HP (Hewlett-Packard) 81531A User Manual

5 (1)

User's Guide

HP 81531A Sensor Module

SERIAL NUMBERS

This manual applies to all instruments.

ABCDE

HP Part No. 81531-90011

Printed in Germany

Second Edition

E0796

Control Serial Number: Edition 2 applies directly to all instruments. Edition 1 : 1st May 1990 : 81531-90011 : E0590

1st September 1990 : 81531-90011 : E0990

Edition 2 : 1st September 1991 : 81531-90011 : E0991 1st September 1992 : 81531-90011 : E0992

1st December 1994 : 81531-90011 : E1294 1st July 1996: 81531-90011 : E0796

HP (Hewlett-Packard) 81531A User Manual

Safety Considerations

Before operation, you should review the instrument and manual, including the red safety page, for safety markings and instructions. You must follow these to ensure safe operation and to maintain the instrument in safe condition.

Initial Inspection

Inspect the shipping container for damage. If there is damage to the container or cushioning, you should keep it until you have checked the contents of the shipment for completeness and veri ed the module both mechanically and electrically.

The Performance Tests give procedures for checking the operation of the module. If the contents are incomplete, mechanical damage or defect is apparent, or if a moduke does not pass the operator's checks, notify the nearest Hewlett-Packard o ce.

Warning To avoid hazardous electrical shock, do not perform electrical tests when there are signs of shipping damage to any portion of the outer enclosure (covers, panels, etc.).

Power Requirements

The HP 81531A operates when installed into the HP 8153A Optical Multimeter mainframe.

Operating Environment

The HP 8153A safety information summarizes the HP 81531A operating environment ranges. In order for the HP 81531A to meet speci cations, the operating environment must be within the limits speci ed in this section.

Input/Output Signals

iii

Caution A maximum of 15V can be applied as an external voltage to any BNC connectors.

Storage and Shipment

The module can be stored or shipped at temperatures between 040 C and +70 C. The module should be protected from temperature extremes that may cause condensation within it.

iv

Contents

C. HP 81531A Speci cations

 

Supplementary Performance Characteristics . . . . . . . . . . .

C-3

 

Analog output: . . . . . . . . . . . . . . . . . . . . . . .

C-3

D.

Performance Tests

 

 

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . .

D-1

 

Equipment Required . . . . . . . . . . . . . . . . . . . . .

D-1

 

Test Record . . . . . . . . . . . . . . . . . . . . . . . . . .

D-2

 

Test Failure . . . . . . . . . . . . . . . . . . . . . . . . . .

D-2

 

Instrument Speci cation . . . . . . . . . . . . . . . . . . . .

D-2

 

Performance Test . . . . . . . . . . . . . . . . . . . . . . .

D-3

 

Accuracy and Linearity Test . . . . . . . . . . . . . . . . .

D-3

 

Test Setup . . . . . . . . . . . . . . . . . . . . . . . . .

D-3

 

I. Accuracy Test . . . . . . . . . . . . . . . . . . . . . .

D-4

 

II. Linearity Test . . . . . . . . . . . . . . . . . . . . . .

D-5

 

III. Noise Test . . . . . . . . . . . . . . . . . . . . . . . .

D-8

E.

Cleaning Procedures

 

 

The Cleaning Kit . . . . . . . . . . . . . . . . . . . . . . . .

E-1

 

Other Cleaning Tools . . . . . . . . . . . . . . . . . . . . .

E-3

 

Preserving Connectors . . . . . . . . . . . . . . . . . . . . .

E-4

 

Cleaning Instrument Housings . . . . . . . . . . . . . . . . .

E-4

 

Correct Cleaning Procedure . . . . . . . . . . . . . . . . . .

E-5

 

Cleaning Cable Connectors . . . . . . . . . . . . . . . . . .

E-5

 

Cleaning Connector Adapters . . . . . . . . . . . . . . . . .

E-6

 

Cleaning Connector Interfaces . . . . . . . . . . . . . . . .

E-6

 

How to clean instruments with a recessed lens interface . . . .

E-7

 

Light dirt . . . . . . . . . . . . . . . . . . . . . . . . .

E-7

 

Heavy dirt . . . . . . . . . . . . . . . . . . . . . . . .

E-7

Index

Contents-1

Figures

D-1.

Accuracy and Linearity Test Setup . . . . . . . . . . . . . .

D-4

D-2.

Addition of results (example) . . . . . . . . . . . . . . . . .

D-7

Contents-2

C

C

HP 81531A Speci cations

Speci cations describe the instrument's warranted performance. Supplementary performance characteristics describe the instrument's non-warranted typical performance.

Because of the modular nature of the instrument, these performance speci cations apply only to this module. You should insert these pages into the appropriate section of the manual.

HP 81531A Speci cations C-1

C

HP 81531A Speci cations

Sensor Element

In GaAs

Wavelength range

800-1700nm

Power range

+3 to -90dBm

Display resolution

 

dBm, dB

0.001

 

0.0001 on printout

 

Watt

0.01 to 10

 

(depending on power range)

 

 

Applicable ber type

9/125 m - 100/140 m, NA 0.3

Uncertainty (Accuracy)

62.5% (1000-1650nm)[1]

Total Uncertainty

65%61.5pW (1000-1650nm)[2]

Linearity

(0 to -70dBm)

18 C to 28 C const. temp

60.015dB61pW

0 C to 55 C const. temp

60.05dB61.5pW

Noise

<1.5pW (1200-1600nm)

peak-peak, avg. time 1sec

 

Dimensions

75mm H, 32mm W, 335mm D (2.8"21.3"213.2")

Weight

net 0.6kg (1.3lbs), shipping 1kg (2.2lbs)

Recalibration period

2 years

Warmup time

20 min.

Information on the traceability of power meters is available on request

[1]at the following reference conditions:

Power level 10 W (-20dBm), Continuous Wave (CW).

Fiber 50 m graded index, NA=0.2, fully excited.

Ambient temperature 23 C65K.

Connector Diamond HMS-10/HP.

At day of calibration (add 0.3% for aging over one year, add 0.6% over two years).

Spectral width of source <10nm.

[2]at the following operating conditions:

Fiber 50 m, NA 0.2.

For NA >0.2 add 1%.

C-2 HP 81531A Speci cations

C

Ambient temperature 0 to 55 C, non-condensing.

Within 1 year after calibration, add 0.3% for second year.

For Biconic Connector add 61%.

Supplementary P erformance Characteristics

Add 1% to total uncertainty for the full wavelength range.

Outside the speci ed wavelength range, the noise increases by up to 5 times the value shown above.

Analog output:

Bandwidth: DC, 300 to 400Hz, depending on range and sensor module. Output voltage: 0 to 2V into open.

Output impedance: 600 typical. Max. input voltage: 610V.

HP 81531A Speci cations C-3