Fresenius 2008T User's troubleshooting manual

4 (4)
Fresenius 2008T User's troubleshooting manual

2008T BlueStarTM Hemodialysis Machine

Operator’s Troubleshooting Guide

TABLE OF CONTENTS

Installation

1

Auto Start

2

Auto Prime

5

Miscellaneous

8

Indications for Use

14

NOTE: This troubleshooting guide is intended to be a supplement to the information in the 2008T Operator’s Manual, P/N 490122, including Warnings and Cautions. Users must read and understand the information in the 2008T Operator’s Manual before using the 2008T BlueStarTM hemodialysis machine.

INSTALLATION

The 2008T BlueStar Standard Upgrade Kit (P/N 191137) installs a different functional software version than expected (e.g., installs version 2.64 instead of version 2.72).

1.Read and follow the installation instructions included with the upgrade kit.

2.The 2008T BlueStar software installation requires a two-step installation process as noted in the instructions. The updater will install a temporary software version which will not match the Flash Drive label during the first installation. The latest functional software version will be installed during the second installation.

2008T BlueStarTM Machine Operator’s Troubleshooting Guide

1

AUTO START

Auto Start not available.

1.2008T BlueStar Premium Upgrade Kit (P/N 191138) must be installed and the Premium features enabled.

2.Once installed, Auto Start is accessible in Service Mode / Maintenance / Scheduler / Auto Start.

Auto Start automatic self-tests do not complete.

1.A conductivity time-out occurs.

2.Select Both Tests to run self-tests manually.

NOTE: The Auto Start Dialysis button will be green and the status box message “Connect Concentrates” will be displayed. Once the acid connector is removed from its port, the machine will wait up to seven minutes for dialysate conductivity and temperature to stabilize. If they do not stabilize, automatic testing will be cancelled, and the self-tests must be run manually.

To reduce the chances of this situation occurring, it is suggested that both the acid and bicarbonate concentrate supplies be connected at the same time to achieve conductivity. Follow facility policies and procedures.