end users are provided with this manual.
Symbols
The following symbols are used for the list of precautions to ensure safety and to prevent personal injury and/or property damage when using the LS-7000 Series. The following symbols alert you to important messages. Be sure to read these messages carefully.
DANGER
WARNING
CAUTION
Important:
Note:
Tips
Reference:
Failure to follow instructions may lead to death or serious injury.
Failure to follow instructions may lead to injury.
Failure to follow instructions may lead to product damage (product malfunctions, etc.).
Provides important precautions and restrictions on proper opera­tion.
Provides additional information on proper operation.
Provides useful information on proper operation.
Provides reference pages.
General precautions
1. No part of this manual may be reprinted or reproduced in any form or by any means without the prior written permission of KEYENCE CORPORATION.
2. The content of this manual is subject to change without notice.
3. KEYENCE has thoroughly checked and reviewed this manual. Please contact the sales office listed at the end of this manual if you have any questions or comments regarding this manual, or if you find an error.
4. KEYENCE assumes no liability for damages resulting from the use of the information in this manual, item 3 above notwithstanding.
5. KEYENCE will replace any incomplete or incorrectly collated manual.
6. All company names and product names in this manual are registered trademarks or trademarks of their respective owners.

Safety Precautions

General precautions
• At startup and during operation, be sure to monitor the functions and performance of the LS-7000 Series.
•Take sufficient safety measures to prevent damage to the human body and/or equip­ment that may be caused if this product should fail to operate properly.
• Do not modify the LS-7000 Series or use it in any way other than as described in the specifications. Its functions and performance are not guaranteed under said conditions.
• When the LS-7000 Series is used in combination with other instruments, its functions and performance may be degraded, depending on the operating conditions and surrounding environment.
• Do not use the LS-7000 Series for the purpose of protecting the human body.
• Do not expose the LS-7000 Series and peripheral devices to sudden temperature change, as this may cause condensation.
Precautions
WARNING
Operation
• Always use the LS-7000 Series at 24 VDC, otherwise a fire, electric shock or product failure may result.
• Prepare the applicable AC power supply cable if you use LS­S11.
• Secure the GND line of the power cable of the AC power supply stand to protective ground, otherwise an electric shock or product failure may result.
• Do not disassemble or modify the LS-7000 Series. This may cause fire or electric shock.
When abnormal conditions are encountered
If the following conditions are encountered, immediately turn off the power. Continuing to use the LS-7000 Series. under abnor­mal conditions may cause fire, electric shock or equipment failure. Contact your nearest KEYENCE sales office for repairs.
•When water or foreign matter enters the controller.
• When the LS-7000 Series. is dropped or the housing is damaged.
• When the controller produces smoke or an abnormal smell.
96M12219
i
CAUTION
Usage
• Be sure to turn off the power to the LS-7000 Series and any connected devices before connecting or disconnecting the cables. Otherwise, the camera and connected devices may be damaged.
• Do not turn off the power while setting a parameter. Otherwise, the settings may be partially or completely lost.
• Do not block the ventilation slots on the LS-7000 Series and peripheral devices. A rise in inner temperature may cause equipment failure.
Proper environment and conditions
To use the LS-7000 Series properly and safely, do not install the LS-7000 Series in locations with the following conditions. Use of this equipment in an improper environment may cause fire, electric shock, or equipment failure.
• Locations with high humidity, a large amount of dust, or poor ventilation
• Locations where the temperature rises excessively due to direct sunlight, etc.
• Locations where corrosive or flammable gas exists
• Locations where the LS-7000 Series is directly subjected to vibration or impact
• Locations where water, oil or chemicals may splash the LS­7000 Series
• Locations where static electricity is easily built up
Countermeasures against Noise
The LS-7000 Series may malfunction or fail to operate due to noise generated from power lines or high-tension lines. If the LS­7000 Series is operated close to such noise sources, use a noise filter, wire the cable of the LS-7000 Series in an independent conduit, or locate the LS-7000 Series in an area with proper insulation from noise.
ii
Note: Influence of Ambient Operating Temperature
Ensure that the ambient operating temperature is constant. A change in the ambient operating temperature may result in measurement errors. If the ambient operating temperature changes by 10°C, it will take approximately 60 minutes for the interior temperature distribution of the LS-7000 Series to be uniform.
Warming Up
Do not operate the LS-7000 Series for approximately 30 minutes after it is turned on. The internal circuitry of the LS-7000 Series is not stable immediately after it is turned on. The measured value drifts gradually as a result.
Influence of Dust and Dirt
A measurement error may result due to dirt, dust, water, or oil in any of the following cases.
• Cover glass: Blow off the dirt, dust, water or oil on the cover glass with clean air. If the cover glass is excessively dirty, wipe the glass with a soft cloth moistened with isopropyl alcohol.
• Target surface: Blow off the dirt, dust, water or oil on the sur­face with clean air or wipe the surface.
• Intrusion of dirt, dust, water or oil blown or sprayed into optical sensing area: Install a protection cover or perform air purging.
Influence of Vibration
The measured value of an object may fluctuate if the object is vibrating. If that happens, increase the averaging number. This will ensure highly accurate measurement.
Targets
The measured value of an object may have an error due to the shape and surface condition. Locate a reference object within the measuring area and use the calibration function of the LS-7000 Series to correct the error if that happens.
Influence of Air Flow
The measured value of an object may fluctuate due to a slow flow of air. In that case, the following countermeasures are effective.
• Protect the measuring head with a cover.
• Use a fan to mix the air vigorously between the transmitter and receiver.
Maintenance
Do not wipe the LS-7000 Series with a damp cloth or a cloth moistened with benzine or paint thinner. This may cause the discoloration or deformation of the LS-7000 Series. If the LS­7000 Series is excessively dirty, use a tightly squeezed cloth moistened with thin neutral detergent. Then wipe the LS-7000 Series with a soft dry cloth.
Precautions for LED
The LS-7010/LS-7030/LS-7070 Measuring Head uses an LED light source. Abide by the following precautions when using the unit.
CAUTION
• Do not look at the LED light source for a long time. The LED used as the light source is classified into class 1 according to IEC60825-1 standards. Basically the light beam of the LED is safe, but do not look at the LED for a long time.
• Do not disassemble the unit. The unit does not incorporate a mechanism to stop the emission of the LED when the unit is disassembled. Never attempt to disassemble the unit, otherwise you will be exposed to LED light.
iii

Precautions for CE Markings

Keyence confirms that the LS-7000 Series meets EC directive requirements. The LS-7000 Series bears CE markings. Keep the following conditions if the LS-7000 Series is used in European countries.
EMC Directives
Keep the following condition so that the LS-7000 Series will satisfy EN61326-1 requirements.
• The power cable and I/O cable connected to the controller are both less than 30m in length.
• Attach the OP-31973 Clamp-type Ferrite Core to the cable between the controller and measuring head. The attaching position is as close as possible to the cable end on the controller side, where the core must be snapped onto the cable to form a one-turn loop.
Ferrite core
OP-31973
Low-voltage Directives
When using the LS-S11 AC Power Supply Stand as a dedicated power supply to the LS­7000 Series, keep the following conditions so that the LS-S11 will satisfy EN61010-1 requirements. The LS-S11 satisfies EN60825-1 (LED Class 1) requirements with no conditions attached.
Environment Conditions
Installation category (Overvoltage category): II (see note 1) Pollution degree: 2 (see note 2)
Operation and Installation Conditions
• The LS-S11 AC Power Stand is a dedicated power supply for the LS-7000 Series. Never use the LS-S11 for other purposes.
• When using the LS-7000 Series in European countries, check that the AC power supply cable satisfies EC directive requirements certified by an authorized third party. Consult your Keyence representative for the AC power supply cable in detail.
• Before turning the LS-7000 Series on, connect the ground terminal of the AC power supply cable to the protective ground terminal of the power supply.
• Be sure to use a time lag fuse rated 250 V, 3.15 A for the replaceable fuse of the LS­S11.
Note 1: Installation categories (overvoltage categories) include impulse overvoltage regu-
Note 2: The pollution degree refers to the level of solid, liquid, or gas contamination that
iv
lations at transient overvoltage levels specified by EC directives. Installation category II indicates a level where power is provided from fixed installations, such as distribution panels.
deteriorates the dielectric strength or surface resistance of the LS-7000 Series. Pollution degree 2 indicates the normal indoor environment (free of non-con­ductive contamination).

About this Manual

The following section provides information on the configuration of each page of this manual along with symbols and terms used in this manual.

Page Configuration and Symbols

Chapter 3 Operation Control
Automatic Zero Function
The automatic zero function sets measuring values to zero (0.00000) instantly. This func­tion is convenient for zero-point calibration for a variety of target. Furthermore, master calibration is possible by using this function in combination with the offset function.
Tips: Describes information that will deepen the user’s comprehension of the manual along with other useful information.
Caution: Indicates information that, if not heeded, could result in relatively serious or minor injury, damage to the LS-7000 Series, or faulty operation.
Note: This function will not be available while the controller is awaiting the result of
measurement, during which the LCD monitor will display “– – –. – – – –.”)
The automatic zero value will not be lost after the LS-7500 Series is turned
Tips
off. The automatic zero value for each program will be stored.
Automatic zero settings are made through the following functional components.
• Control panel
• External terminal
3
• RS-232C interface
Control Panel
If a single measuring value is displayed, an automatic zero setting will be made for the corresponding OUT number only. If both measuring values are displayed, an automatic zero setting will be made for the selected OUT number on the screen.
1
Press the [ZERO] key once. Automatic zero will be set.
To cancel the setting, press the [ZERO] key again.
External I/O Terminal
Automatic zero will be set by short-circuiting the I/O, ZERO, and COM terminals on the rear panel of the controller together. Refer to Section 7 I/O Terminals on page 7-1 for details.
CAUTION
Tips
3-10
• Do not look at the LED light source for a long time. The LED used as the light source is classified into class 1 according to IEC60825-1 standards. Basically the light beam of the LED is safe, but do not look at the LED for a long time
Master Calibration with Offset Function (Automatic Offset Function)
Set the size of a master workpiece to an offset value and make an automatic zero setting while measuring the master workpiece. Then the size of the master workpiece will be displayed as an offset value. Refer to 3-3 Measure­ment of Outer Diameter and Width (with Two Measuring Heads) on page 3-4 for an application example. Refer to Offset on page 5-43 for the offset function in detail.
AREA DISP
HOLD
ZERO
Note: Describes items that need the user’s attention in order to prevent mistakes in the operation of the LS-7000 Series.
The above page sample is for your reference only. It does not coincide with any of
the actual pages of this manual.

Definition of Terms

This manual uses the following terms.
Term Description
LS-7000 Series Refers to the LS-7001 controller and the measuring head as a set. Controller Refers to the LS-7001 controller. Measuring head Refers to the LS-7010/LS-7030/LS-7070 measuring head.
v

Optimum Measurement Methods

Refer to the following methods, and select the optimum measurement method according to the type of target.
Measurement of Outer Diameter
of a Round Bar or Wire
Measurement of Average Outer
Diameter of Round Bar or Wire
Measurement of Sheet Width
D
Measurement of outer diameter and width (with single measuring head) (Page 2-2)
Measurement of Inner
Diameter of a Disc
Measurement of inner diameter and clearance (Page 2-5)
Measurement of Connector
Pins and IC Pins
Y
X
Measurement of outer diameter and width (with two measuring heads in X and Y directions) (Page 2-3)
Measurement of Outer Diameter
and Eccentricity of a Roller
Measurement of outer diameter and eccentricity (Page 2-6)
L
Measurement of outer diameter and width (with two measuring heads for larger objects) (Page 2-4)
Measuring Robot Movement and Positioning of LCD Plate
Measurement of movement and positioning (Page 2-7)
Measurement of pitch (Page 2-8)
vi

Configuration of Manual

Chapter 1 Getting Started
Provides information on precautions and necessary preparations required.
Chapter 2 Easy Setting Guide
Provides information on targets and settings for typical applications respectively.
Chapter 3 Operation Control
Provides information on setting control items on the control panel while the LS-7000 Series is in measurement operation.
Chapter 4 Function Settings
Provides information on the overview of program settings and setting methods.
Chapter 5 Environment Settings
Provides information on basic operation settings.
Chapter 6 I/O Terminals
Provides information on the specifications of the I/O terminals and timing charts.
1
2
3
4
5
6
Chapter 7 RS-232C
Provides information on how to connect external devices and the communications function to operate the LS-7000 Series through the
external devices.
Chapter 8 Specifications
Provides the specifications, characteristics, and external dimensions of the LS-7000 Series.
Chapter 9 Troubleshooting
Provides information on troubleshooting and error messages.
Appendix
Provides a list of options and the index of the manual.
7
8
9 Appendix
vii
Measured va between the r
setting. (See pa

LS-7001 Operation Map

Select the area (See page 3-4)
Press the Up or Down key to select the area.
AREA
The selected area appears on the AREA display.
POSITION monitor
The POSITION monitor is lit according to the position and size of the target.
FOCUS monitor
The FOCUS monitor is lit according to the distance between the transmitter and receiver. To make lit range changes, refer to page 4-14.
Program Settings
Area, calibration, output, or option setting:
Press the key and select the
item with key.
PROG
FUNC
AREA
1 2
FUNC
POSITION
FOCUS
ESCPROG
PRINT
ENT
No.
PROG No.
Environment setting:
Press the key while pressing
the key.
Set item selection:
Press the or key.
Set item cancellation:
Press the key.
Entering set item:
Press the key.
viii
FUNC
PROG
ESC
ENT
Program selection
Press the Up or Down key.
PROG No.
OUT 1
OUT 2
HI
GO
LO
TIM
HI
GO
LO
HOLD
ZERO
LS-7000
Tolerance setting (See page 4-50)
Reference value mode
Upper limit setting:
Reference value setting:
Lower limit setting:
Press the
Press the
Press the
HI
GO
LO
Threshold setting
HI value setting:
HH value setting:
LO value setting:
LL value setting:
Press the
Press the
Press the
Press the
HI
HI
LO
LO
Value change while setting
LO
Press the key below the value
HOLD
ZERO
flashing.
Hold function (See page 3-5)
key.
key.
key.
key.
key 2X.
key.
key 2X.
Measured value display selection
Press the Up or Down key.
DISP
Measured value display varies between the reference value mode and threshold mode for tolerance setting. (See page 3-2)
Press the key.
HOLD
Auto zero function (See page 3-6)
Press the key.
To cancel the function, press the
ZERO
key again.
ZERO
ix

Contents

Safety Precautions ..............................................................................................................i
Precautions for CE Markings ........................................................................................... iv
About this Manual .............................................................................................................. v
Page Configuration and Symbols .................................................................................... v
Definition of Terms .......................................................................................................... v
Optimum Measurement Methods .................................................................................... vi
Configuration of Manual .................................................................................................. vii
LS-7001 Operation Map .................................................................................................. viii
Contents.............................................................................................................................. x
Chapter 1 Getting Started
Outline and Features of LS-7000 Series ....................................................................... 1-2
System Applications ....................................................................................................... 1-3
Package Checks .............................................................................................................. 1-4
LS-7001 ....................................................................................................................... 1-4
LS-7010 ....................................................................................................................... 1-4
LS-7030 ....................................................................................................................... 1-4
LS-7070 ....................................................................................................................... 1-5
LS-C✱✱A ..................................................................................................................... 1-5
LS-S11 ......................................................................................................................... 1-6
Functions and Nomenclature ......................................................................................... 1-7
Controller ..................................................................................................................... 1-7
Measuring Head ........................................................................................................... 1-9
Mounting and Connection ............................................................................................ 1-11
Mounting the Measuring Head ................................................................................... 1-11
Mounting the Controller .............................................................................................. 1-13
Connection .................................................................................................................1-17
Turning AC Power Supply Stand On and Off .............................................................. 1-19
Measurement and Setting Overview ........................................................................... 1-20
Resetting to Initial Status .............................................................................................1-21
Chapter 2 Easy Setting Guide
Measurement of Outer Diameter and Width (with Single Measuring Head) .............. 2-2
Measurement of Outer Diameter and Width (with Two Measuring Heads)................. 2-3
Measurement of Outer Diameter and Width
(with Two Measuring Heads for Larger Objects) ..........................................................2-4
Measurement of Inner Diameter and Clearance ........................................................... 2-5
Measurement of Outer Diameter and Eccentricity ....................................................... 2-6
Measurement of Movement and Positioning ................................................................ 2-7
Measurement of Pitch .....................................................................................................2-8
Chapter 3 Operation Control
Selection of Measured Value Display ............................................................................ 3-2
Area Monitor ....................................................................................................................3-4
Hold Function ..................................................................................................................3-5
Automatic Zero Function................................................................................................ 3-6
Panel Lock Function ....................................................................................................... 3-7
Program Changes ........................................................................................................... 3-8
x
Chapter 4 Function Settings
Flow of Program Settings ............................................................................................... 4-2
Default Values and Possible Setting Ranges ............................................................... 4-4
Copying Program Setting Details and Initialization ..................................................... 4-6
Area Settings ...................................................................................................................4-7
Area Selection .............................................................................................................. 4-9
Focus Monitor ............................................................................................................ 4-14
Area Check ................................................................................................................ 4-15
Difference Check........................................................................................................ 4-16
Function Output Selection .......................................................................................... 4-17
Change of Edge Detection Threshold ........................................................................ 4-18
Calibration Settings ...................................................................................................... 4-20
Calibration Type Selection ......................................................................................... 4-23
Logic Calibration Settings .......................................................................................... 4-23
Two-point Calibration Settings ................................................................................... 4-25
One-point Calibration Settings ................................................................................... 4-26
Initialization of Calibration .......................................................................................... 4-26
Output Settings ............................................................................................................. 4-28
Area Calculation ......................................................................................................... 4-30
Number of Averaging Measurements......................................................................... 4-31
Measuring Mode ........................................................................................................ 4-33
Offset ......................................................................................................................... 4-43
Analog Output Settings .............................................................................................. 4-44
Comparator Output Setting ........................................................................................ 4-47
Tolerance (LIMIT) Settings ........................................................................................... 4-50
Limit Range ................................................................................................................ 4-50
Limit Settings ............................................................................................................. 4-51
Hold Value Elimination ............................................................................................... 4-52
Option Settings ............................................................................................................. 4-54
Tolerance Mode (LIMITS)........................................................................................... 4-56
Timing Mode .............................................................................................................. 4-56
Display Unit ................................................................................................................ 4-57
Minimum Display Unit ................................................................................................ 4-58
I/O Mode .................................................................................................................... 4-59
Units, Default Minimum Values, and Setting Ranges ................................................. 4-60
Chapter 5 Environment Settings
Possible Environment Settings ...................................................................................... 5-2
Default Settings ............................................................................................................ 5-2
Changing Environment Settings .................................................................................. 5-3
Flow of Environment Setting Display ........................................................................... 5-4
Details of Environment Settings .................................................................................... 5-5
Program ....................................................................................................................... 5-5
RS-232C ...................................................................................................................... 5-7
Head .......................................................................................................................... 5-10
Beep ........................................................................................................................... 5-12
Panel Lock ................................................................................................................. 5-13
xi
Chapter 6 I/O Terminals
Nomenclature and Functions of I/O Terminals ............................................................. 6-2
Te r minal Block .............................................................................................................. 6-2
Connector I/O ............................................................................................................... 6-4
Functions ..................................................................................................................... 6-7
Electrical Specifications ............................................................................................... 6-9
Timing Chart ..................................................................................................................6-11
Chapter 7 RS-232C
Specifications .................................................................................................................. 7-2
Communication using Serial Commands ..................................................................... 7-4
Connecting to a PC/PLC Link Unit ............................................................................... 7-4
Connecting to KV Series .............................................................................................. 7-5
Outline of Command Formats ...................................................................................... 7-6
Measurement Serial Commands ................................................................................. 7-7
Setup Serial Commands ............................................................................................ 7-12
Serial Commands to check the LS Settings ............................................................... 7-19
Batch Commands ...................................................................................................... 7-20
Timing Chart .............................................................................................................. 7-21
ASCII Table ................................................................................................................ 7-22
Communication using an External Synchronous Trigger ......................................... 7-23
Chapter 8 Specifications
Specifications .................................................................................................................. 8-2
Controller ..................................................................................................................... 8-2
Measuring Head ........................................................................................................... 8-3
Optional Accessories ................................................................................................... 8-3
Characteristics ................................................................................................................ 8-4
Measurement Range and Accuracy ............................................................................. 8-4
Temperature Characteristics ........................................................................................8-5
Response Delay Time .................................................................................................. 8-6
Focus Characteristics................................................................................................... 8-6
Mode Specifications ....................................................................................................... 8-7
External Dimensions....................................................................................................... 8-8
Chapter 9 Troubleshooting
Troubleshooting Guide ................................................................................................... 9-2
Error Messages ............................................................................................................... 9-6
Appendix
List of Optional Accessories........................................................................... Appendix-2
Setting Record Sheet ....................................................................................... Appendix-3
Environment Setting Record Sheet ................................................................ Appendix-5
Index ................................................................................................................. Appendix-6
xii
Chapter 1
Getting Started
This section provides information on the configuration of the LS-7000 Series, precautions, and necessary preparations required before operating the LS-7000 Series. Familiarize yourself with this section before using the LS-7000 Series.
Outline and Features of LS-7000 Series .............................. 1-2
System Applications ............................................................. 1-3
Package Contents ................................................................ 1-4
Functions and Nomenclature ............................................... 1-7
Mounting and Connection .................................................. 1-11
Supply Stand On and Off ................................................... 1-19
Setting Overview ................................................................ 1-20
Resetting to Initial Status ................................................... 1-21
1-1
Chapter 1 Getting Started

Outline and Features of LS-7000 Series

1
The LS-7000 Series is a high-speed, high-accuracy digital micrometer used for the dimen­sional measurement of objects without coming into contact with the objects. It is such a versatile model that it has a wide range of applications including in-line measurement and offline measurement.
Special Features
High-speed Sampling of 2400 Times per Second
Ensures high-speed sampling that is twice as fast as the sampling speed of conven­tional models. The continuous measurement of extruded products and the in-line measurement of moving workpieces are possible.
Repeat Accuracy of ±0.15
μm
Incorporates the latest optical system, thus ensuring excellent repeat accuracy that is twice as high as the accuracy of conventional models, thus supporting the manufacture of high-precision products.
Connecting Two Measuring Heads for two-channel Simultaneous Measurement
Two measuring heads can be used in combination for the simultaneous measurement of two targets.
Stable Detection of Transparent Objects with Threshold Change
The threshold change function supported by the DE processor makes it possible to detect transparent objects stably.
Principle of Measurement
The high-intensity GaN green LED radiates light, which will be changed into uniform parallel light through the special diffusion unit and collimator lens and emitted to the target in the measuring range. Then the shadow image of the target will appear on the HL-CCD (high-speed linear CCD) through the telecentric optical system. The output incident signal of the HL-CCD will be processed by the DE (digital edge-detection) processor in the controller and CPU. As a result, the dimensions of the target will be displayed and output.
1-2
HL-CCD for measurement use
Receiver Transmitter
A/D conversion
Telecentric optical system
Target
DE processor
Controller
High-intensity GaN green LED
Special diffusion unit
CPU
Collimator lens
Control input Control output BCD output Analog output RS-232C interface

System Applications

Chapter 1 Getting Started
The LS-7000 Series has a wide range of applications when used in combination with commercially available peripheral devices.
Printer
The results of measurement are printed.
RS-232C
PC
Control values or measuring values are retrieved over RS­232C.
RS-232C
LS-7001
Programmable Logic Controller
Control output and measured values are retrieved and programs for measurement timing control are changed.
Comparator output BCD output RS-232C Timing input Auto zero input Reset input Program number selection
1
Comparator output
Indicator Light and Buzzer
Alarms are generated according to the output of discrimination results.
Analog output
Recorder
The results of measurement are recorded.
Timing input
Photoelectric Sensor or Proximity Sensor
Timing input signals are sent whenever objects are detected.
1-3
Chapter 1 Getting Started

Package Checks

1
Check that nothing is missing from the LS-7000 Series package before use.

LS-7001

Controller Four Panel Mounting
Brackets
`
LS-7000 Series User’s Manual

LS-7010

Measuring Head
Allen-head bolt (Five, M3 x 45 with a washer)
Allen-head bolt (Four, M4 x 20)
96M1613
1-4

LS-7030

Measuring Head
Allen-head bolt (Six, M5 x 45 with a washer)
Allen-head bolt (Four, M4 x 20)

LS-7070

Chapter 1 Getting Started
Measuring Head
Allen-head bolt (Six, M4 x 50 with a washer)
Allen-head bolt (Four, M5 x 20)
LS-C✱✱A
Extension Cable (Cable between the controller and measuring head)
1
LS-C3A: 3-m cable
✱✱... LS-C10A: 10-m cable ✱✱... LS-C30A: 30-m cable
Up to two cables are connectable, provided that the total length of the cables does not exceed 40 m.
1-5
1
Chapter 1 Getting Started

LS-S11

Stand Unit Base Unit
Screws (Four, M4 x 8)
Keyence ships each package with utmost care and attention. If there should be any
improper or damaged product, contact your Keyence representative.
1-6

Functions and Nomenclature

Chapter 1 Getting Started
The following section provides information on the functions and nomenclature of the controller and measuring head.

Controller

Control Panel
(Page 1-8, Control Panel)
Controller (Rear Side)
Connector I/O
Used for signal I/O (Page 6-2, Section 6 I/O Pin)
Modular Connector
Used to connect the console connector or RS-232C cable connector. (Page 7-4, 7-5, Section 7 RS-232C)
Terminal Block
Used for signal I/O (Page 6-2, Section 6 I/O Terminals)
1
Controller (Bottom)
Power Input Pins (24 VDC)
Provide 24 VDC power supply.
DC IN
Connect the LS-S11 AC Power Supply Cable.
Extension Cable
Connect the cable between the controller and the measuring head.
1-7
1
Chapter 1 Getting Started
Control Panel
(1)
AREA
(3)
(4)
(5)
1 2
POSITION
FOCUS
OUT 1
OUT 2
HI
GO
LO
TIM
HI
GO
LO
(19) (20)
(21)
(2)
(17)
(6)
(7)
FUNC
ESCPROG
PRINT
ENT
PROG No.
AREA DISP
HOLD
ZERO
LS-7000
(16)
(15)
(8) (10) (9) (11) (12) (18) (13) (14)
(1) Main display: The measured value is displayed. Refer to Selection
of Measured Value Display on page 3-2.
(2) Sub display: Used to display tolerance or measured values while
the LS-7000 Series is in measuring operation. The set value is displayed while in value setting. Refer to Selection of Measured Value Display on page 3-2.
(3) Area monitor display: The number of the selected area is displayed. Refer
to Area Monitor on page 3-4.
(4) Position display: The position of the target in the selected area is
displayed. Refer to Area Monitor on page 3-4.
(5) Focus display: The focus status of the selected area is displayed.
Refer to Area Monitor on page 3-4.
(6)
(Setting) key: Press this key to set program details. Refer to
PROG
Program Changes on page 3-8.
(7)
(Function) key: Press this key to set the setting item. Refer to Section
FUNC
4 Function Settings. (8) (9) (10)
/ key: Press the Up or Down key to select the set item. key: Press this key to finish the present setting.
ESC
key: Press this key to enter the set item or print the meas-
ENT
ured value from the printers. (11) Program number display: The selected program number is displayed. (12) Program selection key: Press this key to select the program number when
measuring. (Page 3-8, Program Change) (13) Area monitor selection key: Press this key to select the area number when
measuring. (Page 3-4, Area Monitor)
1-8
Chapter 1 Getting Started
(14) Display selection key: Press this key to select the main display or sub
display when measuring. (Page 3-2, Switching Display Screen)
ZERO
(15)
key: Press this key to select the auto zero function. (Page
3-6, Automatic Zero Function)
(16)
key: Press this key to select the hold function. (Page 3-5,
HOLD
Hold Function)
(17)
, GO, and LO keys: Press these keys to make tolerance settings. (Page 4-
HI
50, Limit Setting)
and keys: Press these keys to select the set value or item.
(18) (19) Output selection display: Indicates whether the measured value on the main
display is OUT1 or OUT2. (20) Comparator result display: Displays the comparator result of the main display. (21) Timing display: Lit when timing input into the main display turns ON.

Measuring Head

LS-7010
Multi-purpose Mounting Holes
A fixture for targets can be attached.
Receiver (R Head)
Receives light from the transmitter.
Extension Cable
Connected to the controller.
Cover Glass
Transmitter (T Head)
Emits light for measurement.
TR Base
Used to mount the transmitter and receiver.
TR Cables
Used to connect the transmitter
Mounting Holes
Used to mount the measuring head.
1
1-9
1
Chapter 1 Getting Started
LS-7030
Multi-purpose Mounting Holes
A fixture for targets can be attached.
Receiver (R Head)
Receives light from the transmitter.
Extension Cable
Connected to the controller.
LS-7070
Multi-purpose Mounting Holes
A fixture for targets can be attached.
Receiver (R Head)
Receives light from the transmitter.
Cover Glass
Cover Glass
TR Base
Used to mount the transmitter and receiver.
TR Cables
Used to connect the transmitter and receiver.
Mounting Holes
Used to mount the measuring head.
Transmitter (T Head)
Emits light for measurement.
Transmitter (T Head)
Emits light for measurement.
1-10
Mounting Holes
Used to mount the measuring head.
Extension Cable
Connected to the controller.
TR Base
Used to mount the transmitter and receiver.
TR Cables
Used to connect the transmitter and receiver.
Chapter 1 Getting Started

Mounting and Connection

Mount the measuring head and the controller, and connect them with the cables.
Mounting the Measuring Head
The measuring head can be mounted with or without the TR base employed, according to the type of targets and operating environment. The following section provides information on how to mount the measuring head with or without the TR base employed, along with mounting conditions required.
Mounting the Controller
Mount the controller to panels, such as control or operation panels, or to the LS-S11 AC Power Supply Stand. The LS-S11 Power Supply Stand is sold separately.
Connection
After the measuring head and the controller are mounted, connect them with the ca­bles.

Mounting the Measuring Head

Mounting Restrictions
The transmitter and receiver must abide by the following restrictions when the measur­ing head is mounted.
Vertical Position Mismatching Angle Mismatching
1
±1 mm max.
±1 mm max.
The values are the same for LS-7010, LS-7030, and LS-7070.
TransmitterReceiver
TransmitterReceiver
±0.25° max
±0.25° max
±0.25° max
±0.25° max
TransmitterReceiver
TransmitterReceiver
TransmitterReceiver
TransmitterReceiver
1-11
1
Chapter 1 Getting Started
With TR Base
The transmitter and receiver are mounted to the TR base at the time of shipment. This section provides how to install the transmitter and receiver (LS-7030 in this case) that are mounted on the TR base.LS-7010 and LS-7070 can also be installed in the same way.
Use the four, M4x20 Allen-head bolts, which are provided with the package, to mount the TR base. To mount the TR base with the bolts on the surface of the TR base, use the mounting holes on the surface.
To mount the TR base with the bolts on the back of the TR base, use the 10-mm-deep M4 (10-mm-deep M5 for LS-7070) mounting holes on the back.
Without TR Base
Take the following steps to use the transmitter and receiver (LS-7030 in this case) after dismounting them from the TR base. LS-7010 and LS-7070 can also be used in the same way.
1 Loosen the bolts on the back of the
TR base and remove the measuring head.
1-12
2 Mount the measuring head.
Use the side mounting holes on the transmitter and receiver as shown in the illustration. Mount them with the Allen­head bolts provided with the package.
Use the 5-mm-deep M4 bottom mounting holes of the transmitter and receiver as shown in the illustration.

Mounting the Controller

Mounting Restrictions
Be aware of the following items when mounting the controller.
Chapter 1 Getting Started
1
CAUTION
Note: • For the maintenance, operability, and ventilation of the LS-7000
• Do not install the controller upside down.
• Do not block the ventilation louver. Otherwise, the interior heat may cause the controller to malfunction.
Series, separate the controller as far as possible from periph­eral structures or parts.
• The mounting angle must abide by the following restrictions.
30˚
Side
Display panel
• Check that the controller will not be exposed to the heat radia­tion of peripheral devices.
• Separate the controller as much as possible from devices with arc generation, such as electromagnetic switches or non-fuse breakers.
1-13
1
Chapter 1 Getting Started
Mounting to Panel
The controller can be mounted to control panels with the mounting brackets provided with the package. Arrange the following panel mounting dimensions.
207
+1
169
0
+1
124
0
200
Unit: mm
1 Mount the controller from the front
side of the panel.
1-14
2 Attach resin caps to the mounting
brackets.
3 Insert the mounting brackets to the
four mounting holes on the side of the controller respectively.
Mounting holes
Mounting holes
4 Secure the mounting brackets with a
Phillips screwdriver.
Chapter 1 Getting Started
1
CAUTION
• Tighten the mounting screws to a maximum torque of 0.4 Nm. Do not tighten up the mounting screws excessively, otherwise the panel or controller casing may be deformed.
• Do not remove the gasket, otherwise the controller will not satisfy IP64 conditions.
Mounting to AC Power Supply Stand
The following section provides information on how to mount the controller to the AC power supply stand (sold separately). The AC power supply stand consists of a stand unit and base unit. Mount the controller to the stand unit first. Connect the power supply cable and extension cable (i.e., the cable between the controller and measuring head). Then mount them to the base unit.
1 Remove the gasket around the
controller.
Gasket
2 Mount the controller to the stand.
Use the two screws provided with the package and secure the controller on the side.
The stand must be located on the groove from which the gasket was removed.
Secure these potions with screws.
1-15
1
Chapter 1 Getting Started
3 Connect the cable between the con-
troller and measuring head.
If only a single measuring head is used, connect the cable to the HEAD1 connec­tor.
Cable between controller and measuring head
4 Connect the power supply cable of
the AC power supply stand to the controller.
5 Mount the stand unit to the base unit.
Mount the stand unit with the two screws provided with the package.
1-16
(2) Lean the stand unit in the
direction indicated by the arrow. Pay utmost attention so that the connected cable will not be caught by the stand unit.
(1) Lean the stand unit in the
front direction and insert the tab.
(3) Secure these portions
with screws.
Loading...
+ 162 hidden pages