This manual describes how to install the LS-3100(W) Series as well as its
operating procedures and precautions. Please read this manual carefully to
get the best from your LS-3100(W) Series.
Symbols
General Pecautions
The following symbols alert you to important messages. Be sure to read
these messages carefully.
WARNING
CAUTION
Note
• At startup and during operation, be sure to monitor the functions and
performance of the LS-3100(W) series.
• We recommend that you take substantial safety measures to avoid any
damage in the event a problem occurs.
• Do not open or modify the LS-3100(W) series or use it in any way other
than described in the specifications.
• When the LS-3100(W) series is used in combination with other instruments, functions and performance may be degraded, depending on
operating conditions and the surrounding environment.
Failure to follow instructions may lead to injury.
(electric shock, burn, etc.)
Failure to follow instructions may lead to product
damage.
Provides additional information on proper operation.
• Do not use the LS-3100(W) series for the purpose of protecting the
human body.
■ Best Management Practice for Perchlorate Materials — California
only
When you sell, manufacture and/or waste the products containing
perchlorate material in California, the following statement must appear on
the exterior of all outer shipping packages and on consumer packages,
based on California Code of Regulations.
“Perchlorate Material – special handling may apply, See
www.dtsc.ca.gov/hazardouswaste/perchlorate.”
Note: Alternative option is to indicate this statement on your MSDS or
manual accompanying with each product.
This product includes a CR coin type cell containing perchlorate material. If
you ship this product (or your product including this product) to California,
you must ensure to comply with this regulation.
Unpacking
Check that the following items are included in your LS system package.
CAL (Calibration) ......................................................................................96
WARRANTIES AND DISCLAIMERS97
CHAPTER 1
AVOID EXPOSURE
LASER RADIATION
IS EMITTED FROM
THIS APERTURE.
CAUTION-Laser radiation
when open.
DO NOT STARE
INTO BEAM.
LASER SAFETY PRECAUTIONS
1-1. Classification
Laser
FDA (CDRH) 21CFR
Class
Part 1040.10
IEC60825-1Class 1
DIN EN 60825-1Class 1
1-2. Warning Labels
1) Warning label
FDA Class II
CAUTION
LASER RADIATION-DO NOT
STARE INTO BEAM
SEMICONDUCTOR LASER 670nm
MAXIMUM OUTPUT 0.5mW
CLASS II LASER PRODUCT
LS-3034 LS-3033 LS-3032 LS-3033 LS-3036 LS-3060
SO(8073)
Class IClass II
2) Aperture label
FDA Class II
AVOID EXPOSURE
LASER RADIATION
IS EMITTED FROM
THIS APERTURE.
CAUTION-Laser radiation
when open.
DO NOT STARE
INTO BEAM.
1
CHAPTER 1 Preparation for Measurement
CAUTION
AVOID EXPOSURE
LASER RADIATION
IS EMITTED FROM
THIS APERTURE.
CAUTION-Laser radiation
when open.
DO NOT STARE
INTO BEAM.
AVOID EXPOSURE
1-3. Label Location
1-4Safety Consideration
CAUTION
LASER RADIATION-DO NOT
STARE INTO BEAM
SEMICONDUCTOR LASER 670nm
MAXIMUM OUTPUT 0.5mW
CLASS II LASER PRODUCT
1)2)
AVOID EXPOSURE
LASER RADIATION
IS EMITTED FROM
THIS APERTURE.
CAUTION-Laser radiation
when open.
DO NOT STARE
INTO BEAM.
: Aperture
CAUTION
WARNING
Use of controls or adjustments or the performance of procedures other than
those specified herein may result in hazardous radiation exposure.
The possible health hazard is in exposing the eyes to the laser beam.
Damage to the eyes can occur if the operator stares directly into the beam.
Do not look directly at the laser beam.
Follow the safety precautions below to ensure operator safety:
• Operate the LS-3100(W) series only according to the procedures
described in this instruction manual.
Otherwise, injury may occur due to exposure to the laser beam.
• Do not disassemble the sensor head.
Laser emission from the LS-3100(W) series is not automatically stopped if
the sensor head is disassembled. If you disassemble the sensor head for
inspection or repair, you may be exposed to the laser beam. If the LS3100(W) series malfunctions, contact KEYENCE immediately.
• Do not look directly at the laser beam.
Looking directly at the laser beam may result in serious eye injury.
• Protective enclosure
We recommend that you install a protective enclosure around the sensor
head to prevent any person from getting near the sensor head during
operation.
• Protective goggles
We recommend that you wear protective goggles when using the LS3100(W) series.
2
CHAPTER 1 Preparation for Measurement
1-5Safety Features Provided with the LS-3100(W) Series
The LS-3100(W) series has the following safety features:
■ Laser ON alarm LED
Both the sensor head and the controller panel have a visible LED that lights
when laser is ready to be and is being emitted.
LEDs can be checked to see if they are lit even when you are wearing laser
protective glasses.
■ Delay of laser beam emission
To prevent an operator from being exposed to the laser beam, the laser
beam is emitted three seconds after the laser ON alarm LED lights.
■ Laser emission remote control input connector
The laser emission control connector is located on the rear panel of the
control unit. The laser can be turned on or off by a remote control signal
through this connector.
■ Key-operated power switch
The controller power switch can be locked using the attached key.
When the LS-3100(W) controller is OFF, the key can be removed.
■ Laser beam shield
The sensor head transmitter comes with a laser beam shield.
Cover is closed.Cover is open.
LS-3032/3034/3036LS-3032/3034/3036
LS-3033/3033SO(8073)LS-3033/3033SO(8073)
Laser emission
Laser emission
LS-3060LS-3060
Laser emission
3
CHAPTER 1 Preparation for Measurement
1-6. Preparation for Operation
CAUTION
CAUTION
• Do not mount the scanning head to the place where a certain level of
electrical noise is applied or transferred. Isolate the scanning head if
electrical noise may be occured. Otherwise, laser diode may deteriorate
or become damaged.
• Although the LS controller has been thoroughly inspected before shipment, we kindly request that you check it upon purchase for any damage.
Contact the nearest KEYENCE distributor if your LS controller is damaged
in any way.
• Precision components are incorporated in the scanning head, therefore be
sure not to subject it to shocks.
• The scanning head has been precision set before shipment. Always keep
the casing closed to maintain the head's setting.
• The LS controller can operate normally at 0 to 45°C (32 to 113°F) [0 to
40°C (32 to 104°F) for LS-3036/3060] with no freezing and 35 to 85%RH
without condensation. Operation outside these ranges may result in
malfunction.
Do not connect or disconnect any cables while the power is ON. Otherwise,
the laser diode may deteriorate and/or may be damaged.
Note
1. Connect the controller to the scanning head and the transmitter and
receiver using the supplied cables.
LS-3032, LS-3034, LS-3036
LS-3033, LS-3033 SO, LS-3060
The LS-3101(W) can be used with LS-3032, 3033, 3036 and 3060.
4
CHAPTER 1 Preparation for Measurement
2. Ground the controller. (Except for models followed with “W”.)
Grounding the controller to the
grounding bar
3. Check first that the power switch of the controller is OFF, then connect it to
the outlet using the supplied power cable.
Controller (rear view)
LS-3100 Series
Grounding the
controller through the
grounding outlet
LS-3100(W) Series
I
O
4. Check that laser can be safely emitted, then turn the power switch ON.
Allow the controller to warm-up for approximately 30 minutes after the
laser emission indicator LED has lit.
With the LS-3100(W) series, first turn on the POWER switch (at the rear),
and then turn on the LASER switch (on the front panel).
Controller (front view)
LS-3100 Series
LIMT
HIGH GO LOW
1 2
LASER LASER
LS
LASER SCAN DIAMETER
• OFF
• ON
POWER
OFFSET ZERO HOLD LOCK POSITION
LS-3100(W) Series
LIMT
HIGH GO LOW
1 2
LASER LASER
LS
LASER SCAN DIAMETER
•
•
LASER
OFFSET ZERO HOLD LOCK POSITION
SEGSETMODE
AVE LIMIT OF S
PRM FNC CAL
PROG DISP
SEGSETMODE
AVE LIMIT OF S
PRM FNC CAL
PROG DISP
mm
inch
7
8
9
ZERO
4
5
6
HOLD
1
2
3
CLR
0
.
+/-
ENT
mm
inch
7
8
9
ZERO
4
5
6
HOLD
1
2
3
CLR
0
.
+/-
ENT
5
CHAPTER 1 Preparation for Measurement
1-7. Connecting 2 Scanning Heads
CAUTION
Do not connect or disconnect any cables while the power is ON. Otherwise,
the laser diode may deteriorate and/or may be damaged.
Note the following for connecting 2 scanning heads to the controller:
• These heads have been factory-set for use either as the HEAD1 or
HEAD2. Incorrect connection of the scanning heads will, therefore, result
in malfunction. To prevent this, first check the serial No. of each scanning
head, then check the serial No. of the HEAD1 and HEAD2 on the back of
the controller for correct connection. Each of the scanning heads must be
connected respectively to the HEAD1 or HEAD2 connector on the back of
the controller.
Connect the correct scanning
head to the HEAD1 or HEAD2
connector.
Serial No.
Serial No.
Next, follow the steps (1) through (4) on P4. and 5.
6
CHAPTER 2
TRIAL OPERATION
1. Connect the controller and scanning head(s). Make sure that the optical
axis of the transmitter and receiver are aligned before turning the power
ON. Aligning the optical axis is not necessary, however, when using the
single-body type scanning head or when the transmitter and receiver are
secured on the detachable frame.
2. Laser emission starts in approximately 5 seconds.
3. Make sure that the target is correctly positioned in the measurement area.
The measured value will be displayed shortly.
4. The LS controller has been adjusted to the following factory-settings:
Therefore, the displayed value was measured in the above settings. Read
this manual carefully to use the desired settings on your LS controller. To
change each setting, see the table of contents for the corresponding page
No. Some of the most frequently-used operations are given below with the
reference page Nos.
Ref.
• Changing the measurement position of targets .................................. P.32
• Changing the number of measurements for averaging ....................... P.21
• Changing the measurement mode (NORMAL, P-P, PEAK, BOTTOM)P.40
• Measuring two values at a time .......................................................... P.32
• Using two scanning heads to measure (X+Y)/2 .................................. P.32
Do not connect or disconnect any cables while the power is ON. Otherwise,
the laser diode may deteriorate and/or may be damaged.
7
CHAPTER 3 Quick Setup Procedures
CHAPTER 3
QUICK SETUP PROCEDURES
The LS Series is a multi-function, versatile laser scan micrometer. This
chapter describes the procedures for setting the LS-3100(W)/3101(W) by
illustrating typical applications.
This chapter covers the procedures needed to make measurements after the
memory is initialized (after "MEMORY INITIALIZED" is displayed).
Initializing Current Settings ........................................................................... 8
How to Use This Chapter ..............................................................................9
Common Settings .........................................................................................9
Setting the Number of Measurements for Averaging ............................... 9
Measuring Target Displacement within the Measuring Range ..............13
Measuring Edge Movement ...................................................................13
Using LS-3100D(W) Controller with 2 Scanning Heads .............................14
Once a setup is completed, do not initialize the setting again unless
you need to change them. (When initialized, all the current settings
are cleared.)
Turn the power switch "ON" while pressing ENT.
LS-3100 SeriesLS-3100(W) Series
ON
ENT
"MEMORY INITIALIZED" is displayed on the sub-display.
MIENMIOTRIY
All prior settings are cleared and memory initialization is completed.
AL I ZED
ENT
8
How to Use This Chapter
CHAPTER 3 Quick Setup Procedures
3-1. Common Settings
Setting the Number
of Measurements for
Averaging
The switches on the left
side of the front panel
are shown on the left
side of the pages in this
chapter.
The switches on the
right side of the front
panel are shown at the
center of the pages in
this chapter.
See page 93 for Measurement Procedures.
AVE
SET
SET
SET
.
.
.
.
AVE1M0O2V4
AVE
AVE2
AVE4
E
S I MPL1E
All massages on the
sub-display are shown
on the right side of the
pages in this chapter.
512
256
128
64
32
16
8
• After initialization, the number of measurements for averaging is automatically set to 1024. "----" appears on the main display for about 3 seconds,
even after a workpiece has entered the measurement range.
• The sampling time required for one measurement is 2.5 m sec. The
sampling time then required for 1024 measurements is 2.56 sec. (2.5 x
1024). The time required to display an averaged measurement changes
depending on the number of measurements being made.
If highly accurate measurement is needed, select 512 or higher as the
number of measurements for averaging.
9
CHAPTER 3 Quick Setup Procedures
Tolerance Settings
The following procedure shows how to set the upper limit to 10.5 mm, and
lower limits to 9.5 mm.
LIMIT
OK
1
0
.
5
ENT
9
.
5
ENT
HLIO=
=–6600..00000000
HLIO==[
HLIO==[
HLIO=
HLIO=
HLIO=
HLIO=
HLIO=
HLIO=
HLIO=
HLIO=
–60 . 00010
–60 . 001000
[
=
–60 . 01000.0
[
=
–60 .1000.050
=
–1600..50000000
10 . 5000
=[
10 . 5000
=[
10 . 50090
=[
10 . 5090.0
=[
109..5500000
=
]
]
]
]
]
]
9
]
.
]
5
0
▲
" " moves
to the "LO"
position
For details on tolerance setting for 7-step differentiation, refer to pages 23
through 26.
3-2. Measurement Procedures
Outer Diameter
Measurements
■ Applications
• Measuring piston outer diameter
• Measuring flat cable width
•Measuring capacitor outer diameter
■ Setting
After "MEMORY INITIALIZED" is displayed, the segment mode is automatically set to DIA (outer diameter measurement).
For details on AVERAGE setting, refer to "AVERAGE", and for details on
tolerance setting, refer to "Tolerance Settings" shown above.
RT
10
• The number of measurements for averaging is set to 1024 after settings
are initialized.
"----" appears on the display for about 3 seconds, even after a target is
placed within the measuring range.
• In the DIA mode, "----" appears on the main display when no workpiece is
within the measuring range.
CHAPTER 3 Quick Setup Procedures
Gap Measurement 1
(When part of the target is
outside the optical axis)
Gap Measurement 2
(When all of the target is
within the optical axis)
■ Applications
• Measuring gap between rollers in a copy machine
• Measuring gap between magnet roller and doctor blade
TR
■ Setting
1. Initialize settings ("MEMORY INITIALIZED" is displayed.)
2.
SEG
SET
OK
SEG D I A
S EG E DGE1
■ Applications
• Measuring pitch of capacitor leads
• Measuring pitch of IC leads
1
TR
2
3
4
5
6
■ Setting
1. Initialize settings ("MEMORY INITIALIZED" is displayed.)
2.
SEG
SET
SET
SET
3
.
4
SEG D I A
S EG E DGE1
S EG E DGE2
SEG [2 ,3]
SEG [3 ,3 ]
SEG [3 ,3 ]
SEG [3 ,4 ]
OK
ENT
SEG [3 ,4 ]
11
CHAPTER 3 Quick Setup Procedures
Measuring
Roundness
(When target is rotating)
■ Applications
• Detecting uneven rubber roller in a copying machine
• Detecting uneven rubber roller on a printer
RT
■ Setting
1. Initialize settings ("MEMORY INITIALIZED" is displayed.)
2.
MODE
SET
OK
MODE NO MA L
MOD E P–P
R
When the measurement mode is set to "P-P", the displayed value is
retained until HOLD is pressed or until an external signal is input through
the hold synchronous input terminal.
Measuring
Eccentricity
(When target is rotating)
3. Resetting displayed values (When 1. 2 3 4 5 appears on the main
display)
HOLD
HOLD
HOLD
The displayed
Main display
value is reset
to "0".
■ Applications
• Measuring eccentricity of rubber rollers
• Measuring shaft eccentricity for automobile
RT
■ Setting
1. Initialize settings ("MEMORY INITIALIZED" is displayed.)
12
2.
SEG
SET
MODE
SET
OK
SEG D I A
S EG E DGE 1
MODE NOMA L
MODE P–P
CHAPTER 3 Quick Setup Procedures
When the measurement mode is set to "P-P", the displayed value is
retained until HOLD is pressed or until an external signal is input through
the hold synchronous input terminal.
3. Resetting displayed values (When 1. 2 3 4 5 appears on the main
display)
HOLD
Measuring Target
Displacement within
the Measuring Range
Main display
HOLD
The displayed
value is reset
to "0".
■ Applications
•Measuring disk head movement
• Measuring camera lens movement
TR
–
+
■ Setting
1. Initialize settings ("MEMORY INITIALIZED" is displayed.)
Follow the same procedure as that for "Gap Measurement 1".
To reset the current value to "0", press ZERO.
ZERO
HOLD
Measuring Edge
Movement
To reset the displayed value to "0" again, press ZERO twice.
■ Applications
•Measuring paper feed for copying machines
• Controlling sheet edges
• Measuring dot printer wire movement
TR
+
–
■ Setting
1. Initialize settings ("MEMORY INITIALIZED" is displayed.)
Follow the same procedure as that for "Gap Measurement 1".
To reset the current value to "0", press ZERO.
ZERO
To reset the displayed value to "0" again, press ZERO twice.
13
CHAPTER 3 Quick Setup Procedures
3.3 Using LS-3100D(W) Controller with 2 Scanning Heads
Measuring Outer Diameters in the X and Y Axes
The LS-3100D(W) uses 2 scanning heads for this measurement.
■ Applications
• Measuring outer diameter of coated wire
in the X and Y axes
• Measuring outer diameter of extrusion
molded parts
■ Setting
1. Initialize settings
2.
FNC
SET
SET
SET
DISP
DISP
XY==DDIIA
XY==DDIIA
XY==DEIDA
XY==DEIDA
XY==D2I:A
DNIOA
CNHO2R--
A
A
GE1
GE2
D IA
RMA L
AVE1 024
----------
14
OK
In the above setup, measurements from scanning head 1 are displayed on
the main display, and measurements from scanning head 2 are displayed on
the sub-display.
Measurement value in Y-axisMeasurement value in X-axis
For details on tolerance setting, refer to p.23-26, 39, 40, 43 in this manual.
This appendix does not cover all of the LS-3100(W) functions. In addition to
the applications covered in this appendix, the LS-3100(W) can be used for
several other types of measurements.
For more details on functions and measurements, read chapter 4. (A cross
reference between this appendix and chapter 4 is given below.)
Setting number of measurements for averaging .......... p.21
Measuring outer diameter
Gap measurement 1
Gap measurement 2
Measuring roundness
Measuring eccentricityp.30-32, 40-41, 89-96
Measuring target displacement
within the measuring range
Measuring edge movement
Measuring outer diameter in the X and Y axes
CHAPTER 4
FUNCTIONSAND CONTROLS
4-1. Part Names
Controller
Model: LS-3100(W)/3100D(W)
Front Panel
LS-3100 Series
LS-3100(W) Series
12 3456
LS
LASER SCAN DIAMETER
• OFF
• ON
POWER
7
LS
LASER SCAN DIAMETER
•
•
LASER
LIMT
HIGH GO LOW
1 2
LASER LASER
LIMT
HIGH GO LOW
1 2
OFFSET ZERO HOLD LOCK POSITION
LASER LASER
OFFSET ZERO HOLD LOCK POSITION
SEGSETMODE
AVE LIMIT OFS
PRM FNC CAL
PROG DISP
SEGSETMODE
AVE LIMIT OFS
PRM FNC CAL
PROG DISP
mm
inch
7
8
9
ZERO
4
5
6
HOLD
1
2
3
CLR
0
.
+/-
ENT
I
H GF E D C BA0 98
mm
inch
J
7
8
9
ZERO
4
5
6
HOLD
1
2
3
CLR
0
.
+/-
ENT
K
L
M
N
7
1 Laser emission indicator LED
2 Comparator output LEDs
3 Current setting indicator
4 Main display (CH1)
5 Target position indicator
6 Subdisplay (CH2)
7 Key-operated power switch
(LS-3100(W) Series: Key-operated
laser switch)
8 SEGMENT key
9 AVERAGE key
(for number of measurements)
0 PARAMETER key
A PROGRAM key
B MODE key
C LIMIT key (for 3-level
comparator output only)
D FUNCTION key
E SUBDISPLAY key
F UP/DOWN key
G CALIBRATION key
H OFFSET key
I SET key
J Numeric keys
K AUTO ZERO key
L HOLD key
M CLEAR key
N ENTER key
15
CHAPTER 4 Functions and Controls
ST
WXYZ[\
ST
Back Panel
Model: LS-3100
Model: LS-3100(W)
PQ
O
RS - 232C
HEAD 1HEAD 2
TIMMON
OV GND P1 P2 P 3 P4 P5 GND
1 23 4 5 6 7 8 9 10
UV
PQ
O
RS - 232C
HEAD 1HEAD 2
TIMMON
OV GND P1 P2 P 3 P4 P5 GND
1 23 4 5 6 7 8 9 10
PROGRAM SELECT
PROGRAM SELECT
R
CONTROL I/O
LASAER
REMOTE
13 14 15 16 17 18 1911 12
R
LASAER
REMOTE
13 14 15 16 17 18 1911 12
LOWGOHIGH
CONTROL I/O
LOWGOHIGH
85~264V AC
2A
85~264V ACPOWER
O
2A
I
Scanning Heads
Model: LS-3060
UVWXYZ[
O SCANNING HEAD connector port
P SCANNING HEAD connector port
(for second scanning head)
Q RS-232C connector port
R Expansion I/O port (for optional
BCD or GP-IB interface board)
S 36-pin connector port
(Control I/O port)
T Cooling fan
U Analog voltage output terminals
1. Not used for the LS-3100(W) series.
18
THIS PRODUCT..... ...
CONFORMS
LS-3060T
LASAER SCAN DIAMETER
]\
V Hold synchronous input terminal
W Program selector input terminals
X Laser emission control input
terminals
Y Comparator output terminals
Z Earth ground terminal
[ Power cable receptacle
\ Fuse holder (2A)
] POWER switch
To prevent lens from being damage or dust from accumulating on the lens
surface (Spares are available for replacement.)
0 Connector cable between the controller and the scanning head (3 m)
A Transmitter (T)
B Receiver (R)
C 6-pin connector cable between the controller and the scanning head
D 8-pin connector cable between (T) and (R)
E Laser emission indicator LED
F Laser aperture
17
CHAPTER 4 Functions and Controls
4-2. Indicators and Displays
1
2
LIMIT
HIGH GO LOW
11
1 Laser Emission
11
Indicator LEDs
22
2 Comparator Output
22
LEDs
LS
LASER SCAN DIAMETER
Light when laser is being or about to be emitted from the scanning head.
Only LED [1] lights when one scanning head is used. Both LED [1] and LED
[2] light when two scanning heads are used.
Three LEDs, HIGH, LOW, and GO, are provided. HIGH and LOW light
respectively when the measured value is above the preset upper limit and
below the preset lower limit. GO lights when the measured value is within the
range between these two limits.
3456
OFFSET ZERO HOLD LOCKPOSITION
LASER LASER 1 2
mm
inch
33
3 Current Setting
33
Indicator
44
4 Main Display
44
(Channel 1)
55
5 Target Position
55
Indicators
66
6 Subdisplay
66
(Channel 2)
Lights when one of OFFSET, AUTO-ZERO, HOLD, or PANEL LOCK has
been selected.
Shows a measured value.
Indicate where the target is located in the measurement area.
Shows preset values when only one scanning head is used. It shows a
measured value when two scanning heads are used.
A measured value is displayed in 8 digits and updated 10 times per second.
18
4-3. Panel Keys
CHAPTER 4 Functions and Controls
11
1 Numeric Keys
11
A
SEGSETMODE
8
AVELIMITOFS
9
PRMFNCCAL
G
PROGDISP
B
C2F
D50E
7
4
1
0
1
8
5
2
.
9
6
3
+/-
6
ZERO
HOLD
CLR
ENT
4
7
3
Enter numbers, a decimal point, and + or - sign. See the following examples:
• Entering +10.0000
10
.
0000
May be defaulted
7
4
1
0
8
5
2
.
9
6
3
+/-
• Entering -10.0000
22
2 SET Key
22
+ / -
10 . 0000
May be defaulted
"+" and "-" alternate each time +/- is pressed. Note that the "+" sign is not
displayed. This key can be pressed to select a positive or negative number
as long as you have only up to 3-digit integers.
• Entering 0.1234
0.1234
May be defaulted
• When you first enter any number by using numeric keys, [ ] appears
on both sides of this number. This indicates that the key input can be
accepted. Any incorrect entry can be deleted using CLR only in this mode.
Changes parameter, setting, and ON/OFF status.
33
3 CLR Key
33
Deletes the desired number. Press ENT to regain the initial data.
19
CHAPTER 4 Functions and Controls
44
4 ENT Key
44
55
5 UP/DOWN Key
55
66
6 ZERO (Auto-Zero)
66
Key
Registers the desired number.
Turn ON the power while pressing and holding ENT to return to the factory
settings.
➪
See P.89 to P.96 for the factory settings.
When making certain settings, two of the following items, value, mode or ON/
OFF status, appear at the same time on the subdisplay (CH2). Press this key
to move the cursor up and down to change one of the preset value or status.
Resets the displayed value to "0".
1. Press ZERO to reset the displayed and output values to 0.0000 mm or
0.0000 inch. The "ZERO" indicator below the main display lights when this
key is pressed. From this moment, the main display will only show the
changes in measured value based on the value that has been reset.
mm
inch
OFFSET ZERO HOLD LOCK POSITION
2. Press ZERO again to cancel this mode. The "ZERO" indicator goes out
when this key is pressed.
77
7 HOLD Key
77
• This function is available only for the measured value appearing on the
main display (CH1). Use FNC to reset the measured value on the
subdisplay (CH2) to 0.
➪
See P.39
• This key is invalid when the measurement mode is set to "P-P". At this
time, press HOLD to reset the measured value.
➪
See P.20
• It is also possible to remotely reset the measured value or cancel this
mode by sending a signal through one of the control I/O connector pins.
➪
See P.43
Holds the displayed and output values.
1. Press HOLD to hold the displayed and output values just before this key is
pressed. The "HOLD" indicator below the main display lights when this
key is pressed.
mm
inch
OFFSET ZERO HOLD LOCK POSITION
20
2. Press HOLD again to cancel this mode. The "HOLD" indicator goes out
when this key is pressed.
• If the measurement mode is set to "P-P", "PEAK", or "BOTTOM", the
previously measured value is reset when this mode is cancelled.
• It is possible to use this mode only on the main (CH1) display or on both
the main and subdisplay (CH2). To do this, choose the desired mode
using PRM.
CHAPTER 4 Functions and Controls
88
8 AVE Key
88
(Selector key for number
of measurements for
averaging and averaging
method)
Sets the desired number of measurements and averaging method.
1. Use AVE to display the current number and method.
2. To change the number of measurements, use UP/DOWN to move the
cursor up. Then, use SET to specify the number. Each time you press
SET, the number will increment in the following order: "1, 2, 4, 8 ••••• 512,
and 1024". Note also that the number of measurements is set automatically to the currently displayed number. Therefore, you do not need to
press any other key to finalize this number.
AVE
1024
MOV E
3. To change the measurement method, use UP/DOWN to move the cursor
down. Then, use SET to change the averaging method. Each time you
press SET, SIMPLE and MOVE display alternately.
AVE
1024
SIMPLE
See the timing diagrams below for the difference between these two
methods.
* MOVE is valid only when the number of measurements is set to a number
between 16 and 1024.
21
CHAPTER 4 Functions and Controls
99
9 PRM Selector
99
Key
Analog Voltage Range
Displays sequentially the following 9 items: preset analog range, "LIMIT",
"DIGIT SUPPRESS", "HOLD", "BAUD RATE", "UNIT", "BUZZER", "HEAD2",
and "POSITION CHECK". Display the desired item first and make the
necessary changes. See the instructions below for making settings in each
item.
In this mode, measured value is converted into an analog voltage between
-6 V and +6 V. Match the desired measurement range and the above voltage
range. By matching these ranges, the measured value can be converted into
its corresponding voltage.
Determine the upper and lower limits (m) and (n) as shown below. These
values correspond respectively to +6 V and -6 V. This automatically determines the conversion rate of the measurement range against the voltage
range.
Voltage
+6V
n
0
m
Measured value
-6V
1. Use PRM to display the analog voltage range as shown below. UP/DOWN
is not used in this mode.
+6
=
- =-
V6
60. 0V000
0. 0000
6
2. Enter the desired value in mm or inch that corresponds to +6 V.
3. Press ENT to register.
4. When ENT is pressed, the cursor moves down to the second line and the
current preset value corresponding to -6 V disappears.
+6
=
-=
V6
20. 0V000
5. Enter the desired value in mm or inch that corresponds to -6 V.
6. Press ENT to register and finish.
+6
=
-=
V6
20. 0V000
0. 0000
22
7. Use CLR to delete incorrect data and enter the correct one if required.
CHAPTER 4 Functions and Controls
[]
[]
[]
[]
[]
[]
[]
Example of OperationYou are to measure the diameter of round bars. The reference diameter is 10
mm, and you wish to convert a difference of 1 mm into 1 V.
In this case, first set 100.000 mm to 0 V. Then, to represent each 1 mm
change in diameter from the reference by 1 V, set the reference plus 6 mm to
+6 V and the reference minus 6 mm to -6 V. That is, +6 V = 16 mm and -6 V
= 4 mm.
StepKey operationDisplay
1Use PRM to display analog
voltage range.
2Press "1".
3Press "6".
4Press ".".
5Press "0" four times.
6Press ENT
7Press "4".
8Press ".".
+6
=
- =-
V6
+6
V
=
-=
+6V
-=
+6V
-=
+6V
-=
+6V
-=
+6V
-=
+6V
-=
-
V6
=
-
V6
=
-
V6
=
-
V6
=
V6
=
V6
=
V6
60. 0V000
0. 0000
6
6. 0000
0
6. 0000
0
6. 0000
0
1. 00006
6. 0000
0
1. 00006
1. 00006
1. 00006
1
1.
6
4
4
1
6
.
LIMIT
Note
Note
9Press "0" four times.
10Press ENT.
V
=
+6
-=
V6
+6V
=
-=
V6
1. 00006
4 . 0000
1. 00006
4 . 0000
• The LS controller is factory-set to +6 V = 60 mm and -6 V = - 60 mm.
•A range of 0.024 mm to 240 mm can be specified.
• Make sure that the value corresponding +6 V is greater than that for -6 V.
• When your reference value is an integer with no decimals, steps 4, 5, 8
and 9 can be omitted.
You can set the following 3 parameters in LIMIT:
• 3-step or 7-step differentiation
• Tolerance limits when 7-step differentiation is selected
• Comparator output type; Relay or open-collector (if the measured value is
above the upper limit or below the lower limit).
• Use FNC to set the tolerance limits for the measured value appearing on
the subdisplay (CH2). ➪
See P.32
23
CHAPTER 4 Functions and Controls
Tolerance Limits and
Comparator Output
■ 3-step differentiation
Choose the 3-step differentiation first, then set the tolerance limits using
LIMIT. When "ON" appears on the display, the comparator output is sent
through each of the relays and open-collectors (HIGH, GO, or LOW).
➪
See P.63, 64
Tolerance limitConditionOutput
(Relay)(Open-collector)
HI < XHIGHHI1 (CH1)
HI
HI ≥ X ≥ LOGOGO (CH1)
LO
X < LOLOWLO1 (CH1)
Note: X represents a measured value.
■ 7-step differentiation
Choose the 7-step differentiation first, then set the tolerance limits using
PRM. When "ON" appears on the display, the comparator output is sent
through each of the relays and open-collectors (HI3, HI2, HI1, GO, LO1,
LO2, or LO3).
➪
See P.63
Tolerance limitConditionOutput
(Relay)(Open-collector)
HI3 < XHIGHHI3 (CH1)
HI3
HI3 ≥ X > HI2HIGHHI2 (CH1)
HI2
HI2 ≥ X > HI1HIGHHI1 (CH1)
HI1
HI1 ≥ X ≥ LO1GOGO (CH1)
LO1
LO1 > X ≥ LO2LOWLO1 (CH1)
LO2
LO2 > X ≥ LO3LOWLO2 (CH1)
LO3
LO3 > XLOWLO3 (CH1)
Note: X represents a measured value.
• It is possible to set the tolerance limits for both 3-step and 7-step differentiation. However, only the tolerance limits of the type of differentiation you
have selected are valid. Therefore, the comparator output is sent according to these limits.
24
• In 7-step differentiation, measured values can be correctly checked only
when the values assigned to the tolerance limits meet the following
condition: HI3>HI2>HI1>LO1>LO2>LO3
CHAPTER 4 Functions and Controls
Procedure■ Switching between 3-step and 7-step differentiations
1. Use PRM to display "LIMIT" (on the subdisplay).
2. To switch between two types of differentiation, use UP/DOWN to move the
cursor up.
I MILTON7CL ASSES
3. Each time you press SET, "7 CLASSES" and "3 CLASSES" are displayed
alternately. Choose the desired type of differentiation.
■ Turning comparator output function ON/OFF
1. Use PRM to display "LIMIT" (on the subdisplay).
2. Use UP/DOWN to move the cursor down.
I MILTON7CL ASSES
3. Each time you press SET, "ON" and "OFF" are displayed alternately.
4. Display "ON" to obtain the comparator output function, or "OFF" to cancel
this output function.
1. First select 7-step differentiation as mentioned above, then use PRM to
set the tolerance limit.
2. "HI1" and "LO1" will appear on the subdisplay. Use UP/DOWN to move
the cursor up for HI1 or down for LO1.
I MILTON7CL ASSES
3. Enter the desired values and press ENT to register.
4. Press PRM.
5. "HI2" and "LO2" will appear on the subdisplay. Enter the desired values
and press ENT to register.
H2
I
LO
=2-
.
30 000.
0
0
30=000
6. Press PRM to display "HI3" and "LO3", enter the desired values, and
press ENT to register.
25
CHAPTER 4 Functions and Controls
Example of Operation
You are to measure the diameter of round bars. The reference diameter is
10 mm. You wish to perform a 7-step differentiation with the following tolerance limits: ±0.1 mm, ±0.2 mm, ±0.3 mm.
StepKey operationDisplay
1Press PRM to display
"LIMIT".
I MILTON3CL ASSES
2Use UP/DOWN to move
cursor up.
3Use SET to display "7
CLASSES".
I MILTON3CL ASSES
I MILTON7CL ASSES
4Press PRM to display "HI1"
and "LO1".
H1
I
LO
-
=1-
10 000.
.
0
0
10=000
5Enter 10.1000 for HI1 and
9.9000 for LO1, then press
ENT.
I
H1
LO
=
=1
10
00
1
..
9
.
0
009
0
6Press PRM again to display
30=000
"HI2" and "LO2".
I
H2
LO
=2-
.
30 000.
0
0
DIGIT SUPPRESS
7Enter 10.2000 for HI2 and
9.8000 for LO2, then press
ENT.
H2
I
LO
=
=2
10
00
2
..
9
.
0
008
0
8Press PRM again to display
"HI3" and "LO3".
I
H3
LO
=3-
.
60 000.
0
0
60=000
9Enter 10.3000 for HI3 and
9.7000 for LO3, then press
ENT.
H3
I
LO
=
=3
10
..
9.007
00
3
0
0
Deletes fractions from measured value on the display. DIGIT SUPPRESS is
useful when there is a relatively large difference from one measured value to
another. The possibly applicable cases are listed below.
• When the number of measurements for averaging is set to a small number
• When the target is vibrating
•When it is not necessary to obtain precise measurements
This prevents errors involved in reading measured values.
26
• This function can be used for both the main display and subdisplay. It
becomes invalid when the subdisplay is used to show measured values in
P-P, PEAK, or BOTTOM mode.
• Although fractions are not displayed, the target is measured down to
0.0001 mm and comparator output sent based on these measured values.
CHAPTER 4 Functions and Controls
Procedure
HOLD
Procedure
1. Use PRM to display "DIGIT SUPPRESS" (on the subdisplay).
IG
I T
UDS0PPRESSCH20
CH1
2. Use UP/DOWN to move the cursor up or down, then use SET to choose
how many fractions you wish to delete. Each time you press SET, the
subdisplay will show "0", "1", "2", "3", "4", and "0" again.
3. Note that they represent the number of digits to be deleted starting from
the rightmost (smallest) fraction.
Selects whether to use HOLD function (see P.20) only for CH1 or for both
CH1 and CH2.
1. Use PRM to display "HOLD" (on the subdisplay).
C1
OLD
H
H
2. Each time you press SET, "CH1" and "CH2" are displayed alternately.
Display "CH1" only for CH1 or "CH2" for both CH1 and CH2.
BAUD RATE
Procedure
UNIT
Procedure
Selects the data transmission speed via RS-232C interface.
1. Use PRM to display "BAUD RATE" (on the subdisplay).
T1
B
AUD
EA R2 00
2. Each time you press SET, the baud rate will increment from 75 to 19200.
Display the desired choice.
Selects the unit of measurement between mm and inch.
1. Use PRM to display "UNIT" (on the subdisplay).
NIT
mUm
2. Use SET to display "mm" or "inch".
The unit selected in this mode will be applied to other values such as
tolerance limits and calibration values.
27
CHAPTER 4 Functions and Controls
BUZZER
Turns ON/OFF the beep during panel key operation.
Procedure1. Use PRM to display "BUZZER" (on the subdisplay).
UZZNBOER
2. Use SET to display "ON" or "OFF".
HEAD2
Procedure
Selects whether to use two scanning heads or not.
1. Each time you press SET, "ON" and "OFF" are displayed alternately.
2. Display "ON" to use two scanning heads or "OFF" to use just one head.
EADHN2O
The controller must be equipped with the optional PC board for the second
scanning head. If not, an error message "H2-Err" appears on the main
display. In such a case, display "HEAD2 OFF" (on the subdisplay). This will
cause "H-Err" to appear on the main display. Then, turn OFF the power
once, then turn it ON again. You can now perform measurements using only
the first scanning head (HEAD1).
POSITION CHECK
(Model: LS-3033/3060 only)
Procedure
Checks whether the transmitter and receiver are aligned in terms of the
optical axis of the laser. This alignment is required when these units are
removed from the detachable frame during operation.
➪
See "Mounting Scanning Head" in Chapter 5 on P.65 for details.
1. Use PRM to display "POSITION CHECK".
EADH1
OSTPIOINCHECK
2. Align the transmitter and receiver in terms of the optical axis.
3. When two scanning heads are used, press PRM again to display
"HEAD2" above "POSITION CHECK", and align the second set of
transmitter and receiver.
H2
EAD
OSTPIOINCHECK
28
CHAPTER 4 Functions and Controls
[]
00
0 LIMIT
00
(Tolerance limits
for 3-step
differentiation)
Key
This key is used to set the tolerance limits (upper and lower) for measured
values on the main display (CH1). If the measured value is above the upper
or below the lower limit, the comparator output is sent through the corresponding relay and open-collector.
➪
See P.24 for the relationship between the tolerance limits and types of output obtained.
• Use FNC to set the tolerance limits for the subdisplay (CH2).
➪
See P.39
Procedure1. Use LIMIT to display the current tolerance limits.
2. HI and LO represent respectively the upper and lower limits.
3. Use UP/DOWN to move the cursor up or down.
HI
LO=-60 . 0000
60. 0=000
4. Enter the desired tolerance limits.
Make sure that the value for HI is greater than that for LO.
5. Press ENT to register.
Example of Operation
6. Use UP/DOWN again to move the cursor, and enter another tolerance
limit.
7. Use CLR to delete any incorrect entry.
You are to measure the diameter of round bars. The reference diameter is
10 mm. You wish to perform a 3-step differentiation with the tolerance limits
of ±0.1 mm (upper limit=10.1 mm, lower limit=9.9 mm).
StepKey operationDisplay
1Press LIMIT to display
current tolerance limits.
HI
LO=-60 . 0000
60. 0=000
2Enter tolerance limit.
10. 1=000
[
HI
LO=-60 . 0000
]
3Press ENT to register.
10. 1=000
10. 1=000
4Press UP/DOWN to move
cursor.
HI
LO=-60 . 0000
HI
LO=-60 . 0000
5Enter another tolerance
limit.
6Press ENT to register.
HI
LO=-9 . 9000
HI
LO=9. 9000
10. 1=000
10. 1=000
29
CHAPTER 4 Functions and Controls
AA
A SEG (Segment)
AA
Key
Definition of Segment
Note
Targets placed within the measurement area interrupt the passage of the
laser beam, thus creating a light-dark pattern. The border between the light
and dark regions is referred to an "edge". Also, each dark or light region
separated by the edge is called a "segment".
TransmitterReceiver
Edge1
Edge2
Edge3
by laser
Direction of scanning
Targets
Edge4
Edge5 (X-1)
Edge6 (X)
The desired pair of edges is to be selected from those shown above, and the
distance between these two edges measured.
• The edges are numbered in the order that they are scanned by the laser
(from top to bottom).
• The desired pair from among up to 126 edges can be selected.
Description
4 types of segment pairs, DIA, EDGE1, EDGE2, (m,n), can be measured.
The description of measurement for each pair is given below.
DIA: This mode is applicable to measurement of targets’ outer diameters. In
the above figure, the distance between edge 2 and edge (X-1) is measured.
When there is only one target, its outer diameter is measured. When there
are two or more targets, the distance between the upper edge of the upper
most target to the lower edge of the lower most target is measured.
EDGE1: The width of the uppermost region where the laser is not interrupted by the target is measured. This mode is applicable to measurement of
changes in targets’ diameters. In the above figure, the distance between
edge 1 and edge 2 is measured.
EDGE2: The width of the lowermost region where the laser is not interrupted
by the target is measured. In the above figure, the distance between edge
(X-1) and edge X is measured.
(m,n): The distance between two desired edges is measured. When measuring two or more targets, this mode is applicable to measurement of the
outer diameter of each target or to measurement of the gap between two
targets.
In (m,n) mode, both "m" and "n" represent the edge numbers ranging from 1
to 126. Make sure that the number specified for "n" is greater than that for
"m".
30
CHAPTER 4 Functions and Controls
Procedure
1. Press SEG to display the current segment pair (on the subdisplay).
SEGD I A
2. Each time you press SET, the subdisplay will sequentially show "DIA",
"EDGE1", "EDGE2", "(m,n)", then "DIA" again. Display the desired segment pair. You do not need to press any other key. (When two scanning
heads are used, the subdisplay will sequentially show "DIA", "EDGE1",
"EDGE2", "(m,n)", "2:DIA", "2:EDGE1", "2:EDGE2", "2:(m,n)", and "DIA"
again.)
3. To measure the distance between two desired edges, select the (m,n)
mode.
S E G(5,2 )
4. Use SET to display "(m,n)" first.
SEG(,
)
nm
5. Enter the desired edge numbers.
➪
See the example of operation given below for details.
6. Press ENT to register.
Example of OperationYou are to measure the gap between two round bars (between edges 3 and
4). Suppose that the currently registered edges are 2 and 3.
StepKey operationDisplay
1Press SEG.
SEGD I A
2Use SET to display current
edges (2 and 3).
SEG( 2, 3)
3Press "3".
SEG( 3, 3)
4Press ".".
SEG( 3, 3)
5Press "4".
SEG( 3, 4)
6Press ENT.
SEG( 3, 4)
31
CHAPTER 4 Functions and Controls
Note
• If you enter invalid numbers (other than 1 through 126) and press ENT,
the initial edge numbers will remain on the subdisplay.
• Basically, two numbers are entered in this mode. If by mistake you enter
the larger number on the left and smaller number on the right, and press
ENT, the smaller number will appear on the left and larger number on the
right.
S E G(4,3 )
ENT
S E G(3,4 )
• Clear objects can be used as targets only in DIA, EDGE1, and EDGE2
modes, but not in SEG (m,n) mode.
• Use CLR to delete any incorrect values.
• In the above example operation, if you need to change only the first edge
number, that is, from 2 to 1, follow only steps 1, 2, 3, and 6.
BB
B FNC (Function)
BB
Key
Segment Selector
• Similarly, if you need to change only the second edge number, that is,
from 3 to 5, follow only steps 1, 2, 4, 5, and 6.
Is used for the following settings:
• Segment selector
Selection of measurement points (2-point simultaneous measurement)
• Segment check
Inhibiting output of measurements obtained under abnormal conditions
• Output channel
Selection of output channel for displaying measured values X and Y and
calculation of equations using these values
• MODE, OFFSET, and LIMIT for CH2
Selection of mode for CH2, offset value, and tolerance limit
Description of each setting is given below.
In this mode, two segments are to be specified for simultaneous 2-point
measurement. Just as with the SEG key, any of four types of segments can
be specified respectively for segment selectors X and Y. (See P.30.) By
selecting two different segments, the LS controller can measure two different
segments at the same time.
32
CHAPTER 4 Functions and Controls
Procedure 1:
Measuring 2 Segments with 1
Scanning Head
1. Use FNC to display the current segments for X and Y.
XD
=
Y=
IIADA
2. Use UP/DOWN to move the cursor up.
3. Each time you press SET, "DIA", "EDGE1", "EDGE2", "(m,n)", and "DIA"
will sequentially appear on the right of "X=". Display the desired segment.
* "m" and "n" in (m,n) represent two desired edge numbers.
X
=I
DA
DG
EI
EY=
4. When you have selected "(m,n)", enter the desired edge numbers, and
press ENT to register.
➪
See P.31.
Once you select "(m,n)" for X, "(m,n)" will be automatically selected for Y.
X
Y
(
=2, 3
(
=
1, 2
)
)
5. Use UP/DOWN to move the cursor down.
X
Y
(
=2, 3
(
=
1, 2
)
)
Example of Operation
6. Repeat steps 3 and 4.
You are to measure the outer diameter and eccentricity of the round bars
using one scanning head.
Edge 1
Y
Edge 2
X
Edge 3
Edge 4
Transmitter
Receiver
StepKey operationDisplay
1Use FNC to display current
XD
Y=
=
IIADA
segments.
2Use UP/DOWN to move
XD
cursor up.
Y=
=
IIADA
3Use SET to display:
(1) "X= DIA"
(2) "X= (m,n)". Enter 2 for
"m" and 3 for "n", and press
ENT.
XD
=
Y=
X
=2, 3
Y
=
(
(
IIADA
1, 2
(To be continued)
)
)
33
CHAPTER 4 Functions and Controls
Procedure 2:
Using 2 Scanning Heads
4Use UP/DOWN to move
cursor down.
XD
=
Y=
X
=2, 3
Y
=
(
(
IIADA
1, 2
5When you have selected
X
=I
"DIA" in step 3, use SET to
display
DA
DG
EY=
EI
"Y= EDGE1".
When you have selected
"(m,n)" in step 3, enter 1 for
X
Y
(
=
1, 2
(
=2, 3
"m" and 2 for "n", and press
ENT.
Press PRM first, and check that "HEADS2 ON" is displayed on the
subdisplay.
1. Use FNC to display current segments.
2. Use UP/DOWN to move cursor up.
)
)
)
)
Example of Operation
3. Each time you press SET, the subdisplay will sequentially show "DIA",
"EDGE1", "EDGE2", "(m,n)", "2:DIA", "2:EDGE1", "2:EDGE2", "2:(m,n)",
and "DIA" again. Display the desired segment.
* "m" and "n" in (m,n) represent two desired edge numbers.
* "2:" in "2:DIA", for example, represents the second scanning head.
4. When you have selected "(m,n)", enter the desired numbers, and press
ENT to register.
➪
See P.31.
X
Y
(
=2, 3
(
=
1, 2
)
)
5. Use UP/DOWN to move the cursor down.
X
Y
(
=
1, 2
(
=2, 3
)
)
6. Repeat steps 3 and 4.
Measuring the outer diameter of 2 round bars at the same time using 2
scanning heads
StepKey operationDisplay
1Use FNC to display current
XD
segments.
=
Y=
IIADA
34
2Use UP/DOWN to move
cursor up.
3Use SET to display "X=
DIA"
X
Y=
X
Y=
D
=
IIADA
D
=
IIADA
(To be continued)
4Use UP/DOWN to move
cursor down.
CHAPTER 4 Functions and Controls
XD
=
Y=
IIADA
5Use SET to display
Note
• Some combinations of segments cannot be used with the LS controller.
See the table below. ("✓" indicates valid combinations.)
• Even when you have selected the segments using segment selectors X
and Y, these segments will be replaced by those selected using SEG.
When using SEG, be sure to select the same segments again.
"Y=2:DIA".
X
Y=
IIADA:2
D
=
Segment selector X
✓✓✓ ✓ ✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
Segment Check
This mode is applicable to segment mode (m,n). In this mode, you are to
register the number of segments. If the number of segments detected during
measurement is not the same as the registered one, error message " - - - - "
will appear. When this happens, the measured value is ignored.
This prevents incorrect measurements due to dust, drops of oil, or any other
particles on the scanning head. The error message will also appear when
the target is displaced from the measurement area.
•Non-clear targets are being measured.
1
2
3
4
5
6
ReceiverTransmitter
Segment selectors
X=(2. 3)
Y=(4. 5)
The registered number of edges is 6. If more than 6 edges are detected due
to dust as shown below, " - - - - - " will be displayed.
Floating dust
1
2
3
4
5
6
7
ReceiverTransmitter
* The segment check mode does not function with segment pairs in the
DIA, EDGE1, and EDGE2 modes.
35
CHAPTER 4 Functions and Controls
ProcedureCheck first that segment pair, (m,n), has been selected.
1. Use FNC to display "X SEG CHECK OFF".
SCEGXHECK OF F
4
2. Use SET to display "ON" on the right of "SEG CHECK" to activate the
segment check mode.
SCEGXHE CK ON
4
* When the currently selected segment pair is DIA, EDGE1, or EDGE2, "Y
SEG CHECK" will be displayed.
3. Enter the correct number of edges.
4. Press ENT to register.
5. Press FNC to display "Y SEG CHECK OFF".
6. Repeat steps 3 and 4.
Example of Operation
You are to measure the outer diameter of two round bars at the same time
using one scanning head. You need to use the segment check mode to
prevent incorrect measurements. In this case, the number of edges is 6.
1
2
3
4
5
6
ReceiverTransmitter
Segment selectors
X=(2. 3)
Y=(4. 5)
StepKey operationDisplay
1Use FNC to display "X SEG
CHECK OFF".
SCEGXHECK OF F
4
2Press SET.
SCEGXHE CK ON
4
3Enter "6".
4Press ENT to register.
SCEGXHECK ON
SCEGXHECK ON
[]
6
6
36
5Press FNC.
6Press SET.
7Press "6".
8Press ENT to register.
SCEGYHECK OF F
4
SCEGYHECK ON
4
SCEGYHE CK ON
SCEGYHECK ON
[
]
6
6
CHAPTER 4 Functions and Controls
Output Channel
The LS controller is equipped with two output channels (CH1 and CH2)
having their own displays and output terminals for simultaneous 2-point
measurement.
In this mode, you are to select which output channel to display or output
each of the measured values for X and Y. You can also choose to select the
equation containing X and Y as variables and obtain the result.
Types of output for CH1 and CH2
ChannelDisplayOutput
Main display
3 comparator LEDs (HI,
GO, LO)
CH1
3-step comparator output (relay)
➪
See P.64.
7-step comparator output
(control I/O)
Analog voltage output
RS-232C
BCD (option)
GP-IB (option)
➪
➪
See P.49.
➪
See P.62.
See P.68.
➪
See P.70.
➪
See P.47.
Subdisplay3-step comparator output
CH2
(control I/O)
➪
See P.62.
RS-232C
GP-IB (option)
* For the subdisplay you can use DISP to display the current settings,
measured value for CH1 in one of the measurement modes, or measured
value for CH2. Therefore, the subdisplay may not show the measured
value of the segment you have selected. However, the comparator output
and the BCD output will be sent based on the measured value of the
segment selected using the segment selectors.
Procedure1. Use FNC to display the current segment pairs for the output channels.
C HX1=
CH2= Y
2. Use UP/DOWN to move the cursor up.
3. Each time you press SET, the subdisplay will sequentially show "X", "Y",
"X+Y", "X-Y", "(X+Y)/2", "(X-Y)/2", "-X", "-Y", "-(X+Y)", "-(X+Y)/2", and "X"
again.
Choose the desired segment selector or equation.
4. Use UP/DOWN to move the cursor down.
C
HX1=
CH2= Y
5. Repeat step 3.
Equations
It is possible to choose the desired equation having X and Y as variables.
You can display and output the result of the equation through CH1 or CH2 or
both. Eight equations are available as mentioned in step (3) of the above
procedure.
37
CHAPTER 4 Functions and Controls
EquationDisplay or output
-XNegative X or Y value, or Remainder of OFFSET value
-Y
from which X or Y is subtracted (if OFFSET value is
registered)
X+YAddition of X and Y
X-YSubtraction of Y from X
(X+Y)/2Average of X and Y
(X-Y)/2Subtraction of Y from X, then division by 2
-(X+Y)Addition of X and Y, and negative of this result, or
Negative remainder of OFFSET value from which
addition of X and Y is subtracted (if OFFSET value is
registered)
-(X+Y)/2Negative average of X and Y
Negative remainder of OFFSET value from which
average of X and Y is subtracted (if OFFSET value is
registered)
➪
See P.43 or OFFSET value.
Example of Operation
You are to measure the outer diameter of the round bar and its eccentricity
with one scanning head. The measured value of the diameter will be displayed and output through CH1. The measured value of the eccentricity will
be displayed and output through CH2. Suppose that the following parameters have been set for CH1:
After step 10, each time you press FNC, the subdisplay will show the current
measurement mode (one of NORMAL, P-P, PEAK, BOTTOM), OFFSET
value, and tolerance limits. Be sure to select the parameters for CH2 in the
same manner as you do for CH1.
Use FNC to select the MODE, OFFSET, and LIMIT parameters. Note that
only 3-step differentiation is available for CH2 and that its output is sent
through the corresponding control I/O connector pins.
Enter the desired parameters for MODE, OFS (OFFSET), and LIMIT.
➪
See P.40 for MODE, P.43 for OFFSET, and P.29 or LIMIT.
Procedure1. First select the desired segment pair for the segment selectors and
equations for the output channels.
➪
See P.32 and P.35.
2.Press FNC once to display the current measurement mode.
3.Each time you press SET, the subdisplay will show sequentially "NORMAL", "P-P", "PEAK", and "BOTTOM". Display the desired mode.
4.Press FNC again to display the current OFFSET value.
5.Enter the desired value, and press ENT to register.
* OFFSET value is factory-set to "0.0000". Do not change it unless
otherwise required.
6.Press FNC again to display the current tolerance limits.
7.Use UP/DOWN to move the cursor up.
8.Enter the desired value, and press ENT to register.
9.Use UP/DOWN to move the cursor down.
10. Enter the desired value, and press ENT to register.
39
CHAPTER 4 Functions and Controls
Example of OperationContinued from P.39, you are to select "P-P" for measurement mode and
"0.1 mm" for tolerance upper limit. By selecting 0.1 mm, the comparator
output will be sent through CH2 if the measured value is above this limit.
StepKey operationDisplay
11Press FNC once to display
current mode.
HO2MCDE NORMA L
12Use SET to display "P-P".
HO2MCDE P-P
13Use FNC to display current
OFFSET value.
HF2OCFS TE
.0 0000
14Do not change OFFSET
value. Press FNC again to
display current tolerance
CH
60. 02000
H
- 60. 0000
L
limits.
CC
C MODE Key
CC
Switching Measurement
Modes
15Use UP/DOWN to move
cursor up.
CH
60. 02000
H
- 60. 0000
L
16Enter "0.1000".
[]
CH
H
L
0.12000
- 60. 0000
17Press ENT to register.
CH
H
L
0.12000
- 60. 0000
• When you finish setting the parameters, use DISP to show measured
values for CH2 on the subdisplay.
• Even when the subdisplay shows the current settings or measured value
in one of the measurement modes for CH1, the comparator output and
BCD output will be sent based on the measured value for CH2.
• NORMAL
Measured value is displayed or output as is.
• P-P (PEAK to PEAK)
The difference between the maximum (Peak) and minimum (Bottom)
values measured during the specified period is displayed and output.
40
• PEAK (PEAK HOLD)
The maximum value measured during the specified period is displayed
and output.
• BOTTOM (BOTTOM HOLD)
The minimum value measured during the specified period is displayed
and output.
CHAPTER 4 Functions and Controls
• "Specified period" refers to the duration of time from when a measurement
mode is set until when the mode is cancelled by pressing HOLD or inputting a signal through terminal TIM.
• Use FNC to set the measurement mode for CH2.
➪
See P.39.
Procedure
1. Press MODE to display the current measurement mode.
OM DE NORMA L
2. Each time you press SET, the subdisplay will show sequentially "NORMAL", "P-P", "PEAK", and "BOTTOM". Display the desired mode.
DD
D CAL (Calibration)
DD
Key
In this mode, the LS controller can be calibrated to correct possible measurement errors arising from irregularities on the target surface or inclination
of the target. Have ready two gauges of known dimension for calibration.
The LS controller can be calibrated for each of segment selectors X and Y.
Procedure1. Press CAL to calibrate segment selector X. "CAL X" will appear on the
subdisplay. At this time, the segment pair (SEG) remains the same.
However, the number of measurements for averaging (AVE) and measurement mode (MODE) will be set respectively to "1024, MOVE" and
"NORMAL". The LS controller will be reset to the initial parameters when
the calibration is over. Also, the hold and auto-zero functions are cancelled, and the offset function is invalid during calibration.
CL
A
0X000.
0 00 0.
0
0
2. Place a reference gauge in the measurement area to display the measured value (on the main display). Make sure that the displayed value is
not changing before you proceed to the next step.
3. Enter the known dimension of the reference gauge, then press ENT to
register. The cursor will move down, and the initial value will disappear
automatically when you press ENT.
A
CL
1
0
0X000.
4. Place another reference gauge in the measurement area, enter its dimension, and press ENT to register.
5. When the parameters are correctly set, "CALIB OK" will appear on the
subdisplay. If not, "CALIB ERROR" will appear. Check the measured
value and reference gauges for any errors, and repeat steps 1 to 4.
AKI
CLB
O
41
CHAPTER 4 Functions and Controls
6. Press CAL to calibrate segment selector Y. "CAL Y" will appear on the
subdisplay. You can display "CAL Y" by pressing CAL either when "CALIB
OK" appears or when "CAL X" has just been displayed by pressing CAL.
("CAL X" and "CAL Y" display alternately each time you press CAL.)
Note
CL
A
0Y000.
0 00 0.
0
0
7. Repeat steps 2 to 5 just as you do for calibrating segment selector X.
8. When you finish the calibration, press any of SEG, MODE, AVE, LIMIT,
OFS, PRM, FNC, PROG, or DISP to return to the measurement mode.
• To return to the factory calibration, enter "0.0000" on the upper and lower
lines of the subdisplay.
• When "CALIB ERROR" appears on the subdisplay, press CAL to display
the return to the previous status.
• When you use only one of the segment selectors X and Y, you can
calibrate only the selector you use. When you finish the calibration, first
check that "CALIB OK" appears on the subdisplay, then press any of the
keys mentioned in Step 8 to return to the measurement mode.
• When you measure the diameter of two round bars with only one scanning head, the measurement position differs one from the other. To
ensure higher accuracy, it is recommended that the reference gauge be
placed at each position and segment selectors X and Y be calibrated
separately.
• For even higher accuracy, the closer the dimension of the reference
gauge to the target size, the better.
• You cannot enter negative values.
• The value entered on the lower line of the subdisplay must be greater
than that on the upper line. If not, "CALIB ERROR" will appear.
Example of OperationYou are to measure a diameter of round bars that can vary from 10 mm to
20 mm. Two gauges, each having 10.0100 mm and 20.1000 in diameter are
used for calibration.
StepKey operationDisplay
1Press CAL to display "CAL
X".
CL
A
0X000.
0 00 0.
0
0
2Enter "10.0100".
A
CL
[1
0 00 0.
0
]
0
0X010.
3Press ENT to register.
0X010.
Cursor will move down.
A
CL
1
0 00 0.
0
0
42
(To be continued)
CHAPTER 4 Functions and Controls
EE
E DISP
EE
(Subdisplay) Key
4Enter "20.1000".
5Press ENT to register.
"CALIB OK" will appear.
6Press CAL to display
"CAL Y".
7Repeat steps 2 to 5 to
calibrate Y.
One of the following items appears on the subdisplay:
• Status of settings during panel key operation
• Measured value for CH1 in NORMAL, P-P, PEAK, or BOTTOM mode
Measured value in any of the desired mode can be displayed regardless
of the mode set using MODE. The comparator output and BCD output will
be sent based on the mode set using MODE.
CL
A
AKI
CLB
CL0Y
A
0X010.
1
[
0 10 0.
2
O
0 00 0.
0
0
]
000.
0
0
• Current settings for CH1
Current parameters or modes set using SEG, MODE, or AVE can be
displayed.
• Measured values for CH2
Each time you press DISP, the values or parameters mentioned in (2), (3),
and (4) will display alternately.
* These values or parameters cannot be changed on the subdisplay.
FF
F OFS (Offset) Key
FF
Procedure1. Press OFS.
Offset value is set in this mode. This value is added to the measured value,
and the total value is displayed and output. Any value between -999.9999
and 9999.9999 can be specified as an offset value. If the diameter of a round
bar or width of a sheet material is too large to measure with one scanning
head, offset value can be used to measure it using two scanning heads. In
this case, the diameter or width is displayed or output in absolute value.
* Use FNC to set the offset value for CH2.
➪
See P.39.
2. Enter the desired offset value.
3. Press ENT to register.
Example of Operation
43
CHAPTER 4 Functions and Controls
You are to measure the diam-eter of a round bar. This diameter, supposedly 100 mm, is displayed and
output in absolute value.
HEAD1
X
D
Edge1
Edge2
HEAD2
Y
TransmitterReceiver
Edge1
Edge2
Place two scanning heads as shown above.
Enter the following segment pair for the segment selectors and equation for
the output channel:
Segment selectorX= EDGE 1 or EDGE 2
Y= 2: EDGE 1 (when EDGE 2 is selected for X) or
2:EDGE 2 (when EDGE 1 is selected for X)
Output channelCH1= -(X+Y)
With the above parameters, you will obtain values X and Y in the figure.
Value D, diameter of the target, is obtained by subtracting values X and Y
from the width of the measuring area. To obtain the width of the measuring
area, add values X, Y, and D. This width is used as the offset value.
See P.37 for the segment selectors and output channels.
Have ready a target of known diameter. Suppose that its diameter is
100.1000 mm, and that values X and Y are measured to be 35.2345 mm and
36.1234 mm, respectively.
StepKey operationDisplay
1Use FNC to set segment
XE
pair for X and Y.
➪
See P. 32.
=DGE I
Y=
2EDGE 2:
2Use FNC to set equation for
CH1 and CH2. Values X
and Y are obtained.
➪
See P.37.
3Use DISP to display X and
Y. Take a note of these
values.
4Use FNC again to set
equation for CH1 and CH2.
Target diameter is to be
displayed and output
through CH1.
5Press OFS to display
"OFFSET".
6Enter "171.4579".
* 171.4579 is the total of X,
Y, and D.
7Press ENT to register, or
press CLR and reenter
correct value.
C
HX1=
CH2= Y
35 2.345
CH
2 36 1.234
NO
R
C H-1=
CHY2=
OFEF
ST
0 00 0.
OFEF
ST
[
1.47
1
F
OFE
ST
1.47
1
()
X+Y
0
]
5
7
9
5
7
9
44
CHAPTER 4 Functions and Controls
GG
G PROG (Program)
GG
Key
The LS controller can store the currently displayed settings. Up to 10 types
of settings can be stored with each assigned a program number. Each type
of settings can be retrieved through its program number for quicker operation.
* BAUD RATE, BUZZER ON/OFF, HOLD type, HEAD2 ON/OFF, and
POSITION CHECK cannot be stored.
Storing (SAVE) Settings1. Enter the desired parameters.
2. Press PROG to display "PROG".
R OG
P1SALOAD
VE
3. Enter the desired program No.
4. Press ENT to register. At this moment, only the program No. has been
selected.
5. Use UP/DOWN to move the cursor down to "SAVE".
6. Press SET to display "SA OK?". Press ENT to accept or CLR to cancel.
R OGP5SALOAD
OK ?
7. When you enter ENT, the settings will be stored under the selected
program No. When you press CLR, you can resume from step 3.
Example of OperationStoring the current settings under program No.6.
StepKey operationDisplay
1Press PROG.
R OGP2SALOAD
2Enter "6".
P6SALOAD
R OG
3Press ENT.
R OGP6SALOAD
4Use UP/DOWN to move
R OG
cursor down.
P6SALOAD
5Press SET.
R OG
P6SALOAD
VE
][
VE
VE
VE
OK ?
6Press ENT to accept.
R OG
P6SALOAD
VE
45
CHAPTER 4 Functions and Controls
Retrieving (LOAD) Settings1. Press PROG to display "PROG".
2. Enter the desired program No.
R OGP1SALOAD
VE
3. Press ENT to accept. At this moment, only the program No. has been
selected.
4. Use ON/OFF to move the cursor up to "LOAD".
5. Press SET to display "LO OK?". Press ENT to accept or CLR to cancel.
R OGP2SALO OK
6. When you press ENT, the settings stored under the specified program No.
will be retrieved.
7. When you press CLR, you can resume from step 3.
Example of OperationRetrieving program No.3
StepKey operationDisplay
1Press PROG.
2Enter "3".
3Press ENT.
4Use UP/DOWN to move
cursor up to "LOAD".
VE
?
P1SALOAD
R OG
VE
P3SALOAD
P3SALOAD
[]
R OG
VE
R OG
VE
R OGP3SALOAD
VE
46
5Press SET.
6Press ENT to retrieve.
P3SALO OK
R OG
R OG
P3SALOAD
?
VE
VE
• Program can be retrieved by using the program selector input terminals or
control I/O connector port.
➪
See P.62.
• See P.49 for storing and retrieving programs through the RS-232C interface.
Tab le of I/O
Terminals
CHAPTER 4 Functions and Controls
Analog Voltage Output
Analog voltage signal for CH1 can be obtained through the MON terminals.
The LS controller has been factory-set to convert a range from -60 mm to
+60 mm into a voltage range from -6 V to +6V. (No voltage signal is provided
for CH2. If voltage signal for CH2 is required, consult the KEYENCE distributor.)
Analog voltage range: A range between -6 V and +6 V can be set to
match a desired measurement range. Any value between -999.9999 and
9999.9999 can be selected. Note, however, that the difference between the
maximum and minimum measurements (those that you specify to correspond respectively to +6 V and -6 V) must fall within a range between 0.024
mm and 240 mm.
Calibration: When the segment selectors are calibrated, the analog voltage
signal through these terminals and RS-232C interface is automatically sent
based on the calibrated value.
Offset: When the segment selectors are calibrated, the selected offset
value is added to the analog voltage signal sent through these terminals and
RS-232C interface.
The analog voltage which appears on the main display and that which is
output through the MON terminals are always the same.
Comparator Output
For CH1 (Channel 1)
7-step differentiation
CH1
3-step differentiation
Tolerance LimitOutput Device/Terminal
HI3
HI2
HI1
GO
LO1
LO2
LO3
HIGH
GO
LOW
HIGH
GO
LOW
HI1
GO
LO1
HIGH
GO
LOW
CH1
CH1
Control I/O (open-collectors)
Front panel (LEDs)
Comparator output terminals (relays)
Control I/O (open-collectors)
Front panel (LEDs)
For CH2 (Channel 2)
CH2
Only 3-step
differentiation is
available.
HIGH
GO
LOW
HI
GO
LO
CH2
Comparator output terminals (relays)
Control I/O (open-collectors)
47
CHAPTER 4 Functions and Controls
Program Selector Input
Terminals
AUTO ZERO Terminals
The desired program No. can be selected from external equipment using the
following terminals:
<Terminal Block> program selector input:
Programs 1 to 5 selectable
<Control I/O> terminals #2 to #11:
Programs 1 to 10 selectable (program selector input)
Note that when two or more terminals are grounded at the same time, the
program with the smallest number will be selected.
Auto Zero function can be activated or deactivated from external equipment
using the following terminals:
<Control I/O> zero set terminal:
Ground to activate Auto Zero function
<Control I/O> zero reset terminal:
Ground to deactivate Auto Zero function
Auto Zero function can be activated as many times as desired without being
deactivated.
HOLD Terminal
Laser Emission Control
Input Terminal
Hold function can be activated or deactivated from external equipment using
the following terminals:
<Terminal Block> HOLD synchronous input:
Ground to activate and open to deactivate HOLD function
<Control I/O> synchronous input:
Ground to activate and open to deactivate HOLD function
Unlike Auto Zero function, no terminal is provided for deactivating the Hold
function. That is, while each of the above terminals is grounded, the Hold
function remains activated. However, this function remains deactivated when
the terminal is left open.
The laser emission indicator LED lights when this terminal is left open,
stopping the emission of the laser beam. For shipment, a jumper is provided.
<Terminal Block> Laser Remote:
Ground to stop and open to release the emission
48
4-4. I/O Terminal
CHAPTER 4 Functions and Controls
RS-232C I/O Terminal
Specifications
Pin Assignment
The LS controller is provided with the RS-232C interface, which allows the
LS controller to communicate with an external computer. Data transmission
or change of the settings can be remotely performed.
The LS controller employs the RS-232C interface (recommended standard
by EIA: Electronic Industries Association) for data transmission. LS controller
is defined as "MODEM".
Communication parameters
DuplexFull
SynchronizationStart-stop
Data formatASCII
Data length8 bits
Stop bit length1 bit
Parity checkNone
Baud rate75/150/300/600/1200/2400/4800/9600/19200 bps*
*Selectable by setting
D-sub 25 pin female connector
111213
2425
6 7 8 910
20212223
21345
141516171819
Pin 20 is internally connected to the power source via a pull-up resistor. This
pin remains ON unless it is grounded. Therefore, measurement data can be
output regardless of whether pin 20 is connected to external equipment.
Pin No. Modem signal codeModem signalDirection
1ShieldShield (to signal ground)–
2SD (TXD)Input data from externalInput
equipment
3RD (RXD)Output measurement dataOutput
to external equipment
5CS (CTS)ON (High) at power-onOutput
6DR (DSR)ON (High) at power-onOutput
7SG (GND)Signal ground–
8CD (DCD)ON (High) at power-onOutput
20ER (DTR)Enables output ofInput
measurements to external
equipment when ON (High)
49
CHAPTER 4 Functions and Controls
Connector Pin Assignment
Remote Operation
Use straight type cable.
Connection example 1Connection example 2
Pin No.Pin No.Pin No.Pin No.
(PC)(LS)(PC)(LS)
1111
2222
3333
5577
664
776
888
2020
In addition to the pin connections shown here, all identically-numbered pins of PC
and LS controller can be
connected.
11
22
..
..
..
5
20
OO
The settings and data output can be performed using external equipment
such as a computer, instead of the front panel keys of the LS controller. This
operation method is referred to as remote operation. The items which can be
remotely operated are divided into four sections (setting of measurement
conditions, setting of output conditions, output of measurement data and
output of setting status), and are described in the following.
"Format" used in the following description means input form.
Setting Measurement
Conditions
[P$] [S$] [H$] List
Set the measurement conditions for the desired program. The format differs
with the setting item, as described below.
Format:
([P$] [S$] [H$] ) [COMMAND] CR
Head No. (Scanning head No.)
Segment No. (Segment Selector)
Program No.
Note that one or more of the commands in ( ) can be omitted.
When [P$] is omitted, the current program No. will be valid.
When [S$] is omitted, the current segment selector will be valid.
When [H$] is omitted, the current scanning head will be valid.
FunctionCommandDescription
Specifying [P$]P1Specifies program No. 1
(Program No.)P2Specifies program No. 2
P3Specifies program No. 3
P4Specifies program No. 4
P5Specifies program No. 5
P6Specifies program No. 6
P7Specifies program No. 7
P8Specifies program No. 8
P9Specifies program No. 9
PASpecifies program No. 10
Specifying [S$]XSpecifies segment selector X
(Segment selector)YSpecifies segment selector Y
Specifying [H$]1Specifies scanning head No. 1
(Scanning head No.)2Specifies scanning head No. 2
50
CHAPTER 4 Functions and Controls
[COMMAND] List
FunctionCommandDescription
Specifying segmentSSSpecifies segment check ON
checkSRSpecifies segment check OFF
NnSpecifies the number of
segments ("n") for segment check
SpecifyingDSpecifies DIA
measurement segmentESpecifies EDGE1
FSpecifies EDGE2
Sm, nSpecifies values ("m" and "n)
for SEG (m, n) (m < n)
Setting number ofA0Sets 1
measurementsA1Sets 2
for averagingA2Sets 4
Setting channel toJ1Sets CH1 only
which HOLDJ2Sets both CH1 and CH2
function is applied
Setting AUTO ZEROZSSets AUTO ZERO function
functionZRCancels AUTO ZERO function
Setting HOLDHSSets HOLD function
functionHRCancels HOLD function
operation
• When the command for locking of panel key operation is entered, any
change of settings cannot be performed using the panel keys. At this time,
setting indicator LOCK lights.
• If there is an error in the format, all the commands in that format are
disregarded.
• Only one command for setting various functions can be entered for one
format.
51
CHAPTER 4 Functions and Controls
Entry Examples
• To specify "1" for program No., "Y" for segment selector, "1" for scanning
head No., and SEG (3, 4) for measurement segment: "P1Y1S3,4 CR"
• To specify "2" for program No., "X" for segment selector, and "OFF" for
segment check: "P2XSR CR"
• To specify "3" for program No., "X" for segment selector, "1" for scanning
head No., and "DIA" for measurement segment: "P3D CR"
• To set the number of measurements for averaging to 1,024, while not
changing the other settings on the front panel: "AA CR"
Setting Output ConditionsSets the output conditions (processing methods for measured values) for the
desired program. The format differs with the setting item, as described
below.
Format:
( [P$] [C$] ) [COMMAND] CR
Channel No.
[P$] [C$] List
[COMMAND] List
52
Program No.
Note that one or more of the commands in ( ) can be omitted.
When [P$] is omitted, the current program No. will be valid.
When [C$] is omitted, the current channel No. will be valid.
FunctionCommandDescription
Specifying program No.P1Specifies program No. 1
P2Specifies program No. 2
P3Specifies program No. 3
P4Specifies program No. 4
P5Specifies program No. 5
P6Specifies program No. 6
P7Specifies program No. 7
P8Specifies program No. 8
P9Specifies program No. 9
SettingMNSets NORMAL mode
measurement modeMPSets P-P mode
MESets PEAK mode
MBSets BOTTOM mode
Setting toleranceHnSets value ("n") for HIGH
range for 3-level(upper tolerance limit)
differentiationLnSets value ("n") for LOW
Commands for setting functions
(lower tolerance limit)
Setting OFFSETOnSets value ("n") for OFFSET value
value
Setting displayGnSets the number of display digits
digit deletionto be deleted
Note:
• Entry of the bracketed commands ([ ]) can be omitted.
•When no command for specifying program No. is entered, the program currently used on the front panel is specified.
•When no command for specifying output channel is entered, CH1 is
specified.
• When there is an error in the format, all the commands in that format are
disregarded.
• Only one command for setting functions can be entered for one format.
Entry examples
• To specify "3" for program No., CH2 for output channel, and 10.0000 mm
for OFFSET value: "P3WO10.0000 CR"
• To specify 1.2345 mm for LOW1 of tolerance range for 7-level differentiation, while not changing the other settings on the front panel: "TD1.2345
CR"
53
CHAPTER 4 Functions and Controls
Output of Measurement DataSpecifies the output method of the processed measurements which are
displayed on the main display (CH1) and sub-display (CH2), and outputs
them according to the specified method.
Input format: [COMMAND] CR
FunctionCommandDescription
Specifying dataX0Outputs measurement data of CH1 once
output methodX1Outputs measurement data of CH1
of data outputtingcontinuously
X2Outputs measurement data of CH2 once
X3Outputs measurement data of CH2
continuously
X4Outputs measurement data of CH1 and
then that of CH2 once
X5Outputs measurement data of CH1 and
CH2 alternately and continuously
X6Stops continuous output
Output data format:
• At normal measurement
Measurement in "mm"
1st data 2nd data 3rd data 4th data 5th data 6th data 7th data 8th data 9th data 10th data
C
3
'-' or 10 digit210 digit110 digit010 digit
-1
-2
10 digitDecimal point
10 digit-310 digit-410 digit
R
Measurement in "inch"
1st data 2nd data 3rd data 4th data 5th data 6th data 7th data 8th data 9th data 10th data
C
2
'-' or 10 digit110 digit310 digit
-1
10 digit
-2
-3
10 digitDecimal point
10 digit-410 digit-510 digit
R
When there is no 1st, 2nd and 3rd data, signal corresponding to blank
( SPACE ) is output, allowing these digits to be displayed as blank.
• At erroneous measurement
Measurement in "mm" or "inch"
1st data 2nd data 3rd data 4th data 5th data 6th data 7th data 8th data 9th data 10th data
C
---------
R
Measurement data processed on the basis of the current settings is output,
according to the specified command, 10 ms max. after the command has
been received.
When "X4" or "X5" command is entered, the values of CH1 and CH2 obtained when the command is received by the LS are output in order of CH1
and CH2.
54
CHAPTER 4 Functions and Controls
Output of Setting Status
[P$] [Ø$] List
Specifies and outputs the setting status of the desired program.
The format differs with the setting information to be output, as described
below.
Format:
( [P$] [Ø$] ) [COMMAND] CR
Output Channel
Program No.
Note that one or both of the commands in ( ) can be omitted.
When [P$] is omitted, the current program No. will be valid.
When [Ø$] is omitted, CH1 will be specified.
FunctionCommandDescription
SpecifyingP1Specifies program No. 1
program No.P2Specifies program No. 2
P3Specifies program No. 3
P4Specifies program No. 4
P5Specifies program No. 5
P6Specifies program No. 6
P7Specifies program No. 7
P8Specifies program No. 8
P9Specifies program No. 9
PASpecifies program No. 10
Specifying andX?Outputs setting information of
outputtingsegment selector X
setting informationY?Outputs setting information of
segment selector Y
XS?Outputs whether segment check
for segment selector X is set to
ON or OFF
YS?Outputs whether segment check
for segment selector Y is set to
ON or OFF
XN?Outputs preset value of segment
check for segment selector X
YN?Outputs preset value of segment
check for segment selector Y
TA?Outputs preset value of HIGH3 for
7-level differentiation
TB?Outputs preset value of HIGH2 for
7-level differentiation
TC?Outputs preset value of HIGH1 for
7-level differentiation
TD?Outputs preset value of LOW3 for
7-level differentiation
TE?Outputs preset value of LOW2 for
7-level differentiation
(To be continued)
55
CHAPTER 4 Functions and Controls
FunctionCommandDescription
Specifying and
outputting setting
information
TF?Outputs preset value of LOW1for
7-level differentiation
I?Outputs whether comparator output
is set to ON or OFF
T?Outputs differentiation level (3- or
7-level)
AS?Outputs preset value for analog
output voltage scaling
A?Outputs preset value for number of
measurements for averaging
B?Outputs averaging method (simple
or sequential )
U?Outputs measurement unit
( "mm" or "inch" ).
J?Outputs setting information for
HOLD/AUTO ZERO function.
Z?Outputs whether AUTO ZERO
function is set to ON or OFF.
HO?Outputs whether HOLD function
is set to ON or OFF.
C?Outputs which arithmetic operation
method is set
M?Outputs measurement mode
H?Outputs preset upper tolerance limit
for 3-level differentiation
L?Outputs preset lower tolerance limit
for 3-level differentiation
O?Outputs preset OFFSET value
G?Outputs preset number of deleted
display digits
56
CHAPTER 4 Functions and Controls
The output data format from the controller is as follows:
Output data format
● Segment selector
When DIA is specified:
1st data 2nd data 3rd data 4th data
DIAC
R
When EDGE1 is specified:
1st data 2nd data 3rd data 4th data 5th data 6th data
E D G E 1C
R
When SEG (1, 2) for the 2nd scanning head is specified:
1st data 2nd data 3rd data 4th data 5th data 6th data 7th data 8th data 9th data 10th data 9th data
2:S001,00C
R
2
● Segment check
When segment check is set to ON:
1st data 2nd data 3rd data
O NC
R
When value for segment check is set to 4:
1st data 2nd data
4CR
● 7-level differentiation
Example: When HIGH3 is set to 60.0000 mm:
1st data 2nd data 3rd data 4th data 5th data 6th data 7th data 8th data 9th data 10th data
R
SPSP60.0000
C
* For HIGH2, HIGH1, LOW1, LOW2, or LOW3, the data is output in the
same way as the above.
* SP (SPACE) appears in the digit where there is no data.
● Comparator output (ON or OFF)
When it is set to ON:
1st data 2nd data 3rd data
O NC
R
● Differentiation level (3 or 7)
Example: When differentiation level is set to 3:
1st data 2nd data
3CR
● Analog output voltage scaling
Example: When 60.0000 mm is set for +6 V and -60.0000 mm for -6 V:
1st data 2nd data 3rd data 4th data 5th data 6th data 7th data 8th data 9th data 10th data
SPSP60.0000
11th data 12th data 13th data 14th data 15th data 16th data 17th data 18th data 19th data 20th data
-SP60 .000 0C
,
R
* SP (SPACE) appears in the digit where there is no data.
57
CHAPTER 4 Functions and Controls
● Number of measurements for averaging
When 16 is set:
1st data 2nd data 3rd data
1 6C
R
When 1,024 is set:
1st data 2nd data 3rd data 4th data 5th data
1024C
R
● Averaging method
When simple averaging is set:
1st data 2nd data 3rd data 4th data 5th data 6th data 7th data
S I M P L EC
R
When sequential averaging is set:
1st data 2nd data 3rd data 4th data 5th data
M O V EC
R
● Measurement unit
When "mm" is set:
1st data 2nd data 3rd data
m mC
R
When "inch" is set:
i n c hCR
● Channel for HOLD or AUTO ZERO
When HOLD or AUTO ZERO function is set for only CH1:
1st data 2nd data 3rd data 4th data
C H 1C
R
When HOLD or AUTO ZERO function is set for both CH1 and CH2:
1st data 2nd data 3rd data 4th data 5th data 6th data 7th data 8th data
C H 1 , C H 2C
R
● AUTO ZERO function (ON or OFF)
When AUTO ZERO function is set to ON:
1st data 2nd data 3rd data
O NC
R
58
When AUTO ZERO function is set to OFF:
1st data 2nd data 3rd data 4th data
O F FC
R
● HOLD function (ON or OFF)
When HOLD function is set to ON:
1st data 2nd data 3rd data
O NC
R
● Arithmetic operation method
When "X" is set:
1st data 2nd data
XC
R
When "X+Y" is set:
1st data 2nd data 3rd data 4th data
X + YC
R
When "(X+Y)/2" is set:
1st data 2nd data 3rd data 4th data 5th data 6th data 7th data 8th data
(
X+Y
)
/ 2C
CHAPTER 4 Functions and Controls
R
● Measurement mode
When NORMAL is set:
1st data 2nd data 3rd data 4th data 5th data 6th data 7th data
N O R M A LCR
When P-P is set:
1st data 2nd data 3rd data 4th data
P - PC
R
● 3-level differentiation
Example: When upper limit is set to 10.0000 mm and lower limit to -10.0000
mm:
1st data 2nd data 3rd data 4th data 5th data 6th data 7th data 8th data 9th data 10th data
SPSP10.0000CR
1st data 2nd data 3rd data 4th data 5th data 6th data 7th data 8th data 9th data 10th data
-SP10 .000 0CR
*SP (SPACE) appears in the digit where there is no data.
● OFFSET function
Example: When OFFSET value is set to 12.3456 mm:
1st data 2nd data 3rd data 4th data 5th data 6th data 7th data 8th data 9th data 10th data
SPSP12.3456CR
* SP (SPACE) appears in the digit where there is no data.
59
CHAPTER 4 Functions and Controls
● Display digit suppress function
When the number of suppressed display digits is 0:
1st data 2nd data
0C
R
When the number of suppressed display digits is 3:
1st data 2nd data
3C
R
Program Example in BASICProgram 1
Pressing the space key of the computer allows the measured values to
appear on the computer display.
100OPEN "COM1:4800,N,8,1,RS,CS,DS,CD" FOR RANDOM AS #1
110A$=INKEY$
120IF A$<>" " THEN 110
130PRINT #1, "X0"
140INPUT #1, B$
150PRINT B$
160GOTO 110
Program 2
The following settings are specified through the computer: SEG (2, 3), NORMAL mode, upper and lower tolerance limits (20.0000 mm and 10.0000 mm),
number of measurements for averaging (512), and simple averaging mode.
After these settings are stored under program No. 3, the measured values are
displayed.
100OPEN "COM1:4800,N,8,1,RS,CS,DS,CD" FOR RANDOM AS #1
110PRINT #1, "S2, 3"Sets SEG (2, 3)
120PRINT #1, "MN"Sets NORMAL mode
130PRINT #1, "H20"Sets upper limit to 20.0000 mm
140PRINT #1, "L20"Sets lower limit to 10.0000 mm
150PRINT #1, "A9"Sets the number of measurements
for averaging to 512
160PRINT #1, "BS"Sets simple averaging method
170PRINT #1, "SV3"Saves settings into program
No.3
180For I=1 TO 2000: NEXT I
190PRINT #1, "X0"
200PRINT #1, A$
210PRINT A$Display measured values
220GOTO 190
60
CHAPTER 4 Functions and Controls
Program 3
The following settings are specified through the computer: simultaneous
measurement of DIA and EDGE1, 100 samplings for each of DIA and EDG1,
and display of maximum, minimum and average values of these measurements.
100DIM D$ (100), E$ (100)
110OPEN "COM1:4800,N,8,1,RS,CS,DS,CD" FOR RANDOM AS #1
120PRINT #1, "X1D"Sets segment selector X, measurement part 1, and DIA
130PRINT #1, "Y1E"Sets segment selector Y, measurement part 1, and EDGE1
140PRINT #1, "VC0"Sets output channel CH1 and arithmetic operation "X"
150PRINT #1, "VMN"Sets output channel CH1 and NORMAL mode
160PRINT #1, "WC1"Sets output channel CH2 and arithmetic operation "Y"
170PRINT #1, "WMN"Sets output channel CH2 and NORMAL mode
180PRINT "START or END ? (Press [S] or [E] key)"
190PRINT
200A$=INPUT$ (1)
210IF A$="E" OR A$="e" THEN END
220IF A$<> "S" AND A$<> "S" THEN 200
230PRINT #1, "X5"Sets continuous output of CH1 and then CH2 data
240FOR I=1 TO 100
250 INPUT #1, D$ (I)Inputs data
260 INPUT #1, E$ (I)
270NEXT I
280PRINT #1, "X6"Stops continuous output
290 FOR I=1 TO 500 :NEXT I
300IF EOF (1) = 0 THEN INPUT #1, DAMMY$ : GOTO 300
310DMAX=-10000 :DMIN=10000 :DSUM=0
320EMAX=-10000 "EMIN=10000 :ESUM=0
330FOR I=1 TO 100
340D=VAL (D$ (I))
350IF D>DMAX THEN DAMX=D
360IF D<DMIN THEN DMIN=D
370DSUM=DSUM+D
380E=VAL (E$ (I))
390IF E>EMAX THEN EMAX=E
400IF E<EMIN THEN EMIN=E
410ESUM=ESUM+E
420NEXT I
430DAVE=DUSM/100
440EAVE=ESUM/100
450PRINT"[DIA][EDGE]"
460PRINT USING "MAX = ####,#### mm ####, #### mm" ; DMAX, EMAX
470PRINT USING "MIN = ####, #### mm ####, #### mm" ; DMIN, EMIN
480PRINT USING "AVE = ####, #### mm ####, #### mm" ;DAVE, EAVE
490PRINT
500GOTO 180
Displays prompt for
starting or ending
measurement
Processors or discards
data received after
continuous output has
been stopped
Performs arithmetic
operation to obtain
maximum, minimum, and
average values
Displays
maximum,
minimum and
average
values
Note: The baud rate is set to 4800 bps in these 3 examples. Set the baud rate according to your requirements.
61
CHAPTER 4 Functions and Controls
36-pin Connector
(Control I/O)
Specifications
Output
• TTL open-collector (7406 or equivalent)
I C
• Max. applicable voltage: 30 V
• Max. sink current: 40 mA
Input
• TTL voltage level, negative logic
(74LS19 or equivalent)
9Program No. 8 selector input: P827Comparator output for CH1: HI2
10Program No. 9 selector input: P928GND
11Program No. 10 selector input: P1029Comparator output for CH1: HI3
12Synchronous input: TIM30GND
13GND31Comparator output for CH2: LO
14AUTO ZERO input: ZERO SET32GND
15GND33Comparator output for CH2: GO
16AUTO ZERO RESET input:34GND
ZERO RESET
17Comparator output for CH1: LO335Comparator output for CH2: HI
18GND36GND
OUT
I C
62
CHAPTER 4 Functions and Controls
Input Signal
Program (No. 1 to 10) selector input: P1 to P10
• Used to select the desired program from external equipment.
• When pin 2 (3, 4, ••• or 11) is grounded with GND for an instant, program
No. 1 (2, 3, ••• or 10) is selected.
• When two or more pins are grounded at the same time, the program with
the smallest number will be selected.
Synchronous input: TIM
• When pin 12 (TIM) is grounded with pin 13, the measured value is retained. This value can be retained until pin 12 is disconnected from pin 13.
(The measured value is also retained when terminal 3 on the back panel
is grounded with terminal 4).
AUTO ZERO input: ZERO SET
• When pin 14 (ZERO SET) is grounded with pin 15 (ZERO SET pulled
LOW), the measured value is reset to "0" (mm or inch) for use as a
reference. This allows a variation from this reference to be output as a
measured value.
• The AUTO ZERO input uses a one-shot circuit that accepts only the first
LOW signal and ignores the following ones. This input, therefore, must be
reset first to input another AUTO ZERO input signal.
Output Signal
Functions of
Terminal Blocks
Analog Voltage Output
Terminals
AUTO ZERO RESET Input: ZERO RESET
• When pin 16 (ZERO RESET) is grounded with pin 15 (ZERO RESET
pulled LOW), the AUTO ZERO input is cleared, and normal measurement
is resumed.
• The AUTO ZERO RESET input uses a one-shot circuit that accepts only
the first LOW signal and ignores the following ones. This input, therefore
must be reset first before inputting another AUTO ZERO RESET signal.
Comparator outputs: LO3, LO2, LO1, GO, HI1, HI2, HI3, LO, GO, and HI
• Each comparator output is sent through the corresponding open-collector.
• LO 3 to HI3 are used for 7-step differentiation with CH1. LO, GO, and HI
are used for 3-step differentiation CH2 at simultaneous 2-point measurement.
• LO1, GO, and HI1 can be used also for 3-step differentiations.
Analog voltage proportional to the measured value is output.
±6 V is equal to ±60 mm (when voltage range of ±6 V is set to ±60 mm
measuring range).
Output impedance: 0 Ω, Current capacity: 10 mA
63
CHAPTER 4 Functions and Controls
Control Input Terminals
Input specifications
Hold Synchronous Input
Terminals
Internal circuit
+12V
Photocoupler
Example of connection
•Contact (relays or switches)
LS controller
1.8 kΩ
680 Ω
Control input
terminals
GND
Control input terminals
GND
•Solid-state (transistors)Open-collector output
LS controller
Control input
terminals
GND
• When terminal 3 is grounded with terminal 4, the displayed and output
values are retained.
• When terminal 3 is disconnected from terminal 4, normal measurement is
resumed.
• When the LS controller is set to the P-P, PEAK, or BOTTOM mode, the
previous measured value is reset by grounding terminal 3 with terminal 4
and then disconnecting these terminals.
Program Selector Input
Terminals
Laser Emission Control Input
Terminals (LASER REMOTE)
Control Output Terminals
Output Specifications
Comparator Output
Terminals
The desired program No. 1 to 5 can be selected from external equipment
using these terminals.
When terminal 5 (6, 7, 8, or 9) is grounded with terminal 10 for an instant,
program No. 1 (2, 3, 4, or 5) is selected and retrieved.
* When two or more terminals are grounded at the same time, the program
with the smallest number will be selected.
When terminals 11 and 12 are connected, the laser emission indicator LED
on the front panel lights. And, laser is emitted approx. 5 seconds after the
connection is made. When terminal 11 is disconnected form terminal 12, the
laser emission indicator LED goes out, and the laser emission stops.
The LS controller is delivered with a jumper inserted between terminals 11
and 12.
Relay: SPST-NO contact, 250 VAC, 2 A (resistive load)
• HIGH OUT terminals: HI
Comparator output is sent through these terminals when the measured value
is above the upper tolerance limit.
• LOW OUT terminals: LO
Comparator output is sent through these terminals when the measured value
is below the lower tolerance limit.
• GO OUT terminals: GO
Comparator output is sent through these terminals when the measured value
is within the tolerance range.
64
Note
Use a crimp contact for wiring as shown below. Be sure to use a cable with
at least 0.4-mm sheath.
Round-shapeFork-shape
CHAPTER 5
MOUNTING SCANNING HEAD
5-1. Alignment of Optical Axis (LS-3060 and LS-3033 only)
The LS-3060 is delivered with the transmitter and receiver (scanning head)
mounted to the detachable frame. When using the LS-3060 with the transmitter and receiver removed from the frame on when using the LS-3033, be
sure to align them in terms of the optical axis of the laser. (See the figure
below.) Misalignment of the optical axis will cause measurement errors.
(A special optical axis checker is available with the LS-3060 and LS-3033.
Refer to "5-2. Checking Alignment of Optical Axis" on page 66 for the alignment procedures.)
Allowance for
Aligning Transmitter
and Receiver
Deviation in height or width
Within ±1 mm
Transmitter
Within ±1 mm
Receiver
Inclination in height or width
Transmitter
Within ±0.25°
Receiver
Transmitter
Within ±1 mm
Within ±1 mm
Within ±0.25°
Transmitter
Within ±0.25°
Receiver
Within ±0.25°
65
CHAPTER 5 Mounting Scanning Head
5-2. Checking Alignment of Optical Axis (LS-3033 and LS-3060 only)
Not applicable to LS-3033 SO (8073).
Use the POSITION CHECK function to check whether the transmitter and
receiver (scanning head) are aligned in terms of the optical axis.
The procedure is as follows:
1. Use PRM to display "HEAD1" and "POSITION CHECK" on the
subdisplay.
EADH1
OSTPIOINCHECK
2. Attach the optical axis checker to the receiver.
3. When the transmitter and receiver are aligned in terms of the optical axis,
"GOOD" will appear. If not, "P-Err" will appear.
µm
OFFSET LOCKPOSITION
µm
OFFSET LOCKPOSITION
4. When 2 scanning heads are used, press PRM again to display "HEAD2"
and "POSITION CHECK". (To use 2 scanning heads, the optional PC
board for the second scanning head is required.) Then, use the optical
axis checker to check whether the transmitter and receiver of the second
scanning head are aligned in the same manner as those of the first
scanning head.
EADH2
OSTPIOINCHECK
When you use the LS-3033SO (8073), note the following:
1. Optical axis alignment — Install the transmitter and receiver so that the
both reference base edge are aligned. This also completes the optical
alignment. The optical axis checker, which is provided for standard LS3033, cannot be used for LS-3033SO (8073). No optical axis checker is
provided for LS-3033SO (8073) for this reason.
66
2. Transmitter-to-receiver distance — Set this distance to 120 mm.
CHAPTER 5 Mounting Scanning Head
Error Messages
(For Models LS-3060 and
LS-3033)
In the POSITION CHECK mode, the target position indicator LEDs show
whether the transmitter and receiver (scanning head) are aligned in terms of
the optical axis.
DisplayDescription
mm
inch
OFFSET LOCKPOSITION
1234 567
"GOOD" appears, and target position indicator LED
4 lights.
mm
inch
OFFSET LOCKPOSITION
2314567
The transmitter and receiver are aligned .
"P-Err" appears, and LED 3 or 5 lights.
OFFSET LOCKPOSITION
2314567
"P-Err" appears, and LED 1 or 7 lights.
OFFSET LOCKPOSITION
2314567
When the transmitter is positioned higher than the
receiver as shown above, LED 3 lights.
When the transmitter is positioned lower than the receiver, LED 5
lights.
mm
inch
When the transmitter is tilted with the laser emitting
surface facing upward as shown above, LED 1 lights.
When the transmitter is inclined with the surface facing downward,
LED 7 lights.
mm
inch
"P-Err" appears, and all the LEDs light.
When the transmitter and receiver are positioned as
shown above, all the LEDs light.
67
CHAPTER 6
OPTIONAL INTERFACE BOARDS
The optional interface boards shown in this chapter can be factory-attached
by the manufacturer at your request.
6-1. BCD Output
Specifications
Pin Assignment
Input
TTL voltage level, negative logic
(74LS19 or equivalent)
Output
TTL voltage level, positive logic
(74LS04 or equivalent)
• BCD positive logic signal that represents measured values
Measured value is output every 20 ms when the MOVE averaging mode
(number of measurements for averaging: 16 to 1,024) is set, or every
"2.5 ms x number of measurements for averaging" when the SIMPLE
averaging mode is set.
• When the measured value is negative, "1011" is output as the 103 digit.
Strobe Signal (STROBE OUT)Negative logic signal with pulse width of approx. 500 µs that represents
output timing of BCD signal
Input Signal
External Trigger Enable Input
(ETE IN)
InputLevelOutput
HIGH
(Input level is HIGH
even when this pin is
open.)
ETE IN
Measured value is output every 20 ms
when the MOVE averaging mode
(number of measurements for averaging: 16 to 1,024) is set, or every "2.5 ms
x number of measurements for averaging" when the SIMPLE averaging mode
is set.
LOW
Measured value is output only when
EXT TRIG IN signal is received.
Timing Diagram
• When ETE IN is pulled high:
• When ETE IN is pulled low:
Outputs of
measured value
500 µ sec
STROBE OUT
EXT TRIG IN
Outputs of
measured value
STROBE OUT
• When ETE IN is pulled low inputting EXT TRIG IN signal allows measured
values to be output at the desired timing.
• Pull ETE IN low, if STROBE OUT signal cannot be received by external
equipment when ETE IN is pulled high.
T
500 µ sec
0.1 µs min.
T
500 µ sec
500 µ sec
T:
(MOVE averaging)
Number of measurements
for averaging
=16 to 1,024: 20 ms
(SIMPLE averaging)
2.5 ms x number of
measurements for avaraging
T:
(MOVE averaging)
Number of measurements
for averaging = 1 to 8 : 2.5 ms(max.)
Number of measurements
for averaging = 16 to 1,024 : 20 ms(max.)
(SIMPLE averaging)
2.5 ms x number of measurements
for averaging (max.)
69
CHAPTER 6 Optional Interface Boards
External Trigger Input
(EXT TRIG IN)
When EXT TRIG IN is pulled low with ETE IN pulled low, the measured value
is output after a max. of 20 ms (MOVE averaging, number of measurements
for averaging: 16 to 1,024), or a max. of "2.5 ms x number of measurements
for averaging" (SIMPLE averaging). The output value is retained until
EXT TRIG IN is pulled low again.
6-2. GP-IB Interface I/O
Specifications
Pin Assignment
The GP-IB I/O interface conforms to IEEE (Institute of Electrical and Electronics Engineers) standard 488-1978.
Logic level
Number of connectable devices
Total cable length
Cable length between 2 devices
*Varies according to the number of interconnected devices.
Transmission code
ASCII
Logic 0 (HIGH state): +2.4 V min.
Logic 1 (LOW state): +0.4 V max.
(Bidirectional transceiver is used.)
15 max. (including controller)
20 m max.*
4 m max.
12 11 10 9 8 7 6 5 4 3 2 1
24 23 22 21 20 19 18 17 16 15 14 13
Pin No.SignalFunction
1DIO1Data bus
2DIO2Data bus
3DIO3Data bus
4DIO4Data bus
5EOIIndicates that the data transmission is
completed.
6DAVIndicates that the data is valid.
7NRFDIndicates that the data reception is impossible.
8NDACIndicates that the data reception is not
completed.
9IFCInitializes the interface.
10SRQAllows devices on the line to request interrupt to
the controller.
11ATNDetermines whether the data is interface
message or device message.
12ShieldShield
13DIO5Data bus
14DIO6Data bus
15DIO7Data bus
16DIO8Data bus
(To be continued)
70
CHAPTER 6 Optional Interface Boards
Pin No.SignalFunction
17RENAllows the controller to remotely enable or
disable devices on the line.
18GND (DAV)Ground for control bus
19GND (NRFD)Ground for control bus
20GND (NDAC)Ground for control bus
21GND (IFC)Ground for control bus
22GND (SRQ)Ground for control bus
23GND (ATN)Ground for control bus
24Logic GNDSignal ground
Interface Functions
The interface bus functions of the LS controller are as follows:
Be sure to turn OFF the power of the LS controller, then set these switches.
Since the GP-IB allows up to 15 devices to be interconnected, an address
must be assigned to each device. To assign the address, use the DIP
switches on the back panel of the LS controller. (See the figure below). The
talk only mode or delimiter can also be set using these switches.
C
R
+
F
L
12
ON
CRL
Set this DIP switch to ON to enable "L ". (To enable
R
both "C " and "L " or only "C ", set this switch to OFF.)
OFF
5678
4
ON
F
16 8
Use these switches to set address
Use this switch to set talk only mode
Use these switches to set delimiter
F
12
4
R
F
71
CHAPTER 6 Optional Interface Boards
Setting Address
Setting Talk Only Mode
Setting Delimiter
Use the 1st to 5th DIP switches from the right (4, 5, 6, 7 and 8 in the figure)
to set the address. Addresses "0" to "30" can be set. To set the address to
"11", set the DIP switches as shown below.
With the LS controller, the address has been factory-set to "0".
1356784
2
ON
To set the LS controller to the talk only mode, set the 3rd DIP switch from
the left (3 in the figure) to ON (down).
With the LS controller, the talk only mode has been factory-set to OFF (up).
* When the LS controller is set to the talk only mode, outputs are sent only
through CH1.
Use the 1st and 2nd DIP switches from the left (1 and 2 in the figure) to
select the delimiter for data transmission.
• To set the delimiter to "LF":
Set the 1st and 2nd DIP switches respectively to OFF and ON (as shown
below).
1356784
2
ON
• To set the delimiter to "CR":
Set the 1st and 2nd DIP switches respectively to ON and OFF (as shown
below).
1356784
2
ON
• To set the delimiter to "CR + LF":
Set both the 1st and 2nd DIP switches to OFF (as shown below).
* With the LS controller, the delimiter has been factory-set to "CR + LF".
1356784
2
ON
EOI (End of Identify) can be output from the controller simultaneously when
the delimiter is transmitted from the LS controller. Select the appropriate
command to select how EOI is output.
Thus, you can select whether EOI is output simultaneously when the
delimiter is transmitted. In addition, you can output EOI alone from the
controller without transmitting the delimiter from the LS controller.
Interface Messages
72
The responses to commands sent from the controller (computer, etc.) are as
follows:
Command
IFCInterFace CLear
"SDC"Selcted Device CLear
Initializes the interface, selected
device or all devices on the line.
Response
"DCL"Device CLear
"GET"Group Execute
Trigger
Reads measured values into the GPIB buffer.
CHAPTER 6 Optional Interface Boards
Listener-talker
Functions
Setting Functions and
Outputting Measured
Values or Other Data
By sending commands from an external device (computer, etc.) to the LS
controller, you can remotely set the various functions, or output measured
values or other data from the LS controller.
* For sending commands, ASCII character codes can be used.
The commands and formats are the same as those for the RS-232C.
See section "RS-232C" in this manual.
The following commands can be used on the GP-IB in addition to those for
the RS-232C.
CommandDescription
ESTransmits EOI (End of Identify) simultaneously when the
delimiter is transmitted.
EOTransmits EOI (End of Identify) only.
ERCancels EOI (End of Identify) transmission.
QSEnables SRQ (Service Request) transmission.
QRDisables SRQ (Service Request) transmission.
K1Outputs measured value of CH1 once when "GET" is
received.
K2Outputs measured value of CH2 once when "GET" is
received.
K3Outputs measured value of CH1 once and then that of CH2
once when "GET" is received.
Talk Only Mode
• When the power is turned ON, "ES", "QR", and "K1" (see the table above)
are set.
• When "SRQ" transmission is enabled, bits DIO7 and DIO1 are set to "1" if
the transmitted command has a syntax error.
Format:
Commands for setting functions CR *
With the RS-232C the delimiter for data output from the LS controller is
"CR". However, the delimiter varies according to the setting when the GP-IB
is used.
* The delimiter for commands sent from an external device (computer, etc.)
to the LS controller can be any of "CR", "LF", "CR + LF" and "EOI".
To set the LS controller to the talk only mode, set the 3rd DIP switch from
the left (3 in the figure) on the back panel to ON.
When the LS controller is set to the talk only mode, it only talks, continuously outputting the preset values. The listener function does not work.
Note that outputs are sent only through CH1 when the LS controller is in the
talk only mode.
1356784
2
ON
73
CHAPTER 6 Optional Interface Boards
Status Byte
Basic Sequence
When Selecting and Setting
Function
When the LS controller receives serial polling signal from the controller, the
LS controller transmits a status byte.
The content of the status byte is as shown below.
(MSB)(LSB)
DI01DI02DI03DI04DI05DI06DI07DI08
000000
RQS bitBit DI01 is set to "1"
when syntax error
occures
When 2 or more commands are transmitted, loop 1 is executed repeatedly.
Clearing device
Controller
(computer, etc.)
LS
Transmitting command
Analyzing command
Polling
When Reading Output Data
Controller
(computer, etc.)
LS
1
Checking command
for error
No.
Setting data to be
transmitted
Yes.
Setting RQS bit to "1"to
indicate that error occurs
Data cannot be read into the LS controller in the talk only mode.
When 2 or more commands are transmitted, loop 1 is executed repeatedly.
When the command to continuously output measured values (such as "X1")
is transmitted, operation 2 is executed repeatedly.
Clearing device
Transmitting command
Analyzing command
Polling
Reading data
74
1
Checking command
for error
No.
Setting data to be
2
transmitted
Yes.
Setting RQS bit to "1"to
indicate that error occurs
CHAPTER 6 Optional Interface Boards
Example of Program
When connecting LS controller with external computer
Program
The following settings are specified through the computer: DIA (segment
mode), CH1=X (output channel mode), NORMAL (measurement mode),
MOVE (averaging mode), 512 (the number of measurements for averaging).
Then, the measured value is displayed once on the computer.
100Clears interface.
110Sets delimiter to "CR +LF".
120Sets talker address for PC to "1".
Sets listener address for LS controller to "0".
130Transmits commands to LS controller.
"D": Sets "DIA" as segment mode.
140"C0": Sets "CH1=X" as output channel mode.
150"MN": Sets NORMAL as measurement mode.
160"A9": Sets 512 as the number of measurements for
averaging.
170"BM": Sets MOVE as averaging mode.
180"X0": Allows LS controller to output measured value once.
190Receives measured value once from LS controller.
200Displays measured value.
210Terminates program.
The above program provides the following status:
Talker address for PC: 1
Listener address for LS controller: 0
Delimiter sent from LS controller: CR + LF
Talk only mode of LS controller: OFF
Take the following precautions when using the LS controller.
Be sure to turn OFF the power of the LS controller to set the address or
delimiter using the DIP switches on the back panel.
75
CHAPTER 7
HINTSON CORRECT USE
General
GP-IB Interface I/O
• When noise interference may occur in the surroundings, connect the LS
controller to an external device (computer, etc.) using an electromagneticinterference protection cable(s). Be sure to turn OFF the power of the LS
controller before connecting it to an external device.
• Supply power to the computer or other device(s) before supplying power
to the LS controller. If power is supplied to the LS controller first, it may
malfunction due to noise occurring when another device starts up.
• Up to 15 devices can be interconnected using the GP-IB bus.
• The total cable length for connecting the devices should be 20 m or
shorter, provided that it does not exceed "2 m x number of devices". Also,
the length of each cable should not exceed 4 m. When 3 devices are
connected, for example, the maximum total cable length is 6 m, and when
4 devices are connected, it is 8 m. (See the figure below.)
To make long-distance transmission of data, use an appropriate signal
transmission system instead of the GP-IB bus.
Connecting 3 devices
LS
LS
Targets
Measuring Conditions
2m
Connecting 4 devices
LSLS
2m
When using the GP-IB, set the address and delimiter applicable to the
program to be executed or device(s) to be connected.
When the shape or glossiness of targets varies, measurement error may
occur. In such a case, place a target of known dimension in the measuring
area and calibrate the LS controller using CAL. (Two targets of different size
are required.)
When the target is vibrating, measured values may vary significantly. In such
a case, set the number of measurements for averaging to a larger number to
ensure highly accurate measurement.
2m
4m
LS
4m
76
CHAPTER 7 Hints on Correct Use
Compatibility
Operating Environment
• With the LS series, the controller and scanning head are calibrated as a
pair. To satisfy the specifications, therefore, be sure to combine a controller and scanning head having the same serial number.
• The transmitter and receiver are also calibrated as a pair.
• When using 2 scanning heads, connect them respectively to the specified
connector port of the controller.
• Be sure to operate the LS controller in a place where the temperature is
within a range between 0 and 45°C. (0 to 40°C for LS-3036/3060)
• Fluctuation in ambient temperature may cause measurement error.
Always maintain the ambient temperature at a constant level. If the
ambient temperature changes by 10°C, it takes approx. 60 minutes for the
temperature distribution in the LS controller to become uniform.
• Be sure to keep the laser emitting surface (transmitter) and the laser
receiving surface (receiver) free from any substance refracting light (such
as water or oil).
• Blow off dirt or dust on the protective glass using clean air. If dirt or dust
still remains, wipe it off using a soft cloth moistened with alcohol.
Countermeasures against
Noise
• The protective glass can be replaced. See the figure below to replace the
glass.
• Do not mount the scanning head to the place where a certain level of
electrical noise is applied or transferred. Isolate the scanning head if
electrical noise may be occured. Otherwise, laser diode may deteriorate
or become damaged.
• Keep the wiring or connection cable away from high-voltage lines or
power lines to prevent the LS controller from malfunctioning because of
noise.
• Be sure to earth-ground the LS controller through the earth ground terminal (No. 19). Insulating the scanning head is also effective in reducing
noise.
Fuse
Cooling Fan
When the fuse is blown, replace it with the following one:
Midget fuse: F-7142 2A or equivalent
Do not block the cooling fan on the back panel. Blocking it will overheat the
LS controller, causing malfunction.
77
CHAPTER 8
ERROR MESSAGES
H-Err
• If no scanning head is
connected to the controller or the connected scanning head malfunctions,
an error message appears on the main display (CH1).
When 1 scanning head is used:
If no scanning head is connected or the connected scanning head mal
functions, "H-Err" appears.
When 2 scanning heads are used:
When "HEAD2 ON" is selected using PRM:
If no scanning head is connected to the HEAD1 connector port or the
scanning head connected to that port malfunctions, "H-Err" appears.
If no scanning head is connected to the HEAD2 connector port or the
scanning head connected to that port malfunctions, "H2-Err" appears.
When "HEAD2 OFF" is selected using PRM:
Since the LS controller judges that no scanning head is to be connected
to the HEAD2 connector port, "H-Err" appears.
• When "H-Err" or "H2-Err" appears, turn OFF the power. Then, make sure
that the problem with the scanning head is removed or that the scanning
head is connected properly, then turn ON the power.
• Turn OFF the power to cancel error message "H-Err" or "H2-Err".
P-Err
CALIB ERROR
When the transmitter and receiver are not aligned in terms of the optical
axis, "P-Err" appears on the main display (CH1). Align them correctly.
Note: Ignore this message when using model LS-3033 SO.
When the LS controller is not calibrated correctly, "CALIB ERROR" appears
on the main display (CH1). Carefully read the description for the calibration
procedures, and calibrate the LS controller again using a proper gauge.
78
CHAPTER 8 Error Messages
79
CHAPTER 9
TROUBLESHOOTING GUIDE
If any conditions occur that could be considered as equipment failure or
malfunction, follow the inspection procedures given below.
Check
item
Measurement
value
display
Problem
• Measured value is not
displayed.
• Nothing appears on the
display.
"------" is displayed.• Are targets in position?
Details of inspectionCorrective action
• Is the power switch ON?
• Is the AC power cable connected?
• Is the entire scanning area
completely covered?
• Is the appropriate measuring
segment selected?
• Is the scanning head properly
connected?
•Is there any dust or dirt on
the laser emitting/receiving
window?
• Is the laser emission indicator
LED lamp lit?
• Wasn’t "------" displayed
immediately after turning ON
the power switch?
• Wasn’t "------" displayed
immediately after switching
the measurement modes?
• Turn ON the switch.
• Connect the AC power cable.
• Correctly place the targets
within the effective measurement area.
• Correctly select the appropriate measuring segment.
•Turn the power switch OFF,
then after connecting the
scanning head, turn the
power switch ON
• Remove dust or dirt from the
laser emitting/receiving
window.
• Short LASER REMOTE
terminals.
• Wait approximately 10
seconds after turning ON the
power switch before operation.
• Wait approximately 3
seconds after changing
measurement modes before
resuming operation.
Reference
page
p. 5
p. 84
p. 28
p. 15
80
• Wasn’t "------" displayed
immediately after shorting
the laser emission remote
control terminal?
• Isn’t the number of measurements for averaging set to a
large value?
• Is there condensation forming
on the equipment?
• Wait approximately 10
seconds after shorting this
terminal.
• In the Average mode, maximum of approximately 2.6
seconds are required until
the measurement value is
displayed after the target is
positioned in the measurement area.
• Change the location.
p. 21
CHAPTER 9 Troubleshooting Guide
Check
item
Measurement
Value
Display
Problem
The measurement display
value contains unusual
dispersion.
Abnormal measurement values are being
displayed.
Details of inspectionCorrective action
• Are the targets correctly
placed in the measurement
area?
• Is a standard edge placed
within the measurement
area?
•Is there any dust or dirt on
the laser aperture?
•Is there any dust or dirt on
the targets?
• Is there a large fluctuation of
ambient temperature?
• Do the serial numbers match
between the controller and
scanning head?
• Has calibration been
correctly performed?
• Is the laser beam correctly
aligned?
• Is the target tilted or has it
moved from the original
position?
• Are the targets correctly
placed within the measurement area?
• Is a reference edge placed
within the measurement
area?
• Is there any dust or dirt on
the target?
• Isn’t the controller set to the
analog output range setting
mode?
• Correctly place the target
within the measurement area.
• Place a standard edge within
the measurement area.
• Remove any dust or dirt.
• Remove any dust or dirt.
• Maintain the constant operating temperature.
• Connect a scanning head
with the matching serial
number.
• Correctly perform calibration.
• Check whether or not the
laser beam is correctly
aligned in the laser beam
alignment check mode. If the
beam is not correctly aligned,
refer to the error message as
indicated in the display and
adjust the mounting conditions accordingly.
• Correctly place the targets.
• Correctly place the targets
within the measurement area.
• Place a reference edge within
the measurement area.
• Remove any dust or dirt from
the target.
• Set the appropriate measurement mode.
Reference
page
p. 14
p. 41
p. 65
p. 84
Analog
output
"H-Err" is displayed.
No output.
(0 V output)
Voltage above +6 V or
below -6 V is output.
Is the scanning head
properly connected?
Is the measurement value
displayed?
Is the analog output range
correctly set?
Turn the power OFF, then after
correctly connecting the
scanning head, turn the power
ON.
To make sure that the target is
correctly measured, correctly
position the controller and
targets.
Check that the analog output
range is correctly set.
p. 78
p. 22
81
CHAPTER 9 Troubleshooting Guide
Check
item
Computer
output
Problem
Computer data is not
output.
Details of inspectionCorrective action
• Is the baud rate correctly
set?
• Is the cable connector
securely connected to the
port?
• Has the computer been
turned ON before the
controller?
• Is the correct program used?
• Set the controller and computer to the same baud rate.
Note, however, that some
computers may not recognize
correct measurements when
set to a high baud rate.
• Check that the cable connector is securely connected to
the port.
• Be sure to turn ON the computer first, then controller.
• Send a message such as
"X0" from the computer to
access the controller.
Reference
page
p. 27
p. 49
82
APPENDIX A
OPERATING PRINCIPLE
Laser beam emitted by semiconductor is reflected first by the polygonal
mirror, then by the reflector mirror. Laser beam is scanned according to the
rotational angle of the polygonal mirror during this process. Then, scanned
beams pass through the collimator lens so that they run parallel to each
other. These beams are directed at the target and converged onto a
photodiode by another lens. The beams thus collected are converted into an
electric signal corresponding to the light-dark pattern produced.
The dimension of the target is determined by the width of "dark" region
measured in terms of time.
Data
display
Key input
Waveform
shaping
Relay output
Analog output
Photodiode
RAM
ROM
Receiving
lens
Edge
detection
LatchCounter
CPU
Control I/OBCD output*
Collimeter lens
(Fθ lens)
Target
RS232C
Clock
pulse
Photodiode
(for synchronization)
Polygonal mirror
GP-IB*
Reflector
(mirror)
Semiconductor
laser
*Option*Option
83
APPENDIX B
CHARACTERISTICS
Measuring Area vs.
Accuracy
Data
LS-3032
10
TransmitterReceiver
4
20
30
Center of detecting area
±
2 µm±
5 µm
LS-3033
Transmitter
Center of detecting area
20
10
30
20
60
± 2 µm
± 5 µm
When a rod 10 mm in diameter is measured.When a rod 10 mm in diameter is measured with a transmitter-to-
receiver distance set to 120 mm.
LS-3034/3033 SO
5
TransmitterReceiver
2
20
30
Center of detecting area
2µm±
m
5µ
±
LS-3036
10
TransmitterReceiver
4
20
30
Center of detecting area
2 µm±
5 µm
±
When a rod 10 mm in diameter is measured.
LS-3060
100
Transmitter
60
40
60
20
30
When a rod 20 mm in diameter is measured.
84
When a rod 10 mm in diameter is measured.
± 3µm
± 6µm
APPENDIX B Characteristics
Temperature
Characteristics
(Typical)
When a rod 10 mm in diameter is measured at the center of the measuring
area [reference temperature: 20°C (68°F)].
Measured valueSubdisplay: 16 character x 2 line LCD
(Backlight color: yellow-green)
Minimum display unit0.0001 mm (0.00001 inches)
Display range-999.9999 to 9999.9999
Display
Measuring rangemm/inch selectable
Target position indicatorGreen LED (7 levels)
Comparator output indicatorGreen LED x 3 (HI/GO/LO)
Laser emission indicatorGreen LED x 2 (HEAD 1/HEAD 2)
Hold synchronous
Control
input
I/O
Control
Terminal block
output
Control
input
I/O
Control
output
36-pin connector
RS-232C
GP-IB
port
Interface
BCD
Power supply85 to 264 VAC, 50/60 Hz
Power consumption65 VA max.
Ambient temperature0 to +40°C (32 to 104°F), No freezing
Relative humidity35 to 85%, No condensation
WeightApprox. 4.6 kg
1. Available as an option.
LASER REMOTE
PROGRAM SELECT
(Program Nos. 1 to 5)
Analog voltage±6 V, Output impedance: 0 Ω
Control outputRelay output: SPST-NO contact x 3, 250 VAC, 2 A (resistive load)
Hold synchronous
AUTO ZERO SET/RESET
PROGRAM SELECT
(Program Nos. 1 to 5)
7-level
3-level
1.
1.
Non-voltage input (contact/solid state)
TTL voltage level, negative logic
NPN open-collector: 40 mA max. (30 V max.)
Measurement data output and control input
(baud rate: 75 to 19200 bps selectable)
Measurement data output and control input
Measurement data output (8 digits)
86
Scanning Head
APPENDIX C Specifications
TypeFine measuringStandard
Wide
range
Scanning head typeSingle-bodySeparateSingle-bodySeparateSingle-bodySeparate