*TB 9-6625-108-50

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

CALIBRATION PROCEDURE FOR

SLOTTED LINE, HEWLETT-PACKARD, MODELS 805A, 805C, 805D, S810A, G810B, J810B, H810B, AND X810B; AND

IM-92/U AND IM-92A/U

(NASSAU MODEL 230)

Headquarters, Department of the Army, Washington, DC

10 December 1973

REPORTING OF ERRORS

You can help improve this publication by calling attention to errors and by recommending improvements and stating your reasons for the recommendations. Your letter or DA Form 2028, Recommended Changes to Publications, should be mailed directly to Commander, U. S. Army Aviation and Missile Command, ATTN: AMSAM-TMD-EP, Redstone Arsenal, AL 35898-5000. FAX to DSN 7882313 (commercial 256-842-2313). A reply will be mailed directly to you.

 

 

Paragraph

Page

SECTION

I. IDENTIFICATION AND DESCRIPTION

 

 

 

Test instrument identification .............................

1

2

 

Calibration data card (DA Form 2416).................

2

2

 

Calibration description.........................................

3

2

II.EQUIPMENT REQUIREMENTS

Equipment required .............................................

4

2

Accessories required .............................................

5

3

III.CALIBRATION PROCESS FOR SLOTTED LINES, HEWLETT-PACKARAD MODELS 805A, 805C, AND 805D, AND IM-92/U AND IM-92A/U (NASSAU MODEL 230)

Preliminary instructions ......................................

6

4

Residual VSWR ....................................................

7

4

Final procedure.....................................................

8

7

IV. CALIBRATION PROCESS FOR SLOTTED

 

 

LINES, HEWLETT-PACKARD, MODELS

 

 

S810A, G810B, J810B, H810B, AND X810B

 

 

Preliminary instructions ......................................

9

7

Equipment setup ..................................................

10

8

Slope and irregularities ........................................

11

10

Final procedure.....................................................

12

11

TB 9-6625-108-50

SECTION I

IDENTIFICATION AND DESCRIPTION

1.Test Instrument Identification. This bulletin provides instructions for the calibration of Slotted Line, Hewlett-Packard Models 805A, 805C, 805D, S810A, G810B, J810B, H810B, and X810B; and IM-92/U and IM-92A/U (Nassau Model 230). The manufacture's instruction manual was used as the prime data source in compiling these instructions. These slotted lines will be referred to as the "TI" (test instrument) throughout this bulletin.

a.Model Variations. Variations among, models are described in table 1 and in text.

b.Time and Technique. The time required for this calibration is approximately 2 hours per unit, using the microwave technique.

2.Calibration Data Card (DA Form 2416). Maintenance forms, records, and reports which are to be used by calibration personnel at all calibration levels are listed in and prescribed by TM 38-750.

3.Calibration Description. TI parameters and performance specifications which pertain to this calibration are listed in table 1.

Table 1.

Calibration Description

Test instrument

 

 

parameters.

 

Performance specifications

Frequency range

 

500 to 4000 MHz 1 2 3

 

 

2.60 to 3.95 GHz (H-P S810A).

 

 

3.95 to 5.85 GHz (H-P G810B).

 

 

5.30 to 8.20 GHz (H-P J810B).

 

 

7.05 to 10.0 GHz (H-P H810B).

 

 

8.0 to 12.4 GHz (H-P X810B).

Characteristic impedence.5

 

50 ohms 1 3

 

 

46.3 ohms 2

Residual SWR

 

Less than 1.04:11 3

 

 

Less than 1.02:12

Slope and irregularities

 

Less than 1.01:14

1H-P Models 805A and 805C.

2H-P Model 805D.

3IM-92/U and IM-92A/U.

4H-P Models S810A, G810B, J810B, and X810B.

5This specification is for information only and is not verified in this bulletin.

SECTION II

EQUIPMENT REQUIREMENTS

4. Equipment Required. Tables 2 and 4 identify the specific equipment used in this calibration procedure. This equipment is issued with secondary transfer calibration standards set 4931-621-7877 and is to be used in performing this procedure. Alternate

2

TB 9-6625-108-50

items may be used by the calibrating activity when the equipment listed in tables 2 and 4 is not available. The items selected must be verified to perform satisfactorily prior to use and must bear evidence of current calibration. The equipment must meet or exceed the minimum use specifications listed in tables 2 and 4. The accuracies listed in tables 2 and 4 provide a four-to-one accuracy ratio between the standard and TI. Where the four-to-one ratio cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.

5. Accessories Required. The accessories listed in tables 3 and 5 are issued with secondary transfer calibration standards set 4931-621-7877 and are to be used in this calibration procedure. When necessary, these items may be substituted by equivalent items unless specifically prohibited.

Table 2. Minimum Specifications of Equipment Required

 

 

 

 

 

 

 

Manufacturer, model, and

Item

 

Common name

 

Minimum use specifications.

 

part number.

A1

MODULATOR

Range:

2 to 4 GHz

 

Hewlett-Packard, Model

 

 

 

Type: Square wave

 

C108403A (7912321).

A2

PIN MODULATOR

Range:

Square wave modulation:

 

Hewlett-Packard, Model 8732B

 

 

 

 

from 2 to 4 GHz.

 

(7912581).

 

 

 

 

 

 

 

 

A3

SIGNAL SOURCE

Range:

2 to 4 GHz

 

(7923114)

 

 

 

Power output: 10 MW min

 

 

 

 

 

 

 

 

 

 

A4

SLIDING TERMINATION

Range:

2 to 4 GHz

 

Narda, Model 22487 (7923224).

A5

SWR METER

Range:

1.01 to 1.1

 

Hewlett-Packard, Model Y10-

 

 

 

 

 

 

 

415E (7910160-3).

 

 

 

Table 3.

Accessories Required

 

Item

 

Common Name

 

 

Description And Part Number

 

 

 

 

 

 

B1

 

ADAPTER1

 

N plug to N plug (MS-15507-57B)

 

B2

 

ADAPTER

 

N jack to GR900 (7911051) or N plug to GR900

 

 

 

 

(7911052).2

 

 

B3

 

ADAPTER2

 

N jack to 7/8 in., jack (8899714)

 

B4

 

ADAPTER2

 

N plug to 7/8 in., plug (10528698)

 

B5

 

ADAPTER2

 

90° bend with N plug to N jack (MS90156-27D)

B6

 

VARIABLE

 

0 to 15 db, 1 to 4 GHz (7923130)

 

 

 

ATTENUATOR

 

 

 

 

 

B7

 

CABLE

 

18 in., RG-9 ( ) /U with N plug terminations (10519072).

B8

 

CABLE1

 

24 in., RG-58 ( ) /U with BNC plug terminations (10519141).

B9

 

ISOLATOR

 

1 to 4 GHz (7913127-2)

 

1Two required.

2Used for H-P Model 805D.

3

TB 9-6625-108-50

SECTION III

CALIBRATION PROCESS FOR SLOTTED LINES,

HEWLETT-PACKARD MODELS 805A, 805C, AND 805D;

AND IM-92/U AND IM-92A/U (NASSAU MODEL 230)

6.Preliminary Instructions

a. The instructions outlined in this paragraph are preparatory to the calibration process. Personnel should become familiar with the entire section before beginning the calibration.

b. Items of equipment used in this procedure are referenced within the text by common name and item identification number as listed in tables 2 and 3. For the identification of equipment referenced by item numbers prefixed with A, see table 2, and for prefix B, see table 3.

WARNING

HIGH VOLTAGE is used during the performance of this calibration. DEATH ON CONTACT may result if personnel fail to observe safety precautions.

NOTE

Unless otherwise specified, verify the results of each test and take corrective action whenever the test requirement is not met before continuing with the calibration.

c. Connect the equipment as shown in figure 1.

d. Energize the equipment and allow sufficient time for warmup.

e. Adjust signal source (A3) for a 2-GHz leveled output.

7. Residual VSWR

NOTE

The VSWR should be measured on the 30-db range of SWR meter (A5). To set up a reference for this measurement, adjust modulator (A1) frequency, variable attenuator (B6), and the TI probe tuning and depth controls until the SWR meter indicates full scale with minimum TI probe depth.

a.Performance Check

(1)Adjust sliding termination (A4) and the TI until the measured SWR is maximum. Record SWR meter (A5) indication as SWR max.

4