This instruction manual provides necessary information on the operation and maintenance
of the LS-7000 Series along with precautions. Please read the manual carefully and be
sure you understand the information provided before attempting to install and operate the
LS-7000 Series. Keep this manual handy for future reference.
Please make sure that the end users are provided with this manual.
Symbols
The following symbols are used for the list of precautions to ensure safety and to prevent
personal injury and/or property damage when using the LS-7000 Series.
The following symbols alert you to important messages. Be sure to read these messages
carefully.
DANGER
WARNING
CAUTION
Important:
Note:
Tips
Reference:
Failure to follow instructions may lead to death or serious injury.
Failure to follow instructions may lead to injury.
Failure to follow instructions may lead to product damage
(product malfunctions, etc.).
Provides important precautions and restrictions on proper operation.
Provides additional information on proper operation.
Provides useful information on proper operation.
Provides reference pages.
General precautions
1. No part of this manual may be reprinted or reproduced in any form or by any means
without the prior written permission of KEYENCE CORPORATION.
2. The content of this manual is subject to change without notice.
3. KEYENCE has thoroughly checked and reviewed this manual. Please contact the sales
office listed at the end of this manual if you have any questions or comments regarding
this manual, or if you find an error.
4. KEYENCE assumes no liability for damages resulting from the use of the information in
this manual, item 3 above notwithstanding.
5. KEYENCE will replace any incomplete or incorrectly collated manual.
6. All company names and product names in this manual are registered trademarks or
trademarks of their respective owners.
Safety Precautions
■ General precautions
• At startup and during operation, be sure to monitor the functions and performance of
the LS-7000 Series.
•Take sufficient safety measures to prevent damage to the human body and/or equipment that may be caused if this product should fail to operate properly.
• Do not modify the LS-7000 Series or use it in any way other than as described in the
specifications. Its functions and performance are not guaranteed under said
conditions.
• When the LS-7000 Series is used in combination with other instruments, its functions
and performance may be degraded, depending on the operating conditions and
surrounding environment.
• Do not use the LS-7000 Series for the purpose of protecting the human body.
• Do not expose the LS-7000 Series and peripheral devices to sudden temperature
change, as this may cause condensation.
■ Precautions
WARNING
■ Operation
• Always use the LS-7000 Series at 24 VDC, otherwise a fire,
electric shock or product failure may result.
• Prepare the applicable AC power supply cable if you use LSS11.
• Secure the GND line of the power cable of the AC power
supply stand to protective ground, otherwise an electric shock
or product failure may result.
• Do not disassemble or modify the LS-7000 Series.
This may cause fire or electric shock.
■ When abnormal conditions are encountered
If the following conditions are encountered, immediately turn off
the power. Continuing to use the LS-7000 Series. under abnormal conditions may cause fire, electric shock or equipment
failure.
Contact your nearest KEYENCE sales office for repairs.
•When water or foreign matter enters the controller.
• When the LS-7000 Series. is dropped or the housing is
damaged.
• When the controller produces smoke or an abnormal smell.
96M12219
i
CAUTION
■ Usage
• Be sure to turn off the power to the LS-7000 Series and any
connected devices before connecting or disconnecting the
cables. Otherwise, the camera and connected devices may be
damaged.
• Do not turn off the power while setting a parameter. Otherwise,
the settings may be partially or completely lost.
• Do not block the ventilation slots on the LS-7000 Series and
peripheral devices. A rise in inner temperature may cause
equipment failure.
■ Proper environment and conditions
To use the LS-7000 Series properly and safely, do not install
the LS-7000 Series in locations with the following conditions.
Use of this equipment in an improper environment may cause
fire, electric shock, or equipment failure.
• Locations with high humidity, a large amount of dust, or poor
ventilation
• Locations where the temperature rises excessively due to
direct sunlight, etc.
• Locations where corrosive or flammable gas exists
• Locations where the LS-7000 Series is directly subjected to
vibration or impact
• Locations where water, oil or chemicals may splash the LS7000 Series
• Locations where static electricity is easily built up
■ Countermeasures against Noise
The LS-7000 Series may malfunction or fail to operate due to
noise generated from power lines or high-tension lines. If the LS7000 Series is operated close to such noise sources, use a noise
filter, wire the cable of the LS-7000 Series in an independent
conduit, or locate the LS-7000 Series in an area with proper
insulation from noise.
ii
Note:■ Influence of Ambient Operating Temperature
Ensure that the ambient operating temperature is constant. A
change in the ambient operating temperature may result in
measurement errors. If the ambient operating temperature
changes by 10°C, it will take approximately 60 minutes for the
interior temperature distribution of the LS-7000 Series to be
uniform.
■ Warming Up
Do not operate the LS-7000 Series for approximately 30 minutes
after it is turned on. The internal circuitry of the LS-7000 Series is
not stable immediately after it is turned on. The measured value
drifts gradually as a result.
■ Influence of Dust and Dirt
A measurement error may result due to dirt, dust, water, or oil in
any of the following cases.
• Cover glass: Blow off the dirt, dust, water or oil on the cover
glass with clean air. If the cover glass is excessively dirty, wipe
the glass with a soft cloth moistened with isopropyl alcohol.
• Target surface: Blow off the dirt, dust, water or oil on the surface with clean air or wipe the surface.
• Intrusion of dirt, dust, water or oil blown or sprayed into optical
sensing area: Install a protection cover or perform air purging.
■ Influence of Vibration
The measured value of an object may fluctuate if the object is
vibrating. If that happens, increase the averaging number. This
will ensure highly accurate measurement.
■ Targets
The measured value of an object may have an error due to the
shape and surface condition. Locate a reference object within the
measuring area and use the calibration function of the LS-7000
Series to correct the error if that happens.
■ Influence of Air Flow
The measured value of an object may fluctuate due to a slow
flow of air. In that case, the following countermeasures are
effective.
• Protect the measuring head with a cover.
• Use a fan to mix the air vigorously between the transmitter and
receiver.
■ Maintenance
Do not wipe the LS-7000 Series with a damp cloth or a cloth
moistened with benzine or paint thinner. This may cause the
discoloration or deformation of the LS-7000 Series. If the LS7000 Series is excessively dirty, use a tightly squeezed cloth
moistened with thin neutral detergent. Then wipe the LS-7000
Series with a soft dry cloth.
■ Precautions for LED
The LS-7010/LS-7030/LS-7070 Measuring Head uses an LED light source. Abide by the
following precautions when using the unit.
CAUTION
• Do not look at the LED light source for a long time.
The LED used as the light source is classified into class 1
according to IEC60825-1 standards. Basically the light beam of
the LED is safe, but do not look at the LED for a long time.
• Do not disassemble the unit.
The unit does not incorporate a mechanism to stop the emission
of the LED when the unit is disassembled. Never attempt to
disassemble the unit, otherwise you will be exposed to LED light.
iii
Precautions for CE Markings
Keyence confirms that the LS-7000 Series meets EC directive requirements. The LS-7000
Series bears CE markings. Keep the following conditions if the LS-7000 Series is used in
European countries.
■ EMC Directives
Keep the following condition so that the LS-7000 Series will satisfy EN61326-1
requirements.
• The power cable and I/O cable connected to the controller are both less than 30m in
length.
• Attach the OP-31973 Clamp-type Ferrite Core to the cable between the controller and
measuring head. The attaching position is as close as possible to the cable end on
the controller side, where the core must be snapped onto the cable to form a one-turn
loop.
Ferrite core
OP-31973
■ Low-voltage Directives
When using the LS-S11 AC Power Supply Stand as a dedicated power supply to the LS7000 Series, keep the following conditions so that the LS-S11 will satisfy EN61010-1
requirements. The LS-S11 satisfies EN60825-1 (LED Class 1) requirements with no
conditions attached.
Environment Conditions
Installation category (Overvoltage category): II (see note 1)
Pollution degree: 2 (see note 2)
Operation and Installation Conditions
• The LS-S11 AC Power Stand is a dedicated power supply for the LS-7000 Series.
Never use the LS-S11 for other purposes.
• When using the LS-7000 Series in European countries, check that the AC power
supply cable satisfies EC directive requirements certified by an authorized third party.
Consult your Keyence representative for the AC power supply cable in detail.
• Before turning the LS-7000 Series on, connect the ground terminal of the AC power
supply cable to the protective ground terminal of the power supply.
• Be sure to use a time lag fuse rated 250 V, 3.15 A for the replaceable fuse of the LSS11.
Note 1: Installation categories (overvoltage categories) include impulse overvoltage regu-
Note 2: The pollution degree refers to the level of solid, liquid, or gas contamination that
iv
lations at transient overvoltage levels specified by EC directives.
Installation category II indicates a level where power is provided from fixed
installations, such as distribution panels.
deteriorates the dielectric strength or surface resistance of the LS-7000 Series.
Pollution degree 2 indicates the normal indoor environment (free of non-conductive contamination).
About this Manual
The following section provides information on the configuration of each page of this manual
along with symbols and terms used in this manual.
Page Configuration and Symbols
Chapter 3 Operation Control
Automatic Zero Function
The automatic zero function sets measuring values to zero (0.00000) instantly. This function is convenient for zero-point calibration for a variety of target. Furthermore,
master calibration is possible by using this function in combination with the offset function.
Tips: Describes
information that will
deepen the user’s
comprehension of
the manual along
with other useful
information.
Caution: Indicates
information that, if
not heeded, could
result in relatively
serious or minor
injury, damage to
the LS-7000 Series,
or faulty operation.
Note:This function will not be available while the controller is awaiting the result of
measurement, during which the LCD monitor will display “– – –. – – – –.”)
The automatic zero value will not be lost after the LS-7500 Series is turned
Tips
off. The automatic zero value for each program will be stored.
Automatic zero settings are made through the following functional components.
• Control panel
• External terminal
3
• RS-232C interface
■ Control Panel
If a single measuring value is displayed, an
automatic zero setting will be made for the
corresponding OUT number only.
If both measuring values are displayed, an
automatic zero setting will be made for the
selected OUT number on the screen.
1
Press the [ZERO] key once. Automatic
zero will be set.
To cancel the setting, press the [ZERO]
key again.
■ External I/O Terminal
Automatic zero will be set by short-circuiting the I/O, ZERO, and COM terminals on the
rear panel of the controller together.
Refer to Section 7 I/O Terminals on page 7-1 for details.
CAUTION
Tips
3-10
• Do not look at the LED light source for a long time.
The LED used as the light source is classified into class 1
according to IEC60825-1 standards. Basically the light beam of
the LED is safe, but do not look at the LED for a long time
Master Calibration with Offset Function (Automatic Offset Function)
Set the size of a master workpiece to an offset value and make an automatic
zero setting while measuring the master workpiece. Then the size of the
master workpiece will be displayed as an offset value. Refer to 3-3 Measurement of Outer Diameter and Width (with Two Measuring Heads) on page 3-4
for an application example.
Refer to Offset on page 5-43 for the offset function in detail.
AREA DISP
HOLD
ZERO
Note: Describes
items that need the
user’s attention in
order to prevent
mistakes in the
operation of the
LS-7000 Series.
❈ The above page sample is for your reference only. It does not coincide with any of
the actual pages of this manual.
Definition of Terms
This manual uses the following terms.
TermDescription
LS-7000 SeriesRefers to the LS-7001 controller and the measuring head as a set.
ControllerRefers to the LS-7001 controller.
Measuring headRefers to the LS-7010/LS-7030/LS-7070 measuring head.
v
Optimum Measurement Methods
Refer to the following methods, and select the optimum measurement method according to
the type of target.
Measurement of Outer Diameter
of a Round Bar or Wire
Measurement of Average Outer
Diameter of Round Bar or Wire
Measurement of Sheet Width
D
Measurement of outer diameter and
width (with single measuring head)
(Page 2-2)
Measurement of Inner
Diameter of a Disc
Measurement of inner diameter and
clearance (Page 2-5)
Measurement of Connector
Pins and IC Pins
Y
X
Measurement of outer diameter and
width (with two measuring heads in X
and Y directions) (Page 2-3)
Measurement of Outer Diameter
and Eccentricity of a Roller
Measurement of outer diameter and
eccentricity (Page 2-6)
L
Measurement of outer diameter and
width (with two measuring heads for
larger objects) (Page 2-4)
Measuring Robot Movement
and Positioning of LCD Plate
Measurement of movement and
positioning (Page 2-7)
Measurement of pitch (Page 2-8)
vi
Configuration of Manual
Chapter 1Getting Started
Provides information on precautions and necessary preparations
required.
Chapter 2Easy Setting Guide
Provides information on targets and settings for typical applications
respectively.
Chapter 3Operation Control
Provides information on setting control items on the control panel while
the LS-7000 Series is in measurement operation.
Chapter 4Function Settings
Provides information on the overview of program settings and setting
methods.
Chapter 5Environment Settings
Provides information on basic operation settings.
Chapter 6I/O Terminals
Provides information on the specifications of the I/O terminals and
timing charts.
1
2
3
4
5
6
Chapter 7RS-232C
Provides information on how to connect external devices and the
communications function to operate the LS-7000 Series through the
external devices.
Chapter 8Specifications
Provides the specifications, characteristics, and external dimensions of
the LS-7000 Series.
Chapter 9Troubleshooting
Provides information on troubleshooting and error messages.
Appendix
Provides a list of options and the index of the manual.
7
8
9
Appendix
vii
Measured vabetween the r
setting. (See pa
LS-7001 Operation Map
Select the area (See page 3-4)
Press the Up or Down key to
select the area.
AREA
The selected area appears on the
AREA display.
POSITION monitor
The POSITION monitor is lit according to the
position and size of the target.
FOCUS monitor
The FOCUS monitor is lit according
to the distance between the
transmitter and receiver.
To make lit range changes, refer to
page 4-14.
Program Settings
Area, calibration, output, or option
setting:
Press the key and select the
item with key.
PROG
FUNC
AREA
1
2
FUNC
POSITION
FOCUS
ESCPROG
PRINT
ENT
No.
PROG No.
Environment setting:
Press the key while pressing
the key.
Set item selection:
Press the or key.
Set item cancellation:
Press the key.
Entering set item:
Press the key.
viii
FUNC
PROG
ESC
ENT
Program selection
Press the Up or Down key.
PROG No.
OUT 1
OUT 2
HI
GO
LO
TIM
HI
GO
LO
HOLD
ZERO
LS-7000
Tolerance setting (See page 4-50)
Reference value mode
Upper limit setting:
Reference value setting:
Lower limit setting:
Press the
Press the
Press the
HI
GO
LO
Threshold setting
HI value setting:
HH value setting:
LO value setting:
LL value setting:
Press the
Press the
Press the
Press the
HI
HI
LO
LO
Value change while setting
LO
Press the key
below the value
HOLD
ZERO
flashing.
Hold function (See page 3-5)
key.
key.
key.
key.
key 2X.
key.
key 2X.
Measured value display selection
Press the Up or Down key.
DISP
Measured value display varies
between the reference value mode
and threshold mode for tolerance
setting. (See page 3-2)
List of Optional Accessories........................................................................... Appendix-2
Setting Record Sheet ....................................................................................... Appendix-3
Environment Setting Record Sheet ................................................................ Appendix-5
Index ................................................................................................................. Appendix-6
xii
Chapter 1
Getting Started
This section provides information on the configuration of the
LS-7000 Series, precautions, and necessary preparations
required before operating the LS-7000 Series. Familiarize
yourself with this section before using the LS-7000 Series.
Outline and Features of LS-7000 Series .............................. 1-2
System Applications ............................................................. 1-3
Resetting to Initial Status ................................................... 1-21
1-1
Chapter 1 Getting Started
Outline and Features of LS-7000 Series
1
The LS-7000 Series is a high-speed, high-accuracy digital micrometer used for the dimensional measurement of objects without coming into contact with the objects. It is such a
versatile model that it has a wide range of applications including in-line measurement and
offline measurement.
■ Special Features
High-speed Sampling of 2400 Times per Second
Ensures high-speed sampling that is twice as fast as the sampling speed of conventional models. The continuous measurement of extruded products and the in-line
measurement of moving workpieces are possible.
Repeat Accuracy of ±0.15
μm
Incorporates the latest optical system, thus ensuring excellent repeat accuracy that is
twice as high as the accuracy of conventional models, thus supporting the manufacture
of high-precision products.
Connecting Two Measuring Heads for two-channel Simultaneous Measurement
Two measuring heads can be used in combination for the simultaneous measurement
of two targets.
Stable Detection of Transparent Objects with Threshold Change
The threshold change function supported by the DE processor makes it possible to
detect transparent objects stably.
■ Principle of Measurement
The high-intensity GaN green LED radiates light, which will be changed into uniform
parallel light through the special diffusion unit and collimator lens and emitted to the
target in the measuring range. Then the shadow image of the target will appear on the
HL-CCD (high-speed linear CCD) through the telecentric optical system. The output
incident signal of the HL-CCD will be processed by the DE (digital edge-detection)
processor in the controller and CPU. As a result, the dimensions of the target will be
displayed and output.
1-2
HL-CCD for
measurement use
ReceiverTransmitter
A/D
conversion
Telecentric optical
system
Target
DE
processor
Controller
High-intensity GaN
green LED
Special diffusion unit
CPU
Collimator lens
Control input
Control output
BCD output
Analog output
RS-232C interface
System Applications
Chapter 1 Getting Started
The LS-7000 Series has a wide range of applications when used in combination with
commercially available peripheral devices.
Printer
The results of measurement
are printed.
RS-232C
PC
Control values or measuring
values are retrieved over RS232C.
RS-232C
LS-7001
Programmable Logic
Controller
Control output and measured
values are retrieved and
programs for measurement
timing control are changed.
Comparator output
BCD output
RS-232C
Timing input
Auto zero input
Reset input
Program number selection
1
Comparator output
Indicator Light and Buzzer
Alarms are generated according to
the output of discrimination
results.
Analog output
Recorder
The results of
measurement are
recorded.
Timing input
Photoelectric Sensor or
Proximity Sensor
Timing input signals are sent
whenever objects are detected.
1-3
Chapter 1 Getting Started
Package Checks
1
Check that nothing is missing from the LS-7000 Series package before use.
LS-7001
ControllerFour Panel Mounting
Brackets
`
LS-7000 Series
User’s Manual
LS-7010
Measuring Head
Allen-head bolt
(Five, M3 x 45 with a washer)
Allen-head bolt
(Four, M4 x 20)
96M1613
1-4
LS-7030
Measuring Head
Allen-head bolt
(Six, M5 x 45 with a washer)
Allen-head bolt
(Four, M4 x 20)
LS-7070
Chapter 1 Getting Started
Measuring Head
Allen-head bolt
(Six, M4 x 50 with a washer)
Allen-head bolt
(Four, M5 x 20)
LS-C✱✱A
Extension Cable (Cable between the controller and measuring head)
Up to two cables are connectable, provided that the
total length of the cables does not exceed 40 m.
1-5
1
Chapter 1 Getting Started
LS-S11
Stand UnitBase Unit
Screws (Four, M4 x 8)
❈ Keyence ships each package with utmost care and attention. If there should be any
improper or damaged product, contact your Keyence representative.
1-6
Functions and Nomenclature
Chapter 1 Getting Started
The following section provides information on the functions and nomenclature of the
controller and measuring head.
Controller
Control Panel
(Page 1-8, Control Panel)
Controller (Rear Side)
Connector I/O
Used for signal I/O
(Page 6-2, Section 6 I/O Pin)
Modular Connector
Used to connect the console connector
or RS-232C cable connector.
(Page 7-4, 7-5, Section 7 RS-232C)
Terminal Block
Used for signal I/O
(Page 6-2, Section 6 I/O Terminals)
1
Controller (Bottom)
Power Input Pins (24 VDC)
Provide 24 VDC power supply.
DC IN
Connect the LS-S11 AC Power Supply
Cable.
Extension Cable
Connect the cable between the controller and
the measuring head.
1-7
1
Chapter 1 Getting Started
■ Control Panel
(1)
AREA
(3)
(4)
(5)
1
2
POSITION
FOCUS
OUT 1
OUT 2
HI
GO
LO
TIM
HI
GO
LO
(19)
(20)
(21)
(2)
(17)
(6)
(7)
FUNC
ESCPROG
PRINT
ENT
PROG No.
AREA DISP
HOLD
ZERO
LS-7000
(16)
(15)
(8) (10) (9) (11) (12)(18)(13) (14)
(1) Main display:The measured value is displayed. Refer to Selection
of Measured Value Display on page 3-2.
(2) Sub display:Used to display tolerance or measured values while
the LS-7000 Series is in measuring operation. The set
value is displayed while in value setting. Refer to
Selection of Measured Value Display on page 3-2.
(3) Area monitor display:The number of the selected area is displayed. Refer
to Area Monitor on page 3-4.
(4) Position display:The position of the target in the selected area is
displayed. Refer to Area Monitor on page 3-4.
(5) Focus display:The focus status of the selected area is displayed.
Refer to Area Monitor on page 3-4.
(6)
(Setting) key:Press this key to set program details. Refer to
PROG
Program Changes on page 3-8.
(7)
(Function) key:Press this key to set the setting item. Refer to Section
FUNC
4 Function Settings.
(8)
(9)
(10)
/ key:Press the Up or Down key to select the set item.
key:Press this key to finish the present setting.
ESC
key:Press this key to enter the set item or print the meas-
ENT
ured value from the printers.
(11) Program number display:The selected program number is displayed.
(12) Program selection key:Press this key to select the program number when
measuring. (Page 3-8, Program Change)
(13) Area monitor selection key: Press this key to select the area number when
measuring. (Page 3-4, Area Monitor)
1-8
Chapter 1 Getting Started
(14) Display selection key:Press this key to select the main display or sub
display when measuring. (Page 3-2, Switching
Display Screen)
ZERO
(15)
key:Press this key to select the auto zero function. (Page
3-6, Automatic Zero Function)
(16)
key:Press this key to select the hold function. (Page 3-5,
HOLD
Hold Function)
(17)
, GO, and LO keys: Press these keys to make tolerance settings. (Page 4-
HI
50, Limit Setting)
and keys:Press these keys to select the set value or item.
(18)
(19) Output selection display:Indicates whether the measured value on the main
display is OUT1 or OUT2.
(20) Comparator result display:Displays the comparator result of the main display.
(21) Timing display:Lit when timing input into the main display turns ON.
Measuring Head
LS-7010
Multi-purpose
Mounting Holes
A fixture for targets can
be attached.
Receiver (R Head)
Receives light from
the transmitter.
Extension Cable
Connected to the controller.
Cover Glass
Transmitter (T Head)
Emits light for
measurement.
TR Base
Used to mount the
transmitter and receiver.
TR Cables
Used to connect the transmitter
Mounting Holes
Used to mount the
measuring head.
1
1-9
1
Chapter 1 Getting Started
LS-7030
Multi-purpose
Mounting Holes
A fixture for targets can
be attached.
Receiver (R Head)
Receives light from
the transmitter.
Extension Cable
Connected to the controller.
LS-7070
Multi-purpose
Mounting Holes
A fixture for targets can
be attached.
Receiver (R Head)
Receives light from
the transmitter.
Cover Glass
Cover Glass
TR Base
Used to mount the
transmitter and receiver.
TR Cables
Used to connect the
transmitter and receiver.
Mounting Holes
Used to mount the measuring head.
Transmitter (T Head)
Emits light for
measurement.
Transmitter (T Head)
Emits light for
measurement.
1-10
Mounting Holes
Used to mount the measuring head.
Extension Cable
Connected to the controller.
TR Base
Used to mount the
transmitter and receiver.
TR Cables
Used to connect the
transmitter and receiver.
Chapter 1 Getting Started
Mounting and Connection
Mount the measuring head and the controller, and connect them with the cables.
■ Mounting the Measuring Head
The measuring head can be mounted with or without the TR base employed, according
to the type of targets and operating environment.
The following section provides information on how to mount the measuring head with or
without the TR base employed, along with mounting conditions required.
■ Mounting the Controller
Mount the controller to panels, such as control or operation panels, or to the LS-S11 AC
Power Supply Stand. The LS-S11 Power Supply Stand is sold separately.
■ Connection
After the measuring head and the controller are mounted, connect them with the cables.
Mounting the Measuring Head
■ Mounting Restrictions
The transmitter and receiver must abide by the following restrictions when the measuring head is mounted.
Vertical Position MismatchingAngle Mismatching
1
±1 mm max.
±1 mm max.
The values are the same
for LS-7010, LS-7030, and
LS-7070.
TransmitterReceiver
TransmitterReceiver
±0.25° max
±0.25° max
±0.25° max
±0.25° max
TransmitterReceiver
TransmitterReceiver
TransmitterReceiver
TransmitterReceiver
1-11
1
Chapter 1 Getting Started
■ With TR Base
The transmitter and receiver are mounted to the TR base at the time of shipment. This
section provides how to install the transmitter and receiver (LS-7030 in this case) that
are mounted on the TR base.LS-7010 and LS-7070 can also be installed in the same
way.
Use the four, M4x20 Allen-head bolts, which
are provided with the package, to mount the
TR base. To mount the TR base with the
bolts on the surface of the TR base, use the
mounting holes on the surface.
To mount the TR base with the bolts on the
back of the TR base, use the 10-mm-deep
M4 (10-mm-deep M5 for LS-7070) mounting
holes on the back.
■ Without TR Base
Take the following steps to use the transmitter and receiver (LS-7030 in this case) after
dismounting them from the TR base. LS-7010 and LS-7070 can also be used in the
same way.
1 Loosen the bolts on the back of the
TR base and remove the measuring
head.
1-12
2 Mount the measuring head.
Use the side mounting holes on the
transmitter and receiver as shown in the
illustration. Mount them with the Allenhead bolts provided with the package.
Use the 5-mm-deep M4 bottom mounting
holes of the transmitter and receiver as
shown in the illustration.
Mounting the Controller
■ Mounting Restrictions
Be aware of the following items when mounting the controller.
Chapter 1 Getting Started
1
CAUTION
Note:• For the maintenance, operability, and ventilation of the LS-7000
• Do not install the controller upside down.
• Do not block the ventilation louver. Otherwise, the interior heat
may cause the controller to malfunction.
Series, separate the controller as far as possible from peripheral structures or parts.
• The mounting angle must abide by the following restrictions.
30˚
Side
Display
panel
• Check that the controller will not be exposed to the heat radiation of peripheral devices.
• Separate the controller as much as possible from devices with
arc generation, such as electromagnetic switches or non-fuse
breakers.
1-13
1
Chapter 1 Getting Started
■ Mounting to Panel
The controller can be mounted to control panels with the mounting brackets provided
with the package. Arrange the following panel mounting dimensions.
207
+1
169
0
+1
124
0
200
Unit: mm
1 Mount the controller from the front
side of the panel.
1-14
2 Attach resin caps to the mounting
brackets.
3 Insert the mounting brackets to the
four mounting holes on the side of
the controller respectively.
Mounting holes
Mounting holes
4 Secure the mounting brackets with a
Phillips screwdriver.
Chapter 1 Getting Started
1
CAUTION
• Tighten the mounting screws to a maximum torque of 0.4 Nm.
Do not tighten up the mounting screws excessively, otherwise
the panel or controller casing may be deformed.
• Do not remove the gasket, otherwise the controller will not
satisfy IP64 conditions.
■ Mounting to AC Power Supply Stand
The following section provides information on how to mount the controller to the AC
power supply stand (sold separately). The AC power supply stand consists of a stand
unit and base unit. Mount the controller to the stand unit first. Connect the power supply
cable and extension cable (i.e., the cable between the controller and measuring head).
Then mount them to the base unit.
1 Remove the gasket around the
controller.
Gasket
2 Mount the controller to the stand.
Use the two screws provided with the
package and secure the controller on the
side.
The stand must be located
on the groove from which
the gasket was removed.
Secure these
potions with screws.
1-15
1
Chapter 1 Getting Started
3 Connect the cable between the con-
troller and measuring head.
If only a single measuring head is used,
connect the cable to the HEAD1 connector.
Cable between
controller and
measuring head
4 Connect the power supply cable of
the AC power supply stand to the
controller.
5 Mount the stand unit to the base unit.
Mount the stand unit with the two screws
provided with the package.
1-16
(2) Lean the stand unit in the
direction indicated by the
arrow. Pay utmost attention so
that the connected cable will
not be caught by the stand unit.
(1) Lean the stand unit in the
front direction and insert
the tab.
(3) Secure these portions
with screws.
Loading...
+ 162 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.